Product specification
Constant climate chamber
HPP750eco
Maximum energy efficiency in continuous
operation - tailor-made for stability
studies according to ICH guidelines,
stability tests for cosmetics and foods as
well as environmental testing and
material testing
On this page, you can find all the essential technical data on the Memmert stability chamber HPP. Our customer
relations team will be pleased to help if you want further information. If you should require a customised special
solution, please contact our technical specialists at [email protected].
Page 1/5 | 27.07.2021 Memmert GmbH + Co. KG | Tel. +49 (0) 9122/925-0 | E-Mail [email protected] | 2
Memmert Constant climate chamber HPP750eco - en
Temperature
Setting temperature range without light, with humidity: +5 to +70 °C
Setting temperature range with light, with humidity: +5 to +70 °C
Setting temperature range with light, without humidity: 0 to +70 °C
Setting temperature range without light, without humidity: 0 to +70 °C
Working temperature range with light, without or with humidity: 15 to +40 °C
Working temperature range without light, with humidity: +5 (at least 20 below ambient temperature) to +70 °C
Working temperature range without light, without humidity: 0 (at least 20 below ambient temperature) to +70 °C
Setting accuracy 0.1 °C
temperature
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an
error
Humidity
Setting range humidity 10 - 90 % rh
Humidity humidity supply with distilled water from external tank by self-priming pump
Humidification humidification by hot steam generator
Dehumidification dehumidification by cold trap using the Peltier technology
Setting accuracy humidity 0.5 % rh
Control technology
ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition
TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian
adjustable parameters temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones,
summertime/wintertime
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE adapting the distribution of the heating performance of the upper and lower heating circuit from -50 %
to +50 %
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable values each, temperature and humidity
Communication
Interface Ethernet LAN, USB
Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes
via Ethernet interface or USB port
Page 2/5 | 27.07.2021 Memmert GmbH + Co. KG | Tel. +49 (0) 9122/925-0 | E-Mail [email protected] | 2
Memmert Constant climate chamber HPP750eco - en
Safety
Temperature control over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter
TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the
setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating
function is switched off in case of overtemperature, cooling function in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature and humidity control
Alarm visual and acoustic
Heating concept
Peltier energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)
Standard equipment
Internals 2 stainless steel grid(s), electropolished
Standard accessories Water tank including connection hose (110-750: 2,5 liter, 1060-2200: 10 liter)
Works calibration certificate standard values +10 °C and +37 °C, 60 % rh at +30 °C
Door inner glass doors
Door fully insulated stainless steel doors with2-point locking (compression door lock)
Stainless steel interior
Dimensions w(A) x h(B) x d(C): 1040 x 1200 x 600 mm (d less 32 mm for fan - Peltier)
Volume 749 l
Max. number of internals 14
Max. loading of chamber 200 kg
Max. loading per internal 30 kg
Textured stainless steel casing
Dimensions w(D) x h(E) x d(F): 1224 x 1720 x 755 mm (d +56mm door handle & +111mm Peltier element)
Installation on lockable castors
Housing rear zinc-plated steel
Electrical data
Voltage 230 V, 50/60 Hz
Electrical load approx. 1400 W
Voltage 115 V, 50/60 Hz
Electrical load approx. 1400 W
Page 3/5 | 27.07.2021 Memmert GmbH + Co. KG | Tel. +49 (0) 9122/925-0 | E-Mail [email protected] | 2
Memmert Constant climate chamber HPP750eco - en
Ambient conditions
Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance
from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances
must not be less than 5 cm.
Ambient temperature 16 ºC to 40 ºC
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2
Packing/shipping data
Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. w x h x d: 1330 x 1910 x 1050 mm
carton
Net weight approx. 208 kg
Gross weight carton approx. 279 kg
The Peltier-cooled climate chamber HPPeco is
specifically designed for stability studies
according to ICH guidelines (Q1A) as well as
stability tests for cosmetics and food. With the
large working range ascending from the freezing
point, the climate chamber is also ideal for
working materials testing and environmental tests
in industry.
Page 4/5 | 27.07.2021 Memmert GmbH + Co. KG | Tel. +49 (0) 9122/925-0 | E-Mail [email protected] | 2
Memmert Constant climate chamber HPP750eco - en
Standard units are safety-approved and bear the test marks
Page 5/5 | 27.07.2021 Memmert GmbH + Co. KG | Tel. +49 (0) 9122/925-0 | E-Mail [email protected] | 2