pHORP8500A Operation Manual - Simple Version - 20130410
pHORP8500A Operation Manual - Simple Version - 20130410
Operation manual
Notes before Operation
1. Notice before operation:
(1) Carefully read the relevant parts of this manual before installation and operation to
(2) Improper installation and unsuitable flow speed will cause big measurement error, so
(3) This instrument is precise electrochemical measurement, and its installation and
(4) If for some special conditions, please consult our Technical Department for further
details.
2. Maintenance items:
(1) The indicators’ quality guarantee is one year from the date of purchasing. During this
(2) pH and ORP sensors are consumer goods which without quality guarantee. Because,
(3) Manufacturer offers the maintenance service for whole life of the sold meters.
(4) If the damage of the meter is caused by the following reasons, it is out of the
maintenance service:
E、 The cable damage and rupture caused by improper installation and usage.
personally.
G、 The measurement inaccurate caused by privately cutting the sensor cable.
H、 The inner broken wire caused by indiscreetly disassembling.
I、 The consumptive materials are not within the scope the warranty.
I CONCEPTION..................................................................................................................................1
I FEATURES .................................................................................................................................1
II APPLICATIONS .......................................................................................................................2
III TECHNICAL INDEX. .................................................................................................................3
II INSTALLATION ..........................................................................................................................4
APPENDIX I ......................................................................................................................................19
APPENDIXⅡ.....................................................................................................................................19
I Conception
pH/ORP-8500A is our newly type pH/ORP meter which with color screen display.
The meter with 32-bit processor, ARM core operation system.
i Features
3.5 inch 320×240 TFT color screen.
Virtual pointer or Digital style
interface display. Multi-parameter
display.
OFFLINE/INLINE calibration
selection, 6 point buffer solution
calibration.
Automatic/manual temperature
compensation function. The meter
temperature range could up to 100℃。
Configuration
pH/ORP/temperature,
(4~20)mA
current support instrument/transmitter
mode.
1
DC24V Power supply polarity
internal automatic identification
EMC design with better ability on
anti-interference.
ii Applications
High level water quality management. Automatic running.
High pure water /ultra pure water management control. The meter is used for
Electronics, electric power, pharmaceutical, fine chemical, clinical medical science,
life science research
The meter is used for Industrial process solution salinity analysis of Water
treatment, desalination, concentrate management, circulating cooling water,
industrial coating, large-scale water treatment equipment
The meter is used for inline monitoring of Metallurgy, petrochemical,
industrial cleaning.
2
iii Technical index.
Medium temperature
(0.0~50.0)℃
compensation
Temperature
NTC10k temperature components
components
Resistance
50mA(max)AC/DC 30V
load
Data
RS485,modbus protocol
communication
3
Consumption <5.5W
Working Temperature:(0~50)℃
environment Relative Humidity :≤85%RH( non- condensing )
Temperature :(-20~60)℃
Storage
Relative Humidity:≤85%RH(non- condensing)
Installation Panel.
II Installation
i Meter installation
4
3. When disassemble the meter, please take off the clamp slowly, at the same
time, hold the meter carefully.
5
TEMP-I Temperature wire
TEMP-G Connect with the ground.
NC Empty terminal
485A/485B/485GND RS485 port
1st channel instrument mode, instrument internal
I1+/I1-
power supply
1st channel transmitter mode, conditioning
T1+/T1-
module external power supply.
2nd channel instrument mode, instrument internal
I2+/I2-
power supply
2nd transmitter mode, conditioning module
T2+/T2-
external power supply.
SWITCH1/ 1st channel photoelectronic switch control
SWITCH1' contact/ Automatic polarity identification
SWITCH2/ 2nd channel photoelectronic switch control
SWITCH2' contact / Automatic polarity identification
SWITCH3/ 3rd channel photoelectronic switch control
SWITCH3' contact / Automatic polarity identification
24VA Power input interface , connect with DC 24V,
24VB non-polarity.
EARTH Grounding
b. Installation method
6
P34A adjustment flow device Immersion installation
A. Regular questions:
1. User should cutoff the flow before replacing sensor.
2. Fast flow and air bulbs will cause instable measurement.
3. Higher pipeline pressure will damage the sensor
4. There will be big error when testing pure water or ultra pure water.
5. The sensor will fail very soon if for testing pure water.
6. Conductivity will rise if electrode electrolyte diffuses in water through
liquid junction.
7. Negative pressure will damage the sensor.
8. Water particles and fibrous material will damage the sensor.
B. Measurement equipment
1.The patent product(patent No:CN200820077131.X)P34,installed
between pipeline and sensor.
2.The sensor can be replaced without cutting off water flow.
3.The reading will be accurate and stable when testing pure water. The
working life is longer.
4.There is no electrode electrolyte diffuses in water
7
[NOTE 2]:Installation notes
• llation angle Less than 30 degree Inverse installation is incorrect Horizontal installation is incorrect
•
III Adjustment
Reading type and value
Temperature
Date
Trans.current
Keys
8
Key board introduction
9
Normal
user
Digital style
Engineering user
10
Measurement Engineer user
Measure Control Current
interface Operator RS485 Password Time
Display Language Style
Manu. Regulate
Engineer
Transmitting parameter.
Double channels transmitting parameter setting which could
configure the parameter output.
Communication parameter.
RS485 baud rate and address setting. Please find the
protocol on Appendix.
Password setting.
11
System will go back to measurement interface automatically in case of no
operation within three minutes in parameters check or parameters setting
menu.
12
Measure and calibration
Calibrations
pH/ORP sensors are electrochemical and their sensitivity decreases with influence of
time and medium. In order to get an accurate measurement, it is suggested to often
calibrate sensor’s slope. The calibration period relays on the influence from the
measured medium.
1 Six points buffer solution pH(10.00/9.18/7.00/6.86/4.01/4.00) can be
recognizable at calibration.
2.Two points slope calibration can be selected when used for acid or alkaline media
for a long time.
3.Long time stored pH sensor must be staying in KCl solution for 12 hrs before
calibration.
Rinse in diluted hydrochloric acid for two minutes is recommended at calibration for
dry and sensitive glass bulb.
2.The indicator is furnished with off-line calibration with mV input to pH sensor.For
details pls refer to the instructions to Off-line Calibration.
a. Buffer Solution Calibration
1.According to the range ,select the buffer stuff .According to the instruction,
make up buffer solution.
Place the cleaned sensor into the present buffer solution and press
button .Interface will display” sensor outputs mV” ,pls press button
to save.
13
System will go into the next sub menu after calibration. In the mean
time ,place the cleaned sensor in the buffer solution indicated by the indicator
and complete every calibration in turn.
“Pls replace the sensor” will be displayed in the screen when there is
something wrong with calibration. Pls check whether the makeup of the buffer
solution is correct or not and recalibrate.
The next calibration will not be done until the present calibration is made
successfully.
14
enter the interface of Off-line calibration.
Input the corresponding mV value to pH10.00 ,press to save and enter the next
calibration menu.
“Pls replace the sensor” will be displayed in the screen when there is something
wrong with calibration. . Input corresponding mV value once again . The next
calibration will not be done until the present calibration is made successfully.
15
c. Maintenance of sensors
To avoid long time store dry, the sensor should be kept in the protection cap
with KCl solution in 3.0mol/L.
Timely clean the sensor and calibrate the same on the indicator.
If the above methods are invalid to reset slope, it means that the sensor should
be replaced.
If the surface of platinum is polluted and formed oxidation film, the platinum
could be polished with sand paper and toothpaste, then clean it with distilled
water
Place the senor into 3.5 mol/L KCl solution for 6 hours before usage.
Buffer solution gets different value at different temperature so pls confirm the
temperature of buffer solution at calibration.
17
measurement installation installation position
is on a low
side
A. Connect the DC ammeter with
Difference at
A. instrument fault instrument to measure the loop
transmitting current, check the displayed value of
B. incorrect setting on PLC the instrument and judge the reason.
data
engineering value B. Check the mA loop resistance and
reset the receiving module migration
z Complete Set
z Transmitting controller 1pc(including one pair of fast installation
clamp/back cover )
z Instruction manual 1pc
z Temperature sensor 1 pc(optional)
I Order direction
24V electronic transducer is optional part. Pls specify the power supply of industrial
site before placing an order.
18
appendix I
appendixⅡ
Build up signal collection and linkage control system through configurable two
channels mA transmitting and three channels photoelectronic relay .mA transmitting
is divided into instrument mode and transmitting mode.
19
External connection in Instrument mode
20
Photoelectric switch connection:
21
Dosing pump controlled by Photoelectric switch
22
23