Vibration Measuring Instruments CL-173-01-E

The document discusses vibration measuring instruments produced by EMIC Corporation. It summarizes several charge amplifiers and accelerometers that the company manufactures for measuring vibration. The amplifiers can be connected to piezoelectric or integrated electronics accelerometers and support multi-channel vibration testing and data collection via USB communication.

Uploaded by

81Satyam Ghayal
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
93 views

Vibration Measuring Instruments CL-173-01-E

The document discusses vibration measuring instruments produced by EMIC Corporation. It summarizes several charge amplifiers and accelerometers that the company manufactures for measuring vibration. The amplifiers can be connected to piezoelectric or integrated electronics accelerometers and support multi-channel vibration testing and data collection via USB communication.

Uploaded by

81Satyam Ghayal
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 16

V I B R A T I O N

M E A S U R I N G
INSTRUMENTS
The accurate measurement
is necessary to obtain an
effective test result.
is the specialized manufacturer of
vibration measuring instrument and
test equipment.

To go for a more advanced measuring instrument


Since its foundation, EMIC CORPORATION has been consistently to this day involved in the business
of developing and manufacturing the measuring instrument and test equipment used in the vibration
and environmental area. During this time, we have strived as a specialized maker to amass the technical
know-how while receiving warm instruction and encouragement from the customer. We desire to
contribute to the progress in science and technology as much as possible by creating a state of the
art measuring instrument and test equipment based on such a know how.

Quality Assurance
Measuring instrument, source of data analysis.
We keep the importance of quality assurance in mind and expend all possible means to it.
Our goal is to offer the product which can be satisfied with a customer by assuring its quality from prod-
uct design and material procurement to after-sale service.

As all-round manufacturer of vibration products


We are proud of high quality of vibration and environmental test equipment and displacement meter as
well as vibration meter. We would appreciate it very much if you could consider the purchase of them at
the same time.

1
Charge Amplifier, 6001-AHD and 6002-A P3・4

ACP-12 (AC Power Supply Unit with USB Communication Feature) P5

EM-995 Power Supply for Accelerometers with built-in Pre-amplifier P6

505 and 509 Series Charge Amplifier Type Vibrometer P7

Technical Information of Charge Amplifier P8・9

Application Note for Accelerometer P9・10

540 Series Piezoelectric Type Accelerometer P11・12

700 Series Piezoelectric Type Accelerometer P13・14

2
Charge Amplifier  Vibration Measuring Instrument
High Performance with Slim Body

Features:
● Matched to both piezoelectric accelerometer and
accelerometer with built-in electronics.
● Enhanced filter according to JIS B 0907, 6 steps for
both L.P.F and H.P.F.
● 36 mm width of slim body facilitates to configure
multi-channel measurement system
● Optional AC power supply unit (USB communication
feature) allows the PC to load measured data

Charge Amplifier 6001-AHD


● 3 measuring modes with integrating network:
acceleration, velocity, and displacement
● Wide sensitivity range for setting accelerometer:
0.01000 pC/(m/s2) to 999.9 pC/(m/s2)
● Broadband frequency response (acceleration measurement):
0.5 Hz to 90 kHz

Accelerometer

CHARGE AMPLIFIER

Vibration Test System 6001-AHD Data Logger


Front Rear

3
Charge Amplifier 6002-A

● By saving the storing space to put away 2ch in one unit


multi-channel measurement becomes possible.
● Broadband due to acceleration measurement from 0.5 Hz to 30 kHz

Accelerometer

CHARGE AMPLIFIER

Vibration Test System 6002-A Data Logger


Front Rear

Model 6001-AHD 6002-A

Matched Piezoelectric Accelerometer (Max. input: 100,000 pC)


accelerometer Accelerometer with Built-in Pre-amplifier (Constant-current drive: 24 V/4 mA)
Max. input 1 2
Sensitivity 0.01000 to 99.99 pC/(m/s2) 0.100 to 99.9 pC/(m/s2)
setting range 0.1000 to 999.9 mV/(m/s2) 0.100 to 99.9 mV/(m/s2)
Measuring mode Acceleration : m/s2 Velocity : mm/s Displacement : mm Acceleration : m/s2
Max. measurable acceleration 10000 m/s2 10000 m/s2
Acceleration measuring range 0.1/0.3/1/3/10/30/100/300/1000/3000/10000 m/s2
Max. measurable velocity 10000 mm/s -
Velocity measuring range 0.1/0.3/1/3/10/30/100/300/1000/3000/10000 mm/s -
Max. measurable displacement 1000 mmp-p -
Displacement measuring range 0.01/0.1/0.3/1/3/10/30/100/300/1000 mm -
Acceleration : 0.5 Hz to 90 kHz ±15% Acceleration : 0.5 Hz to 30 kHz ±10%
Frequency 1.0 Hz to 80 kHz ±5% 1.0 Hz to 20 kHz ±5%
characteristics V e l o c i t y : 3 Hz to 1 kHz ±5%
Displacement : 3 Hz to 200 Hz ±10%
AC output Max. 1 V0-p full scale Max. 1 V0-p full scale
DC output Max. 1 V0-p full scale -
Attenuation: -18 dB/oct Butterworth, OFF/3/10/20/50/100/300 Hz according to
High-pass filter
JIS B 0907-1989 Characteristic of Butterworth Filter (HPF: 10 Hz, LPF: 1 kHz)
Attenuation: -18 dB/oct Butterworth, 100/300/1k/3k/5k/10k Hz according to JIS
Low-pass filter
B 0907-1989 Characteristic of Butterworth Filter (HPF: 10 Hz, LPF: 1 kHz)
Display value Root mean square (rms) EQ PEAK(rms×√2) EQp-p(EQ PEAK×2)
Noise level 0.05 pCrms (Less than 10 mV/rms)
Quantity of connecting unit Up to 12 units
Communication USB2.0 (An AC power supply unit with USB communication function sold separately is necessary.)
function RS-232C
Input power DC 9 to 15 V
Dimensions 36W×149H×240D mm
Mass 1 kg or app.
Ambient conditions Temperature: -10 to +50℃, Humidity: 20 to 85%RH without dewdrop

4
ACP-12 (AC Power Supply Unit with USB Communication Function)

Features:

Power supply up to 12 units : To connected charge amplifier units


Settings by PC via USB : Various settings of charge amplifier can be set remotely
Acceptance up to 24 channels : Up to 24 channels can be set remotely

Configuration Example: 6001-AHD, 6002-A, ACP-12

Specifications
Input power Wide input range of 100/200 V system, voltage: AC 85 to 265 V, frequency: 47 to 66 Hz
Output capacity More than 4.0 A @ DC+12 V±5%, up to 12 units in mixed combination of 6001-A and 6002-A
Communication External communication (See software specifications for isolated USB 2.0, full speed and communication specifications)
Front Rear

O p t i o n
Description Model Remarks
Power Supply Select a suitable model for your application from the followings.
Standard AC Adapter ADP- 03 for driving up to 3 units Input voltage: AC 85 to 265 V
Medium AC Adapter ADP- 06 for driving up to 6 units Input voltage: AC 85 to 265 V
Large AC Adapter ADP-12 for driving up to 12 units Input voltage: AC 85 to 265 V
AC Power Supply Unit
ACP-12 for driving up to 12 units Input voltage: AC 85 to 265 V
with USB communication function
Dry Battery Unit (Upcoming product) DBP- 03 for driving up to 3 units Houses 8 size C batteries
Secondary Cell Unit (Upcoming product) BAP-03 for driving up to 3 units Charging type
Car Battery Adapter CAP-01 for driving up to 3 units -
Remote Software
Independent Remote SVM01 Compatible with 6001-A
Remote for Plurality of Units SVM02 Compatible with 6001-A and 6002-A
Remote for Plurality of Units matched to AHD SVM03 Compatible with 6001-A, 6002-A and 6001-AHD
Connection Kit In connecting the unit the suitable connecting plate and cable for your application are required.
Connection Plate CON - ** Fill in the quantity of unit for your application on“**”
.
Connection Cable COB - ** Example: For connecting 2 units, models are“CON-02”and“COB-02”respectively.
Grip Kit CAR-03 1 set of foot can be attached for every 3 units.
Rubber Foot - 1 set of foot can be attached for every 3 units.
Rack-mount Kit
Rack Mounting Board RAC - 01 Board for putting the unit in a rack
Rack Mounting Bracket RAF - 01 Bracket for securing the mounting bracket to a rack
Blank Panel RAP-01 Blindfold board for insufficient quantity of the unit
Accelerometer
General-Purpose 731-B Center stud mount type with side connector
Small 710-D Center stud mount type with side connector, most suitable for modal analysis
General-Purpose Triaxial 712-B3 Small structure, high sensitivity, simultaneous measurement in three axes
Small Triaxial 703-B3 Small structure, simultaneous measurement in three axes
Ultrasamall Triaxial 702-B3 Ultrasmall, simultaneous measurement of lightweight specimen in three axes

SVM Remote Software SVM Remote Control Measuring System using ACP-12
Communication Remote Operation
This software makes it possible to remotely Power Supply Monitoring of Measurement

control settings on the panel of the 6001-A


charge vibrometer and 6002-A 2ch charge
CHARGE AMPLIFIER CHARGE AMPLIFIER CHARGE AMPLIFIER AC-POWER
SUPPY
ACP-12

amplifier through USB interface via ACP-12. ALARM


CURRENT
TEMP
VOLT

Since the ACP-12 can connect up to 12 up to 12 units


RESET

connectable
units (total quantity of 6001-A and REM.

Internal Bus
6002-A), measuring channel can be
POWER

USB
remotely controlled up to 24 channels.
Charge Amplifier Microsoft Windows PC
6001-AHD/6002-A AC Power Supply Unit
※The multi-channel measuring system can be
configured up to 24 channels with 24 6001-A units
ACP-12 Remote Window
and two ACP-12 units, but it is acceptable even if one All remote settings of connected charge
ACP-12 unit does not have the communication feature. amplifier, transducer type, charge
※Please contact us about the multi-channel configura- sensitivity, measuring mode, filter, etc..
tion from 25 to 99 channels. Accelerometer and monitoring can be performed.

5
EM-995 Power Supply for
Accelerometer with Built-in Pre-amplifier
4 Channel Constant Current Power Supply for Accelerometer
with Built-in Pre-amplifier
The power supply unit for supplying power to the accelerometer with built-in pre-amplifier.
The accelerometer cable can be extended using a coaxial cable because of low output imped-
ance of the unit.

Specifications
Model PS-504
Input channels 4 channels
I/O Connector BNC Receptacle
I/O Gain × 1 (fixed)
Lower limit frequency within -3 dB at 0.2 Hz, within -5% at 0.7 Hz
Upper limit frequency Upper limit frequency of Accelerometer w/ built-in pre-amplifier
Max. voltage DC 24 V
Output current DC 4mA
Input power AC100V±10%
Max. consumption 10 VA
Operating environments Temperature: 0 to 50 ℃, Humidity:95%RH w/o dewdrop
Outer dimensions 200W×55H×140D mm
Mass 1.3 kg

Charge Amplifier Vibration Controller,


Accelerometer Accelerometer Cable
Vibrometer Analyzer, etc.
Microdot Plug - BNC Plug
Low-noise Cable

Piezoelectric Type Pre-charge Amplifier Vibration Controller


Accelerometer 10-32UNF BNC 504series DCS-98000MJ
(Charge Output) Data Logger,
Microdot Plug - Microdot Plug Microdot to BNC Analyzer, etc.
Low-noise Cable Adapter

10-32UNF 10-32UNF BNC


Microdot to BNC
Mini-microdot Plug - Microdot Plug Adapter
Low-noise Cable

5-44UNF 10-32UNF BNC

Mini-microdot Plug - Microdot Plug


Low-noise Cable
Pre-charge Amplifier
from 6001-AHD 6002-A
5-44UNF 10-32UNF Accelerometer

Microdot Plug - Microdot Plug


Low-noise Cable CHARGE AMPLIFIER

from
Accelerometer
10-32UNF 10-32UNF

Microdot Plug - Microdot Plug


Cable
Accelerometer w/
Pre-amplifier to Input Terminal
(Voltage Output) 10-32UNF 10-32UNF on rear panel
Microdot to BNC
Microdot Plug - Microdot Plug Adapter
Cable
Power Supply for Accelerometers
with Built-in Pre-amplifier
10-32UNF 10-32UNF BNC

Microdot Plug - BNC Plug


Cable

10-32UNF BNC

Microdot Plug - BNC Plug


Cable

10-32UNF BNC
BNC-BNC BNC-BNC
Relay Adapter Coax Cable(Max. recommended length: 25 m)

BNC

6
505 & 509 Series Charge Amplifier Type Vibrometer

505-CBP
505-CB Unit
Charge Amplifier Type Vibrometer for measuring Accelera-
tion, Velocity and Displacement.
Up to 6 channels can be combined in order to accommo-
date a request from the multi-channel measurement.

509-CBP
509-CB Unit
Charge Amplifier Type Vibrometer for measuring Accelera-
tion, Velocity and Displacement.
505-CBP 509-CBP
Up to 6 channels can be combined in order to accommo-
date a request from the multi-channel measurement.

Features:
● Since the output voltage is not affected by the capacitance of an
accelerometer cable, the cable length can be changed as you please.
● The vernier dial can readily calibrate the amplifier to a specific
accelerometer sensitivity.
● Since there are four switching positions of cut-off frequency in both
high-pass and low-pass filter, the analysis can be easily performed.
● Two output systems of AC and DC output can drive all the recorders. Integration of 505-CB Units
● A recorder can be easily calibrated with an internal calibration signal.
● Two kinds of power supply, providing both AC(100 V) and DC(12 V)
power and providing only AC(100 V) power are available.

505-CB-3A3 509-CB-3A3
505-CB-6B6 509-CB-6B6
Model 505-CBP 505-CB-1A1 505-CB-3B3 509-CBP 509-CB-1A1 509-CB-3B3
505-CB-6C6 509-CB-6C6
505-CB-3C3 509-CB-3C3
Matched
Piezo-electric Type Acceleromter Piezo-electric Type Acceleromter
Accelerometer
Max. number of inputs 1 1 3 6 1 1 3 6
Max. acceleration measurement 10000 m/s2 10000 m/s2
Acceleration measuring range 0.3※1/3/10/100/300/1000/3000/10000 m/s2 0.3※1/3/10/100/300/1000/3000/10000 m/s2
Max. velocity 10000 cm/s(5 to 500 Hz)

measurement 100 cm/s(200 to 3k Hz)
Velocity 0.3※1/1/3/10/30/100/300/1000/3000/10000 cm/s(5 to 500 Hz)

measuring range 0.03/0.1/0.3/1/3/10/30/100 cm/s(200 to 3k Hz)
Max. displacement 1000 mmp-p(5 to 100 Hz)

measurement 100 mmp-p(50 to 1k Hz)
Displacement 0.3/1/3/10/30/100/300/1000 mmp-p(5 to 100 Hz)

measuring range 0.03/0.1/0.3/1/3/10/30/100 mmp-p(50 to 1000 Hz)
Frequency response Less than -1 dB (1 Hz to 30 kHz) Less than -1 dB (1 Hz to 30 kHz)
Noise level 0.01 pCrms+0.005 pCrms/1000 pF 0.01 pCrms+0.005 pCrms/1000 pF
Max. charge input ±100,000 pC ±100,000 pC
High-pass filter 5 Hz,20 Hz,50 Hz,200 Hz 5 Hz,20 Hz,50 Hz,200 Hz
(Attenuation: -12 dB/oct) (Attenuation: -12 dB/oct)
Low-pass filter 500 Hz,1000 Hz,5000 Hz,10000 Hz 500 Hz,1000 Hz,5000 Hz,10000 Hz
(Attenuation: -12 dB/oct) (Attenuation: -12 dB/oct)
AC output Max. 1 V0-p Max. 1 V0-p
DC output DC0 to 5 V DC0 to 5 V
AC power supply AC100V 50/60Hz
AC100V 50/60Hz AC100V 50/60Hz AC100V 50/60Hz AC100V 50/60Hz AC100V 50/60Hz AC100V 50/60Hz AC100V 50/60Hz
DC power supply DC12V 300 mA DC12V 300 mA DC12V 300 mA DC12V 600 mA※2 DC12V 300 mA DC12V 300 mA DC12V 300 mA DC12V 600 mA※2
Matched power DMA-500 DMB-500 DMA-500 DMB-500
supply unit ー DMA-500 DMB-500 DMC-500※2 ー DMA-500 DMB-500 DMC-500※2
DMC-500※2 DMC-500※2
Dimensions 120W×180H×260D 180W×180H×260D 300W×180H×260D 480W×180H×260D 120W×180H×260D 180W×180H×260D 300W×180H×260D 480W×180H×260D
Housing model ー CAC-501 CAC-503 CAC-506 ー CAC-501 CAC-503 CAC-506
Mass 3.5 kg ※3 ※3 ※3 2.0 kg ※3 ※3 ※3
※1 The measuring range is available only when the accelerometer sensitivity is from 1 to 10 pC(m/s2).
※2 The DMC-500 provides the AC power only and does not do the DC power supply.
※3 As for the details, contact us.

7
Technical Information of Charge Amplifier
● Theory of Operation
The charge amplifier converts the charge output from a piezoelectric accelerometer
(vibration sensor) into a voltage signal and amplifies it.

Since the current is obtained by differentiating the electrical charge with respect to time (I = dQ/dt), the charge amplifier is
suitable for conditioning a minute signal like the current amplifier (I/V converter). Let Vout, Qs, Cs be the output voltage, output
charge from the accelerometer, accelerometer capacitance respectively, then, the simple equivalent circuit model among
charge and voltage gives the following equation.
Qs
Qs
V out =
Cs
The cutoff frequency FH of a high-pass filter can be defined V out
by a load resistance RL as follows; Transducer RL
(Accelerometer)
1
FH =
Cs
2π・  ・RL
Since the resistance RL controls the low frequency response of the circuit, it shall be generally large. For this reason, an
integration circuit using an operational amplifier is absolutely essential so that electric capacitance connected to the
accelerometer does not have an influence on the charge sensitivity. The parallel connection of the accelerometer and cable
capacitance, difference in electric capacitance between accelerometers and electric properties of cable such as material,
structure, length worsen precision in the measurement.
Qf

Refer to the electric equivalent circuit model as shown in the right, Cf


where QS is the electric charge generated by the accelerometer, Q0
the electric charge accumulated in the accelerometer (Cd), Cf the RF
feedback capacitance, Cd the capacitance of the accelerometer, e -

the open circuit voltage, RF the feedback resistance and A the gain +
of the operational amplifier, respectively. Then the relational e
Operational V out
Qs Cd Cc Amplifier
expressions are established between these parameters as follows; (Gain: A)

Q0 Qf
V out = - A e e = e - V out =
Cd+ Cc Cf
Cd+ Cc 1 Vibration Sensor Cable Signal Conditioner
Q0 = + Qf ・ (Accelerometer)
Cf 1+ A
Qf 1
e = + ・
Cf 1+ A
Qf A
V out = - ・
Cf 1+ A

Since the gain of the operational amplifier is quite large, the above electrical charge and voltages can be approximated as follows;

Q 0 ≒ 0
e ≒ 0
Qf
V out = -
Cf
The effect of varying input accelerometer or cable capacitance turns out to be insignificant.

● NOTES
How to handle:
In order to ensure reliable performance of the charge amplifier the user must be familiar with the type and characteristics of an acceler-
ometer for use with it and select the most suitable accelerometer for the application. Two types of accelerometer are available,“piezo-
electric accelerometer”and“accelerometer with built-in electronics”
.
※ As for the details of accelerometer, see“Application Note for Accelerometer”at pages 9 and10. Since the charge amplifier is
precision equipment, please pay attention to the following points.

● Since a dirty connector or cable may have a bad effect on measurement accuracy, please clean them.
● Please take care of handling the connector so that static electricity will not apply to it because there may be a possibility of being failure or damaged.
● It is impossible to disconnect the accelerometer with built-in electronics under the condition that the power is supplied through the charge amplifier.
● In measuring the velocity and displacement please set the cutoff frequency of the high-pass filter to 3 Hz to remove the fluctuation of meter
indication in the low frequency range. The indicated value will become stable.
● Please check the sensitivity, measuring range, and settings of the filter before commencing the measurement.
● Please use the accelerometer whose mass is 1/50 of a measuring object. If the accelerometer is heavy, there is a possibility of changing
its dynamical characteristics. Depending on the configuration its vibration level may reach 100 times of excitation level due to resonance.
As such is the case, please select the suitable accelerometer whose shock limit meets the requirement for your application. If failed to
select a proper accelerometer, the accelerometer may be damaged or charge amplifier failure.

8
Application Note for Accelerometer
● Introduction

Accelerometer
The accelerometer is used for measuring acceleration of an object. Gravity, vibration/motion, and shock can be measured as three
prominent types of kinematic measurable amount. The positional change of the mass caused by acceleration and low acceleration
can be measured with some accelerometers; therefore, static acceleration (response at 0 Hz) and gravity can be detected. Other
type of accelerometer has high frequency characteristics enabling to measure vibrations up to 20 kHz. They are generally referred
to as acceleration pickup, acceleration sensor, and accelerometer.

Accelerometer Type … The commonly used accelerometers are described as below.

1 Piezoelectric Type Accelerometer


The accelerometer having a piezoelectric element generates the charge being proportional to the acceleration to which the
accelerometer is subjected. The output charge will be converted into the voltage output with the charge amplifier to measure
acceleration. The output charge is very low in the order of picocoulomb, thereby being electrostatically affected by bending or
twisting of a cable giving a false output signal. For this reason, the low noise cable is essentially used.

 Features of Piezoelectric Accelerometer


● High sensitivity・Small ● Broadband (measurable from low to high frequency) ● Large measuring range (from very weak vibration
to high acceleration) ● High mechanical strength ● Superior environment immunity ● Power supply is unnecessary except for
accelerometer with built-in electronics ● Wide temperature range (extreme cold, high temperature)

2 Accelerometer with Built-in Electronics


It says the piezoelectric accelerometer which has a built-in charge to voltage converter.

 Features of Piezoelectric Accelerometer with Built-in Electronics


● High sensitivity despite small size because of built-in charge to voltage converter ● Built-in filter defines any frequency response
● Other types of instrument cable than low noise cable are acceptable ● Sensitivity deterioration and noise due to cable length are
small ● Immune to humidity because of low impedance

Structure of Piezoelectric Type Accelerometer Structure of Shear-type Accelerometer

M M
M K K
Mass
Mass
K M Piezoelectric Element
Piezo- K

electric
Element

● NOTES
How to handle:
In order to measure vibrations accurately please mount the accelerometer with enough care. When it is not mounted and fixed
properly, its high frequency response and dynamic range will be limited not to provide reliable measurement of vibrations. In
mounting and fixing the accelerometer the mounting stiffness between accelerometer and vibrating body shall be increased.

◆ Fastening with Bolt


The mounting method with the bolt is the best solution. If this
method is acceptable to your application, we recommend it to Fixing Bolt
Fixing Bolt
all the applications. The mounting surface shall be flattened
Accelerometer
Accelerometer
out not so as to form any clearance and tip the accelerometer,
Mounting Surface Mounting Surface
then, please securely fasten the accelerometer with the bolt.
If the mounting surface is not quite smooth, the high
<Mounting with Bolt ①> <Mounting with Bolt ②>
frequency response may be improved by applying a thin layer If the accelerometer is the center stud If the accelerometer has a tapped
mount type, please provide a tapped hole at its bottom, please provide a
of silicon grease to the surface. hole in the mounting surface to fix it tapped hole in the mounting surface
with the fixing bolt. to fix it with the fixing bolt.

9
◆ Sticking Accelerometer with Adhesive Cement
If drilling a tapped hole for directly mounting the accelerometer to vibrating
Accelerometer
body is not available, we recommend fixing with an epoxy-based or
Hexagon
alpha-cyanoacrylate (instant) adhesive after cleaned the joint area to Cap Nut
remove dirt, liquid, oil etc.. Proper mounting using an adhesive can provide Bonding
Bonding
Plane Plane
the same frequency response as a bolt. An optional hexagon cap nut may
prevent the mounting surface or tapped hole from being damaged in <Mounting Method with Adhesive ①> <Mounting Method with Adhesive ②>
Directly bonding accelerometer to Bonding through the hexagon cap nut.
disconnecting the stuck accelerometer. It has a tapped hole for attaching mounting surface The tapped hole is provided in its top
surface and its bottom surface is flat
the accelerometer at one side, while the other side is flat to attach to a
for bonding.
vibrating object; furthermore, additional machining to the flat surface is
Insulating <Mounting Method with Adhesive ③>
also available. We do not recommend the soft adhesive because its high Plate
In bonding the accelerometer, after
frequency response becomes worse. Since the mounting stiffness with Bonding insulated, to the mounting surface, an
Plane insulating plate such as ceramic,
the adhesive may, beyond its rated temperature, be drastically reduced or
epoxy plate, etc. is put between them.
the stuck accelerometer may break away, please take care of the
performance guarantee temperature condition of adhesives.

Fixing Nut
Accelerometer
Accelerometer

Bonding Bonding
Insulated Plane Insulated Plane
Mounting Stud
Mounting Stud
<Mounting Method with Insulated Mounting Stud ①>
<Mounting Method with Insulated Mounting Stud ②>
If the accelerometer has a tapped hole at its bottom, it can be
If the accelerometer is a center stud mounting type, it can be
mounted after provided a tapped hole in the mounting surface by
mounted after provided a tapped hole in the mounting surface by
electrically insulating it with a dedicated insulated mounting stud.
electrically insulating it with the dedicated insulated mounting stud.

◆ Electrical Insulation
The accelerometer shall be isolated from a measuring object when the noise comes to be mixed in due to the difference in electric potential between
measuring object and instrument or due to ground loop formed by direct mounting of the accelerometer. To mount the accelerometer by isolating from
the measuring object the insulated mounting stud (option) or screw/washer shall be employed. Joining with an adhesive through ceramic or epoxy
insulator is also acceptable. Direct mounting method using an adhesive may break down the electric insulation due to thickness of bonding layer or
surrounding moisture. In trying to isolate and mount the accelerometer, please use the dedicated insulated mounting stud or joint it through the
insulator.

◆ Fixing with Wax


When the same accelerometer should be used at many measuring points, wax makes it easier to attach and remove that. Its operating temperature
range, however, should be limited to lower than 40℃ or app. because the stiffness of wax decreases at higher temperature. It is limited to the application
using a light accelerometer or measuring relative low acceleration.

◆ Mounting with Magnet Holder


If the mounting surface of the measuring object is ferromagnetic material, the accelerometer can be mounted with a magnet holder. Its mounting
stiffness depends on the condition of the mounting surface, total mass of the accelerometer and magnet holder, and magnetic force.

<Connection Treatment of Low-noise Cable>


Figure 1
When accelerometer cable bends or becomes deformed mechanically, the noise signal due to static electricity is generated
because the output impedance of the piezoelectric accelerometer is very high. When the coaxial cable is mechanically shield core insulator

distorted or twisted, small separations between the shield and insulator layer will occur as shown in Figure-1 and a
capacitor will be formed locally. This part becomes electrically charged through friction. If the charge amplifier receives this -
- - -
-
+ +
charge, it generates errors because it cannot be separated from the correct measurement charge. This phenomenon is ++ +

known as the so-called triboelectric effect. To reduce this effect as much as possible the surface of an insulator in a special
low-noise cable is processed with a conductive film (carbon layer). As this is the case, the capacitor cannot be formed in
the low-noise cable even if the insulator moves away from the shield conductor to drastically reduce the charge due to the
triboelectric effect. The low-noise cable is most widely used with piezoelectric accelerometers. Commonly used instrument
coaxial cables are not suitable for this application. Even if the low-noise cable is used, it shall be clamped as firmly as
possible to the vibrating specimen so that it may not be subjected to any mechanical force or may not whip. The accelerom-
eter with built-in electronics does not always have to use the low-noise cable because its output impedance is low.

<Instructions on Handling Accelerometers>


In particular, if failed to follow the ground rules in the below, the accelerometer would be failure or damaged. Please handle it with enough care.
※Do not use it beyond its operating temperature range. ※Do not use the general-purpose accelerometer in water, oil, and other liquid. However, a waterproof type is also available. Please
contact us. ※When it is dropped or struck it will be subjected to higher acceleration than its shock limit. Take extra care not so as to apply the excessive shock force to it. ※Firmly fix the
accelerometer not so as to be loosened. And tighten the connector not so as to be loosened. ※Do not apply the acceleration of more than its rated vibration limit. Even low acceleration whose
frequency is close to the resonance frequency of the accelerometer may cause it to be failure or damaged. Take extra care. ※Take care not to intrude moisture, oil content, adhesive, solvent
into the inside of a connector. ※Clamp the cable to the measuring object because it may generate noise or break when excessive force or vibration is applied to the cable. ※The sensitive axis
of the accelerometer is uniaxial except for triaxial type. And the mounting or joining surface of the accelerometer cannot be changed, therefore, mount it to take the sensitive direction. ※When
the accelerometer is mounted using an alpha-cyanoacrylate (instant) adhesive, use a solvent such as acetone solution or special depleting agent to disconnect it. ※Do not use the
accelerometer with built-in electronics beyond its rated voltage range. ※Contact us or our representative in advance if your application shall be performed in unusual environments.

10
540 Series Piezoelectric Type Accelerometers
Charge output being proportional to acceleration with the piezoelectric element in the acceler-
ometers.
Correspondence about small size, lightweight, high sensitivity, wide frequency range, and
acceleration from ultra-low level to high-level is possible.
Available for various vibration measurements including the sophisticated vibration/shock mea-
surement and test control.

Small/ Light/ Standard Top


Category Small/ Light/ High Frequency Small/ Light/ Female Screw
High Frequency Connector
Model 540 - E 540 - DA 540 - DS 540 - DAF 540 - DSF 541- DT

Outline

Hex14
Hex12

Hex12
∅6.8
Hex12

Hex12

Microdot Connector
Microdot Connector Microdot Connector (10-32UNF)
10.5

(10-32UNF) (10-32UNF) ∅13.8


5.5

(4.5)
∅11.7 ∅11.7
Microdot Connector Microdot Connector
(10-32UNF) (10-32UNF)

15
Overall Size (mm) Mini-microdot ∅11.7 ∅11.7
(11)

(11)

25
Connector
20

20
3

9
16
16
Hex7

M6, Depth: 5 M6, Depth: 5


5

5
1.5

1.5

M3, Depth: 1.6


M6 M6

M6, Depth: 5

Dimensions HEX7×10.5mm HEX12×16mm HEX12×16mm HEX12×20mm HEX12×20mm HEX14×25mm

Titan case, small, Aluminum case, Stainless case, Aluminum case, Stainless case, Stainless case,
Construction light, most suitable small, light, male s m a l l , m a l e s m a l l , l i g h t , small, female small, female
& Application
for light object screw screw female screw screw screw

Sinusoidal Vibration Limit 49000m/s2 49000m/s2 16300m/s2

Shock Limit 50000m/s2 50000m/s2 16000m/s2

Mass 2g 5g 10g 6.5g 13g 32g

Frequency Response 20k Hz ※1 10k Hz ※1 8k Hz ※1

Charge Sensitivity pC/(m/s2) 0.3±20% 1.0±15% 5.0±15%

Mounted Resonance about 70 kHz about 60 kHz about 40 kHz

Upper Limit Temperature (from -20℃) +150 ℃ +140 ℃

Mode Design Shear Compression

Capacitance 900pF 500pF 1000pF

Transverse Sensitivity Less than 5 %

Piezoelectric Material Pb(Z・Ti)O3

Case Material Titan Aluminum Stainless Aluminum Stainless

Mounting Method Female M3D 1.6 Male M6P1L5 Female M6P1D5


Accessory Cable & Adapter AC7020-BM BLM-001 AC8020-AB
(Microdot-BNC)
※1 Lower limit of frequency response depends on an available charge amplifier.

11
Standard Side Standard Side Standard High Standard Super-High
Category Standard
Connector Connector Temperature High Sensitivity Sensitivity
Model 541- DS 541- DSH 541- D4S 541- D4SH 541- ES 546 - D

Outline

Microdot Connector Microdot Connector Microdot Connector Microdot Connector Microdot Connector Microdot Connector
(10-32UNF) (10-32UNF) (10-32UNF) (10-32UNF) (10-32UNF) (10-32UNF)
Hex14

Hex14

Hex14

Hex14

Hex14

(4.5) Hex38
(4.5)

(4.5)

(4.5)

(4.5)

(4.5)
∅13.7
∅13.8 ∅13.8 ∅13.8 ∅13.8
∅36

(17)
(16)

(16)

(16)

(16)

(30)
Overall Size (mm)
32
30

29

30

29

65
35
15

27
14

13

14

13
9.5

8.5

9.5

8.5

M6, Depth: 5 M6, Depth: 5 M4, Depth: 5 M4, Depth: 5 M6, Depth: 5 9 M8, Depth: 8

Dimensions HEX14×30mm HEX14×29mm HEX14×30mm HEX14×29mm HEX14×32mm HEX38×65mm

S t a i n l e s s c a s e , S t a i n l e s s c a s e , S t a i n l e s s c a s e , Stainless case, standard, S t a i n l e s s c a s e , S t a i n l e s s c a s e ,


Construction standard, female screw, standard, female screw, standard, female female screw, high acceler- s t a n d a r d , f e m a l e female screw, ultra-
& Application
high acceleration high temperature screw ation, high temperature screw, high sensitivity high sensitivity

Sinusoidal Vibration Limit 16300m/s2 9800m/s2 16300m/s2 9800m/s2 980 m/s2

Shock Limit 16000m/s2 10000m/s2 1000m/s2

Mass 35g 43g 500g

Frequency Response 7k Hz ※1 5k Hz ※1 2k Hz ※1

Charge Sensitivity pC/(m/s ) 2


5.0±15% 5.0±20% 5.0±15% 5.0±20% 10±15% 500±15%

Mounted Resonance about 30 kHz about 25 kHz about 23 kHz about 7 kHz

Upper Limit Temperature (from -20℃) +140 ℃ +250 ℃ +140 ℃ +250 ℃ +140 ℃ +80 ℃

Mode Design Compression

Capacitance 1000pF 10000pF

Transverse Sensitivity Less than 5 %

Piezoelectric Material Pb(Z・Ti)O3

Case Material Stainless

Mounting Method Female M6P1D5 Female M4P0.7D5 Female M6P1D5 Female M8P1.25D8
Accessory Cable & Adapter AC8020-AB(H) for High AC8020-AB(H) for High
(Microdot-BNC) AC8020-AB AC8020-AB AC8020-AB
Temperature Temperature
※1 Lower limit of frequency response depends on an available charge amplifier.

12
700 Series Piezoelectric Type Accelerometers
The piezoelectric accelerometer is used in combination with the charge amplifier. The 700
Series Accelerometers employ the shear mode construction for the piezoelectric sensing
element, the shear mode which provides a stable output against environments of rapid tem-
perature change. This stable characteristic is suitable for a wide range of application, in partic-
ular, for measuring the vibration mode of small and lightweight material, and low acceleration
of a building.

Category Small/Light Triaxial

Model 701- B 710 - D 702 - B3 703 - B3 712 - B3

Outline

11
Y-axis 7
φ0.8 Low-noise Cable 17.5
Y-axis

2.25±0.1
LN-008(Black), X-axis 11

15.25 (4.5)
Length: 0.3 m×3 2.25±0.1
Hex 6

10.8 (3.1) Microdot Jack on One End Y

7
11
X Y
z
8

11
17.5
Y Z X
∅8

Y
Y
7

X X-axis
3.8

X
X
2-∅2.2 15.25±0.1(4.5)
Tube Y
Mini-microdot Connector
Z-axis (Mini-microdot Connector) X
(5-44UNF) Mini-microdot Connector
2 2 (5-44UNF)
Overall Size (mm) (5-44UNF) Z
5.5

Z
X Z-axis (Mini-microdot Connector)
0.5

∅6.4 ∅6
(3.7)

0.4 (4.1) 0.5

(10-32UNF)

9 (4.5)
Microdot Jack ∅9
3.7
5

Identification Mark for


X, Y and Z respectively
4.5

2.5

Z
∅5

4.5
300 X
Z Z

X
∅5 Black Tube

Dimensions 6HEX×8.2mm φ8×5mm 8×7×5.5mm 11×11×7mm 17.5×17.5×9mm


Ultrasmall, light, most Center stud mounting, Ultrasmall, simultaneous Ultrasmall, simulta- Small, high sensitivi-
Construction suitable for light mea- side connector, suitable triaxial measurement of neous triaxial mea- ty, simultaneous
& Application
suring object for modal analysis lightweight material surement triaxial measurement

Sinusoidal Vibration Limit 5000m/s2

Shock Limit 10k m/s2

Mass 0.7g 1.9g 1.2g 3.7g 14g

Frequency Response up to 10k Hz±1dB※1 up to 20k Hz±3dB※1 up to 20k Hz±3dB※1 up to 12k Hz±3dB※1 up to 8k Hz±1dB※1

Charge Sensitivity pC/(m/s2) 0.061±20% 0.2±15% 0.04±25% 0.061±20% 0.347±20%

Mounted Resonance about 60 kHz about 35 kHz about 25 kHz

Upper Limit Temperature (from -20℃) -50 to +160 ℃

Mode Design Shear Annular Shear Shear

Capacitance 650pF±25% 1200pF±20% 250pF±25% 650pF±25% 750pF±25%

Transverse Sensitivity Less than 5 %

Piezoelectric Material Pb(Zr・Ti)O3

Case Material Titan Stainless Titan

Mounting Method Adhesive Thru for M2, adhesive Adhesive Thru for M2, adhesive
Accessory Cable & Adapter AC-7020-BM(BLM-001) AC-8030-AB×3 AC-7020-BM(BLM-001)×3 AC-8030-AB×3
(Microdot-BNC)
※1 Lower limit of frequency response depends on an available charge amplifier.

13
Description Model Length Outline View

Microdot Plug - BNC Plug 10-32UNF BNC


AC-8030-AB 3m Low-noise Cable

Microdot Plug - Microdot Plug


Accelerometer Cable AC-8030-AM 3m 10-32UNF 10-32UNF
Low-noise Cable

Mini-microdot Plug - Microdot Plug 5-44UNF 10-32UNF


AC-7020-BM 2m Low-noise Cable

Microdot Jack - BNC Plug 10-32UNF


AC-8030-CB 3m Low-noise Cable female
BNC

Extension Cable
Microdot Jack - Microdot Plug 10-32UNF
AC-8030-CM 3m Low-noise Cable female
10-32UNF

Category Water-proof General Purpose Large Output

Model 720 - BW 730 - B 731- B 740 - B 760 - B

Outline

Hex24
Hex17

18.5 (4.5)

7
18.2 (22)

x1
He

(4.5)
∅15

Microdot Connector Microdot Connector


(10-32UNF) (10-32UNF)
Microdot Connector
∅4.3 ∅16.8 (25)
Microdot Connector (10-32UNF)
(10-32UNF)
(5.5)
8

∅23.8
0.5

∅17.5

∅10 ∅16.8
4.5

16

Overall Size (mm)

(11) 19
4.5
0.5(7.5)4

Microdot Connector on One End ∅4.4


12

30
(10-32UNF)
0.5

6
∅2 ∅10 ∅10
φ2 Low-noise Cable 18.7
9.8
0.5

LN-020
5

M5 P=0.8 ∅10
1000 Depth: 5
Hex17

M5 P=0.8
M8 P=1.25
Depth: 5
Depth: 5

Dimensions φ15×8mm 17HEX×21.7mm φ17.5×9.8mm 17HEX×12mm 24HEX×30mm


Center stud mounting, for small Top connector type, Center stud mount- Side connector, for For measuring low
Construction place, waterproof of more than most suitable for multi- ing, side connector, small place acceleration of building
& Application
10 kg/cm2 at room temperature point measurement easy cabling and structural object

Sinusoidal Vibration Limit 5000m/s2 1250m/s2

Shock Limit 10k m/s2 15k m/s2 10k m/s2 2.5k m/s2

Mass 11g 13.8g 13.5g 15.2g 96.6g

Frequency Response up to 8k Hz±1dB※1 up to 7k Hz±1dB※1 up to 3.5k Hz±1dB※1

Charge Sensitivity pC/(m/s2) 1.33±20% 2.04±20% 3.67±20% 4.28±20% 35±20%

Mounted Resonance about 26 kHz about 37k Hz about 38k Hz±5 about 25k Hz±4 about 13.5k Hz±4

Upper Limit Temperature (from -20℃) -20 to +120 ℃ -50 to +160 ℃ -20 to +120 ℃

Mode Design Annular Shear Shear Annular Shear Shear

Capacitance 1200pF±25% 900pF 1600pF±25% 900pF±25% 1500pF±25%

Transverse Sensitivity Less than 5 %

Piezoelectric Material Pb(Zr・Ti)


O3

Case Material Stainless Titan Stainless

Mounting Method M4 thru M5 Screw M4 thru M5 Screw M8 Screw


Accessory Cable & Adapter Integral: 10 m
(Microdot-BNC) w/BNC Plug AC-8030-AB
※1 Lower limit of frequency response depends on an available charge amplifier.

14
Head office :
Shinagawa-ku, Tokyo 141-0031
Tel: +81-3-3494-1221
http://www.emic-net.co.jp/

A-PLACE Gotanda Bldg., 3F, 2-27-3 Nishi-Gotanda,

FAX: +81-3-3494-1288
Nagoya office : Tomei Grand Bldg., 6F, 30 Isha 2-chome Mitoh-ku,
Contact:
MEASURING
MicrosoftⓇ, WindowsⓇ, and WindowsⓇ 7 Professional are a registered trademark of Microsoft.

Nagoya, Aichi-prefecture 465-0093


Tel: +81-52-753-6308 FAX: +81-52-753-6328
Osaka office : Hanahara 5 Bldg., 6 Fl., 8-17 Nishinakajima 7-chome
Yodogawa-ku, Osaka, 532-0011
Tel: +81-6-6886-0451 FAX: +81-6-6886-0454
Mishima factory : 11 Heiseidai Mishima, Shizuoka 411-0042
Tel: +81-55-988-8411 FAX: +81-55-988-2223
Service Centre : 11 Heiseidai Mishima, Shizuoka 411-0042
Tel: +81-55-988-8411 FAX: +81-55-987-1477

※ Specifications are subject to change without notice for improvement.

Published: June 2016


CL-173-01-E
(PreCL-106-08-J)