Ralco: Operating Instructions
Ralco: Operating Instructions
Plug located on the right side of the collimator Pin 1 and 2 - free
(see Figure 3 page 13of this Manual):
Pin 3 + 20 volts AC/DC Lamp
Pin 5 + push-button
Pin 6 - push-button
OPERATING INSTRUCTIONS:
By pressing the area marked with the light symbol the light field, simulating the X-Ray field,
and a laser light, aligning the collimator with the Potter Bucky, will operate for 30 seconds.
The laser light can be screened by the displacement of the plastic door.
The collimator can be rotated manually plus/minus 45°. Perform the rotation by turning the
collimator.
The unit has variable filtration: Select the desired filtration by rotating the black disk
protruding on the left side of the collimator.
so that the inner coin is lighted as completely as possible and the outer coin is lighted as
little as possible. The points of tangency will define the edges of the light field. The extra
four coins in the lower left corner will provide the means for orienting the film.
Set the technique factors at the X-Ray generators to produce a density of about 1 (about
50 kVp, 5 mAs).
Make an exposure, remove the cassette and process the film. Use the test film to check
the following alignment requirements:
NOTE: Because of heel effect, the field toward the cathode will be slightly less sharp
than the other three sides . This is normal and cannot be corrected by adjustment. In
addition, an X-Ray tube of 12 ° or less target angle will produce an asymmetrically
shaped field when a large field size is used at short FFD (SID), because of anode
cutoff effect. This is normal and may not be corrected by adjustment.
Regulations state that the total misalignment of the light field to the X-Ray field in either the X
(cross table) or Y (long table) direction must be less than 2% of the FFD. In this case, that is
less than 20 mm (0.80"). If the test film shows that the light field, shown by the coins
shadows, is congruent to the X-Ray field, shown by the shadow of the collimator shutters, to
within one coin diameter, if the diameter is less than 20 mm., then the alignment complies
with the regulations.
However, the degree of precision required by the regulation is of a low order, and a much
more precise degree of congruency can be obtained. The recommended maximum deviation
is one fourth; you are urged to adjust for the greatest obtainable degree of congruency.
If misalignment is detected in both X and Y directions, check that the spacing from focal spot
to collimator mounting surface is 80 mm. +/- 1 mm. (3.15" +/- .039"). If the spacing must be
changed, repeat the test film exposure after the adjustment.
If the collimator mount spacing is correct, but adjustment is still required, do not move the
collimator relative to the X-Ray tube, but proceed as follows:
Place the test film on the face of the cassette over the white paper and place the cassette
in the position originally marked.
Check the correct orientation of the film by the coin shadows.
Using the images of the collimator shutters as reference for the shape and size of the X-
Ray field, adjust the light field to be congruent to it.
To move the light field in the long table direction, first remove the front top cover. In the
top right side on the front of the collimator are located a screw head, locking the mirror to
the side of the collimator, and two adjuster cams. Loosen the screws, one locking the