Technical Information Liquiline CM442R/CM444R/ CM448R

This document provides technical specifications for the Liquiline CM442R/CM444R/CM448R cabinet controller, which can monitor and control industrial processes with up to eight measuring channels based on digital Memosens technology. It has flexible inputs and outputs that allow connection of sensors, actuators and fieldbuses. The document describes the device architecture, communication protocols, power supply requirements, input and output specifications, installation guidelines, and performance characteristics.

Uploaded by

kythuat04 nte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
116 views56 pages

Technical Information Liquiline CM442R/CM444R/ CM448R

This document provides technical specifications for the Liquiline CM442R/CM444R/CM448R cabinet controller, which can monitor and control industrial processes with up to eight measuring channels based on digital Memosens technology. It has flexible inputs and outputs that allow connection of sensors, actuators and fieldbuses. The document describes the device architecture, communication protocols, power supply requirements, input and output specifications, installation guidelines, and performance characteristics.

Uploaded by

kythuat04 nte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 56

TI01112C/07/EN/09.

19 Products Solutions Services


71478328
2019-11-30

Technical Information
Liquiline CM442R/CM444R/
CM448R
Cabinet controller with a maximum of eight measuring
channels based on digital Memosens technology

For monitoring and controlling processes in industry and the environmental


sector
Application • Optional digital or analog inputs/outputs
• Maximum process safety thanks to standardized operating
• Food and beverages
concept across all devices in the Liquiline, sampler and
• Life science
analyzer platform
• Water and wastewater
• Fast commissioning thanks to:
• Chemical industry
• Memosens: lab-calibrated sensors & hot plug-and-play
• Preconfigured Liquiline transmitters
Your benefits • Easy extension and adaptation
• Highly flexible: • Minimum inventory:
• Able to connect up to 8 Memosens sensors • Cross-platform, modular concept (e.g. identical modules
• Mathematics functions calculate new measured values irrespective of parameters)
• Digital fieldbuses (HART, PROFIBUS, Modbus, • Integration into FieldCare and W@M facilitates effective
Ethernet/IP, PROFINET) and integrated web server asset management
• Choice of cleaning function, controller and alarm relay
Liquiline CM442R/CM444R/CM448R

Table of contents

Function and system design . . . . . . . . . . . . . . . . . . . 4 External power supply . . . . . . . . . . . . . . . . . . . . . . . . . 22


Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 PFM function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Application example . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Auxiliary voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Test voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Device architecture . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Slot and port assignment . . . . . . . . . . . . . . . . . . . . . . . . 6
Order of the modules . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Current outputs, active . . . . . . . . . . . . . . . . . . . . . . 22
Basic rule for hardware upgrades . . . . . . . . . . . . . . . . . . . 6 Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Determining the hardware delivery status . . . . . . . . . . . . . 7 Signal characteristic . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Terminal diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Electrical specification . . . . . . . . . . . . . . . . . . . . . . . . . 23
Device configuration using the example of a CM442R- Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
**M1A1F0* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Function diagram CM442R . . . . . . . . . . . . . . . . . . . . . . . 9 Relay outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Device configuration using the example of a CM444R- Electrical specification . . . . . . . . . . . . . . . . . . . . . . . . . 24
**M42A1FA* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Function diagram CM444R . . . . . . . . . . . . . . . . . . . . . . 11
Device configuration using the example of a CM448R-
**26A1* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . 24
Function diagram CM448R . . . . . . . . . . . . . . . . . . . . . . 13 HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
PROFIBUS DP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Modbus RS485 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Communication and data processing . . . . . . . . . . . 14 Modbus TCP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Ethernet/IP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Dependability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 PROFINET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Reliability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Web server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Maintainability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Measured variables . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Measuring ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . 28
Types of input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Cable specification for optional display cable . . . . . . . . . . 28
Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Connecting optional modules . . . . . . . . . . . . . . . . . . . . . 31
Protective ground connection . . . . . . . . . . . . . . . . . . . . 33
Sensor connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Digital inputs, passive . . . . . . . . . . . . . . . . . . . . . . . 20
Electrical specification . . . . . . . . . . . . . . . . . . . . . . . . . 20
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Performance characteristics . . . . . . . . . . . . . . . . . . 35
Nominal input current . . . . . . . . . . . . . . . . . . . . . . . . . 20 Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
PFM function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Reference temperature . . . . . . . . . . . . . . . . . . . . . . . . . 35
Test voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Measured error for sensor inputs . . . . . . . . . . . . . . . . . . 35
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Measured error for current inputs and outputs . . . . . . . . . 35
Frequency tolerance of digital inputs and outputs . . . . . . . 35
Resolution of current inputs and outputs . . . . . . . . . . . . . 35
Current input, passive . . . . . . . . . . . . . . . . . . . . . . . 20
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Signal characteristic . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Internal resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Test voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Mounting on DIN rail as per IEC 60715 . . . . . . . . . . . . . . 35
Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Mounting the external display . . . . . . . . . . . . . . . . . . . . 37
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Ambient temperature range . . . . . . . . . . . . . . . . . . . . . 37
Linearization/transmission behavior . . . . . . . . . . . . . . . . 22 Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Digital outputs, passive . . . . . . . . . . . . . . . . . . . . . 22
Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Electrical specification . . . . . . . . . . . . . . . . . . . . . . . . . 22
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

2 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . 38
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Degree of contamination . . . . . . . . . . . . . . . . . . . . . . . 39

Mechanical construction . . . . . . . . . . . . . . . . . . . . 39
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
External display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Local operation via external, optional display . . . . . . . . . . 43
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Language packages . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Certificates and approvals . . . . . . . . . . . . . . . . . . . 46


 mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
EAC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
cCSAus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Marine approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Ordering information . . . . . . . . . . . . . . . . . . . . . . . 46
Product page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Product Configurator . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Measuring cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Additional functionality . . . . . . . . . . . . . . . . . . . . . . . . 51
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Other accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

Endress+Hauser 3
Liquiline CM442R/CM444R/CM448R

Function and system design


Measuring system The overview shows examples of measuring systems. Other sensors and assemblies can be ordered
for conditions specific to your application (www.endress.com/products).

Measuring point pH value or ORP


A complete measuring system comprises: pH measurement in drinking
• TransmitterLiquiline water (→ Fig.)
• Optional display • Retractable assembly Cleanfit
• Sensors with Memosens technology CPA871
• Assemblies to suit the sensors used • Sensor Orbisint CPS11D
• Measuring cable CYK10
ORP in drinking water
• Dipfit CYA112 immersion
assembly
• Sensor Orbisint CPS12D
• Measuring cable CYK10
1 2
Conductivity
Inductive conductivity
measurement in wastewater
treatment
• Sensor Indumax CLS50D
• Sensor fixed cable
Conductive conductivity
measurement in power plant
cooling water
• Sensor Condumax CLS15D
5 4 3 • Measuring cable CYK10
Oxygen
Oxygen in aeration basins
A0042875
• Dipfit CYA112 immersion
1 Cabinet installation (without sensor and signal assembly
cables) • Holder CYH112
1 Optional display (back) • Sensor
2 Liquiline • COS61D (optical) with fixed
3 External power unit (CM444R and CM448R only) cable (→ Fig.)
4 Power cable (to be provided by the customer) • COS51D (amperometric),
5 Display cable cable CYK10

Nitrate and SAC Turbidity and interface


Nitrate in wastewater Turbidity in industrial water
• Sensor CAS51D-**A2 with fixed • Sensor Turbimax CUS51D with
cable fixed cable (→ Fig.)
• Dipfit CYA112 immersion • Assembly Flowfit CUA250
assembly • Spray head CUR3 (optional)
• Holder CYH112 Interface in the primary clarifier
SAC in the wastewater treatment • Sensor Turbimax CUS71D
outlet • Assembly CYA112
• Sensor CAS51D-**2C2 with • Holder CYH112
fixed cable
• Dipfit CYA112 immersion
assembly
• Holder CYH112
Disinfection Ion-selective electrodes
Free available chlorine (and pH) in Ammonium and nitrate
drinking water measurement in the aeration basin
• Sensor CCS142D • Sensor CAS40D with fixed cable
• Sensor CPS11D • Holder CYH112
• Measuring cable CYK10
• Flow assembly CCA250

4 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Application example • Transmitter CM444R-AAM44A0FM6 with:


4 x Memosens, Modbus TCP, 2 digital inputs and 2 digital outputs, 2 x relays for cleaning/limit
value, 2 x analog current input
• pH and temperature with CPS11D, item 1 (www.endress.com/cps11d)
• Chlorine with CCS142D, item 4 (www.endress.com/ccs142d)
• 2 x conductivity, inductive measurement with CLS50D, item 2 and 3 (www.endress.com/cls50d)
• 1 x measuring range switching for conductivity via Modbus module
• Flow assembly CCA250 with optional proximity switch INS (www.endress.com/cca250)
• Chlorine regulation with dosing interrupted if no flow: proximity switch via digital input of DIO
module, flow feedforward control (via digital or analog input), PFM-controlled dosing pump via
digital output of DIO module

1 4

Outlet 1
11:12:13 29.06.2013
CH 1:1 pH 6.99 pH
CH 1:1 Temperature 25.1 °C
CH 1:2 Cond I 9.02 mS/cm 3
2 CH 2:1 Cond I 13.41 mS/cm
CH 2:2 Chlorine 9.03 mg/l
CH 3:1 Current Input [Flow] 15.1 nA
MENU Field Data Manager

10.00
mS/cm

8.00
mS/cm

29.05.2013 29.06.2013
10:02:20 11:09:15

12.00
pH

6.00
pH
29.05.2013 29.06.2013
10:02:20 11:09:15

A0025347

2 Measuring point in the CIP process

Data retention
• Storage of all measured values, incl. values of external sources, in the non-volatile memory
(data logbook)
• Data called up on site via user-defined measuring menu and load curve display of the data logbook
• Transmission of data by ethernet, CDI interface or SD card and storage in a tamper-proof database
(Field Data Manager)
• Data export to csv file (for Microsoft Excel)

Endress+Hauser 5
Liquiline CM442R/CM444R/CM448R

Device architecture
Slot and port assignment
Slots
BASE2-E 2DS 2DS 2DS 2AI 2AO 4R
Ports

1
1

1
1

1
2

2
2

2
2
Sensor 1

Sensor 1

Sensor 1

3
2
1

Sensor 2

Sensor 2

Sensor 2
1

4
2

3 Slot and port assignment of the hardware modules


• Inputs are assigned to measuring channels in the
ascending order of the slots and ports.
Adjacent example:
"CH1: 1:1 pH glass" means:
Channel 1 (CH1) is slot 1 (basic module) : Port 1 (input
1), pH glass sensor
• Outputs and relays are named according to their
function, e.g. "current output", and are displayed in
ascending order with the slot and port numbers

4 Slot and port assignment on the


display

Order of the modules Depending on the version ordered, the device is supplied with a number of electronic modules, which
are assigned in a specific sequence in ascending order to slots 0 to 7. If you do not have a particular
module, the next moves up automatically:
• The basic module (which is always present) always occupies slots 0 and 1
• Fieldbus module 485
• Memosens input module 2DS (DS = digital sensor)
• Extension module for digital inputs and outputs DIO (DIO = digital input and output)
• Current input module 2AI (AI = analog input)
• Current output module 4AO or 2AO (AO = analog output)
• Relay modules AOR, 4R or 2R (AOR = analog output + relay, R = relay)
Modules with 4 ports are connected before modules of the same type with 2 ports.

Basic rule for hardware Please note the following if upgrading the device:
upgrades • The sum of all current inputs and outputs may not exceed 8!
• A maximum of two "DIO" modules may be used.

6 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Determining the hardware You must be aware of the type of modules and the number of them supplied with the device you
delivery status have ordered to determine the delivery status of your Liquiline.
• Basic module
One basic module in all versions. Always occupies slots 0 and 1.
• Fieldbus module
Optional, and only one fieldbus module is possible.
• Input modules
• Must be clearly assigned to the number of optional inputs ordered.
• Examples:
2 current inputs = module 2AI
4 Memosens inputs = 2 inputs with basic module + module 2DS with 2 further inputs
• Current outputs and relays
Various module combinations can exist.
The following table will help you determine which modules your device has, depending on the type
and number of outputs.

Relays

Current outputs 0 2 4

2 - 1 x 2R 1 x 4R

4 1 x 2AO 1 x AOR 1 x 2AO + 1 x 4R

6 1 x 4AO 1 x 4AO + 1 x 2R 1 x 4AO + 1 x 4R

8 1 x 4AO + 1 x 2AO 1 x 4AO + 1 x 2AO + 1 x 2R 1 x 4AO + 1x 2AO + 1 x 4R

‣ Sum up the number of modules and sort them according to the specified sequence →  6.
 This will give you the slot assignment for your device.

Terminal diagram The unique terminal name is derived from:


Slot no. : Port no. : Terminal
Example, NO contact of a relay
Device with 4 inputs for digital sensors, 4 current outputs and 4 relays
• Base module BASE2-E (contains 2 sensor inputs, 2 current outputs)
• 2DS module (2 sensor inputs)
• 2AO module (2 current outputs)
• 4R module (4 relays)

Slot 1 2 3 4
BASE2-E 2DS 2AO 4R
Relay 1
41

Slot 4 Port 1 Pin 41


43 42

Relay 2
41
43 42
41

Relay 3
43 42

Relay 4
41
43 42

A0039621

5 Creating a terminal diagram using the example of the NO contact (terminal 41) of a relay

Endress+Hauser 7
Liquiline CM442R/CM444R/CM448R

Device configuration using


the example of a CM442R-
**M1A1F0*

A0041706

Ordered basic device (example) • Order code CM442R-**M1A1F0*


• Functionality: 1 x Memosens, 2 current outputs without HART

Extension options without additional Activation with activation code:


modules • Second Memosens input (71114663)
• HART (71128428)
• EtherNet/IP + web server (71449914)
• Modbus TCP + web server (71449915)
• PROFINET + web server (71449901)
• Web server (71449918)

Extension options by using an PROFIBUS DP or Modbus RS485 with module 485 incl. activation
extension module in free slot 2 code:
• PROFIBUS DP (71140888)
• Modbus RS485 (71140889)

 Ifdisabled!
module 485 is retrofitted, any existing current outputs are

Additional inputs or outputs, relays:


• Module 2AI (71135639): 2 current inputs
• Module 2AO (71135632): 2 current outputs
• Module AOR (71111053): 2 current outputs, 2 relays
• Module 2R (71125375) or 4R (71125376): 2 or 4 relays
• Module DIO (71135638): 2 digital inputs and 2 digital outputs

Basic rule for extensions The sum of all current inputs and outputs may not exceed 8.

Restrictions if using CUS71D sensors Only one CUS71D can be connected. The second Memosens input
for interface measurement may not be used.

Product Configurator www.endress.com/cm442r

8 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Function diagram CM442R ~ =


31 L L+
HART optional
mA N L-
1 32
PE
6

31
mA
2 32 7
87 85
88
86
3 97 8
98

87 85
88
86
3 97 8
98
optional
41

42
4 9
43
optional
41
31
mA
42
2 32 10
43

31 41
mA
42
2 32 10
43

optional optional
41
23 mA
42
5 24 10
43

41
23 mA
42
5 24 10
43
optional optional

45 47
+
46 48

45 47
+
46 48
– 11
91
+
– 92

91
+
92
optional –

A0039427

6 Function diagram CM442R


1 Current output 1:1, + HART (optional) 6 Power supply
2 Current outputs (2 x optional) 7 Service interface
3 2 x Memosens input (1 x optional) 8 Power supply, fixed cable sensors
4 PROFIBUS DP/Modbus/Ethernet (optional) 9 Alarm relay
5 2 x current input (optional) 10 2 or 4 x relays (optional)
11 2 digital inputs and outputs (optional)

Endress+Hauser 9
Liquiline CM442R/CM444R/CM448R

Device configuration using


the example of a CM444R-
**M42A1FA*

Ordered basic device (example) • Order code CM444R-**M42A1FA*


• Functionality:
• 4 x Memosens (2 on BASE2-E module + 2 on an extension
module 2DS)
• PROFIBUS communication (module 485)
• Web server (BASE2-E module)
• 2 current outputs without HART (on BASE2-E module)
• 2 current inputs (module 2AI)
3 slots are still free in this example. More or fewer slots can be free in
other versions.

Extension options without additional None


modules

Modification options without Communication type changed by entering activation code. This
additional modules disables the communication type used previously!
Modbus RS485 + web server (71135636)
Retrofit by removing module 485 and entering the activation code
for:
• Modbus TCP + web server (71449915)
• EtherNet/IP + web server (71449914)
• PROFINET + web server (71449901)
• HART (71128428)
• Web server (71449918)

Extension options by using extension Only the following is possible for the example above:
modules in free slots 5-7 • Module 2R (71125375) or 4R (71125376): 2 or 4 relays
• Module DIO (71135638): 2 digital inputs and 2 digital outputs
If extending to 8 measuring channels:
Module 2DS (71135631): 2 Memosens inputs
Additional inputs or outputs and relays if fieldbus module 485 is
removed:
• Module 2AO (71135632): 2 current outputs
• Module AOR (71111053): 2 current outputs, 2 relays
• Module 2R (71125375) or 4R (71125376): 2 or 4 relays
• Module DIO (71135638): 2 digital inputs and 2 digital outputs

 Ifused
module 485 is removed and an Ethernet-based fieldbus is
via BASE2-E module, a maximum of up to 6 current
outputs can be operated in addition. Only two current outputs
are possible with module 485.

Basic rule for extensions The sum of all current inputs and outputs may not exceed 8.

Restrictions if using CUS71D sensors • In the case of CM444R, every combination of Memosens sensors
for interface measurement (max. 4) is possible.
• An extension to CM448R is not advisable as the maximum number
of Memosens inputs when using CUS71D remains limited to 4.

Product Configurator www.endress.com/cm444r

10 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Function diagram CM444R ~ =


31 HART optional L L+
mA N L-
1 32 6
PE

31
mA
2 32 7
optional
87 85
88
3 97 86 8
98
87 85
88
3 97 86 8
98

87 85
88
3 97 86 8
98
87 85
88
3 97 86 8
98
optional
41
42
4 9
43
optional

31 41
mA
42
2 32 10
43

31 41
mA
42
2 32
43 10
optional optional

31 41
mA
42
2 32 10
43

31 41
mA
42
2 32 10
43
optional optional

31
mA 23 mA
2 32 24 5

31
mA 23 mA
2 32 24 5
optional optional

45 47 +
+
46 48
– –
45
+ 47
46 +
– 48
91 – 11
+
– 92
91
+
– 92 optional

A0039431

7 Function diagram CM444R


1 Current output 1:1, + HART (both optional) 6 Power supply
2 Max. 7 x current output (optional) 7 Service interface
3 Memosens input (2 x standard + 2 x optional) 8 Power supply, fixed cable sensors
4 PROFIBUS DP/Modbus/Ethernet (optional) 9 Alarm relay
5 2 x current input (optional) 10 2 or 4 x relays (optional)
11 2 digital inputs and outputs (optional)

Endress+Hauser 11
Liquiline CM442R/CM444R/CM448R

Device configuration using


the example of a CM448R-
**26A1*

Ordered basic device (example) • Order code CM448R-**26A1*


• Functionality:
• 6 x Memosens (2 on BASE2-E module + 2 on two 2DS extension
modules)
• PROFIBUS communication (module 485)
• Web server (BASE2-E module)
3 slots are still free in this example. More or fewer slots can be free in
other versions.

Extension options without additional Activation code for the use of current outputs of the basic module:
modules 2 current outputs (71140891)

Modification options without Communication type changed by entering activation code. This
additional modules disables the communication type used previously!
Modbus RS485 + web server (71135636)
Retrofit by removing module 485 and entering the activation code for
communication via the BASE2 module:
• Modbus TCP + web server (71449915)
• EtherNet/IP + web server (71449914)
• PROFINET + web server (71449901)
• HART (71128428)
• Web server (71449918)

Extension options by using extension If extending to 8 measuring channels:


modules in free slots 5-7 Module 2DS (71135631): 2 Memosens inputs
Additional inputs or outputs, relays:
• Module 2AO (71135632): 2 current outputs
• Module 2AI (71135639): 2 current inputs
• Module AOR (71111053): 2 current outputs, 2 relays
• Module 2R (71125375) or 4R (71125376): 2 or 4 relays
• Module DIO (71135638): 2 digital inputs and 2 digital outputs

 Ifused,
module 485 is removed and an Ethernet-based fieldbus is
a maximum of up to 6 current outputs can be operated in
addition. Only two current outputs are possible with module
485.

Basic rule for extensions The sum of all current inputs and outputs may not exceed 8.

Restrictions if using CUS71D sensors The maximum number of Memosens inputs that can be used is
for interface measurement limited to 4! Here, every combination of CUS71D and other
Memosens sensors is then possible.

Product Configurator www.endress.com/cm448r

12 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Function diagram CM448R 31 ~ =


HART optional L L+
mA
1 32
N L- 6
PE
31
mA
2 32 7
optional
87 85
88
3 97 86 8
98
87 85
88
3 97 86 8
98
87 85
88 8
3 97 86
98
87 85
88 8
3 97 86
98
87 85
88
3 97 86 8
98
87 85
88
3 97 86 8
98
87 85
88 8
3 97 86
98
87 85
88 8
3 97 86
98
optional
31 41
mA
42
2 32 43
9
31 41
mA
42
2 32
43
10
optional
41
31
mA
42
2 32 43 10
optional
31 41
mA
42
2 32 10
43
optional
41
31
mA 42
2 32 43
10
optional
31
mA
2 32 4
optional optional
23 mA
+
45
24 5
– 46
45 23 mA
11 +
– 46 24 5
optional
91 47
+ +
– 92 48

91 47 +
+
92 optional 48
– –

A0039429

8 Function diagram CM448R


1 Current output 1:1, + HART (both optional) 7 Service interface
2 Max. 7 x current output (optional) 8 Power supply, fixed cable sensors
3 Max 8 x Memosens input (2 x of which are 9 Alarm relay
optional)
4 PROFIBUS DP/Modbus/Ethernet (optional) 10 2 or 4 x relays (optional)
5 2 x current input (optional) 11 2 digital inputs and outputs (optional)
6 Power supply

Endress+Hauser 13
Liquiline CM442R/CM444R/CM448R

Communication and data processing


Communication protocols:
Fieldbus systems
• HART
• PROFIBUS DP (Profile 3.02)
• Modbus TCP or RS485
• PROFINET
• Ethernet/IP
Only one type of Fieldbus communication can ever be active. The last activation code entered
decides which bus is used.
The device drivers available make it possible to perform a basic setup and display measured
values and diagnostics information via the fieldbus. A full device configuration via the fieldbus
is not possible.
Extension module 485 and current outputs
For PROFIBUS DP, and Modbus RS485 communication protocols:
• CM442R
Current outputs cannot be used in parallel. Any existing current outputs are deactivated with the
installation of 485.
• CM444R/CM448R
A maximum of 2 current outputs can be used in parallel.
Ethernet functionality via Base2 module and current outputs
• CM442R
A maximum of 2 current outputs can be used in parallel.
• CM444R and CM448R
A maximum of 6 current outputs can be used in parallel.
Bus termination on the device
• Via slide switch at bus module 485
• Displayed via LED "T" on bus module 485

Dependability
Reliability Memosens MEMO SENS

Memosens makes your measuring point safer and more reliable:


• Non-contact, digital signal transmission enables optimum galvanic isolation
• No contact corrosion
• Completely watertight
• Sensor can be calibrated in a lab, thus increasing the availability of the measuring point in the
process
• Intrinsically safe electronics mean operation in hazardous areas is not a problem.
• Predictive maintenance thanks to recording of sensor data, e.g.:
• Total hours of operation
• Hours of operation with very high or very low measured values
• Hours of operation at high temperatures
• Number of steam sterilizations
• Sensor condition

Heartbeat diagnostics
• Heartbeat diagnostics screen with graphic indicators for the health of the device and sensor and
with a maintenance or (sensor-dependent) calibration timer
• Heartbeat status information on the health of the device and the condition of the sensor
• : Sensor/device condition or maintenance timer > 20 %; no action is required
• : Sensor/device condition or maintenance timer > 5 ≤ 20 %, maintenance not yet urgent but
should be scheduled
• : Sensor/device condition or maintenance timer < 5 %, maintenance is recommended
• The Heartbeat sensor condition is the assessment of the calibration results and the sensor
diagnostic functions.

14 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

An unhappy smiley can be due to the calibration result, the measured value status or to the
operating hours limit having been exceeded. These limits can be configured in the sensor setup in a
way that adapts the Heartbeat diagnostics to the application.
Heartbeat and NAMUR category
The Heartbeat status indicates the sensor or device condition while the NAMUR categories (F, C, M,
S) assess the reliability of the measured value. The two conditions can correlate but do not have to.
• Example 1
• The number of remaining cleaning cycles for the sensor reaches 20% of the defined maximum
number. The Heartbeat symbol changes from to . The measured value is still reliable so the
NAMUR status signal does not change.
• If the maximum number of cleaning cycles is exceeded, the Heartbeat symbol changes from
to . While the measured value can still be reliable, the NAMUR status signal changes to M
(maintenance required).
• Example 2
The sensor breaks. The Heartbeat status changes immediately from to and the NAMUR
status signal also changes immediately to F (failure).

Heartbeat Monitoring
Sensor data from Memosens sensors are transmitted via the EtherNet/IP, PROFINET, PROFIBUS DP,
HART, Modbus RTU and Modbus TCP fieldbus protocols. These data can be used for predictive
maintenance, for instance.
Examples include:
• Total hours of operation
• Hours of operation with very high or very low measured values
• Hours of operation at high temperatures
• Number of steam sterilizations
• Sensor identification
• Calibration information
For detailed information on "Ethernet/IP communication", see the product pages on the Internet
(→ SD01293C).
For detailed information on "Modbus communication", see the product pages on the Internet (→
SD01189C).
For detailed information on "PROFINET communication", see the product pages on the internet
(→ SD02490C).
For detailed information on "PROFIBUS communication", see the product pages on the Internet
(→ SD01188C).
More detailed information on HART communication is provided on the product pages on the
Internet (→ SD01187C).

Heartbeat Verification
Heartbeat Verification makes it possible to verify the correct operation of the measuring device
without interrupting the process. This verification can be documented anytime.

Sensor Check System (SCS)


The Sensor Check System (SCS) monitors the high impedance of the pH glass. An alarm is issued if a
minimum impedance value is undershot or a maximum impedance is exceeded.
• Glass breakage is the main reason for a drop in high impedance values
• The reasons for increasing impedance values include:
• Dry sensor
• Worn pH glass membrane
For the SCS, upper and lower limit values can be enabled or disabled independently of one
another.

Process Check System (PCS)


The process check system (PCS) checks the measuring signal for stagnation. An alarm is triggered if
the measuring signal does not change over a specific period (several measured values).
The main causes of stagnating measured values are:
• Contaminated sensor, or sensor outside of medium
• Sensor defective
• Process error (e.g. through control system)

Endress+Hauser 15
Liquiline CM442R/CM444R/CM448R

Self-monitoring functions
Current inputs are deactivated in the event of overcurrent and reactivated once the overcurrent
stops. Board voltages are monitored and the board temperature is also measured.

USP and EP
The limit functions for pharmaceutical water in accordance with USP and EP specifications are
implemented in the software for conductivity measurements:
• "Water for Injection" (WFI) as per USP <645> and EP
• "Highly Purified Water" (HPW) as per EP
• "Purified Water" (PW) as per EP
The uncompensated conductivity value and the temperature are measured for the USP/EP limit
functions. The measured values are compared against the tables defined in the standards. An alarm
is triggered if the limit value is exceeded. Furthermore, it is also possible to configure an early
warning alarm that signals undesired operating states before they occur.

ChemocleanPlus
Freely programmable sequence control
• e.g. for automatic sensor cleaning in retractable assemblies for reliable measurement results in
processes with a high risk of contamination
• Individual, time-based activation of 4 outputs e.g. relays
• Starting, stopping or pausing of activities via digital input or fieldbus signals e.g. from limit
position switches

Maintainability Modular design


The modular transmitter design means it can be easily adapted to suit your needs:
• Retrofit extension modules for new or extended range of functions, e.g. current outputs, relays and
digital communication
• Upgrade to maximum eight-channel measurement
• Optional: M12 sensor connector for connecting any kind of Memosens sensor
• Optional: CDI connector for external access to the service interface (avoids having to unscrew the
housing cover)

A0042325 A0042346

9 CM444R: example  10 Extension module

16 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Data logger function

• Adjustable scan time: 1 to 3600 s (1 h)


• Data logbooks:
• Max. 8 data logbooks
• 150,000 entries per logbook
• Graphic display (load curves) or numerical list
• Calibration logbook: max. 75 entries
• Hardware version logbook:
• Hardware configuration and modifications
• Max. 125 entries
• Version logbook:
• E.g. software updates
• Max. 50 entries
• Operation logbook: max. 250 entries
• Diagnostics logbook: max. 250 entries

A0015032

 11 Data logbook: Graphic display


Logbooks remain unchanged even after a software update.

SD card
The exchangeable storage medium enables:
• Quick and easy software updates and upgrades
• Data storage of internal device memory (e.g. logbooks)
• Transfer of complete configurations to a device with an identical setup (backup function)
• Transfer of configurations without the TAG and bus address to devices with an identical setup
(copy function)
• Saving of screenshots for documentation purposes
Endress+Hauser offers industry-approved SD cards as accessories. These memory cards provide
maximum data security and integrity.
Other SD cards up to a maximum weight of 5 g can also be used. However, Endress+Hauser does not
accept any responsibility for the data security of such cards.

External signals for device control and for activating external devices
Hardware options, e.g. module "DIO" with 2 digital inputs and 2 digital outputs or fieldbus module
"485" enable the following:
• via a digital input signal
• measuring range switching for conductivity (upgrade code required, see accessories)
• switching between different calibration datasets in the case of optical sensors
• an external hold
• a cleaning interval to be triggered
• switching on and off a PID controller, e.g. via the proximity switch of the CCA250
• the use of the input as an "analog input" for pulse-frequency modulation (PFM)
• via a digital output signal
• the static transmission (similar to a relay) of diagnostic states, point level switch states etc.
• the dynamic transmission (comparable to a non-wearing "analog output") of PFM signals, e.g. to
control dosing pumps.

Endress+Hauser 17
Liquiline CM442R/CM444R/CM448R

FieldCare and Field Data Manager


FieldCare
Configuration and asset management software based on FDT/DTM technology
• Complete device configuration when connected via FXA291 and service interface
• Access to a number of configuration parameters and identification, measuring and diagnostic data
when connected via HART modem
• Logbooks can be downloaded in CSV format or binary format for "Field Data Manager" software
Field Data Manager
Visualization software and database for measuring, calibration and configuration data
• SQL database which is protected against manipulation
• Functions to import, save and print out logbooks
• Load curves to display measured values

A0016009

 12 Field Data Manager: Load curves

Virtual process values (mathematical functions)


In addition to "real" process values, which are provided by connected physical sensors or analog
inputs, mathematical functions can be used to calculate a maximum of 8 "virtual" process values.
The "virtual" process values can be:
• Output via a current output or a fieldbus
• Used as a controlled variable
• Assigned as a measured variable to a limit switch
• Used as a measured variable to trigger cleaning
• Displayed in user-defined measuring menus
The following mathematical functions are possible:
• Calculation of pH from two conductivity values according to VGB 405 RL, e. g. in boiler feedwater
• Difference between two measured values from different sources, e. g. for membrane monitoring
• Differential conductivity, e. g. for monitoring the efficiency of ion exchangers
• Degassed conductivity, e. g. for process controls in power plants
• Redundancy for monitoring two or three redundant sensors
• rH calculation based on the measured values of a pH and an ORP sensor
• Calculation of the remaining capacity of a cation exchanger
• Formula editor

18 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Concentration tables
When the device is delivered from the factory, tables are saved in the device that allow inductive
conductivity measurements to be converted to concentrations of certain substances. 4 user-defined
tables are also possible.

The following factory concentration tables are available:


NaOH 0 to 15 % 0 to 100 ˚C (32 to 212 ˚F)
NaOH 25 to 50% 2 to 80 ˚C (36 to 176 ˚F)
HCl 0 to 20 % 0 to 65 ˚C (32 to 149 ˚F)
HNO3 0 to 30 % 2 to 80 ˚C (36 to 176 ˚F)
H2SO4 0.5 to 27 % and 35 to 85 % 0 to 100 ˚C (32 to 212 ˚F)
H2SO4 93 to 100 % 10 to 115 ˚C (50 to 239 ˚F)
H3PO4 0 to 40 % 2 to 80 ˚C (36 to 176 ˚F)
NaCl 0 to 26 % 2 to 80 ˚C (36 to 176 ˚F)

Safety Real-time clock


The device has a real-time clock, which is buffered by a button cell battery if the power supply fails.
This ensures that the device continues to keep the correct date and time when it is restarted and that
the time stamp for the logbooks is correct.

Data security
All settings, logbooks etc. are stored in a non-volatile memory to ensure that the data are retained
even in the event of a disruption to the power supply.

Measuring range switching for conductivity


• Can be used in CIP processes e.g. for safe monitoring of phase separations
• Switching between 4 complete parameter sets:
• Conductivity operating mode
• Concentration tables
• Temperature compensation
• Output signal range
• Limit value switch
• Via digital inputs or fieldbus

Measured value compensation for oxygen and conductivity


• Pressure or temperature compensation
• Input signals from external sensors via current input or fieldbus
• Signals from connected temperature sensors

Password protection
Password-protected login
• For remote operation via web server
• For local operation

Process safety
Two independent PID controllers
• One- or two-sided control
• Limit switches
• 4 cleaning programs which can be programmed independently of each other

IT security
Our warranty is valid only if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the settings.
IT security measures, which provide additional protection for the device and associated data transfer,
must be implemented by the operators themselves in line with their security standards.

Endress+Hauser 19
Liquiline CM442R/CM444R/CM448R

Input
Measured variables → Documentation of the connected sensor

Measuring ranges → Documentation of the connected sensor

Types of input • Digital sensor inputs for sensors with Memosens protocol
• Analog current inputs (optional)
• Digital inputs (optional)

Input signal Depending on version:


• Max. 8 x binary sensor signal
• 2 x 0/4 to 20 mA (optional), passive, potentially isolated from one another and from the sensor
inputs
• 0 to 30 V

Cable specification Cable type


Memosens data cable CYK10 or sensor fixed cable, each with cable end sleeves or M12 round-pin
connector (optional)

Cable length
Max. 100 m (330 ft)

Digital inputs, passive


Electrical specification • drawing power (passive)
• Galvanically isolated

Span • High: 11 to 30 V DC
• Low: 0 to 5 V DC

Nominal input current max. 8 mA

PFM function Minimum pulse width: 500 µs (1 kHz)

Test voltage 500 V

Cable specification Max. 2.5 mm2 (14 AWG)

Current input, passive


Span > 0 to 20 mA

Signal characteristic Linear

Internal resistance Non-linear

Test voltage 500 V

20 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Output
Output signal Depending on version:
• 2 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
• 4 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
• 6 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
• 8 x 0/4 to 20 mA, active, galvanically isolated from one another and from the sensor circuits
• Optional HART communication (only via current output 1:1)

HART

Signal encoding FSK ± 0.5 mA via current signal

Data transmission rate 1200 baud

Galvanic isolation Yes

Load (communication resistor) 250 Ω

PROFIBUS DP/RS485

Signal encoding EIA/TIA-485, PROFIBUS DP-compliant acc. to IEC 61158

Data transmission rate 9.6 kBd, 19.2 kBd, 45.45kBd, 93.75 kBd, 187.5 kBd, 500 kBd,
1.5 MBd, 6 MBd, 12 MBd

Galvanic isolation Yes

Connectors Spring terminal (max. 1.5 mm), bridged internally (T‐function),


optional M12

Bus termination Internal slide switch with LED display

Modbus RS485

Signal encoding EIA/TIA-485

Data transmission rate 2,400, 4,800, 9,600, 19,200, 38,400, 57,600 and 115,200
baud

Galvanic isolation Yes

Connectors Spring terminal (max. 1.5 mm), bridged internally (T‐function),


optional M12

Bus termination Internal slide switch with LED display

Ethernet and Modbus TCP

Signal encoding IEEE 802.3 (Ethernet)

Data transmission rate 10/100 MBd

Galvanic isolation Yes

Connection RJ45

IP address DHCP (default) or configuration via menu

EtherNet/IP

Signal encoding IEEE 802.3 (Ethernet)

Data transmission rate 10/100 MBd

Galvanic isolation Yes

Connection RJ45

IP address DHCP (default) or configuration via menu

Endress+Hauser 21
Liquiline CM442R/CM444R/CM448R

PROFINET

Signal encoding IEEE 802.3 (Ethernet)

Data transmission rate 100 MBd

Galvanic isolation Yes

Connection RJ45

Name of station Via DCP protocol using the configuration tool (e.g. Siemens
PRONETA)

IP address Via DCP protocol using the configuration tool (e.g. Siemens
PRONETA)

Signal on alarm Adjustable, as per NAMUR Recommendation NE 43


• In measuring range 0 to 20 mA (HART is not available with this measuring range):
Failure current from 0 to 23 mA
• In measuring range 4 to 20 mA:
Failure current from 2.4 to 23 mA
• Factory setting for failure current for both measuring ranges:
21.5 mA

Load Max. 500 Ω

Linearization/transmission Linear
behavior

Digital outputs, passive


Electrical specification • Passive
• Open collector, max. 30 V, 15 mA
• Maximum voltage drop 3 V

External power supply When using an onsite auxiliary voltage supply and an onsite digital input:
Recommended minimum auxiliary voltage = 3 V + VIHmin
(VIHmin = minimum input voltage required (high-level input voltage)

PFM function Minimum pulse width: 500 µs (1 kHz)

Auxiliary voltage Electrical specification


• Galvanically isolated
• Unregulated, 24 V DC
• Max. 50 mA (per DIO module)

Test voltage 500 V

Cable specification Max. 2.5 mm2 (14 AWG)

Current outputs, active


Span 0 to 23 mA
2.4 to 23 mA for HART communication

Signal characteristic Linear

22 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Electrical specification Output voltage


Max. 24 V
Test voltage
500 V

Cable specification Cable type


Recommended: shielded cable
Cable specification
Max. 2.5 mm2 (14 AWG)

Endress+Hauser 23
Liquiline CM442R/CM444R/CM448R

Relay outputs
Electrical specification Relay types
• 1 single-pin changeover contact (alarm relay)
• 2 or 4 single-pin changeover contacts (optional with extension modules)
Maximum load
• Alarm relay: 0.5 A
• All other relays: 2.0 A
Relay switching capacity

Base module (Alarm relay)


Switching voltage Load (max.) Switching cycles (min.)

230 V AC, cosΦ = 0.8 to 1 0.1 A 700,000

0.5 A 450,000

115 V AC, cosΦ = 0.8 to 1 0.1 A 1,000,000

0.5 A 650,000

24 V DC, L/R = 0 to 1 ms 0.1 A 500,000

0.5 A 350,000

Extension modules
Switching voltage Load (max.) Switching cycles (min.)

230 V AC, cosΦ = 0.8 to 1 0.1 A 700,000

0.5 A 450,000

2A 120,000

115 V AC, cosΦ = 0.8 to 1 0.1 A 1,000,000

0.5 A 650,000

2A 170,000

24 V DC, L/R = 0 to 1 ms 0.1 A 500,000

0.5 A 350,000

2A 150,000

Cable specification Max. 2.5 mm2 (14 AWG)

Protocol-specific data
HART Manufacturer ID 11h

Device type 155Dh

Device revision 001h

HART version 7.2

Device description files (DD/DTM) www.endress.com/hart


Device Integration Manager DIM

Device variables 16 user-definable and 16 predefined device variables, dynamic


variables PV, SV, TV, QV

Supported features PDM DD, AMS DD, DTM, Field Xpert DD

24 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

PROFIBUS DP Manufacturer ID 11h

Device type 155Dh

Profile version 3.02

GSD files www.endress.com/profibus


Device Integration Manager DIM

Output values 16 AI blocks, 8 DI blocks

Input variables 4 AO blocks, 8 DO blocks

Supported features • 1 MSCY0 connection (cyclical communication, master class 1


to slave)
• 1 MSAC1 connection (acyclical communication, master class 1
to slave)
• 2 MSAC2 connections (acyclical communication, master class
2 to slave)
• Device lock: The device can be locked using the hardware or
software.
• Addressing using DIL switches or software
• GSD, PDM DD, DTM

Modbus RS485 Protocol RTU/ASCII

Function codes 03, 04, 06, 08, 16, 23

Broadcast support for function codes 06, 16, 23

Output data 16 measured values (value, unit, status), 8 digital values (value,
status)

Input data 4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information

Supported features Address can be configured using switch or software

Modbus TCP TCP port 502

TCP connections 3

Protocol TCP

Function codes 03, 04, 06, 08, 16, 23

Broadcast support for function codes 06, 16, 23

Output data 16 measured values (value, unit, status), 8 digital values (value,
status)

Input data 4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information

Supported features Address can be configured using DHCP or software

Ethernet/IP Log EtherNet/IP

ODVA certification Yes

Device profile Generic device (product type: 0x2B)

Manufacturer ID 0x049Eh

Device type ID 0x109Ch

Polarity Auto-MIDI-X

Connections CIP 12

I/O 6

Explicit message 6

Multicast 3 consumers

Minimum RPI 100 ms (default)

Endress+Hauser 25
Liquiline CM442R/CM444R/CM448R

Maximum RPI 10000 ms

System integration EtherNet/IP EDS

Rockwell Add-on-Profile Level 3, Faceplate for Factory


Talk SE

IO data Input (T → O) Device status and diagnostic message with


highest priority
Measured values:
• 16 AI (analog input) + Status + Unit
• 8 DI (discrete input) + Status

Output (O → T) Actuating values:


• 4 A0 (analog output) + status + unit
• 8 DO (discrete output) + Status

PROFINET Protocol "Application layer protocol for decentral device periphery and
distributed automation", PNIO Version 2.34

Communication type 100 MBit/s

Conformance Class Conformance Class B

Netload Class Netload Class II

Baud rate Automatic 100 Mbps with full-duplex detection

Cycle times From 32 ms

Device profile Application interface identifier 0xF600


Generic device

PROFINET interface 1 port, Realtime Class 1 (RT_CLASS_1)

Manufacturer ID 0x11h

Device type ID 0x859C Dh

Device description files (GSD) Information and files under:


• www.endress.com
On the product page for the device: Documents/Software →
Device drivers
• www.profibus.com
On the website under Products/Product Finder

Polarity Auto-polarity for automatic correction of crossed TxD and RxD


pairs

Supported connections • 1 x AR (IO Controller AR)


• 1 x AR (IO-Supervisor Device AR connection allowed)
• 1 x Input CR (Communication Relation)
• 1 x Output CR (Communication Relation)
• 1 x Alarm CR (Communication Relation)

Configuration options for measuring device • Web browser


• Manufacturer-specific software (FieldCare, DeviceCare)
• Device master file (GSD), can be read out via the integrated
web server of the measuring device

Configuration of the device name DCP protocol

26 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Supported functions • Identification & Maintenance


Simple device identification via:
• Process control system
• Nameplate
• Measured value status
The process variables are communicated with a measured
value status
• Blinking feature (FLASH_ONCE) via the local display for
simple device identification and assignment
• Device operation via operating tools (e.g. FieldCare,
DeviceCare)

System integration For information on system integration, see the Operating


Instructions
• Cyclic data transmission
• Overview and description of the modules
• Status coding
• Startup configuration
• Factory setting

Web server The Web server enables full access to the device configuration, measured values, diagnostic
messages, logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G routers with a user-
defined IP address.

TCP port 80

Supported features • Remote-controlled device configuration(1 session)


• Save/restore device configuration (via SD card)
• Logbook export (file formats: CSV, FDM)
• Access to Web server via DTM or Internet Explorer
• Login
• Web server can be switched off

Power supply
Supply voltage CM442 R
Depending on version:
• 100 to 230 V AC, 50/60 Hz
Maximum permitted fluctuation of mains supply voltage: ± 15 % of nominal voltage
• 24 V AC/DC, 50/60 Hz
Maximum permitted fluctuation of mains supply voltage: + 20/- 15 % of nominal voltage
CM444 R and CM448 R
Depending on the version, via external DIN rail power unit:
• 100 to 230 V AC, 50/60 Hz
Maximum permitted fluctuation of mains supply voltage: ± 15 % of nominal voltage 1)
• 24 V DC
Maximum permitted fluctuation of mains supply voltage: + 20/- 15 % of nominal voltage 1)

NOTICE
The device does not have a power switch!
‣ Provide a protected circuit breaker in the vicinity of the device at the place of installation.
‣ The circuit breaker must be a switch or power switch, and must be labeled as the circuit breaker
for the device.
‣ At the supply point, the power supply must be isolated from dangerous live cables by double or
reinforced insulation in the case of devices with a 24 V supply voltage.

1) *Specifications only apply if used with power unit supplied by manufacturer.

Endress+Hauser 27
Liquiline CM442R/CM444R/CM448R

Power consumption CM442 R


Depending on supply voltage
• 100 to 230 V AC and 24 V AC:
Max. 55 VA
• 24 V DC:
Max. 22 W
CM444 R and CM448 R
Depending on supply voltage
• 100 to 230 V AC:
Max. 150 VA 1)
• 24 V DC:
Max. 59 W 1)

Fuse Fuse not exchangeable

Overvoltage protection Integrated overvoltage/lightning protection as per EN 61326


Protection category 1 and 3

Cable specification for Length of display cable provided:


optional display cable 3 m (10 ft)
Maximum permitted length of a display cable:
5 m (16.5 ft)

28 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Electrical connection Basic module

1 2 3 1 2 3

9 4 10 4

8 5 9 5

7 6 8 7 6
A0040639 A0040640

 13 Basic module BASE2-H or -L (two-channel  14 Basic module BASE2-E (four- and eight-
device) channel device)

1 Power supply for digital fixed cable sensors with 1 Power supply for digital fixed cable sensors with
Memosens protocol Memosens protocol
2 SD card slot 2 SD card slot
3 Slot for display cable 1) 3 Slot for display cable 1)
4 Ethernet interface 4 Ethernet interface
5 Connections for 2 Memosens sensors 5 Connections for 2 Memosens sensors
6 Current outputs 6 Current outputs
7 Power connection 7 Socket for internal supply cable 2)
8 Service interface 8 LEDs
9 Alarm relay connection 9 Service interface
10 Alarm relay connection
1) For optional external display.
2) Power supply to DIN rail power unit.

Endress+Hauser 29
Liquiline CM442R/CM444R/CM448R

Connecting supply voltage for CM442R

Sensor supply
1 PK 85 +

Display
GY 86 –
2 PK 85 +

SD
GY 86 –

Ethernet
41

Service
Alarm
42
43
+ BN 87 87 BN +

Sensor 2
Sensor 1
– WH 88 88 WH –
A GN 97 97 GN A
B YE 98 98 YE B

0/4 ... 20 mA

HART
+
PE N/- L/+

L/+ L 31
1

Power
N/– N 32 –
PE * 31 +
2 32 –

A0039665 A0039625

 15 Connecting power supply on the BASE2-H or -  16 Overall wiring diagram for BASE2-H or -L
L
H Power unit 100 to 230 VAC
L Power unit 24 VAC or 24 VDC

Connecting supply voltage for CM444R and CM448R


Sensor supply

1 PK 85 + Display
GY 86 –
2 PK 85 +
SD

*
GY 86 –
Ethernet

41
Service
Alarm

42
43
+ BN 87 87 BN +
Sensor 2
Sensor 1

– WH 88 88 WH –
+ RD A
(internal)

A GN 97 97 GN
Power

- BK B YE 98 98 YE B
L N
0/4 ... 20 mA

HART

31 +
1
32 –
31 +
2 32 –
Power

L/+ N/– PE B

L/+ N/– *

A0039668 A0039624

 17 Connecting power supply with BASE2-E  18 Overall wiring diagram for BASE2-E and
external power unit (B)
* Assignment depending on power unit, make sure
to connect correctly
The two device versions may only be operated with the power unit supplied and the power unit
cable. Also pay attention to the information in the operating manual supplied for the power
unit.

30 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Connecting optional modules With extension modules you can purchase additional functions for your device.

NOTICE
Unacceptable hardware combinations (due to conflicts in power supply)
Incorrect measurements or total failure of the measuring point as a result of heat build-up or
overloading
‣ If you are planning to extend your controller, make sure the resulting hardware combination is
permitted (Configurator at www.endress.com/CM442R or .../CM444R or .../CM448R).
‣ Remember that the sum of all current inputs and outputs may not exceed 8.
‣ Make sure not to use more than 2 "DIO" modules. More "DIO" modules are not permitted.
‣ Please contact your Endress+Hauser sales center should you have any questions.
Overview of all the modules available
Module name

AOR 2R 4R 2DS DIO

• 2 x 0/4 to 20 mA • 2 relays • 4 relays • 2 digital sensor • 2 digital inputs


analog outputs • Order No. • Order No. inputs • 2 digital outputs
• 2 relays 71125375 71125376 • 2 power supply with auxiliary
• Order No. systems for digital voltage
71111053 sensors • Order No.
• Order No. 71135638
71135631

31 41 41 85 47
Sensor supply
0/4 ... 20 mA

+1 1+ PK + 1
Relay 1
Relay 1

32 – 42 42 48
– 86 GY –
43 43 85 47 +
31 +
2 2+ PK 2
41 41
32 – – 86 GY 48
Relay 2


Relay 2

42 42
43 43 + 87 BN 45 +
1
Sensor 1

41
– 88 WH 46 –
Relay

41
42 1 A 97 GN 45 +
Relay 3

42 98 YE 46 – 2
43 B
43
41 + 87 BN 91 +
Sensor 2

41 1
– 88 WH
Relay

Relay 4

92
42 2 –
42 A 97 GN 91 +
43 43 98 YE 2
B 92

Endress+Hauser 31
Liquiline CM442R/CM444R/CM448R

Module name

2AO 4AO 2AI 485

24 23
24 23
• 2 x 0/4 to 20 mA • 4 x 0/4 to 20 mA • 2 x 0/4 to 20mA • Ethernet (web
analog outputs analog outputs analog inputs server or Modbus
• Order No. • Order No. • Order No. TCP)
71135632 71135633 71135639 • 5V power supply
for PROFIBUS DP
termination
• RS485 (PROFIBUS
DP or Modbus
RS485)
• Use of BASE2
module disables
Ethernet port of
module 485
• Order No.
71135634

31 31 23
Ethernet

+1 +1 +
0/4 ... 20 mA
0/4 ... 20 mA

0/4 ... 20 mA

1
32 32 24
– – –
1
PWR

31+ 31 + 23+ 2
DIP switch

2 2 2 4
8
32 32 24
T COM SF BF

on

– – – 16
32
31 64
+3 128/SW
0/4 ... 20 mA

32
– Service

31 + Termi- DGND 82
4 nation VP 81
32
– 96 96'
DP/
95 RS485 95'
99 99'

PROFIBUS DP (module 485)


Contacts 95, 96 and 99 are bridged in the connector. This ensures that PROFIBUS
communication is not interrupted if the connector is disconnected.

32 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Protective ground connection

A0025366

 19 Mounting rail for functional ground connections

Sensor connection Sensors with Memosens protocol


Sensor types Sensor cable Sensors

Digital sensors without With plug-in • pH sensors


additional internal power connection and • ORP sensors
supply inductive signal • Combined sensors
transmission • Oxygen sensors (amperometric and optical)
• Conductivity sensors with conductive measurement
of conductivity
• Chlorine sensors (disinfection)

Fixed cable Conductivity sensors with inductive measurement of


conductivity

Digital sensors with Fixed cable • Turbidity sensors


additional internal power • Sensors for interface measurement
supply • Sensors for measuring the spectral absorption
coefficient (SAC)
• Nitrate sensors
• Optical oxygen sensors
• Ion-sensitive sensors

The following rule applies if connecting CUS71D sensors:


• CM442R
• Only one CUS71D is possible; an additional sensor is not permitted.
• The second sensor input may also not be used for another type of sensor.
• CM444R
No restrictions. All the sensor inputs can be used as required.
• CM448R
• If a CUS71D is connected, the number of sensor inputs that can be used is limited to a maximum
of 4.
• Of these, all 4 inputs can be used for CUS71D sensors.
• Every combination of CUS71D and other sensors is possible, provided that the total number of
connected sensors does not exceed 4.
Connection
Direct connection of the sensor cable to the terminal connector of the sensor module 2DS or of the
base module-L, -H or -E (→  20 ff.)

Endress+Hauser 33
Liquiline CM442R/CM444R/CM448R

PK

86 85 86 85

86 85 86 85
GY

BN

98 97 88 87

98 97 88 87
WH
GN GN
YE YE

Sensor Sensor

A0039629 A0039622

 20 sensors without additional supply voltage  21 sensors with additional supply voltage

2DS
PK
85
1

GY
86
85
2

86

Sensor 1 Sensor
98 97 88 87

GN
YE

Sensor 2
BN
98 97 88 87

WH
GN Sensor
YE

A0033206

 22 sensors with and without additional supply


voltage at sensor module 2DS

34 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Performance characteristics
Response time Current outputs
t90 = max. 500 ms for an increase from 0 to 20 mA
Current inputs
t90 = max. 330 ms for an increase from 0 to 20 mA
Digital inputs and outputs
t90 = max. 330 ms for an increase from low to high

Reference temperature 25 °C (77 °F)

Measured error for sensor → Documentation of the connected sensor


inputs

Measured error for current Typical measured errors:


inputs and outputs < 20 µA (with current values < 4 mA)
< 50 µA (with current values 4 to 20 mA)
at 25 ˚C (77˚ F) each
Additional measured error depending on the temperature:
< 1.5 µA/K

Frequency tolerance of ≤ 1%
digital inputs and outputs

Resolution of current inputs < 5 µA


and outputs

Repeatability → Documentation of the connected sensor

Installation
Mounting on DIN rail as per NOTICE
IEC 60715 Incorrect mounting location in the cabinet, spacing regulations not observed
Possible malfunctions as a result of heat buildup and interference from neighboring devices!
‣ Do not position the device directly above sources of heat. The temperature specification must be
observed.
‣ The components are designed for convection-based cooling. Avoid heat buildup. Ensure openings
are not covered, e.g. by cables.
‣ Observe the specified distances to other devices.
‣ Physically separate the device from frequency converters and high-voltage devices.
‣ Recommended installation direction: horizontal. The specified ambient conditions, and
particularly the ambient temperatures, only apply for horizontal installation.
‣ Vertical orientation is also possible. However, this requires additional fixing clips at the place of
installation to hold the device in position on the DIN rail.
‣ Recommended installation of power unit for CM444R and CM448R: to the left of the device.

Endress+Hauser 35
Liquiline CM442R/CM444R/CM448R

The following minimum clearance specifications must be observed:


• Distances at the side in relation to other devices incl. power units and to the wall of the cabinet:
at least 20 mm (0.79 inch)
• Distance above and below the device and depth distance (to control cabinet door or other devices
installed there):
at least 50 mm (1.97 inch)

min. 50 (1.97)

min. 50 (1.97)
min. 50 (1.97)

min. 20 (0.79)

min. 20 (0.79)

A0039735

 23 Minimum clearance in mm (in)

Wall mounting
83 (3.27) 205.5 (8.09)
100 (3.94)

100 (3.94)
75 (2.95)

75 (2.95)
4 x ! 4.5 (0.18) 4 x ! 4.5 (0.18)

A0025370

 24 Drilling pattern for wall mounting in mm (in)

36 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Mounting the external The mounting plate also serves as the drilling template. The marks on the side help you mark
display the position of the drill holes.

140 (5.51)
105 (4.13)
4 x min. Ø 8 (0.31)

min. Ø 15 (0.59) a

140 (5.51)
105 (4.13)
b

A0025371

 25 Mounting plate of external display, dimensions in mm (in)


a Retaining tab
b Production-related recesses, no function for the user

Environment
Ambient temperature range CM442R
0 to 60 ˚C (32 to 140 ˚F)
CM444R
• Generally 0 to 55 ˚C (32 to 130 ˚F), with the exception of packages under the second list item
• 0 to 50 ˚C (32 to 120 ˚F) for the following packages:
• CM444R-**M40A7FI*+...
• CM444R-**M40A7FK*+...
• CM444R-**M4AA5F4*+...
• CM444R-**M4AA5FF*+...
• CM444R-**M4AA5FH*+...
• CM444R-**M4AA5FI*+...
• CM444R-**M4AA5FK*+...
• CM444R-**M4AA5FM*+...
• CM444R-**M4BA5F4*+...
• CM444R-**M4BA5FF*+...
• CM444R-**M4BA5FH*+...
• CM444R-**M4BA5FI*+...
• CM444R-**M4BA5FK*+...
• CM444R-**M4BA5FM*+...
• CM444R-**M4DA5F4*+...
• CM444R-**M4DA5FF*+...
• CM444R-**M4DA5FH*+...
• CM444R-**M4DA5FI*+...
• CM444R-**M4DA5FK*+...
• CM444R-**M4DA5FM*+...

Endress+Hauser 37
Liquiline CM442R/CM444R/CM448R

CM448R
• Generally 0 to 55 ˚C (32 to 130 ˚F), with the exception of packages under the second list item
• 0 to 50 ˚C (32 to 120 ˚F) for the following packages:
• CM448R-***6AA*+...
• CM448R-***8A4*+...
• CM448R-***8A5*+...
• CM448R-**28A3*+...
• CM448R-**38A3*+...
• CM448R-**48A3*+...
• CM448R-**58A3*+...
• CM448R-**68A3*+...
• CM448R-**26A5*+...
• CM448R-**36A5*+...
• CM448R-**46A5*+...
• CM448R-**56A5*+...
• CM448R-**66A5*+...
• CM448R-**22A7*+...
• CM448R-**32A7*+...
• CM448R-**42A7*+...
• CM448R-**52A7*+...
• CM448R-**62A7*+...
• CM448R-**A6A5*+...
• CM448R-**A6A7*+...
• CM448R-**B6A5*+...
• CM448R-**B6A7*+...
• CM448R-**C6A5*+...
• CM448R-**C6A7*+...
• CM448R-**D6A5*+...
• CM448R-**D6A7*+...
External display (optional)
–20 to 60 ˚C (0 to 140 ˚F)

Storage temperature –25 to 85 °C (–13 to 185 °F)

Humidity DIN rail device


5 to 85%, not condensing
External display (in installed state)
5 to 95%, not condensing

Degree of protection DIN rail device


IP20 shock protection
External display
IP66 front-panel, when installed correctly including seal for housing door

Climate class As per IEC 60654-1: B2

Vibration resistance Environmental tests


Vibration test based on DIN EN 60068-2, October 2008
Vibration test based on DIN EN 60654-3, August 1998

Wall mounting
Frequency range 10 to 150 Hz (sinusoidal)
Amplitude 10 to 12.9 Hz: 0.75 mm
12.9 to 150 Hz: 0.5 g 1)
Test duration 10 frequency cycles/ spatial axis, in 3 spatial axes (1 oct./min)

1) g ... gravitational acceleration (1 g ≈ 9.81 m/s2)

Electromagnetic Interference emission and interference immunity as per EN 61326-1:2013, Class A for Industry
compatibility

38 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Electrical safety IEC 61010-1, Class I equipment


Low voltage: overvoltage category II
Environment < 2000 m (< 6562 ft) above MSL

Degree of contamination DIN rail device


The product is suitable for pollution degree 2.

Optional display
The product is suitable for pollution degree 4.

Mechanical construction
Dimensions CM442R

97 (3.82)

89 (3.50)

91 (3.58)
7.5 (0.30)

TS 35

83 (3.27)

100 (3.94)
114 (4.49)
75 (2.95)

TS 35

99 (3.90)
A0039729

 26 Dimensions in mm (inch)

Endress+Hauser 39
Liquiline CM442R/CM444R/CM448R

CM444R and CM448R

205.5 (8.09)

114 (4.49)
100 (3.94)

75 (2.95)
TS 35

97 (3.82)

89 (3.50)

91 (3.58)
7.5 (0.30)
TS 35

220 (8.66)
A0039730

 27 Dimensions in mm (inch)

40 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Optional display

54 (2.13)
39.5 (1.56)
144 (5.67) 144 (5.67)

A0025346

 28 Dimensions in mm (inch)

External power units (CM444R and CM448R only)


Depending on the version ordered, a power unit for connection to 230 V or 24 V is supplied. There
are two delivery variants for each version (cannot be selected). The factory-preferred variant is
shown on the left in each case.

90 (3.54)
121 (4.76)

32 (1.26) 125 (4.92) 40 (1.58) 99 (3.99)

A0025738 A0025739

 29 Power unit 230 V  30 Power unit 230 V (optional)

32 (1.26) 102 (4.02)


45 (1.77) 100 (3.94)

out
124 (4.88)
75 (2.95)

in

A0025784 A0025786

 31 Power unit 24 V  32 Power unit 24 V (optional)

Endress+Hauser 41
Liquiline CM442R/CM444R/CM448R

Service display (accessories)


The service display comprises:
• Portable display (same dimensions as under "Optional display")
• Cover to protect the display and to hook it onto the (open) cabinet door

193 (7.60)
123 (4.84) 36 (1.42)
B 13.5 (0.53)

B-B

A A
157 (6.18)
119 (4.69)

143 (5.63)
6
R1

40 (1.57)
B
47 (1.85)

144.5 (5.69)
A-A

A0025343

 33 Dimensions of the service display cover in mm (inch)

Weight depending on the version:


CM442R (fully configured) Approx. 0.45 kg (1 lbs)
CM444R and CM448R (fully configured) Approx. 0.95 kg (2.1 lbs)
Individual module Approx. 0.06 kg (0.13 lbs)
External display (excluding cables) Approx. 0.56 kg (1.2 lbs)
Service display cover 0.46 kg (1 lbs)
External power unit (CM444R, CM448R) 0.27 to 0.42 kg (0.60 to 0.92 lbs), depending on the power
unit variant

Materials DIN rail housing PC-FR

Display cover PC-FR

Display seal EPDM

Soft keys EPDM

Module covers PBT GF30 FR

Terminal strip Nickel-plated brass

Ground terminals Stainless steel 1.4301 (AISI304)

Threaded fasteners Stainless steel 1.4301 (AISI304)

Mounting plate (optional display) Stainless steel 1.4301 (AISI304)

Securing screws (optional display) Steel, galvanized

Cover for service display (accessories) EPDM

42 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Operability
External display Graphic display:
• Resolution: 240 x 160 pixel
• Back light with switch-off function
• Red display background for alarms alerts users to errors
• Transflective display technology for maximum contrast even in bright environments
• User-definable measuring menus mean you can always keep track of the values that are important
for your application.

Operating concept The simple and structured operating concept sets new standards:
• Intuitive operation with the navigator and soft keys
• Fast configuration of application-specific measurement options
• Easy configuration and diagnosis thanks to plain-text display
• All languages that can be ordered are available in every device

A0025229
 35 Plain-text menu
 34 Operation via external, optional display

Local operation via external,


1
optional display

Outlet 1
11:09:15 29.01.2013
CH 1:1 pH 6.99 pH
CH 1:1 Temperature 25.1 °C
CH 1:2 Nitrate 9.02 mg/l
CH 2:1 SAC 2.02 1/m
CH 2:2 Turbidity 1022 FNU
CH 3:1 Current Input [Flow] 15.1 nA
MENU

2 3
A0025231

 36 Overview of operation
1 Display (with red display background in alarm condition)
2 Soft keys (function depends on menu)
3 Navigator (jog/shuttle and press/hold function)

Endress+Hauser 43
Liquiline CM442R/CM444R/CM448R

Remote operation Via HART (e.g. via HART modem and FieldCare)

250 ... 500 Ω


31 2
32
USB
1

3
A0039620

 37 HART using modem


1 Device module Base2-L, -H or -E: current output 1 with HART
2 HART modem for connection to PC, e.g. Commubox FXA191 (RS232) or FXA195 1) (USB)
3 HART handheld terminal
1) Switch position "on" (substitutes the resistor)

Via PROFIBUS DP

Master class 1
(PLC, DCS)
Master class 2 (FieldCare, PDM)

PROFIBUS DP
(RS485)
T T

FXA291

Ethernet

A0039617

 38 PROFIBUS DP
T Terminating resistor

44 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Via Modbus RS485

Modbus Master / Gateway

Modbus RS485
(RTU, ASCII)
T T

FXA291

Ethernet

A0039615

 39 Modbus RS485
T Terminating resistor

Via Ethernet: web server/Modbus TCP/PROFINET/EtherNet/IP

Ethernet client
FieldCare/
Webbrowser

Ethernet

FXA291

A0039616

 40 Modbus TCP or EtherNet/IP or PROFINET

Language packages The language selected in the product structure is the operating language preset at the factory. All
other languages can be selected using the menu.
• English (US)
• German
• Chinese (Simplified, PR China)
• Czech
• Dutch
• French
• Italian
• Japanese
• Polish
• Portuguese
• Russian
• Spanish
• Swedish
• Turkish

Endress+Hauser 45
Liquiline CM442R/CM444R/CM448R

• Hungarian
• Croatian
• Vietnamese
The availability of other languages can be checked via the product structure at www.endress.com/
cm442R or .../cm444R or .../cm448R.

Certificates and approvals


 mark The product meets the requirements of the harmonized European standards. As such, it complies
with the legal specifications of the EU directives. The manufacturer confirms successful testing of the
product by affixing to it the  mark.

EAC The product has been certified according to guidelines TP TC 004/2011 and TP TC 020/2011 which
apply in the European Economic Area (EEA). The EAC conformity mark is affixed to the product.

cCSAus The device has been certified with regard to its electrical safety and for NI Class I Div. 2 cCSAus
explosion-proof environments. It meets the requirements in accordance with:
• CLASS 2252 06 - Process Control Equipment
• CLASS 2252 86 - Process Control Equipment - Certified to US Standards
• CLASS 2258 03 - Process Control Equipment - Intrinsically Safe and Non-incendive Systems - For
Hazardous Locations
• CLASS 2258 83 - Process Control Equipment - Intrinsically Safe and Non-incendive Systems - For
Hazardous Locations - Certified to US Standards
• FM3600
• FM3611
• FM3810
• UL50E
• IEC 60529
• CAN/CSA-C22.2 No. 0
• CAN/CSA C22.2 No. 94
• CSA Std. C22.2 No. 213
• CAN/CSA-C22.2 No. 61010-1
• CAN/CSA-C22.2 No. 60529
• UL/ANSI/ISA 61010-1
• ANSI - ISA 12 12 01

‣ The device must be installed in a housing or cabinet where access is only possible with a tool or
key.

Marine approvals A selection of the devices and sensors have type approval for marine applications, issued by the
following classification societies: ABS (American Bureau of Shipping), BV (Bureau Veritas), DNV-GL
(Det Norske Veritas-Germanischer Lloyd) and LR (Lloyd’s Register). Details of the order codes of the
approved devices and sensors, and the installation and ambient conditions, are provided in the
relevant certificates for marine applications on the product page on the Internet.

Ordering information
Product page www.endress.com/cm442r
www.endress.com/cm444r
www.endress.com/cm448r

Product Configurator On the product page there is a Configure button to the right of the product image.
1. Click this button.
 The Configurator opens in a separate window.

46 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

2. Select all the options to configure the device in line with your requirements.
 In this way, you receive a valid and complete order code for the device.
3. Export the order code as a PDF or Excel file. To do so, click the appropriate button on the right
above the selection window.

For many products you also have the option of downloading CAD or 2D drawings of the
selected product version. Click the CAD tab for this and select the desired file type using
picklists.

Scope of delivery The scope of delivery comprises:


• 1 multichannel controller in the version ordered
• 1 mounting plate
• 1 external display (if selected as an option) 2)
• 1 DIN rail power unit incl. Cable (CM444R and CM448R only)
• 1 printed copy of the Operating Instructions for DIN rail power unit
• 1 printed copy of the Brief Operating Instructions in the language ordered

Accessories
The following are the most important accessories available at the time this documentation was
issued.
‣ For accessories not listed here, please contact your Service or Sales Center.

Measuring cable Memosens data cable CYK10


• For digital sensors with Memosens technology
• Product Configurator on the product page: www.endress.com/cyk10
Technical Information TI00118C

Memosens data cable CYK11


• Extension cable for digital sensors with Memosens protocol
• Product Configurator on the product page: www.endress.com/cyk11
Technical Information TI00118C

Sensors Glass electrodes


Orbisint CPS11D
• pH sensor for process technology
• Optional SIL version for connecting to SIL transmitter
• With dirt-repellent PTFE diaphragm
• Product Configurator on the product page: www.endress.com/cps11d
Technical Information TI00028C

Memosens CPS31D
• pH electrode with gel-filled reference system with ceramic diaphragm
• Product Configurator on the product page: www.endress.com/cps31d
Technical Information TI00030C

Ceraliquid CPS41D
• pH electrode with ceramic junction and KCl liquid electrolyte
• Product Configurator on the product page: www.endress.com/cps41d
Technical Information TI00079C

2) The external display can be selected as an option in the order structure or ordered as an accessory at a later stage.

Endress+Hauser 47
Liquiline CM442R/CM444R/CM448R

Ceragel CPS71D
• pH electrode with reference system including ion trap
• Product Configurator on the product page: www.endress.com/cps71d
Technical Information TI00245C

Memosens CPS171D
• pH electrode for bio-fermenters with digital Memosens technology
• Product Configurator on the product page: www.endress.com/cps171d
Technical Information TI01254C

Orbipore CPS91D
• pH electrode with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps91d
Technical Information TI00375C

Orbipac CPF81D
• Compact pH sensor for installation or immersion operation
• In industrial water and wastewater
• Product Configurator on the product page: www.endress.com/cpf81d
Technical Information TI00191C

Enamel pH electrodes
Ceramax CPS341D
• pH electrode with pH-sensitive enamel
• Meets highest demands of measuring accuracy, pressure, temperature, sterility and durability
• Product Configurator on the product page: www.endress.com/cps341d
Technical Information TI00468C

ORP sensors
Orbisint CPS12D
• ORP sensor for process technology
• Product Configurator on the product page: www.endress.com/cps12d
Technical Information TI00367C

Ceraliquid CPS42D
• ORP electrode with ceramic junction and KCl liquid electrolyte
• Product Configurator on the product page: www.endress.com/cps42d
Technical Information TI00373C

Ceragel CPS72D
• ORP electrode with reference system including ion trap
• Product Configurator on the product page: www.endress.com/cps72d
Technical Information TI00374C

Orbipac CPF82D
• Compact ORP sensor for installation or immersion operation in process water and wastewater
• Product Configurator on the product page: www.endress.com/cpf82d
Technical Information TI00191C

Orbipore CPS92D
• ORP electrode with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps92d
Technical Information TI00435C

48 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

pH ISFET sensors
Tophit CPS441D
• Sterilizable ISFET sensor for low-conductivity media
• Liquid KCl electrolyte
• Product Configurator on the product page: www.endress.com/cps441d
Technical Information TI00352C

Tophit CPS471D
• Sterilizable and autoclavable ISFET sensor for food and pharmaceutics, process engineering
• Water treatment and biotechnology
• Product Configurator on the product page: www.endress.com/cps471d
Technical Information TI00283C

Tophit CPS491D
• ISFET sensor with open aperture for media with high dirt load
• Product Configurator on the product page: www.endress.com/cps491d
Technical Information TI00377C

pH and ORP combined sensors


Memosens CPS16D
• Combined pH/ORP sensor for process technology
• With dirt-repellent PTFE diaphragm
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps16D
Technical Information TI00503C

Memosens CPS76D
• Combined pH/ORP sensor for process technology
• Hygienic and sterile applications
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps76d
Technical Information TI00506C

Memosens CPS96D
• Combined pH/ORP sensor for chemical processes
• With poison-resistant reference with ion trap
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cps96d
Technical Information TI00507C

Conductivity sensors with inductive measurement of conductivity


Indumax CLS50D
• High-durability inductive conductivity sensor
• For standard and hazardous area applications
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cls50d
Technical Information TI00182C

Indumax H CLS54D
• Inductive conductivity sensor
• With certified, hygienic design for foodstuffs, beverages, pharmaceuticals and biotechnology
• Product Configurator on the product page: www.endress.com/cls54d
Technical Information TI00508C

Endress+Hauser 49
Liquiline CM442R/CM444R/CM448R

Conductivity sensors with conductive measurement of conductivity


Condumax CLS15D
• Conductive conductivity sensor
• For pure water, ultrapure water and hazardous area applications
• Product Configurator on the product page: www.endress.com/CLS15d
Technical Information TI00109C

Condumax CLS16D
• Hygienic, conductive conductivity sensor
• For pure water, ultrapure water and Ex applications
• With EHEDG and 3A approval
• Product Configurator on the product page: www.endress.com/CLS16d
Technical Information TI00227C

Condumax CLS21D
• Two-electrode sensor in plug-in head version version
• Product Configurator on the product page: www.endress.com/CLS21d
Technical Information TI00085C

Memosens CLS82D
• Four-electrode sensor
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cls82d
Technical Information TI01188C

Oxygen sensors
Oxymax COS22D
• Sterilizable sensor for dissolved oxygen
• With Memosens technology or as an analog sensor
• Product Configurator on the product page: www.endress.com/cos22d
Technical Information TI00446C

Oxymax COS51D
• Amperometric sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos51d
Technical Information TI00413C

Oxymax COS61D
• Optical oxygen sensor for drinking water and industrial water measurement
• Measuring principle: quenching
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos61d
Technical Information TI00387C

Memosens COS81D
• Sterilizable, optical sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos81d
Technical Information TI01201C

50 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Disinfection sensors
CCS142D
• Membrane-covered amperometric sensor for free chlorine
• Measuring range 0.01 to 20 mg/l
• With Memosens technology
• Product Configurator on the product page: www.endress.com/ccs142d
Technical Information TI00419C

Ion-selective sensors
ISEmax CAS40D
• Ion selective sensors
• Product Configurator on the product page: www.endress.com/cas40d
Technical Information TI00491C

Turbidity sensors
Turbimax CUS51D
• For nephelometric measurements of turbidity and solids in wastewater
• 4-beam scattered light method
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cus51d
Technical Information TI00461C

Turbimax CUS52D
• Hygienic Memosens sensor for turbidity measurement in drinking water, process water and in
utilities
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cus52d
Technical Information TI01136C

SAC and nitrate sensors


Viomax CAS51D
• SAC and nitrate measurement in drinking water and wastewater
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cas51d
Technical Information TI00459C

Interface measurement
Turbimax CUS71D
• Immersion sensor for interface measurement
• Ultrasonic interface sensor
• Product Configurator on the product page: www.endress.com/cus71d
Technical Information TI00490C

Additional functionality Hardware extension modules


Kit, extension module AOR
• 2 x relay, 2 x 0/4 to 20 mA analog output
• Order No. 71111053
Kit, extension module 2R
• 2 x relay
• Order No. 71125375
Kit, extension module 4R
• 4 x relay
• Order No. 71125376

Endress+Hauser 51
Liquiline CM442R/CM444R/CM448R

Kit, extension module 2AO


• 2 x 0/4 to 20 mA analog output
• Order No. 71135632
Kit, extension module 4AO
• 4 x analog output 0/4 to 20 mA
• Order No. 71135633
Kit, extension module 2DS
• 2 x digital sensor, Memosens
• Order No. 71135631
Kit, extension module 2AI
• 2 x 0/4 to 20 mA analog input
• Order No. 71135639
Kit, extension module DIO
• 2 x digital input
• 2 x digital output
• Auxiliary voltage supply for digital output
• Order No. 71135638
Kit, extension module 485
• Can be extended to PROFIBUS DP or Modbus RS485. This requires an additional activation code
which can be ordered separately.
• Order No. 71135634
Upgrade kit, extension module 485 with PROFIBUS DP
• Extension module 485
• PROFIBUS DP (+ Ethernet configuration)
• Order No. 71140888
Upgrade kit, extension module 485 with Modbus RS485
• Extension module 485
• Modbus RS485 (+ Ethernet configuration)
• Order No. 71140889

Firmware and activation codes


SD card with Liquiline firmware
• Industrial Flash Drive, 1 GB
• Order No. 71127100
You must quote the serial number of the device when ordering the activation code.

Activation code for digital HART communication


Order No. 71128428
Activation code for PROFIBUS DP
Order No. 71135635
Activation code for Modbus RS485
Order No. 71135636
Activation code for PROFINET + web server for BASE2
Order No. 71449901
Activation code for Ethernet/IP + web server for BASE2
Order No. 71449914
Activation code for Modbus TCP + web server for BASE2
Order No. 71449915
Activation code for web server for BASE2
Order No. 71449918
Kit CM442R: activation code for 2nd digital sensor input
Order No. 71114663
Kit CM444R/CM448R: upgrade code for 2 x 0/4 to 20 mA for BASE2-E
On request
Activation code for feedforward control
• Requires current input or fieldbus communication
• Order No. 71211288

52 Endress+Hauser
Liquiline CM442R/CM444R/CM448R

Activation code for measuring range switch


• Requires digital inputs or fieldbus communication
• Order No. 71211289
Activation code for ChemocleanPlus
• Requires relays or digital outputs or fieldbus communication and optional digital inputs
• Order No. 71239104
Activation code for Heartbeat Verification and Monitoring
Order No. 71367524
Activation code for ion exchanger operating time
• Configure the mathematical function
• Order No. 71367531
Activation code for mathematics
• Formula editor
• Order No. 71367541

Software Memobase Plus CYZ71D


• PC software to support laboratory calibration
• Visualization and documentation of sensor management
• Sensor calibrations stored in database
• Product Configurator on the product page: www.endress.com/cyz71d
Technical Information TI00502C

Field Data Manager Software MS20


• PC software for central data management
• Visualization of series of measurements and logbook events
• SQL database for secure data storage
FieldCare SFE500
• Universal tool for field device configuration and management
• Supplied with a complete library of certified DTMs (Device Type Manager) for operation of Endress
+Hauser field devices
• Order according to product order structure
• www.endress.com/sfe500

Other accessories External display 3)


Graphic display
• For installation in the control cabinet door or panel
• Order No. 71185295
Service display
• Portable, for commissioning
• Order No. 71185296

SD card
• Industrial Flash Drive, 1 GB
• Order No. 71110815

Communication-specific accessories
Commubox FXA195
Intrinsically safe HART communication with FieldCare via the USB port
Technical Information TI00404F

Commubox FXA291
Connects the CDI interface of measuring devices with the USB port of the computer or laptop
Technical Information TI00405C

3) The external display can be selected as an option in the product structure or ordered subsequently as an accessory.

Endress+Hauser 53
Liquiline CM442R/CM444R/CM448R

Wireless HART adapter SWA70


• Wireless device connection
• Easily integrated, offers data protection and transmission safety, can be operated in parallel with
other wireless networks, minimum cabling complexity
Technical Information TI00061S

System components
RIA14, RIA16
• Field display unit for integration into 4-20 mA circuits
• RIA14 in flameproof metal enclosure
Technical Information TI00143R and TI00144R

RIA15
• Process display unit, Digital display unit for integration into 4-20 mA circuits
• Panel mounting
• With optional HART communication
Technical Information TI01043K

54 Endress+Hauser