Capture User Manual v3.0 Us-A
Capture User Manual v3.0 Us-A
Contact us
TECHNICAL SUPPORT and AFTER SALES
ADDRESS
Miniparc du Verger – Bat. H
1 rue Terre Neuve
91940 Les Ulis
FRANCE
https://support.m2m-ndt.com
www.m2m-ndt.com
© Eddyfi Europe
1, rue de terre neuve, Bâtiment H, 91940 Les Ulis, FRANCE
Eddyfi Europe. M2M, Mantis, Capture, Enlight and their associated logos are trademarks or registered
trademarks of Eddyfi Technologies in France and/or other countries. Eddyfi Technologies
reserves the right to change product offerings and specifications without notice.
CAPTURE_user_manual_v3.0_US-A | 2019-03-20
CAPTURE manual #3
Summary
INTRODUCTION .................................................................................................. 8
PART 1. CAPTURE GENERAL OVERVIEW ................................................. 10
1. HARDWARE OVERVIEW .......................................................................... 10
1.1 GEKKO overview ................................................................................. 10
1.2 MANTIS overview ................................................................................ 12
2. HOME PAGE .............................................................................................. 13
3. TOUCH SCREEN ....................................................................................... 14
3.1 Navigating the touch screen................................................................. 14
3.1.1 Top banner ................................................................................... 14
3.1.2 Backward/validation buttons ......................................................... 15
3.2 Touch screen tips ................................................................................. 16
3.2.1 Multiple-choice buttons ................................................................. 16
3.2.2 On-screen keyboard ..................................................................... 16
3.2.3 Invalid values ................................................................................ 17
4. GENERAL SETTINGS................................................................................ 18
4.1 System parameters .............................................................................. 19
4.2 UT preferences .................................................................................... 20
4.3 CAPTURE upgrade .............................................................................. 20
5. LAUNCH AN APPLICATION ..................................................................... 20
5.1 Select a Wizard .................................................................................... 21
5.1.1 The Wizards «Conventional UT» .................................................. 21
5.1.2 The Wizards « Beam forming phased array »............................... 21
5.1.3 The Wizards « Total focusing method (TFM) » ............................. 22
5.1.4 The Wizards « Adaptive total focusing method (ATFM)» .............. 23
5.1.5 The Wizards « Multi-groups techniques» ...................................... 23
5.2 Create an application file...................................................................... 24
5.3 Open an application file ....................................................................... 25
5.4 Application and inspection files management ...................................... 26
6. IMPORT/EXPORT ...................................................................................... 27
6.1 Import files ........................................................................................... 27
6.1.1 Import application or inspection files ............................................. 27
6.1.2 Import pictures .............................................................................. 29
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #4
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #5
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #6
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #7
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #8
INTRODUCTION
The M2M GEKKO is the first portable phased-array system on the market able to drive matrix arrays
and use real-time Total Focusing Methods (TFM). In addition to standard Phased-array and
UT techniques, the GEKKO handles most advanced inspection methods, and benefits from
the latest CIVA features natively embedded. The rugged design and touchscreen make the
battery powered GEKKO ideal for field use.
The M2M MANTIS is completing the portable phased array system family.
The M2M CAPTURE software has been developed to drive both GEKKO and MANTIS for all levels of
operators. The step-by-step interface simplifies the use of menus and reduces the risk of
errors.
CAPTURE has been designed to be upgradable and to allow easy integration of new techniques.
Upgrades are the means to:
• Ensure compatibility with the latest UT inspection techniques. For example, to handle flexible
probes and other real-time adaptive techniques.
• Allow integration of new applications (Note: EDDYFY TECHNOLOGIES can tailor or develop
new applications on demand).
The CAPTURE User Manual is a step-by-step guide to navigate through the unit from input and
inspection parameters through data acquisition and display.
At each step, the parameters that can be accessed are described and explained.
Recommendations, tips and warnings are marked with the following icons:
Ref: CAPTURE_user_manual_v3.0_US-A
GEKKO manual #9
PART 1.
Capture
General Overview
Réf: GEKKO_user_manual_v1.2_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #10
1. Hardware overview
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #11
For more information about GEKKO connectors, front panel and hardware description, please refer to
GEKKO_TECHNICAL_DOCUMENT_AX available online at https://support.m2m-ndt.com/
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #12
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #13
2. Home page
Press the ON/OFF button to turn the system ON.
On/Off
2 6
3
1
4 5
7 8
10
11 12
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #14
3. Touch screen
A few simple gestures - press, swipe, drag - are all you need to use the unit. The following icons are
used in the manual:
CAPTURE software follows a simple and intuitive colour code, which guides the operator and offers
an overview of the actions already completed and parameters that remain to be defined.
In general:
- orange colour is used to guide the operator to the next step.
- light grey colour is used to indicate the current step or the current parameter selection.
- dark grey is used to indicate the steps or parameters previously set.
- red colour is used to indicate an invalid numeric value.
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #15
In all configuration files, navigation through the GUI is from left to right using the arrows located in the
top banner (Figure 6). The arrows accessibility depends on the context:
The changes are activated by pressing the Next button in main panels and OK button in sub-panels. If
the Cancel button is pressed the parameters are reset to their previous values (Figure 8).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #16
CAPTURE GUI offers several editorial tools according to the operator and setup needs.
Press the multiple-choice button to display the drop-down selection list (Figure 9), then press the
correct field to validate the selection.
For the double-option buttons, the selected field is displayed in white text on a grey background
(Figure 10).
Press on any white field box to edit a text or a numeric value (Figure 11). Text or numbers can be
entered using the on-screen keyboard (Figure 12).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #17
Use the on-screen keyboard settings to change the numeric value or text.
In CAPTURE, each parameter has a validity range and inconsistency with other parameters is
checked. An invalid or inconsistent value is highlighted in red (Figure 14). As long as fields
highlighted in red are not edited, it is not possible to access to the next step.
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #18
4. General settings
From the Home panel, press to access to the General settings.
General settings panel provides access to the System parameters and UT preferences (Figure 15).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #19
1: Localization
▪ Language
The available languages in CAPTURE are Chinese (CH), German (DE), English (EN), Spanish (ES),
French (FR), Italian (IT), Japanese (JP), Portuguese (PT) and Russian (RU).
2: Screen
▪ Brightness
▪ Touchscreen
If the touch screen is not behaving correctly, the recommendation is to Calibrate then Linearize.
During the calibrating and linearization process, a blue cross will appear and flash. It is
recommended to use the stylus to precisely press on the crosses.
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #20
4.2 UT preferences
3: Units
▪ Length
Length measurements expressed inch or mm.
▪ Amplitude
Amplitude measurements expressed in % FSH or in dB.
4: A-Scan
▪ Representation
The A-Scan time base can be expressed in time-of-flight, sound path or true-depth.
▪ Signal
5: Calibration TCG
▪ Reference level
Adjust the calibration reference level and the tolerance threshold in %.
Press Upgrade (6) to install new releases of CAPTURE software when available. The upgrade file
Capture_Update_x.x.x.cap should be stored in the root (not in a sub-folder) of a USB flash
drive. When the USB drive is plugged, the detection of the upgrade is automatic and will be
done when confirming “Upgrade to version Vx.x”.
5. Launch an application
Several possibilities are offered to launch an application. The first one is to create a new application
from the Wizard panel. The second is to run an existing application from the Applications
panel. The third possibility is to open an inspection file from the Inspection files panel, which
contains all settings information. This will open the inspection in the Analysis panel, then end
the Analysis panel to run the same application or modify the settings.
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #21
The Wizard panel contains the generic file source of techniques from which one application can be
derived.
They present all single techniques available and offer some multi-group templates.
• Dual element transducer: conventional transmitting and receiving separate element technique
Pressing Beam forming phased array (Figure 17) highlights the different phased array techniques
available.
➢ Dual linear array: dual linear array probe (DLA) with sectorial and/or linear scan technique.
➢ Dual matrix array: dual matrix array probe (DMA) with skewed sector scans technique.
➢ Linear array: linear array probe with sectorial and/or linear scan technique.
➢ Matrix array: matrix array probe with skewed sector scans technique.
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #22
Pressing 3 – Total focusing method (TFM) (Figure 18) highlights the different TFM techniques.
➢ TFM Dual linear array: dual linear array probe (DLA) with TFM reconstruction.
➢ TFM Dual matrix array: dual matrix array probe (DMA) with skewed TFM possibility.
➢ TFM Matrix array: matrix array probe with skewed TFM possibility.
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #23
Pressing Adaptive total focusing method (ATFM) highlights the template for ATFM Linear Array.
This module is optional and available only on GEKKO units.
Pressing Multi-groups techniques (Figure 19) highlights the different combinations of technique
available.
• 1 PA + TOFD: one phased array probe and one pair of TOFD probes.
• 1 PA multi-salvos: one phased array probe combining electronic and sectorial scanning.
• 2 PA + Dual: two phased array probes and one dual element transducer.
• 2 PA + TOFD: two phased array probes and one pair of TOFD probes.
• 2 PA multi-salvos: two phased array probes, each one combining electronic and sectorial
scanning.
It is possible to change the probe configuration, to add or delete probes and to add or remove
salvos (groups) from any wizards except TFM.
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #24
To create a new application, select a technique or a combination of techniques in the Wizards list. The
right side of the panel shows a schematic of associated technique or combination.
When selected, the Start button of the bottom banner turns from light grey to orange.
Press Start to create the Application. Then it requires a folder location and application name to save
the new application (Figure 20).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #25
The default location is in the root of the Applications panel. Otherwise, select any existing application
folder or press to create a new folder as the application location.
Press the application name to rename the application. The default application name is the Wizard
name.
Press OK to switch to application setup screens.
Application files can be organised and edited from the application panel (Part 1 Section 5.4).
On first run, the Applications tab contains four single-technique applications located in the Tests
folder. All the new applications (generated from the Wizards, imported or edited – Part 1
Section 6) will be stored at this location.
To open an application file, select the application then press Start. If the application is located in an
application folder, press on the folder then select the application file (Figure 21).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #26
In contrast to the Wizards, Applications and Inspection files can be organised, edited, imported…
Depending on the type of element selected (file, folder, or none), some editing buttons are
available (Figure 22).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #27
The available file management buttons – accessible in the Applications and the Inspection files
panels - are surrounded by an orange border in figure 20, and allow to:
Cut a file
Paste a file
6. Import/Export
CAPTURE allows importation and exportation of pictures, applications, inspection files and inspection
reports. Importation and exportation procedures are described in the next two chapters.
Application and inspection files can be imported from a USB flash drive. Select the Applications or
Inspection files panel then press to import related files (Figure 23).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #28
If an application or an inspection file is selected, the button is not active. Press the
Applications or Inspections tab to deselect all or select a folder before importation.
When the pop-up window appears, select the USB flash drive in which the file or file folder is located
(Figure 24).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #29
Navigate through the folders to select the file then press Import (Figure 25).
The file or file folder is now available in the CAPTURE home panel.
Some external hard drive with USB port requires some power to run properly. Its mays occur
that the unit odes not deliver enough current to access the hard drive.
Prefer USB memory stick or see the document Installation Guide for Connectivity
explaining how to connect the unit to a computer which is available here:
https://support.m2m-ndt.com/gekko/documentation.html
The CAPTURE GUI can be customised and dedicated to a specific inspection using a picture. This
feature is available from the Equipment panel, after pressing the pencil icon of the specimen,
probe or scanner picture (Figure 29 – 3), to associate a personalised image or photo to the
equipment.
To export applications or inspection files, select a file in the list. To export a set of applications or
inspections, select the associated folder then press (Figure 26) to open the exportation
pop-up.
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #30
Select the USB flash drive and navigate through the folders to select the appropriate location then
press Export.
At any time, press the Screen Copy ( ) button located on the unit front panel to perform a
screenshot. The screenshot manager is available from the Home page. (Figure 5 – 5) :
Once in the screenshot manager pop up window, click on the images to select them (ticking their box
in green) and use the buttons to export, delete and rename the selected screenshot(s).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #31
Press to rename the selected screenshot (available when only one screenshot is selected)
Although each CAPTURE software update is rigorously tested before release, undetected issues may
slip through the net.
If a serious error occurs, a pop up will notice the user and the software will automatically generate an
error report. As for the screen copies, it is possible to export them to a USB flash drive.
To access the export pop up window, click on the dedicated button from the Home page (Figure 5 – 4)
During exportation, a file named like this one « [0xEA8F2E1C]_ErrorReport_2019-02-18-15-49-
02.capture_errlog» and containing any unexpected error reports is automatically created on
the USB flash drive and the error report files are erased from the unit.
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #32
Please send these files with a brief description of how the error occurred to
When the unit switches on after the error, a pop up allows the user to recover the last application file
(Figure 28).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE GENERAL OVERVIEW CAPTURE manual #33
PART 2.
Equipment
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #34
PART 2. EQUIPMENT
The Equipment panel appears when any application is loaded. It is the first step of the application
definition.
The Equipment panel shows the specimen, probe and scanner sub-panels (Figure 29). A summary of
the current equipment is presented. When all equipment is correctly set, press Next to go to
the Settings panel.
1: Top banner – Equipment step
2: Equipment sub-panels (Specimen / Probe / Scanner)
3: Edit the equipment picture
4: Summary of the current parameters
5: Load equipment (Part 2 Section 1.5)
6: Edit / check the settings
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #35
1. Specimen
Press Edit from the Specimen sub-panel (Figure 29 - 6) to define or check the specimen settings.
This section contains all the defining elements of the part to be inspected: part geometry, weld
preparation details and material properties (Figure 30).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #36
1.1 Geometry
Press Geometry tab to access the geometrical properties of the specimen to be inspected.
Following geometries are available:
• Plate
• Cylinder
• Nozzle
• T/Y
There are some restrictions according to the unit versions.
1.1.1 Plates
1.1.2 Cylinders
Cylinder parameters are length, external radius, thickness and angular sector.
The essential parameters for cylinders are the thickness and the external radius.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #37
1.1.3 Nozzle
This allows the definition of the full nozzle geometry (except weld) and a real-time display of the cross-
section of the nozzle superimposed to sectorial data.
Nozzle geometry is available from the dedicated application template “PA – Nozzle inspection” and for
all wizard techniques except TFM, ATFM and TOFD.
Press the box “Nozzle configuration” to fully define the nozzle geometry.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #38
Press the shape drop-down menu to select Fillet weld (Figure 33)
This allows the definition of the T or Y shape geometry, including weld preparation. Chamfer and back
strip settings are available from the Weld tab.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #39
For analysis, it is then possible to display rebounds off the geometry to detect where the indications
are coming from. It also helps the user to determine if an echo is related to the specimen
geometry or to a real defect.
Filet weld geometry is available from the dedicated application template “PA – Fillet weld inspection”
and for all wizard techniques except TFM, ATFM and TOFD.
1.2 Weld
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #40
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #41
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #42
The presence and the orientation of the weld leads to a default orientation of the
transducer(s) on the component, always perpendicular to the axis of the weld.
When a cylindrical component without weld is selected, it is still possible to select the
orientation of the probe by selecting “Circum.” or “Longi.”
Press Modify Weld (Figure 35– 4) to open a graphical wizard that allows defining 25 bevel shape,
classified by: symmetry of the bevel (Figure 41) and bevel profile (Figure 42).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #43
Press the left or right arrows to select another weld profile. The sketch and parameters of each bevel
are updated when scrolling.
Adjust the weld profile parameters then press OK to confirm or press Cancel to quit the Modify Weld
menu.
This feature allows superimposing a CAD image of a specimen to the ultrasonic data. This feature is
particularly useful when one wants to analyse the origin of the various echoes and distinguish
between geometry echoes and echoes from defects.
In CAPTURE, dxf compatibility is limited to 2D dxf images containing segments and arcs. The feature
is also limited to an image import, which means it is still necessary to define a basic
specimen (plane or cylinder, with or without a weld) to position the probe and calculate the UT
paths.
As mentioned in introduction, dxf file import is at this time limited to 2D image import and not true CAD
compatibility.
Please see below the main advantages and limitations of the dxf image feature:
▪ Automatic positioning of the dxf image on the specimen: from a corner of the specimen or from
the weld centreline if a weld is defined. It is up to the user to define correctly the origin of the
dxf file to match the origin of the geometry.
▪ Automatic adjustment along the length direction if plane geometry, auto-adjustment along the
angular sector in case of cylindrical geometry.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #44
▪ Geometry shape, thickness, width and eventually radius (if cylindrical shape chosen) have to
be manually adjusted in order to match the dxf image profile.
▪ Probe positioning, delay-law computation and signal visualization are still dependant on the
specimen. So, it is not possible to visualize reflexion off the surfaces of the dxf image.
▪ dxf image appears in the 3D view and in the corrected B-scan views (S-Scan, E-Scan and T-
Scan)
▪ Both the dxf file and the geometry of the component appear in the various views. It is not
possible to remove the component defined as a specimen in CAPTURE.
dxf image import is available from the Specimen panel, in the Geometry tab (Figure 30– 6).
Press to import a dxf file. When the pop-up window appears, select the USB flash drive in which
the file or file folder is located (Figure 43).
You can import several dxf files and superimpose them on your specimen profile
Press to delete dxf file. When the pop-up window appears, select the dxf file(s) then press OK.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #45
1.3.3 Examples
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #46
Figure 46: dxf image with sectorial scan and focal laws
The S-Scan, E-Scan and TFM are all static images meaning that they are displayed for a fixed
position of the dxf file. These views don’t change even if an encoder is attached to the probe.
If one wants to see the data move along the dxf file, the trajectory must be defined along the
inspection plane. In the Configuration/Motion/Reference panel the Orientation of the probe
must be set to 0 and in the Trajectory panel the main axis must be along X. It is now possible
to see the ultrasonic data dynamically in a 3D view of the component.
The Figure 47 shows the Focal laws panel after the dxf import of a tapered weld with a weld bead. The
thickness of the component was modified to be equal to the thickness entered in the dxf file.
As explained earlier, both the component and the dxf files are represented at the same time so
we see the front and backwall surfaces of the plane component defined in CAPTURE and the
dxf file.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #47
Figure 47: Specimen panel after dxf import and without adjusting
With the dxf image, an operator will be able to analyse properly echoes coming from defects located
along the tapered bottom surface and not confuse them for defects inside the weld.
To use a block dxf image as the stainless steel block in Figure 48, one needs to adjust the thickness
and the width of the specimen.
One can see that the echoes are superimposed with the SDH and the backwall of the dxf file
(Figure 49).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #48
If the thickness of the specimen does not match the dxf import, some tools and calculations
can’t be used. For example, the rebound tool (in the 3D view and in analysis), the
PCS calibration and the half-skip or the one skip mode in TFM depend on the
thichness defined in the Specimen panel.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #49
1.4 Material
Press the Material tab to define the material and the wave velocities in the material (Figure 50).
Press Velocity calibration (Figure 50 – 1) to open the velocity measurement assistant in the
specimen (Figure 51).
This wizard provides step-by-step calibration instructions to measure the speed of sound in a given
material from a block of known thickness and geometry.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #50
5
4
8
7
6
1: UT parameters
Set the delay, range and gain to visualize the backwall echoes on the A-scan. Adjust the gain so that
the strongest echo is not saturated and has an amplitude equal to 80% FSH.
2: Instructions
Follow the instructions to achieve the measurement.
If it is not possible to capture two echoes in the measurement gate when the amplitude of the
strongest echo is displayed at 80% FSH, it is possible to make two sequential measurements.
For the first echo, set the amplitude to 80% FSH and position the calibration gate on this echo
only; then press Measure. For the second measurement, increase the gain so that the next
echo is visible. Position the calibration gate on this echo and press Measure again.
2 Section 2). It is also possible to load another probe from the library database with the
button located below.
4: Parameters to edit
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #51
Load a transducer from the library or reset the default probe. By default, with a Phased Array probe
the software will use an aperture of 16 elements without any focal law applied, i.e. the natural
refracted angle will be generated. Select the type of wave (longitudinal or transverse) to
consider for the calibration, set the distance d1 and d2 in mm according to the instruction
sketch.
6: 80% button.
Use this button will adjust automatically the gain to get the echo in the gate at 80% FSH. This button is
not active if the signal in the gate is saturated.
Once the material name or velocities are updated, press to record a new material in the library
(Figure 50 - 4).
Press to open the material library (Figure 50 - 3). Swipe the list and select a material then press
OK (Figure 52). Material can be removed by pressing .
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #52
Once a new specimen has been defined, press to record it in the specimen library (Figure 53).
Name the specimen and select the destination folder then press OK (Figure 54)
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #53
By default, no specimen folder is present. They can be created from the specimen library
(Part 2 Section 1.5.2)
In the specimen panel, press (Figure 55) to open the specimen library (Figure 56).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #54
1: Specimen list.
2: Summary of the selected specimen.
3: Delete the specimen.
4: Create a new folder.
5: Cut / paste specimen.
6: Select a specimen then press OK to load it.
Organize the specimen library by creating specific folders then moving each specimen file in the
dedicated folder (use the cut and paste buttons to move them).
Open directly the specimen library by pressing from the Equipment panel (Figure 29– 5).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #55
2. Probe
Once the specimen has been defined or loaded from the library, press Edit on the probe subpanel to
check or modify the probe details. (Figure 57).
This section contains all the elements related to the probe definition and the UT setup for the pulser
and the signal processing (Figure 58).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #56
2.1 Configuration
Configuration tab allows defining the probe configuration and to set the pulser and signal processing
parameters (Figure 58).
1. Frequency in MHz
- For phased-array: minimum frequency = 0.4MHz; maximum frequency = 20MHz.
- for conventional probes: minimum frequency = 0.4MHz; maximum frequency= 25MHz.
2. Transmit/receive configuration: Pulse-echo, Dual (PART 2 Section 2.2.3) or TOFD.
2.1.2 Digitizer
For multiple probes, the Sampling frequency and Averaging are global values and are applied
to all the probes similarly.
2.1.3 Pulser
Transmission voltage:
- for phased-array:
minimum voltage = 24V (12V for Gekko Rev C);
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #57
Battery life is affected by the voltage selected; lower voltages will extend the battery life.
For phased arrays, this panel provides an amplitude balancing assistant to ensure that all elements
provide the same output. This wizard is also available in the Probe dicing tab and is described
in the next section.
Press the Dicing tab (Figure 59) to set the probe footprint and element parameters.
For conventional probes, the tab is named Geometry (Figure 64).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #58
Select Linear footprint to define a linear phased-array probe (Figure 59). Parameters to be set for
linear PA probes are:
• 1st channel: define the channel number of the system connected to the 1st element of the
linear array.
• Length: total length of the linear array, automatically calculated from the number of elements,
the pitch and the gap.
Select Matrix footprint to define a rectangular matrix array probe (Figure 60).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #59
Parameters for matrix arrays are defined in 2 dimensions, relative to the array length and width:
• Number of elements: from 2 to 32 or 64 according to the unit.
• Pitch (p): distance between the centres of two adjacent elements.
• Gap (g): space between two adjacent elements.
The total number of elements (lines x columns) cannot exceed 64 or 128 according to the unit
Three buttons are available (horizontal or vertical symmetry, row-column inversion) to allow the
definition of any standard matrix array numbering system. A sketch showing the two-scale axis
and the top-left element highlighted eases probe definition.
For linear or matrix arrays, an amplitude balancing assistant is available from the Configuration or
Dicing tab.
Press Amplitude balancing (Figure 59– 1) to open the PA probes balancing wizard (Figure 62).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #60
The amplitude balancing screen (Figure 62) displays two data windows on the left-hand side of the
screen. The top-left window displays the B-scan made with elementary A-scans. The top-right
window displays the A-scan, with a signal encompassed within the white gate, corresponding
to the vertical line in the echo-dynamic window (lower left).
Instructions can be displayed by pressing Instructions. (Figure 63)
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #61
3. Position the gate on the backwall echo in the A-Scan view to visualize a response for all
channels in the echo-dynamic view.
4. Press 80% to set the response of the higher amplitude signal to 80% FSH.
5. Press Balance or press Reset to start over.
6. Press OK to accept the amplitude balancing or press Cancel to discard changes
12 dB is the maximum gain available. When more than 12 dB are needed (dead elements),
the pulser is disabled.
Click on Channel status to check the gain values and/or disable the channels you want. Green colour
is set for less than 12dB compensation. Others are red.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #62
For both cases, set the diameter or the length and width of the piezo-electric element then the physical
channel (LEMO output) to which the conventional probe is connected.
To define a TOFD probe, select first a Circular or Rectangular crystal footprint in the Geometry tab
(Figure 64 and Figure 65), then set TOFD configuration in the Configuration tab (Figure 58).
In the geometry tab, define the emission and the reception channels (Figure 66).
TOFD probes may be connected using the conventional channels 1 up to 4. These channels have a
maximum voltage of 200V. Nevertheless, if the user needs to extend the number of TOFD
channels, it is possible by using the Phased Array channels. In this case the maximum
available voltage is 120V (100V for Gekko Rev C). To use Phased Array channels to perform
TOFD inspections it’s advisable the use of a pre-amplifier in order to improve the RSB.
Upon request a Split Box with 4 LEMO 00 connectors is available as accessory. For more information,
please contact your local sales representative.
The user should then select Connector = PAUT 128 LEMO Splitter or Connector = PAUT 64 LEMO
Splitter upon the case.
The PAUT 64 LEMO Splitter has 2 PA connector connected with the channel 1-30 and 33-62. The
LEMO connectors are available using the channels 31, 32, 63, 64.
The PAUT 128 LEMO Splitter has 2 PA connector connected with the channel 1--62 and 65-126. The
LEMO connectors are available using the channels 63, 64, 127, 128.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #63
For UT Conventional channels, is advisable to avoid the 1-2 and 3-4 channels combinations
to limit the crosstalk between channels.
TRL (Transmit Receive Longitudinal) Probes have separate transmit and receive transducers which
are acoustically isolated from each other, therefore transmitted sound waves do not interfere
with received sound waves, this gives advantages in near surface inspection and inspection of
austenitic, coarse grained and dissimilar materials.
TRL probes are traditionally conventional UT probes composed of two crystals set at a specific roof
angle and therefore having a fixed focal depth. Now with developments in Phased Array we
are able to use Dual Linear Arrays (DLA) and Dual Matrix Arrays (DMA) coupled to an
acoustically isolated wedge, giving us the opportunity to combine the benefits of phased array
and TRL probes.
It is possible to create applications that utilize each of these; TRL (conventional UT), DLA & DMA and
even combine advanced phased array imaging techniques such as TFM. These options are
available from each dedicated wizard but can also be set by selecting “DUAL” in the
Configuration tab.
The presence of wedge is mandatory for ALL dual probes (conventional, linear or matrix).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #64
3
2
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #65
The footprint setup and probe definitions are set for the transmitting probe and are applied to
both transmitting and receiving parts.
Select Linear footprint to define a Dual linear phased-array probe (Figure 68). Parameters to be set
are identical as for linear PA probes:
• 1st channel configuration: define the channel number of the system connected to the 1st
element of each linear array
• Length: total length of the linear array, automatically calculated from the number of elements,
the pitch and the gap.
Press the button Numbering (4) to open the assistant (Figure 69). A sketch showing the two-scale
axis and the top-left element highlighted eases probe definition.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #66
In case of ‘Independent’:
'Transmission' should be set to 1;
'Reception' should be set to number of elements per array +1.
Select Matrix footprint to define a Dual matrix array probe (Figure 70).
Parameters for matrix arrays are defined in 2 dimensions, relative to the array length and width.
Parameters to be set are identical as for matrix array probes:
• Number of elements: from 2 to 32 according to the unit version. The total number of elements
(lines x columns) cannot exceed 32 or 64 according to the unit version.
• Pitch (p): distance between the centres of two adjacent elements.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #67
As for Matrix array, a numbering assistant is available for Dual matrix arrays from Dicing tab.
Press the button Numbering (Figure 70) to open the assistant (Figure 71). A sketch showing the two-
scale axis and the top-left element highlighted eases probe definition.
Three buttons are available (horizontal or vertical symmetry, row-column inversion) to allow the
definition of any standard matrix array numbering system.
There are two predefined layouts in the drop-down menu 'identical' and 'symmetrical'
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #68
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #69
For both cases, set the diameter or the length and width of the piezo-electric element then the physical
channel (LEMO output) to which the Transmission and the Reception element is connected.
Avoid to use the 1-2 and 3-4 channels combinations to limit the crosstalk between channels.
2.3 Wedge
Press the Wedge tab to define a wedge or to set a transducer without a wedge (Figure 74).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #70
Select With wedge then press Angled for the Type of wedge to define an inclined wedge (Figure 74).
Parameters to be set for angled wedges are:
• Incident or refracted angle: the other field is automatically updated according to the wave type
selected (see below) and the velocities for the wedge and material (material velocity is as
previously defined in specimen sub-panel (Part 2 section 1.4)).
• Wave type: select the type of wave used for the refracted angle (it will select the correct
specimen velocity for the calculation).
• Wedge dimensions: set the length and width (Figure 75). The wedge height is editable in the
References tab (Part 2 Section 2.4).
• Longitudinal wave velocity: set the correct wedge velocity or use the wedge velocity calibration
assistant to calculate the correct wedge velocity (Part 2 Section 2.3.3).
For linear PA probes, there is a specific assistant to calculate automatically the correct
incident angle and height of the wedge (Part 2 Section 2.4.2)
2.3.2 L0 wedge
Select With wedge then press L0 for the Type of wedge to define a flat wedge (Figure 77).
Parameters to be set for L0 wedges are:
• Wedge dimension: set the height, length and width of the wedge (Figure 76).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #71
• Longitudinal wave velocity: set the correct wedge velocity or use the wedge velocity calibration
assistant to calculate the correct wedge velocity (Part 2 Section 2.3.3).
Dual transducers require specific wedges with additional parameters: Roof angle, Gap and Squint.
Defining a small thickness like 0.1mm is equivalent to define a transducer without wedge.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #72
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #73
If a wedge has been defined, it is necessary to define the correct LW velocity value.
You can get this value from the wedge manufacturer. In most cases, wedges are built from REXOLITE
(LW velocity = 2350m/s) or PERSPEX (LW velocity = 2650m/s).
CAPTURE offers the ability to accurately measure the wedge LW velocity using the dedicated wedge
velocity calibration assistant.
Press Velocity (Figure 74– 1) to open the velocity calibration assistant for angled or flat wedges
(Figure 81).
2
2
1
1
5
5
3 4
3 4
Figure 81: Wedge velocity assistant
This wizard provides step-by-step calibration instructions to measure LW velocity for removable
wedges.
3: Adjust the gain for the strongest echo amplitude to be 80% FSH.
4: Press Measure.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #74
The measurement can be achieved in two steps, by positioning the gate and measuring successively
each backwall echo time-of-flight.
Press to record a new wedge in the library (Figure 74– 7), or press to open the wedge library
( Figure 82).
2.4 References
Press the References tab to define the probe origin and set the wedge height and front length in case
of angled wedge (Figure 83).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #75
If a probe without a wedge or with a L0 wedge has been defined, the References tab is only dedicated
to the coordinate system reference setup (Figure 84): at the front, at the middle (beam exit
point) or at the back of the wedge.
If an angled wedge has been defined, the normalized height (H) and the front length (l) of the wedge
have to be set in this tab (Figure 83).
For PA probes, CAPTURE offers three different ways to define the normalized height “H”:
A. Height between the middle of the probe and the beam exit point (id CIVA).
B. Height measured at the middle of the probe face.
C. Height to the centre of the first probe element.
D.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #76
For phased array probes, CAPTURE provides a calculation assistant to measure the incident angle
and the height of the wedge for linear arrays only. A delay path correction is available for
conventional probes.
This wizard is compatible with flat, AOD and COD contoured wedges as shown below (Figure 85).
Press Angle/Height (Figure 83– 1) to open the wedge angle/ height assistant (Figure 86).
1
2
4 5
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #77
The top left window displays the B-scan for all activate elements individually. The middle window
displays the A-scan signal of the single element currently selected, by the orange cursor in the
B-scan view.
• Make sure that the small white crosses are displayed on the B-Scan view. It means that the
wedge echoes are correctly detected, otherwise readjust the gate or the gain value.
For conventional probes with angled wedges, two options for the normalized height “H” are available
(Figure 87):
A. Height between the middle of the probe and the beam exit point (e.g. CIVA).
B. Height measured at the middle of the probe face.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #78
For TOFD probes, the probe reference is the centre of the two probes (Figure 88).
CAPTURE provides a path-delay correction for conventional probes using the measurement of the
height of the wedge.
Press Zero calibration from the References tab (Figure 87/ Figure 88) to open time base calibration
for conventional probes (Figure 89).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #79
1: UT parameters.
2: Probe setup summary.
3: Instructions to make the measurement.
4: A-Scan view.
5: Transmission channel selection (additionally, reception channel selection for TOFD or Dual
transducers).
6: Wedge height.
7: Adjust the amplitude to 80% FSH.
8: Achieve or reset the measurement.
9: Validate or cancel the modifications.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #80
A total of six different probes can be managed by CAPTURE. A pair of TOFD probes is considered as
one probe. Probes can be added when the probe icon is available as indicated in the
following panel (Figure 90). When selected, a probe can be changed to another type in the
Configuration and Dicing tab or removed by pressing .
By default, the probe added is a copy from the previous one, with the same parameters and
active channels. CAPTURE auto-checks the availability of channels and highlights
the part to edit in orange in case of incompatibility.
When a probe is selected, press to save it in the specimen library (Figure 91).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #81
Name the probe and select the destination folder as for specimen then press OK (Part 2
Section1.5.1).
In the probe panel, press (Figure 91) to open the probe library (Figure 92).
Probe library can be managed as the specimen library (Part 2 Section 1.5.2).
The probe library contains probes from Ekoscan, GE, Imasonic, Karl Deutsch, Krautkramer,
M2M, Metalscan, Olympus, Sonatest, Sonaxis, Zetec.
Library is fully customizable to add your own probes.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #82
3. Scanner
Press Edit on the Scanner panel to check or modify the scanner characteristics (Figure 93).
Three encoder inputs can be read by CAPTURE on GEKKO and MANTIS EXPERT and
MANTIS MASTER. Other MANTIS versions have only 2 encoder inputs.
This section allows user to define a 1, 2 or 3 axis encoded scanner or a time encoded inspection.
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #83
. Press Calibration (Figure 94 – 3) to open the encoder resolution calibration assistant (Figure 95).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #84
2: Press to start the calibration. Move the encoder from the above-mentioned distance. Once
achieved, press to stop and measure the accurate resolution.
Scanner library can be managed as the specimen library (Part 2 Section 1.5.1).
Ref: CAPTURE_user_manual_v3.0_US-A
EQUIPMENT CAPTURE manual #85
PART 3.
Settings
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #86
PART 3. SETTINGS
This panel is dedicated to the main acoustic setups for the CAPTURE.
The setting panel is divided into subpanels which differ according to the NDT technique used:
- Specific settings for Phased array probes
- Specific settings for TOFD probes
- Specific settings for Pulse Echo
- Specific settings for Total Focusing Method
In the following chapters, the GUI is described for each single NDT technique.
The advanced concepts of salvos, multi-probes and multi-salvos are described in a dedicated chapter
called multi-salvos setup.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #87
The subpanel Focal laws is dedicated to compute the delay laws and set the probe positioning.
1 2
4
3
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #88
- Sound path;
- Projection;
The Definition parameters are dependent on the scanning type. Capture software supports the
following type of scanning:
- Sectorial Scan: where the active aperture is fixed, and electronically the beam angles are
changed within a defined angular sector;
- Linear Scan: where the ultrasonic beam angle is fixed, and the beam is moved electronically,
through the active apertures of the array;
- Compound Scan: combination between a sectorial and a linear scan. The beam angle and the
active aperture will change consistently, allowing to increase the ultrasonic beam coverage
from a single probe position;
- CIVA laws: where the user may import a set of focal laws previously calculated in CIVA
software.
The following example illustrates different inspection strategies applied for a weld inspection
using the different type of scanning that are available on GEKKO/MANTIS
In the 3D view (3), the rays are computed and displayed following the definition which can be set by
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #89
Direct (half skip) W - transmission (double bounce) V - transmission (full skip) Manual (N-skips)
Direct (half skip) W - transmission (double bounce) V - transmission (full skip) Manual (N-skips)
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #90
on in order to select or unselect the elements in the aperture (Figure 102). T/R pattern
can be identical or independent.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #91
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #92
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #93
The following example illustrates the interlacing between the selected aperture and the angle
to form the compound scan:
The CIVA laws tab is dedicated to handle delay law configuration that are not implemented natively in
the Capture software (for example, algorithms other than Angle and Depth algorithms).
Figure 105 illustrates one example based on a CIVA model with one inspection using immersion. In
this example were used a set of focal laws with focalization points located over the inside
diameter of the cylindrical component:
In Figure 106 we illustrate how to load this configuration. The file to be imported should be the CIVA
folder file *.CIVA.
The piece and probe defined in CAPTURE should be the same as defined on CIVA software;
There are some limitations on the focal laws generated by CIVA that can be imported, namely:
- The group of emitting and receiving elements should be the same. The current version doesn’t
allow to configure pitch and catch configurations;
- The focal laws can have one sequence and multiple shots or multiple sequences and a single
shot. Multiple sequences with multiples shots simultaneously cannot be handled by CAPTURE;
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #94
1: Button to select the *.CIVA file folder to be loaded. The folder should be located on a USB drive;
2: The software retrieves the following information from the CIVA folder:
- Wave type LW (Longitudinal Waves) or SW (Shear Waves) ;
- Shot Number (Number of shots)
- Type (Scan Type): Sectorial Scanning, Linear Scanning
Positioning of the probe over the specimen to be inspected is fully managed in the motion subpanel of
the configuration panel (Part 4).
The index offset box manages the distance between the specimen reference (plotted as a green dot)
and the probe reference (plotted as a red dot)
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #95
- + - +
- +
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #96
For nozzle configurations, the probe is always referenced in relation to the Beam exit point.
90° 270°
0°
Increment axis
Scan axis
Scan axis
180°
On Focusing panel, the user may define the use of focalization. The focus points can be set
according one of the followings:
- True-depth: focus points are set along the defined depth range; the algorithm ensures the
beam angle AND the focal depth are respected;
- Sound path: focus points are set with a constant sound path; the algorithm ensures that the
uncorrected sound path is the same for every angle;
- Projection: focus points are set along a line oriented in any direction;
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #97
TRUE-DEPTH
The algorithm used for focalization is the same as CIVA optimized Focalization and Depth focusing. It
is defined by two parameters:
- Initial depth: which defines the focus initial depth. This value will be linked to Initial angle
defined on Configuration panel;
- Final depth: which defines the focus final depth. This value will be linked to Final angle
defined on Configuration panel;
The Angle range and the depth range are discretized in rigorously equal steps as shown in the next
picture:
SOUND PATH
The algorithm used for constant sound path guaranties that the uncorrected sound path is the same
for every angle.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #98
The following example shows the results obtained with a sound path focalization located at 50 mm.
All the side drilled holes at this position are well detected. We can see the influence of the
beam focalization outside this position. For the holes located at 25mm the beam distortion is
evident, leading to an image and SNR relatively poor.
Figure 111: Sound path Focusing (Focus sound path 50mm – ASTM E2491 Block)
PROJECTION
Projection focusing allows to focus along a line oriented in any direction. It is defined by two
parameters:
- Projection: which defines the distance between the reference point of the wedge (red dot)
and the centre of the line of focusing points;
- Tilt: which defines the orientation of the line of focusing points. It is possible to enter negative
values;
Projection focusing allows to focus along chamfers. For example, the acoustic beam will be focused
along the first bevel preparation (probe side) on full skip and along the second bevel
preparation (probe opposite side) on half skip. This scanning strategy will improve the
detection of planar defects oriented along the fusion line as lack of fusion. Figure 112 and
Figure 113 illustrates the benefices of such type of focusing applicable to a V shape weld
preparation.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #99
Matrix arrays allow to steer the acoustic beam on 3D plans of incidence, improving the detectability of
misoriented discontinuities.
The Configuration panel has the same parameters described previously. With this type of probes, the
sweep mode is restricted to Sectorial scan. It is also possible to import CIVA laws that were
previously configured with CIVA software.
On Focusing panel, the user can adjust the sweep angles along the skew direction:
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #100
1.1.9 3D views
The 3D view shows 3D management buttons. Table 1 describes the associated actions.
Display the specimen viewing angle Select the viewing angle of the specimen
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #101
One signal gate, containing the A-Scan data, is automatically set and displayed in white (Figure 116).
Swipe the gate rectangle on the A-Scan view to move the gate position. The Corrected view is
automatically updated.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #102
1: Four additional peak gates (saving amplitude and time of flight) can be added by pressing .
One additional Sync gate is available, if either one of those four gates will be synchronized.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #103
5
3
1
2
4
1: The Detection type can be selected from the drop-down menu. These detections can be
associated to a positive, negative or absolute value of the RF signals (Figure 119). The
Detection type drop down menu allows selecting the:
• Max peak: this option stores the amplitude and time-of-flight value associated to the maximum
echo above the threshold.
• First peak: this option stores the amplitude and time–of-flight value associated to the first
echo above the threshold.
• Last peak: this option stores the amplitude and time-of-flight value associated to the last echo
above the threshold.
• Max flank: this option stores the amplitude and time-of-flight value right above the gate
threshold associated to the Max echo.
• First flank: this option stores the amplitude and time-of-flight value right above the gate
threshold associated to the first echo.
Max peak First peak Last peak Max flank First flank
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #104
The A-scan is stored for all range of the time base independently the gate position and gate
detection mode. A-Scan information will be stored if the recording mode is set to All data, in
Recording panel (Section 4)
2: The Phase can be selected from the drop-down menu. The following options are available:
• Positive: this option will consider only the positive sign of the RF signal for the detection
according with one of the above detection methods;
• Negative: this option will consider only the negative sign of the RF signal for the detection
according with one of the above detection methods;
• Absolute: this option will consider either the positive or negative sign of the RF signal for the
detection according with one of the above detection methods;
3: The Synchronized button can be selected to synchronize the gate with a reference signal.
A synchronized Gate is typically used for L0 inspection when using an immersion tank or a water
wedge (Hydroform type or Roller probe)
The variation in the height of the water column needs to be compensated to have a true depth
measurement
A synchronized gate is positioned at the front surface echo, it is being used to measure the reflexion
off that surface. The second gate is synchronized with this gate GSync and moves with it.
4: The Colour drop-down menu allows to change the gate colour
5: The Mode button can be selected to choose between true-depth and sound path gate setting
Often the front surface echo appears saturated. The 16-bit dynamic of the GEKKO/MANTIS
allows to still measure the maximum up to 800% FSH. If the value goes to 800%, it is
recommended to switch the synchronization to a First Threshold
By default, Amplitude calibration is disabled (Off selected). Press On to activate the amplitude
calibration panel.
Press Edit to access the Amplitude calibration subpanels (Figure 120).
• Advanced TCG: this option will used when different TCG curves will be applied for each
shot/sequence; This option is normally used when the configuration uses a phased array
probe with a wedge for S-Scan and E-Scan configuration;
• Basic TCG: this option will used when the same TCG curve will be applied for all the
shots/sequences; This option is normally used when the configuration uses a phased array
probe with a 0° LW configuration, as well for conventional probes;
• DGS: this option will used when to calculate the ERS- Equivalent Reflector Size in comparison
with a response obtained from a flat bottom hole oriented perpendicularly with the beam.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #105
TCG calibration is a Time-Corrected-Gain wizard which is designed to equalize the amplitude of all
A-scans from the different focal laws over the sound path range, using a collection of reflectors
located at different depths.
This is the same function as a conventional single channel instrument; The main difference is in this
case is that is applicable to all focal laws or A-scans simultaneously.
The basics of phased-array TCG calibration is to scan over the reference reflectors and to compute a
TCG curve per shot/sequence. The TCG compensation will set all reflectors to the same
amplitude previously defined. The default reference value is 80% and the tolerance value is
5%. These two values can be modified in the Preference panel (Part 1 Section 4.2).
The TCG in CAPTURE can compensate between -6 dB and +20 dB meaning that signals can be
saturated as much as 200% FSH.
However, when CAPTURE applies negative compensation, it changes the dynamic of the acquisition.
If – 6dB are applied in the TCG then the dynamic will be 400% instead of 800%. It is thus
recommended to bring the strongest signal at 80% FSH by adjusting the gain.
Basic TCG calibration will apply the same TCG curve to all the shots/sequences. This means that will
no individual compensation per shot/sequence will exists;
Basic TCG is usually used for LW 0° inspection like corrosion mapping and composite inspection for
which all the sequences have the same delay laws. It can also be used to set the TCG curve
for a conventional monolithic or TR probe;
In the following example, a calibration block with four side-drilled holes located at depths of 10mm,
20mm, 30mm and 40mm will be used to demonstrate the use of this tool.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #106
• Press Play
• Move the probe front and backwards to record the amplitude and TOF for each one of
the sequences. An image with all superimposed E-Scans is being displayed. The A-Scan view
displays the recorded signal for the selected shot/sequence, and the respective echo dynamic;
If necessary, to restart the TCG recording press . This will clear all recorded signals from
the E-Scan and A-Scan view
• Press Detect
Capture will detect automatically the peaks over the threshold and display them in a list.
It is possible to delete some points if needed;
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #107
• Press Apply
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #108
It is possible to create manually a TCG by adding points pressing the sign. The operator will be
prompt about the sound path and gain for the point to be added:
It is possible to Save/Load the TCG. If the operator uses the same probe, same wedge, the same
material, and the same focal laws, it might simpler to load a TCG previously done
In the following example, four side-drilled holes located at depths of 15mm, 30mm, 45mm and 60mm,
using ISO 19675 calibration block will be used to demonstrate the use of this tool.
• Press Play
• Move the probe front and backwards to record the amplitude and TOF for each one of
the shots/sequences. An image with all superimposed S-Scans is being displayed. The A-
Scan view displays the recorded signal for the selected shot/sequence, and the respective
echo dynamic.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #109
Corner Echo
1st Hole 1st Hole
2nd Hole
2nd Hole
rd
3 Hole
3rd Hole
4th Hole
4th Hole
In this step the user will define zones to retrieve the amplitude and TOF data from the superimposed
S-Scan view:
• The echo dynamic displays the response of the hole for all shots (angles)
• Add as many zones as required by the procedure and the required thickness to be
inspected
• Adjust the Threshold to set the gate record level (set typically to 10% - 20%).
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #110
The corner echo from the R50 reflector from block ISO 19675, does not allow to add computation zones
over the 3rd and 4th holes.
After Compute press , to return to the TCG Recording menu, and thus record the 3rd and 4th
holes, from a different probe position, avoiding to record the corner echo.
NOTE: Since current version of Capture, does not handle with different gain settings during the TCG
recording, the gain setting should not be changed.
• After Compute Press , returning to the TCG Recording menu, in order to record
again the amplitude response for the 1st and 2nd hole. The aim is to check if the TCG
compensation was set properly;
• Press Play
• Move the probe front and backwards to record the amplitude and TOF for each one of
the shots/ sequences
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #111
• Press Next
• Verify if the echo dynamic for each one of the zones, fits inside the shadow green area
For the case of this example, press to repeat the entire procedure for the 3rd and 4th hole. The
final result is illustrated on Figure 130, where the 4 holes are set at 80% ±5% for all the shots
(angles) and depth range:
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #112
1st Hole
2nd Hole
3rd Hole
4th Hole
Figure 131: Advanced TCG calibration – Focal Laws Compatibility during TCG load
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #113
DGS is a sizing technique that relates the amplitude of the echoes and return an estimation of the
equivalent size by comparison with a flat bottom hole located at a position with the same
sound path. This value is known as Equivalent Reflector Size or ERS.
Traditionally this technique involves the comparison between the echo amplitude from a reflector with
reference curves using abacus curves that were available either in printed forms or directly
embedded in the instruments. One abacus curve is associated to each transducer
(propagation media, crystal shape and dimension, frequency).
The limitation of this technique is frequent since it will not work for the probes for which there are no
abacus available as is the case of most phased-array probes.
Capture software uses CIVA modelling tool to estimate the ERS for each position in the S-Scan and
for any probe including phased-array probes
DGS calibration wizard allows to express signal amplitude in ERS diameter. By recording a reference
amplitude value from a SDH for one shot/sequence at one know sound path, the software will
be able to calculate the ERS value for all shots/sequences, and for any sound paths.
• Select the reference angle and position the gate on SDH echo;
• Set the correct SDH diameter;
• Optimize the echo amplitude and then press Compute REF ;
Once computed, the A-Scan view will be frozen indicating the ERS value calculated (Figure 132).
Press Reset to repeat the measurement or press OK to apply computation.
The ERS value will then be displayed either in inspection as in analysis mode.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #114
2. TOFD Settings
The TOFD panel is divided into three sub-panels. The first one addresses the PCS calibration, the
second one regards gate settings and the third one is for amplitude calibration (Figure 133).
The TOFD calibration panel is used to measure the PCS – Probe Centre Separation of the TOFD
probes. For TOFD inspection, PCS value is required to retrieve the depth of the indications
(Figure 134).
In this menu, the operator can access the following calibration parameters.
1: Specifications of the probe used are called from the probe panel (Part 2 Section 2.2.4);
2: The PCS value is computed by default, from the refraction angle considering the Rays intersection
located at 2/3 of the thickness. This PCS value may be edited manually, or may be computed
using the TOFD calibration procedure, described above;
3: Two gates are available. The first one is used to detect the lateral wave echo. The second one is
used to detect the backwall echo. For each gate, detection mode and threshold may be
specified by the user (Part 3 Section 1.2);
4: Two views may be displayed: A-Scan and mechanical B-Scan views;
5: Pressing the TOFD calibration button calibrates the TOFD settings; (Part 3 Section 2.1.2);
If no PCS calibration is performed, hyperbolic cursors and depth scale are not available
in acquisition.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #115
1 4
Figure 135) and the threshold type (Figure 136) of the lateral wave and back wall echoes, press the
button .
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #116
Available detection modes are the same as the ones presented in Part 3 Section 1.2.
Available threshold types are positive, negative and absolute, respectively enabling the detection of
positive peaks, negative peaks and absolute value peaks.
There are three methods to perform a TOFD calibration with Capture, i.e. to convert the TOFD scale
on the A-Scan and B-Scan from TOF to depth (mm).
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #117
If both lateral wave and back wall echoes can be observed, check the two boxes and position each
gate on the respective echo. Then press the TOFD Calibration button. PCS is automatically
computed from the TOF retrieved from the gate information. TOFD scale is now calibrated and
may be displayed in mm in the A-Scan view (Figure 137). Press the Reset button to erase the
calibration.
The input parameters used for this calibration are the thk - thickness and the vLW - sound
speed.
Based on the measurement of tLW – time of flight of lateral wave and tBW – time of flight of
backwall echo the software may compute the PCS and the probe delay according to:
Figure 137: TOFD Calibration (Lateral wave, Backwall echo and thickness)
The TOFD depth scale will be computed based on PCS, the probe delay and the vLW -
sound speed, according to:
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #118
If only the lateral wave echo can be observed, the TOFD scale may be calibrated using this echo and
the PCS value. To achieve this, check only the Lateral wave box and position the gate onto
the associated echo. Measure and enter the PCS value that matches with the probe centre
separation probe. Press the TOFD Calibration button to calibrate and adjust TOFD scale
(Figure 138). The scale of the A-scan view may now be displayed in mm.
The input parameters used for this calibration method are the PCS and the vLW - sound
speed.
Based on the measurement of tLW – time of flight of lateral wave, the software may
compute the delay.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #119
If only the Backwall echo can be observed, the TOFD scale may be calibrated using this echo and the
PCS value. To achieve this, only check box the Backwall echo and position the gate over this
signal. Measure and enter the PCS value that matches with the probe centre separation
probe. Press the TOFD Calibration button to calibrate and adjust the TOFD scale (Figure
139).
The input parameters used for this calibration method are the PCS, the thk - thickness and
the vLW - sound speed.
Based on the measurement of tBW – time of flight of Backwall echo, the software may
compute the delay.
Press the Edit button to enter the Gate panel (Figure 140).
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #120
Gate management in TOFD and conventional modes (Figure 141) are the same as gate management
in phased-array mode (Part 3 Section 1.2.).
1 2
1: The operator may add up to five gates by pressing the button (max number of gates is 4)
2: Each gate parameter may be set by pressing the button. Pressing the button deletes the
selected gate.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #121
In TOFD mode, the signal is always displayed in RF mode and the button located at the bottom
left of the A-Scan enables to choose between millimeters (depth scale) and microseconds
(time-of-flight scale).
Since TOFD technique is a non-amplitude technique, the amplitude calibration is not normally required
to be performed. Nevertheless, the equipment allows to perform such compensation. To
enable the Amplitude calibration, press On, and then press the Edit button to enter the panel
(Figure 142).
The amplitude calibration process for this module follows the same steps that are applicable to
conventional UT probes, and it will be described in detail in Section 3.2.
3. Pulse-Echo settings
The Pulse-Echo settings panel is divided into two sub panels. The first part addresses gate settings
and the second one deals with amplitude calibration (Figure 143).
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #122
The edition of gate settings shown in Figure 144 is the same as described in TOFD mode Part 3
Section 2.2.
The choice of the scale unit – between mm and µs – should be made in the System preferences panel
(Part 1 Section 44).
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #123
Amplitude calibration is the same as described in the TOFD section (Part 3 Section 2.3), see Figure
146.
In pulse echo mode, the software allows to display the ultrasonic signals either in TCG representation
as well as DAC curve drawn over the screen, as a conventional UT device. The Figure 145
illustrates the same UT signal displayed by both methods. To switch between both
representations, press the button as desired.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #124
The Total Focusing Method, also known as TFM, is a high-resolution imaging algorithm achieved into
two steps: first the data collection (Full Matrix Capture) and then the image processing (TFM
itself).
Full Matrix Capture (FMC) is a specific data-acquisition process using ultrasonic phased-array probes
(Figure 147 and Figure 148). For an array of N elements, each element is successively used
as the transmitter, while all other elements are used a receiver. Figure 147 illustrates the
sequencing of the shots fired by a phased-array probe used with FMC. The data are organized
in a matrix Sij that contains all the acquired signals. Sij contains the signal transmitted by
element i and received by element j, as illustrated in Figure 149.
Figure 147: Transmission and reception by the first element of the probe
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #125
For a trigger mechanical position, once the full data is acquired, this is FMC matrix is full, all the data
are ready for processing. The signal processing is called as total focusing method (TFM). A
computation zone is defined for the data reconstruction. This zone of interest is meshed, and
for each point in the grid, the correspondent time of flight from the ith emitter to the jth receiver
is calculated for a specific mode of wave propagation (Figure 149).
The TFM algorithm consists in coherently summing all the signals sij(t) from the dataset to focus at
every points of a Region of Interest (ROI) in a specimen.
Mathematically this can be expressed as:
where tij(P) represents the theoretical time-of-flight corresponding to the propagation time between the
ith transmitter and the jth receiver, through point P.
In CAPTURE, this reconstruction is performed in a 256x256 pixels zone with a rate of 25 frames/sec;
each pixel corresponding to one of the focused points of the ROI.
In the Settings panel, enter the sub-panel named TFM settings to edit the TFM parameters
(Figure 151).
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #126
Press the Edit button to specify TFM options. The panel shown in Figure 152 is now active.
3
5
6
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #127
1: The following wave propagation computation modes are natively embedded in CAPTURE software:
- Wave Type L (Longitudinal Waves)
LL: longitudinal waves in Half Skip;
LLL: longitudinal waves in Pitch-And-Catch configuration;
LLLL: longitudinal waves in Full Skip;
- Wave Type T (Shear Waves)
TT: shear waves in Half Skip;
TTT: shear waves in Pitch-And-Catch configuration;
TTTT: shear waves in Full Skip;
- Conversion mode techniques
TLT: shear-longitudinal-shear in Pitch-And-Catch configuration;
TLL: shear-longitudinal-longitudinal in Pitch-And-Catch configuration;
LTT: longitudinal-shear-shear in Pitch-And-Catch configuration;
Ultrasonic paths are used by CIVA to calculate the times-of-flight to process the TFM images. These
travel paths only deal with waves propagating inside the component.
2: The computation Zone is the area where the mesh will be applied, and the algorithm will apply the
TFM reconstruction.
The Depth and X-Offset parameters defines the position of the zone of interest. The reference is the
middle of the red line at the top of the zone of interest as illustrate in Figure 153
3: Index offset and Orientation allow to position the referential of the transducer with respect to the
component referential.
4: Gain; The maximum analogue value is limited to 40dB. It is recommended to set the gain between
30 and 40 dB to have a good detection capability.
5: Press the Calculate button to start CIVA calculation and to see the real-time TFM display.
6: Press the Adjust button to adjust automatically the gain;
7: Press the TFM options button to adjust image resolution, aperture and computation speed (TFM
mode); it displays the panel as seen on Figure 154.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #128
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #129
For the modes with rebounds, i.e. all available modes except LL and TT, is better to define
the zone of interest within the component.
When the thickness is not perfectly known, or if the thickness is varying, it is most efficient
to compute the TFM image in direct mode (LL or TT), and define an area of
interest >2x thickness
By default, Amplitude calibration is disabled (Off selected). Press On to activate the amplitude
calibration panel.
Press Edit to access the Amplitude calibration subpanels.
The Basic and Advanced TCG described earlier is also available for TFM technique.
Advanced TCG works exactly the same as for phased array except that the operator needs to move
the probe slower. With a frequency of 20-25 Hz it is necessary to move at around 2 mm/s.
The advantage of TCG applied to TFM is that it works great with a focused beam as the TFM focuses
the energy everywhere.
In the following example, four side-drilled holes located at depths of 10mm, 20mm, 30mm and 40mm,
will be used to demonstrate the use of the tool.
• Press Play
• Move the probe front and backwards to record the amplitude and TOF associated to
each one of the pixels. An image with all superimposed T-Scans (TFM-Scan) is being
displayed.
The vertical echo dynamic view displays the recorded signal along the vertical column located
at the middle of the TFM image;
• Press Detect
Capture detects automatically the peaks over the threshold and display them in the list. It is possible to
delete some points if needed;
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #130
• Press Apply
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #131
• Press Play
• Move the probe front and backwards to record the amplitude and TOF associated to
each one of the pixels. An image with all superimposed T-Scans (TFM-Scan) is being
displayed.
The vertical echo dynamic view displays the recorded signal along one vertical column of the
TFM image, selectable by the user.
• Press Next
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #132
In this step the user will define zones to retrieve the amplitude and TOF data from the superimposed
T-Scan (TFM-Scan) view:
• Position the zone in the T-Scan to encompass the response of the 1st hole
• The echo dynamic displays the maximum response of the hole for all the pixels along
the vertical direction
• Add as many zones as required by the procedure and the required thickness to be
inspected
• Adjust the Threshold to set the gate record (set typically to 10% - 20%).
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #133
• After Compute Press , returning to TCG Recording menu, in order to record again
the amplitude response of the reflectors. The aim is to check if the TCG compensation was set
properly;
• Press Play
• Move the probe front and backwards to record the amplitude and TOF for each one of
the pixels;
• After finish, press Stop
• Press Next
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #134
• Verify if the echo dynamic for each one of the zones, fits inside the green shadow area;
1st Hole
2nd Hole
3rd Hole
4th Hole
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #135
5. Multi-salvo setup
A multi-salvo setup allows performing multiple configurations with different ultrasonic settings with the
same phased-array probe at the same time.
CAPTURE allows a multi-salvo setup with up to 6 probes, phased-array probes each performing
multiple salvos as well as single-crystal probes used in pulse-echo or TOFD. By extension, the
TOFD or Pulse-Echo modes defined in the Probe Tab are defined as unique salvos.
CAPTURE can manage up to eight salvos. Table 2 shows the various combinations.
Combined
wit PA TFM TOFD PE
h
PA x x x
TFM
TOFD x x x
PE x x x
Table 2: Multi-salvo combinations
Due to the required processing power, the TFM can only be applied to one salvo (specific
wizard).
Multi-salvos can be defined in the Settings panel, when the icons and are available at the
bottom of the screen (Figure 163). The creation of multi-salvo applications requires to start
from a Combination Wizard.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #136
When a salvo is selected, press the button to add a salvo to the transducer. The new salvo will
copy automatically the parameters previously selected. Then, select the salvo and modify the
delay laws to fix a new set of laws.
When several salvos are available for one transducer, press the button to remove the selected
salvo.
Ref: CAPTURE_user_manual_v3.0_US-A
SETTINGS CAPTURE manual #137
PART 4.
Configuration
Ref: CAPTURE_user_manual_v3.0_US-A
GEKKO manual #138
PART 4. CONFIGURATION
All configuration panels, required to start an Inspection, are included in this part: the probe positioning
and the control trajectory can be set in the Motion panel, the layouts and the content of
acquisition view can be organized in the Display panel and the Pulse Repetition Frequency
(PRF) can be adjusted from the Configuration summary panel (Figure 164).
This panel is dedicated to present the configuration panels that are required to be configured before
starting the Inspection. The configuration panel is divided into the following subpanels:
1: Motion: where the user sets the probes positioning, the scanning trajectory, and the Pulse
Repetition Frequency (PRF);
2: Display: where the user configures the display panels;
3: Recording: where the user configures the recording mode and defines the information that will
be included on the test report;
The Configuration panel is independent from the techniques previously defined.
1 2 3
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #139
1. Motion
The subpanel Motion contains two tabs: References and Trajectory. The References tab allows
positioning of each probe with respect to the probe holder reference point and the specimen
reference point (Part 4 Section 1.1). The Trajectory tab defines the scanning trajectory (Part 4
Section 1.2).
With phased-array probes, the orientation and the index offset can be defined in the
Settings panel (Part 3 Section 1.1.5.).
1.1 References
In this tab, each probe or group of probes location can be managed. In addition to the 3D visualization
window, three main sections are available:
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #140
Each probe is associated to a red dot representing the probe reference in the 3D visualization window;
After select the probe on the right panel, the associated probe (pair of probes for TOFD) will be
highlighted in blue on the 3D view;
The values Index offset, Scan offset and Orientation, are independently linked with each one of the
probes, represented in 3D view by the red dot;
2: Scan
The Scan offset or group offset (yellow dot), allow to add an offset applied to all set of the probes with
respect to the specimen reference (green dot);
To define a probe as the group reference, select a probe, and press ; When a probe is
used as group reference, the yellow dot is merged with a red dot linked with the probe;
3: Specimen reference
Depending on the specimen type, the origin can be selected with the drop-down menu, Specimen
border or Specimen centre;
Probe Ref.
Scan Ref.
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #141
1
Specimen Ref. Scan offset
Index offset
Probe Ref.
Scan Ref.
2
1.2 Trajectory
The trajectories menu will be dependent on the type of component to be inspected, as well the type of
scanner to be used during the inspection.
The 1D trajectory is automatically defined when the selected scanner only has one axis.
The 3D representation is a live view, i.e. when the user moves the scanner, the probe displacement
can be seen directly over the 3D view. This allows the user to verify the probe path and check
the direction of scanning.
1: Orientation
The direction of scanning should be defined, to match along X or Y direction;
2: Distance
The Distance is the total length of the acquisition area (expressed in mm). For pipe components, the
user needs to calculate and enter the perimeter of the pipe;
3: Step
The Step is the acquisition step between two consecutive data points;
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #142
4: Scanning direction
The acquisition direction may be reversed by checking the box Reverse the trajectory on the 3D
view;
5: Auto Stop
The acquisition will be automatically stopped after the user records all data along the defined distance
by checking Automatic acquisition stop;
6: Encoder Reset
2 3
4
The 1D trajectory + increment is similar to 1D trajectory, but it allows to add a discrete increment in the
index direction. The increment will be added manually, either by software or by using the index
button in the scanner, if available.
Normally, for the case of linear scan with 0° LW, the increment step value is about 90% of the sweep
range of the probe. This overlap allows to assure the complete beam coverage along the
material to be inspected;
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #143
Figure 169: One-axis application (Corrosion mapping with manual triggered overlapping)
In this example the increment Step was defined to 33mm, and the Distance to 165mm, which means
that the coverage area of 165mm will be filled after 5 steps of 33mm, as illustrated in Figure
169
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #144
In this example the Step was defined to 35mm, and the Distance to 70mm; The acquisition will be
performed with 3 steps (Increment Y=0, Y=35 and Y=70mm), ensuring the full weld coverage
with a single acquisition file, without the need of posterior post-processing to merge the data.
The Figure 171 illustrates the achieved weld coverage using this type of scanning strategy:
Figure 170: One-axis application (Weld application with manual triggered overlapping)
Figure 171: Weld coverage for scan with index overlapping (Thickness: 25mm)
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #145
The 2D trajectory is automatically defined when the selected scanner has two axis, 3 axis and 1-axis
+ increment.
The parameters are identical to the 1D trajectory except that the user has to define the Distance and
Step for both the Scan and Index axis.
While for the trajectory 1D trajectory + increment the index values are pre-set, for the
2D trajectory the index value will be the value provided by the encoder linked to the index
axis of the scanner.
On Figure 172, the Scan axis is being feed by Coder 1, and the Index axis by Coder 2.
Nevertheless, the increment Step value needs to be set for the equipment trigger the acquisition after
the index have been moved of that value.
Figure 172: Two-axis application (Corrosion mapping with encoder triggered overlapping)
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #146
When using a 3-axis rotating arm, the trajectory is similar to a 2D trajectory after calibration of the
scanner
It is possible recalibrate it by pressing the “Set Sinus” button at the bottom of the screen
Regarding the trajectory, set the scan axis (along R) and second axis distance and step as for
standard 2D Mapping application.
When using the nozzle inspection module, the acquired data axis and respective trajectory paths will
rely on the Inspection face where the inspection will be performed:
- For Set-In nozzle -> Inspection face: Main;
- For Set-On nozzle -> Inspection face: Secondary;
-
The Inspection face is defined on tab Equipment -> Specimen -> Weld.
For further information consult Part 2 Section 1.2.;
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #147
Ɵ – Scan axis position. For further information consult Part 3 Section 1.1.5.
In Figure 174 is illustrated the trajectory panel for an inspection that is being performed from the main
component. In this case the trajectory was set to scan 360°, i.e. all around the circumferential
perimeter of the nozzle connection. The increment Step was defined to 10mm, and the
Distance to 100mm, which means that the coverage area of 100mm will be filled after 10
steps of 10mm.
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #148
When the inspection is being performed from the Secondary component, the scanner type
needs to be changed from 3-axis nozzle scan to 1-axis scanner;
The PRF is the frequency at which the pulsers are firing, it is expressed as the inverse in time
between two excitations, i.e., between two shots (sectorial scan) or two sequences (linear
scan).
The frequency to perform a complete E-Scan or S-Scan is the PRF divided by the number of
sequences or shots within that group. For Multigroup, it is the total number of sequences or
shots that is considered for all the groups.
The PRF is directly related to how fast the probe or probes can move either manually or connected to
a scanner without missing data.
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #149
The maximum scanning speed is limited by the value of the PRF setting. To scan faster, the user
should increase as much as possible the PRF value. The maximum scanning speed can be
calculated by the following expression:
By other hand, the time between two consecutive firing cannot be less the required time that the
sound takes to propagate for the time base range. The maximum PRF that can be applied
during an acquisition without generate “ghost echoes” may calculated by the following
expression:
Capture software calculates automatically the maximum PRF and the maximum scanning speed that
can be achievable with the present configuration taking into consideration:
The PRF can be accessed through Motion panel as well Gates, TCG and Inspection panels. Press
to access to PRF management window located at top right of the screen;
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #150
2. Display
2.1 Layout
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #151
1: Up to four specific tabs are available for the inspection panel to display a large variety of ultrasonic
images dispositions. Press on the button to add a layout panel.
2: Name or rename each layout panel.
3: Each tab can be managed individually and a maximum of six different windows is configurable per
tab. Press the left or right arrows to switch the available layouts (Figure 177).
4: A colour code is associated to each window. For the different views, use the drop-down menus to
select the salvo, the gate and the type of view.
The available views depend from the configuration that is being used.
Table 4.3 summarises the views that are currently available, for all the components, trajectories and
inspection techniques available on CAPTURE:
Trajectory Phased Array TFM UT TOFD
Views
B-Scan Mec.
B-Scan Mec.
EchoDyn Z
EchoDyn X
Side View
Side View
Top View
Map X-Y
Map R-
A-Scan
A-Scan
A-Scan
A-Scan
C-Scan
D-Scan
C-Scan
D-Scan
C-Scan
E-Scan
S-Scan
T-Scan
1D
2D
3D
3D
3D
Component
X X X X X X X X X X X X X X X X X X X X
Plane
X X X X X X X X X X X X X X X X
X X X X X X X X X X X X
Fillet Weld
X X X X X X X X X X X
X X X X X X X X X X X X X X X X X X X X
Cylindrical (Circular Orientation)
X X X X X X X X X X X X X X X X
X X X X X X X X X X X X X X X X X X X X
Cylindrical (Longitudinal Orientation)
X X X X X X X X X X X X X X X X
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #152
To simplify the reading of the data from the different views, the scales are codified by colours to be
easier to read the physical values that encompasses each one of the views, according the
following convention:
In the following paragraphs we will describe shortly each one of the views.
The A-Scan view is the basic representation of any ultrasonic signal. This view represents the
ultrasonic data, where the Y-Axis is Amplitude of the ultrasonic signal [% FSH] or [dB], and
the X-axis is the TOF [µs] or [mm] or the True-Depth [mm];
The Figure 15 illustrate this data representation, where the X-axis is codified as True-Depth, since this
axis is coloured as dark blue.
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #153
The Mechanical B-Scan view is a stack of A-Scan views that is used to represent the ultrasonic data
from the conventional UT techniques as well from the TOFD technique. The colour scheme of
the view provides information from the signal amplitude, the Y-Axis the TOF [µs], and the X-
axis the Mechanical Scan position [mm];
The E-Scan (Electronic Scanning) and the S-Scan (Sectorial Scanning) are the basic views to
represent the ultrasonic data from the Phased Array technique. This view is an angle
corrected view, where the ultrasonic signals are represented considering the refraction angles
for each one of the shots/sequences.
The colour scheme of the view provides information from the signal amplitude, the Y-Axis represents
the True-Depth [mm], and the X-axis represent the Index position [mm];
Although the data representation is similar, normally it is designated as E-Scan for the Linear Scan
inspection, and S-Scan for the Sectorial Scan inspection.
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #154
2.6 C-Scan
The C-Scan view represents the uncorrected view of the ultrasonic data. The C-Scan view displays
always the data from one specific Gate and consider the detection mode of the Gate (First
Echo, Max Echo, etc).
The colour scheme of the view provides either the information from the signal amplitude or the time of
flight or the depth. The Y-Axis represents the Angle or the Sequence, and the X-axis
represent the Mechanical Scan position [mm];
2.7 D-Scan
The D-Scan view represents the uncorrected view of the ultrasonic data. For each selected Angle or
Sequence, one D-Scan view is available.
The colour scheme of the view provides information from the signal amplitude, the Y-Axis represents
the Depth [mm] or the Time of Flight [µs], and the X-axis represent the Mechanical Scan
position [mm];
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #155
The Top-View represents the corrected top projection of the ultrasonic data.
The colour scheme of the view provides either the information from the signal amplitude or the time of
flight or the depth. The view represents a true X-Y projection over specimen [mm];
This view projects the ultrasonic data directly over the component top surface in a corrected projection
that consider the different parameters as Mechanical Position, Index Offset, Probe Orientation,
Refracted Angles, Skew Angles and Surface curvature.
This view uses the raw A-Scan data for the reconstruction, so it might be heavy in terms of
computational processing. By this reason, we recommend to use this view only with Capture
PC version.
The Side-View represents the corrected side projection of the ultrasonic data.
The colour scheme of the view provides information from the signal amplitude. The view represents a
true X-Z projection (when the scanning is done along X) or Y-Z projection (when the
scanning is done along Y) over specimen [mm];
This view projects the ultrasonic data directly over the component side surface in a corrected
projection that consider the different parameters as Mechanical Position, Index Offset, Probe
Orientation, Refracted Angles, Skew Angles and Surface curvature.
This view uses the raw A-Scan data for the reconstruction, so it might be heavy in terms of
computational processing. By this reason, we recommend to use this view only with Capture
PC version.
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #156
The Map X-Y / Map R- view represents the uncorrected view of the ultrasonic data. This view is used
when the acquisition uses one primary axis as the Mechanical Scan axis, and a secondary
axis as the Increment or Index axis. This type of acquisition is used for example in the butt-
weld inspection when the full coverage is achieved from 2 or more index positions, or for the
weld nozzle inspection when the scanner has 3 degrees of freedom. For those cases the C-
Scan is not suitable to display the ultrasonic data and it is replaced by these views.
The Map X-Y / Map R- views display always the data from one specific Gate and consider the
detection mode of the Gate (First Echo, Max Echo, etc). The view Map X-Y / Map R- (at fixed
shot) displays the ultrasonic data for the selected Angle or Sequence.
The colour scheme of the view provides either the information from the signal amplitude or the time of
flight or the depth. The Y-Axis represents the Index position, and the X-axis represent the
Mechanical Scan position [mm];
The 3-D view displays the E-Scan (Electronic Scanning) or the S-Scan (Sectorial Scanning)
superimposed over the 3-D view of the component.
Réf: GEKKO_user_manual_v1.2_US-A
GEKKO manual #157
PART 5.
Inspection
Réf: GEKKO_user_manual_v1.2_US-A
INSPECTION CAPTURE manual #158
PART 5. INSPECTION
Once all the application parameters have been set, CAPTURE allows the inspection to be performed.
The inspection panel is divided into 3 mains zones (Figure 187).
1.1 UT parameters
The default banner shows the current UT settings: delayed time-base sweep (delay), time base range
(range) and Gain. These values have been set in the Gate or Amplitude calibration panel and
cannot be modified. Nevertheless, it is possible to add or subtract gain before launching the
inspection using the +/- box (Figure 188).
Ref: CAPTURE_user_manual_v3.0_US-A
INSPECTION CAPTURE manual #159
The UT parameters displayed are linked to the current salvo: select another salvo to update the UT
parameters.
Press to switch the top banner display and visualize indicator measurement (Figure 189).
This measuring tool allows to access the live values for amplitude, time-of-flight or true depth.
Press on any of the four live gates values to change the type and source of the indicator. A pop-up
menu shows the parameters to be set (Figure 190).
Select in the drop-down menu the salvo and the gate for the measurement.
Press the drop-down menu to select the value to be displayed (Amplitude, Time-of-flight or Depth).
If Amplitude value is set, measurement can be displayed in dB or % with the associated drop down.
When Current shot box is unselected, the value displayed by the indicator is the gate measurement
according to the type of gate setup (first peak, first threshold…), for all sequences (in case of
linear scan) or all shots (in case of sector scan).
When Current shot box is checked, the value displayed by the indicator is the gate measurement
according to the type of gate setup (first peak, first threshold…), for the current selected shot
or sequence (selected from the S-Scan or E-Scan view).
For conventional or TOFD configurations, this parameter is not available. Information mentioned in the
indicator box are detailed below (Figure 191).
Ref: CAPTURE_user_manual_v3.0_US-A
INSPECTION CAPTURE manual #160
For matrix configurations, the skew information is displayed in the bottom right of each indicator box
(Figure 192).
Regardless of whether or not the Current shot box is checked, the measurement always considers all
the available skews.
Ref: CAPTURE_user_manual_v3.0_US-A
INSPECTION CAPTURE manual #161
For each view, generic and specific tools are available from the toolbar buttons. The axis colour code
gives also information to the operator. All these features are presented in the Part 6 Section 4.
3. Inspection management
The Bottom banner is dedicated to data acquisition and management (Figure 194).
4 1 2 3
2: Press to reset the encoder value. The current encoder position is displayed in the adjacent box.
4: Press to access Cursors settings or Offsets settings. This access is also managed by
pressing the physical button MODE on the unit front panel.
Once started, and buttons turn to and buttons (Figure 195) and some actions and
buttons are disabled until the end of the inspection.
When the inspection is completed or stopped by pressing the Square, the following is displayed:
Ref: CAPTURE_user_manual_v3.0_US-A
INSPECTION CAPTURE manual #162
If Save or Save & Analysis is selected, a pop up asks to save the inspection file (Figure 196). Name
the inspection file and select the destination folder (located in the Inspection tab in the Home
page) then press OK to validate.
Once the file is saved without Analysis, the following window appears (Figure 197) allowing to:
• Go back to Settings
• Go back to Configuration
• Cancel and remain in the Inspection Panel
Ref: CAPTURE_user_manual_v3.0_US-A
INSPECTION CAPTURE manual #163
In the Inspection Panel, by pressing Scan, it gives access to Cursors settings and Offset settings.
Ref: CAPTURE_user_manual_v3.0_US-A
INSPECTION CAPTURE manual #164
1. Cursor settings:
1 cursor
2 cursors
3 cursors
2. Offset settings:
Ref: CAPTURE_user_manual_v3.0_US-A
INSPECTION CAPTURE manual #165
PART 6.
Data analysis
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #166
3 1
• Indicators
• Analysis boxes
• Specific tools associated to the various views (colour scales, cursors, rebound…)
• Inspection report
1. Layout edition
By clicking on this icon , it is possible to edit and modify the layout, in the same way as it is
described in the PART 4 2.1.
1
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #167
2. Indicators
The indicators presented in Part 5 Section 1.2 are also available in the Data Analysis. It works similarly
except that instead of returning information about the live signals, they return information about
the recorded signals.
3. Analysis boxes
3.1 Overview
Each indication can be measured in amplitude, length and height thanks to the analysis boxes.
These analysis files are compatible with ENLIGHT and CIVA Analysis for advanced analysis
tools, signal processing.
Then drag the analysis box surrounding the indication in a B-scan or C-scan view (Figure 200). It is not
possible to define an analysis box in an A-scan view.
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #168
Press the icon, at the left bottom of the screen, to list the indications (Figure 201). For each
indication in the list, a square with or without an eye is present at the end of the line. This field
allows activating or deactivating the display of the analysis box in the various views.
Press the icon, at the right bottom of the pop-up window, to add data about the selected indication
or on the to remove the indication from the list.
Select the type of indication in the Tag part of the pop-up window (Figure 202). It is possible to add
remarks or to associate a picture by pressing .
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #169
The picture can be loaded from the screen captures done with the unit (Figure 203) or they can be
imported from a USB flash drive by pressing the button (Part 1 section 6)
The data in the indication list are included in the inspection report by default.
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #170
4. View tools
Table 4 lists all the buttons accessible in the CAPTURE from the views toolbar as well as their
associated availability, with the exception of button available in the 3D views which are
explained in PART 3 Section 1.1.9.
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #171
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #172
Table 5 summarizes the various colour codes available in CAPTURE and their signification.
5. TOFD Analysis
For the salvos that contains TOFD data; a specific menu is available. To access this menu, press the
button “Mode” and select the TOFD option as shown in – TOFD ToolsFigure 204.
The basic TOFD calibration tool allows to recalculate the wedge delay, to match the starting depth value
with the lateral wave signal. To perform the calibration, adjust the gate over the lateral wave
using either the A-Scan or the Mechanical B-Scan view (Start, Width and Threshold), and select
the A-Scan data line that will be used as the reference signal for the calibration. The Figure 205
illustrates the procedure:
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #173
Figure 205 –A-Scan selection as refence for the Wedge Delay calibration
The detection mode and the phase of the signal can be adjusted accessing the icon tools. We
recommend using the same detection mode and phase that were used during the PCS
calibration. Figure 206 illustrates the TOFD settings menu:
After adjusting the gate settings, to perform the calibration press the button ON, and if the result is
satisfactory press OK to accept the changes. The light blue scales (time of flight) became dark
blue and the scale now displays the depth units. Moreover, the TOFD parabolic cursors are
displayed, allowing to dimension properly the indications. The Figure 207 illustrates the result:
In addition to WD-Wedge Delay, the software allows two other calibration modes:
- WD + Lateral Synchro
- WD + Lateral Synchro + Lateral Wave Removal
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #174
The first option proceeds to align all the A-Scan data, using the detection peak from the lateral wave for
all the A-Scans. To activate this option, change the calibration mode from Wedge Delay to WD
+ Lateral Synchro. It is advisable to carefully adjust the gate parameters as the Start, Width
and Threshold to achieve good synchronization results. The Figure 208 illustrates the result:
Figure 208 TOFD Calibration Mode = Wedge Delay + Lateral Synchro (WD+LS)
After the calibration, the TOFD hyperbolic cursors are displayed. The shape of this cursors depends of
the scannnig type Non-Parallel or Paralell, that were previously defined in the menu Motion.
Note: Non-Parallel Scan = Along X
Parallel Scan = Along Y
This cursors are normally used to dimension the flaw length and height and shown in Figure 209 below:
Figure 209 – Length and Height measuring using the parabolic cursors
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #175
6. Inspection report
Press (Figure 210) to open the inspection report (Figure 211).
Ref: CAPTURE_user_manual_v3.0_US-A
DATA ANALYSIS CAPTURE manual #176
On the left hand-size, one can select or unselect the elements to be put in the report.
Scroll the screen to check the content of the report (Figure 212).
Press to save and export the inspection report to a USB flash drive (Figure 213).
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #177
LIST OF ABBREVIATIONS
PE: pulse-echo
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #178
LIST OF FIGURES
Figure 1: GEKKO connectors overview 10
Figure 2: GEKKO front-panel buttons 11
Figure 3: MANTIS front-panel buttons 12
Figure 4: MANTIS connectors overview 12
Figure 5: Home panel 13
Figure 6: Navigation arrows 14
Figure 7: Edit button 15
Figure 8: Backward/Forward buttons 15
Figure 9: Drop-down list 16
Figure 10: Double option button 16
Figure 11: Editable blank box 16
Figure 12: On-screen keyboard 16
Figure 13: Digital scroll wheel 17
Figure 14: Invalid value 17
Figure 15: General settings 19
Figure 16: Wizards “Conventional UT” 21
Figure 17: Wizards “Standard phased array” 22
Figure 18: Wizards “Total Focusing Method (TFM)” 23
Figure 19: Wizards “Multi-groups techniques” 24
Figure 20: Application creation pop-up 25
Figure 21: Open an application file 26
Figure 22: Applications/Inspection files buttons 26
Figure 23: Import files 28
Figure 24: File-import pop-up 28
Figure 25: File selection 29
Figure 26: Export files 30
Figure 27: Screenshot manager 31
Figure 28: Application recovery 32
Figure 29: General overview of the equipment panel 34
Figure 30: Specimen definition 35
Figure 31: Cylindrical geometry 36
Figure 32: Nozzle selection 37
Figure 33: Fillet weld geometry 38
Figure 34: Chamfer and back strip setting 39
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #179
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #180
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #181
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #182
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #183
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #184
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #185
LIST OF TABLES
Ref: CAPTURE_user_manual_v3.0_US-A
CAPTURE manual #186
Ref: CAPTURE_user_manual_v3.0_US-A