Determination of Halogens in Coal Using Combustion Ion Chromatography
Determination of Halogens in Coal Using Combustion Ion Chromatography
Because coal is a solid that is not readily solubilized, accessing and quantifying
its constituents requires some form of sample preparation prior to analysis.
1
Combustion methods have been developed that ensure gases. The combustion by-product gases, including HX
complete decomposition of solid and semi-solid samples and SO2/SO3, are passed through an aqueous absorbing
and concomitant release of halogens. These methods solution and directly injected into the IC instrument,
include the use of the Wickbold apparatus, combustion thereby eliminating the sample matrix and any associated
(oxygen) bombs, the Schöniger flask, microwave-induced interferences (Figure 1). The absorbing solution used is
combustion (MIC), and automated furnace combustion dependent on the anions of interest. Typically, deionized
systems such as the one described here.5 After the (DI) water is used for halides or an aqueous hydrogen
sample is combusted, detection can include precipitation peroxide solution if halides and sulfur are being analyzed.
with silver nitrate followed by gravimetric analysis6 or The hydrogen peroxide is required to ensure all the sulfur
measurement by ion chromatography (IC),7 coulometric anions are oxidized to sulfate prior to injection into the
titration,8 or ion selective electrodes9. IC. The fully automated process of CIC saves time and
labor, increases reproducibility, and eliminates the need to
Combustion IC (CIC) uses the thermochemical dispose of hazardous chemicals used in acid digestions or
decomposition of organic matter in the presence of oxygen back extractions.
(oxidative pyrolysis) to convert the analytes of interest to
Combustion tube
Suction
2 Sample burning 4 Absorbent is injected into the IC system
1 Absorbent H20
In this application note, coal samples were combusted • Thermo Scientific Dionex EGC III KOH cartridge
in the Mitsubishi® AQF-2100H system and then fluoride (P/N 074532)
and chloride were accurately and precisely determined
• Thermo Scientific Dionex CR-ATC Continuously
using the Thermo Scientific Dionex ICS-2100 Ion
Regenerated Anion Trap Column (P/N 060477)
Chromatography System with a Thermo Scientific™
Dionex™ IonPac™ AS15 anion-exchange column set. • Thermo Scientific™ Dionex™ AERS™ 500 Anion
Electrolytically Regenerated Suppressor, 4 mm
Equipment (P/N 082540)
• Mitsubishi Automatic Quick Furnace AQF-2100H
− Solid Auto Sampler ASC-240S Software
• Thermo Scientific™ Dionex™ Chromeleon™
––Horizontal Furnace HF-210
Chromatography Data System (CDS) software,
- AQF-2100H Mullite Pyrolysis Set (P/N AQ3QM5) version 7.2
determined
Concentrationby
= CIC
( 100 –are referred to as wet unless it has
Dry
been dried. To determine
% Moisture
)
100 dry values, weigh coal into a
heat resistant container and then dry in an oven at
>100 °C for ~1.5 h. Reweigh the sample and calculate
the
Dilution = ( (
moisture
(Areacontent and dry) concentration:
of Direct Injection)
(Absorption Solution Volume) 1000
(Area of Dilute Injection)
* ) *
% Moisture = 7
Weight Before Drying
Concentration
Dry Concentration = 100 – % Moisture
(100
)
4
Confirmation of Calibration Good precision was achieved as evidenced by the relative
To confirm the validity of the IC calibration, fluoride was standard deviation (RSD) values calculated from the three
determined in triplicate 50 mg aliquots of National replicates (0.2–5.5). In addition to fluoride and chloride,
Institute of Standards and Technology (NIST) standard peaks were detected at 7.3 and 8.2 min. These were likely
reference material 163516 (subbituminous coal) (Table 3). sulfite and sulfate, respectively. The amount of sulfur
The 27.6 mg/kg (2.1% RSD) value was in agreement with present in CIC samples can be determined by adding
that of the certified reference material (25.9 ± 3.3 mg/kg). hydrogen peroxide (H2O2) to the absorption solution at a
Chloride and bromide were detected in this material, but concentration that is high enough to ensure complete
they were not certified. oxidation of sulfite to sulfate. Caution: g/kg sulfur can be
typically found in coal, which would require the use of an
Determination of Halogens in Coal absorbent containing H2O2 in molar excess ( ≥ 900 mg/L).
Because the moisture content was not determined, the Using H2O2 at this concentration can result in the
concentrations reported here are the wet values. Coal appearance of contaminants that interfere with fluoride
samples contained 60–200 mg/kg fluorine, while chlorine determination and could damage the column and other
was generally present at higher levels and had a broader instrument components. Because water (with no H2O2)
range (30–1400 mg/kg; Figure 4 and Table 3). Bromine was was used as the absorbent for the data presented here,
detected in some of the coal samples but the multiple peaks corresponding to the sulfur-containing
concentrations were below the lowest level standard species were observed.
(0.1 mg/L).
Conclusion
This application note demonstrates that chlorine and
52:34 /54+=54"')
$
)52:348
33/* fluorine can be precisely and accurately determined in
2:+49$5:7)+ /54+=!)'797/*-+
2:+4954) 3 !
25<#'9+
33/4
coal using combustion ion chromatography. Analysis was
$
40&52:3+
52%+36
? automated by using the Mitsubishi AQF-2100H system
+9+)9/54 $:667+88+*)54*:)9/;/9>
/54+=#$ 8:667+8857 in combination with the Dionex ICS-2100 IC system that
7+)>)2+35*+
$'362+ 2'41
@5'28'362+8
@
paired reagent-free IC (RFIC) with a Dionex IonPac AS15
column set.
/4:9+8
"+'18 References
2:57/*+
3- 1. American Coal Foundation. Fast Facts About Coal, http://teachcoal.org/fast-facts-
.257/*+
'7(54'9+$:2,/9+$:2,'9+
about-coal (accessed Sept. 28, 2015).
2. Environmental Health and Engineering, Inc. Emissions of Hazardous Air Pollutants
Figure 4. Separation of anions in absorption solution from coal from Coal-Fired Power Plants, Needham, MA, 2011.
combustion.Analyte concentrations have not been adjusted for the
3. U.S. Environmental Protection Agency. Mercury and Air Toxics Standards (MATS).
mass and dilution (see Table 3 for these values).
http://www3.epa.gov/mats/ (accessed Sept. 28, 2015).
4. United States Geological Survey. Mercury and Halogens in Coal—Their Role in
Determining Mercury Emissions From Coal Combustion. Fact Sheet 2012–3122,
Table 3. Fluorine and chlorine results in coal samples. Reston, VA, 2012.
5. Mello, P.A., Barin, J.S., Duarte, F.A., Bizzi, C.A., Diehl, L.O., Muller, E.I., and
Fluorine Chlorine Flores, E.M. Analytical Methods for the Determination of Halogens in Bioanalytical
Sample
mg/kg RSD mg/kg RSD Sciences: a Review. Anal. Bioanal. Chem., 2013, 405, 7615-7642.
6. ASTM D808-11. Standard Test Method for Chlorine in New and Used Petroleum
NIST 1635 27.6 2.1 -- --
Products (High Pressure Decomposition Device Method), 2011.
1 187 1.5 723 2.5 7. Zhang, S., Zhao, T., Wang, J., Qu, X, Chen, W. and Han, Y. Determination of
2 221 0.23 28.6 0.42 Fluorine, Chlorine and Bromine in Household Products by means of Oxygen Bomb
Combustion and Ion Chromatography.
3 75.9 3.2 1420 3.3 J Chromatogr. Sci., 2013, 51, 65-69.
4 77.5 1.2 211 0.4 8. ASTM D6721 - 01(2015). Standard Test Method for Determination of Chlorine in
5 56.5 5.0 1230 1.2 Coal by Oxidative Hydrolysis Microcoulometry, 2001.
9. ASTM D5987 - 96(2007). Standard Test Method for Total Fluorine in Coal and
6 195 3.3 81.7 5.5
Coke by Pyrohydrolytic Extraction and Ion Selective Electrode or Ion Chromatograph
n=3 Methods, 1996.
10. Thermo Fisher Scientific. Product Manual for Eluent Generator Cartridges. Doc. No.
065018-05, Sunnyvale, CA, 2014.
11. Thermo Fisher Scientific. Product Manual for CR-ATC. Doc. No. 031910-03,
Sunnyvale, CA, 2010.
12. Thermo Fisher Scientific. Product Manual for the Dionex ERS 500 Suppressor. Doc.
No. 031956-09, Sunnyvale, CA, 2013.
13. Thermo Fisher Scientific. Dionex ICS-2100 Ion Chromatography System Operator’s
Manual. Doc. No. 065291, Sunnyvale, CA, 2012.
14. Thermo Fisher Scientific. Dionex Integrion HPIC System Operator’s Manual. Doc.
No. 22153-97003 Sunnyvale, CA 2016
15. Mitsubishi Chemical Analytech, Co., Ltd.
Instruction Manual for Automatic Quick Furnace AQF-2100H. Doc. No.
ZN2AQHE-00, Chigasaki, Kanagawa Prefecture, Japan, 2010.
16. National Institute of Standards and Technology. Certificate of Analysis, Standard
Reference Material 1635, Trace Elements in Coal (Subbituminous), Gaithersburg,
MD, 2000.
©2017 Thermo Fisher Scientific Inc. Mitsubishi is a registered trademark of Mitsubishi Corp. All other trademarks are the property of Thermo
Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is
not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications,
terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
AN1145-EN 0117S