0% found this document useful (0 votes)
657 views74 pages

1720-E Manual

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
657 views74 pages

1720-E Manual

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 74

Catalog Number 60100-18

Hach sc100™ 1720E


Low Range Turbidimeter
Instrument Manual
04/04 3ed
© Hach Company, 2003. All rights reserved. Printed in the U.S.A.
Catalog Number 60100-18

Hach sc100™ 1720E Low Range Turbidimeter


Instrument Manual

04/04 3ed

© Hach Company, 2003, 2004. All rights reserved. Printed in the U.S.A. eac/dp
Table of Contents
Section 1 Specifications ............................................................................................................................................... 4

Section 2 General Information ..................................................................................................................................... 6


2.1 Safety Information ..................................................................................................................................................... 6
2.2 General Product Information ..................................................................................................................................... 6
2.3 Theory of Operation................................................................................................................................................... 7

Section 3 Installation..................................................................................................................................................... 8
3.1 Mechanical Installation .............................................................................................................................................. 8
3.1.1 Controller Dimension Illustrations .................................................................................................................... 9
3.1.2 Using the Optional Sun Shield ....................................................................................................................... 11
3.1.3 Mounting the Controller.................................................................................................................................. 12
3.2 Electrical Installation................................................................................................................................................ 14
3.2.1 Installation in Conduit..................................................................................................................................... 14
3.2.3 Wiring for Power at the Controller .................................................................................................................. 15
3.3 Alarms and Relays .................................................................................................................................................. 18
3.3.1 Connecting the Relays ................................................................................................................................... 18
3.3.2 Connecting the Analog Outputs ..................................................................................................................... 18
3.4 Connecting/Wiring the Sensor Cable ...................................................................................................................... 19
3.5 Connecting the Optional Digital Output ................................................................................................................... 21
3.6 Turbidimeter Installation Information ....................................................................................................................... 24
3.6.1 Mounting the Turbidimeter Body .................................................................................................................... 24
3.6.2 Installing the Head Assembly......................................................................................................................... 24
3.7 Installing a Sample Line .......................................................................................................................................... 24
3.8 Sample Connections ............................................................................................................................................... 25

Section 4 Operation..................................................................................................................................................... 26
4.1 Using the Keypad .................................................................................................................................................... 26
4.2 Controller Display Features ..................................................................................................................................... 27
4.2.1 Important Key Presses................................................................................................................................... 27
4.3 Instrument Setup ..................................................................................................................................................... 28
4.3.1 Software Text Abbreviation Conventions ....................................................................................................... 28
4.3.2 Adjusting Display Contrast ............................................................................................................................. 28
4.3.3 Specifying the Displayed Language............................................................................................................... 29
4.3.4 Setting the Time and Date ............................................................................................................................. 29
4.4 Changing the Sensor Name .................................................................................................................................... 30
4.4.1 Setting Up System Security ........................................................................................................................... 31
4.5 Output Options ........................................................................................................................................................ 32
4.5.1 Output Setup Menu (from System Setup) ...................................................................................................... 32
4.5.2 Hold/Transfer Outputs.................................................................................................................................... 33
4.5.3 Release Outputs ............................................................................................................................................ 33
4.6 Relay Options .......................................................................................................................................................... 34
4.6.1 Relay Setup Menu (from System Setup) ....................................................................................................... 34
4.7 Data and Event Logging Options............................................................................................................................. 36
4.7.1 Data Logging Options .................................................................................................................................... 36
4.8 Digital Network Options ........................................................................................................................................... 36
4.9 Menu Structure ........................................................................................................................................................ 36
4.9.1 Sensor Diagnostics Menu .............................................................................................................................. 36
4.9.2 Sensor Setup Menu ....................................................................................................................................... 37
4.9.3 System Setup Menu....................................................................................................................................... 37
4.9.4 Test/Maint Menu ............................................................................................................................................ 38

2
Table of Contents

Section 5 System Startup ........................................................................................................................................... 40


5.1 General Operation ................................................................................................................................................... 40
5.2 Starting Sample Flow .............................................................................................................................................. 40
5.3 Sensor Calibration and Verification ......................................................................................................................... 40
5.3.1 User-prepared Calibration.............................................................................................................................. 41
5.3.2 Calibration with StablCal® ...............................................................................................................................................................42
5.4 Instrument Verification............................................................................................................................................. 43
5.4.1 Dry Verification............................................................................................................................................... 44
5.4.2 Wet Verification .............................................................................................................................................. 45
5.5 Calibration and Verification History ......................................................................................................................... 46

Section 6 Maintenance................................................................................................................................................ 47
6.1 Maintenance Schedule ............................................................................................................................................ 47
6.2 Removing a Sensor from the System...................................................................................................................... 47
6.3 Reinstalling a Sensor on the System....................................................................................................................... 47
6.4 Cleaning .................................................................................................................................................................. 48
6.4.1 Cleaning the Controller .................................................................................................................................. 48
6.4.2 Cleaning the Photocell Window ..................................................................................................................... 48
6.4.3 Cleaning the Turbidimeter Body and Bubble Trap ......................................................................................... 48
6.4.4 Replacing the Lamp Assembly ...................................................................................................................... 49
6.5 Fuse Replacement .................................................................................................................................................. 51

Section 7 Troubleshooting.......................................................................................................................................... 52
7.1 Error Codes ............................................................................................................................................................. 52
7.2 Warnings ................................................................................................................................................................. 52
7.3 Event Codes ............................................................................................................................................................ 54

Section 8 Replacement Parts and Accessories........................................................................................................ 55

Section 9 How to Order ............................................................................................................................................... 57


9.1 Information Required ............................................................................................................................................... 57
9.2 International Customers .......................................................................................................................................... 57
9.3 Technical and Customer Service (U.S.A. only) ....................................................................................................... 57

Section 10 Repair Service........................................................................................................................................... 58

Section 11 Limited Warranty ...................................................................................................................................... 59

Section 10 Compliance Information .......................................................................................................................... 60

Appendix A ModBUS Register Information .............................................................................................................. 62

Index ............................................................................................................................................................................. 70

3
Section 1 Specifications

Specifications are subject to change without notice.

Table 1 Turbidimeter Specifications


Nephelometric light scatter at 90 degrees relative to the incident light beam. The incident light
Method of Detection beam is composed of a tungsten filament light source with a color temperature between 2200 and
3000K.
Range 0.001–100 nephelometric turbidity units (NTU)

Accuracy ± 2% of reading or ± 0.015 NTU (whichever is greater) from 0 to 40 NTU;


± 5% of reading from 40 to 100 NTU
Better than 1% 0–40 NTU on formazin. Allows for accurate calibration at high turbidity values.
Linearity
Temperature dependent ±2 °C.
Resolution (Displayed) 0.0001 NTU up to 9.9999 NTU; 0.001 NTU from 10.000 to 99.999 NTU

Repeatability Better than ±1.0% of reading or ±0.002 NTU, whichever is greater for each range.
For a full-scale step change, initial response in 1 minute, 15 seconds. Varies with flow rate,
see the table below. The response time is also dependent on the signal averaging time, which is
user selectable.

Flow Rate
% Step Change
Response Time 750 500 250
10 1¼ minutes 1½ minutes 2½ minutes
50 2 minutes 2½ minutes 6 minutes
90 3½ minutes 3½ minutes 9 minutes
99 4 minutes 5 minutes 12 minutes

Sample Flow Required 200 to 750 mL/minute (3.2 to 11.9 gal/hour)

Sensor Storage
–20 to 60 °C (–4 to 140 °F); 95% relative humidity, non-condensing.
Temperature

Operating Temperature 0 to 50 °C (32–122 °F) for single sensor system, 0 to 40 °C (32–104 °F) for two sensor system

Sample Temperature Range 0 to 50 °C (32–122 °F)

Operating Humidity 5 to 95% non-condensing

Power Requirements 12 V dc ±5%, 12.5 watts maximum

Sample Inlet Fitting ¼-inch NPT female. ¼-inch compression fitting (supplied)

Drain Fitting ½-inch NPT female, ½-inch hose barb (supplied)

Signal Average (Filter) Time no averaging, 6, 30, 60, and 90 seconds, user selectable. Default is 30 seconds.

Sensor Dimensions Turbidimeter body and cap: 25.4 x 30.5 x 40.6 cm (10 x 12 x 16 inches)

Sensor Cable Length 2 m (6.6 ft); Optional 7.62 m (25 ft) extension cable. Maximum cable length is 9.62 m (31.6 ft).
Cable: 105 °C, 300 V, PVC jacket
Sensor Cable Rating
Wires: 22 AWG, PVC jacket
Mounting Options Wall; floor stand

Shipping Weight 1720E Turbidimeter and Controller: 6.31 kg (13.5 lb); 1720E Turbidimeter only: 4.71 kg (10 lb)

4
Specifications

Table 1 Turbidimeter Specifications (continued)


1. StablCal® (stabilized formazin) – primary or wet calibration of the instrument
2. Formazin – user-prepared primary or wet calibration of the instrument
Calibration Methods
3. Multi-sensor calibration – Performed with a specialized calibration procedure for up to eight
sensors on a single set of fresh StablCal® standards.
1. StablCal® (stabilized formazin) – recommended for verification in the appropriate application
range of measurement. For regulatory verification, standards of 0.1. 0.3. 0.5 and 1.0 NTU
Verification (Wet) Method are available.
2. Formazin – fresh user-prepared standard
1. ICE-PIC™ Verification Module with factory-set values of 20.0 or 1.0 ±25%. Unique value is
Verification (Dry) Method assigned when dry verification is done immediately after calibration and is used as pass/fail
criteria for subsequent verifications.
1. Mandatory before calibration
Recommended Cleaning
2. Optional before verification
Intervals
3. Mandatory upon verification failure
Languages English (default), German, Spanish, Nederlands

Installation Environment Indoor

Primary Compliance Method USEPA 180.1; Hach Method 8195; ASTM D 6698; Standard Methods 2130B

Limit of Detection 0.0032 NTU (according to criteria specified by ISO 15839)

Table 2 Controller Specifications


Microprocessor-controlled measuring unit with measured value display,
Component Description
temperature display (for some parameters), and menu-driven system
–20 to 60 °C (–4 to 140 °F); 95% relative humidity, non-condensing with
Controller Operating Temperature
sensor load <7 W; –20 to 40 °C (–4 to 104 °F) with sensor load <25 W
Controller Storage Temperature –20 to 70 °C (–4 to 158 °F); 95% relative humidity, non-condensing

Enclosure NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish


100–230 V ac ±10%, 50/60 Hz;
Power Requirements
Power: 11 W with 7 W sensor load, 35 W with 25 W sensor load
Pollution Degree/Installation Category II; II
Two (Analog outputs, each selectable for 0–20mA or 4–20 mA), maximum
Outputs impedance 500 ohm. Output span programmable over any portion of the 0–100
NTU range. Optional digital network connection1. Infrared Data Acquisition (IrDA).
Three SPDT, user-configurable contacts rated 100–230 V ac, 5 Amp
Relays
resistive maximum.
Controller Dimensions ½ DIN—144 x 144 x 150 mm (5.7 x 5.7 x 5.9 inches)

Controller Weight 1.6 kg (3.5 lb)

1. See Replacement Parts and Accessories on page 55.

5
Section 2 General Information

2.1 Safety Information


Please read this entire manual before unpacking, setting up, or operating this
equipment. Pay attention to all danger and caution statements. Failure to do so
could result in serious injury to the operator or damage to the equipment.

To ensure that the protection provided by this equipment is not impaired, do not
use or install this equipment in any manner other than that specified in
this manual.

Use of Hazard Information


DANGER: Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.

CAUTION: Indicates a potentially hazardous situation that may result in minor or


moderate injury.

Note: Information that requires special emphasis.

Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or damage to
the instrument could occur if not observed.

This symbol, if noted on the instrument, references the instruction manual for operation
and/or safety information.

This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or
electrocution exists.

This symbol, if noted on the product, indicates the need for protective eye wear.

This symbol, when noted on the product, identifies the location of the connection for Protective Earth
(ground).

2.2 General Product Information


The controller enclosure is NEMA4X/IP66-rated and has a corrosion-resistant
finish designed to withstand corrosive environmental constituents such as salt
spray and hydrogen sulfide. The controller display shows the current turbidity
reading if connected to a single sensor, or two readings when two sensors
are connected.

The 1720E Turbidimeter is a continuous-reading nephelometric turbidimeter


designed for low-range turbidity monitoring. This process turbidimeter is capable
of measuring turbidity from 0.001 to 100.0 NTU. Calibration is based on formazin,
the primary turbidity reference standard adopted by the APHA Standard Methods
for the Examination of Water and Wastewater and the U.S. Environmental
Protection Agency (EPA) and on StablCal® which is also recognized as a primary
standard.

6
General Information

2.3 Theory of Operation


The 1720E Turbidimeter measures turbidity by directing a strong beam of
collimated light from the sensor head assembly down into the sample in the
turbidimeter body. Light scattered at 90° relative to the center line of incident light
by suspended particles in the sample is detected by the submerged photocell
(see Figure 1).

Figure 1 90 Degree Detector

The amount of light scattered is proportional to the turbidity of the sample. If the
turbidity of the sample is negligible, little light will be scattered and detected by the
photocell and the turbidity reading will be low. High turbidity, on the other hand, will
cause a high level of light scattering and result in a high reading.

Sample enters the turbidimeter body and flows through the baffle network of the
bubble trap. The flow allows bubbles to either cling to surfaces of the baffle system
or rise to the surface and vent to atmosphere. After traveling through the bubble
trap, sample enters the center column of the turbidimeter body, rises into the
measuring chamber and spills over the weir into the drain port. A reading is taken
once per second.

7
Section 3 Installation

DANGER
Only qualified personnel should conduct the installation tasks described in
this section of the manual. The 1720E/sc100 product configuration is not
intended for installation in hazardous locations.

Figure 2 Controller Mounting Components

1 2 5
4
6
sc100 3
7

10

1. Controller 6. Lock washer, ¼-inch I.D. (4), Cat. No. 8H1336


2. Mounting foot for panel mounting (2), 7. Flat washer, ¼-inch I.D. (4), Cat. No. 8H1346
Cat. No. 1000B4F3222
3. Bracket for panel and pipe mounting, 8. Pan head screws (4), M6 x 1.0 x 20 mm, Cat. No. 58674-00
Cat. No. 1000C4F3217-101
4. Gasket for panel mounting, rubber, 9. Pan head screws (4), M6 x 1.0 x 100 mm, Cat. No. 5867500
Cat. No. 1000A4F3249-101
5. Hex nut, M6 (4), Cat. No. 5867300 10. Pan head screws (4), M6 x 1.0 x 150 mm, Cat. No. 5867600

Table 3 Customer-supplied Items


Item
14-AWG wire for electrical power connections in conduit or if allowed by local electrical codes, 115 or 230 V ac power cord plus a
NEMA 4X-rated strain relief
High-quality, shielded instrumentation cable for connecting the analog outputs plus a NEMA 4X-rated strain relief
Mounting hardware for the sensor
Sun shield for mounting configurations where the sun strikes the front of the display (available from the manufacturer, order
separately). See Figure 8 on page 11.
Common hand tools

3.1 Mechanical Installation


Install in an environment that is protected from corrosive fluids. The sensor is
adversely affected by ClO2. Install the sensor in an area well ventilated from any
corrosive liquids or gasses.

8
Installation

3.1.1 Controller Dimension Illustrations

Figure 3 Controller Dimensions

144.0 mm (5.67 inches) 150.0 mm (5.91 inches)

sc100

144.0 mm (5.67 inches)

Figure 4 Controller Mounting Dimensions

144.02 mm (5.67 inches)


72.01 mm
(2.84 inches)

72.01 mm
(2.84 inches) 40.14 mm
(1.58 inches)

80.27 mm
(3.16 inches)
144.02 mm
(5.67 inches)

40.14 mm (1.58 inches) M6 x 1.0

80.27 mm
(3.16 inches)

9
Installation

Figure 5 Panel Mount Cut-out Dimensions


144 mm (5.67 inches)
(ref only)
75 mm (2.955 inches)
(ref only)

4 mm 135 m
inches) (5.31 in
only)

72 mm
(2.835 inches) 67.4 mm
(ref only) (2.65 inches)

66.67 mm (2.625 inches)

133 mm (5.25 inches)

Figure 6 Conduit Hole Dimensions

144 mm (5.67 inches)


75.07 mm (2.955 inches) 68.96 mm (2.715 inches)

28.57 mm (1.125 inches) 28.57 mm (1.125 inches)

150 mm
(5.91 inches)
99.31 mm (ref only)
(3.91 inches)

127 mm
50.8 mm (5.00 inches)
(2.00 inches)

25.4 mm (1.00 inch) 6.35 mm (0.25 inch)


25.4 mm (1.00 inch) 15.24 mm (0.60 inch)

10
Installation

3.1.2 Using the Optional Sun Shield


The optional sun shield was designed to increase the readability of the display
by screening it from direct sunlight. See Replacement Parts and Accessories on
page 55 for ordering information.

Figure 7 Sun Shield Kit Components

1 5

3 6

1. Sun shield 5. Pipe mounting brackets (2), includes items 6 and 7,


Cat. No. 9H1079
2. Pan head screws, M6 x 1.0 x 12 mm (6), 6. Hex/slotted head screw, 5/16-inch x 1.0-inch (supplied with item
Cat. No. 200-1025 number 5)
3. Lock washers, ¼-inch I.D. (2), Cat. No. 8H1336 7. Square nut, 5/16-inch (supplied with item number 5)
4. Hex nuts, M6 x 1.0 (2), Cat. No. 5867300 8. Uni-strut, 27 cm (10.5 inch) length, Cat. No. 276F1227

Figure 8 Mounting the Controller in the Sun Shield

1 5

2
2

1. Sun shield 4. Hole pattern for mounting controller 7. Slide mounting brackets into the
2. Uni-strut (rotate 90° as required) 5. Pipe (vertical or horizontal as required) uni-strut as shown. Place the
mounting brackets around the
3. Pan head screw, lock washer (2 each) 6. Hex/slotted head screw and square nut pipe and fasten the hardware.

11
Installation

3.1.3 Mounting the Controller


Attach the controller to a rail or wall or mount it in a panel. Supplied mounting
hardware is shown in Figure 9, Figure 10, and Figure 11.

Figure 9 Vertical or Horizontal Pipe Mounting the Controller

1 2

4
5
6

1. Controller 4. Flat washer, ¼-inch I.D. (4), Cat. No. 8H1346


2. Pipe (vertical or horizontal) 5. Hex nut, M6 (4), Cat. No. 5867300
3. Bracket, pipe mounting, Cat. No. 1000C4F3217-101 6. Pan head screw, M6 x 1.0 x 100 mm (4), Cat. No. 5867400

Figure 10 Wall Mounting the Controller

3 4

1. Controller 4. Pan head screw, M6 x 1.0 x 20 mm (4), Cat. No. 5867400


2. Bracket, Cat. No. 1000C4F3217-101 5. Customer-supplied hardware for wall mounting
3. Lock washer, ¼-inch I.D., Cat. No. 8H1336

12
Installation

Figure 11 Panel Mounting the Controller

1 2 3

6 7

9
8

11 10

1. Controller 7. Lock washer, ¼-inch I.D., (4) Cat. No. 8H1336


2. Gasket, rubber, panel mount, Cat. No. 1000A4F3249-101 8. Hex nut (4), Cat. No. 5867300
3. Panel (maximum thickness is 9.5 mm (3/8 inch)) 9. Flat washer (4), Cat. No. 8H1346
4. Mounting Foot (2), Cat. No. 1000B4F3222 10. Pan head screw, M6 x 1.0 x 150 mm (4), Cat. No. 5867600
5. Mounting bracket, controller, Cat. No. 1000C4F3217-101 11. It may be necessary to remove the sensor connectors. See
6. Pan head screw (4), Cat. No. 5867400 procedure below.

To remove the sensor connectors before inserting the controller enclosure into the
panel cut-out:

1. Disconnect power to the controller.

1. Disconnect the wires at terminal block J5, see Figure 21 on page 21.

2. Loosen and remove the nut securing the sensor connector inside the
enclosure. Remove the sensor connector and wires. Repeat step 1 and 2 for
the other sensor connector.

3. After the controller is in place in the panel, reinstall the sensor connectors and
reconnect the wiring to terminal J5 as shown in Figure 21 on page 21.

13
Installation

3.2 Electrical Installation


DANGER High-voltage wiring for the controller is conducted behind the high voltage
The instrument must be installed barrier in the controller enclosure. The barrier must remain in place unless a
by qualified technical personnel qualified installation technician is installing wiring for power, alarms, or relays.
for adherence to all applicable See Figure 12 for barrier removal information.
electrical codes. The 1720E/sc100
product configuration is not
intended for installation in
hazardous locations.

3.2.1 Installation in Conduit


In hard-wired electrical applications, the power and safety ground service drops
for the instrument must be 18 to 12 AWG. See Figure 13 on page 15 for strain
relief and conduit opening sealing plug information. See section 3.2.3 on page 15
for wiring information.

3.2.2 Installation Using a Power Cord


DANGER Where permitted by local electrical codes, a sealing-type strain relief to
Use of power cords is not maintain the NEMA 4X/IP66 environmental rating and a power cord less than
permitted in hazardous locations. 3 meters (10 feet) in length with three 18-gauge conductors (including a safety
ground wire) can be used, see Replacement Parts and Accessories on page 55.
See Figure 13 on page 15 for strain relief and conduit opening sealing plug
assembly. See section 3.2.3 on page 15 for wiring information.

Figure 12 Removing Voltage Barrier

J1

S1 J3
J2 1
1 + DATA
PROBES

2 – DATA
U5 3 SERVICE REQUEST
4 +V
J4
F1

F2

5 GND
6
ANALOG OUTPUTS

U9
J5 1 + OUT 2
2 – OUT 2
3 SHIELD/CHASSIS GND
4 + OUT 1

NETWORK
PCB
CONNECTOR
5 – OUT 1 2
NC COM NO NC COM NO NC COM NO
INTERFACE RELAY A RELAY B RELAY C
CARD
DANGER - EXPLOSION HAZARD
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS. FIELD WIRING
INSULATION MUST
DANGER - RISQUE D'EXPLOSION BE RATED TO
J6 NE PAS DEBRANCHER TANT QUE LE EST SOUS
80° C MINIMUM
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX

1. High voltage barrier 2. Unsnap the barrier latch then pull out to remove the barrier.

14
Installation

Figure 13 Using the Optional Strain Relief and Conduit Plug

2 3

1. Power cord strain relief 2. Conduit strain relief 3. Conduit opening sealing plug

Figure 14 Proper Wire Preparation and Insertion

1
2

1. Strip ¼-inch of insulation. 2. Seat insulation against connector with no bare wire exposed.

3.2.3 Wiring for Power at the Controller


Wire the instrument for line power by hard-wiring in conduit or by wiring to a power
cord if local code allows. Regardless of the type of wire used, the connections are
made at the same terminal. A local disconnect designed to meet local electrical
code is required and must be identified for all types of installation. See Figure 16
and Figure 17 on page 17 for suggested local disconnect configurations.

1. Obtain appropriate fittings with NEMA 4X/IP66 environmental rating.

2. Loosen the screws using a phillips-head screwdriver and open the hinged
controller cover.

3. Remove the high-voltage barrier (see Figure 12 on page 14).

4. Insert the wires through the strain relief fitting or conduit hub located in the
right-rear access hole in the bottom of the enclosure. Tighten the strain relief if
used, to secure the cord.

5. Properly prepare each wire (Figure 14) and insert each wire into the terminal
according to Table 4. Tug gently after each insertion to ensure the connection
is secure.

6. Seal any unused openings in the controller box with conduit opening sealing
plugs, see Replacement Parts and Accessories on page 55.

7. Reinstall the high-voltage barrier and latch to secure.

15
Installation

Table 4 Power Wiring Information


Terminal Number Terminal Description Wire Color Code for North America Wire Color Code for Europe
1 Hot (L1) Black Brown
2 Neutral (N) White Blue
3 Protective Earth (PE) Green Green w/yellow tracer

Figure 15 Wiring Connections

1 2 3 4 5

J1

S1 J3
J2
14

13 U5

J4 6

F1

F2
U9
J5

NETWORK 7
12 INTERFACE
NC COM NO
RELAY A
NC COM NO
RELAY B
NC COM NO
RELAY C
CARD

J6

11 10 9 8

1. J1—Network connector 8. Sensor connector


2. J2—Header for optional network interface card 9. Sensor connector
3. J5—Relay A connector 10. J6—Analog output (4–20 mA) connector
4. J6—Relay B connector 11. J5—Sensor connector for hard-wiring
5. J7—Relay C connector 12. Position for network interface card
6. Fuses (F1, F2) 13. Service port
7. J8—ac Power connections 14. Sensor terminator selector/service port configuration

16
Installation

Figure 16 Local Disconnect for Power Cord

J1

S1 J3
J2
1

U5

J4

F1

F2
U9
J5

2
NETWORK
NC COM NO NC COM NO NC COM NO
INTERFACE RELAY A RELAY B RELAY C
CARD

J6

1. Power terminal 2. Power cord strain relief

Figure 17 Local Disconnect for Hard-wired Line Power

J1

S1 J3
J2
1

U5

J4
F1

F2

U9
J5

2
NETWORK
NC COM NO NC COM NO NC COM NO
INTERFACE RELAY A RELAY B RELAY C
CARD

J6

1. Power terminal 2. Conduit strain relief

17
Installation

3.3 Alarms and Relays


The controller is equipped with three unpowered relays rated 100–230 V ac,
50/60 Hz, 5 amp resistive maximum.

3.3.1 Connecting the Relays


The relay connector accepts 18–12 AWG wire (as determined by load
application). Wire gauge less than 18 AWG is not recommended.

Danger: Relay loads must be The controller contains three relays designed for use with high voltage
resistive. User must externally (greater than 30V-RMS and 42.2V-PEAK or 60 V dc). Refer to Figure 18 for
limit current to the relays to connection information. The wiring is not designed for low voltage connections.
5 Amps by use of a fuse or Relay must not be powered from the same wiring used to power the controller.
breaker. See section 4.6 on page 34 for relay setup details.

Danger: ac power terminals are The Normally Open (NO) and Common (COM) relay contacts will be connected
designed for single wires. Do not when an alarm or other condition is active. The Normally Closed (NC) and
use more than one wire in each Common relay contacts will be connected when an alarm or other condition
terminal. is inactive or when power is removed from the controller.

Figure 18 Alarm and Relay Connections

NC COM NO

J1

S1 J3
J5 J6 J7
J2

1 2 3
U5

RELAY A RELAY B RELAY C


F1

F2

J4
U9
J5

NETWORK
NC COM NO NC COM NO NC COM NO
INTERFACE RELAY A RELAY B RELAY C
CARD

J6

Disconnect
Power

3.3.2 Connecting the Analog Outputs


Two isolated analog outputs (1 and 2) are provided, see Figure 19. Each output
can be set to 0–20 or 4–20 mA, and can be assigned to represent the measured
parameter or secondary measurement such as temperature. Make connections
with twisted-pair shielded wire and connect the shield at the controlled component
end or at the control loop end. Do not connect the shield at both ends of the cable.
Use of non-shielded cable may result in radio frequency emission or susceptibility
levels higher than allowed. Maximum loop resistance is 500 ohm. Refer to
section 4.5 on page 32 for output software setup.

18
Installation

Make wiring connections at the analyzer end as shown in Figure 19.

Table 5 Output Connections at Terminal Block J6


Recorder Wires Circuit Board Position
Output 2 + 1
Output 2 – 2
Shield 3
Output 1 + 4
Output 1 – 5

Figure 19 Analog Output Connections

J1

S1 J3
J2

1 + DATA

PROBES
2 – DATA
U5 3 SERVICE REQUEST
4 +V

F1

F2
J4 5
6
GND

ANALOG OUTPUTS
U9
J5 1 + OUT 2
2 – OUT 2
3 SHIELD/CHASSIS GND
4 + OUT 1
PCB 5 – OUT 1
CONNECTOR
NETWORK
NC COM NO NC COM NO NC COM NO
INTERFACE RELAY A RELAY B RELAY C
CARD

DANGER - EXPLOSION HAZARD


DO NOT DISCONNECT WHILE CIRCUIT IS LIVE FIELD WIRING
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
INSULATION MUST
DANGER - RISQUE D'EXPLOSION BE RATED TO
J6 NE PAS DEBRANCHER TANT QUE LE EST SOUS 80° C MINIMUM
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX

1
2
3
4
5

J6

3.4 Connecting/Wiring the Sensor Cable


The sensor cable is supplied with a keyed quick-connect fitting for easy
attachment to the controller, see Figure 20. Retain the connector cap to seal the
connector opening in case the sensor must be removed.

The 1720E sensor cable may be extended by a maximum of 7.62 m (25 ft),
see Replacement Parts and Accessories on page 55.

Modify the controller for sensor hard-wiring as follows:

1. Remove all power to the controller.

2. Open the controller cover.

19
Installation

3. Disconnect and remove the existing wires between the quick connect and
terminal block J5, see Figure 21 on page 21.

4. Remove the quick connect fitting and wires and install the threaded plug on
the opening to maintain the environmental rating.

Table 6 Wiring the Sensor at Terminal Block J5


Terminal Number Terminal Designation Wire Color
1 Data (+) Blue
2 Data (–) White
3 Service Request No Connection
4 +12 V dc Brown
5 Circuit Common Black
6 Shield Shield (grey wire in existing quick disconnect fitting)

Figure 20 Attaching the Sensor using the Quick-connect Fitting

J1

S1 J3
J2

1 + DATA
PROBES

2 – DATA
U5 3 SERVICE REQUEST
4 +V
F1

F2
J4 5
6
GND
ANALOG OUTPUTS

U9
J5 1 + OUT 2
2 – OUT 2
3 SHIELD/CHASSIS GND
4 + OUT 1
PCB 5 – OUT 1
CONNECTOR
NETWORK
NC COM NO NC COM NO NC COM NO
INTERFACE RELAY A RELAY B RELAY C
CARD

DANGER - EXPLOSION HAZARD


DO NOT DISCONNECT WHILE CIRCUIT IS LIVE FIELD WIRING
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
INSULATION MUST
DANGER - RISQUE D'EXPLOSION BE RATED TO
J6 NE PAS DEBRANCHER TANT QUE LE EST SOUS 80° C MINIMUM
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX

5. Cut the connector from the sensor cable.

6. Reinstall the plug on the sensor access opening to maintain the


environmental rating.

7. Strip the insulation on the cable back 1-inch. Strip ¼-inch of each individual
wire end.

8. Wire as shown in Table 6.

20
Installation

9. Pass the cable through conduit and a conduit hub or a strain relief fitting and
an available access hole in the controller enclosure. Tighten the fitting.

10. Close and secure the cover.

Figure 21 Hard-wiring the Sensor

J1

S1 J3
J2

1 + DATA

PROBES
2 – DATA
U5 3 SERVICE REQUEST
4 +V
J4 5 GND

F1

F2
6

ANALOG OUTPUTS
U9
J5 1 + OUT 2
2 – OUT 2
3 SHIELD/CHASSIS GND
4 + OUT 1
PCB 5 – OUT 1
CONNECTOR
NETWORK
NC COM NO NC COM NO NC COM NO
INTERFACE RELAY 1 RELAY 2 RELAY 3
CARD

DANGER - EXPLOSION HAZARD


DO NOT DISCONNECT WHILE CIRCUIT IS LIVE FIELD WIRING
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
INSULATION MUST
DANGER - RISQUE D'EXPLOSION BE RATED TO
J6 NE PAS DEBRANCHER TANT QUE LE EST SOUS 80° C MINIMUM
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX

Disconnect
Power

From Probe

3.5 Connecting the Optional Digital Output


At this time, the manufacturer supports Modbus RS485 and Modbus RS232
communication protocols. The optional digital output card is installed in the
location indicated in Figure 22 on page 22. Terminal block J1 provides user
connection to the optional network card. The terminal connection is based on the
selected network card. Refer to the instructions supplied with the network card for
more details.

Table 3-1 Network Connections at Terminal Block J1


Terminal N
AquaTrend Network Modbus RS485 2-wire Modbus RS485 4-wire Modbus RS2321 Profibus
umber
1 Network A D+ RD + Rx A–
2 Network B D– RD – — B1 +
3 Network A — TD + Tx A2 –
4 Network B — TD – — B2 +
5 NC Common Common Common Common
6 NC No connection No connection No connection No connection
7 Shield Shield Shield Shield Shield

1. See Figure 22

21
Installation

Figure 22 RS232 Connection to Customer-supplied Computer 9-pin D Subminiature Connector

Common 5
9
Not used 4
8
Rx 3
7
Tx 2
6
Not used 1

Figure 23 Network Card Position in the Controller

J1

S1 J3
J2

U5

F1

F2
J4
U9
J5

NETWORK
NC COM NO NC COM NO NC COM NO
INTERFACE RELAY A RELAY B RELAY C
CARD

J6

2 3 4

1. J2—Network card header 2. Mounting hole (3) 3. Network card placement 4. J1 Terminal

22
Figure 24
Allow at least 254 mm (10.00 inches) above the instrument
for removal of the head assembly.

312.48 mm (12.30 inches)

238.35 mm (9.38 inches)


273.05 mm (10.75 inches)
127.8 mm
19.72 mm 157.00 mm
(5.0 inches)
(0.78 inches) (6.18 inches)
1720E Dimensions

Cable
2 meters (6.6 feet)
POWER STATUS

2 Slots for
0.25² fastener
Sample Inlet
¼-NPT to ¼-Compression fitting

Sample Drain
½-NPT to ½-ID Tubing hose Barb

343.67 mm (13.53 inches)

255.3 mm (11.05 inches)

217.2 mm (8.55 inches)

143.2 mm
40.6 mm (5.64 inches)
(1.60 inches) 72.04 mm (2.84 inches)

107.95 mm (4.25 inches)


164.2 mm (6.46 inches)
172.4 mm (6.79 inches)
193.7 mm (7.625 inches)
Service Drain
1.00 I.D. 195.6 mm (7.70 inches)

23
Installation
Installation

3.6 Turbidimeter Installation Information


The turbidimeter body is designed for wall-mounting (although it may be mounted
on the optional floor stand). The turbidimeter sensor must be mounted within
six feet of the controller unless an extension cable is used. Maximum cable length
is 9.6 m (31 feet).

3.6.1 Mounting the Turbidimeter Body


Locate the turbidimeter as close to the sampling point as possible. A shorter
distance for the sample to travel results in a faster response time.

Clean the turbidimeter body and bubble trap before installation using the
instructions supplied in section 6.4.3, Cleaning the Turbidimeter Body and Bubble
Trap on page 48. Slotted mounting brackets are integral parts of the turbidimeter
body. Install customer-supplied hardware appropriate for the installation
environment using the criteria detailed below:

• Install in a location that is isolated from vibration.

• Allow at least 22 cm (approximately 10 inches) clearance for removal of the


head assembly and bubble trap cover from the top of the turbidimeter body.

• Leave enough room below the turbidimeter body to remove the bottom plug
and to place a container under the drain when calibrating or cleaning.

Note: Make sure the top of the • Install two ¼-20 bolts 10-3/4 inches apart (on center). Leave at least ¼-inch of
turbidimeter body is level. the bolt head exposed.

• Make sure the bolts are installed level.

Slide the slotted mounting brackets of the turbidimeter body onto the bolts.

3.6.2 Installing the Head Assembly


After the turbidimeter body has been mounted, install the bubble trap cover, then
place the head assembly on the turbidimeter body with the label facing the front.
Move the head assembly back and forth slightly to ensure it is properly seated on
the body of the instrument. Failure to properly seat the head will result in light
leakage and erroneous readings.

The rear portion of the head assembly has a molded “lip” which may be used to
hang the head assembly on the turbidimeter body edge for routine maintenance.

3.7 Installing a Sample Line


DANGER One-fourth inch O.D. rigid or semi-rigid tubing is recommended for sample lines.
This turbidimeter is not designed Run them as directly as possible between the turbidimeter body and the sampling
for use in hazardous locations or point to minimize sample flow lag time.
with samples that are flammable
or explosive in nature. If any Install sample line taps into larger process pipes to minimize interference from
sample solution other than water
air bubbles or pipeline bottom sediment. A tap projecting into the center of the
is used in this product, test the
pipe is ideal. Figure 25 shows examples of sample tap installations.
sample/product compatibility to
ensure user safety and proper
product performance. Note: When setting the flow rate, take care to avoid sweeping air “micro-bubbles” through
the internal bubble trap. Observe the sample flow inside the turbidimeter body. If small
air bubbles can be seen flowing up through the center, reduce the flow rate.

24
Installation

Figure 25 Sampling Techniques


Air (Typical)

Sediment (Typical)

Poor Poor Good Best

3.8 Sample Connections


Sample inlet and drain connections are made on the turbidimeter body. The
sample inlet fitting installed in the body is a ¼-inch NPT x ¼-inch compression
fitting. One additional fitting supplied with the instrument is a ½-inch NPT-to-hose
fitting for use with ½-inch ID flexible plastic tubing on the drain.

Note: For samples with high solids The required flow rate is 200 to 750 mL/minute (4.0 to 11.9 gal/hour). Flow rate
content (high turbidity), operate at into the turbidimeter may be controlled with a flow restriction device on the inlet
the highest flow rate possible. For line. Flow rates below 200 mL/min will reduce response time and cause
samples with low solids content (low inaccurate readings. Flow rates above 750 mL/min will cause the turbidimeter to
expected turbidity), operate at a low overflow, indicating the flow rate is too high.
flow rate (200–300 mL/min).

Figure 26 Sample Connections

POWER STATUS

1. Sample Inlet, ¼-28 NPT x ¼-inch Compression fitting 2. Drain, ½-inch NPT fitting 3. Service Drain

25
Section 4 Operation

4.1 Using the Keypad


The front of the controller is shown in Figure 27. The keypad consists of the eight
keys described in Table 4.

Figure 27 Front of the Controller

1
5
sc100

2 6

3 7

1. Instrument Display 4. Right, Left, Up, and Down keys 7. Enter key
2. Back key 5. IrDA (Infrared Data Acquisition) window
3. Menu key 6. Home key

Table 4 Controller Key Functions/Features


Number Key Function

2 Move back one level in the menu structure.

Move to the main menu from other menus. This key is not active in menus where a selection or other
3
input must be made.

4 Navigate through the menus, change settings, and increment and decrement digits.

Move to the Main Measurement screen from any other screen. This key is not active in menus where a
5
selection or other input must be made.

6 Accept an input value, updates, or accepts displayed menu options.

26
Operation

4.2 Controller Display Features


When the controller is in measurement mode, measurements for each connected
sensor are displayed.

The display will flash on startup, when the hold outputs function has been
activated, and when the filter function (signal average) is changed to a different
value.

An active system warning will cause the warning icon (a triangle with an
exclamation point inside) to be displayed on the right side of the display.

Figure 28 Display

1
SENSOR NAME 4

2
0.023 NTU
5

6
3 OUTPUT1: 11.23 mA

1. Status bar. Indicates the sensor name and status of relays. The relay
4. Energized relay indicator
letter is displayed when the relay is energized.
2. Main measurement 5. Warning icon area
3. Secondary measurement/output information 6. Measurement units

4.2.1 Important Key Presses

• Press the HOME key then the RIGHT or LEFT key to display two readings
when two sensors are connected. Continue to press the RIGHT or LEFT key
to toggle through the available display options as shown below.

RTC:MM/DD/YY SENSOR NAME: SENSOR NAME: MAIN MEASURE SENSOR NAME SENSOR NAME
02 SENSOR NAME: Turbidity: Turbidity:

24:00:00 0.023 ppm NTU 0.031 NTU


0.023
SENSOR NAME:
0.031
NTU

NTU
TEMP:
0.023

22.9°C
NTU
TEMP:
0.031

22.9°C
NTU

OUTPUT1: 11.23 mA OUTPUT1: 11.23 mA OUTPUT1: 11.23 mA

• Press the UP and DOWN keys to toggle the status bar at the bottom of the
measurement display to display the output information.

• In Menu mode, an arrow may appear on the right side of the display to
indicate that more items are available. Press the UP or DOWN key
(corresponding to the arrow direction) to display additional menus.

MAIN MENU SYSTEM SETUP SYSTEM SETUP SYSTEM SETUP


SENSOR DIAG OUTPUT SETUP DISPLAY SETUP SECURITY SETUP
SENSOR SETUP RELAY SETUP SECURITY SETUP LOG SETUP
SYSTEM SETUP NETWORK SETUP LOG SETUP CALCULATION
TEST/MAINT DISPLAY SETUP CALCULATION ERROR HOLD MODE

27
Operation

4.3 Instrument Setup

4.3.1 Software Text Abbreviation Conventions

Abbreviation Meaning Abbreviation Meaning


Adj Adjust P/F Pass/Fail
Cal Calibration Pass Password
Cont. Continue Preped Prepared
Cyl Cylinder SN Serial Number
Dflt Default Std Standard
Diag Diagnostic Temp Temperature
Int Internal Ver Verification
Meas. Measurement Xfer Transfer

4.3.2 Adjusting Display Contrast


Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SYSTEM SETUP

3 DISPLAY SETUP

4 — ADJ CONTRAST

5 (+0–50)

6 MAIN MENU or Main Measurement Screen —

28
Operation

4.3.3 Specifying the Displayed Language


Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SYSTEM SETUP

3 DISPLAY SETUP

4 LANGUAGE

5 select language

6 Main Menu or Main Measurement Screen —

4.3.4 Setting the Time and Date

4.3.4.1 Setting the Time

Note: The time is available only in 24-hour (military) format.

Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SYSTEM SETUP

3 DISPLAY SETUP

4 SET DATE/TIME

5 highlight TIME

select character to edit


6
choose appropriate number

7 Main Menu or Main Measurement Screen —

29
Operation

4.3.4.2 Setting the Date Format and Date


Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SYSTEM SETUP

3 DISPLAY SETUP

4 SET DATE/TIME

5 — highlight DATE FORMAT

6 choose appropriate date format

7 Highlight DATE

select character to edit —


8
choose appropriate number

9 Main Menu or Main Measurement Screen —

4.4 Changing the Sensor Name


Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SENSOR SETUP

3 highlight sensor of interest if more than one sensor is attached

4 CONFIGURE

5 EDIT NAME

select character to edit —


6
choose appropriate alpha/numeric digit

7 Main Menu or Main Measurement Screen —

30
Operation

4.4.1 Setting Up System Security


The sc100 has a passcode feature to restrict unauthorized access to Network
Setup, Security Setup, Log Setup, and Test/Maint menus. In addition, the
passcode also regulates function selection for relay options. The passcode is
factory set to sc100 (the five digits must be followed by a space to remove the
trailing asterisk). The passcode may be changed, see section 4.4.1.1.

The following two options are available:

Disabled: All configuration settings and calibrations can be changed. This is the
default setting.

Enabled: All configuration settings can be displayed but not changed. Network
Setup, Security Setup, Log Setup, and Test/Maint menus cannot be accessed
without the passcode.

Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SYSTEM SETUP

3 SECURITY SETUP

4 — SET PASSCODE

5 highlight ENABLED

6 Main Menu or Main Measurement Screen —

4.4.1.1 Editing the Passcode


If the passcode is enabled, it may be edited. The passcode can consist of up to six
digits (alpha and/or numeric and available characters). If a passcode is forgotten,
obtain the Master passcode from the Technical Consulting Services Department,
see Technical and Customer Service (U.S.A. only) on page 57.

Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SYSTEM SETUP

3 highlight SECURITY SETUP

ENTER (current or default) PASSCODE —


4
EDIT PASSCODE

31
Operation

Step Select Menu Level/Instructions Confirm

5 Change the existing passcode

6 Main Menu or Main Measurement Screen —

4.5 Output Options


The controller provides two isolated analog outputs (Output 1 and Output 2).
Customize the outputs using the table in section 4.5.1 on page 32.

4.5.1 Output Setup Menu (from System Setup)


1. Select OUTPUT 1 or 2
SELECT SOURCE
Press ENTER to access a list of all connected sensors. Choose the sensor to associate with the output.
SET PARAMETER

Highlight the appropriate displayed parameter and press ENTER.

SET FUNCTION
Select LINEAR CONTROL for current output to track the measurement value. Select PID CONTROL for the sc100 to operate as
a PID controller.
SET TRANSFER
Each analog output is normally active, responding to the measured value of its assigned parameter. However, during calibration,
each output can be transferred to this preset transfer value. Default: 4mA; Range: 0–20
SET FILTER

Allows the user to average the analog outputs over time Default: 0; Range: 0–120 seconds

SCALE 0 mA/4 mA

Select 0 mA or 4 mA for minimum current (outputs will be set to 0–20 mA or 4–20 mA).

ACTIVATION
FUNCTION set to LINEAR CONTROL
If LINEAR CONTROL was selected in SET FUNCTION, set the low and the high values for the current output here.
Defaults: Low = 0; High = 100; Low Value Range: 0–100, High Value Range: 0–100.
FUNCTION set to PID CONTROL
If PID CONTROL was selected in SET FUNCTION, configure the PID Control as follows:
1. Set MODE: AUTO or MANUAL. Manual output default: 100%
2. Set PHASE: DIRECT or REVERSE controller operation.
3. SET SETPOINT: enter the set point the PID control will control the process to. Default: 100; Range: 0–100
4. PROP BAND: control the proportional band for the PID control. Default: 5.00; Range: 0–1000
5. INTEGRAL: control the integral action time period in minutes. Range: 0–999
6. DERIVATIVE: control the settings for the rate control. Range: 0–999

32
Operation

4.5.2 Hold/Transfer Outputs


When cleaning or servicing the instrument, the analog outputs can be held at the
last measured values. To hold the output until released:

Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 TEST/MAINT

Enter Passcode if enabled

3 HOLD OUTPUTS

4 — SET OUTMODE

5 Select HOLD OUTPUTS or XFER OUTPUTS

6 SET CHANNELS

7 Select ALL or 1720E

8 ACTIVATION

9 — LAUNCH

reading will
10 Main Menu or Main Measurement Screen
flash

During calibration, the analog outputs can remain active, be held, or be


transferred to a preset mA value. When output hold or transfer is enabled during a
calibration, the hold or transfer is automatically released when calibration is
completed. See section 5.3, Sensor Calibration and Verification on page 40.

4.5.3 Release Outputs


Step Select Menu Level Confirm

1 MAIN MENU —

2 TEST/MAINT

3 HOLD OUTPUTS

4 ACTIVATION

33
Operation

Step Select Menu Level Confirm

5 — RELEASE

6 Main Menu or Main Measurement Screen —

4.6 Relay Options


Step Select Menu Level Confirm

1 MAIN MENU —

2 SYSTEM SETUP

3 RELAY SETUP

4 — Customize the options using the information in section 4.6.1 —

4.6.1 Relay Setup Menu (from System Setup)


1. Select Relay A, B, or C
SELECT SOURCE
Choose from the available options (none, installed sensors, real time clock (RTC)).
SET PARAMETER
Choose from the available options.
SET FUNCTION
Source set to sensor
Alarm: Operates relays in response to the measured parameter. Contains separate High and Low Alarm points, deadbands, and
ON/OFF delay. Defaults: Low = 0.000 NTU, high = 100.00 NTU, low deadband = 5.000 NTU, high deadband = 5.000 NTU, on/off
delays default to zero seconds; Range: 0–999 sec.
Feeder Control: Operates in response to the measured parameter. Can be set for phasing, set point, deadband, overfeed timer,
and ON/OFF delay.
Event Control: Controls a cleaning system (or equivalent) on a timed basis.
Warning: Activated when the analyzer detects a sensor warning.
Source set to RTC
Timer: Sets the timer for a cleaning system (or equivalent). Controls the output hold, interval, duration and off delay.
SET TRANSFER
Normally, each control or alarm relay is active, responding to the measured value of its assigned parameter. During calibration,
however, the relay can be transferred to a preset on/off state to suit the application requirements. Select Energize or De-energize
and press ENTER.

34
Operation

4.6.1 Relay Setup Menu (from System Setup) (continued)


ACTIVATION
Function set to ALARM
Sets the value where the relay will turn on in response to decreasing measured value. For example: if the
LOW ALARM low alarm is set for 1.0 and the measured value drops to 0.9, the relay will be activated.
Range: 0.00–100 NTU
Sets the value where the relay will turn on in response to increasing measured value. For example: if the
HIGH ALARM high alarm is set for 4.0 and the measured value increases to 4.2, the relay will be activated.
Range: 0.00–100 NTU
Sets the range where the relay remains on after the measured value increases above the low alarm value.
LOW DEADBAND Default is 20% of the range. For example: if the low alarm is set for 1.0 and the low deadband is set for 0.5,
then the relay remains on between 1.5 and 1.0. Range: 0.00–100 NTU
Sets the range where the relay remains on after the measured value decreases below the high alarm value.
HIGH DEADBAND For example: if the high alarm is set for 4.0 and the high deadband is set for 0.5, then the relay remains on
between 3.5 and 4.0. Range: 0.00–100 NTU
OFF DELAY Sets a time to delay the relay from normally turning off. Off Delay Range: 0–999 seconds
ON DELAY Sets a time to delay the relay from normally turning on. On Delay Range: 0–999 seconds
Function set to FEEDER CONTROL
“High” phase assigns the relay setpoint to respond to an increasing measured value; conversely, a “Low”
PHASE
phase assigns the relay setpoint to respond to a decreasing measured value.
SET SETPOINT Sets the value where the relay will turn on. Default: 100 NTU; Range: 0–100
Sets the range where the relay remains on after the measured value decreases below the setpoint value
DEADBAND (high phase relay) or increases above the setpoint value (low phase relay). Default: 5 NTU;
Range: 0–100 NTU
OVERFEED
Sets the time to limit how long the relay can remain “on.” Default: 5 minutes; Range: 0–999 minutes
TIMER
OFF DELAY Sets a time to delay the relay from normally turning off. Default: 0 seconds; Range: 0–999 seconds
ON DELAY Sets a time to delay the relay from normally turning on. Default: 0 seconds; Range: 0–999 seconds
Function set to EVENT CONTROL
“High” phase assigns the relay setpoint to respond to increasing measured value; conversely, a “Low” phase
PHASE
assigns the relay setpoint to respond to decreasing measured value.
SET SETPOINT Sets the value where the relay will turn on. Default: 100; Range: 0–100
Sets the range where the relay remains on after the measured value decreases below the setpoint value
DEADBAND (high phase relay) or increases above the setpoint value (low phase relay). Default: 5 NTU;
Range: 0–100 NTU,
OnMax TIMER Sets the time to limit the time the relay can remain “on.” Default: 0 seconds; Range: 0–999 seconds
OffMax TIMER Sets a time to delay the relay from normally turning off. Default: 0 seconds; Range: 0–999 seconds
OnMin TIMER Sets the time to limit the time the relay can remain “on.” Default: 0 seconds; Range: 0–999 seconds
OffMin TIMER Sets the time to limit the time the relay can remain “off.” Default: 0 seconds; Range: 0–999 seconds
Function set to TIMER CONTROL (RTC selected in SELECT SOURCE)
HOLD OUTPUTS Set OUTMODE to select output hold operation and select the channels that cause the outputs to be held.
INTERVAL Set the off time for the relay. Default: 5 minutes; Range: 0–999 minutes
DURATION Set the on time for the relay. Default: 30 seconds; Range: 0–999 seconds
Set the time for additional hold/output time after the relay has been turned off. Default: 1 second;
OFF DELAY
Range: 0–999 seconds
Function set to WARNING CONTROL
Set the warning level that will trigger a relay. Range: 0–32 (warnings assigned to 1-9 for 1720E). For
example: Set the warning level to 0 to allow all warnings to trigger the relay; set the warning level to 5 to
WARNING LEVEL
allow warnings 6 and above to trigger the relay. Set the warning level to 9 or greater to not trigger the relay
on any warning. See Table 6 on page 52 for a full list of warnings.

35
Operation

4.7 Data and Event Logging Options


The sc100 provides two data logs (one for each sensor) and two event logs (one
for each sensor). The data logs store the measurement data at selected intervals.
The event log stores a variety of events that occur on the devices such as
configuration changes, alarms, and warning conditions. The data logs are stored
in a packed binary format and the event logs are stored in a CSV format. The logs
can be downloaded through either the digital network port or the IrDA port using
the file transfer program available from the manufacturer.

The default datalogging frequency is 15 minutes. If the datalogging frequency is


set to 15 minute intervals, the instrument can continue to store data for
approximately six months.

4.7.1 Data Logging Options

Sensor Data Log:


Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SENSOR SETUP

3 Highlight sensor of interest if more that one sensor is attached.

4 CONFIGURE

5 DATALOG INTERVAL
(select from 30 sec., 1 min., 5 min., 10 min., 15 min.)

6 Main Menu or Main Measurement Screen —

4.8 Digital Network Options


The sc100 provides two digital communication methods with the controller
(the digital network port and the IrDA port). Either of the digital ports can be used
to access setup data, measurement data, or data/event logs. For the features
available for each individual digital network port, refer to the instruction sheet
supplied with the selected network card.

4.9 Menu Structure


4.9.1 Sensor Diagnostics Menu
SELECT SENSOR
ERROR LIST Displays a list of errors that are present. See section 7.1 on page 52.
WARNING LIST Displays a list of warnings that are present. See section 7.2 on page 52.

36
Operation

4.9.2 Sensor Setup Menu


CALIBRATE
SELECT SENSOR (if more than one sensor is attached)
USER PREPD CAL Calibration using 4000 NTU stock solution diluted to 20.00 NTU formazin.
STABLCAL CAL Calibration using 20 NTU StablCal Stabilized Formazin Standard
VERIFICATION Perform a verification, set the pass/fail criteria, and view the verification history.
0 ELECTRONICS Zero electronics
SET DFLT GAIN Return instrument to default calibration.
View the last 12 entered calibrations. Press the ENTER key to move to the next history entry.
CAL HISTORY
See section 5.5 on page 46 for more information.
CONFIGURE
BUBBLE REJECT Choose Yes or No to enable/disable bubble reject. Default: Yes
Choose no averaging or specify the amount of time for signal averaging. Available options are: no
SIGNAL AVG
averaging, 6 sec., 30 sec., 60 sec., or 90 sec. Default is 30 seconds.
Select the appropriate measurement units to display. Choose from mg/L, NTU, TE/F, and FTU.
MEAS UNITS
Default: NTU
Enter up to a 12-digit name in any combination of symbols and alpha or numeric characters. Press
EDIT NAME ENTER when the entry is complete. The name will be displayed on the status line above the
measurement value on the main display.
SET RESOLUTION Set the number of significant digits to display. Default is three significant digits.
Choose the amount of time between saving data points to the data log. Default: 15 min.; Options: 30
DATALOG INTRVL
sec., 1 minute, 5 minutes, 10 minutes, or 15 minutes.
DIAG/TEST
SOFTWARE VERS. Displays the software version number.
DRIVER VERS Displays the software driver version number.
SERIAL NUMBER Displays the serial number of the sensor.
INT TEMP Displays the internal temperature of the sensor electronics in °C.
DEFAULT SETUP Restores the sensor’s factory default settings and invalidates the current calibration.
POWER CHECK Displays the electrical statistics for the sensor.
CAL VALUE Displays the gain and zero electronics values for the current calibration.

4.9.3 System Setup Menu


OUTPUT SETUP (see section 4.5 on page 32 for expanded menu information)
SELECT OUTPUT 1 or 2
SELECT SOURCE Press ENTER to access a list of all connected sensors and select the sensor that will drive the output.
SET PARAMETER Press ENTER to select from the displayed parameters.
Select LINEAR CONTROL for current output to track the measurement valve. Select PID CONTROL for
SET FUNCTION
the sc100 to operate as a PID controller.
Each analog output is normally active, responding to the measured value of its assigned parameter.
SET TRANSFER
However, during calibration, each output can be transferred to this preset transfer value.
SET FILTER Average measurements over time (0–120 seconds). Default: 0 seconds.
SCALE 0 mA/4 mA Select 0 mA or 4 mA for minimum current (outputs will be set to 0–20 mA or 4–20 mA).
ACTIVATION Dependent on Function selected previously. See section 4.5 on page 32 for additional information.

37
Operation

4.9.3 System Setup Menu (continued)


RELAY SETUP (See section 4.6 on page 34 for expanded menu information.)
SELECT RELAY A, B, or C
SELECT SOURCE Select from none, any connected sensor, or the real time clock (RTC).
SET PARAMETER Press ENTER to select from the displayed parameters.
Select from the available options to customize the relay functions. See section 4.6.1 on page 34 for
SET FUNCTION
additional details.
SET TRANSFER Sets the relay to Energize or De-energize (user-selectable).
ACTIVATION Activate the relays from this menu (dependent on Function selected).
NETWORK SETUP (this menu appears only if a network card is installed in the controller)
MODBUS Highlight sc100 Analyzer, or either connected sensor then press ENTER to select. Choose a number
ADDRESS between 1 and 247 as the address (each source must have a different address) then press ENTER.
BAUD RATE Select a baud rate of 9600, 19200, 38.4K, 57.6K, or 115.2K. Default: 19200
STOP BITS Select 1 or 2 stop bits. Default: 1
MODBUS MODE Select RTU or ASCII. Default: RTU
DATA ORDER Select NORMAL or SWAPPED.
DISPLAY SETUP
Use the UP and DOWN keys to increase or decrease the contrast, see section 4.3.2 on page 28.
ADJ CONTRAST
Range = 0–50
The default is English. Choose from the available options to allow all menus to appear in the selected
LANGUAGE
language.
Use this menu to select the date format and to set the date and time (24-hour (military) format), see
SET DATE/TIME
section 4.3.4 on page 29.
SECURITY SETUP (Enter a 6-digit passcode)
SET PASSCODE
ENABLE Enables system security. See section 4.4.1 on page 31.
DISABLE Disables system security. See section 4.4.1 on page 31.
LOG SETUP (Not used for 1720E system. Enable datalogging from the sensor setup menu for 1720E)
DATALOG SETUP Set up datalogging of data and events. See section 4.7.1 on page 36.
ERROR HOLD MODE
HOLD OUTPUTS Holds outputs when unable to communicate with the sensor.
XFER OUTPUTS Goes to transfer state when unable to communicate with the sensor.

4.9.4 Test/Maint Menu


STATUS
Indicates the status of each relay and indicates which sensors are connected to the controller.
OUTPUT CAL
SELECT OUTPUT 1 or 2
Calibrate Analog Output by specifying values to correspond to 4 mA and 20 mA.
HOLD OUTPUTS
SET OUTMODE Choose Hold Outputs or Xfer Outputs.
SET CHANNELS Choose any individual attached sensor or all attached sensors to be held or transferred.
ACTIVATION Select Launch or Release.

38
Operation

4.9.4 Test/Maint Menu (continued)


OVERFEED RESET
Reset the overfeed time out.
TEST OUTPUT
SELECT OUTPUT 1 or 2
User selectable mA value. 0–20 mA
TEST RELAY
SELECT RELAY A, B, or C
Energize or de-energize the selected relay.
RESET CONFIG
Reset to default configuration of the controller
SIMULATION
SELECT SOURCE, SET PARAMETER, SET SIM VALUE
Simulate sensor measurement values for testing the outputs and relays.
SCAN SENSORS
Manually scans for sensors to determine if sensors have been added or removed.
MODBUS STATS
Indicates the communication statistics for use with an external network.
CODE VERSION
Indicates the controller software version.

39
Section 5 System Startup

5.1 General Operation


Plug the sensor into the unpowered controller by aligning the orientation tab on
the cable connector with the channel in the controller connector. Push in and turn
to secure the connection. Tug gently to check the connection.

After all plumbing and electrical connections have been completed and checked,
place the head on the body and supply power to the system. Ensure the head is
seated on the body when power is applied, since dark readings are measured at
this time. If power is applied while the sensor head is off the turbidimeter body,
cycle the power with the sensor head on the body.

The first time a controller is powered up, a language selection menu will appear.
The user must select the correct language from the displayed options.
Use the UP and DOWN keys to highlight the appropriate language and press
ENTER to select.

Following language selection and upon power-up, the controller will search for
connected sensors. The display will show the main measurement screen. Press
the MENU key to access the menus.

5.2 Starting Sample Flow


Start sample flow through the instrument by opening the sample supply valve.
Allow the turbidimeter to run long enough for the tubing and body to become
completely wetted and the reading on the display to stabilize. One to two hours or
longer may be required initially for complete stabilization. Allow measurements to
become stable through adequate conditioning before completing instrument
settings or performing calibrations.

5.3 Sensor Calibration and Verification


The manufacturer offers two EPA-approved calibration methods one using user-
prepared formazin and the other using StablCal® stabilized formazin. Two
verification methods (wet and dry) are also offered.

The 1720E Turbidimeter is factory-calibrated using StablCal® Stabilized Formazin


before shipment. The instrument must be recalibrated before use to meet
published accuracy specifications. In addition, recalibration is recommended after
any significant maintenance or repair and at least once every three months during
normal operation. The turbidimeter body and bubble trap must be thoroughly
cleaned and rinsed before initial use and prior to each calibration.

Tips to achieve the most accurate calibrations:

• Optimum performance is achieved when calibration is performed in the


turbidimeter body. Accurately prepare the standard then add it to the
turbidimeter body at the appropriate step in the procedure. Do not prepare the
standard in the body.

• Stop sample flow, drain, and clean the turbidimeter body before beginning the
calibration procedure.

• Always clean the photocell window per the instructions in section 6.4.2 on
page 48. Rinse the photocell with deionized water and dry with a soft, lint-free
cloth before calibrating.

40
System Startup

• Always clean the turbidimeter body or calibration cylinder per the instructions
in section 6.4.3 on page 48. Rinse with deionized water before calibrating.

• Store the calibration cylinder upside-down to minimize contamination between


calibrations.

• Pour the calibration standard into the turbidimeter body at the inflow end (left
side when facing the instrument).

• Gently invert StablCal standards for 1 minute before opening. Do not shake.
This ensures a consistent turbidity of the standard.

• If the 20.0 NTU StablCal standard is allowed to sit in the calibration cylinder or
turbidimeter body for more than 15 minutes, it must be remixed (gently swirled
in the calibration cylinder) before use to ensure a consistent turbidity.

• Discard all standards after use per the instructions on the container.
Never transfer the standard back into its original container. Contamination
will result.

• Always recalibrate after restoring default settings.

5.3.1 User-prepared Calibration


Before starting the calibration, read and apply the tips in section 5.3.

Follow the procedure as written (using 1 L of deionized water and 5.0 mL of


4000 NTU formazin) if using a calibration cylinder for calibration.

If using the turbidimeter body for the user-prepared calibration follow the
procedure below using 20 NTU formazin in step 6c. Prepare the standard as
follows:

1. Stop the sample flow, then drain and clean the body.

2. Prepare a 20 NTU standard by adding 5.0 mL of 4000 NTU formazin to a


1-L flask. Dilute to the mark with deionized water and invert gently to mix.

3. Drain the deionized water and pour the prepared 20 NTU standard into the
turbidimeter body at step 6c. Do not add additional 4000 NTU formazin.

Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SENSOR SETUP

3 SELECT SENSOR (if more than one sensor is connected)

4 CALIBRATE

5 — USER PREPD CAL

41
System Startup

Step Select Menu Level/Instructions Confirm

OUTPUT MODE
6
Select ACTIVE, HOLD, or TRANSFER

Stop sample flow. Drain body and clean the body and bubble trap.
a
FILL CYL WITH 1 L DI WATER. REPLACE HEAD.

b Measured reading (based on a gain of 1.0) displayed

(Remove head)
c
ADD 5 ML OF 4000 NTU FORMAZIN INTO CAL CYLINDER.

d Measured reading (based on a gain of 1.0) displayed

GOOD CAL!
e GAIN: X.XX
ENTER TO CONT
(to store)

Verify CAL?
f to verify
(see Note below)

exit no verify

Select VERIFICATION type (begin at step 7 in section 5.4.1 on page 44 or


7
section 5.4.2 on page 45) or enter initials to complete calibration.

8 — RETURN SENSOR TO MEASURE MODE

9 Main Menu or Main Measurement Screen —

Note: If a dry verification is performed directly after a calibration, the measured value is
assigned as the expected value for future verifications (when using the dry verification
device with the same serial number). As long as the verification exists within the verification
history, the expected value will be retained. Otherwise, the expected value will be the
nominal labeled value associated with the dry verification device.

5.3.2 Calibration with StablCal®


Before starting the calibration, read and apply the tips in section 5.3 on page 40.

Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SENSOR SETUP

42
System Startup

Step Select Menu Level/Instructions Confirm

3 — CALIBRATE

4 STABLCAL CAL

OUTPUT MODE
5
Select ACTIVE, HOLD, or TRANSFER

Drain/clean/rinse the turbidimeter body or cal cylinder.


6 a
POUR 20 NTU STD INTO CYL/BODY. REPLACE HEAD

b Measured reading and reading based on 1.0 gain displayed

GOOD CAL!
c GAIN: X.XX
ENTER TO CONT
(to store)

d Verify CAL? (see Note below) to verify/

exit no verify

Select VERIFICATION type (begin at step 7 in section 5.4.1 on page 44 or section 5.4.2
e
on page 45) or enter initials to complete calibration.}

7 — RETURN SENSOR TO MEASURE MODE

8 Main Menu or Main Measurement Screen —

Note: If a dry verification is performed directly after a calibration, the measured value is
assigned as the expected value for future verifications (when using the dry verification
device with the same serial number). As long as the verification exists within the verification
history, the expected value will be retained. Otherwise, the expected value will be the
nominal value associated with the dry verification device.

5.4 Instrument Verification


Instrument verification is intended as a simple check to ensure turbidimeter
functionality between calibrations. A verification is initially performed directly after
a calibration and subsequent independent verifications are referenced to the initial
verification. The pass/fail criteria is set and subsequent verifications are deemed
good or bad, when compared to the initial verification. All verifications are based
on the current calibration and must be repeated when the instrument is
recalibrated or when the pass/fail criteria is not met.

43
System Startup

Two types of verifications are offered. The dry verification is performed using a
“dry” calibration device such as the ICEPIC™. A wet verification is performed
using a standard with a predetermined value such as StablCal® Stabilized
Formazin or a user-prepared standard with a value that has been verified on an
independent device such as a laboratory turbidimeter.

5.4.1 Dry Verification


Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SENSOR SETUP

3 CALIBRATE

4 VERIFICATION

5 — PERFORM VER

OUTPUT MODE
6
Choose ACTIVE, HOLD, or TRANSFER

7 — VER TYPE Select DRY

DRY VERIFY
8
select 1 NTU STD or 20 NTU STD or verify SN of previously used calibration device

Set Head on Standard

Reading Displayed

(to accept)

GOOD VER!

(to store)

9 ENTER INITIALS (user input)

10 — RETURN SENSOR TO MEASURE MODE

11 MAIN MENU or Main Measurement Screen —

44
System Startup

5.4.2 Wet Verification


Before starting the verification, read and apply the appropriate tips in section 5.3
on page 40.

Step Select Menu Level/Instructions Confirm

1 MAIN MENU —

2 SENSOR SETUP

3 — CALIBRATE

4 VERIFICATION

5 — PERFORM VER

OUTPUT MODE
6
Choose ACTIVE, HOLD, TRANSFER

Select VER TYPE


7
Select WET

Enter Std Turbidity

a. DRAIN AND CLEAN SENSOR BODY. ENTER TO CONT

8 b. POUR STANDARD INTO CYL/BODY. PLACE HEAD ON. ENTER TO CONT

c. Reading Displayed

(to accept)

d. GOOD VER!

9 ENTER INITIALS

10 RETURN SENSOR TO MEASURE MODE

11 Main Menu or Main Measurement Screen —

45
System Startup

5.5 Calibration and Verification History


The calibration and verification history logs contain information on the last
12 calibrations and the last 12 verifications. The calibration history log shows the
gain value, the time and date of the calibration, and the initials of the operator
performing verification.

Note: Restoring default settings The calibration history log is accessed from the Calibrate menu. The verification
from the DIAG/TEST menu will history log is accessed from the Verification menu (a submenu of the Calibrate
return the turbidimeter to its non menu).
calibration state (gain = 1.0) but it
will not remove the previous Each verification history entry shows the serial number of the verification device
calibration history from memory.
(dry verification) or the value of the verification standard (wet verification), the time
and date of the verification, and the initials of the operator performing the
verification.

Scroll through the entries by pressing the ENTER key. After scrolling through all
12 histories, the display will return to the calibration menu level.

When the instrument is received from the factory, default values or blank spaces
will be shown for the calibration and verification history information. Those values
will be replaced with real data as the history log is filled.

The data is retained as first in, first out. When the log is full, the newest entry is
stored and the oldest entry in the log is deleted.

46
Section 6 Maintenance

DANGER
Only qualified personnel should conduct the maintenance tasks described
in this section of the manual.

6.1 Maintenance Schedule

Maintenance Task Frequency


Before each calibration and as needed. Depends on sample
Clean the sensor
characteristics.
Calibrate Sensor (as required by regulatory agency) Per agency-dictated schedule.

Scheduled periodic maintenance requirements of the 1720E are minimal and


include calibration and cleaning of the photocell window, bubble trap, and body.
Check and clean the bubble trap and turbidimeter body (as described in
section 6.4.3) if visual inspection shows that it is necessary. Perform other
maintenance on a regular basis; experience will dictate scheduling and may
depend on the installation, sample type, and season.

It is very important to maintain the cleanliness of the interior and exterior of the
turbidimeter body, head assembly, the integral bubble trap, and the surrounding
area. Doing so will ensure accurate, low-level turbidity measurements.

Clean the body before calibration and verification (especially when measurements
are being made at 1.0 NTU or lower).

Check and/or perform a calibration periodically (as experience dictates) using one
of the methods described in section 5.3 on page 40. A calibration history menu
option is available under Sensor Setup/Calibrate.

6.2 Removing a Sensor from the System


Prior to physically removing a sensor from the system, record all user defined
settings such as relays, signal averaging, etc. Disconnect the sensor connector at
the controller.

6.3 Reinstalling a Sensor on the System


To return the system to normal operation following a software upgrade or sensor
repair perform the following procedure:

1. Detach all sensors from the sc100 controller.

2. From the main menu, press the down arrow key to highlight TEST/MAINT.
Press ENTER.

3. Use the down arrow key to scroll to SCAN SENSORS and press ENTER.

4. Remove attached sensors by selecting the corresponding serial number or


select “All”.

5. Power down the sc100 then attach the 1720E.

Note: Clean sensors before 6. Supply power to the sc100. The system will initialize automatically.
reinstallation on the system.

47
Maintenance

6.4 Cleaning

6.4.1 Cleaning the Controller


With the enclosure securely closed, wipe the exterior with a damp cloth.

6.4.2 Cleaning the Photocell Window


Occasional cleaning of the photocell window is required. The frequency will
depend on the nature and concentration of dissolved and suspended solids in the
sample. Biological activity is a primary factor in mineral scale deposit on the
window and the amount differs with sample temperature. In general, more growth
will occur in warm temperatures and less in cold.

Note: Take care to not scratch the Inspect the photocell window often to determine cleaning needs. Remove any
photocell window. organic growth or film on the photocell window before standardization or
calibration. Use a cotton swab and isopropyl alcohol or a mild detergent (such as
Liqui-nox®) to remove most sediment and dirt. Mineral scale buildup may require
cleaning with a mild acid applied with a cotton swab followed by a detergent wash.
Do not use abrasive cleaners.

6.4.3 Cleaning the Turbidimeter Body and Bubble Trap


Sediment may collect in the turbidimeter body after extended use. Noise
(fluctuation) in the reading could indicate the need to clean the body and/or bubble
trap. The 1720E bubble trap and bottom plate may be removed to make cleaning
easier. Drain and clean the turbidimeter body before each calibration. Establish a
regular schedule or perform cleaning as determined by visual inspection.

Cleaning the Turbidimeter Body

Note: The turbidimeter body, bubble 1. Turn off sample flow to the turbidimeter body.
trap, and detector must be cleaned
before each calibration. 2. Remove the head assembly and bubble trap cover from the body. Remove the
bubble trap by lifting it vertically. Set it aside to be cleaned separately.

3. Drain the body by removing the plug from the bottom of the body.

4. Replace the drain plug and fill the body to the weir with cleaning solution. This
cleaning solution can consist of dilute chlorine solution (25 mL of household
bleach in 3.78 liters of water) or a laboratory detergent such as Liqui-nox
(1 mL detergent in 1 liter of water).

5. Use a soft brush to clean the inside surfaces of the body.

6. Remove the drain plug again and thoroughly flush the turbidimeter body with
ultra-filtered deionized water. Clean and replace the plug.

Cleaning the Bubble Trap

1. Prepare a cleaning solution (as in step 4 above) in a container large enough


to submerge the entire bubble trap.

2. Using a test tube brush such as Cat. No. 690-00, clean each surface.

3. Rinse the bubble trap thoroughly with ultra-filtered deionized water and
reinstall it in the turbidimeter body.

4. Replace the bubble trap cover and head assembly on the top of the body.

48
Maintenance

5. Restore sample flow to the instrument.

6. Calibrate the instrument using one of the methods in section 5.3 on page 40.

If the above cleaning procedures have been performed and the turbidimeter
readings are still noisy, the bottom plate and gasket may need to be removed and
cleaned. Carefully perform the following procedure to ensure the turbidimeter
body integrity is maintained.

1. Turn off sample flow to the turbidimeter body.

2. Remove the head assembly, bubble trap cover, and bubble trap (by lifting it
vertically) from the body.

3. Drain the body by removing the plug from the bottom of the body.

4. Lift the body off of its mounting screws.

5. With the body turned upside-down, remove the two Phillips-head screws
holding the bottom plate.

6. Lift the bottom plate off the body; set the gasket aside for use in reassembly.

7. Use a soft brush and a dilute cleaning solution (as prepared above) to clean
the bottom plate and inside surfaces of the turbidimeter body. Rinse the entire
body and bottom plate with ultra-filtered deionized water.

8. Reassemble by inserting the gasket into the molded channel in the


bottom plate.

9. Fit the bottom plate onto the turbidimeter body.

10. Reinstall both screws and carefully tighten to 15 inch-lb maximum.

11. Reinstall the turbidimeter onto the wall mounting screws.

12. Replace the bubble trap, bubble trap cover, and head assembly on the top of
the body.

13. Restore sample flow to the instrument.

6.4.4 Replacing the Lamp Assembly


The Lamp Assembly is located on the head assembly. Under normal use, Hach
recommends replacing the lamp once a year to maintain peak performance.
Replacement bulbs have been “burned-in” at the factory and are ready for
installation and use.

To change the lamp, refer to Figure 29 on page 50 and perform the


following steps:

1. Disconnect power to the controller to remove all power to the turbidimeter.

2. Disconnect the lamp leads by unplugging the connector.

3. After the bulb has cooled, remove as follows:

49
Maintenance

a. Wear cotton gloves to protect your hands and to avoid fingerprints on


the bulb.

b. Grasp the bulb.

c. Twist the bulb in a counterclockwise direction, pulling out slightly, until it is


released from the housing.

d. Pull the lamp leads and connector through the hole in the lamp housing.

Do not touch the new bulb with bare hands. Etched glass and reduced lamp life
will result. Wear cotton gloves or grasp the lamp assembly with a tissue to avoid
contamination. If contamination occurs, clean the glass bulb portion with isopropyl
alcohol.

Replace the bulb by reversing the above instructions. The bulb base only fits one
way; align the notch in the metal bulb flange with the hole in the lamp holder.

Figure 29 Lamp Replacement

1
3

1. Lamp Housing 2. Lamp Assembly 3. Photo Detector

50
Maintenance

6.5 Fuse Replacement


The controller contains two mains fuses. Failed fuses are an indication that an
equipment problem could exist. Problem resolution and fuse replacement should
only be performed by qualified service personnel. Refer to Figure 30 and perform
the following steps to replace the fuses:

1. Disconnect power to the controller (including power to relays if powered).

2. Open the hinged controller cover by completely loosening all four captive
screws in the cover.

3. Remove the high voltage barrier; pull out on the lever of the captive fastener
then pull straight out on the barrier. Set the barrier aside for reinstallation.

4. Remove the fuses and install new fuses of the same type and rating
(T, 1.6A, 250 V, slow blow).

5. Reinstall the high voltage barrier.

6. Close the controller cover and hand-tighten the four screws.

7. Reconnect all power to the instrument.

Figure 30 Fuse Replacement

J1 1

S1 J3
J2

U5

J4
F1

F2

U9
J5

NETWORK
NC COM NO NC COM NO NC COM NO
INTERFACE RELAY A RELAY B RELAY C
CARD

J6

1. Fuses F1 and F2, 1.6 A

51
Section 7 Troubleshooting

7.1 Error Codes


When a sensor is experiencing an error condition, the sensor reading on the
measurement screen will flash and all relays and analog outputs associated with
the sensor will be held.

Highlight the Probe Diag menu and press ENTER. Highlight Errors and press
ENTER to determine the case of the error. Errors are defined in Table 5.

Table 5 Error Codes


Displayed Error Definition
A/D Fail Failed A/D converter. Call the Service Department.
The light source has failed. See section 6.4.4 on page 49 for
Laser Fail
instructions for replacing it.
Flash Fail Datalog and event log will not work.
PMT Fail
Low Signal Error

7.2 Warnings
A sensor warning will leave all menus, relays, and outputs functioning normally,
but will cause a warning icon to flash on the right side of the display. Highlight the
Probe Diag menu and press ENTER to determine the cause of the error.
Warnings are defined in Table 6.

A warning may be used to trigger a relay and users can set warning levels to
define the severity of the warning. Errors are defined in Table 6.

Table 6 Warning Codes


Warning
Displayed Warning Definition/Resolution
Number

1 Dark Reading Warning Dark reading detects too much light.


Sensor head internal temperature is higher than specified. Contact the Service
2 Temperature Warning
Department.
Sensor data log is full. No additional data will be logged until sensor log is
3 Data Log Full Warning
downloaded into controller memory.
Sensor data log is full. No additional data will be logged until sensor log is
4 Event Log Full Warning
downloaded into controller memory.
Monitored voltage is outside the range of 4.5 – 5.5 V @ microprocessor input ->
5 5 Volt Warning
698 to 854 counts from the microprocessor ADC.
Monitored voltage is outside the range of 9.08 – 14.3 V @ microprocessor pin ->
6 Voltage in Warning
279 to 435 counts from the microprocessor ADC.
Monitored voltage is outside the range of 3.96 – 4.48 V @ microprocessor pin ->
7 Laser V Warning
614 to 385 counts from the microprocessor ADC.
Monitored voltage is outside the range of 1.67 – 2.75 V @ microprocessor pin ->
8 Laser I Warning
39 to 64 counts from the microprocessor ADC.
Activated when the sensor is not in normal measurement mode (such as when in
9 Output Mode Not Normal
calibration or verification mode).
10 PMT Voltage Warning

11 Reference Voltage Warning

52
Troubleshooting

Table 7 presents sensor warnings displayed in the Event Log, possible causes,
and corrective actions.

Table 7 Troubleshooting
Sensor Warning Possible Cause Corrective Action
Lamp burned out Replace the lamp. See section 6.4.4 on page 49.
Lamp unplugged Restore connection
Bad Lamp +12 V connection loose at controller Restore connection
Dislodged lamp Reinstall lamp
Bad circuit board in turbidimeter head Contact the Customer Service Department.
See Cleaning the Photocell Window on page 48.
Photocell coated/dirty
Contact the Customer Service Department.
Photocell wires disconnected Reconnect wires
Replace photocell
Photocell broken/cracked
Low Signal Contact the Customer Service Department.
Lens coated/dirty Clean the lens using isopropyl alcohol and a cotton swab.
Obstructed light path Remove obstruction
Sample turbidity >100 NTU Switch to a high range turbidimeter
See Bad Lamp causes above See Bad Lamp corrective actions above
See the controller wiring instructions in section 3.2,
Improper wiring at controller
Electrical Installation on page 14.
Turbidimeter head cable shortened to
Bad System Voltage Contact the Customer Service Department.
improper length
Fluctuation in voltage Turn instrument power off and back on.
Bad circuit board in turbidimeter head Contact the Service Department
Fluctuation in voltage Turn instrument power off and back on.
A/D Converter Timeout
Bad circuit board in turbidimeter head Contact the Customer Service Department.
Light Leak—Turbidimeter head not on Make sure the turbidimeter head is on the turbidimeter body
turbidimeter body or calibration cylinder and properly aligned and repower instrument or perform
during Power Up or Zero Electronics ZERO ELECTRONICS in the CALIBRATION MENU.
Light Leak—Turbidimeter head not
High Dark Counts properly aligned on the turbidimeter Make sure the turbidimeter head is properly aligned and
body or calibration cylinder during repower instrument or perform ZERO ELECTRONICS in
Power Up or the CALIBRATION MENU.
Zero Electronics
Photocell broken/cracked Contact the Customer Service Department.

The following table presents additional malfunctions which may not be recorded in
the Event Log.

Table 8 Additional Malfunctions Not Recorded in the Event Log


Symptom Possible Cause Corrective Action
Continuous Underrange Calibration standards not in the Verify the accuracy of calibration standards and recalibrate the
(negative reading) correct order or incorrect dilution instrument. See Low Signal in Table 7.
Continuous Overrange Calibration standards not in the Verify the accuracy of calibration standards and recalibrate the
(100 NTU) correct order or incorrect dilution instrument.

53
Troubleshooting

Table 8 Additional Malfunctions Not Recorded in the Event Log (continued)


Symptom Possible Cause Corrective Action
Erratic Readings Calibration standards have the Verify the accuracy of calibration standards and recalibrate the
same value instrument.
Inadequate bubble removal from Increase the signal averaging time to a longer interval.
sample Make sure the Bubble Reject feature is turned on.
Slow the flow of sample into the instrument.
High Readings Deionized water turbidity is Clean the instrument.
greater than 0.5 NTU Access Calibration History for turbidity value of ultra-filtered water.
Verify the flow is between 200–750 mL/min.
Recalibrate the instrument.

7.3 Event Codes


Event codes are not displayed on the screen of the sc100 and must be
downloaded from the event log to be viewed. See section 4.7 on page 36 for
additional information. Troubleshooting actions are given in section Table 7 on
page 53.

Table 9 Event Log List


Event Event # Data1 Data2 Data3
0 = OFF
Bubble Reject Change 0 — —
1= ON
0 = No avg
1 = 6 sec
Filter Size Change 1 2 = 30 sec — —
3 = 60 sec
4 = 90 sec
0 = 30 sec
1 = 1min
Data Log Interval Change 2 2 = 5 min — —
3 = 10 min
4 = 15 min
Power On 3 1 — —
Calibration 4 Cal Gain Initials —
Verification 5 Expected Value Meas Value Initials
Dark Reading Warning 6 Measured Value — —
Temperature Warning 7 Temperature Value — —
16 = +5V high
32 = +5V low
Voltage Warning 8 — —
64 = V in high
128 = V in low
1 = lamp V high
2 = lamp V low
Lamp Warning 9 — —
4 = lamp I high
8 = lamp I low
A/D Fail 10 1 — —
Lamp Fail 11 2 — —
Flash Fail 12 3 — —
1 = Hold
Output Mode Change 13 — —
2 = Transfer

54
Section 8 Replacement Parts and Accessories

Replacement Items
Item Cat. No.
Fuse, T, 1.6 A, 250 V .......................................................................................................................................... 52083-00
Instruction Manual, sc100 Controller, English .................................................................................................... 58600-18
Instruction Manual, 1720E Turbidimeter System, English .................................................................................. 60100-18
Installation kit, sc100 Controller.......................................................................................................................... 58672-00
Lamp Assembly .................................................................................................................................................. 18950-00
sc100 Controller ................................................................................................................................................. 58600-00

Optional Accessories
Cable, sensor extension, 7.7 m (25 ft)................................................................................................................ 57960-00
Cap, Connector Receptacle ...................................................................................................... each ................ 52100-00
Deionized (demineralized) water .................................................................................................. 1 L .................... 272-56
Digital Output Card for ModBUS RS232 communication ................................................................................... 59200-00
Digital Output Card for ModBUS RS485 communication ................................................................................... 59200-01
Drain plug for the 1720E body ................................................................................................... each ................ 44116-00
Filter, 0.45 µm, to produce ultra-filtered water for cleaning and calibration ............................... each ................ 26705-00
Filter, 0.2 µm, to produce ultra-filtered water for calibration standard preparation .................... each ................ 23238-10
Formazin Calibration Kit includes:
Calibration Cylinder, TenSette® Pipet, 4000 NTU Formazin Primary Standard (500 mL).... each ................ 44156-00
Floor Stand ............................................................................................................................... each ................ 57432-00
Flow meter, 500–700 mL/min .................................................................................................... each ................ 40282-00
ICE-PIC Module for Calibration and Calibration Verification
20 NTU ................................................................................................................................. each ................ 52250-00
1 NTU ................................................................................................................................... each ................ 52215-00
Lid, Bubble Trap, 1720E ............................................................................................................ each ............... 52012-00
LonWorks Card.................................................................................................................................. ................ 59200-02
Photocell Replacement Kit for the 1720E .................................................................................. each ................ 52180-00
Pipet tips for 19700-01 TenSette Pipet ................................................................................... 50/pkg ................ 21856-96
Pipet tips for 19700-10 TenSette Pipet ................................................................................... 50/pkg ................ 21997-96
Plug, conduit opening ......................................................................................................................................... 58687-00
Power Cord with strain relief, 115 V.................................................................................................................... 54488-00
Power Cord with strain relief, 230 V ................................................................................................................... 54489-00
Power Cord Kit, 10A-125V, 1.8 m (6 ft), UL/CSA listed ............................................................. each ................ 46306-00
Power Cord Kit, 10A-230V, 1.8 m (6 ft), European-style plug, VDE approved .......................... each ................ 46308-00
StablCal Calibration Set for the 1720 Series Turbidimeter
Includes: StablCal Standards, < 0.1 NTU, 20.0 NTU........................................................... 1 L/each ................ 26596-00
StablCal Standard, 0.1 NTU ......................................................................................................... 1 L ................ 27233-53
StablCal Standard, 0.3 NTU ......................................................................................................... 1 L ................ 26979-53
StablCal Standard, 0.5 NTU ......................................................................................................... 1 L ................ 26980-53
StablCal Standard, 1.0 NTU ......................................................................................................... 1 L ................ 26598-53
Strain relief, Heyco .................................................................................................................................................. 16664
Sun shield........................................................................................................................................................... 58690-00
Swabs, Cotton, presterilized for cleaning the photodetector ................................................ 100/pkg ................ 25543-00
TenSette Pipet, 0.1 to 1.0 mL .................................................................................................... each ................ 19700-01
TenSette Pipet, 1.0 to 10.0 mL .................................................................................................. each ................ 19700-10
Tubing, Inlet, ¼ inch O.D., Polyethylene ................................................................................per foot ................ 51322-00
Tubing, Outlet, ½ inch I.D., ¾ inch O.D., Tygon R3603..........................................................per foot ................ 51263-00

55
Replacement Parts and Accessories

Calibration and Verification Standards and Accessories


Item Qty Cat. No.
Calibration Cylinder....................................................................................................................each.................44153-00
Formazin Calibration Standards
Formazin, 4000 NTU Stock Solution................................................................................ 500 mL...................2461-49
Calibration/Verification Modules
ICE-PIC™ Module, 1 NTU .................................................................................................1 each.................52215-00
ICE-PIC™ Module, 20 NTU ...............................................................................................1 each.................52250-00
StablCal® Calibration Standards
StablCal® Stabilized Formazin Standard, 1 NTU .................................................................... 1 L.................26598-53
StablCal® Stabilized Formazin Standard, 20 NTU ................................................................. 1 L.................26601-53
StablCal® Stabilized Formazin Standard, <0.1 NTU ............................................................... 1 L.................26597-53
StablCal® Stabilized Formazin Set, four 1-L bottles 20-NTU and four 1-L bottles <0.1 NTU1 L each ...........26596-00
StablCal® Stabilized Formazin Standard, 40 NTU ............................................. 1 gallon (3.78 L).................27463-56
StablCal® Stabilized Formazin Standard, 0.1 NTU ............................................ 1 gallon (3.78 L).................27233-56

56
Section 9 How to Order

U.S.A. Customers
By Telephone:
6:30 a.m. to 5:00 p.m. MST
Monday through Friday
(800) 227-HACH (800-227-4224)

By Fax:
(970) 669-2932

By Mail:
Hach Company
P.O. Box 389
Loveland, Colorado 80539-0389 U.S.A.

Ordering information by e-mail: [email protected]

9.1 Information Required


• Hach account number (if available) • billing address

• Your name and phone number • Shipping address

• Purchase order number • Catalog number

• Brief description or model number • Quantity

9.2 International Customers


Hach maintains a worldwide network of dealers and distributors. To locate the
representative nearest you, send e-mail to [email protected] or contact:

Hach Company World Headquarters; Loveland, Colorado, U.S.A.


Telephone: (970) 669-3050; Fax: (970) 669-2932

9.3 Technical and Customer Service (U.S.A. only)


Hach Technical and Customer Service Department personnel are eager to
answer questions about our products and their use. Specialists in analytical
methods, they are happy to put their talents to work for you.

Call 1-800-227-4224 or e-mail [email protected]

57
Section 10 Repair Service

Authorization must be obtained from Hach Company before sending any items for
repair. Please contact the Hach Service Center serving your location.

In the United States:


Hach Company
Ames Service
100 Dayton Avenue
Ames, Iowa 50010
(800) 227-4224 (U.S.A. only)
FAX: (515) 232-3835

In Canada:
Hach Sales & Service Canada Ltd.
1313 Border Street, Unit 34
Winnipeg, Manitoba
R3H 0X4
(800) 665-7635 (Canada only)
Telephone: (204) 632-5598
FAX: (204) 694-5134
E-mail: [email protected]

In Latin America, the Caribbean, the Far East, the


Indian Subcontinent, Africa, Europe, or the Middle East:
Hach Company World Headquarters,
P.O. Box 389
Loveland, Colorado, 80539-0389 U.S.A.
Telephone: (970) 669-3050
FAX: (970) 669-2932
E-mail: [email protected]

58
Section 11 Limited Warranty

Hach Company warrants its products to the original purchaser against any defects that are due
to faulty material or workmanship for a period of one year on the controller and two years on the
1720E Sensor from date of shipment unless otherwise noted.

In the event that a defect is discovered during the warranty period, Hach Company agrees that, at
its option, it will repair or replace the defective product or refund the purchase price, subject to
the pro-rated schedule above, excluding original shipping and handling charges. Any product
repaired or replaced under this warranty will be warranted only for the remainder of the original
product warranty period.

This warranty does not apply to consumable products such as chemical reagents; or
consumable components of a product, such as, but not limited to, lamps and tubing.

Contact Hach Company or your distributor to initiate warranty support. Products may not be
returned without authorization from Hach Company.

Limitations
This warranty does not cover:

• Damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or
undeclared), terrorism, civil strife or acts of any governmental jurisdiction
• Damage caused by misuse, neglect, accident or improper application or installation

• Damage caused by any repair or attempted repair not authorized by Hach Company

• Any product not used in accordance with the instructions furnished by Hach Company

• Freight charges to return merchandise to Hach Company

• Freight charges on expedited or express shipment of warranted parts or product

• Travel fees associated with on-site warranty repair


This warranty contains the sole express warranty made by Hach Company in connection with its
products. All implied warranties, including without limitation, the warranties of merchantability
and fitness for a particular purpose, are expressly disclaimed.

Some states within the United States do not allow the disclaimer of implied warranties and if this
is true in your state the above limitation may not apply to you. This warranty gives you specific
rights, and you may also have other rights that vary from state to state.

This warranty constitutes the final, complete, and exclusive statement of warranty terms and no
person is authorized to make any other warranties or representations on behalf of Hach
Company.

Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the
exclusive remedies for the breach of this warranty. On the basis of strict liability or under any
other legal theory, in no event shall Hach Company be liable for any incidental or consequential
damages of any kind for breach of warranty or negligence.

59
Section 10 Compliance Information

Hach Co. certifies this instrument was tested thoroughly, inspected and found to
meet its published specifications when it was shipped from the factory.

The Model sc100 with 1720E Sensor has been tested and is certified as
indicated to the following instrumentation standards:

Product Safety
UL 61010A-1 (ETL Listing # 65454)
CSA C22.2 No. 1010.1 (ETLc Certification # 65454)
Certified by Hach Co. to EN 61010-1 Amds. 1 & 2 (IEC1010-1) per
73/23/EEC, supporting test records by Intertek Testing Services.

Immunity
This equipment was tested for Industrial level EMC per:

EN 61326 (EMC Requirements for Electrical Equipment for Measurement,


Control and Laboratory Use) per 89/336/EEC EMC: Supporting test records by
Hach Company, certified compliance by Hach Company.

Standards include:

IEC 1000-4-2:1995 (EN 61000-4-2:1995) Electro-Static Discharge Immunity


(Criteria B)
IEC 1000-4-3:1995 (EN 61000-4-3:1996) Radiated RF Electro-Magnetic Field
Immunity (Criteria A)
IEC 1000-4-4:1995 (EN 61000-4-4:1995) Electrical Fast Transients/Burst
(Criteria B)
IEC 1000-4-5:1995 (EN 61000-4-5:1995) Surge (Criteria B)
IEC 1000-4-6:1996 (EN 61000-4-6:1996) Conducted Disturbances Induced
by RF Fields (Criteria A)
IEC 1000-4-11:1994 (EN 61000-4-11:1994) Voltage Dip/Short Interruptions
(Criteria B)

Additional immunity Standard/s include:

ENV 50204:1996 Radiated Electro-Magnetic Field from Digital Telephones


(Criteria A)

Emissions
This equipment was tested for Radio Frequency Emissions as follows:

Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipment for measurement,


control and laboratory use—EMC requirements) Class “A” emission limits.
Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test
Center (A2LA # 0905-01) and certified compliance by Hach Company.

Standards include:

EN 61000-3-2 Harmonic Disturbances Caused by Electrical Equipment


EN 61000-3-3 Voltage Fluctuation (Flicker) Disturbances Caused by Electrical
Equipment

60
Compliance Information

Additional Emissions Standard/s include:

EN 55011 (CISPR 11) Class “A” emission limits

Canadian Interference-causing Equipment Regulation, IECS-003, Class A


Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test
Center (A2LA # 0905-01) and certified compliance by Hach Company.

This Class A digital apparatus meets all requirements of the Canadian


Interference- Causing Equipment Regulations.

Cet appareil numÈrique de la classe A respecte toutes les exigences du


RËglement sur le matÈriel brouilleur du Canada.

FCC PART 15, Class “A” Limits


Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test
Center (A2LA # 0905-01) and certified compliance by Hach Company.

This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions:

(1) this device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause undesired
operation.

Changes or modifications to this unit not expressly approved by the party


responsible for compliance could void the user's authority to operate the
equipment.

This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with
the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference, in which case the user will be required to correct the interference at
his own expense. The following techniques of reducing the interference problems
are applied easily.

1. Disconnect the Model sc100 Controller from its power source to verify that it is
or is not the source of the interference.

2. If the Model sc100 Controller is connected into the same outlet as the device
with which it is interfering, try another outlet.

3. Move the Model sc100 Controller and 1720E sensor away from the device
receiving the interference.

4. Reposition the device receiving the interference.

5. Try combinations of the above.

61
Appendix A ModBUS Register Information

Table 10 Controller ModBUS Registers


Group Name Tag Name Register # Data Type Length R/W Description
Calculated Value calculated from two sensor
Measurements 40001 Float 2 R
Value measurements
Unsigned
Setup Language 40003 1 R/W Current System Language
Integer
Current Data Display Format
Unsigned
Setup Date Format 40004 1 R/W (0 = DD/MM/YY; 1 = MM/DD/YY;
Integer
2 = DD-MM-YY; 3 = MM-DD-YY)
Error Hold Unsigned Error Mode Hold State (0 = Hold outputs;
Setup 40005 1 R/W
Mode Integer 1 = Transfer outputs to predefined value)
Unsigned Selects data source for this output
Setup/Analog Output 1 Source 40006 1 R/W
Integer (0 = None; 2 = sensor; 4 = Calculation)
Unsigned Selects sensor source when Source =
Setup/Analog Output 1 Sensor Select 40007 1 R/W
Integer Sensor (0 = sensor1; 1 = sensor2)
Measurement Unsigned Selects measurement on the sensor
Setup/Analog Output 1 40008 1 R/W
Select Integer (0 = Meas1.. 3 = Meas4)
Unsigned Selects output type
Setup/Analog Output 1 Type 40009 1 R/W
Integer (0 = Linear output; 1 = PID control)
Setup/Analog Output 1 Transfer Value 40010 Float 2 R/W Sets the transfer value
Unsigned Sets the output filter value in seconds
Setup/Analog Output 1 Filter 40012 1 R/W
Integer (0 to 120 sec.)
0mA - 4mA Unsigned Selects 0mA/4mA for min output
Setup/Analog Output 1 40013 1 R/W
Select Integer (0 = 0mA; 1 = 4mA)
Setup/Analog Output
Min Setting 40014 Float 2 R/W Sets the min output value
1/Linear
Setup/Analog Output
Max Setting 40016 Float 2 R/W Sets the max output value
1/Linear
Setup/Analog Output Unsigned
PID Mode 40018 1 R/W Sets the PID mode (0 = auto; 1 = manual)
1/PID Integer
Setup/Analog Output PID Manual Sets the PID manual output value
40019 Float 2 R/W
1/PID Set (0.0 to 100.0%)
Setup/Analog Output
PID Setpoint 40021 Float 2 R/W Sets the PID setpoint
1/PID
Setup/Analog Output Unsigned Sets the PID phase (0 = Direct;
PID Phase 40023 1 R/W
1/PID Integer 1 = Reverse)
PID
Setup/Analog Output
Proportional 40024 Float 2 R/W Sets the PID proportional band
1/PID
Band
Setup/Analog Output PID Integral Unsigned
40026 1 R/W Sets the PID integral time (min)
1/PID Time Integer
PID
Setup/Analog Output Unsigned
Derivative 40027 1 R/W Sets the PID derivative time (min)
1/PID Integer
Time
Unsigned Selects data source for this output
Setup/Analog Output 2 Source 40028 1 R/W
Integer (0 = None; 2 = Sensor; 4 = Calculation)
Unsigned Selects sensor source when Source =
Setup/Analog Output 2 Sensor Select 40029 1 R/W
Integer Sensor (0 = Sensor1; 1 = Sensor2)
Measurement Unsigned Selects measurement on the Sensor
Setup/Analog Output 2 40030 1 R/W
Select Integer (0 = Meas1 .. 3 = Meas4)

62
ModBUS Register Information

Table 10 Controller ModBUS Registers (continued)


Group Name Tag Name Register # Data Type Length R/W Description
Unsigned Selects output type (0 = Linear output;
Setup/Analog Output 2 Type 40031 1 R/W
Integer 1 = PID control)
Setup/Analog Output 2 Transfer Value 40032 Float 2 R/W Sets the transfer value
Unsigned Sets the output filter value in seconds
Setup/Analog Output 2 Filter 40034 1 R/W
Integer (0 to 120 sec)
0mA - 4mA Unsigned Selects 0mA/4mA for min output
Setup/Analog Output 2 40035 1 R/W
Select Integer (0 = 0mA; 1 = 4mA)
Setup/Analog Output
Min Setting 40036 Float 2 R/W Sets the min output value
2/Linear
Setup/Analog Output
Max Setting 40038 Float 2 R/W Sets the max output value
2/Linear
Setup/Analog Output Unsigned
Mode 40040 1 R/W Sets the PID mode (0 = auto; 1 = manual)
2/PID Integer
Setup/Analog Output Sets the PID manual output value
Manual Set 40041 Float 2 R/W
2/PID (0.0 to 100.0%)
Setup/Analog Output
Setpoint 40043 Float 2 R/W Sets the PID setpoint
2/PID
Setup/Analog Output Unsigned Sets the PID phase
Phase 40045 1 R/W
2/PID Integer (0 = Direct; 1 = Reverse)
Setup/Analog Output Proportional
40046 Float 2 R/W Sets the PID proportional band
2/PID Band
Setup/Analog Output Unsigned
Integral Time 40048 1 R/W Sets the PID integral time (min)
2/PID Integer
Setup/Analog Output Derivative Unsigned
40049 1 R/W Sets the PID derivative time (min)
2/PID Time Integer
Selects data source for this relay
Unsigned
Setup/Relay 1 Source 40050 1 R/W (0 = None; 1 = Real Time Clock;
Integer
2 = Sensor; 4 = Calculation)
Selects Sensor source when
Unsigned
Setup/Relay 1 Sensor Select 40051 1 R/W Source = Sensor
Integer
(0 = Sensor1; 1 = Sensor2)
Measurement Unsigned Selects measurement on the Sensor
Setup/Relay 1 40052 1 R/W
Select Integer (0 = Meas1 .. 3 = Meas4)
Selects the relay type
Unsigned
Setup/Relay 1 Type 40053 1 R/W (0 = Alarm; 1 = Control;
Integer
2 = Status; 3 = Timer; 4 = Event)
Transfer Unsigned Selects the transfer value for the relays
Setup/Relay 1 40054 1 R/W
Setting Integer (0 = De-energized; 1 = Energized)
Setup/Relay 1/Alarm High Alarm 40055 Float 2 R/W Sets the high alarm setpoint
Setup/Relay 1/Alarm Low Alarm 40057 Float 2 R/W Sets the low alarm setpoint
High
Setup/Relay 1/Alarm 40059 Float 2 R/W Sets the high alarm deadband
Deadband
Low
Setup/Relay 1/Alarm 40061 Float 2 R/W Sets the low alarm deadband
Deadband
Unsigned
Setup/Relay 1/Alarm On Delay 40063 1 R/W Sets the on delay time
Integer
Unsigned
Setup/Relay 1/Alarm Off Delay 40064 1 R/W Sets the off delay time
Integer
Setup/Relay 1/Control Setpoint 40065 Float 2 R/W Sets the controller setpoint

63
ModBUS Register Information

Table 10 Controller ModBUS Registers (continued)


Group Name Tag Name Register # Data Type Length R/W Description
Unsigned Sets the controller phase
Setup/Relay 1/Control Phase 40067 1 R/W
Integer (0 = Low; 1 = High)
Setup/Relay 1/Control Deadband 40068 Float 2 R/W Sets the controller deadband
Overfeed Unsigned
Setup/Relay 1/Control 40070 1 R/W Sets the overfeed timer value (mins)
Timer Integer
Unsigned
Setup/Relay 1/Control On Delay 40071 1 R/W Sets the on delay time (sec)
Integer
Unsigned
Setup/Relay 1/Control Off Delay 40072 1 R/W Sets the off delay time (sec)
Integer
Reset
Unsigned
Setup/Relay 1/Control Overfeed 40073 1 R/W Resets the overfeed timer
Integer
Timer
Setup/Relay 1/Event Setpoint 40074 Float 2 R/W Sets the event setpoint
Unsigned
Setup/Relay 1/Event Phase 40076 1 R/W Sets the event phase (0 = Low; 1 = High)
Integer
Setup/Relay 1/Event Deadband 40077 Float 2 R/W Sets the event deadband
Unsigned
Setup/Relay 1/Event On Max Time 40079 1 R/W Sets the max on time (mins)
Integer
Unsigned
Setup/Relay 1/Event On Min Time 40080 1 R/W Sets the min on time (mins)
Integer
Unsigned
Setup/Relay 1/Event Off Max Time 40081 1 R/W Sets the max off time (mins)
Integer
Unsigned
Setup/Relay 1/Event Off Min Time 40082 1 R/W Sets the min off time (mins)
Integer
Sets which Sensor outputs are affected
Unsigned
Setup/Relay 1/Timer Hold Type 40083 1 R/W during timer on time (0 = None;
Integer
2 = Selected Sensor; 13 = All Sensors)
Selects which Sensor outputs are being
Unsigned held/transferred during the timers on time
Setup/Relay 1/Timer Sensor Select 40084 1 R/W
Integer (this is used when Hold type is set for
single Sensor)
Unsigned Selects hold outputs vs. set transfer value
Setup/Relay 1/Timer Hold Mode 40085 1 R/W
Integer during timers on time
Unsigned
Setup/Relay 1/Timer Duration Time 40086 1 R/W Sets the timer on duration time (sec)
Integer
Unsigned Sets the period between timer on events
Setup/Relay 1/Timer Period Time 40087 1 R/W
Integer (mins)
Sets the time the affected Sensor outputs
Unsigned
Setup/Relay 1/Timer Off Delay 40088 1 R/W are held/transferred after the timer turns
Integer
off (sec)
Unsigned Sets the status level which will trigger the
Setup/Relay 1/Status Level 40089 1 R/W
Integer relay
Selects data source for this relay
Unsigned
Setup/Relay 2 Source 40090 1 R/W (0 = None; 1 = Real Time Clock;
Integer
2 = Sensor; 4 = Calculation)
Unsigned Selects Sensor source when Source
Setup/Relay 2 Sensor Select 40091 1 R/W
Integer = Sensor (0 = Sensor1; 1 = Sensor2)
Measurement Unsigned Selects measurement on the Sensor
Setup/Relay 2 40092 1 R/W
Select Integer (0 = Meas1 .. 3 = Meas4)

64
ModBUS Register Information

Table 10 Controller ModBUS Registers (continued)


Group Name Tag Name Register # Data Type Length R/W Description
Unsigned Selects the relay type (0 = Alarm; 1 =
Setup/Relay 2 Type 40093 1 R/W
Integer Control; 2 = Status; 3 = Timer; 4 = Event)
Transfer Unsigned Selects the transfer value for the relays
Setup/Relay 2 40094 1 R/W
Setting Integer (0 = De-energized; 1 = Energized)
Setup/Relay 2/Alarm High Alarm 40095 Float 2 R/W Sets the high alarm setpoint
Setup/Relay 2/Alarm Low Alarm 40097 Float 2 R/W Sets the low alarm setpoint
High
Setup/Relay 2/Alarm 40099 Float 2 R/W Sets the high alarm deadband
Deadband
Low
Setup/Relay 2/Alarm 40101 Float 2 R/W Sets the low alarm deadband
Deadband
Unsigned
Setup/Relay 2/Alarm On Delay 40103 1 R/W Sets the on delay time
Integer
Unsigned
Setup/Relay 2/Alarm Off Delay 40104 1 R/W Sets the off delay time
Integer
Setup/Relay 2/Control Setpoint 40105 Float 2 R/W Sets the controller setpoint
Unsigned Sets the controller phase
Setup/Relay 2/Control Phase 40107 1 R/W
Integer (0 = Low; 1 = High)
Setup/Relay 2/Control Deadband 40108 Float 2 R/W Sets the controller deadband
Overfeed Unsigned
Setup/Relay 2/Control 40110 1 R/W Sets the overfeed timer value (mins)
Timer Integer
Unsigned
Setup/Relay 2/Control On Delay 40111 1 R/W Sets the on delay time (sec)
Integer
Unsigned
Setup/Relay 2/Control Off Delay 40112 1 R/W Sets the off delay time (sec)
Integer
Reset
Unsigned
Setup/Relay 2/Control Overfeed 40113 1 R/W Resets the overfeed timer
Integer
Timer
Setup/Relay 2/Event Setpoint 40114 Float 2 R/W Sets the event setpoint
Unsigned
Setup/Relay 2/Event Phase 40116 1 R/W Sets the event phase (0 = Low; 1 = High)
Integer
Setup/Relay 2/Event Deadband 40117 Float 2 R/W Sets the event deadband
Unsigned
Setup/Relay 2/Event On Max Time 40119 1 R/W Sets the max on time (mins)
Integer
Unsigned
Setup/Relay 2/Event On Min Time 40120 1 R/W Sets the min on time (mins)
Integer
Unsigned
Setup/Relay 2/Event Off Max Time 40121 1 R/W Sets the max off time (mins)
Integer
Unsigned
Setup/Relay 2/Event Off Min Time 40122 1 R/W Sets the min off time (mins)
Integer
Sets which Sensor outputs are affected
Unsigned
Setup/Relay 2/Timer Hold Type 40123 1 R/W during timer on time (0 = None;
Integer
2 = Selected Sensor; 13 = All Sensors)
Selects which Sensor outputs are being
Unsigned held/transferred during the timers on time
Setup/Relay 2/Timer Sensor Select 40124 1 R/W
Integer (this is used when Hold type is set for
single Sensor)
Unsigned Selects hold outputs vs. set transfer value
Setup/Relay 2/Timer Hold Mode 40125 1 R/W
Integer during timers on time

65
ModBUS Register Information

Table 10 Controller ModBUS Registers (continued)


Group Name Tag Name Register # Data Type Length R/W Description
Unsigned
Setup/Relay 2/Timer Duration Time 40126 1 R/W Sets the timer on duration time (sec)
Integer
Unsigned Sets the period between timer on events
Setup/Relay 2/Timer Period Time 40127 1 R/W
Integer (mins)
Sets the time the affected Sensor outputs
Unsigned
Setup/Relay 2/Timer Off Delay 40128 1 R/W are held/transferred after the timer turns
Integer
off (sec)
Unsigned Sets the status level which will trigger the
Setup/Relay 2/Status Level 40129 1 R/W
Integer relay
Selects data source for this relay
Unsigned
Setup/Relay 3 Source 40130 1 R/W (0 = None; 1 = Real Time Clock;
Integer
2 = Sensor; 4 = Calculation)
Unsigned Selects Sensor source when Source =
Setup/Relay 3 Sensor Select 40131 1 R/W
Integer Sensor (0 = Sensor1; 1 = Sensor2)
Measurement Unsigned Selects measurement on the Sensor
Setup/Relay 3 40132 1 R/W
Select Integer (0 = Meas1 .. 3 = Meas4)
Unsigned Selects the relay type (0 = Alarm; 1 =
Setup/Relay 3 Type 40133 1 R/W
Integer Control; 2 = Status; 3 = Timer; 4 = Event)
Transfer Unsigned Selects the transfer value for the relays
Setup/Relay 3 40134 1 R/W
Setting Integer (0 = De-energized; 1 = Energized)
Setup/Relay 3/Alarm High Alarm 40135 Float 2 R/W Sets the high alarm setpoint
Setup/Relay 3/Alarm Low Alarm 40137 Float 2 R/W Sets the low alarm setpoint
High
Setup/Relay 3/Alarm 40139 Float 2 R/W Sets the high alarm deadband
Deadband
Low
Setup/Relay 3/Alarm 40141 Float 2 R/W Sets the low alarm deadband
Deadband
Unsigned
Setup/Relay 3/Alarm On Delay 40143 1 R/W Sets the on delay time
Integer
Unsigned
Setup/Relay 3/Alarm Off Delay 40144 1 R/W Sets the off delay time
Integer
Setup/Relay 3/Control Setpoint 40145 Float 2 R/W Sets the controller setpoint
Unsigned Sets the controller phase
Setup/Relay 3/Control Phase 40147 1 R/W
Integer (0 = Low; 1 = High)
Setup/Relay 3/Control Deadband 40148 Float 2 R/W Sets the controller deadband
Overfeed Unsigned
Setup/Relay 3/Control 40150 1 R/W Sets the overfeed timer value (mins)
Timer Integer
Unsigned
Setup/Relay 3/Control On Delay 40151 1 R/W Sets the on delay time (sec)
Integer
Unsigned
Setup/Relay 3/Control Off Delay 40152 1 R/W Sets the off delay time (sec)
Integer
Reset
Unsigned
Setup/Relay 3/Control Overfeed 40153 1 R/W Resets the overfeed timer
Integer
Timer
Setup/Relay 3/Event Setpoint 40154 Float 2 R/W Sets the event setpoint
Unsigned
Setup/Relay 3/Event Phase 40156 1 R/W Sets the event phase (0 = Low; 1 = High)
Integer
Setup/Relay 3/Event Deadband 40157 Float 2 R/W Sets the event deadband
Unsigned
Setup/Relay 3/Event On Max Time 40159 1 R/W Sets the max on time (mins)
Integer

66
ModBUS Register Information

Table 10 Controller ModBUS Registers (continued)


Group Name Tag Name Register # Data Type Length R/W Description
Unsigned
Setup/Relay 3/Event On Min Time 40160 1 R/W Sets the min on time (mins)
Integer
Unsigned
Setup/Relay 3/Event Off Max Time 40161 1 R/W Sets the max off time (mins)
Integer
Unsigned
Setup/Relay 3/Event Off Min Time 40162 1 R/W Sets the min off time (mins)
Integer
Sets which Sensor outputs are affected
Unsigned
Setup/Relay 3/Timer Hold Type 40163 1 R/W during timer on time (0 = None;
Integer
2 = Selected Sensor; 13 = All Sensors)
Selects which Sensor outputs are being
Unsigned held/transferred during the timers on time
Setup/Relay 3/Timer Sensor Select 40164 1 R/W
Integer (this is used when Hold type is set for
single Sensor)
Unsigned Selects hold outputs vs. set transfer value
Setup/Relay 3/Timer Hold Mode 40165 1 R/W
Integer during timers on time
Unsigned
Setup/Relay 3/Timer Duration Time 40166 1 R/W Sets the timer on duration time (sec)
Integer
Unsigned Sets the period between timer on events
Setup/Relay 3/Timer Period Time 40167 1 R/W
Integer (mins)
Sets the time the affected Sensor outputs
Unsigned
Setup/Relay 3/Timer Off Delay 40168 1 R/W are held/transferred after the timer turns
Integer
off (sec)
Unsigned Sets the status level which will trigger
Setup/Relay 3/Status Level 40169 1 R/W
Integer the relay
Unsigned Sets the Modbus mode
Comm/Net Card Mode 40170 1 R/W
Integer (0 = RTU; 1 = ASCII)
Sets the Modbus baud rate (0 = 9600;
Unsigned
Comm/Net Card Baud 40171 1 R/W 1 = 19200; 2 = 38400; 3 = 57600;
Integer
4 = 115200)
Unsigned
Comm/Net Card Stop Bits 40172 1 R/W Sets the number of stop bits (1,2)
Integer
Unsigned Sets the register data order for floats
Comm/Net Card Data Order 40173 1 R/W
Integer (0 = Normal; 1 = Reversed)
Min
Unsigned Sets the minimum response time
Comm/Net Card Response 40174 1 R/W
Integer (0 to 30 sec)
Time
Max
Unsigned Sets the maximum response time
Comm/Net Card Response 40175 1 R/W
Integer (100 to 1000 sec)
Time
Comm/Net Unsigned
sc100 40176 1 R/W Sets the sc100 Modbus Address
Card/Addresses Integer
Comm/Net Unsigned
Sensor 1 40177 1 R/W Sets the Sensor 1 Modbus Address
Card/Addresses Integer
Comm/Net Unsigned
Sensor 2 40178 1 R/W Sets the Sensor 2 Modbus Address
Card/Addresses Integer
Good Unsigned
Comm/Net Card/Stats 40179 2 R/W Number of good messages
Messages Integer
Bad Unsigned
Comm/Net Card/Stats 40181 2 R/W Number of failed messages
Messages Integer
Comm/Net Card/Stats % Good Mesg 40183 Float 2 R/W % of good messages

67
ModBUS Register Information

Table 10 Controller ModBUS Registers (continued)


Group Name Tag Name Register # Data Type Length R/W Description
Unsigned Sets the Modbus mode (0 = RTU;
Comm/Service Port Mode 40185 1 R/W
Integer 1 = ASCII)
Sets the Modbus baud rate (0 = 9600;
Unsigned
Comm/Service Port Baud 40186 1 R/W 1 = 19200; 2 = 38400; 3 = 57600;
Integer
4 = 115200)
Unsigned
Comm/Service Port Stop Bits 40187 1 R/W Sets the number of stop bits (1,2)
Integer
Unsigned Sets the register data order for floats
Comm/Service Port Data Order 40188 1 R/W
Integer (0 = Normal; 1 = Reversed)
Min
Unsigned Sets the minimum response time
Comm/Service Port Response 40189 1 R/W
Integer (0 to 30 sec)
Time
Max
Unsigned Sets the maximum response time
Comm/Service Port Response 40190 1 R/W
Integer (100 to 1000 sec)
Time
Comm/Service Unsigned
sc100 40191 1 R/W Sets the sc100 Modbus Address
Port/Addresses Integer
Comm/Service Unsigned
Sensor 1 40192 1 R/W Sets the Sensor 1 Modbus Address
Port/Addresses Integer
Comm/Service Unsigned
Sensor 2 40193 1 R/W Sets the Sensor 2 Modbus Address
Port/Addresses Integer
Comm/Service Good Unsigned
40194 2 R/W Number of good messages
Port/Stats Messages Integer
Comm/Service Bad Unsigned
40196 2 R/W Number of failed messages
Port/Stats Messages Integer
Comm/Service
% Good Mesg 40198 Float 2 R/W % of good messages
Port/Stats
Comm/Sensor/ Sensor1 Good Unsigned
40200 2 R/W Number of good messages
Stats Messages Integer
Comm/Sensor/ Sensor1 Bad Unsigned
40202 2 R/W Number of failed messages
Stats Messages Integer
Comm/Sensor/ Sensor1
% Good Mesg 40204 Float 2 R/W % of good messages
Stats
Comm/Sensor/ Sensor2 Good Unsigned
40206 2 R/W Number of good messages
Stats Messages Integer
Comm/Sensor/ Sensor2 Bad Unsigned
40208 2 R/W Number of failed messages
Stats Messages Integer
Comm/Sensor/ Sensor2
% Good Mesg 40210 Float 2 R/W % of good messages
Stats
Output1 4mA Unsigned
Calibration 40212 1 R/W Calibration counts for the 4mA output 1
count Integer
Output1 Unsigned
Calibration 40213 1 R/W Calibration counts for the 20mA output 1
20mA count Integer
Output2 4mA Unsigned
Calibration 40214 1 R/W Calibration counts for the 4mA output 2
count Integer
Output2 Unsigned
Calibration 40215 1 R/W Calibration counts for the 20mA output 2
20mA count Integer

68
ModBUS Register Information

Table 11 Sensor ModBUS Registers


Register Data
Group Name Tag Name Length R/W Units (U) Range
# Type
Measurements Cal Gain 40013 float 2 R none 0.5 to 2.0
Verification PF Criteria 40062 Integer 1 R/W — —
Measurements Turbidity 40001 Float 2 R NTU 0/100

Diagnostics Temperature 40005 Float 2 R Deg C

Diagnostics Dark Reading 40009 Float 2 R NTU 0/100


Diagnostics Raw Turbidity 40011 Float 2 R NTU —
Diagnostics Lamp Voltage 40018 Float 2 R Volts —
Diagnostics Lamp Current 40020 Float 2 R Amps —
Diagnostics Plus 5V 40022 Float 2 R Volts —
Diagnostics Voltage In 40024 Float 2 R Volts —
Setup Software Version 40015 Float 2 R — —
Setup Bubble Rej 40017 Integer 1 R/W — On/Off

30 sec, 1 min,
Setup DataLog Interval 40026 Integer 1 R/W Sec or Min 5 min, 10 min,
15 min

Setup Sensor Name 40027 String 6 R/W — —


no averaging,
Setup Filter Size 40033 Integer 1 R/W sec
6, 30, 60, 90
Setup Sensor Ser Num 40036 String 6 R/W — 12 digits
Setup Output Mode 40042 Integer 1 R/W — —

Setup Set Resolution 40061 Integer 1 R/W decimal places 4, 3, or 2

69
Index

Numerics H
90 Degree Detector ....................................................... 7 Head Assembly ........................................................... 24

A I
Accuracy ........................................................................ 4 Installation ..................................................................... 8
Alarms ......................................................................... 18 Controller .............................................................. 12
Analog Outputs ............................................................ 18 Panel ..................................................................... 10
Panel Mounting ..................................................... 13
B Pipe Mounting ....................................................... 12
Bubble trap .................................................................... 7 Wall Mounting ....................................................... 12
bubble trap .................................................................... 7
K
C Key
Calibration ................................................................... 40 Functions .............................................................. 26
History ................................................................... 46
StablCal ................................................................ 42 L
User-prepared ....................................................... 41 Language
Calibration Methods ...................................................... 5 Displayed .............................................................. 29
Cleaning ...................................................................... 48
Components M
Customer Supplied ................................................. 8 Maintenance
System .................................................................... 8 Cleaning ................................................................ 48
Controller Dimensions ................................................... 9 Fuse Replacement ................................................ 51
Schedule ............................................................... 47
D ModBUS RS232 .......................................................... 21
Data Logging ............................................................... 36 Mounting ..................................................................... 24
Date
Setting ................................................................... 30 O
Digital Output ............................................................... 21 Output
Dimensions Connections .......................................................... 19
Conduit Hole ......................................................... 10 Setup .............................................................. 32, 37
Display ......................................................................... 27 Outputs .......................................................................... 5
Contrast ................................................................ 28 Hold/Transfer ........................................................ 33
Release ................................................................. 33
E
Electrical Installation .................................................... 14 P
Conduit .................................................................. 14 Photocell Window ........................................................ 48
Connecting/Wiring the Sensor Cable .................... 19 Power Requirements ..................................................... 4
Hard-wiring the Sensor ......................................... 19
Power Cord ........................................................... 14 R
Enclosure, Controller ..................................................... 6 Range ............................................................................ 4
Event Codes ................................................................ 54 Relay
Event Logging ............................................................. 36 Setup .............................................................. 34, 38
Setup Menu .......................................................... 34
F Relays ......................................................................... 18
Flow Rate .................................................................... 25 Connecting ............................................................ 18
Fuse Specifications .......................................................... 5
Rating .................................................................... 51 Response Time ............................................................. 4
Replacement ......................................................... 51
S
Safety Information ......................................................... 6
Sample
Connections .......................................................... 25
Flow .................................................................. 4, 40

70
Index

Sample Line Installation .............................................. 24 System Power ............................................................. 40


Security System Startup ............................................................ 40
Passcode .............................................................. 31
Setup ............................................................... 31, 38 T
System .................................................................. 31 Test/Maintenance Menu .............................................. 38
Sensor Theory of Operation ...................................................... 7
Setup ..................................................................... 37 Time
Starting Sample Flow .................................................. 40 Setting ................................................................... 29
Sun Shield ................................................................... 11 Turbidimeter Body ....................................................... 48
System
Setup ..................................................................... 37 V
Verification .................................................................. 40

W
Warnings
Sensor .................................................................. 53
Wire preparation .......................................................... 15

71
Visit http: //www.hach.com

You might also like