Chapter 4
Chapter 4
22 710 300 53
25 600 350 45 For aperture sizes above about 1 mm, perforated plate
30 500 400 38
sieves are often used, with round or square holes
(Figure 4.3). Square holes are arranged in line with the
center points at the vertices of squares, while round holes
are arranged with the centers at the apices of equilateral
triangles (Anon., 2000b).
nominal aperture size. A fuller comparison of several
standards is given in Napier-Munn et al. (1996).
Wire-cloth screens are woven to produce nominally
4.3.2 Choice of Sieve Sizes
uniform square apertures within required tolerances
(Anon., 2000a). Wire cloth in sieves with a nominal aper- In each of the standard series, the apertures of consecutive
ture of 75 μm and greater are plain woven, while those in sieves bear a constant relationship to each other. It has
cloths with apertures below 63 μm may be twilled long been realized that a useful sieve scale is one in
(Figure 4.1). which the ratio of the aperture widths of adjacent sieves
Standard test sieves are not available with aperture is the square root of 2 (O2 5 1.414). The advantage of
sizes smaller than about 20 μm. Micromesh sieves are such a scale is that the aperture areas double at each
available in aperture sizes from 2 μm to 150 μm, and are sieve, facilitating graphical presentation of results with
made by electroforming nickel in square and circular particle size on a log scale.
mesh. Another popular type is the “micro-plate sieve,” Most modern sieve series are based on a fourth root of
which is fabricated by electroetching a nickel plate. The 2 ratio (4O2 5 1.189) or, on the metric scale, a tenth root
apertures are in the form of truncated cones with the small of 10 (10O10 5 1.259), which makes possible much closer
circle uppermost (Figure 4.2). This reduces blinding (a sizing of particles.
particle passing the upper opening falls away unhindered) For most size analyses it is usually impracticable and
but also reduces the percentage open area, that is, the per- unnecessary to use all the sieves in a particular series. For
centage of the total area of the sieving medium occupied most purposes, alternative sieves, that is, a O2 series, are
by the apertures. quite adequate. However, over certain size pffiffiffi ranges of par-
Micro-sieves are used for wet or dry sieving where ticular interest, or for accurate work the 4 2 series may be
accuracy is required in particle size analysis down to the used. Intermediate sieves should never be chosen at ran-
very fine size range (Finch and Leroux, 1982). The toler- dom, as the data obtained will be difficult to interpret.
ances in these sieves are much better than those for In general, the sieve range should be chosen such that
woven-wire sieves, the aperture being guaranteed to no more than about 5% of the sample is retained on the
within 2 μm of nominal size. coarsest sieve, or passes the finest sieve. (The latter
94 Wills’ Mineral Processing Technology
FIGURE 4.3 Arrangement of square and round holes in perforated plate sieves.
Sieve Size Range (μm) Sieve Fraction Nominal Aperture Size (μm) Cumulative (%)
aperture, and brushing, even with a soft hair brush, of The results of a sieving test should always be plotted
sieves finer than about 150 μm aperture tends to distort graphically to assess their full significance (Napier-Munn
the individual meshes. et al., 1996).
Wet sieving can be used on material already in the There are many ways of recording the results of sieve
form of slurry, or it may be necessary for powders which analysis, the most common being that of plotting cumula-
form aggregates when dry-sieved. A full description of tive undersize (or oversize) against particle size, that is,
the techniques is given in BS 1796 (Anon., 1989b). the screen aperture. Although arithmetic graph paper can
Water is the liquid most frequently used in wet sieving, be used, it suffers from the disadvantage that points in the
although for materials which are water-repellent, such as region of the finer aperture sizes become congested. A
coal or some sulfide ores, a wetting agent may be necessary. semi-logarithmic plot avoids this, with a linear ordinate
The test sample may be washed down through a nest for percentage oversize or undersize and a logarithmic
of sieves. At the completion of the test the sieves, abscissa for particle size. Figure 4.6 shows graphically the
together with the retained oversize material, are dried at a results of the sieve test tabulated in Table 4.3.
suitable low temperature and weighed. It is not necessary to plot both cumulative oversize
(or coarser than) and undersize (finer than) curves, as
they are mirror images of each other. A valuable quan-
4.3.4 Presentation of Results tity that can be determined from such curves is the
“median size” of the sample. This refers to the mid-
There are several ways in which the results of a sieve test
point in the size distribution, X50, where 50% of the par-
can be tabulated. The three most convenient methods are
ticles are smaller than this size and 50% are larger, also
shown in Table 4.3 (Anon., 1989b). Table 4.3 shows for
called the 50% passing size.
the numbered columns:
Size analysis is important in assessing the performance
1. The sieve size ranges used in the test. of grinding circuits. The product size is usually quoted
2. The weight of material in each size range. For exam- in terms of one point on the cumulative undersize curve,
ple, 1.32 g of material passed through the 250 μm this often being the 80% passing size, X80. On Figure 4.6
sieve, but was retained on the 180 μm sieve: the mate- the X80 is about 110 μm. In a grinding circuit it is com-
rial therefore is in the size range 2250 to 1180 μm. mon to refer to the 80% passing size of the feed and
3. The weight of material in each size range expressed as product as F80 and P80, respectively, and to T80 as the
a percentage of the total weight. transfer size between grinding units (e.g., SAG mill
4. The nominal aperture sizes of the sieves used in the test. product going to a ball mill). Although the X80 does not
5. The cumulative percentage of material passing show the overall size distribution of the material, it does
through the sieves. For example, 87.5% of the material facilitate routine control of the grinding circuit. For
is less than 125 μm in size. instance, if the target size is 80% 2250 μm (80% finer
6. The cumulative percentage of material retained on the than 250 μm), then for routine control the operator needs
sieves. only screen a fraction of the mill product at one size. If
96 Wills’ Mineral Processing Technology
FIGURE 4.6 Sieve analysis presented as semi-log plot (Table 4.3). FIGURE 4.7 Example of Gates-Gaudin-Schuhmann type plot for the
set of laboratory grind times in Appendix IV.
it is found that, say, 70% of the sample is 2250 μm, plant). An example G-G-S plot is shown in Figure 4.7 for
then the product is too coarse, and control steps to increased grinding time in a laboratory ball mill (data
remedy this can be made. given in Appendix IV). The figure shows a series of
Many curves of cumulative oversize or undersize roughly parallel (or self-similar) linear sections of the size
against particle size are S-shaped, leading to congested distributions. This result, showing a consistent behavior,
plots at the extremities of the graph. More than a dozen is quite common in grinding (and other) studies (but see
methods of plotting in order to proportion the ordinate are Chapter 5 for exceptions) and supports the use of a single
known. The two most common methods, which are often metric to describe the distribution, for example, the 80%
applied to comminution studies where non-uniform size passing size, X80 (Austin et al., 1984).
distributions are obtained, are the Gates-Gaudin- Self-similar distributions such as in Figure 4.7
Schuhmann (Schuhmann, 1940) and the RosinRammler become a unique distribution by plotting against a
(Rosin and Rammler, 1933) methods. Both methods are reduced particle size, typically X/X50 where X50 is the
derived from attempts to represent particle size distribu- 50% passing size or median size. If it is known that data
tion curves by means of equations. This results in scales obtained from the material usually yield a linear plot on
which, relative to a linear scale, are expanded in some loglog axes, then the burden of routine analysis can be
regions and contracted in others. greatly relieved, as relatively few sieves will be needed
In the Gates-Gaudin-Schuhmann (G-G-S) method, to check essential features of the size distribution. Self-
cumulative undersize data are plotted against sieve aper- similarity can be quickly identified, as a plot of, say,
ture on loglog axes. This frequently leads to a linear % 2 75 μm versus % 1150 μm will yield a single curve
trend from which data can be interpolated easily. The lin- if this property holds.
ear trend is fitted to the following: Plotting on a loglog scale considerably expands
the region below 50% in the cumulative undersize
X
logP 5 α log 1 log100 (4.1) curve, especially that below 25%. It does, however,
K
severely contract the region above 50%, and especially
which can be written as: above 75%, which is a disadvantage of the method
α (Figure 4.8).
X The RosinRammler method is often used for repre-
P 5 100 (4.2)
K senting the results of sieve analyses performed on mate-
rial which has been ground in ball mills. Such products
where P is the cumulative undersize (or passing) in per- have been found to obey the following relationship
cent, X the particle size, K the apparent (theoretical) top (Austin et al., 1984):
size (i.e., when P 5 100%) obtained by extrapolation, and
α a constant sometimes referred to as the distribution 100 2 P 5 100 expðbXÞn (4.3)
coefficient (Austin et al., 1984). This method of plotting
is routinely used in grinding studies (laboratory and where b and n are constants.
Particle Size Analysis Chapter | 4 97
TABLE 4.4 Results of Screen Analysis to Evaluate the Suitability for Treatment by Gravity Concentration
(1) Size Range (μm) (2) Weight (%) (3) Assay (% Sn) Distribution Size (μm) Cumulative Cumulative
(% Sn) oversize (Wt %) distribution (% Sn)
1422 9.7 0.02 0.9 422 9.7 0.9
2422 1 300 4.9 0.05 1.2 300 14.6 2.1
2300 1 210 10.3 0.05 2.5 210 24.9 4.6
2210 1 150 23.2 0.06 6.7 150 48.1 11.3
2150 1 124 16.4 0.12 9.5 124 64.5 20.8
2124 1 75 33.6 0.35 56.5 75 98.1 77.3
275 1.9 2.50 22.7
100.0 0.21 100.0
Table 4.4 shows the results of a screen analysis per- Conversion factors between methods will vary with
formed on an alluvial tin deposit for preliminary evaluation sample characteristics and conditions, and with size where
of its suitability for treatment by gravity concentration. the distributions are not self-similar. For spheres, many
Columns 1, 2, and 3 show the results of the sieve test and methods will give essentially the same result (Napier-
assays, which are evaluated in the other columns. It can be Munn, 1985), but for irregular particles this is not so.
seen that the calculated overall assay for the material is Some approximate factors for a given characteristic size
0.21% Sn, but that the bulk of the tin is present within the (e.g., X80) are given below (Austin and Shah, 1983;
finer fractions. The results show that, for instance, if the Napier-Munn, 1985; Anon., 1989b)—these should be
material was initially screened at 210 μm and the coarse used with caution:
fraction discarded, then the bulk required for further proces-
sing would be reduced by 24.9%, with a loss of only 4.6% Conversion Multiplying Factor
of the Sn. This may be acceptable if mineralogical analysis Sieve size to Stokes’ diameter 0.94
shows that the tin is finely disseminated in this coarse frac- (sedimentation, elutriation)
tion, which would necessitate extensive grinding to give
Sieve size to projected area diameter 1.4
reasonable liberation. Heavy liquid analysis (Chapter 11) on (microscopy)
the 2210 μm fraction would determine the theoretical (i.e.,
Sieve size to laser diffraction 1.5
liberation-limited, Chapter 1) grades and recoveries possi-
ble, but the screen analysis results also show that much of Square mesh sieves to round hole 1.2
the tin (22.7%) is present in the 275 μm fraction, which sieves
constitutes only 1.9% of the total bulk of the material. This
indicates that there may be difficulty in processing this 4.4.1 Stokes’ Equivalent Diameter
material, as gravity separation techniques are not very effi-
cient at such fine sizes (Chapters 1 and 10). In sedimentation techniques, the material to be sized is
dispersed in a fluid and allowed to settle under carefully
controlled conditions; in elutriation techniques, samples
4.4 SUB-SIEVE TECHNIQUES are sized by allowing the dispersed material to settle
Sieving is rarely carried out on a routine basis below against a rising fluid velocity. Both techniques separate
38 μm; below this size the operation is referred to as sub- the particles on the basis of resistance to motion in a
sieving. The most widely used methods are sedimentation, fluid. This resistance to motion determines the terminal
elutriation, microscopy, and laser diffraction, although velocity which the particle attains as it is allowed to fall
other techniques are available. in a fluid under the influence of gravity.
There are many concepts in use for designating parti- For particles within the sub-sieve range, the terminal
cle size within the sub-sieve range, and it is important to velocity is given by the equation derived by Stokes (1891):
be aware of them, particularly when combining size dis-
tributions determined by different methods. It is prefera-
d 2 gðρs 2 ρl Þ
ble to cover the range of a single distribution with a v5 (4.5)
single method, but this is not always possible. 18 η
Particle Size Analysis Chapter | 4 99
h=v1 d2 1
5 5 2 5
h=v d 2 d12
1 2 ðd1 =dÞ2
with the faster-settling particles, reporting to the apex The optical microscope method is applicable to parti-
opening, while the remainder of the liquid, together with cles in the size range 0.8150 μm, and down to 0.001 μm
the slower settling particles, is discharged through the using electron microscopy.
vortex outlet and into the next cyclone in the series. Basically, all microscopy methods are carried out on
There is a successive decrease in the inlet area and vortex extremely small laboratory samples, which must be col-
outlet diameter of each cyclone in the direction of the lected with great care in order to be representative of the
flow, resulting in a corresponding increase in inlet veloc- process stream under study.
ity and an increase in the centrifugal forces within the In manual optical microscopy, the dispersed particles
cyclone. This results in a successive decrease in the limit- are viewed by transmission, and the areas of the magni-
ing particle-separation size of the cyclones. fied images are compared with the areas of circles of
The cyclosizer is manufactured to have definite limit- known sizes inscribed on a graticule.
ing separation sizes at standard values of the operating The relative numbers of particles are determined in
variables, viz. water flow rate, water temperature, particle each of a series of size classes. These represent the size
density, and elutriation time. To correct for practical oper- distribution by number from which it is possible to calcu-
ation at other levels of these variables, a set of correction late the distribution by volume and, if all the particles
graphs is provided. have the same density, the distribution by weight.
Complete elutriation normally takes place after about Manual analysis of microscope slides is tedious and
20 min, at which time the sized fractions are collected by error prone; semi-automatic and automatic systems have
discharging the contents of each apex chamber into sepa- been developed which speed up analyses and reduce the
rate beakers. The samples are dried, and weighed to deter- tedium of manual methods (Allen, 1997).
mine the mass fractions of each particle size. The development of quantitative image analysis has
made possible the rapid sizing of fine particles. Image
analyzers accept samples in a variety of forms—photo-
4.4.4 Microscopic Sizing and Image Analysis graphs, electron micrographs, and direct viewing—and
are often integrated in system software. Figure 4.15 shows
Microscopy can be used as an absolute method of particle the grayscale electron backscatter image of a group of
size analysis, since it is the only method in which individ- mineral particles obtained with a scanning electron micro-
ual mineral particles are observed and measured (Anon., scope; grains of chalcopyrite (Ch), quartz (Qtz), and epi-
1993; Allen, 1997). The image of a particle seen in a dote (Epd) are identified in the image. On the right are
microscope is two-dimensional and from this image an plotted the size distributions of the “grains” of the mineral
estimate of particle size must be made. Microscopic siz- chalcopyrite (i.e., the pieces of chalcopyrite identified by
ing involves comparing the projected area of a particle the instrument, whether liberated or not) and the “parti-
with the areas of reference circles, or graticules, of known cles” in which the chalcopyrite is present. The plots are
sizes, and it is essential for meaningful results that the based on the analysis of several hundred thousand parti-
mean projected areas of the particles are representative of cles in the original sample, and are delivered automati-
the particle size. This requires a random orientation in cally by the system software. Image analysis of this kind
three dimensions of the particle on the microscope slide, is available in many forms for the calculation of many
which is unlikely in most cases. quantities (such as size, surface area, boundary lengths)
FIGURE 4.15 Using image analysis to estimate the size distributions of particles and mineral grains (Courtesy JKMRC and JKTech Pty Ltd).
104 Wills’ Mineral Processing Technology
FIGURE 4.17 The laser diffraction instrument principle (from Napier-Munn et al., 1996; Courtesy JKMRC, The University of Queensland).
for most imaging methods, for example, optical, electron. count representing the number of particles larger than
(See also Chapter 17.) some determinable volume proportional to the appropriate
threshold setting. By taking a series of counts at various
4.4.5 Electrical Impedance Method amplification and threshold settings, data are directly
The Beckman Coulter Counter makes use of current obtained for determining number frequency against vol-
changes in an electrical circuit produced by the presence ume, which can be used to determine the size distribution.
of a particle. The measuring system of a Coulter Counter As the instrument measures particle volume, the
is shown in Figure 4.16. equivalent diameter is calculated from that of a sphere of
The particles, suspended in a known volume of electri- the same volume, which can be a more relevant size mea-
cally conductive liquid, flow through a small aperture sure than some alternatives. The instrument is applicable
having an immersed electrode on either side; the particle in the range 0.41,200 μm.
concentration is such that the particles traverse the aper-
ture substantially one at a time.
4.4.6 Laser Diffraction Instruments
Each particle passage displaces electrolyte within the
aperture, momentarily changing the resistance between In recent years, several instruments based on the diffraction
the electrodes and producing a voltage pulse of magnitude of laser light by fine particles have become available,
proportional to particle volume. The resultant series of including the Malvern Master-Sizer and the Microtrac. The
pulses is electronically amplified, scaled, and counted. principle is illustrated in Figure 4.17. Laser light is passed
The amplified pulses are fed to a threshold circuit hav- through a dilute suspension of the particles which circulate
ing an adjustable screen-out voltage level, and those through an optical cell. The light is scattered by the parti-
pulses which reach or exceed this level are counted, this cles, which is detected by a solid state detector which
Particle Size Analysis Chapter | 4 105
FIGURE 4.21 The original and segmented images of crusher feed and product, with the estimated size distributions (Split Engineering’s Split-
Online system) (Courtesy Split Engineering).
analysis are modifications and improvements of previous Finch, J.A., Leroux, M., 1982. Fine sizing by cyclosizer and micro-sieve.
systems, or adaptations of off-line systems for on-line CIM Bull. 75 (839), 235240.
usage. Harris, C.C., 1971. Graphical presentation of size distribution data: An
assessment of current practice. Trans. Inst. Min. Metall., Sec. C. 80,
C133C139.
REFERENCES Hathaway, R.E., Guthnals, D.L., 1976. The continuous measurement of
particle size in fine slurry processes. CIM Bull. 766 (Feb.), 6471.
Allen, T., 1997. fifth ed. Particle Size Measurement, vol. 1. Chapman
Heywood, H., 1953. Fundamental principles of sub-sieve particle size
and Hall, London. UK.
measurement. Recent Developments in Mineral Dressing: a
Anon, 1989a. British Standard 1017-1: 1989. Sampling of coal and coke.
Symposium Arranged by the Institution. Institute of Mining and
Methods for sampling of coal.
Metallurgy (Great Britain), Imperial College of Science &
Anon, 1989b. British Standard 1796-1: 1989. ISO 2591-1: 1988. Test
Technology, London, England. pp.3158.
sieving. Methods using test sieves of woven wire cloth and perfo-
Kongas, M., Saloheimo, K., 2009. Online slurry particle size analyzers.
rated metal plate.
In: Malhotra, D., et al., (Eds.), Recent Advances in Mineral
Anon, 1993. British Standard 3406-4: 1993. Methods for the determina-
Processing Plant Design. SME, Littleton, CO, USA, pp. 419432.
tion of particle size distribution. Guide to microscope and image
Maron, R., et al., 2014. Low maintenance, real-time tracking of particle
analysis methods.
size in individual hydrocyclones for process optimization. Proc. 11th
Anon, 2000a. British Standard 410-1: 2000. ISO 3310-1: 2000. Test
Encuentro International de Mantenedores de Plantas Mineras
sieves. Technical requirements and testing. Test sieves of metal wire
(MAPLA) Conf., Paper: BIO535, Santiago, Chile. 111.
cloth.
Napier-Munn, T.J., 1985. The determination of the size distribution of
Anon, 2000b. British Standard 410-2: 2000. ISO 3310-2: 2000. Test
ferrosilicon powders. Powder Tech. 42 (3), 273276.
sieves. Technical requirements and testing. Test sieves of perforated
Napier-Munn, T.J., et al., 1996. Mineral Comminution Circuits: Their
metal plate.
Operation and Optimisation (Appendix 3). Julius Kruttschnitt
Anon, 2001a. British Standard ISO 13317-1: 2001. Determination of par-
Mineral Research Centre (JKMRC), The University of Queensland,
ticle size distribution by gravitational liquid sedimentation methods.
Brisbane, Australia.
General principles and guidelines.
Price, E.W., 1962. Stokes’ law and the Haultain infrasizer unit. Ind. Eng.
Anon, 2001b. British Standard ISO 13317-2: 2001. Determination of par-
Chem. Process Des. Dev. 1 (1), 7980.
ticle size distribution by gravitational liquid sedimentation methods.
Rosin, P., Rammler, E., 1933. The laws governing the fineness of pow-
Fixed pipette method.
dered coal. J. Institute Fuel. 7, 2936.
Austin, L.G., Shah, I., 1983. A method for inter-conversion of microtrac
Schuhmann, R., Jr., 1940. Principles of comminution, I-Size distribution and
and sieve size distributions. Powder Tech. 35 (2), 271278.
surface calculations. Technical Publication. No. 1189, AIME, 111.
Austin, L.G., et al., 1984. Process Engineering of Size Reduction: Ball
Smith, A., et al., 2010. Ultrasonic extinction for full concentration, real
Milling. AIME, New York, NY, USA.
time particle size analysis in the mining industry. Proc. 2nd
Bernhardt, C., 1994. Particle Size Analysis: Classification and
International Congress on Automation in the Mining Industry
Sedimentation Techniques, vol. 5. Chapman & Hall, London, UK.
(Automining), Santiago Chile. 110.
Cirulis, D., Russell, J., 2011. Cyclone monitoring system improves
Stokes, S.G.G., 1891. Mathematical and Physical Papers, vol. 3.
operations at KUC’s Copperton concentrator. Eng. Min. J (E&MJ).
Cambridge University Press Warehouse, London, UK.
212 (10), 4449.
Westendorf, M., et al., 2015. Managing cyclones: A valuable asset, the
Coghill, P.J., et al., 2002. On-line measurement of particle size in min-
Copper Mountain case study. Mining Eng. 67 (6), 2641.
eral slurries. Miner. Eng. 15 (1-2), 8390.