0% found this document useful (0 votes)
587 views5 pages

Standard Operating Procedure (SOP) For Shimadzu UV-2600 UV-VIS Spectrophotometer

This document provides the standard operating procedure for using a Shimadzu UV-2600 UV-VIS spectrophotometer. It outlines 16 steps for instrument startup, zeroing the instrument, collecting a baseline, running samples, saving data files, and shutting down. Key steps include connecting to the instrument software, setting method parameters like wavelength range, zeroing the instrument, collecting a baseline with a reference sample, running samples and saving spectra individually with file names and comments. [END SUMMARY]
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
587 views5 pages

Standard Operating Procedure (SOP) For Shimadzu UV-2600 UV-VIS Spectrophotometer

This document provides the standard operating procedure for using a Shimadzu UV-2600 UV-VIS spectrophotometer. It outlines 16 steps for instrument startup, zeroing the instrument, collecting a baseline, running samples, saving data files, and shutting down. Key steps include connecting to the instrument software, setting method parameters like wavelength range, zeroing the instrument, collecting a baseline with a reference sample, running samples and saving spectra individually with file names and comments. [END SUMMARY]
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

Standard Operating Procedure (SOP) for Shimadzu UV-2600 UV-VIS Spectrophotometer

1. Turn of the UV-2600 UV-VIS spectrophotometer.

2. Launche the UVProbe 2.43 software

3. Click the Connect button on the bottom of the window and wait until it has performed all of its
tests (they should all pass and have green dots next to them.

All passed

click OK

4. Click on the Edit method tab in the tool bar (the green M in a yellow circle)
5. Variables to change will typically only be wavelength range

If other parameters need to be changed we will discuss it.

6. Zero the instrument by clicking the Auto Zero button at the bottom of the window.

7. Create a baseline using a solution that contains everything in your sample besides the analyte
(or as best that you can). This is your reference.
Put your reference solution into a cuvette and place it into the front sample holder in the
instrument. Make sure that clear sides of the cuvette is placed into the instrument so that the
light will pass through (notice that the cuvettes have two clear and two semi-opaque sides, be
careful to use the clear sides). With the reference in the sample holder, close the door and click
the Baseline button next to the Auto Zero.

8. When the baseline scan is done you will hear a prompt. Open the door and remove your
reference and place your first sample into the same sample holder and click the Start button
two over from the Baseline button. Once again make sure the top door is closed.
9. When the spectrum has been taken and converted into absorbance units you will see a new
window pop up:

This is where you will give the sample a file name. To the right of the File blank is three dots,
click this to make sure you are storing your data where you want. It should be stored in the data
folder, if you want you can back a folder within the data folder with your name on it.

And fill out the comment section with information about the sample.
Note: this does not save your data just sets up the raw data to be manipulated or saved as the
file time you so choose.

10. Now your raw data will be displayed on the graph on the right hand side of the window. As long
as you do not change any parameters or use a different solvent for your sample you can use the
background you have already generated and run more samples by putting in your new sample
and hitting the Start button. This will place all of your raw data onto one graph. Remember to
save each spectrum as with its own file name and comment.
11. Once you have taken some spectra your window will look something like this:

Notice that many of the solutions were too concentrated, if you have an absorbance that is
greater than 2 you are much too concentrated. An absorbance of 2 is telling you that 99% of the
light was attinuated, or another way of putting it, only 1% of the light made it through. Notice
that all spectra are displayed. If you want to look at only one you can either uncheck all of the
spectra but the one you want or better way is to click on the graph menu at the top and click the
active button which will only display the spectrum that is active. While in the active graph setup,
to observe another spectrum simply double click on its name in the list to the right of the active
spectrum.

12. We want to save the data in two formates, one that can be displayed using this software and
one that you can import into Excell. While in the active spectrum mode and having only one
spectrum displayed, click on file and save as, make sure you are in the folder that you want your
data saved in then save the file (already named) as “Save as type: Spectrum Files (*.spc)”. Then
do it again and save as “Save as type: Data Print Table (*.txt). This last one will be able to be
opened using Excell.
Do both of theses saves to all of your data.

13. There are a number of things you can do to manpulate graph displayed. Right click on the graph
and there are two features to take note of; Auto scale and cross hair. Auto Scale will make sure
all of the data is displayed and cross hair lets you move around within the window and see what
wavelength and abosrbance a peak has.

14. Your data will be stored somewhere near here:


C:\Program Files (x86)\Shimadzu\UVProbe\Data\Eddingsaas
You should transfer all of your saved data onto a USB flash drive.

15. When done taking data click on the Disconnect button at the bottom of the window.

16. After disconnecting you can exit the program then turn off the instrument.

You might also like