MCFG Software Installation Guide

This document provides instructions for installing and using MCFG software to configure MEDICAN alarm devices. The key steps are: 1. Install drivers and MCFG software on a Windows PC connected to the MEDICAN alarms via USB-CAN interface. 2. Use the MCFG software to discover connected MEDICAN alarms by serial number and SID. 3. Various configuration and firmware update procedures can then be performed for individual alarms, which require power cycling the alarms after each change. Proper connections and device identification are important to ensure changes are made to the correct alarms. The document provides troubleshooting tips and details various MCFG functions like erasing configurations, uploading new ones, and changing

Uploaded by

Jeremy BACH
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
171 views15 pages

MCFG Software Installation Guide

This document provides instructions for installing and using MCFG software to configure MEDICAN alarm devices. The key steps are: 1. Install drivers and MCFG software on a Windows PC connected to the MEDICAN alarms via USB-CAN interface. 2. Use the MCFG software to discover connected MEDICAN alarms by serial number and SID. 3. Various configuration and firmware update procedures can then be performed for individual alarms, which require power cycling the alarms after each change. Proper connections and device identification are important to ensure changes are made to the correct alarms. The document provides troubleshooting tips and details various MCFG functions like erasing configurations, uploading new ones, and changing

Uploaded by

Jeremy BACH
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Installation manual

for MCFG software


MCFG (V1.3)

INDEX Page

MCFG SETUP
 Installation Procedure…….…………………………………………………………………………. 2 1
MCFG LOGIC
 Operation…………………………………………….……………………………………………….. 3 2

WORKING WITH MCFG


 Discovering MEDICAN alarms……..…………………………..………………………………….. 4 3
 What if serial number and SID number don’t appear……………………………………………. 5
 Erasing the Configurations…………………………………………………………………………. 6
 Uploading new Configurations……………………………………….………………………………. 7
 Language settings………………………………………………….……………………………….. 8
 Uploading and updating new Firmware…………………………….…………………………….. 9
 How to look for the Firmware version on signalization…………….……………………………. 10

CHANGES
 About SID numbers………………………………………………….……………………………… 11 4
 Changing SID numbers……………………………………………….……………………………. 11
 Procedure for changing SID numbers….………………………….…………………………….. 12

CLEARING
 Clearing logbook…………………………………………………..……………………………….... 13 5

INSTRUCTION
MCFG (V1.3)

INSTALLATION PROCEDURE 1

MCFG SETUP
For proper installation and operation of software, please follow these steps. All setup files
are located on MEDICOP CD media. Software require PC with Windows XP or later and
Microsoft .NET Framework 2.0 SP2.

STEP CAN-BUS INTERFACE

1 Install IXXAT driver (USB-to-CAN),


“vci_3_2_2_2098.exe”.

Connect USB-CAN interface to PC.


Connect CANBUS connector to supply
connection board on MEDICAN (red
mark on picture of supply connection
board). If the interface is connected
properly, a green LED (USB) should be
turned on at interface.

Before we continue with installation the green LED (USB) must be turned on. If the red LED
is on, then there are problems with USB-CAN driver. Restart PC, reconnect interface.

SUPPLY CONNECTION BOARD CAN-BUS INTERFACE

2 Install Run-time engine “[Link]”.

3 Install MCFG configurator “[Link]” in “MCFG_installer” folder.

4 Run the program and perform working test for MCFG.

INSTRUCTION Page 2
MCFG (V1.3)

2
OPERATION

MCFG LOGIC
The diagram below shows the logic of the MCFG program.

BEFORE SETTINGS SETTINGS PROCEDURES AFTER SETTINGS

ERASING THE CONFIGURATIONS

LOADING NEW CONFIGURATIONS

LANGUAGE SETTINGS
DISCOVER POWER OFF / ON
CHANGING SID NUMBERS

CLEARING LOGBOOK

UPLOADING AND UPDATING NEW FIRMWARE

Before we start with each individual setting We can make settings only on one MEDICAN After each individual setting we must first
procedure, we must discover all MEDICAN alarms alarm at a time. disconnect CAN – BUS cable from MEDICAN
and select the one on which we want to make supply connection board and then power OFF /ON
changes. main supply.

INSTRUCTION Page 3
MCFG (V1.3)

DISCOVERING MEDICAN ALARMS 3

WORKING WITH MCFG


When we run MCFG program, we have to wait for the green light in box ONLINE to
appear.

STEP

1 Select CONFIG tab (tab is


selected by default when we
start the program).

2 Press DISCOVER button.

The serial number and SID


number of MEDICAN alarm
should appear.

3 TIMEOUT (time for


discovering the MEDICAN
alarms) is factory set to
300ms.

If we have more than 20


MEDICAN alarms on
network, we have to enlarge
TIMEOUT to 1000ms.

SID and serial number should appear in the section that is shown on the next picture.

STEP

When MCFG is discovering


alarms, the green (CAN)
LED on USB-CAN
interface is blinking. This
means that data exchange
is in progress.

If the SID number is larger


than 99, we have to press
tab “DROP DOWN” to see
complete serial number.

INSTRUCTION Page 4
MCFG (V1.3)

WHAT IF SERIAL NUMBER AND SID NUMBER DON'T APPEAR 3

WORKING WITH MCFG


In case that serial and SID number don’t appear, something is not right with: connections,
MEDICAN alarm or drivers for interface. On USB-CAN interface the red LED is on.

TROUBLESHOOTING

1 Close the MCFG program and all other software that potentially use interface
driver (MiniMon from IXXAT).

2 Check power supply on MEDICAN alarm.

3 Check all connections: PC – USB-CAN interface – MEDICAN alarm.

After we have eliminated problems, we again run MCFG program and repeat the process
of discovering the MEDICAN alarms.

With configuration software MCFG we can change:

CHANGES

1 Configurations (number of gases, alarm settings…)

2 Languages

SID numbers – UNDER CERTAIN CONDITIONS (chapter about SID changes)

IMPORTANTE NOTICE

Because every customer can have his own requests about configurations, config
files are send to the individual customer by e-mail or CD (as customer wants).
When updating configurations we suggest that only one MEDICAN alarm is on the
CAN network. This way we eliminate possibility to update the wrong one.

INSTRUCTION Page 5
MCFG (V1.3)

ERASING THE CONFIGURATIONS 3

WORKING WITH MCFG


Before we uploading new configurations, we should erase current settings.

Follow the steps:

STEP

1 Select CONFIG tab.

2 Press DISCOVER button.

The serial number and SID


number of MEDICAN alarm
should appear.

3 Select correct SID and serial


number.

4 Press ERASE CONFIGURATION


button.

3 Confirm erasing with OK.

After we confirm succesfull


erasing MEDICAN alarm will
restart and the LCD on alarm will
be empty.

4 Disconnect CAN-BUS cable and main power supply from supply connection board of MEDICAN.

IMPORTANTE NOTICE

We have to individually ERASE CONFIGURATION on each MEDICAN alarm we


want to load new complete configuration.
Only in case when updating part of configuration you don’t erase existing one.

INSTRUCTION Page 6
MCFG (V1.3)

UPLOADING NEW CONFIGURATIONS 3


Follow the steps:

WORKING WITH MCFG


STEP

1 Select CONFIG tab.

2 Press DISCOVER button.

The serial number and SID


number of MEDICAN alarm
should appear.

3 Select correct SID and serial


number.

4 Browse on PC for new


Configuration file.

5 Press UPDATE button.

On the bottom of MCFG window


we can observe updating
progress.

After the updating progress is


complete, MEDICAN alarm will
automatically restart.

6 Disconnect CAN-BUS cable and main power supply from supply connection board of MEDICAN.

IMPORTANTE NOTICE

During uploading green Running button is on. If RED Failed button appear you
need to power OFF/ON alarm and repeat uploading procedure.

INSTRUCTION Page 7
MCFG (V1.3)

LANGUAGE SETTINGS 3

WORKING WITH MCFG


Follow the steps:

STEP

1 Select CONFIG tab.

2 Press DISCOVER button.

The serial number and SID


number of MEDICAN alarm
should appear.

3 Select correct SID and serial


number.

4 Select FONT tab.

5 Browse on PC for new


LANGUAGE file.

6 Press UPLOAD GRAPHICAL


DATA button.

On the bottom of MCFG


window we can observe
updating progress.

While uploading is in
progress, don’t touch
MEDICAN’ buttons.

8 Disconnect CAN-BUS cable and main power supply from supply connection board of MEDICAN.

IMPORTANTE NOTICE

During uploading green Running button is on. If RED Failed button appear you
need to power OFF/ON alarm and repeat uploading procedure.
After we uploaded language file we have to switch to next menu on MEDICAN
alarm - to refresh changes on alarm screen.

INSTRUCTION Page 8
MCFG (V1.3)

UPLOADING AND UPDATING NEW FIRMWARE 3

WORKING WITH MCFG


Follow the steps:

STEP

1 Select CONFIG tab.

2 Press DISCOVER button.

The serial number and SID


number of MEDICAN alarm
should appear.

3 Select correct SID and serial


number.

4 Select PROGRAM tab.

5 Browse on PC for new


FIRMWARE file.

6 Press PROGRAM button to


start upload.

Upload progress last


approximately 3 minutes.

After that the MEDICAN


alarm needs 1-2 minutes to
reset itself.

6 Disconnect CAN-BUS cable and main power supply from supply connection board of MEDICAN.

IMPORTANTE NOTICE

During uploading green Running button is on. If RED Failed button appear
you need to power OFF/ON alarm and repeat uploading procedure.

INSTRUCTION Page 9
MCFG (V1.3)

HOW TO LOOK FOR THE FIRMWARE VERSION ON MEDICAN ALARM 3

WORKING WITH MCFG


Follow the steps:

STEP

1 Disconnect main power supply on supply connection board of MEDICAN alarm.

2 Press and HOLD the SERVICE MUTE BUTTON on the back side of alarm display modul. At the
same time connect main power supply back on the MEDICAN supply connection board.

3 We entered the service menu where the version of FIRMWARE is displayed in first line.

INSTRUCTION Page 10
MCFG (V1.3)

ABOUT SID NUMBERS 4

CHANGES
SID number is a name of individual MEDICAN alarm – STATION ID. A unique name (SID)
is assigned by MEDICOP.

SID numbers have very important role in MEDICAN alarms network. It is very important
that each alarm have its own SID number. If two or more alarms in CAN network will have
the same SID number, a problem will appear at data exchange.

In case those MEDICAN alarms are NOT CONNECTED in CAN network, they can have
the same SID numbers.

When we have COPY of individual MEDICAN alarm, we can talk about MOTHER –
DAUGHTER relationship. The original alarm is MOTHER, copy is a DAUGHTER.
Daughter will show the same parameters that are shown by mother. In this kind of
configuration it is important not to change MOTHER’S SID.

CHANGING SID NUMBERS

KIMPORTANTE NOTICE

We can change SID number only with the knowledge of


manufacturer – MEDICOP.

CAN network We can change SID number with number that no other MEDICAN
SID number signalization have.

(If we want to change SID for the second time or on primary


(MEDICOP) number, or some other number, first we have to reset that
alarm with disconnecting the power supply. Otherwise, CAN network
will not work.) THIS APPLIES FOR FIRMWARE VERSIONS BEFORE
VERSION 2.19!

MOTHER It is NOT ALLOWED to change SID number on MOTHER signalization;


DAUGHTER otherwise DAUGHTER will lose reference number from which it
SID numbers receives information data to show parameters.

INSTRUCTION Page 11
MCFG (V1.3)

PROCEDURE FOR CHANGING SID 4

CHANGES
Follow the steps:

STEP

1 Select CONFIG tab.

2 Press DISCOVER button.

The serial number and SID


number of MEDICAN alarm
should appear.

3 Select correct SID and serial


number.

4 We type in a new SID


number and press Set NEW
SID.

New SID will appear next to


serial number of MEDICAN
alarm.

5 Disconnect CAN-BUS cable and main power supply from supply connection board of MEDICAN.

IMPORTANTE NOTICE

DON’T CHANGE SID NUMBERS UNLESS IT IS NECESSARY

INSTRUCTION Page 12
MCFG (V1.3)

CLEARING LOGBOOK 5

CLEARING
When we will make some configuration changes, language or other parameters changes,
powering ON and OFF alarm, CAN error… All those events MEDICAN alarm will log in
Logbook. User can erase all previous events in Logbook anytime.

STEP

1 Select CONFIG tab.

2 Press DISCOVER button.

The serial number and SID


number of MEDICAN alarm
should appear.

3 Select correct SID and serial


number.

4 Press CLEAR LOGBOOK


button.

5 Confirm clearing with OK.

6 Confirm with OK when finished.

7 Disconnect CAN-BUS cable and main power supply from supply connection board of MEDICAN.

IMPORTANTE NOTICE

After we cleared LOGBOOK or made other configuration changes we have to switch to next
menu on MEDICAN alarm - to refresh changes on alarm screen.

INSTRUCTION Page 13