Igc-Medical Advances Inc. Quadrature Lower Extremity Coil 472GE-64 472GE-42 Compatible With General Electric Signa 1.5T, 1.0T MRI Systems 07/11/03 Rev.12
Igc-Medical Advances Inc. Quadrature Lower Extremity Coil 472GE-64 472GE-42 Compatible With General Electric Signa 1.5T, 1.0T MRI Systems 07/11/03 Rev.12
QUADRATURE LOWER
EXTREMITY COIL
472GE-64
472GE-42
07/11/03
Rev.12
Proper performance of this coil is guaranteed only while the coil is being
used on the MR system (hardware/software level) specified at the time
of purchase. Upgrades or other modifications to the system software
and/or hardware may affect compatibility. Prior to upgrading your MR
system, please contact IGC-Medical Advances Inc. Customer Service
Department to discuss coil compatibility issues. Failure to do so may
void your warranty.
©
US Headquarters:
IGC-Medical Advances Inc.
10437 Innovation Drive
Milwaukee, WI 53226-4815 USA
(414) 258-3808
800-657-0891
Fax: (414) 258-4931
E-mail: [email protected]
Web: www.medadv.com
A subsidiary of
Intermagnetics General Corporation
NOTICE:
Attention, Consult Accompanying Documents TRANSPORT AND STORE THIS PRODUCT UNDER
THE FOLLOWING ENVIRONMENTAL CONDITIONS
Type BF Equipment
ONLY, FOR A PERIOD NOT EXCEEDING 4 WEEKS:
Class II Equipment
AMBIENT TEMPERATURE OF -34°C TO + 60°C
Conforms to European Medical Device Directive 93/42/EEC RELATIVE HUMIDITY OF 15% TO 95%
(Non-Condensing)
ATMOSPHERIC PRESSURE OF 765hPa TO 1011hPa
M472GE Rev.12
CONTENTS i
Safety Training........................................................................................ 1-1
Quality Assurance........................................................................ 1-1
Indications.................................................................................... 1-1
Contraindications ......................................................................... 1-1
Precautions................................................................................... 1-1
Cautions ....................................................................................... 1-2
Emergency Procedures................................................................. 1-3
Patient Preparation
& Positioning Patient Preparation ....................................................................... 4-1
General Patient Positioning Guidelines ....................................... 4-1
Troubleshooting/
Maintenance Troubleshooting ........................................................................... 6-1
Receiving No Signal .................................................................... 6-1
Image Quality............................................................................... 6-2
Artifacts........................................................................................ 6-3
Maintenance................................................................................. 6-4
Fuse/Cable Continuity Check Procedure ..................................... 6-5
Spare Parts Listing, Quadrature Knee/Foot Coil Assembly ........ 6-7
Spare Parts Listing, Latch and Dual-Lock Assemblies ............... 6-8
M472GE Rev.12
SAFETY 1-1
TRAINING This manual contains detailed information on the set-up, positioning
and use of the IGC-Medical Advances Inc. (MAI) coil. The
instructions should be read carefully and thoroughly before
attempting to scan patients with the coil.
QUALITY ASSURANCE The procedure described in the Quality Assurance Section of this
manual should be performed upon receipt of the coil to establish a
baseline for coil performance. The procedure should be repeated at
regular intervals.
M472GE Rev.12
1-2 SAFETY
CAUTIONS The following general warning statements apply to scanning with a
magnetic resonance system. For further details, review the warnings
included in your MR system operations manual.
• Route all cables out of the magnet so that they do not contact
the patient.
M472GE Rev.12
SAFETY 1-3
EMERGENCY
PROCEDURES In the unlikely event that a coil creates smoke, sparks or makes an
unusually loud noise or if the patient requires emergency assistance:
M472GE Rev.12
COIL INSTRUCTIONS 2-1
SHIPPING LIST
Table 2-1 lists the Quadrature Lower Extremity Coil parts. Please note that
only the coil and configuration ordered will be supplied. Check that all parts
ordered have been shipped.
TABLE 2-1
M472GE Rev.12
2-2 COIL INSTRUCTIONS
Quadrature Lower
Extremity Coil
COIL DESCRIPTION
1. The top half of the coil disconnects from the bottom half for ease of
patient entry and exit.
2. The base allows you to move the coil left or right into alignment
with the anatomy to be scanned.
M472GE Rev.12
COIL INSTRUCTIONS 2-3
PAD DESCRIPTION
Four pads are supplied with the Quadrature Lower Extremity Coil. One pad
supports the knee during examinations of the knee, one pad supports the
foot/ankle during foot/ankle examinations, one pad stabilizes the forefoot,
and one pad supports the unaffected extremity. The pads are designed to
raise the knee or ankle to isocenter in the coronal plane. If more padding is
necessary to align the extremity with isocenter or to increase patient
comfort, use additional pads from the department.
COIL POSITIONING
3. Release the coil centering lever(s) and slide the coil laterally to its
desired position. Secure the coil in its desired position to ensure that
the coil will not move during imaging.
4. Remove the top half of the coil. Apply downward pressure on the
latches located on either side of the coil to release the latches and
separate the two halves of the coil.
6. Put the top half of the coil back in place. Be sure to align the coil
pins on the top half with the receiver holes on the coil bottom half
(note that there is only one way in which it can connect). Secure the
anterior portion of the coil.
M472GE Rev.12
2-4 COIL INSTRUCTIONS
7. Raise the patient table to “full up” position. Ensure that the RF
cables do not get caught in the table or system while the table is
being raised.
SYSTEM/COIL
CONNECTION
COIL ALIGNMENT
1. Turn on the alignment lights and align the axial plane to the
alignment marks on the anterior portion of the coil.
2. If the coil is offset laterally, record the numerical amount from the
numbered scale on the coil base. Enter the offset distance on the
scan range page of your protocol.
M472GE Rev.12
COIL INSTRUCTIONS 2-5
SYSTEM COIL
SELECTION
If you have not had a specific soft key configured for the Quadrature Lower
Extremity Coil, select the Coil Selection appropriate for your system:
4.X Systems:
(1.5T ONLY) Select the [OTHER COILS] soft key on the patient
parameters page.
5.X Systems:
(1.5T ONLY) Select the [OTHER COILS] soft key on the patient
parameters page.
5.X Systems:
(1.0T ONLY) Select the [OTHER COILS] soft key on the patient
parameters page.
LX Systems:
(1.5T, 1.0T) At the “Patient Position” area, select [...] to the right
of the Coil type-in field. A pop-up window of
available coils will appear.
Select [Accept].
NOTE: For systems with 9.0 software or higher, the name for this
coil is: QUADKNEE
M472GE Rev.12
2-6 COIL INSTRUCTIONS
Choosing a coil selection that transmits with the body coil WILL
damage the coil! If you are not confident that a coil selection is a
Transmit/Receive selection, perform the following simple test to help
confirm if the selection is correct.
If any of the results from the previous tests raise concerns about using a
particular system coil selection with the Quadrature Lower Extremity Coil,
have your G.E. Service Engineer evaluate the situation further!
M472GE Rev.12
COIL INSTRUCTIONS 2-7
NOTE: If you wish, contact your G.E. Service Engineer to configure a coil selection
specifically for the Quadrature Lower Extremity Coil. The values entered
should be identical to the default values for the 1.5T and 1.0T G.E.
Transmit/Receive extremity coil. These values are included in
Table 2-2. It is not imperative for a specific coil selection to be created for
the Quadrature Lower Extremity Coil to function properly so long as a
Transmit/Receive Extremity Coil is selected.
TABLE 2-2
M472GE Rev.12
QUALITY ASSURANCE 3-1
RECONSTITUTION INSTRUCTIONS
FOR PHANTOM BOTTLE
2. Fill the phantom bottle approximately half full with distilled water.
The proper amount of sodium chloride (NaCl) and copper sulfate
(CuSO4⋅5H2O) has been added to the bottle.
6. Fill the bottle completely with distilled water so that no air bubbles
are present (all the way to the rim).
INITIAL SNR
MEASUREMENT
2. Position the coil on the cradle, raise the table and connect the coil
according to the steps outlined on Pages 2-2 and 2-3 of the COIL
INSTRUCTIONS Section of this manual.
3. Insert the phantom into the coil (with the top half of the coil
connected) so that the length of the phantom runs along the axis of
the coil. Use the knee pad, supplied with the coil, to position the
phantom in the coil center. Make sure that the coil is centered on the
coil positioning support. (Figure 3-1)
M472GE Rev.12
3-2 QUALITY ASSURANCE
Figure 3-1
5. Prescribe and scan the protocol found on the next page. Be sure to
choose the proper system coil selection as detailed on Page 2-5.
M472GE Rev.12
QUALITY ASSURANCE 3-3
SCAN PROTOCOL
6. Display the image and check for artifacts. If no artifacts are visible,
go to Step 7. If artifacts are present, eliminate them by checking for
the following:
• Check that the color coded cables are correctly attached and
that the top half of the coil is fully closed.
8. Generate an ROI from the display menu and record the Signal Mean
Value in Table 3-1 on Page 3-4.
9. Move the cursor to the background, to the top right of the screen, out
of the phantom image.
10. Generate another ROI from the display menu on the touch screen
and record the Noise Standard Deviation value of the cursor area in
Table 3-1 on Page 3-4.
M472GE Rev.12
3-4 QUALITY ASSURANCE
11. Calculate the Signal-to-Noise ratio from the following equation and
record in Table 3-1 below:
Table 3-1
Once the Initial SNR Measurement has been completed and the values
recorded in Table 3-1, the Quadrature Lower Extremity Coil is ready to be
used for clinical imaging.
PERIODIC QUALITY
ASSURANCE CHECK
1. Follow the steps previously described in this section for Initial SNR
Measurement to obtain a phantom image from the coil.
M472GE Rev.12
QUALITY ASSURANCE 3-5
3. Record the Signal Mean Value from the phantom ROI in Table 3-2,
Periodic SNR Data on Page 3-6.
4. Record the Standard Deviation value of the noise ROI in Table 3-2,
Periodic SNR Data on Page 3-6.
7. Multiply the value from Step 10 by 100, and record this percentage
in column 5 of Table 3-2 on Page 3-6.
8. If this percentage is not greater than 85%, then the coil may be
damaged or there may be a problem with your system. Follow the
troubleshooting instructions in the Troubleshooting/Maintenance
Section of the manual.
M472GE Rev.12
3-6 QUALITY ASSURANCE
TABLE 3-2
1 2 3 4 5
DATE SIGNAL MEAN NOISE SNR VALUE VALUE IN
VALUE STANDARD COLUMN 4
DEVIATION DIVIDED BY
VALUE ORIGINAL
VALUE X 100
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
> 85%
M472GE Rev.12
PATIENT PREPARATION & POSITIONING 4-1
PATIENT
PREPARATION
GENERAL PATIENT
POSITIONING GUIDELINES
1. Position the coil at the end of the table closest to the bore. Check to
see that the cable from the coil is not too far away to plug in. Place
the table mat on the cradle superior to the coil to provide a more
even positioning surface.
2. Keep the base of the coil centered on the cradle. Offset the coil
within the base to adjust for patient size and comfort. Lock the coil
down when in position and record the offset distance for use during
the exam.
3. Be sure not to pinch any gowns or bedding material between the coil
halves when attaching the top half of the coil. This would cause poor
image quality and possibly result in damage to the coil.
M472GE Rev.12
4-2 PATIENT PREPARATION & POSITIONING
Refer to the images below for the two most common applications using the
Quadrature Lower Extremity Coil, knee imaging (Figure 4-1) and
ankle/foot imaging (Figure 4-2).
M472GE Rev.12
PROTOCOLS 5-1
GENERAL PROTOCOL
CONSIDERATIONS
Proceed to scan using your designated protocols and/or the ones included in
the following section. You may use the protocols supplied as is or modify
them to suit your needs keeping in mind the inherent tradeoffs involved in
MR parameter selection. The parameters achievable on your scanner may
vary slightly from those supplied here due to differences in system
software.
M472GE Rev.12
5-2 PROTOCOLS
Sagittal Sagittal Sagittal Coronal
Axial Coronal
Oblique Oblique Oblique Oblique Sagittal 3D
PARAMETERS Gradient
FSE PD FSE T2 Routine SE FSE PD
Oblique
GRE
Localizer T1
w/ Fat Sat w/ Fat Sat w/ Fat Sat w/ Fat Sat
IMAGE MODE 2D 2D 2D 2D 2D 2D 3D
PULSE SEQUENCE Gre,FC Spin Echo Spin Echo Spin Echo Spin Echo Spin Echo Gre
Fast,NPW,
IMAGING OPTIONS FC Fast,NPW NPW Fast,TRF - Fast,Gx,FC
TRF
PSD TYPE IN - - - - - - -
AUTOSHIM On On On On On On On
PHASE CORRECTION Off On On - On - -
RECEIVE BANDWIDTH
16 16 16 16 16 16 16
(KHz)
FOV (cm) 16 14 14 14 14 14 14
SCAN THICKNESS (mm) 5 4 4 4 4 4 1.5
INTERSCAN SPACE (mm) 1 0 0 0 0 0 0
EXPLICIT / GRAPHIC Rx. I55 - S55 Gx Gx Gx Gx Gx Gx
EXPLICIT SATS SI SI SI SI SI SI SI
FAT / WATER SAT (Y/N) - Fat Fat Fat Fat - -
MATRIX (Freq X Phase) 256 x 160 256 x 224 256 x 224 256 x 224 256 x 224 256 x 224 256 x 224
FREQUENCY DIRECTION RL AP AP AP SI SI SI
PHASE FOV 1 1 1 1 1 1 1
NEX 1 2 2 2 2 2 1
M472GE Rev.12
PROTOCOLS 5-3
Sagittal Sagittal Sagittal Coronal
Axial Coronal
Oblique Oblique Oblique Oblique Sagittal 3D
PARAMETERS Gradient
FSE PD FSE T2 Routine SE FSE PD
Oblique
GRE
Localizer T1
w/ Fat Sat w/ Fat Sat w/ Fat Sat w/ Fat Sat
IMAGE MODE 2D 2D 2D 2D 2D 2D 3D
PULSE SEQUENCE Gre Spin Echo Spin Echo Spin Echo Spin Echo Spin Echo Gre
Fast,NPW,
Fast,NPW, NPW,VB, Fast,TRF, Fast,Gx,
IMAGING OPTIONS FC,ED TRF,VB EDR
VB,ED EDR EDR VB,EDR
ED
PSD TYPE IN - - - - - - -
AUTOSHIM On On On On On On On
PHASE CORRECTION - On On - On - -
RECEIVE BANDWIDTH
16 8 8 16 8 16 6
(KHz)
MATRIX (Freq X Phase) 256 x 160 256 x 224 256 x 224 256 x 224 256 x 224 256 x 224 256 x 224
FREQUENCY DIRECTION RL AP AP AP SI SI SI
PHASE FOV 1 1 1 1 1 1 1
NEX 1 2 2 2 2 2 1
M472GE Rev.12
5-4 PROTOCOLS
Coronal
Sagittal Sagittal Axial
Coronal Sagittal Oblique Sagittal 3D
PARAMETERS T1 FSE PD
FSE PD
Oblique
T1 Fast IR GRE
Localizer w/ Fat Sat T1
w/ Fat Sat
IMAGE MODE 2D 2D 2D 2D 2D 2D 3D
PULSE SEQUENCE Spin Echo Spin Echo Spin Echo IR Spin Echo Spin Echo Gre
IMAGING OPTIONS - - Fast,TRF Fast,TRF Fast,TRF - -
PSD TYPE IN - - - - - - -
AUTOSHIM On On On On On - On
PHASE CORRECTION - - On On On - -
RECEIVE BANDWIDTH
16 16 16 16 16 16 16
(KHz)
FOV (cm) 16 14 14 14 14 10 18
SCAN THICKNESS (mm) 5 3 3 4 3 4 1.5
INTERSCAN SPACE (mm) 1 0 1 0 1 0 0
EXPLICIT / GRAPHIC Rx. Offset Gx Gx Gx Gx Gx Gx
EXPLICIT SATS - SI SI SI SI SI -
FAT / WATER SAT (Y/N) - - Fat - Fat - -
MATRIX (Freq X Phase) 256 x 160 256 x 224 256 x 224 256 x 192 256 x 224 256 x 224 256 x 192
FREQUENCY DIRECTION SI SI SI SI SI RL SI
PHASE FOV 1 1 1 1 1 1 1
NEX 1 2 2 2 2 2 1
M472GE Rev.12
PROTOCOLS 5-5
Coronal
Sagittal Sagittal Axial
Coronal Sagittal Oblique Sagittal 3D
PARAMETERS T1 FSE PD
FSE PD
Oblique
T1 Fast IR GRE
Localizer w/ Fat Sat T1
w/ Fat Sat
IMAGE MODE 2D 2D 2D 2D 2D 2D 3D
PULSE SEQUENCE Spin Echo Spin Echo Spin Echo IR Spin Echo Spin Echo Gre
Fast,EDR, Fast,EDR, Fast,EDR,
IMAGING OPTIONS EDR EDR,VB EDR,VB EDR,VB
VB,TRF VB,TRF VB,TRF
PSD TYPE IN - - - - - - -
AUTOSHIM On On On On On - On
PHASE CORRECTION - - On On On - -
RECEIVE BANDWIDTH
16 12 12 12 12 12 10
(KHz)
FOV (cm) 16 14 14 14 14 14 18
SCAN THICKNESS (mm) 5 3 3 4 4 4 1.5
INTERSCAN SPACE (mm) 1 0 0 0 0 0 0
EXPLICIT / GRAPHIC Rx. Offset Gx Gx Gx Gx Gx Gx
EXPLICIT SATS - SI SI - SI SI -
FAT / WATER SAT (Y/N) - - Fat - Fat - -
MATRIX (Freq X Phase) 256 x 160 256 x 224 256 x 224 256 x 192 256 x 224 245 x 224 245 x 192
FREQUENCY DIRECTION SI SI SI SI SI RL SI
PHASE FOV 1 1 1 1 1 1 1
NEX 1 2 2 2 2 2 1
M472GE Rev.12
5-6 PROTOCOLS
IMAGE MODE 2D 2D 2D 2D 3D
PULSE SEQUENCE Spin Echo Spin Echo Spin Echo Spin Echo Spin Echo
IMAGING OPTIONS - Fast,TRF Fast,VB VB Fast,VB
PSD TYPE IN - - - - -
AUTOSHIM On On On On On
PHASE CORRECTION - On On - On
RECEIVE BANDWIDTH
16 16 10 10 10
(KHz)
MATRIX (Freq X Phase) 256 x 160 256 x 224 256 x 224 256 x 224 256 x 224
FREQUENCY DIRECTION RL SI SI SI RL
PHASE FOV 1 1 1 1 1
NEX 1 2 2 2 2
M472GE Rev.12
PROTOCOLS 5-7
IMAGE MODE 2D 2D 2D 2D 2D
PULSE SEQUENCE Spin Echo Spin Echo Spin Echo Spin Echo Spin Echo
Fast,TRF, Fast,TRF, Fast,EDR,
IMAGING OPTIONS EDR EDR
EDR,VB EDR,VB VB
PSD TYPE IN - - - - -
AUTOSHIM On On On On -
PHASE CORRECTION - On On - On
RECEIVE BANDWIDTH
16 12 10 10 10
(KHz)
MATRIX (Freq X Phase) 256 x 160 256 x 224 256 x 224 256 x 224 256 x 224
FREQUENCY DIRECTION RL SI SI SI SI
PHASE FOV 1 1 1 1 1
NEX 1 2 2 2 2
M472GE Rev.12
TROUBLESHOOTING/MAINTENANCE 6-1
TROUBLESHOOTING
The following is a list of common problems and solutions for those problems. If you cannot solve a
problem by following the procedures in the manual, contact IGC-Medical Advances Inc. Customer
Service at 1-800-657-0891 between the hours of 7:30 AM and 5:30 PM, (United States Central Standard
Time) Monday through Friday to arrange for service/repair or to speak with an Applications Specialist.
RECEIVING NO SIGNAL
PROBLEM: You are unable to pre-scan or are scanning and yet receiving no signal.
POSSIBLE SOLUTION: Verify that you have selected the appropriate system coil selection. Refer to
Page 2-5 for additional information.
Verify that the cable is correctly connected to the system and that it is
completely plugged in.
Verify that the landmark is correct and that the cradle has not unlatched.
Verify that the scan locations and any FOV offsets are correct.
If you still cannot get a signal, try to scan (transmit and receive) with the
body coil. For this test, be sure to remove the imaging coil from the
magnet bore before you scan with the body coil. If you still receive no
signal the problem probably lies with your MR system. If the body coil scan
is satisfactory, attempt a scan using both another coil of the exact same
type (receive-only, phased array or transmit/receive, whichever applies)
and the same system coil selection. If the scan completes successfully, there
is probably a problem with the IGC-Medical Advances Inc. coil. Contact
IGC-Medical Advances Inc. for further assistance. If you are unable to scan
with the substitute coil, there may be a system problem related to this
particular coil type.
M472GE Rev.12
6-2 TROUBLESHOOTING/MAINTENANCE
IMAGE QUALITY
PROBLEM: The ratio obtained in the periodic quality assurance check is not greater than
85%, or the image quality is not what you expect it should be, given the
parameters selected.
POSSIBLE SOLUTION: Review the selected protocol. If you are performing the Periodic Quality
Assurance, be sure your protocol is identical to the protocol provided on
Page 3-3 of this manual. If you are performing diagnostic images, you may
need to increase NEX or FOV.
Verify that there are no metal or ferromagnetic objects close to the coil,
patient or magnet (i.e., safety pin, hair pin).
Verify that your center frequency is within the frequency adjustment range
for your system.
Verify that the R1, R2 and TG values from the pre-scan are within normally
expected ranges.
Cleaning the electrical contacts that connect the upper and lower halves
with Isopropyl Alcohol and a cotton swab has been demonstrated to
improve the coil’s SNR on occasion.
M472GE Rev.12
TROUBLESHOOTING/MAINTENANCE 6-3
ARTIFACTS
POSSIBLE SOLUTION: Verify that there is no metal present in the area being scanned in or on the
patient.
PROBLEM: Some or all of the images appear shaded or exhibit uneven signal or
banding.
POSSIBLE SOLUTION: Confirm that no metallic objects are located nearby, outside the FOV. This
is especially important on images utilizing Fat Saturation.
If Fat Saturation is being used, verify that the CFA fine adjustment has been
optimized.
For quadrature coils: Verify that the coil is positioned with the cable exiting towards the bore and
that the system and coil polarity match.
For phased array coils: Confirm that all phased array channels are functioning properly. Sites that
have a Research Key may perform the following procedure themselves.
Those sites without a Research Key will need to have their G.E. Field
Engineer carry out the procedure utilizing a Service Key.
If the above checks out, it is possible the coil has failed. Contact
IGC-Medical Advances Inc. for assistance.
M472GE Rev.12
6-4 TROUBLESHOOTING/MAINTENANCE
MAINTENANCE
STORAGE Coils should be stored and used at the same room temperature as your MR
system.
INSPECTION Inspect the coil weekly for mechanical breakage/damage. DO NOT USE A
COIL WHICH HAS SUSTAINED MECHANICAL DAMAGE. Return the
coil to IGC-Medical Advances Inc. for service/repair. This coil contains no
user serviceable parts. All repairs must be performed by factory trained
personnel.
CLEANING The cleaning solutions listed below have been tested and are recommended
(with comments noted) for cleaning the coil(s) and strap(s). Spray or pour
the cleaning liquid onto a soft cotton cloth and proceed to clean.
Solution Comments
1. Warm water Safe for all areas
of the coil(s) or strap(s)
M472GE Rev.12
TROUBLESHOOTING/MAINTENANCE 6-5
FUSE/CABLE CONTINUITY
CHECK PROCEDURE
1. Remove four (4) nylon screws from the cover of the interface box
assembly. (Figure 6-1)
3. Using the DMM, verify if the fuse is blown. If the fuse is blown,
replace the fuse.
5. Verify open (No short circuit) using DMM between J1 and J2.
6. Verify open (No short circuit) using DMM between J3 and J4.
7. Flex the cable while checking to test for intermittent shorting in the
cable.
8. If the cable fails either test, replace the cable assembly. Affix the
cover to the interface box using four (4) nylon screws and return for
repair.
9. If the cable passes the tests, affix the cover to the interface box using
four (4) nylon screws.
M472GE Rev.12
6-6 TROUBLESHOOTING/MAINTENANCE
Figure 6-1
Figure 6-2
M472GE Rev.12
TROUBLESHOOTING/MAINTENANCE 6-7
SPARE PARTS LISTING
Figure 6-3
M472GE Rev.12
6-8 TROUBLESHOOTING/MAINTENANCE
SPARE PARTS LISTING
Figure 6-4
M472GE Rev.12