Agilent Cary 8454 UV-Visible Spectroscopy System: Installation Guide
Agilent Cary 8454 UV-Visible Spectroscopy System: Installation Guide
UV-Visible Spectroscopy
System
Installation Guide
Notices
© Agilent Technologies, Inc. 2002, Warranty receive no greater than Restricted Rights as
2003-2008, 2011, 2013, 2014-2016 defined in FAR 52.227-19(c)(1-2) (June
The material contained in this docu- 1987). U.S. Government users will receive
No part of this manual may be reproduced in
ment is provided “as is,” and is sub- no greater than Limited Rights as defined in
any form or by any means (including elec-
tronic storage and retrieval or translation
ject to being changed, without notice, FAR 52.227-14 (June 1987) or DFAR
into a foreign language) without prior agree- in future editions. Further, to the max- 252.227-7015 (b)(2) (November 1995), as
ment and written consent from Agilent imum extent permitted by applicable applicable in any technical data.
Technologies, Inc. as governed by United law, Agilent disclaims all warranties,
States and international copyright laws. either express or implied, with regard Safety Notices
to this manual and any information
Manual Part Number contained herein, including but not
limited to the implied warranties of CAUTION
G1115-90009 merchantability and fitness for a par-
ticular purpose. Agilent shall not be A CAUTION notice denotes a haz-
Edition liable for errors or for incidental or ard. It calls attention to an operat-
11/16 consequential damages in connec- ing procedure, practice, or the like
tion with the furnishing, use, or per-
Agilent Technologies, Australia (M) Pty Ltd
formance of this document or of any that, if not correctly performed or
679 Springvale Road adhered to, could result in damage
Mulgrave, Victoria, Australia 3170 information contained herein. Should
Agilent and the user have a separate to the product or loss of important
www.agilent.com written agreement with warranty data. Do not proceed beyond a
terms covering the material in this CAUTION notice until the indicated
Research Use Only document that conflict with these
terms, the warranty terms in the sep- conditions are fully understood and
This product may be used as a component
of an in vitro diagnostic system if the sys- arate agreement shall control. met.
tem is registered with the appropriate
authorities and complies with the relevant Technology Licenses
regulations. Otherwise, it is intended only
for general laboratory use.
The hardware and/or software described in WARNING
this document are furnished under a license
and may be used or copied only in accor-
Software Revision A WARNING notice denotes a
dance with the terms of such license.
hazard. It calls attention to an
This handbook is for B.05.xx revisions
of the Agilent ChemStation software, Restricted Rights Legend operating procedure, practice, or
where xx is a number from 00 If software is for use in the performance of a the like that, if not correctly per-
through 99 and refers to minor revi- U.S. Government prime contract or subcon- formed or adhered to, could result
sions of the software that do not tract, Software is delivered and licensed as in personal injury or death. Do not
affect the technical accuracy of this “Commercial computer software” as
handbook. proceed beyond a WARNING
defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR notice until the indicated condi-
2.101(a) or as “Restricted computer soft- tions are fully understood and
ware” as defined in FAR 52.227-19 (June met.
1987) or any equivalent agency regulation or
contract clause. Use, duplication or disclo-
Microsoft is a U.S. registered trade-
sure of Software is subject to Agilent Tech-
mark of Microsoft Corporation.
nologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
1 Fast Installation
Use this chapter as an overview of the different actions you must perform to
install your UV-Visible spectroscopy system. If you are an experienced user of
analytical equipment from Agilent Technologies, this may be all you need. If
you need more information, this chapter guides you to other relevant areas in
this or another handbooks.
4 Installing Accessories
This chapter describes how to install the simple mechanical and optical
accessories that are available to expand the capabilities of your UV-Visible
spectrophotometer.
A Safety Information
This chapter contains relevant safety information.
4 Installing Accessories 45
Thermostattable Cell Holder 46
Installing a Thermostattable Cell Holder 47
Cell Stirring Module 48
Installing the Cell Stirring Module 49
Tubing Connections and Operation 50
Long Path Cell Holder 52
Installing the Long Path Cell Holder 53
Installing a Sample Cell 53
Optical Filter Wheel 55
Installing the Filter Wheel Assembly 57
Removing the Filter Wheel Assembly 58
A Safety Information 97
Protective Earth 97
Disconnecting Device 97
UV Light Source 98
Equipment Class 98
Installation category 98
Pollution level 98
Index 99
1
Fast Installation
Installing Preconfigured Agilent Cary 8454 UV-Visible Systems 10
Installing Non-Preconfigured Agilent Cary 8454 UV-Visible
Spectrophotometer Systems 11
Installing Software Only 13
Agilent Technologies 9
1 Fast Installation
Installing Preconfigured Agilent Cary 8454 UV-Visible Systems
If you have purchased one of these systems, the software is pre-installed and
preconfigured. You need only install and switch on the hardware.
✔ Ensure you have space and an appropriate power supply.
For more information, see “Site Requirements for Your
Spectrophotometer” on page 17.
✔ Unpack the system—check for damage and contents against packing list.
For more information, see “Unpacking Your Spectrophotometer” on
page 20.
✔ Install your UV-Visible spectrophotometer.
• Connect line power to the rear.
• Connect the waste tubing underneath the sample pan.
For more information, see “Installing your UV-Visible
Spectrophotometer” on page 21.
✔ Install your computer and printer.
• Connect the keyboard and mouse.
• Connect the monitor to line power and to the computer.
• Connect the printer to line power and to the computer.
• Connect the computer to the spectrophotometer with the LAN cable.
For more information, see “Installing Your Computer” on page 26 and
Chapter 6, “LAN Communication, Installation, Connection and
Configuration”.
✔ Switch on the spectrophotometer (make sure the sample area is clear), wait
2 minutes for the lamps to ignite. When you are ready to operate your
spectrophotometer, switch on the computer and printer.
For more information, see “Starting the Spectrophotometer” on page 27.
✔ Switch on the spectrophotometer (make sure sample area is clear) and wait
two minutes for the lamps to ignite. When you are ready to operate your
spectrophotometer, switch on the computer and printer.
For more information, see “Starting the Spectrophotometer” on page 27.
✔ Install the Agilent ChemStation software.
For more information, see “Installing the Agilent ChemStation Using the
DVD” on page 63.
✔ From the Windows Start menu, start the Agilent ChemStation software.
For more information, see “Starting the Agilent ChemStation Software” on
page 30.
✔ Ensure that your software and spectrophotometer operate properly by
taking a blank measurement or performing the automatic self-test in the
diagnostic section of your software.
For more information, see “Operational Checkout” on page 30.
✔ If you need information about solvents, sample preparation, blank
measurements, cuvette handling or how to insert a cuvette, see Chapter 3,
“Good Measurement Practices”.
✔ Install any accessories.
For simple accessories see Chapter 4, “Installing Accessories”; for complex
accessories, see the instructions supplied with the accessory or the
Installation and Maintenance section of the Agilent ChemStation online
help.
✔ Unpack the software—check for damage and check the contents against the
packing list.
✔ Ensure that your computer and printer meet the requirements for proper
operation of the software.
For more information, see “Minimum PC Requirements for
Agilent ChemStation” on page 60.
✔ Switch on the spectrophotometer (make sure the sample area is clear) and
wait two minutes for the lamps to ignite. When you are ready to operate
your spectrophotometer switch on the computer and printer.
✔ From the Windows menu install the Agilent ChemStation software.
For more information, see “Installing the Agilent ChemStation Using the
DVD” on page 63.
✔ From the Windows Start menu start the Agilent ChemStation software.
For more information, see “Starting the Agilent ChemStation Software” on
page 30.
✔ Ensure that your software and spectrophotometer operate properly, by
taking a blank measurement or by performing the automatic self-test in the
diagnostic section of your software.
For more information, see “Operational Checkout” on page 30.
2
Installation and Start Up
Site Requirements for Your Spectrophotometer 17
Unpacking Your Spectrophotometer 20
Installing your UV-Visible Spectrophotometer 21
Installing Your Computer 26
Starting the Spectrophotometer 27
Starting the Agilent ChemStation Software 30
Operational Checkout 30
Agilent Technologies 15
2 Installation and Start Up
Power considerations
Your spectrophotometer can operate on a line voltage of 100–240 V AC with a
line frequency of 50 or 60 Hz. Maximum power consumption is 220 VA.
The instrument power supply accepts any line voltage supplied in the above
mentioned range. Consequently there is no voltage selector in the rear of the
spectrophotometer. There are no externally accessible fuses, because
automatic electronic fuses are implemented in the power supply.
To disconnect the instrument from line power, disconnect the power cord from the
WARNING
supply. The power supply still uses some power, even if the power switch on the
front panel is turned off.
The computer and InkJet or LaserJet printers accept a nominal line voltage of
110 or 220 V AC with a line frequency of 50 or 60 Hz. For more information,
see the documentation that comes with your computer and/or printer.
Connecting your instrument to a line voltage that is higher than specified causes a
WARNING
shock hazard, and damage may occur to your instrument.
Power cords
Different power cords are offered as options with the spectrophotometer. The
female end of each of the power cords is identical. This connects to the
power-input socket at the rear of the spectrophotometer. The male end of each
of the power cords is different and designed to match the wall socket of a
particular country or region.
Do not replace the mains power cord with one that is inadequately rated. When replacing
NOTE the mains power cord, use only a mains power cord with a rating equivalent to the one
supplied with the instrument. See the instrument installation guide for detailed mains
power cord rating information.
Always operate your instrument from a power outlet which has a ground
WARNING
connection. Always use the power cord designed for your region.
Space
The spectrophotometer is 34.4 cm (13.5 inches) wide, 56.0 cm (22.0 inches)
deep and 18.5 cm (7.3 inches) high (see Figure 1 on page 22). It weighs 14 kg
(30.86 lbs) and will fit on almost any desk or laboratory bench. It needs an
additional 2.5 cm (1.0 inches) of space on either side to allow for the
circulation of air for cooling, and approximately 8 cm (3.1 inches) in the rear
for electrical connections.
The computer, including the keyboard, is approximately 50 cm (19.7 inches)
wide and 60 cm (23.6 inches) deep.
Additional space is required for accessories like pumps and autosamplers.
Environment
Table 1 Suitable environmental conditions for the 8454 UV-Vis
If you use the instrument at environmental temperatures higher than 50°C (122°F),
WARNING
the rear of the instrument may get hot.
Do not store, ship or use your instrument under conditions where temperature
CAUTION
fluctuations could cause condensation within the instrument. Condensation will
damage the electronics. If your instrument was shipped in cold weather, leave it in its
box and allow it to warm up slowly to room temperature to avoid condensation.
Instrument overview
The front view of the spectrophotometer is shown in Figure 1 which shows the
sample compartment open. Unlike conventional instruments the UV-Visible
spectrophotometer does not suffer from ambient stray light. The open sample
area makes it easier to access it generally and to connect tubing to a flow cell
or thermostattable cell holder. The spectrophotometer is shipped with the
standard single-cell cell holder already in place. Standard and accessory cell
holders can be removed and replaced in seconds with few or no tools.
Pushbuttons
Line power switch with
green light
Indicator
The line power switch is located at the lower left part of the instrument.
Pressing it in turns on the instrument. It stays pressed in and shows a green
light when the instrument is turned on. When the line power switch stands out
and the green light is off, the instrument is turned off.
On the front panel of the spectrophotometer is an indicator which will display
different colors depending of the actual condition of the instrument.
• Green: the instrument is ready to measure.
• Green, blinking: the instrument is measuring.
• Yellow: the instrument is in busy state, for example, turning one of the
lamps on or if both lamps are switched off.
• Red: Error condition, that is, the spectrophotometer does not pass one of
the self tests which are run when the spectrophotometer is turned on or an
error occurred during operation. In this case the UV-Visible operating
software gives a detailed error message and possible explanations are in
the online help system. Additionally the Reference Manual gives
information about troubleshooting.
GPIB connector
and switches
Remote connector MIO slot for
LAN interface
Multicell transport
connector
RS 232 connector
• The multi-cell connector allows you to connect the cable which comes from
the multicell transport.
• The GPIO (general purpose input/output) connector allow you to connect
your sipper and autosampler or other accessories.
• The remote connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
common shut down, prepare etc.
• The RS 232 connector may be used to control the spectrophotometer from a
computer through RS 232 connection, using appropriate software (for
future use). This connector needs to be activated by the configuration
switch module next to the GPIB connector. The software needs the
appropriate drivers to support this communication. See your software
documentation for further information.
• The MIO slot for the Lan interface.
• The accessory board slot is reserved for future use.
• The power input socket accepts a line voltage of 100–240 V AC with a line
frequency of 50 or 60 Hz. Maximum power consumption is 220 VA. There is
no voltage selector on your spectrophotometer because the power supply
has wide ranging capability. There are no externally accessible fuses,
because automatic electronic fuses are implemented in the power supply.
The security lever at the power input socket prevents that the
spectrophotometer cover is taken off when line power is still connected.
On the right side of the instrument there is a door for exchanging the lamps.
Behind this plastic door there is another sheet metal door. Two independent
safety light switches are implemented. They automatically turn off the lamps
when the sheet metal door is opened.
2 Connect the waste tubing which comes with your spectrophotometer to the
waste outlet underneath the sample pan.
3 Put the other end of the waste tubing into the waste container.
4 Make sure that the waste tubing has a constant down slope from the
spectrophotometer to the waste container.
Detailed instructions for installing your computer are specific to the type and
configuration of the computer and are provided with it. This section gives a
checklist of the main installation procedures; refer to the documentation that
was delivered with your computer for detailed information.
Before you turn on your spectrophotometer, make sure the light path through
the sample compartment is clear. All filters should be removed from the light
path. Anything blocking or absorbing part of the light beam could cause an
error indication when you turn on the spectrophotometer.
The power switch is located at the left front of the spectrophotometer. When
starting UV-Visible application software wait until the completion of the
self-test procedures for the spectrophotometer. Otherwise the application
software may not recognize the spectrophotometer.
Turning On
1 turn on your spectrophotometer with the pushbutton at the lower left part
of the instrument. The pushbutton will stay pressed in and light green if
power is present.
Indicator
Instrument Warm-up
Technically, once your spectrophotometer has passed the self-tests it is ready
to run samples. For best results you may want to let the spectrophotometer
warm up for at least 1 hour. This time allows the optical system to stabilize
and should result in better measurements. If the instrument was not stored at
room temperature before it was turned on, allow more time for the instrument
to stabilize.
In general, the life of your spectrophotometer will be increased by leaving it on
all the time but switching off the lamps when the instrument is not in use.
lamp intensity of the deuterium lamp and lifetime of lamps decrease with use.
Operational Checkout
3
Good Measurement Practices
General Considerations 32
Inserting a Cell 44
Agilent Technologies 31
3 Good Measurement Practices
General Considerations
General Considerations
There are many factors that can affect the results of your measurements. This
section provides brief discussions of some of the more important ones.
Spectrophotometer Design
The sample compartment of the spectrophotometer is open. Unlike
conventional instruments, the Agilent Cary 8454 UV-Visible
spectrophotometer does not suffer from ambient false light. The open sample
area is easier to access, and it is easier to connect tubing to a flow cell or
thermostattable cell holder.
Making Measurements
Blank (Reference) and Sample Measurement
Your spectrophotometer is a single beam instrument so you must measure a
blank before you measure a sample. For high accuracy measurements, the
blank and the sample measurement should closely follow each other.
In general, a blank measurement should be repeated as often as is practical.
Even in a thermally stable environment, a blank measurement should be taken
every half hour to ensure accurate results.
Chemically, the only difference between the blank and the sample should be
the presence of the analyte(s). For measurements with liquid samples, the
blank should be a sample cell filled with the solvent you plan to use.
Light Beam in
Parallel
Cell Walls
Light Beam in
Non-parallel
Cell Walls
Figure 5 Shift of the instrument light beam due to non parallel cell walls
* Quartz cells with black apertures smaller than 2 mm, when used with a multicell
CAUTION
transport, can lead to measurements of poor reproducibility.
Flow Cells
We recommend a sipper system with a flow cell for obtaining high precision
measurements. Using a flow cell eliminates the necessity of moving the cell
between blank measurement and sample measurement. Also, the cell can be
rinsed thoroughly with the solution to be measured.
The design of the flow cell should minimize entrapment of bubbles and flow
channeling to provide the most reliable results.
Cleaning Cells
The fats in fingerprints are significant absorbers in the UV region and, if left
on optical surfaces, can cause erroneous results. Wipe off all fingerprints and
contaminants before using a sample cell.
Use only high quality lens tissues (part number 9300-0761) and never dry the
inside of a cell with lens tissues. Dry the inside of the cell with pressurized,
oil free air, that prevents the cell from contamination by tissue particles, or
rinse the cell with blank or sample solution. Floating particles in the cell will
deflect the light beam and so lead to a very poor quality of the measured
spectrum.
0.01
0.00
Absorbance -0.01
-0.02
-0.03
-0.04
-0.05
200 400 600 800 1000
Wavelength [nm]
Figure 9 Spectrum taken with floating particles in the light path
Lens tissues for glasses or other uses often contain detergents or lubricants
which can affect your measurements. If possible avoid cleaning the faces of
your cell between blank and sample measurements.
Handling Cells
Always install a cell so that it faces the same direction to minimize problems
with cell non-uniformity. For best results with microcuvettes, leave your
sample cell clamped in position throughout the measurement sequence.
Solutions should be removed and replaced by pipette or use flow cells.
If glass pasteur pipettes are used, make sure that the optical windows of the cell are
CAUTION
not touched or scratched by the pipette.
Solvents
Your choice of solvents should be based primarily on the solvent’s absorbance
characteristics over the wavelengths of interest, its suitability as a solvent for
the analyte, and on experimental conditions. Table 2 lists common solvents
and the lower limit of their useful wavelength range.
200–210 nm Cyclopentane
n-Hexane
Glycerol
2,2,4-Trimethylpentane
Methanol
245–260 nm Chloroform
Ethyl acetate
Methyl formate
280–290 nm Benzene
Toluene
m-Xylene
Many of the solvents in Table 2 are hazardous. Be sure you fully understand their
WARNING
properties before using them.
When using volatile solvents such as acetone or methylene chloride, make sure
that the sample cell is stoppered. Evaporation of a solvent can change the
solute concentration or cause solution noise due to solute convection
currents. Both of these will affect the accuracy of your measurements. We also
recommend stirring and temperature control when you use volatile solvents.
When using water as solvent we recommend using UV grade or HPLC grade
water to reduce unwanted absorbance from the water. If you are using the
sipper/sampler system the water should be degassed to avoid bubble
formation in the flow cell, especially if the water comes from a pressurized
water supply.
Sample Preparation
The sample cell should be rinsed three to five times with your intended solvent
before you fill it with the pure solvent that will be used in the measurement.
Turning the cell upside down on a small stack of absorbent tissues will help
remove any residual solvent. This treatment will minimize contamination from
previous experiments.
Samples which contain colloidal dispersions, dust or other particulate matter
should be filtered, centrifuged or allowed to settle. If not, the overall
attenuation-of-transmittance spectrum due to light scattering and/or
reflection will hide the spectral information from the analyte.
Photosensitive Samples
A few substances are very photosensitive. They degrade or undergo
photochemical reactions if exposed to light. This can be easily seen by a
decrease of sample absorbance over time.
Use of Filters
The shorter wavelength, higher-energy UV light is most likely to degrade
photosensitive samples. If you have a problem, you can selectively block
portions of the UV spectrum with a UV cut-off filter. An optical filter wheel
assembly with three cut-off filters is available for the spectrophotometer. The
cut-off wavelength of the filter you choose should be low enough that it does
not eliminate important spectral information but high enough that it blocks
the light that could degrade your sample. If you use a filter with your samples,
you must use the same filter when you make your blank measurement.
0.0015
Absorbance 0.0010
0.0005
0.0000
-0.0005
-0.0010
Wavelength [nm]
Figure 10 Example of a blank on water showing a good baseline
0.01
0.00
Absorbance
-0.01
-0.02
-0.03
-0.04
-0.05
200 400 600 800 1000
Wavelength [nm]
Figure 11 Example of a blank on water with bubbles causing a poor baseline
If your blank or spectra show artifacts similar to the one in Figure 11, see “Solvents” on
NOTE page 39 to optimize the measurement procedure.
Inserting a Cell
Your spectrophotometer is shipped with the standard single-cell cell holder
you must first install in the sample compartment. This cell holder
accommodates standard cells or flow cells. To insert a sample cell (cuvette) in
the cell holder:
1 Move the locking lever to its up position. 2 Insert the sample cell, making sure you orient it
correctly. The frosted (non-clear) sides of the
sample cell should not be in the path of the light
beam.
3 Lock the sample cell in place by pushing the Small volume flow cells and particularly any cells
locking lever back down. with less than a 2 mm aperture may require use of
the optional adjustable cell holder. The adjustable
cell holder helps you ensure the cells are properly
centered in the light path.
4
Installing Accessories
Thermostattable Cell Holder 46
Cell Stirring Module 48
Long Path Cell Holder 52
Optical Filter Wheel 55
Agilent Technologies 45
4 Installing Accessories
Thermostattable Cell Holder
The Agilent 89054A thermostattable cell holder (see Figure 12) is a single-cell
cell holder with a manifold around the cell. Water from a thermostatted water
bath can be circulated through the manifold to hold the sample cell at a
uniform temperature. Up to 1 cm open-topped cells and flow cells can be used
in a thermostattable cell holder.
Tubing connector
The Agilent 89055A cell stirring module shown in Figure 13, mounts inside the
base of the thermostattable cell holder. It enables magnetic induction stirring
inside a standard 1 cm sample cell. The cell stirring module contains a
plastic-encased magnetic impeller that can be driven by water or pressurized
air. When you put a magnetic stirring bar (9301-1161) into the filled sample
cell, the stirring bar rotates in tandem with the impeller magnet, thus stirring
the sample.
4 Tighten the four screws (flathead, M36 mm) that hold the cell stirring
module in place.
5 Attach the thermostattable cell holder to its base.
6 Mount the thermostattable cell holder back in the sample compartment.
If you need a faster stirring speed than can be obtained by connecting the
manifold and cell stirring module in series, you can connect separate supplies
to the manifold and stirring module.
B
A
D
C
Connection Comment
The Agilent 89076A long path cell holder is a sample cell holder which can be
adjusted to hold most cylindrical and rectangular sample cells with path
lengths of up to ten centimeters. It uses the same mounting system as the
other sample cell holders.
Cell stops
The long path cell holder includes built-in cell stops for accurate positioning of
1, 2, 5, and 10-cm sample cells.
The optical filter wheel (08451-60302) can be mounted on the long path cell
holder in the same way it is mounted on other cell holders, see “Optical Filter
Wheel” on page 55 for details. However, instead of a screwdriver, a 2.5-mm hex
wrench is required to install the optical filter wheel on the long path cell
holder.
Cell stops
Cell clamps
Position cell here
Clamp screws
2 If your sample cell is not one of the standard lengths, position the cell stops
as for a long path sample cell, i.e. move all cell stops in Figure 17 to the left.
3 Because the light beam is collimated, absolute positioning of the sample cell
is not important; however, consistent cell positioning gives more consistent
results. The cell stops make it easier to consistently position your sample
cells.
4 Install the sample cell between the clamps. Slide it against the cell stop. If
your sample cell is not one of the standard lengths, position it so the clamps
are gripping the mid-point of the cell.
5 Allow the clamps to close against the sample cell, see Figure 18. Tighten the
four screws that hold the sample cell in place.
Figure 18 Installed 10-cm cylindrical cell in the long path cell holder
The filter wheel rotates to four different positions. Position zero is empty and
permits 100% transmittance at all light wavelengths. Positions one, two and
three have filters which absorb varying amounts of the UV spectrum.
Figure 20 shows the transmission characteristics of each filter.
0
100
80
1
60
40 2
20 3
Position Filter
0 None
1 UV roll-off
The choice of when to use a filter and which filter to use should be made only
after considering all the factors that affect your measurements. If you decide
that a filter is needed, you should choose a filter that transmits the
wavelengths needed for analytical information and that blocks the
wavelengths that contribute to sample degradation. Often, your final choice
will be a compromise since sometimes even the light at the wavelengths of
analytical interest can affect your sample.
If you do not need a filter for a group of measurements, you can move the filter
wheel to position zero. This permits full light transmission. When you will not
be using the filter wheel for an extended period of time you should remove the
filter wheel assembly from the cell holder and store it in a clean plastic bag.
The optical filters of the filter wheel must be kept clean. Dirty filters interfere
with light transmission and lower the spectrophotometer performance. These
filters may be cleaned with isopropanol and photographic lens tissues.
Screw
Clamp
5
Installing the UV-Visible Software
Agilent ChemStation Computer Details 60
Installing the Agilent ChemStation Using the DVD 63
Agilent Technologies 59
5 Installing the UV-Visible Software
Agilent ChemStation Computer Details
The software will run with the minimum amount of RAM installed, but system performance
NOTE will depend on what other processes are running.
Processor 3 GHz
If you are upgrading from a previous version of Agilent ChemStation, you must uninstall the
NOTE old version of ChemStation first.
Scheduled Processes
Some scheduled processes, like virus checking or defragmenting, may cause
instrument communication problems. Due to these problems, any automated
measurement will be aborted. To prevent problems of this kind please disable
any scheduled event.
Number Format
To avoid wrong values displayed or printed, the International settings must
include the correct number format. Set the Number Format Decimal Separator
to "." (decimal point) in Control Panel > Region and Settings.
Printer
Logon to the PC with the ChemStationSystemUser account and then make
sure that a default printer is configured in your Microsoft Windows operating
system. This is done via the Control Panel application.
If no default printer is configured, you may encounter the following problems:
• the 'Copy To Clipboard' menu may have an error
• in the 'Custom Report View' there may be problems with new templates
• the 'Print Preview' will not work
Overview
This section explains how to use the Agilent Technologies UV-Visible
ChemStation Software Products DVD to:
• install the Agilent ChemStation for the first time
• upgrade/repair an existing Agilent ChemStation software
• install an additional Agilent ChemStation module
Local Administrator rights on the target PC are required to perform the installation.
NOTE
• C:\Chem32\uvexe
• C:\Chem32\1
• C:\Chem32\2
• C:\Chem32\3
• C:\Chem32\4
• C:\Chem32\TMP
• C:\Chem32\NonInst
• C:\Chem32\ChemStation.ini
Only files installed by the product DVD are removed. All additional files such as method and
NOTE data files are not automatically removed. By default they are located in the respective
instrument directory C:\Chem32\1.
If the Security Pack had been installed, please see your manual Agilent ChemStation
NOTE Security Pack for UV-Visible Spectroscopy - User's Guide for details.
During the installation some of your configuration files will be changed. If you
don't intend to use any ChemStation software you can remove the
ChemStation.ini file located in the operating system folder, C:\Windows by
default.
Installation Procedure
Use the following procedure if you are using the DVD to install or upgrade the
Agilent ChemStation software.
Before installing the software:
• Set up the computer hardware.
• To connect your Spectrophotometer via LAN to the Computer see
Chapter 6, “LAN Communication, Installation, Connection and
Configuration”
If ChemStation is already installed and you are only adding a new module (also called an
NOTE Add-on), you must ensure the following conditions are met before following the procedure
below. If you are installing ChemStation and all modules at the same time, skip to the
installation procedure.
6 Select the additional modules you want to install and click Next.
Note that the Security-Pack Add-On and the ECM Add-On are mutually
exclusive. One or the other can be installed, but not both.
7 Enter the License Registration Number for each selected module in the
entry line and click Add License. When all required licenses are added,
press NEXT.
The applicable product and license numbers are printed on your Software
Certificate and Registration Packet (see the example license registration
label).
P LE REVISION CODE
A.08.01
S AM REGISTRATION
NUMBER
DU08E7777A
If you selected the G3770AA or G5182AA ECM access module, establish your ECM access
NOTE using Internet Explorer before you first launch UV-Visible ChemStation.
If these details are not correct, you will not be able to start ChemStation.
CAUTION
3 Click Add.
4 Make sure the location is the PC, and type ChemStationSystem in the object
name field, click OK and click OK again.
5 Highlight the ChemStationSystem domain user that was just added and
then select Full Control and click OK.
6 Repeat this for all of the folders described in the section above.
Autosamplers
If you have an Agilent XY autosampler, your Agilent ChemStation software
requires a special driver and a cable. See the readme.txt for further
information.
6
LAN Communication, Installation,
Connection and Configuration
Installing and Configuring TCP/IP on the PC 76
Installing the LAN Interface in your UV-Visible Spectrophotometer 79
Connecting the Spectrophotometer to the PC 80
Using the Agilent BootP Service 82
Using the Agilent ChemStation Configuration Editor to Configure the
Spectrophotometer 91
Troubleshooting LAN Communication 94
The first five sections of this chapter describe all necessary installations,
connections and configurations of the LAN communication.
The last section of this chapter helps you if the LAN communication does not
work properly.
Agilent Technologies 75
6 LAN Communication, Installation, Connection and Configuration
Installing and Configuring TCP/IP on the PC
3 In the ‘Local Area Connection Properties’ dialog (see Figure 27) click
Configure... to make sure the device is working properly.
If the device is not working properly or no network interface card is configured, please refer
NOTE to the hardware manual of your network card and the PC for how to install and configure a
network interface card.
4 In the ‘This connection uses the following items:’ section of this dialog make
sure the ‘Internet Protocol Version 4(TCP/IPv4)’ is available and checked.
If the TCP/IP protocol is not installed, use the Install button, select Protocol from the list of
NOTE network components to install and press Add.
From the ‘Manufacturers:’ select ‘Microsoft’ and from the ‘Network Protocol:’ select the
‘Internet Protocol Version 4(TCP/IPv4)’ and press ‘OK’. If required, reboot your computer.
If the PC is connected to your companies LAN, do not change any setting. Note down the
NOTE ‘Subnet Mask’ if available. Ask your responsible IT consultant for this mask if not available
and request a fixed IP address for your instrument, if you want to use the existing LAN for
the instrument connection.
For further details please see“Connecting the Spectrophotometer to the PC”.
Gateway, DNS and WINS IP addresses do not need to be configured if you do not connect
NOTE to any other part of a network.
7 Turn off the Microsoft Windows Firewall and reboot the computer.
The JetDirect card (G1846A) and the TalkToLab card (G1369C) are the LAN
interfaces used to physically connect the spectrophotometer to the PC via a
LAN cable. To install the LAN interface card:
1 Switch off the spectrophotometer.
2 Remove the metal cover at the upper right of the rear panel.
3 If the system is on an isolated LAN, the following defaults can be used:
192.168.254.10 for the PC IP address, 255.255.255.0 for the subnet mask,
and 192.168.254.11 for the spectrophotometer. (note: to use these defaults,
DIP switches 5 and 6 on the LAN card must be "ON").
4 Plug the LAN interface card into the slot and screw down the two mounting
screws.
Do not use the standard (non-crossover) Ethertwist 10BaseT cable supplied with the LAN
NOTE interface card G1846A for direct single instrument connection.
Do not use the black 10 ft. long crossover twisted pair cable (part number 5183-4649) in an
NOTE installation with multiple instruments.
Instruments
If the system is on an isolated LAN and the default IP address 192.168.254.11 for the
NOTE spectrophotometer is used, BootP is not necessary.
If the system is on an isolated LAN and the default IP address 192.168.254.11 for the
NOTE spectrophotometer is used, BootP is not necessary.
Addresses
Before installing and configuring BootP Service, you must know the IP
addresses of the computer and instruments, the subnet mask, and the
gateway.
Table 6 Example
Device Address
PC 10.10.10.1
Spectrophotometer 10.10.10.101
Gateway 10.10.10.1
If the system is on an isolated LAN and the default IP address 192.168.254.11 for the
NOTE spectrophotometer is used, BootP is not necessary.
Before configuring BootP Service, you must know the IP addresses of the
computer and instruments, the subnet mask, and the gateway. The MAC or
hardware address of a LAN interface is a world wide unique identifier. No
other network device will have the same MAC address. The MAC address is
printed on the card or can be read out during the configuration procedure:
If the PC has been rebooted after installing the Agilent BootP Service, the Agilent BootP
NOTE Service is started automatically during the next startup of the system. To change BootP
settings, the service must be stopped, the changes made, then restarted. The details are
documented within this section.
When this screen opens the first time, the default settings from installation
are displayed.
6 To edit the TabFile, select Maintain bootp tabfile?
The default tab file was created at installation and is located at Program
Files\Common Files\Agilent Shared\BootP\bin\TabFile. It contains
configuration information entered on this screen.
If necessary you can change the BootPlog File Location by using the browse
button on the right (a valid BootP tabfile must exist).
8 Perform your necessary changes by e.g. creating a log file entry of a new
instrument and edit the new instrument using the Launch Manager.
9 Deselect Do you want to log bootp requests?
10 Click OK to save the values or Cancel to discard them. The program ends.
11 To restart the BootP Service:
For Windows 7: Start > Control Panel > System and Security >
Administrative Tools >Services.
For Windows 10: Right-click on Start and then select Control Panel >
System and Security > Administrative Tools > Services.
12 Right click Agilent BootP Service.
13 Select Start.
14 Close the Services and Administrative Tools screen.
15 This completes the configuration.
#8454 lab
UV-Vis:\
ht=1:\
ha=0030D30D3277:\
ip=<ip address - ask your system administrator>
sm=255.255.255.0:\
gw=<gateway address - ask your system administrator>
vm=auto:\
hn:\
bs=auto:\
T145=64:T146=01:T147=01
If LAN connection is used, Microsoft TCP/IP protocol must be installed and configured, the
NOTE LAN interface Card must be installed, the spectrophotometer must be connected to the PC,
and a bootp server must be installed and configured before the spectrophotometer(s) can
be used. See “Installing and Configuring TCP/IP on the PC” on page 76 for details.
4 Select the Spectrophotometer 8454 in the Modules tab, select the LAN
option, and click Add to add to the Selected Modules box.
5 Select Identify by IP Address and type in the IP address of the LAN
interface card. (If a DNS server is installed you can also define the host
name.)
7 Click OK to exit the Device Configuration dialog box and save your
configuration changes by selecting File > Save.
8 Exit the Configuration Editor by selecting File > Exit.
9 Reboot the PC and power cycle the instrument.
10 Wait until the front panel light on the instrument turns green. Then start
the Agilent ChemStation to check if the software communicates with the
spectrophotometer. In case the red error message line Power Fail shows up,
refer either to the following troubleshooting section or repeat the
installation procedure.
DHCP server
Make sure that no DHCP server runs in the same subnet as the instrument(s),
as these servers also may respond to a bootp request and do not set the buffer
handling of the LAN interface card correctly, i.e. T145=64:T146=01:T147=01.
They also may assign a different IP address to the instrument each time it is
started.
If a DHCP server is running in the same subnet, make sure that fixed IP addresses are
NOTE reserved for your instruments. Ask you IT consultant to make sure that no IP address
conflict may occur.
A
Safety Information
Agilent products must only be used in the manner described in the Agilent
product user guides. Any other use may result in damage to the product or
personal injury. Agilent is not responsible for any damages caused, in whole
or in part, by improper use of the products, unauthorized alterations,
adjustments or modifications to the products, failure to comply with
procedures in Agilent product user guides, or use of the products in violation
of applicable laws, rules or regulations.
Protective Earth
Shock Hazard
WARNING
Connecting the Agilent Cary 8454 UV-Vis to a power source that is not equipped with
a protective earth contact creates s shock hazard for the operator and can damage
the instrument. Always operate your instrument from a power outlet which has a
ground connection.
Disconnecting Device
Shock Hazard
WARNING
To disconnect the instrument from line power, disconnect the power cord from the
supply. The power supply still uses some power and a shock hazard remains inside
the instrument, even if the power switch on the front panel is turned off.
For safety reason do not position the equipment so that it is difficult to operate the
disconnecting device.
Agilent Technologies 97
A Safety Information
UV Light Source
Eye Hazard
WARNING
Eye damage may result from directly viewing light produced by deuterium lamps.
Always turn off the deuterium lamps before opening the lamp door on the
instrument. Direct viewing of ultraviolet rays without protection causes permanent
blindness or vision impairment.
Equipment Class
The Agilent Cary 8454 UV-Vis instrument is suitable for indoor use only and is
classified suitable under Equipment class I category.
Installation category
The installation category is II, based on IEC61010-1. The installation category
implies the regulation for impulse withstand voltage. It is also called the 'Over
voltage category'. 'II' applies to electrical equipment with a nominal supply
voltage up to 300 V.
Pollution level
The pollution level is 2, based on IEC61010-1. Pollution level describes the
degree to which a solid, liquid or gas that deteriorates dielectric strength is
adhering. '2' applies to a normal indoor atmosphere, where only
non-conductive pollution occurs.
U
useful wavelength range of solvents, 39
UV grade water, 40
V
video socket on computer, 26
viscous solutions, 41
volatile solvents, 40
voltage (line power), 17
W
waste container, 25
water, 39
wedge shaped cells, 34
In This Book
This handbook describes the
steps needed to install your
new Agilent Cary 8454
UV-Visible spectroscopy
system.
Most of the handbook deals
with installing a complete
system, that is,
spectrophotometer, computer
and printer. However, in the
second half of this handbook
you will find additional
information about installing
mechanical and optical
accessories, accessory boards,
GPIB connections and
configuration, and software
installation.
11/16
*G1115-90009*
*G1115-90009*
G1115-90009
Agilent Technologies