The document outlines preventative maintenance checklists for two types of nebulizers - jet nebulizers and ultrasonic nebulizers. The checklists include inspecting equipment for damage, cleaning interior and exterior components, checking for proper operation of controls and functions, and replacing filters or cleaning parts as needed to ensure safe and effective operation of the equipment.
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0 ratings0% found this document useful (0 votes)
502 views1 page
Nebulizer Preventative Maintenance
The document outlines preventative maintenance checklists for two types of nebulizers - jet nebulizers and ultrasonic nebulizers. The checklists include inspecting equipment for damage, cleaning interior and exterior components, checking for proper operation of controls and functions, and replacing filters or cleaning parts as needed to ensure safe and effective operation of the equipment.
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1
EQUIPMENT
Nebulizer
Preventative
Maintenance
Preventive
Maintenance
Checklist
Jet
Nebulizer
1. Inspect
exterior
of
equipment
for
damaged
or
missing
hardware.
2. Inspect
the
power
cord,
strain
relief
and
plug/s
for
any
signs
of
damage.
3. Turn
unit
off,
open
user
accessible
covers
and
inspect
unit
for
damage.
4. Clean
unit
interior
components
and
exterior
with
vacuum
or
compressed
air.
5. Inspect
interior
for
signs
of
corrosion
or
missing
hardware.
Repair
as
required.
6. Inspect
electrical
components
for
signs
of
excessive
heat
or
deterioration.
7. Inspect
water
jet
assembly
including
orifice.
Check
for
water
leakage.
8. Check
gaskets
and
O-‐rings.
9. Verify
correct
operation
of
all
buttons,
controls,
displays
and/or
indicators.
10. Verify
correct
operation
of
unit
in
all
functional
modalities.
11. Clean
exterior
of
unit
including
all
accessories,
cables,
controls
and
displays.
12. Change
air
intake
filter
when
it
is
gray
or
dirty.
Ultrasonic
Nebulizer:
1. Inspect
exterior
of
equipment
for
damaged
or
missing
hardware.
2. Inspect
the
power
cord,
strain
relief
and
plug/s
for
any
signs
of
damage.
3. Turn
unit
off,
open
user
accessible
covers
and
inspect
unit
for
damage.
4. Clean
unit
interior
components
and
exterior
with
vacuum
or
compressed
air.
5. Inspect
interior
for
signs
of
corrosion
or
missing
hardware.
Repair
as
required.
6. Inspect
electrical
components
for
signs
of
excessive
heat
or
deterioration.
7. Inspect
oxygen
inlet
orifice
for
restrictions.
8. Inspect
water
jet
assembly
including
orifice.
Check
for
water
leakage.
9. Check
gaskets
and
O-‐rings.
10. Verify
correct
ultrasonic
output
for
sharp
water
cone
and
misting.
11. Verify
correct
operation
of
low-‐water
cut-‐off
circuit.
12. Tune
RF
output
if
needed.
13. Verify
correct
operation
of
all
buttons,
controls,
displays
and/or
indicators.
14. Verify
correct
operation
of
unit
in
all
functional
modalities.
15. Clean
exterior
of
unit
including
all
accessories,
cables,
controls
and
displays.
16. Clean
air
filter.
17. Clean
nebulizer
chamber
and
lid
daily
or
between
each
patient.
18. Clean
hoses
and
tubes
daily.