BET Manual ENG (2020)
BET Manual ENG (2020)
EQUIPMENT INFORMATION 4
Shipment Inspection....................................................................................................................... 4
Specifications .................................................................................................................................. 5
BACKGROUND 6
Chemisorption............................................................................................................................... 14
PRINCIPLES OF OPERATION 15
Calibration .................................................................................................................................... 19
OPERATING PROCEDURE 22
Start Up......................................................................................................................................... 23
Introduction .................................................................................................................................. 31
Shipment Inspection
Your PMI Automated BET Sorptometer was carefully inspected and
tested prior to shipping. Please make sure that the instrument and accessories
have arrived in good condition. Claims for any damage must be made against the
delivering carrier, but PMI should also be notified so that we can assist you as
much as possible.
Notes on Installation
This manual is not designed to be a tutorial on the operation of the BET
Sorptometer, but rather is designed as a reference manual and a guide to the
theory of operations. (Though it has been found to be adequate in certain cases
where installation was not requested by the end user.) All automated BET
Sorptometers come with on-site installation by a factory representative within the
continental United States. It is recommended that the manual be read by the
people who are to operate the instrument before the installer arrives, but that they
do not attempt to actually operate the instrument until guided by the installer. If,
when reading the manual, you have any questions about the instrument, don't
hesitate to contact PMI for further information. The installer is well versed in all
aspects of the instrument and can best answer your questions.
0-10 Torr
resolution 1 part in 20,000
accuracy 0.15% of reading
If the constant C is sufficiently large (much greater than one), the BET
equation [1] can be simplified to obtain an estimate of Nm from a single
adsorption measurement:
P P 1
N m,sp = N 1 - , for 1 > >> [5]
P0 P0 C
where the subscript sp denotes the single-point BET method. The multi- and
single-point BET surface areas can be compared by substituting Nm (from
equation [1]) or Nm,sp into equation [4]:
Sg - Sg,sp 1 - (P P0 )
sp = = [6]
Sg 1 + (C - 1)(P P0 )
where sp is the relative error in the single-point BET surface area, and is
tabulated as a function of C and P/P0 in Table I. In the single-point method, the
adsorption measurement is made at a relative pressure between 0.2 and 0.3 so that
the error in the surface area will be only a few percent if C is greater than about
50 (as it often is for nitrogen adsorption). Thus, the additional time and effort
required in the multi-point BET method may not be justified for many
applications.
The adsorption measurements are usually made at -195.8°C (the normal boiling
point of nitrogen), and substitution of numerical values into equation [8] gives the
corresponding relationship between the critical pore radius and the relative
pressure:
-1 − 13
P P
rp,c (in ? = 9.57 ln 0 + 6.05ln 0 [10]
P P
Values of the adsorbed film thickness, Kelvin radius, and critical pore radius
calculated from these equations are tabulated as a function of relative pressure in
Table II.
In this way, Vp and Sp can be determined for each decrement in P/P0 (which
corresponds to a certain range of pore radii) so that the pore volume and pore
surface area distributions are obtained. The total pore volume (Vp) and total
surface area (Sp) are then given by:
Vp = (Vp ) [13]
rp
S p = (S p ) [14]
rp
This total surface area is based upon the assumption that the pores are cylindrical
and does not include any contribution from pores smaller than about 15 Å in
diameter (for which the Kelvin equation should not be applied); thus, it is less
reliable than the BET surface area.
for Nitrogen
Adsorption/Desorption at -196°C
P/P0 t(Å) rK(Å) rp,c(Å)
where N/Nm is the fractional coverage of adsorption sites and K is the adsorption
equilibrium constant, which is a function of temperature. Note that 0 < N/Nm < 1
since chemisorption is limited to monolayer coverage.
where the subscripts n-1 and n denote any two consecutive data points. This
generalized equation reduces to equation [16] when n=1 (i.e. for the first data
point after the system has been evacuated), since Pf,0 = Pze = 0 and N0 = 0.
The system volumes VA, Vit, and Vtt are determined through calibration
tests (described below), while Vs can be calculated from the mass of the sample if
its true (absolute) density is known or can be measured via gas pycnometry. Gas
pycnometry is conveniently performed at room temperature prior to the
adsorption measurements. Although helium is recommended, other gases such as
nitrogen or krypton can be used for pycnometry since they behave nearly ideally
and are essentially non adsorbing at room temperature and low pressures. The
procedure is quite simple: the system is evacuated to a residual gas pressure Pr,
the reference volume is pressurized to Pi, and the isolation valves are opened to
allow the gas to equilibrate at Pf. The true volume of solid in the sample can then
be obtained from the following equation:
Pi - Pf
Vs = Vit + Vtt - VA [18]
Pf - Pr
If desired, the user can repeat these tests to verify the instrument calibration.
Values for Vit and Vtt must be determined by the user for each sample cell
and adsorption temperature of interest. This calibration is performed prior to the
adsorption test, using an empty sample cell and the appropriate adsorbate gas.
With the sample cell at room temperature (it), the system is evacuated to
essentially zero pressure (Pze), the reference volume is pressurized to Pi, and the
isolation valves are opened to allow the gas to equilibrate at Pf,it. The empty
sample cell is then cooled or heated to the desired adsorption temperature (tt), and
the new equilibrated gas pressure Pf,tt is measured. Simultaneous solution of the
following equations yields the values for Vit and Vtt:
(P - P )V = (P
i f,it A f,it - Pze )(Vit + Vtt ) [20]
Vit Vtt
(P - P ) V
i f,tt
A
= (Pf,tt - Pze ) + [21]
it it tt
It is important to note that each set of values for Vit and Vtt applies to a specific
sample cell and adsorption temperature, and must be used accordingly. The
following equation may also be useful:
Vit + Vtt = VB + VC [22]
There are a large number of computers, either an IBM PC or compatible, that you
can choose from to run the software. The minimum system required to run the
BETWIN Software System is an 80286. This will slow down the speed of the
software considerably, but it is a valid system.
We recommend that you have at least a 80386sx-20 (the 20 stands for the
processor speed 20 MHz). This gives you more processing power and the
BETWIN Software System runs much faster.
To further increase system performance you might wish to consider the following
options:
• a more powerful computer system (e.g., an 80486DX-50)
• a math co-processor (this should be matched to your computer system
and include: 80287, 80387sx, Wietek, etc.)
• more memory (the performance of Windows improves with an
increase in memory)
• a GUI Accelerator (Graphical User Interface) (these are new graphics
cards available to increase the speed of graphics display)
All of the above items will increase your computer's interaction with Windows,
thus increasing the speed at which your software runs.
After the test, the final sample weight can be determined by re-weighing
the sample cell immediately after removing it from the instrument (the initial and
final sample weights may differ somewhat, due to the removal of moisture and
adsorbed gases during pretreatment). The gas pressure within the sample cell
should always be adjusted to essentially atmospheric pressure before attempting
to remove the cell from the instrument. This is done automatically by the
computer at the end of any automated test.
The sample cell should always be evacuated before removing the Dewar
flask after the test, to prevent the potential buildup of pressure due to desorption
of gas from the sample as it warms to room temperature. This is done
automatically by the computer at the end of any automated test.
Introduction
The BETWIN Software System was developed by Porous Materials, Inc.
in response to an increasing number of requests for a software package that would
run under the Microsoft Windows 3.x environment. For a more detailed
description of the requirements needed to make use of BETWIN please see the
section of this manual entitled Computer Hardware and Software
Requirements.
The following portion of this manual describes the various functions of the
BETWIN Software System and how to make use of them. This information is
also available from within the BETWIN Software System by using the Help
menu or buttons found on many of the screens within BETWIN.
The above image is a copy of the Main Screen for the BETWIN Software
System. It displays information on the current user, default gas, and default units,
as well as the menu bar that controls the software system itself.
Each item along the menu bar (File, Execute, Report, Modify, Setup,
Help) when activated displays a menu of further options which control the BET
Sorptometer, report on collected data, modify testing parameters, etc. These
menus will be explained in detail below.
Select User
Using the Select User option, you may switch between existing users'
settings. For more information regarding what a user is, please see New User
below.
The Select User option gives you a list of all existing users and asks you
to select one of these to attain a list of system defaults.
The BETWIN software system now supports multiple users, each one
capable of setting some of the machine parameters to deal with the type of test
they wish to run. This saves setup time and effort when a different user wishes to
run a test.
The computer stores the default directories, top temperature, the settings
of the last test run by that user, etc. for each user.
When creating a new user you are asked to enter a name by which the
system may identify this person (the name must be eight letters or less). This
name is used to create a sub-directory within the USERS directory (i.e. if you
entered the name SAM and your BETWIN directory was C:\BETWIN the
computer would create the directory C:\BETWIN\USERS\SAM). The computer
then copies the needed default files into the new user's directory.
You are also asked to select which of the defined gasses you wish to use
as this user's default gas. The gasses you have to select from are stored in the
GASSES sub-directory of your BETWIN directory. Any gas supplied with the
BETWIN Software system, or created using the utility Create a New Gas, and
stored within the GASSES sub-directory is available for selection as your default
gas.
Finally you are asked to select the units you wish your pressure values to
be displayed in. The three most often used pressure units are Torr, PSIA, and
kiloPascals.
There are also three buttons at the bottom of the New User form. They
are:
CANCEL - this aborts the user creation process.
This option allows you to change the default gas for the current user.
Because different gasses have different characteristics, this may affect the options
available on other forms (i.e. the Automatic Test).
You are asked to select the name of the gas you wish to use from the
supplied list (these gas files are stored in the GASSES sub-directory within your
BETWIN directory). This gas will be the one used by the system for tests and
calibrations unless changed before the test starts or you again change the default
gas.
The Select Units option allows you to select a different unit of pressure.
The selected unit of measure (i.e., PSIA, Torr, etc.) is used to display pressure
values. For more information on what these units mean please see the section
entitled Pressure Units.
The form itself consists of a list box and two buttons. The list box allows
you to select from any of the previously created pressure units. To select a unit
highlight the pressure units you wish to use by clicking on that unit name.
This option, when selected, exits the BETWIN Software System and
returns control to the Microsoft Windows 3.x environment.
When this option is selected, the user is presented with the above form to
confirm the fact that he/she wishes to leave the BETWIN Software System.
Automated Test
The Automatic Test Settings screen is the screen that appears when
Automated Test is selected from the Execute Menu unless there is currently no
The sample screen above differs from the actual screen for a single
chamber BET. The box near the top which indicates the testing unit for which the
settings are being displayed is not normally present on a single chamber BET. It
is visible in the form above for display purposes.
The Automatic Test Settings screen allows the user to run a test using the
previously selected options with a minimum of input. The information requested
is as follows:
The following information is for sample chamber -
This frame is visible only on multiple unit BET's. The numbers
visible coincide with the testing units on the BET. If the circle
beside a given number is displayed with a dot in it as in the
example above the information being displayed is for that unit. To
switch units simply click on the desired unit number.
Weight of Sample (gm) -
The weight of the sample to be tested is entered here. The weight
displayed when the form is first displayed is that of the last sample
on that unit.
Sample ID -
This field contains the ID of the previous sample that was tested on
this unit. It allows you to identify individual samples by giving
them unique ID's.
The black bar in the center of the screen divides the information displayed
into that which deals with individual test units and that which deals with all tests.
The information above the bar, which selects which unit you are working with,
deals with the entire unit. These include Instrument Temperature, Room
Temperature, whether you wish to calibrate the instruments gauges before
proceeding with the test, etc.
The information below the bar deals with the unit you are currently
working with. This information includes the sample chamber data file, the name
of the file you wish to use to store test data, whether or not you wish to take
intermediate data points, etc.
The exact options available to you depend upon the version number of
your BET Sorptometer and the gas you are using to run the test. Listed below are
all the options available when running an Automatic Test.
The following options affect the test on all units (for multiple unit
systems):
Instrument Temperature -
This is the internal temperature of the BET Sorptometer.
You must have at least one of the three tests (adsorption, desorption, or
multipoint) set to a valid file for at least one of the sample chambers. Setting
them all to SKIP for all of the sample chambers will generate an error message
when you try to leave this screen.
The four command buttons in the lower right corner of the screen are used
for the following:
START TEST(s) -
When you are done setting up your test(s), clicking on Start
Test(s) switches you to the Test Status screen which allows you to
keep track of what the test running on each unit is doing.
RESET -
The Test Status screen keeps you informed of what the test is doing at any
given point. The exact appearance of the screen depends upon how many testing
units your system has. There will be one column of information for each of the
units your BET Sorptometer has (i.e. 1-6).
To control the test there are a several command buttons located on the
screen. The major buttons are located along the bottom edge of the screen and
along the left edge. When the Test Status screen is loaded only two of these six
buttons will be active. The two are the BEGIN and HELP buttons. The others
will be inactive and the captions on them will be at half intensity and will remain
so unless you enter HOLD mode. There are also two other buttons located in the
left side of the screen, which, depending upon what version of machine you have
and whether or not you have a diffusion or turbo pump, may not be visible. The
following is a list of the these buttons and their functions:
BEGIN -
This button starts the actual test program using the options you set
earlier using the Automated Test Control Screen. This button is
the default button when the screen loads so that you only need to
For more information on the test procedures please see the next
section entitled Automated Test Description.
Manual Control allows you to control the various valves, heaters, lifters,
etc. without actually running a test. You may also get feed-back from the
instrument on the pressure reading from various gauges, position of the lifter,
temperature, etc.
There are separate sections for controls that affect individual testing units
and for those that affect the entire instrument.
Not all options are available on all machines, so there may be what appear
to be blank spots on the control screen. As an example, if you do not have a
diffusion pump as part of your system, the Diffusion Pump controls will not be
visible when you start manual control. In the case of valves that are not present in
your BET (i.e. Valve V which is present only in Version 5.2 and later machines),
they will appear as only half intensity and you will not be able to select them.
This is also true of the Unit Number section if your BET Sorptometer has less
than six units.
The two buttons in the lower right hand corner of the Manual Control screen are:
QUIT - returns you to the portion of the program from which Manual
Control was called.
HELP - calls the Help system.
Before you can run a test or even calibrate the internal volumes of a
machine you must have accurate gauge readings. This routine resets the zero
points and cross-over points of the gauges so that the readings you attain from
them are accurate.
All the activity for this section takes place on one form. You are first
asked to select which testing unit(s) that you wish to calibrate. If your BET
Sorptometer has more than one unit you may calibrate as many as you wish by
clicking on the appropriate unit's number. (Example: If your BET Sorptometer
has three units and you wish to calibrate units one and three but not two, you
would select those two units leaving unit two's check box empty. During the
calibration process unit two would be ignored.)
You are also asked whether or not you wish to do an absolute calibration.
If you select YES you are also asked to enter the current atmospheric pressure.
This entered pressure is used to gain a more accurate calibration of the
instruments by using this pressure value as a set point to calibrate the gauges.
Absolute calibration also requires that there NOT be a sample chamber
installed on the unit(s) that are being calibrated.
Before you can run a test on a sample you must place that sample into a
sample chamber. To attain accurate results from that test you must calibrate that
sample chamber so that the computer knows the proper volumes for that chamber.
This routine is used to determine the correct volumes needed by the program to
run a test using a new sample chamber.
This procedure uses two separate screens. The first is used to set up the
procedure (which unit to use, gas inlet valve, etc.). The second is a status form
that allows the user to keep track of what the machine is doing and also displays
the results of the calibration when completed.
To setup the calibration you need to select which unit you will be
calibrating the sample chamber on (you may only run one calibration at a time),
which gas inlet valve to use, the PI pressure for the gas you are using (only if the
gas doesn't have a P0 value), room temperature, the instrument temperature, and
whether you are using liquid nitrogen (for cold gas), and if you are using a hot gas
or not liquid nitrogen, what the testing temperature is.
When you click BEGIN the first form will be unloaded and the
Calibration Status Screen will appear.
There are also six buttons that appear on the form, one of which is not always
visible, another which is used for four separate purposes. The buttons and their
meanings are as follows:
BEGIN - starts the actual calibration process. Once you have
clicked on BEGIN the caption for this button changes to
Hold.
Hold - places the BET Sorptometer into hold mode. This
pauses the calibration process, enables the Move Lifter,
Manual Control, and Abort Calibration buttons, and
changes the caption for this button to Resume.
Version 5.2 and later machines also have valve V which closes off the
vacuum valve so as not to damage it if the pressure is too high, and the P0 system
has been moved to between valves 4 and 7. The first of the additional volumes
includes the internal volume with valve V open. The second is similar to the first
but valve 7 is closed and valve 4 is open. This gives you the testing volume
including the P0 section up to the P0 valve.
The two sample chambers that you use should be as different in size as
possible. The greater the difference, the better the results you will get. (Example:
if you use two chambers of 18.9 cc and 24.6 cc you will not get as accurate a
measure as if you had used one of 9.6 cc and another of 30.3 cc)
This procedure uses two separate screens. The first is used to set up the
procedure (which unit to use, gas inlet valve, etc.). The second is a status form
that allows the user to keep track of what the machine is doing and also displays
the results of the calibration when completed.
To set up the calibration you need to select which unit you will be
calibrating the sample chamber on (you may only run one calibration at a time),
which gas inlet valve to use, the PI pressure for the gas you are using (only if the
gas doesn't have a P0 value), room temperature, the instrument temperature, and
whether you are using liquid nitrogen (for cold gas), if you are using a hot gas or
not liquid nitrogen, what the testing temperature is, and the volumes of both
sample chambers in cc's.
When you click BEGIN the first form will be unloaded and the
Calibration Status Screen will appear. For more information on the Status
Screen see the above section entitled Calibration Status Screen.
Execute Report
The Execute Report option allows you to attain printed reports on the
data collected during an Automated Test or entered using the Manually Create
a Data File option under the Modify Menu. The image above represents the
report form after the Advanced button has been pressed.
Starting with the BETWIN Software System you are now able to select
only those reports and/or graphs you wish to have printed (i.e. you can choose to
have only the adsorption graphs and the BET Surface Area Analysis report). This
saves printing time because it will only do the calculations for those sections that
you wish printed.
There are five options/sets of options that control what reports are printed,
where they are printed to, and, if you are doing graphs of any of the reports, the
appearance of those graphs. The five report options which need to be set are as
Output Device
The Output Device option allows you to select where the reports
and graphs you select will be displayed or stored. The three options for
output are:
Printer -
Selecting Printer will send the report and graphics to the
systems default printer.
Screen -
This option causes the report and associated graphics to be
printed to the computer screen
Text File -
The report is stored as ASCII text to a disk file. The
computer will bring up the File Selector for you to choose
a file to store the report in or to enter a new file name.
Outputting to a Text File does not allow you to do graphics.
Plotter -
This selection will output only the graphs of the reports
selected. Currently the plotter must be a Hewlett-Packard
7475A or compatible.
Each time you select a valid data file, the number of files you have
selected so far will be checked. If you have selected fewer than 10 data
files, a box will appear in the center of the screen asking you if you wish
to enter more reports. This box also includes the number of files you are
allowed to choose (10) and the number you have selected to that point.
If you wish to include another data file in this report click on Yes,
otherwise select No. If you select Yes, the File Selector will be displayed
again so that you may select another file.
There are also three buttons that appear on the left edge of the Report form.
These buttons perform the following tasks:
BEGIN - begins printing the report using the options selected by the user.
The program will first call the File Selector so that the user
may select a data file(s) from which to attain the data for the
report. It will then use that file(s) to generate the requested
reports and graphs, outputting them to the selected device.
HELP - calls the BETWIN Help System and displays information on
the Report Setup Screen.
CANCEL - cancels the report program without printing any reports and
returns you to the main screen for the BETWIN software
system.
20 DUTCHMILL RD.
For more information on the Notepad program use the Help available
within Notepad application.
There are many factors that control an Automated Test. Some of these
factors have been put into a parameter list to allow easy user modification. For
each gas, there is a default parameter list that is used when running a test with this
gas, and other parameter lists that can be made default by using the parameter
editor.
The following are the options you are allowed to change which control an
Automated Test:
Minimum PI Undershoot Count & Maximum PI Undershoot Count -
When trying to reach a PI value, the optimal pressure is calculated,
and then allowable minimum and maximum pressures are
There is also a series of seven buttons down the right side of the form.
The buttons and their meanings starting at the top are:
The third section consists of the main form with a frame entitled Heading
Information and the following two buttons:
CANCEL - ends the creation of the data file without saving any
information to the file.
HELP - calls the BET Sorptometer Help system and displays
information on the steps necessary to Manually Create a Data
File.
The Heading Information is that information that appears at the top of a
data file before the data itself. The exact information requested depends upon the
gas you are using. The information requested is as follows:
Enter the P0 value, or 0 for default / Enter the Surface Area Factor,
or 0 to skip -
The first item you are asked to enter is the P0 Value for this test.
If you want to use the default value, enter a 0 instead. If the gas
you are using can not be used to calculate pore volume, it will not
have a P0 Value. In this case, enter the Surface Area Factor, or 0
to skip Chemisorption Surface Area calculations.
Enter the outgassing vacuum (microns) -
You are also asked to enter the outgassing vacuum in microns.
Normally, the automated system sets this value to 20. Enter
whatever value you like. This value is not used in calculations,
and is just printed on the report.
Enter the outgassing temperature (C) -
Next enter the temperature of the sample when the vacuum was at
the value previously entered, just before the test was started. As
with the vacuum reading, this value is not used in any calculations,
and is just presented in the report as information to the reader.
Enter the zero correct factor (0 for true pressure readings) &
Enter the scale correct factor (1 for true pressure readings) -
The next two inputs are for the zero correction factor and the
pressure scale factor. These are used for ease of entering pressure
values gathered on a unit that does not give direct readout in PSIA.
Note that if your software has been set for alternate units, such as
kPa, and you are entering data in these alternate units, then you
would set the pressure scale factor to 1.
When the entries for the above topic have been entered to your satisfaction,
click on the Done button within the Heading Information frame to continue to
the next section.
After clicking the Done key from the previous section, that frame will be
disabled and the items dealing with the fourth section will become visible. The
fourth section consists of a frame for use in entering data points and a list of PI
and PF values that have already been entered. The list is empty to begin with but
eventually fills up as you add points to the data file. If you enter more points than
there are room for in the list box, simply click on the arrows or bar that appear in
the list box. To enter data points follow the directions as they appear in the Data
Entry frame. Each time you click on ENTER, the data will be saved and a copy
of the point will be displayed in the list box.
After you have answered these questions and are satisfied with the
responses click on the DONE button within this section of the screen. This
finishes writing the test information to the data file.
You may now go to the Report Menu to setup and run a report using this
data.
Note: If the data you entered was not from a real test, strange
results can happen in the report program. You can also use any
text editor to edit a data file, which will have an extension .BDT,
but there are certain lines that are added by this program which
have to be the values they are. If you create or edit a data file
using a text editor, you may be stopped when submitting it to the
report program with a message to the effect that this data file was
not created on the system, and returned to the main menu.
Use of the Curve Fit utility allows you to smooth out the curve of the data
collected during an Automated Test or manually entered using Manually Create
File.
You are first asked to enter the name of the data file you wish to curve fit.
The File Selector is called to allow you to do this. Once you have selected a
valid data file you are presented with a form and asked to set the following three
options:
Percentage Error Allowed (0-100) -
This is used to determine if a data point is within the allowed range
of difference from the norm.
Number of Points to Use for Fitting (3-10) -
This value is used to determine the number of points the program
will use to fit a point to the curve. The data points are taken on a
sliding selection from 3 to 10 points (i.e. if you enter a 5 the
program will first use points 1 through 5 then 2 through 6, etc.) and
a polynomial will be generated using these points.
Maximum Distance Between Any Two Data Points in _______ -
The label for this option depends upon whether the gas you used
for the test could do Pore Volume. If it could and you asked for
Pore Volume, you are asked to enter the maximum distance
between data points in P/P0. If you did not ask for, or the gas is
not capable of doing Pore Volume, you will be asked to enter the
distance in terms of the pressure units you have chosen to use. The
value you enter will be used to determine whether or not to
generate an interpolated data point.
You are also presented with three buttons. They perform the following tasks:
BEGIN - starts the actual Curve Fit procedure.
A good starting point is to use a percentage error of 0.1 and 5 for the
number of points used. This will use points 1 through 5 to fit points 1, 2 , and 3;
points 2 through 6 will be used to fit point 4, etc.
As the points are fit to the curve, the box at the bottom of the form labeled
Percent Complete will slowly change color to show you how much of the curve
fitting has been completed. When the program is finished curve fitting the data,
the File Selector is again displayed so that you may enter a file to output the
newly fitted data file into.
The first call to the File Selector is for the file which contains the raw data.
If this file already contains interpolated data, you will receive a warning message
and will be asked if you want to continue. It is best to interpolate only raw data.
Interpolating already interpolated data is less accurate, but can be done.
Next you will be asked for the interpolation list, which should already
have been created by using the New Isotherm List option from the Setup Menu.
After selecting the isotherm list file, the data will be interpolated and you will be
asked to enter the name of the new data file. It is best not to write over the old
data file in case you ever want to re-interpolate it, or show the raw data. When
done, you will be returned to the main menu. From there you can run a report on
your new interpolated data file.
For more information on what each field means please see the section on
Create New Gas.
The New Multipoint List option allows you to create a file containing a
list of P/P0 values to be used by BETWIN as target pressures when performing a
Multipoint Test during an Automated Test. If the gas you have selected as your
default gas is not capable of doing Surface Area calculations, you will be
presented with an error message to this affect and returned to the Main Menu.
The top half of the form consists of information dealing with the creation
of a Multipoint List including the valid range of P/P0 points for the default gas
(i.e. for Nitrogen the range is 0.04 to 0.35).
To create the list simply enter P/P0 values in ascending order within the
allowed range and either click on the Enter button or press the RETURN key on
the keyboard. For each value you enter in this manner the value will be stored in
the Multipoint List file and displayed in the Previous Values list box. This list
contains all of the previously entered P/P0 values.
The four command buttons that appear on the form perform the following
tasks:
HELP - calls the BETWIN Help system and displays information on
the New Multipoint List section.
When creating a New Isotherm List File you must first enter the name
you wish to use to store this file. You are presented with the File Selector so that
you may enter a file name and where to store the file. After entering a valid file
name the main form for this option will be displayed.
The exact appearance of this form depends upon the current default gas. If
the gas you are using is not capable of doing Pore Radius calculations, a message
to this affect will appear at the top of the screen informing you of this fact and
telling you that you need to enter the raw PF values instead of Pore Radius values.
If the gas you are using can do Pore Radius calculations this message box will be
empty.
If your gas is not capable of doing Pore Radius calculations, the area
below the above mentioned box will be empty. If your gas is capable of doing
Pore Radius calculations, you are asked to select which form you wish to enter the
values in: Radius Values or P/P0 Values. You must select which type of value
you will be entering before you start entering value. Once you start entering
values you may not switch to the other style.
The next message box informs you of what type of values the program is
looking for (i.e. pressure targets, P/P0 Values, etc.) and the range of valid points.
Again, the exact contents of this field depend upon the gas you are using and/or
the type of value you have selected to enter.
The actual values stored in the list are not the actual radius values, but the
P/P0 values that, when put through a complex formula, become the radius values.
This formula can not be inverted, so this program uses an approximating
algorithm for the conversion. It is done this way so that the two programs that
need this information do not have to go through this conversion process every
time they need radius values. The process does not take much time for an
individual point, and since each P/P0 is calculated when you enter a radius value,
the time used is minimal. You may notice that there is a slight delay after
entering each radius point before the next prompt appears. This conversion
process varies for each gas. A file created for tests using one gas can not be used
for tests using the other gas. If the gas you are using can not generate pore radius
values, then you will have to enter the values as direct pressure values (in PSIA).
As mentioned above, there is a list box just to the right of the Enter key
which keeps track of all the currently entered values.
The three other command buttons located on the right side of the form and
their meanings are as follows:
DONE - ends the isotherm file and saves it to the previously selected
file name. This will happen automatically if you enter a value
that is equal to the upper limit for entered values.
HELP - calls the BETWIN Help system and displays information on
the New Isotherm List section.
CANCEL - ends the Isotherm List File creation process without
saving the list to disk.
The list box mentioned above lists the commands in the order they have
been entered so that you may keep track of what commands you have already
entered.
Finally, there are two command buttons that do not enter commands into
the pretreatment file. They are:
CANCEL - ends the Pretreatment File Creation process and returns
you to the main screen without saving the pretreatment file.
Before any gas can be used with the system, either to run a test, create a
data file manually, or enter any form of parameter list, it must first be described
through this section. This form is also used for the option Edit Old Gas from the
Modify Menu, except that from the modify option you are not able to change the
name of the gas or the one letter extension assigned to it.
The following two questions are asked only if you are creating a new gas:
Enter the single letter or number to be used to signify this gas -
Each gas has a single letter or number that acts as its "handle".
This is usually the first letter of the name of the gas, though it can
be anything that is not already being used by another gas. This
handle is only used internally to designate different types of files.
(For example, isotherm list files all have the three letter extension
beginning IS, but the third letter is the handle of the gas for which
they were meant to be used.) You can see these handles being
used if you look at the three letter extensions on the files listed in
the automated test description list. You select each file by its name
(up to 8 letters), but when the full file name is displayed on the
screen, it also has the disk drive designation (and sub-directory, if
any), plus the extension, which usually contains the gas handle as
the last letter.
If the "handle" is already in use you will be informed of this later
and given the opportunity to change it.
The next four questions deal with the gas and what type of sample it will be use
with:
Enter the Name of this Gas to be used in reports -
This name can be as long as you like, and contain virtually any
characters. This name will be used when printing the report. Only
the file name and the handle need to be unique. You can use the
same official name for more than one gas without causing any
problem with the program, though it may cause confusion when
comparing reports.
Enter the density of the gas at STP -
Simply enter the density at STP of the gas you are creating a file
for.
Enter the molecular weight of the gas -
Enter the molecular weight of the gas you are entering.
This gas is to be used with what type of sample -
This will determine the type of calibration used in sample
chambers. If you select Cooled, then this gas is assumed to be
used with liquid nitrogen. If you select Heated, then this gas is
assumed to be used with a heater.
The next two questions control other questions so they have been separated into
their own frames.
Can this gas be used to calculate Surface Area?
If the gas can be used to calculate surface area select Yes. If the
answer is Yes, you will need to answer four additional questions.
The three questions within the Surface Area Frame are:
Enter the cross-sectional area of one molecule (Sq. Ang)
-
Enter the cross-sectional area, in square angstroms,
of one molecule of the gas you are entering/editing.
Enter the lower range of P/P0 for multi-point &
Enter the upper range of P/P0 for multi-point -
There are also three command buttons on the right side of the form. These
buttons are:
If the single character you selected to designate this gas is already in use,
you will be informed of this fact and asked if you wish to replace the current gas.
If you select YES the old gas file is replace with the new information. If you
select NO you will be returned to the Gas Parameter File Editor/Creation form
so that you may change the designating character.
If you choose to replace an old designator, you will be asked if you wish
to erase the old Isotherm lists, Multipoint lists, etc. created under the old gas.
Simply check off those file types you wish to erase and click on the Continue
button. This will erase all existing files of the type you selected which use the
designator of the old gas. If this new gas file is for a totally new gas, you will
want to erase all the old isotherm lists, as they will no longer be valid. If the new
file is basically the same gas, with just something small (like the name) changed,
then the isotherm lists will still be valid.
Using this option, a user may define a new unit of measure for pressure
readings. This unit of measure is what BETWIN uses to display pressure readings.
BETWIN comes with three previously created units of pressure. They are PSIA,
Torr, and KiloPascals.
The calculations that BETWIN performs, dealing with pressure, use the
pressure unit Torr. The pressure readings that are stored in a data file are in Torr.
The conversion factor is used when the pressure reading is to be displayed. Using
the currently defined pressure units, you get the following equivalencies: 760
Torr = 14.7 PSIA = 101 kiloPascals.
To define a new unit of pressure you need to enter a name, of exactly 4
characters in length without spaces, by which BETWIN will use as the pressure
label (i.e. Torr, Atms, k_Pa, etc.). This name is entered in the Unit Name field.
The Unit Conversion field is for the conversion value. Examples of
conversion values are displayed in the box at the top of the form. This is the
value that, when a Torr value is multiplied by this value, you get the equivalent
value in the unit you are creating.
Example: If you wish to measure pressure in Atmospheres, type
Atms (or similar name) into the Unit Name field. The Unit
Conversion field would contain the value attained from the
following formula:
Atmospheres 1 Atms
= = 0.001315789
Torr 760 Torr
There are also three command buttons located at the bottom of the form.
These buttons perform the following commands:
CANCEL - ends the creation process without saving the conversion
information.
About BETWIN
Help
Starts the BETWIN Help System and displays the main topics list. This
system allows you to attain help on a each of the major sections of the BETWIN
Software System as well as the associated programs.