U3751

Spectrum Analyzer

Compact Size and High Performance


An 8 GHz Spectrum Analyzer Based on a New Concept

● Frequency range: 9 kHz to 8 GHz


● High throughput, twice as fast as our conventional models
● High accuracy: ±0.8 dB between 10 MHz and 3.1 GHz

● Standard configuration includes:

10 MHz to 8 GHz pre-amplifier


Two-channel USB interface
LAN port

U3751
The U3751 is a portable 8 GHz
spectrum analyzer that can be
used in digital appliance produc-
tion lines, in installation and
maintenance of CATV and wire-
less LAN hotspots, and for a va-
riety of other purposes.
The U3751 provides high-speed
throughput twice*1 as high as
that provided by our conventional
models. It also contributes to
reduced tact time, an important
consideration in equipment in-
stallation on production lines.

High Throughput Higher Overall Amplitude Accuracy

Conventional system Measuring With a digital IF section and ADVANTEST's original circuit
time
throughput time of 875 technologies, the U3751 provides remarkably high overall am-
ms is reduced by more plitude accuracy:
60%*1 to 350 ms in the ±0.8 dB (10 MHz to 3.1 GHz) ADC Digital RBW Digital Det

U3751 (on GPIB)*2. This ±1.0 dB (3.1 to 8 GHz)


Our conventional model
increased speed helps
RF Input
greatly to reduce test costs Measuring time Digital IF Block
on production lines and
other facilities.

High Speed Calibration PASS/FAIL Judgment

Calibration is an essential requirement for improved accuracy Upper limits and lower limits can be set for limit lines on the
of measurement data. With the latest techniques in circuit in- U3751 screen, thereby enabling the discriminator to make
tegration, the U3751 requires less calibration time and fewer PASS/FAIL judgments on trace data.
calibration steps.

2 U3751-1E Jan. ’05


With a digital interface, the
U3751 has superb power meas-
urement repeatability, guarantee-
ing a overall amplitude accuracy
of ±0.8 dB or better. In addition,
the U3751 is designed to operate
with batteries and warm up with-
in five minutes, offering ease of
use outdoors and in the field.

Highly Sensitive Measurement Five-Minute Warm-Up

In its standard configuration, the The U3751 can warm


Stable measurement

U3751 has a built-in pre-amplifier up within five minutes. 5


min.

for boosting the signal level to a It requires less time to


maximum of 8 GHz. This helps in reach its operating tem-
analysis of faint signals of 5 GHz or perature and can quickly
Our conventional
more, such as those in wireless LAN, be prepared for highly model
ETC, or similar systems. accurate measurement.
Noise level: Time

-135 dBm at 5 GHz, typ. (Pre-Amp. ON)

Operating Time of 2.5 Hours*3 with the Battery Pack USB Interface Included in the Standard Configuration

The U3751 supports three types of power: AC (100 V/200 V), A screenshot can be saved to a USB memory device by simply
DC (+11 V to +17 V), and power from a battery pack. The specifying USB as the COPY device using soft keys. The
battery pack can be detached and easily replaced with a spare. U3751 supports the BMP and PNG formats for saving data.
*1: Comparison with our conventional models This feature enables measurement data to be gathered and pas-
*2: In a sample setup where channel power measurement results are transferred with
certain frequency and span settings specified
ted into reports in a PC environment. (USB version 1.1 is
*3: Typical value at room temperature and without any option attached supported.)
U3751-1E Jan. ’05 3
Functions for Full-Fledged Analysis

The U3751 is a complete spectrum analyzer containing all of


our high-frequency measuring technologies. It combines the
analysis functions of a full-featured spectrum analyzer with ease
of operation. The U3751 can handle all kinds of measurement
demands on its own.

RMS Detection Supported Three-Trace Independent Detection


For more accurate power measurement of broadband modula- The U3751 has a function for displaying three independent
tion, the U3751 supports RMS detection as well as conven- traces performed simultaneously and a variety of wave-detec-
tional sample detection. With RMS detection and digital IF tion modes (RMS, peak, sample, et al.). The combination of
technology, the U3751 is a portable spectrum analyzer capable the function and the calculation methods listed above facili-
of highly accurate power measurement. tates such tasks as antenna adjustments.

ISDB-T channel power measurement sample Sample of simultaneous Posi/RMS/Sample traces

Pre-Amplifier for Signal Boosts up to 8 GHz Channel Setting Function


The U3751 includes a pre-amplifier for boosting the signal The channel setting function is useful for measuring telecom-
level to a maximum of 8 GHz, enabling highly sensitive mea- munication and broadcast channels over radio waves. The
surement. U3751 has two types of channel setting methods: calculation
Noise level: -135 dBm at 5 GHz typ. (Pre-Amp. ON) and table. The channel setting function can handle differences
among broadcast channels in various countries.

Measurement sample of a faint 5.8 GHz signal Bluetooth channel measurement sample
4 U3751-1E Jan. ’05
Various Measurement Functions
Marker functions: Other measurement functions:
Multi-marker (10 markers)/delta marker/peak search Channel power, Total power, Average power, OBW, ACP,
Different types of detectors: Spectrum emission mask, Spurious measurement, Noise-Hz
Normal, Posi, Nega, Sample, RMS conversion, frequency counter, and more

Harmonics Measurement Function Zoom Functions


The harmonics measurement function of the U3751 is suitable A window and F-F mode are used for easy analysis of a specific
for high-frequency testing of radio devices. With input of the signal from a broadband measurement. RBW settings can be
frequency of a fundamental wave or setting of a marker, the changed, so broadband and narrowband analyses of measured
U3751 can easily be used to make a high-frequency measure- signals can be performed quickly. In addition, F-T mode, T-T
ment. mode, and other types of signal analysis are available.

Sample of harmonic distortion measurement on a transmitter with a 400 MHz Dual-window broadband and narrowband measurement sample
bandwidth

ACP Measurement Function Spectrum Emission Mask Function


This function measures ratios between carrier power and adja- PASS/FAIL judgment made with spectrum masks or limit lines
cent channel power. Up to five adjacent channels can be speci- is a good way to increase productivity in the production of dig-
fied, and channel bands can be specified. ital appliances. The Spectrum Emission Mask (SEM) function
facilitates measurement of licensed radio waves on wireless
LANs and other media.

3GPP ACLR measurement sample IEEE802.11b S.E.M measurement sample


U3751-1E Jan. ’05 5
Specifications Dynamic range
Frequency Displayed average noise level: Frequency: 10 MHz to 8 GHz;
Ref. level: <-45 dBm; RBW: 300 Hz
Frequency range: 9 kHz to 8 GHz Band 0, preamplifier: off: -118 dBm + 2 f (GHz) dB
Frequency band: 9 kHz to 3.1 GHz (Band 0) Band 1, preamplifier: off: -117 dBm + 1 f (GHz) dB
3.0 to 8 GHz (Band 1) Band 0, preamplifier: on: -133 dBm + 3 f (GHz) dB
Preamplifier: 10 MHz to 8 GHz Band 1, preamplifier: on: -134 dBm + 1.3 f (GHz) dB
Frequency reference Gain compression (1 dB): Frequency: 10 MHz to 8 GHz
Aging: 2 x 10-6/year Preamplifier: off: >-8 dBm
Temperature stability: 2.5 x 10-6 (0 to 50°C) Preamplifier: on: >-25 dBm
Frequency counter Second harmonic distortion: <-70 dBc (preamplifier: off;
Resolution: 1 Hz to 1 kHz mixer level: -40 dBm; frequency: >200 MHz)
(RBW: <100 kHz, input signal level: >-50 dBm, <-75 dBc typ. (preamplifier: off;
CW and single signal) mixer level: -30 dBm; frequency: >300 MHz)
Frequency stability Third order intermodulation: -50 dBc (frequency: 10 MHz to 8 GHz;
Residual FM (zero span): <60 Hz p-p/100 ms preamplifier: off; mixer level: -20 dBm;
(internal frequency reference) 2-signal separation: 200 kHz)
Frequency span Image/Multiple/
Range: 0, 10 kHz to Full Out-of-band response: <60 dBc (image suppression: ON)
Accuracy: <±1% of Span Residual responses: <-80 dBm (frequency: 10 MHz to 8 GHz;
Spectrum purity: -85 dBc/Hz, offset: 10 kHz; span: <200 kHz preamplifier: off)
Resolution bandwidth
Range: 300 Hz to 3 MHz (1-3 steps)
RBW accuracy: ±12% Inputs/Outputs
Video bandwidth range: 10 Hz to 3 MHz (1-3 steps)
RF Input
Connector: N type female
Sweep Impedance: 50Ω (nominal)
VSWR: <1.7 : 1 (<3.0 GHz), input attenuator: >10 dB
Sweep time <2.0 : 1 (>3.0 GHz), input attenuator: >10 dB
Sweep time: 20 ms to 1000 s (spectrum mode) Calibration output
50 µs to 1000 s (zero span) Connector: BNC female
Accuracy: <±2% (zero span) Impedance: 50Ω (nominal)
Sweep mode: REPEAT, SINGLE Frequency: 20 MHz
Trigger Level: -20 dBm
Source: Free, Video, EXT, IF Frequency reference input
Connector: BNC female
Amplitude range Impedance: 50Ω (nominal)
Frequency [MHz]: 1, 1.544, 2.048, 5, 10, 12.8, 13, 13.824, 14.4,
Measurement range: Noise to +30 dBm 15.36, 15.4, 16.8, 19.2, 19.44, 19.6608,
Maximum input level: Input attenuator: ≥10 dB 19.68, 19.8, 20, 26
+30 dBm (preamplifier off) Level: 0 to +16 dBm
+13 dBm (preamplifier on) External trigger input
±15 VDC max. Connector: BNC female
Input attenuator range: 0 to 50 dB by 10 dB steps Impedance: 10 kΩ (nominal), DC coupled
Display range: 100, 50, 20, 10, 5 dB, Linear Trigger level: 0 to +5 V
Unit: dBm, dBmV, dBµV, dBµVemf, dBpW, W, V 21.4 MHz IF Output
Reference level range: -140 to 40 dBm Connector: BNC female
Detector: Normal, Posi-peak, Nega-peak, Sample, RMS Impedance: 50Ω (nominal)
Level: Approx. mixer input level: +10 dB (at 20 MHz)
Amplitude accuracy Battery mount
Connector: Antonbauer QR mount
Calibration signal External DC input
Frequency: 20 MHz Connector: XLR-4
Level: -20 dBm Voltage range: +11 to +17 V
Accuracy: ±0.3 dB GPIB: IEEE-488 bus connector
Scale fidelity USB: USB1.1
Log: ±0.5 dB/10 dB Video output Connector: D-sub 15-pin female
±0.5 dB/80 dB LAN Connector: RJ45 type, 10/100 base -T
±0.2 dB/1 dB
Level measurement accuracy:After Cal., preamplifier: off;
temperature range: 20 to 30°C; General specifications
input attenuator: 10 dB; REF: 0 dBm; Operating environment range
and input signal level: -10 to -50 dBm Temperature: 0 to +50°C
±1.5 dB (9 kHz to 10 MHz) Humidity: Relative humidity: 85% or less
±0.8 dB (10 MHz to 3.1 GHz) (without condencation)
±1.0 dB (3.1 to 8 GHz) Storage environment range: -20 to +60°C, Relative humidity: 85% or less
AC power input: Automatic switching to 100 VAC or 200 VAC
100 VAC: 100 to 120 VAC, 50/60 Hz
200 VAC: 200 to 240 VAC, 50/60 Hz
U3751 Web Demonstration DC power input: DC: +11 to +17 V
More detailed information on the U3751 spectrum analyzer Power consumption: 100 VA or less (A.C. operation)
70 W or less (D.C. operation)
and its features is available on the Internet.
Mass: 5.6 kg or less (without option)
Access the URL below to read it.
Dimensions: approx. 308 (W) x 175 (H) x 209 (D) mm
http://green.advantest.co.jp/techinfo_e/www_e/ (without protrusion)
approx. 337 (W) x 190 (H) x 307 (D) mm
demonstration_e/U3751/index.html (with handles, feet, and protectors)

6 U3751-1E Jan. ’05


Option 20 High stability frequency reference Ordering information
Reference frequency stability Main unit
Aging: ±2 x 10-8/day Spectrum analyzer: U3751
±1 x 10-7/year
Accessories
Warm-up drift (nominal): ±5 x 10-8
Power cable: A01412
(+25°C, 10 min. after turning the power on)
Input cable: A01037-0300
Temperature drift: ±5 x 10-8 (0 to +40°C, with reference to 25°C)
N-BNC adapter: JUG-201A/U
Ferrite core: ESD-SR-120
Option
High-stability frequency reference crystal: OPT.20
Tracking generator: OPT.74
Bluetooth is a trademark owned by Bluetooth SIG, Inc., U. S. A.
Accessories (optional)
Battery pack: A870008
Charger: A870009
75Ω impedance converter: ZT-130NC
Please be sure to read the product manual thoroughly before using the products. DC power cable: A114020
Specifications may change without notification. Carrying case: A129001
Transit case: A129002

6.5-inch color liquid


Extension-key block crystal display (LCD) Soft-key block Exclusive function keys Front Panel

Operation-key block

Main function keys

POWER USB CAL OUT TG OUT RF INPUT


Control keys

Rear Panel

GPIB
Battery receptacle

DC INPUT
IF OUT

EXT TRIG INPUT USB VIDEO OUT


EXT REF INPUT LAN

U3751-1E Jan. ’05 7


ADVANTEST CORPORATION Korea: Taiwan: North America, Canada, Mexico:
Shin-Marunouchi Center Advantest Korea Co., Ltd. Advantest Taiwan Inc. Advantest America, Inc.
Building, 1-6-2 Marunouchi, 22BF, Kyobo KangNam Tower, No.1 Alley 17, Lane 62, New Jersey Office
Chiyoda-ku, Tokyo 100-0005, 1303-22, Seocho-Dong, Chung-Ho Street, Chu-Pei City, 258 Fernwood Avenue Edison,
Japan Seocho-Ku, Seoul #137-070, Hsin Chu Hsien, Taiwan R.O.C NJ 08837, U.S.A.
Tel: +81-3-3214-7500 Korea Tel: +886-3-5532111 Tel: +1-732-346-2600
http://www.advantest.co.jp Tel: +82-2-532-7071 Fax: +886-3-5541168 Fax: +1-732-346-2610
Fax:+82-2-532-7132 http://www.advantest.com/
Singapore, Malaysia, Thailand, Indonesia,
instruments
China: Philippines, Vietnam:
Advantest (Suzhou) Co., Ltd. Advantest (Singapore) Pte. Ltd. Europe:
Shanghai Branch Office 438A Alexandra Road, #08-03/06 ROHDE & SCHWARZ Europe GmbH
5F, No.46 Building, No.555, Gui Alexandra Technopark Singapore Mühldorfstraße 15
Ping Road, Caohejing Hi-Tech 119967 D-81671 München, Germany
Area, Shanghai, China 200233 Tel: +65-6274-3100 (P.O.B. 80 14 60
Tel: +86-21-6485-2725 Fax: +65-6274-4055 D-81614 München, Germany)
Fax:+86-21-6485-2726 Tel: +49-89-4129-13711
Fax: +49-89-4129-13723
Beijing Branch Office
406/F, Ying Building, Quantum
Plaza, No. 23 Zhi Chun Road,
Hai Dian District, Beijing,
China 100083
Tel: +86-10-8235-3377
Fax: +86-10-8235-6717

© 2004 ADVANTEST CORPORATION


Printed in Japan Bulletin No.U3751-531E Jan. ’05 I