This document summarizes the specifications of the Wilcoxon 787A general purpose accelerometer. It has a sensitivity of 100 mV/g and can measure accelerations up to 80 g peak. It has a frequency response of 1-5,000 Hz within 10% and 0.5-10,000 Hz within 3 dB. Key features include versions for hazardous areas, a low profile, API 670 compliance, and manufacturing in an ISO 9001 certified facility. It has a hermetic sealing, measures 145 grams, and has a stainless steel case.

Uploaded by

Jhon Baldeon
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
102 views1 page

787A - Ficha Técnica - Acelerómetro Propósito General Con Salida Lateral

This document summarizes the specifications of the Wilcoxon 787A general purpose accelerometer. It has a sensitivity of 100 mV/g and can measure accelerations up to 80 g peak. It has a frequency response of 1-5,000 Hz within 10% and 0.5-10,000 Hz within 3 dB. Key features include versions for hazardous areas, a low profile, API 670 compliance, and manufacturing in an ISO 9001 certified facility. It has a hermetic sealing, measures 145 grams, and has a stainless steel case.

Uploaded by

Jhon Baldeon
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

General purpose accelerometer

787A
SPECIFICATIONS
Sensitivity, ± 5%, 25°C 100 mV/g
Acceleration range, VDC > 25 V 80 g peak
Amplitude nonlinearity 1%
Frequency response: ± 10% 1 - 5,000 Hz
 ± 3 dB 0.5 - 10,000 Hz
Resonance frequency 22 kHz
Transverse sensitivity, max 5% of axial Key features
Temperature response: –55°C –20%
  +120°C +10% • Certified versions available for
use in hazardous areas
Power requirement:
Voltage source 18 - 30 VDC • Low profile
Current regulating diode 2 - 10 mA • API 670 compliant
Electrical noise, equiv. g: • Manufactured in ISO 9001 facility
Broadband   2.5 Hz to 25 kHz 700 µg
Spectral 10 Hz 10 µg/√Hz
  100 Hz 5 µg/√Hz
  1,000 Hz 5 µg/√Hz
Output impedance, max 100 Ω
Bias output voltage 12 VDC
Grounding case isolated, internally shielded MIL-C-5015
2 PIN
CONNECTOR
Temperature range –55° to +120°C
Vibration limit 500 g peak
Shock limit 5,000 g peak
Electromagnetic sensitivity, equiv. g, max 70 µg/gauss
Sealing hermetic
Base strain sensitivity, max 0.002 g/μstrain
Sensing element design PZT, shear
Weight 145 grams
Case material 316L stainless steel 1/4-28 MOUNTING
SCREW

1/4-28 captive hex head screw,


Mounting
0.046” diameter safety wire hole
Output connector 2 pin, MIL-C-5015 style
Recommended cabling J9T2A / J9F
Accessories supplied: 1/4-28 captive hex head screw; calibration data (level 2)

Connections
Function Connector pin
power/signal A
common B
ground shell

Note: Due to continuous process improvement, specifications are subject to change without notice.
This document is cleared for public release.

Wilcoxon Sensing Technologies 8435 Progress Drive Tel: +1 (301) 330-8811 buy.wilcoxon.com
An Amphenol Company Frederick, MD 21701 Fax: +1 (301) 330-8873 www.wilcoxon.com
USA [email protected]
98493 Rev.H.2 09/18