Technical Information Liquiline CM442R/CM444R/ CM448R
Technical Information Liquiline CM442R/CM444R/ CM448R
Technical Information
Liquiline CM442R/CM444R/
CM448R
Cabinet controller with a maximum of eight measuring
channels based on digital Memosens technology
Table of contents
2 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . 38
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Degree of contamination . . . . . . . . . . . . . . . . . . . . . . . 39
Mechanical construction . . . . . . . . . . . . . . . . . . . . 39
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
External display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Local operation via external, optional display . . . . . . . . . . 43
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Language packages . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Ordering information . . . . . . . . . . . . . . . . . . . . . . . 46
Product page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Product Configurator . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Measuring cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Additional functionality . . . . . . . . . . . . . . . . . . . . . . . . 51
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Other accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Endress+Hauser 3
Liquiline CM442R/CM444R/CM448R
4 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
1 4
Outlet 1
11:12:13 29.06.2013
CH 1:1 pH 6.99 pH
CH 1:1 Temperature 25.1 °C
CH 1:2 Cond I 9.02 mS/cm 3
2 CH 2:1 Cond I 13.41 mS/cm
CH 2:2 Chlorine 9.03 mg/l
CH 3:1 Current Input [Flow] 15.1 nA
MENU Field Data Manager
10.00
mS/cm
8.00
mS/cm
29.05.2013 29.06.2013
10:02:20 11:09:15
12.00
pH
6.00
pH
29.05.2013 29.06.2013
10:02:20 11:09:15
A0025347
Data retention
• Storage of all measured values, incl. values of external sources, in the non-volatile memory
(data logbook)
• Data called up on site via user-defined measuring menu and load curve display of the data logbook
• Transmission of data by ethernet, CDI interface or SD card and storage in a tamper-proof database
(Field Data Manager)
• Data export to csv file (for Microsoft Excel)
Endress+Hauser 5
Liquiline CM442R/CM444R/CM448R
Device architecture
Slot and port assignment
Slots
BASE2-E 2DS 2DS 2DS 2AI 2AO 4R
Ports
1
1
1
1
1
2
2
2
2
2
Sensor 1
Sensor 1
Sensor 1
3
2
1
Sensor 2
Sensor 2
Sensor 2
1
4
2
Order of the modules Depending on the version ordered, the device is supplied with a number of electronic modules, which
are assigned in a specific sequence in ascending order to slots 0 to 7. If you do not have a particular
module, the next moves up automatically:
• The basic module (which is always present) always occupies slots 0 and 1
• Fieldbus module 485
• Memosens input module 2DS (DS = digital sensor)
• Extension module for digital inputs and outputs DIO (DIO = digital input and output)
• Current input module 2AI (AI = analog input)
• Current output module 4AO or 2AO (AO = analog output)
• Relay modules AOR, 4R or 2R (AOR = analog output + relay, R = relay)
Modules with 4 ports are connected before modules of the same type with 2 ports.
Basic rule for hardware Please note the following if upgrading the device:
upgrades • The sum of all current inputs and outputs may not exceed 8!
• A maximum of two "DIO" modules may be used.
6 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Determining the hardware You must be aware of the type of modules and the number of them supplied with the device you
delivery status have ordered to determine the delivery status of your Liquiline.
• Basic module
One basic module in all versions. Always occupies slots 0 and 1.
• Fieldbus module
Optional, and only one fieldbus module is possible.
• Input modules
• Must be clearly assigned to the number of optional inputs ordered.
• Examples:
2 current inputs = module 2AI
4 Memosens inputs = 2 inputs with basic module + module 2DS with 2 further inputs
• Current outputs and relays
Various module combinations can exist.
The following table will help you determine which modules your device has, depending on the type
and number of outputs.
Relays
Current outputs 0 2 4
2 - 1 x 2R 1 x 4R
‣ Sum up the number of modules and sort them according to the specified sequence → 6.
This will give you the slot assignment for your device.
Slot 1 2 3 4
BASE2-E 2DS 2AO 4R
Relay 1
41
Relay 2
41
43 42
41
Relay 3
43 42
Relay 4
41
43 42
A0039621
5 Creating a terminal diagram using the example of the NO contact (terminal 41) of a relay
Endress+Hauser 7
Liquiline CM442R/CM444R/CM448R
A0041706
Extension options by using an PROFIBUS DP or Modbus RS485 with module 485 incl. activation
extension module in free slot 2 code:
• PROFIBUS DP (71140888)
• Modbus RS485 (71140889)
Ifdisabled!
module 485 is retrofitted, any existing current outputs are
Basic rule for extensions The sum of all current inputs and outputs may not exceed 8.
Restrictions if using CUS71D sensors Only one CUS71D can be connected. The second Memosens input
for interface measurement may not be used.
8 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
31
mA
2 32 7
87 85
88
86
3 97 8
98
87 85
88
86
3 97 8
98
optional
41
42
4 9
43
optional
41
31
mA
42
2 32 10
43
31 41
mA
42
2 32 10
43
optional optional
41
23 mA
42
5 24 10
43
41
23 mA
42
5 24 10
43
optional optional
45 47
+
46 48
–
45 47
+
46 48
– 11
91
+
– 92
91
+
92
optional –
A0039427
Endress+Hauser 9
Liquiline CM442R/CM444R/CM448R
Modification options without Communication type changed by entering activation code. This
additional modules disables the communication type used previously!
Modbus RS485 + web server (71135636)
Retrofit by removing module 485 and entering the activation code
for:
• Modbus TCP + web server (71449915)
• EtherNet/IP + web server (71449914)
• PROFINET + web server (71449901)
• HART (71128428)
• Web server (71449918)
Extension options by using extension Only the following is possible for the example above:
modules in free slots 5-7 • Module 2R (71125375) or 4R (71125376): 2 or 4 relays
• Module DIO (71135638): 2 digital inputs and 2 digital outputs
If extending to 8 measuring channels:
Module 2DS (71135631): 2 Memosens inputs
Additional inputs or outputs and relays if fieldbus module 485 is
removed:
• Module 2AO (71135632): 2 current outputs
• Module AOR (71111053): 2 current outputs, 2 relays
• Module 2R (71125375) or 4R (71125376): 2 or 4 relays
• Module DIO (71135638): 2 digital inputs and 2 digital outputs
Ifused
module 485 is removed and an Ethernet-based fieldbus is
via BASE2-E module, a maximum of up to 6 current
outputs can be operated in addition. Only two current outputs
are possible with module 485.
Basic rule for extensions The sum of all current inputs and outputs may not exceed 8.
Restrictions if using CUS71D sensors • In the case of CM444R, every combination of Memosens sensors
for interface measurement (max. 4) is possible.
• An extension to CM448R is not advisable as the maximum number
of Memosens inputs when using CUS71D remains limited to 4.
10 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
31
mA
2 32 7
optional
87 85
88
3 97 86 8
98
87 85
88
3 97 86 8
98
87 85
88
3 97 86 8
98
87 85
88
3 97 86 8
98
optional
41
42
4 9
43
optional
31 41
mA
42
2 32 10
43
31 41
mA
42
2 32
43 10
optional optional
31 41
mA
42
2 32 10
43
31 41
mA
42
2 32 10
43
optional optional
31
mA 23 mA
2 32 24 5
31
mA 23 mA
2 32 24 5
optional optional
45 47 +
+
46 48
– –
45
+ 47
46 +
– 48
91 – 11
+
– 92
91
+
– 92 optional
A0039431
Endress+Hauser 11
Liquiline CM442R/CM444R/CM448R
Extension options without additional Activation code for the use of current outputs of the basic module:
modules 2 current outputs (71140891)
Modification options without Communication type changed by entering activation code. This
additional modules disables the communication type used previously!
Modbus RS485 + web server (71135636)
Retrofit by removing module 485 and entering the activation code for
communication via the BASE2 module:
• Modbus TCP + web server (71449915)
• EtherNet/IP + web server (71449914)
• PROFINET + web server (71449901)
• HART (71128428)
• Web server (71449918)
Ifused,
module 485 is removed and an Ethernet-based fieldbus is
a maximum of up to 6 current outputs can be operated in
addition. Only two current outputs are possible with module
485.
Basic rule for extensions The sum of all current inputs and outputs may not exceed 8.
Restrictions if using CUS71D sensors The maximum number of Memosens inputs that can be used is
for interface measurement limited to 4! Here, every combination of CUS71D and other
Memosens sensors is then possible.
12 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
A0039429
Endress+Hauser 13
Liquiline CM442R/CM444R/CM448R
Dependability
Reliability Memosens MEMO SENS
Heartbeat diagnostics
• Heartbeat diagnostics screen with graphic indicators for the health of the device and sensor and
with a maintenance or (sensor-dependent) calibration timer
• Heartbeat status information on the health of the device and the condition of the sensor
• : Sensor/device condition or maintenance timer > 20 %; no action is required
• : Sensor/device condition or maintenance timer > 5 ≤ 20 %, maintenance not yet urgent but
should be scheduled
• : Sensor/device condition or maintenance timer < 5 %, maintenance is recommended
• The Heartbeat sensor condition is the assessment of the calibration results and the sensor
diagnostic functions.
14 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
An unhappy smiley can be due to the calibration result, the measured value status or to the
operating hours limit having been exceeded. These limits can be configured in the sensor setup in a
way that adapts the Heartbeat diagnostics to the application.
Heartbeat and NAMUR category
The Heartbeat status indicates the sensor or device condition while the NAMUR categories (F, C, M,
S) assess the reliability of the measured value. The two conditions can correlate but do not have to.
• Example 1
• The number of remaining cleaning cycles for the sensor reaches 20% of the defined maximum
number. The Heartbeat symbol changes from to . The measured value is still reliable so the
NAMUR status signal does not change.
• If the maximum number of cleaning cycles is exceeded, the Heartbeat symbol changes from
to . While the measured value can still be reliable, the NAMUR status signal changes to M
(maintenance required).
• Example 2
The sensor breaks. The Heartbeat status changes immediately from to and the NAMUR
status signal also changes immediately to F (failure).
Heartbeat Monitoring
Sensor data from Memosens sensors are transmitted via the EtherNet/IP, PROFINET, PROFIBUS DP,
HART, Modbus RTU and Modbus TCP fieldbus protocols. These data can be used for predictive
maintenance, for instance.
Examples include:
• Total hours of operation
• Hours of operation with very high or very low measured values
• Hours of operation at high temperatures
• Number of steam sterilizations
• Sensor identification
• Calibration information
For detailed information on "Ethernet/IP communication", see the product pages on the Internet
(→ SD01293C).
For detailed information on "Modbus communication", see the product pages on the Internet (→
SD01189C).
For detailed information on "PROFINET communication", see the product pages on the internet
(→ SD02490C).
For detailed information on "PROFIBUS communication", see the product pages on the Internet
(→ SD01188C).
More detailed information on HART communication is provided on the product pages on the
Internet (→ SD01187C).
Heartbeat Verification
Heartbeat Verification makes it possible to verify the correct operation of the measuring device
without interrupting the process. This verification can be documented anytime.
Endress+Hauser 15
Liquiline CM442R/CM444R/CM448R
Self-monitoring functions
Current inputs are deactivated in the event of overcurrent and reactivated once the overcurrent
stops. Board voltages are monitored and the board temperature is also measured.
USP and EP
The limit functions for pharmaceutical water in accordance with USP and EP specifications are
implemented in the software for conductivity measurements:
• "Water for Injection" (WFI) as per USP <645> and EP
• "Highly Purified Water" (HPW) as per EP
• "Purified Water" (PW) as per EP
The uncompensated conductivity value and the temperature are measured for the USP/EP limit
functions. The measured values are compared against the tables defined in the standards. An alarm
is triggered if the limit value is exceeded. Furthermore, it is also possible to configure an early
warning alarm that signals undesired operating states before they occur.
ChemocleanPlus
Freely programmable sequence control
• e.g. for automatic sensor cleaning in retractable assemblies for reliable measurement results in
processes with a high risk of contamination
• Individual, time-based activation of 4 outputs e.g. relays
• Starting, stopping or pausing of activities via digital input or fieldbus signals e.g. from limit
position switches
A0042325 A0042346
16 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
A0015032
SD card
The exchangeable storage medium enables:
• Quick and easy software updates and upgrades
• Data storage of internal device memory (e.g. logbooks)
• Transfer of complete configurations to a device with an identical setup (backup function)
• Transfer of configurations without the TAG and bus address to devices with an identical setup
(copy function)
• Saving of screenshots for documentation purposes
Endress+Hauser offers industry-approved SD cards as accessories. These memory cards provide
maximum data security and integrity.
Other SD cards up to a maximum weight of 5 g can also be used. However, Endress+Hauser does not
accept any responsibility for the data security of such cards.
External signals for device control and for activating external devices
Hardware options, e.g. module "DIO" with 2 digital inputs and 2 digital outputs or fieldbus module
"485" enable the following:
• via a digital input signal
• measuring range switching for conductivity (upgrade code required, see accessories)
• switching between different calibration datasets in the case of optical sensors
• an external hold
• a cleaning interval to be triggered
• switching on and off a PID controller, e.g. via the proximity switch of the CCA250
• the use of the input as an "analog input" for pulse-frequency modulation (PFM)
• via a digital output signal
• the static transmission (similar to a relay) of diagnostic states, point level switch states etc.
• the dynamic transmission (comparable to a non-wearing "analog output") of PFM signals, e.g. to
control dosing pumps.
Endress+Hauser 17
Liquiline CM442R/CM444R/CM448R
A0016009
18 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Concentration tables
When the device is delivered from the factory, tables are saved in the device that allow inductive
conductivity measurements to be converted to concentrations of certain substances. 4 user-defined
tables are also possible.
Data security
All settings, logbooks etc. are stored in a non-volatile memory to ensure that the data are retained
even in the event of a disruption to the power supply.
Password protection
Password-protected login
• For remote operation via web server
• For local operation
Process safety
Two independent PID controllers
• One- or two-sided control
• Limit switches
• 4 cleaning programs which can be programmed independently of each other
IT security
Our warranty is valid only if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the settings.
IT security measures, which provide additional protection for the device and associated data transfer,
must be implemented by the operators themselves in line with their security standards.
Endress+Hauser 19
Liquiline CM442R/CM444R/CM448R
Input
Measured variables → Documentation of the connected sensor
Types of input • Digital sensor inputs for sensors with Memosens protocol
• Analog current inputs (optional)
• Digital inputs (optional)
Cable length
Max. 100 m (330 ft)
Span • High: 11 to 30 V DC
• Low: 0 to 5 V DC
20 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Output
Output signal Depending on version:
• 2 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
• 4 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
• 6 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
• 8 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
• Optional HART communication (only via current output 1:1)
HART
PROFIBUS DP/RS485
Data transmission rate 9.6 kBd, 19.2 kBd, 45.45kBd, 93.75 kBd, 187.5 kBd, 500 kBd,
1.5 MBd, 6 MBd, 12 MBd
Modbus RS485
Data transmission rate 2,400, 4,800, 9,600, 19,200, 38,400, 57,600 and 115,200
baud
Connection RJ45
EtherNet/IP
Connection RJ45
Endress+Hauser 21
Liquiline CM442R/CM444R/CM448R
PROFINET
Connection RJ45
Name of station Via DCP protocol using the configuration tool (e.g. Siemens
PRONETA)
IP address Via DCP protocol using the configuration tool (e.g. Siemens
PRONETA)
Linearization/transmission Linear
behavior
External power supply When using an onsite auxiliary voltage supply and an onsite digital input:
Recommended minimum auxiliary voltage = 3 V + VIHmin
(VIHmin = minimum input voltage required (high-level input voltage)
22 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Endress+Hauser 23
Liquiline CM442R/CM444R/CM448R
Relay outputs
Electrical specification Relay types
• 1 single-pin changeover contact (alarm relay)
• 2 or 4 single-pin changeover contacts (optional with extension modules)
Maximum load
• Alarm relay: 0.5 A
• All other relays: 2.0 A
Relay switching capacity
0.5 A 450,000
0.5 A 650,000
0.5 A 350,000
Extension modules
Switching voltage Load (max.) Switching cycles (min.)
0.5 A 450,000
2A 120,000
0.5 A 650,000
2A 170,000
0.5 A 350,000
2A 150,000
Protocol-specific data
HART Manufacturer ID 11h
24 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Output data 16 measured values (value, unit, status), 8 digital values (value,
status)
Input data 4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
TCP connections 3
Protocol TCP
Output data 16 measured values (value, unit, status), 8 digital values (value,
status)
Input data 4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Manufacturer ID 0x049Eh
Polarity Auto-MIDI-X
Connections CIP 12
I/O 6
Explicit message 6
Multicast 3 consumers
Endress+Hauser 25
Liquiline CM442R/CM444R/CM448R
PROFINET Protocol "Application layer protocol for decentral device periphery and
distributed automation", PNIO Version 2.34
Manufacturer ID 0x11h
26 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Web server The Web server enables full access to the device configuration, measured values, diagnostic
messages, logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G routers with a user-
defined IP address.
TCP port 80
Power supply
Supply voltage CM442 R
Depending on version:
• 100 to 230 V AC, 50/60 Hz
Maximum permitted fluctuation of mains supply voltage: ± 15 % of nominal voltage
• 24 V AC/DC, 50/60 Hz
Maximum permitted fluctuation of mains supply voltage: + 20/- 15 % of nominal voltage
CM444 R and CM448 R
Depending on the version, via external DIN rail power unit:
• 100 to 230 V AC, 50/60 Hz
Maximum permitted fluctuation of mains supply voltage: ± 15 % of nominal voltage 1)
• 24 V DC
Maximum permitted fluctuation of mains supply voltage: + 20/- 15 % of nominal voltage 1)
NOTICE
The device does not have a power switch!
‣ Provide a protected circuit breaker in the vicinity of the device at the place of installation.
‣ The circuit breaker must be a switch or power switch, and must be labeled as the circuit breaker
for the device.
‣ At the supply point, the power supply must be isolated from dangerous live cables by double or
reinforced insulation in the case of devices with a 24 V supply voltage.
Endress+Hauser 27
Liquiline CM442R/CM444R/CM448R
28 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
1 2 3 1 2 3
9 4 10 4
8 5 9 5
7 6 8 7 6
A0040639 A0040640
13 Basic module BASE2-H or -L (two-channel 14 Basic module BASE2-E (four- and eight-
device) channel device)
1 Power supply for digital fixed cable sensors with 1 Power supply for digital fixed cable sensors with
Memosens protocol Memosens protocol
2 SD card slot 2 SD card slot
3 Slot for display cable 1) 3 Slot for display cable 1)
4 Ethernet interface 4 Ethernet interface
5 Connections for 2 Memosens sensors 5 Connections for 2 Memosens sensors
6 Current outputs 6 Current outputs
7 Power connection 7 Socket for internal supply cable 2)
8 Service interface 8 LEDs
9 Alarm relay connection 9 Service interface
10 Alarm relay connection
1) For optional external display.
2) Power supply to DIN rail power unit.
Endress+Hauser 29
Liquiline CM442R/CM444R/CM448R
Sensor supply
1 PK 85 +
Display
GY 86 –
2 PK 85 +
SD
GY 86 –
Ethernet
41
Service
Alarm
42
43
+ BN 87 87 BN +
Sensor 2
Sensor 1
– WH 88 88 WH –
A GN 97 97 GN A
B YE 98 98 YE B
0/4 ... 20 mA
HART
+
PE N/- L/+
L/+ L 31
1
Power
N/– N 32 –
PE * 31 +
2 32 –
A0039665 A0039625
15 Connecting power supply on the BASE2-H or - 16 Overall wiring diagram for BASE2-H or -L
L
H Power unit 100 to 230 VAC
L Power unit 24 VAC or 24 VDC
1 PK 85 + Display
GY 86 –
2 PK 85 +
SD
*
GY 86 –
Ethernet
41
Service
Alarm
42
43
+ BN 87 87 BN +
Sensor 2
Sensor 1
– WH 88 88 WH –
+ RD A
(internal)
A GN 97 97 GN
Power
- BK B YE 98 98 YE B
L N
0/4 ... 20 mA
HART
31 +
1
32 –
31 +
2 32 –
Power
L/+ N/– PE B
L/+ N/– *
A0039668 A0039624
17 Connecting power supply with BASE2-E 18 Overall wiring diagram for BASE2-E and
external power unit (B)
* Assignment depending on power unit, make sure
to connect correctly
The two device versions may only be operated with the power unit supplied and the power unit
cable. Also pay attention to the information in the operating manual supplied for the power
unit.
30 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Connecting optional modules With extension modules you can purchase additional functions for your device.
NOTICE
Unacceptable hardware combinations (due to conflicts in power supply)
Incorrect measurements or total failure of the measuring point as a result of heat build-up or
overloading
‣ If you are planning to extend your controller, make sure the resulting hardware combination is
permitted (Configurator at www.endress.com/CM442R or .../CM444R or .../CM448R).
‣ Remember that the sum of all current inputs and outputs may not exceed 8.
‣ Make sure not to use more than 2 "DIO" modules. More "DIO" modules are not permitted.
‣ Please contact your Endress+Hauser sales center should you have any questions.
Overview of all the modules available
Module name
31 41 41 85 47
Sensor supply
0/4 ... 20 mA
+1 1+ PK + 1
Relay 1
Relay 1
32 – 42 42 48
– 86 GY –
43 43 85 47 +
31 +
2 2+ PK 2
41 41
32 – – 86 GY 48
Relay 2
–
Relay 2
42 42
43 43 + 87 BN 45 +
1
Sensor 1
41
– 88 WH 46 –
Relay
41
42 1 A 97 GN 45 +
Relay 3
42 98 YE 46 – 2
43 B
43
41 + 87 BN 91 +
Sensor 2
41 1
– 88 WH
Relay
Relay 4
92
42 2 –
42 A 97 GN 91 +
43 43 98 YE 2
B 92
–
Endress+Hauser 31
Liquiline CM442R/CM444R/CM448R
Module name
24 23
24 23
• 2 x 0/4 to 20 mA • 4 x 0/4 to 20 mA • 2 x 0/4 to 20mA • Ethernet (web
analog outputs analog outputs analog inputs server or Modbus
• Order No. • Order No. • Order No. TCP)
71135632 71135633 71135639 • 5V power supply
for PROFIBUS DP
termination
• RS485 (PROFIBUS
DP or Modbus
RS485)
• Use of BASE2
module disables
Ethernet port of
module 485
• Order No.
71135634
31 31 23
Ethernet
+1 +1 +
0/4 ... 20 mA
0/4 ... 20 mA
0/4 ... 20 mA
1
32 32 24
– – –
1
PWR
31+ 31 + 23+ 2
DIP switch
2 2 2 4
8
32 32 24
T COM SF BF
on
– – – 16
32
31 64
+3 128/SW
0/4 ... 20 mA
32
– Service
31 + Termi- DGND 82
4 nation VP 81
32
– 96 96'
DP/
95 RS485 95'
99 99'
32 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
A0025366
Endress+Hauser 33
Liquiline CM442R/CM444R/CM448R
PK
86 85 86 85
86 85 86 85
GY
BN
98 97 88 87
98 97 88 87
WH
GN GN
YE YE
Sensor Sensor
A0039629 A0039622
20 sensors without additional supply voltage 21 sensors with additional supply voltage
2DS
PK
85
1
GY
86
85
2
86
Sensor 1 Sensor
98 97 88 87
GN
YE
Sensor 2
BN
98 97 88 87
WH
GN Sensor
YE
A0033206
34 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Performance characteristics
Response time Current outputs
t90 = max. 500 ms for an increase from 0 to 20 mA
Current inputs
t90 = max. 330 ms for an increase from 0 to 20 mA
Digital inputs and outputs
t90 = max. 330 ms for an increase from low to high
Frequency tolerance of ≤ 1%
digital inputs and outputs
Installation
Mounting on DIN rail as per NOTICE
IEC 60715 Incorrect mounting location in the cabinet, spacing regulations not observed
Possible malfunctions as a result of heat buildup and interference from neighboring devices!
‣ Do not position the device directly above sources of heat. The temperature specification must be
observed.
‣ The components are designed for convection-based cooling. Avoid heat buildup. Ensure openings
are not covered, e.g. by cables.
‣ Observe the specified distances to other devices.
‣ Physically separate the device from frequency converters and high-voltage devices.
‣ Recommended installation direction: horizontal. The specified ambient conditions, and
particularly the ambient temperatures, only apply for horizontal installation.
‣ Vertical orientation is also possible. However, this requires additional fixing clips at the place of
installation to hold the device in position on the DIN rail.
‣ Recommended installation of power unit for CM444R and CM448R: to the left of the device.
Endress+Hauser 35
Liquiline CM442R/CM444R/CM448R
min. 50 (1.97)
min. 50 (1.97)
min. 50 (1.97)
min. 20 (0.79)
min. 20 (0.79)
A0039735
Wall mounting
83 (3.27) 205.5 (8.09)
100 (3.94)
100 (3.94)
75 (2.95)
75 (2.95)
4 x ! 4.5 (0.18) 4 x ! 4.5 (0.18)
A0025370
36 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Mounting the external The mounting plate also serves as the drilling template. The marks on the side help you mark
display the position of the drill holes.
140 (5.51)
105 (4.13)
4 x min. Ø 8 (0.31)
min. Ø 15 (0.59) a
140 (5.51)
105 (4.13)
b
A0025371
Environment
Ambient temperature range CM442R
0 to 60 ˚C (32 to 140 ˚F)
CM444R
• Generally 0 to 55 ˚C (32 to 130 ˚F), with the exception of packages under the second list item
• 0 to 50 ˚C (32 to 120 ˚F) for the following packages:
• CM444R-**M40A7FI*+...
• CM444R-**M40A7FK*+...
• CM444R-**M4AA5F4*+...
• CM444R-**M4AA5FF*+...
• CM444R-**M4AA5FH*+...
• CM444R-**M4AA5FI*+...
• CM444R-**M4AA5FK*+...
• CM444R-**M4AA5FM*+...
• CM444R-**M4BA5F4*+...
• CM444R-**M4BA5FF*+...
• CM444R-**M4BA5FH*+...
• CM444R-**M4BA5FI*+...
• CM444R-**M4BA5FK*+...
• CM444R-**M4BA5FM*+...
• CM444R-**M4DA5F4*+...
• CM444R-**M4DA5FF*+...
• CM444R-**M4DA5FH*+...
• CM444R-**M4DA5FI*+...
• CM444R-**M4DA5FK*+...
• CM444R-**M4DA5FM*+...
Endress+Hauser 37
Liquiline CM442R/CM444R/CM448R
CM448R
• Generally 0 to 55 ˚C (32 to 130 ˚F), with the exception of packages under the second list item
• 0 to 50 ˚C (32 to 120 ˚F) for the following packages:
• CM448R-***6AA*+...
• CM448R-***8A4*+...
• CM448R-***8A5*+...
• CM448R-**28A3*+...
• CM448R-**38A3*+...
• CM448R-**48A3*+...
• CM448R-**58A3*+...
• CM448R-**68A3*+...
• CM448R-**26A5*+...
• CM448R-**36A5*+...
• CM448R-**46A5*+...
• CM448R-**56A5*+...
• CM448R-**66A5*+...
• CM448R-**22A7*+...
• CM448R-**32A7*+...
• CM448R-**42A7*+...
• CM448R-**52A7*+...
• CM448R-**62A7*+...
• CM448R-**A6A5*+...
• CM448R-**A6A7*+...
• CM448R-**B6A5*+...
• CM448R-**B6A7*+...
• CM448R-**C6A5*+...
• CM448R-**C6A7*+...
• CM448R-**D6A5*+...
• CM448R-**D6A7*+...
External display (optional)
–20 to 60 ˚C (0 to 140 ˚F)
Wall mounting
Frequency range 10 to 150 Hz (sinusoidal)
Amplitude 10 to 12.9 Hz: 0.75 mm
12.9 to 150 Hz: 0.5 g 1)
Test duration 10 frequency cycles/ spatial axis, in 3 spatial axes (1 oct./min)
Electromagnetic Interference emission and interference immunity as per EN 61326-1:2013, Class A for Industry
compatibility
38 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Optional display
The product is suitable for pollution degree 4.
Mechanical construction
Dimensions CM442R
97 (3.82)
89 (3.50)
91 (3.58)
7.5 (0.30)
TS 35
83 (3.27)
100 (3.94)
114 (4.49)
75 (2.95)
TS 35
99 (3.90)
A0039729
26 Dimensions in mm (inch)
Endress+Hauser 39
Liquiline CM442R/CM444R/CM448R
205.5 (8.09)
114 (4.49)
100 (3.94)
75 (2.95)
TS 35
97 (3.82)
89 (3.50)
91 (3.58)
7.5 (0.30)
TS 35
220 (8.66)
A0039730
27 Dimensions in mm (inch)
40 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Optional display
54 (2.13)
39.5 (1.56)
144 (5.67) 144 (5.67)
A0025346
28 Dimensions in mm (inch)
90 (3.54)
121 (4.76)
A0025738 A0025739
out
124 (4.88)
75 (2.95)
in
A0025784 A0025786
Endress+Hauser 41
Liquiline CM442R/CM444R/CM448R
193 (7.60)
123 (4.84) 36 (1.42)
B 13.5 (0.53)
B-B
A A
157 (6.18)
119 (4.69)
143 (5.63)
6
R1
40 (1.57)
B
47 (1.85)
144.5 (5.69)
A-A
A0025343
42 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Operability
External display Graphic display:
• Resolution: 240 x 160 pixel
• Back light with switch-off function
• Red display background for alarms alerts users to errors
• Transflective display technology for maximum contrast even in bright environments
• User-definable measuring menus mean you can always keep track of the values that are important
for your application.
Operating concept The simple and structured operating concept sets new standards:
• Intuitive operation with the navigator and soft keys
• Fast configuration of application-specific measurement options
• Easy configuration and diagnosis thanks to plain-text display
• All languages that can be ordered are available in every device
A0025229
35 Plain-text menu
34 Operation via external, optional display
Outlet 1
11:09:15 29.01.2013
CH 1:1 pH 6.99 pH
CH 1:1 Temperature 25.1 °C
CH 1:2 Nitrate 9.02 mg/l
CH 2:1 SAC 2.02 1/m
CH 2:2 Turbidity 1022 FNU
CH 3:1 Current Input [Flow] 15.1 nA
MENU
2 3
A0025231
36 Overview of operation
1 Display (with red display background in alarm condition)
2 Soft keys (function depends on menu)
3 Navigator (jog/shuttle and press/hold function)
Endress+Hauser 43
Liquiline CM442R/CM444R/CM448R
Remote operation Via HART (e.g. via HART modem and FieldCare)
3
A0039620
Via PROFIBUS DP
Master class 1
(PLC, DCS)
Master class 2 (FieldCare, PDM)
PROFIBUS DP
(RS485)
T T
FXA291
Ethernet
A0039617
38 PROFIBUS DP
T Terminating resistor
44 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Modbus RS485
(RTU, ASCII)
T T
FXA291
Ethernet
A0039615
39 Modbus RS485
T Terminating resistor
Ethernet client
FieldCare/
Webbrowser
Ethernet
FXA291
A0039616
Language packages The language selected in the product structure is the operating language preset at the factory. All
other languages can be selected using the menu.
• English (US)
• German
• Chinese (Simplified, PR China)
• Czech
• Dutch
• French
• Italian
• Japanese
• Polish
• Portuguese
• Russian
• Spanish
• Swedish
• Turkish
Endress+Hauser 45
Liquiline CM442R/CM444R/CM448R
• Hungarian
• Croatian
• Vietnamese
The availability of other languages can be checked via the product structure at www.endress.com/
cm442R or .../cm444R or .../cm448R.
EAC The product has been certified according to guidelines TP TC 004/2011 and TP TC 020/2011 which
apply in the European Economic Area (EEA). The EAC conformity mark is affixed to the product.
cCSAus The device has been certified with regard to its electrical safety and for NI Class I Div. 2 cCSAus
explosion-proof environments. It meets the requirements in accordance with:
• CLASS 2252 06 - Process Control Equipment
• CLASS 2252 86 - Process Control Equipment - Certified to US Standards
• CLASS 2258 03 - Process Control Equipment - Intrinsically Safe and Non-incendive Systems - For
Hazardous Locations
• CLASS 2258 83 - Process Control Equipment - Intrinsically Safe and Non-incendive Systems - For
Hazardous Locations - Certified to US Standards
• FM3600
• FM3611
• FM3810
• UL50E
• IEC 60529
• CAN/CSA-C22.2 No. 0
• CAN/CSA C22.2 No. 94
• CSA Std. C22.2 No. 213
• CAN/CSA-C22.2 No. 61010-1
• CAN/CSA-C22.2 No. 60529
• UL/ANSI/ISA 61010-1
• ANSI - ISA 12 12 01
‣ The device must be installed in a housing or cabinet where access is only possible with a tool or
key.
Marine approvals A selection of the devices and sensors have type approval for marine applications, issued by the
following classification societies: ABS (American Bureau of Shipping), BV (Bureau Veritas), DNV-GL
(Det Norske Veritas-Germanischer Lloyd) and LR (Lloyd’s Register). Details of the order codes of the
approved devices and sensors, and the installation and ambient conditions, are provided in the
relevant certificates for marine applications on the product page on the Internet.
Ordering information
Product page www.endress.com/cm442r
www.endress.com/cm444r
www.endress.com/cm448r
Product Configurator On the product page there is a Configure button to the right of the product image.
1. Click this button.
The Configurator opens in a separate window.
46 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
2. Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.
3. Export the order code as a PDF or Excel file. To do so, click the appropriate button on the right
above the selection window.
For many products you also have the option of downloading CAD or 2D drawings of the
selected product version. Click the CAD tab for this and select the desired file type using
picklists.
Accessories
The following are the most important accessories available at the time this documentation was
issued.
‣ For accessories not listed here, please contact your Service or Sales Center.
Memosens CPS31D
• pH electrode with gel-filled reference system with ceramic diaphragm
• Product Configurator on the product page: www.endress.com/cps31d
Technical Information TI00030C
Ceraliquid CPS41D
• pH electrode with ceramic junction and KCl liquid electrolyte
• Product Configurator on the product page: www.endress.com/cps41d
Technical Information TI00079C
2) The external display can be selected as an option in the order structure or ordered as an accessory at a later stage.
Endress+Hauser 47
Liquiline CM442R/CM444R/CM448R
Ceragel CPS71D
• pH electrode with reference system including ion trap
• Product Configurator on the product page: www.endress.com/cps71d
Technical Information TI00245C
Memosens CPS171D
• pH electrode for bio-fermenters with digital Memosens technology
• Product Configurator on the product page: www.endress.com/cps171d
Technical Information TI01254C
Orbipore CPS91D
• pH electrode with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps91d
Technical Information TI00375C
Orbipac CPF81D
• Compact pH sensor for installation or immersion operation
• In industrial water and wastewater
• Product Configurator on the product page: www.endress.com/cpf81d
Technical Information TI00191C
Enamel pH electrodes
Ceramax CPS341D
• pH electrode with pH-sensitive enamel
• Meets highest demands of measuring accuracy, pressure, temperature, sterility and durability
• Product Configurator on the product page: www.endress.com/cps341d
Technical Information TI00468C
ORP sensors
Orbisint CPS12D
• ORP sensor for process technology
• Product Configurator on the product page: www.endress.com/cps12d
Technical Information TI00367C
Ceraliquid CPS42D
• ORP electrode with ceramic junction and KCl liquid electrolyte
• Product Configurator on the product page: www.endress.com/cps42d
Technical Information TI00373C
Ceragel CPS72D
• ORP electrode with reference system including ion trap
• Product Configurator on the product page: www.endress.com/cps72d
Technical Information TI00374C
Orbipac CPF82D
• Compact ORP sensor for installation or immersion operation in process water and wastewater
• Product Configurator on the product page: www.endress.com/cpf82d
Technical Information TI00191C
Orbipore CPS92D
• ORP electrode with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps92d
Technical Information TI00435C
48 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
pH ISFET sensors
Tophit CPS441D
• Sterilizable ISFET sensor for low-conductivity media
• Liquid KCl electrolyte
• Product Configurator on the product page: www.endress.com/cps441d
Technical Information TI00352C
Tophit CPS471D
• Sterilizable and autoclavable ISFET sensor for food and pharmaceutics, process engineering
• Water treatment and biotechnology
• Product Configurator on the product page: www.endress.com/cps471d
Technical Information TI00283C
Tophit CPS491D
• ISFET sensor with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps491d
Technical Information TI00377C
Memosens CPS76D
• Combined pH/ORP sensor for process technology
• Hygienic and sterile applications
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps76d
Technical Information TI00506C
Memosens CPS96D
• Combined pH/ORP sensor for chemical processes
• With poison-resistant reference with ion trap
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps96d
Technical Information TI00507C
Indumax H CLS54D
• Inductive conductivity sensor
• With certified, hygienic design for foodstuffs, beverages, pharmaceuticals and biotechnology
• Product Configurator on the product page: www.endress.com/cls54d
Technical Information TI00508C
Endress+Hauser 49
Liquiline CM442R/CM444R/CM448R
Condumax CLS16D
• Hygienic, conductive conductivity sensor
• For pure water, ultrapure water and Ex applications
• With EHEDG and 3A approval
• Product Configurator on the product page: www.endress.com/CLS16d
Technical Information TI00227C
Condumax CLS21D
• Two-electrode sensor in plug-in head version version
• Product Configurator on the product page: www.endress.com/CLS21d
Technical Information TI00085C
Memosens CLS82D
• Four-electrode sensor
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cls82d
Technical Information TI01188C
Oxygen sensors
Oxymax COS22D
• Sterilizable sensor for dissolved oxygen
• With Memosens technology or as an analog sensor
• Product Configurator on the product page: www.endress.com/cos22d
Technical Information TI00446C
Oxymax COS51D
• Amperometric sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos51d
Technical Information TI00413C
Oxymax COS61D
• Optical oxygen sensor for drinking water and industrial water measurement
• Measuring principle: quenching
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos61d
Technical Information TI00387C
Memosens COS81D
• Sterilizable, optical sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos81d
Technical Information TI01201C
50 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
Disinfection sensors
CCS142D
• Membrane-covered amperometric sensor for free chlorine
• Measuring range 0.01 to 20 mg/l
• With Memosens technology
• Product Configurator on the product page: www.endress.com/ccs142d
Technical Information TI00419C
Ion-selective sensors
ISEmax CAS40D
• Ion selective sensors
• Product Configurator on the product page: www.endress.com/cas40d
Technical Information TI00491C
Turbidity sensors
Turbimax CUS51D
• For nephelometric measurements of turbidity and solids in wastewater
• 4-beam scattered light method
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cus51d
Technical Information TI00461C
Turbimax CUS52D
• Hygienic Memosens sensor for turbidity measurement in drinking water, process water and in
utilities
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cus52d
Technical Information TI01136C
Interface measurement
Turbimax CUS71D
• Immersion sensor for interface measurement
• Ultrasonic interface sensor
• Product Configurator on the product page: www.endress.com/cus71d
Technical Information TI00490C
Endress+Hauser 51
Liquiline CM442R/CM444R/CM448R
52 Endress+Hauser
Liquiline CM442R/CM444R/CM448R
SD card
• Industrial Flash Drive, 1 GB
• Order No. 71110815
Communication-specific accessories
Commubox FXA195
Intrinsically safe HART communication with FieldCare via the USB port
Technical Information TI00404F
Commubox FXA291
Connects the CDI interface of measuring devices with the USB port of the computer or laptop
Technical Information TI00405C
3) The external display can be selected as an option in the product structure or ordered subsequently as an accessory.
Endress+Hauser 53
Liquiline CM442R/CM444R/CM448R
System components
RIA14, RIA16
• Field display unit for integration into 4-20 mA circuits
• RIA14 in flameproof metal enclosure
Technical Information TI00143R and TI00144R
RIA15
• Process display unit, Digital display unit for integration into 4-20 mA circuits
• Panel mounting
• With optional HART communication
Technical Information TI01043K
54 Endress+Hauser