Dental Unit and Chair: Important
Dental Unit and Chair: Important
OPERATING INSTRUCTIONS
This Instruction Manual describes how to use the SP-CLEO II.
A
W
IMPORTANT
Thoroughly read this Instruction Manual before use, and use the SP-CLEO II according to
the instructions.
Keep this Instruction Manual carefully, and read it again whenever it is needed.
This product is a dental unit used for dental care.
People other than dentists and other dental professionals should not use this product.
TABEL OF CONTENTS
Environmental Requirements
Ambient Temperature Operating +5℃ - +40℃ Storage -10℃ - +50℃
Humidity 10 % - 80%
Atmospherical Pressure 600 hPa - 1060 hPa
Important Notes
In case of the troubles, please contat Takara Belmont offices or your dealers.
Do not disassemble or attempt to repair.
Disassembly, repair or modifications shoud only be done by a qualified repair technician.
Attempts at disassembly, repair or modifications may lead to abnormal operation and accidents.
Chair last position Chair auto return Chair preset1 Chair preset2
To Recline To raise
To raise the chair To lower the chair
the backrest the backrest
Chair last position Chair auto return Chair preset1 Chair preset2
Fiber optic
handpiece light Syringe Function Store
on//off
Caution
It means “caution, Type B
warnings, or Manufacturer Applied Parts
possibility to
danger”.
1. SAFETY PRECAUTIONS
For proper operation, please carefully read through the Safety Precautions before using the product.
Items described herein are provided in order to ensure safe and appropriate use of the product, and to prevent
property damage or personal injury to the operator and other persons.
Furthermore, the precautions describe potential hazards due to improper handling and classify hazards according to
degree of damage and level of urgency.
Please be sure to read the items and precautions as they both relate to safety.
Division of caution level Magnitude of danger and damage and extent of urgency
WA R N I N G This word refers to precautions that if ignored may result in fatal or serious injuries.
This word refers to precautions that if ignored may result in minor or moderate injury
CAUTION
or property damage.
NOTICE This word refers to precautions regarding safety.
WA R N I N G
1. Be sure to switch off the breaker for the equipment when it will not be in use for a long period of time.
When the equipment will not be in use for a long period of time, such as at the end of the day or a non-business
day, be sure to switch off the applicable breaker in the office and disconnect the power plug from the outlet.
Failure to heed this warning may result in electrical leakage or fire due to deterioration of insulation.
2. Be sure to turn off the main switch at the end of the day and during break times.
Be sure to turn off the main switch at the end of the day and during break times. This will prevent
malfunctions and resulting hazards due to accidental contact with the equipment.
Only authorized personnel should disassemble, repair or modify the equipment. Failure to heed this
warning may cause an accident, failure, electrical shock, or fire.
Be sure to ground the equipment firmly and correctly. (To ground the equipment, contact a qualified
technician.) Failure to heed this warning may result in electrical shock during failure or electrical
leakage.
6. Be sure to turn off the main switch when a high frequency surgical instrument (HF-Surg) is used.
Use of a HF-Surg may cause the equipment to malfunction due to generated noise. Be sure to turn off the
main switch when using the HF-Surg.
Use caution towards patients with a pacemaker. If the patient experiences anything unusual while
product is operating, immediately turn off the power of the product and discontinue use.
Using the product without performing maintenance and inspection programs may result in bodily
injury or property damage.
Refer to pages 43 for instructions on maintenance and inspections.
-1-
1. SAFETY PRECAUTIONS
WA R N I N G
9. Be sure to mount the mirror cover.
The mirror cover for the dental light must be attached whenever the light is used. Touching the light
bulb may result in burns.
Refer to the operating instructions on the dental light for details.
10. Be sure to turn off the power when replacing the light bulb.
The power must be turned off when replacing the dental light bulb. Failure to heed this warning may
result in electrical shock.
The specified halogen light bulb must be used.
The bulb and the bulb holder may be hot. Wait until they become cool before replacing the bulb.
Do not touch the halogen bulb with bare hands.
Refer to the operating instructions on the dental light for details.
CAUTION
1. The system must be operated or handled by a qualified dentist or by dental staff under the supervision of a
Persons other than dentists or other dental professionals must not use the system.
Be careful to avoid accidents due to mischief or careless operations by the patient (children in particular).
Keep an eye on the patient.
-2-
1. SAFETY PRECAUTIONS
CAUTION
8. Request for flushout of water staying in the unit
In order to maintain the quality of the treatment water to stably supply water to the handpieces, be sure
to thoroughly flush out any remaining water in the unit before starting treatment. Refer to the pages for
the flushing out procedures.
9. Cautions when shifting/moving the instrument table:
When moving the instrument table, pay attention to the surrounding area. Failure to heed this warning
may result in injuries by the tips of the handpieces.
The instrument table must be moved with the handle of the unit.
10. Do not place hot objects on the instrument table.
Do not place hot objects on the instrument table. Failure to heed this warning may result in deformation or
discoloration of the table.
11. Be sure to wipe chemicals off immediately when spilled on the system.
If chemicals or water are spilled on the system, immediately wipe off with a dry, soft towel. Failure to
heed this warning may result in system malfunction, electrical leakage, stains, or rust on the equipment.
12. Take care not to activate the water heater without water in it.
Activating the water heater without water in it will burn out the heating element, resulting in fire.
13. Be sure to use the cover over the scaler tip.
After using the scaler, be sure to use the designated tip cover (for models with a cover) before placing
it in the holder. Without the cover, there is a possibility of injury caused by the scaler tip.
14. Observe the following when handling handpieces:
Change the rotation speeds of the micro-motor and detach the burr of the handpieces only after motor
rotation comes to a complete stop for safety purposes. Refer to the operating instructions that come with
each handpiece for details.
15. Observe the following when handling the film viewer:
Handle the film viewer carefully, since a fluorescent lamp is encased. Be sure to turn off the lamp when not
in use.
16. Observe the following concerning the cart hose:
Do not tread on the cart hose. Failure to heed this warning may result in damage to the hose, as well
as people tripping over.
17. Cautions when adjusting the height of the cart-type instrument table:
After adjusting the height, be sure to fasten the loosened lock screw. Failure to heed this warning may
cause the table to shift, resulting in an accident.
18. Pay attention to entrapment during automatic operation of the treatment unit.
During automatic operation of the treatment unit, pay attention to surrounding areas before initiating
operations. Failure to heed this warning may result in breakage of the backrest, stool, or instrument table.
Furthermore, check that the patient is in the correct position and always monitor the patient during system
operation.
19. Caution when cleaning the control panel (membrane switch):
Seepage of sterilization liquid into the back of the membrane sheet will cause switch failure. Be sure to use a
paper towel applied with sterilization liquid to clean the membrane sheet surface.
-3-
1. SAFETY PRECAUTIONS
CAUTION
20. Observe the following when cleaning the Spittoon bowl:
Never use sandpaper, wire wool, or detergents containing abrasive compounds to clean the Spittoon bowl.
Do not use overly acidic or alkaline pipe cleaning agents, as they may corrode metals.
21. Pay attention to water discharge when detaching the Spittoon bowl.
When detaching the Spittoon bowl, be sure to turn off the main switch; otherwise, the water supply sensor
will sense hands or objects during the detachment, which will activate the bowl flush mechanism. Water
may spray out and leak into areas that may cause rust.
22. Caution when cleaning the plastic covers:
When cleaning the plastic covers, use only the desinfectant (Durr FD333) recommended by this company
or a diluted neutral detergent, or the plastic cover may crack.
After cleaning, make sure to wipe off all residue with a dry and soft cloth.
23. Air pressure of the clean water system (optional) must be less than 0.2Mpa.
The air pressure to the clean water system must be regulated to less than 0.2Mpa. Excessive pressure may
rupture the clean water system.
24. Do not use water for the clean water system (optional) other than purified water, distilled water or pure water.
For the clean water system use only purified water, distilled water or pure water.
Do not use such as mouth wash or electrolytic solution which may cause clogging in tubings or adverse
effect on inner valves and instruments.
25. Close the water main valve at the end of the day.
Make sure to close the water main valve at the end of the day to prevent a water leakage.
NOTICE
1. Operational check of the air compressor:
If air is not supplied to the product, the system will not function even if the main switch is
turned on. Before operating the product, turn on the power to the air compressor.
2. The air turbine must be of the type with a water check valve.
The air turbine must have a water check valve. If the turbine is not the correct type, contact your
dealer or our company.
The function of the flow adjustment valve/knob is to regulate, not to stop, air/water flow.
Do not overly tighten the valve/knob or the knob may break.
-4-
1.SAFETY PRECAUTIONS
① ② ⑥
③
⑧
④ A
W
⑩
⑦ ⑨
⑤
① Be aware of contact with the assistant’ s ⑥ Take care not to be trapped by the lower part
instrument holder. of the legrest.
The upper body of the patient must not go over the Do not put your feet under the legrest.
cuspidor unit.
⑦ Take care not to be trapped by moving parts
② The cuspidor unit’ s rotation of the headrest.
Make sure that the patient does not lean on the Do not allow hands, fingers, or hair to become
cuspidor unit, as pushing on the spittoon bowl causes it entangled in the moving parts of the headrest.
to swing outwards.
⑧ Take care not to be trapped by the lower part
③ Ensure that the mirror cover is securely in place. of the seat.
The mirror cover must be installed securely. Do not put hands or feet into the gap in the lower
part of the seat.
④ Take care not to be trapped by moving parts of the
backrest. ⑨ Check the locking mechanism of the primary
Do not put hands or feet into the gap between the arm.
backrest and the seat. Check that, when moving the instrument table to
the back of the backrest, the chair lock indicator
⑤ Take care not to be trapped by the armrest. on the control panel (membrane switch) blinks
Do not operate the chair with the right armrest swung red, and operation of the chair is halted.
out 180 degrees.
⑩ Be aware of interference between the chair
and the table.
Do not move the table under the moving parts of
the chair.
-5-
2. REQUEST FOR WATER FLUSHOUT (DISCHARGE)
E type:Standard E
Precautions for water quality safety A type:Option A
CAUTION
To maintain the water quality for treatment and for stable water supply to the handpieces, ensure to carry out
flushout of the residual water in the unit before starting treatment.
At the beginning of the week or in the morning, that is, a long period of idle time, the residual water in tubings
and water heater in the unit tends to permit bacteria being propageted.
Ensure to carry out flushout of water lines in the unit before starting treatment, for the safe treatment and for the
operation of the handpiece without any problem.
It is also recommended to flush out the residual water in the unit and in handpiece tubings with fresh water, after
the treatment in the morning and also in the evening to prevent bacteria propagation.
Standard length of time required for flushout of water lines in the unit
Handpiece line About 40 seconds per each of turbine, motor, scaler and syringe.
Air Turbine (About 40 seconds also in case of flushout of all of them in a batch.)
Micro Motor CAUTION
Scaler Apply flushout to the scaler with the body kept fitted. Failure
Syringe (for both doctor / assistant) may result otherwise.
Cuspidor line
Cupfiller (water heater) About 5 minutes for cleaning water in cupfiller line.
Bowl flush
Flushout procedure E
Handpiece line Mode I: For Flushing out Mode II: For Flushing
Pick up one handpiece out of the holder, and hold it the Handpiece water line out the cupfiller and
over the cuspidor bowl with the turbine kept as it is only cuspidor bowl after
and the contra angle connected to the motor. As for flushout procedures, handpiece flushout
(In case the flushout function is provided, pick up please refer to Page 25 As for flushout
all the handpieces in a lump from the holder, and selection and operation of procedures, please refer
hold them on the cuspidor bowl.) flushout Mode I. to Page 25 Selection and
operation of flushout
Cuspidor line
Mode II.
Cupfiller (water heater)
Bowl flush
Flushout procedure A
Handpiece line
Pick up one handpiece out of the holder, and hold it
over the cuspidor bowl with the turbine kept as it is * Applicable when flush-out function was selected as
and the contra angle connected to the motor. optional.
( *In case the flushout function is provided, pick up Turn the toggle switch on and flush-out the handpiece
all the handpieces in a lump from the holder, and
hold them over the cuspidor bowl.)
-6-
3. OVERALL VIEW AND MAJOR COMPONENTS
Cuspidor Bowl
Balance Arm
Foot Controller
Doctor Instruments Holder
Manual Switch
Auto Switch
Main Switch
-7-
3. OVERALL VIEW AND MAJOR COMPONENTS
Handle Handle
Cuspidor Section
Cupfiller Nozzle
Cuspidor Bowl
Light Post
Front Panel
-8-
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
1. Chair Section
1-1. Main Switch
Main Switch
1-2. Armrest ① ②
* For returning the armrest to the former position, observe the procedure, which is opposite to that of lock
release. That is, turn the armrest to the normal working position, and then lock the armrest in position by
pushing it down diagonally forward, which is opposite to the direction mentioned in item 1 above.
Make sure to confirm that the armrest is in the locked safe position before use.
CAUTION
The armrest approaches the floor when the backrest is tilted as
shown to the left. Never operate the chair in the state where the
armrest is turned out of the normal working position. A part of
the human body or any other article may be caught between the
armrest and floor otherwise.
1-3. Headrest
Headrest
1) Height Adjustment Release Button
Press down or pull up the headrest to the desired height. Bar
2) Angle Adjustment
Grasp the headrest release button on headrest mechanism
and move to desired position.
CAUTION
Before moving the headrest, confirm that the patient is placing his/her head in a normal position,
and never leave the patient unsupervised while the headrest is in motion.
Do not use the headrest in a position where the patient feels painful.
Before moving the headrest, confirm that there is no obstacle around it lest a part of the human body
or any other article be caught between the headrest and backrest.
-9-
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
1-4. Stick Switch
Auto Switch
Auto Switch
1) Preset Control
The chair has two preset positions.
(Preset-1 and Preset-2) Momentarily depress the
auto mode stick switch to left side, the chair will
Manual Switch
move to the preset-1 position automatically.
(Preset-2 is operated by right side.)
2) Auto Return
Momentarily depress the auto mode stick switch to
upward, the chair will return to the initial position.
(The seat is fully lowered and the backrest is in the upright position.),
3) Last Position Memory
Momentarily depress the auto mode stick switch to downward at the reclined backrest position (treatment position),
the backrest will raise to the mouth rinsing position automatically. Momentarily depress the auto mode stick switch
to downward again, the backrest will recline to the previous treatment position automatically.
4) Emergency Stop
During automatic procedure (Preset, Auto return and Last position memory), depress of any side of the
stick switch will cancel the automatic movement immediately.
Note : Do not depress the auto mode stick switch for over 3 seconds, because the memorized position
in auto mode may be changed.
Manual Switch
1) Seat Lifting
Keep depressing the manual mode stick switch to upward until the seat is lifted up to the desired position.
2) Seat Lowering
Keep depressing the manual mode stick switch to downward until the seat is lowered to the desired position.
3) Backrest Reclining
Keep depressing the manual mode stick switch to left side until the backrest is reclined to the desired position.
4) Backrest Raising
Keep depressing the manual mode stick switch to right side until the backrest is rased up to the desired position.
- 10 -
Handpiece Pressure Gauge
When the main switch is turned on, the MAIN LED on the doctor
membrane switch panel lights up in green. This state permits
motion of the unit, chair and dental light.. The unit can be moved Lit in green
When the main switch is turned on, all the lamps on the doctor Indicator
- 11 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
2-3. Handpiece E A
When the main switch is turned on, whether handpieces
are accurately fitted in instrument holders or not is checked
automatically. If any handpiece is not accurately fitted in its
instrument holder at this occasion, the subject holder number
(counted as No. 1, 2 and so forth from facing left excluding
syringe) is displayed on the indicator and the buzzer sounds for
a few seconds. Accurately fit the handpiece in the displayed
ge
rin
holder.
Sy
.1
O
.2
N
NO
If the handpiece is accurately fitted while the buzzer sounds
.3
O
.4
N
O
(for a few seconds), the buzzer stops, and the section become
N
operative.
If the buzzer stops in the mid-way of the work, turn off the main
switch, accurately fit the handpiece and turn on the main switch
again about 5 seconds later. Then, use of the section is permitted.
GE
RIN
SY .1
NO .2
NO .3
NO .4
NO
The handpiece is actuated by picking it up from the handpiece holder and operating the foot controller.
See pages 25 to 30 for operation of foot controller, micromotor and electric scaler.
CAUTION
Holder Support Arm / Hndpiece Holder
Do not adjust the holder support arm and handpiece
holder.
Because the angle adjustment of the holder is fixed at Handpiece Holder
Holder Support Arm
the point of installation, the holder support arm will
be damaged in case it is moved too hard.
- 12 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
* With each one of these knobs, the flow rate decreases when
the knob is turned clockwise, and the flow rate increases
when the knob is turned counterclockwise.
4 6 7 8 9
②
Doctor table front arrow view
Increase Decrease
CAUTION
Do not place a load of over 3 kg on the table, or table damage,
malfunction and failure may be caused.
NOTICE: Noise might be produced due to impact at the time of table stop if the table is moved
upward or downward without pressing the brake release button. Make sure to carry out table height
adjustment while the brake release button is pressed.
- 13 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
Lock
NOTICE: Turn off the light when the filmviewer is not used.
Switch
2-11. Panorama size filmviewer (optional) E EA A
The screen light is lit when the switch is turned to the upper
Switch
position, and the screen light is off when the switch is turned to
the lower position.
- The screen brightness increases when the brightness control dial
is turned upward, and the screen brightness decreases when the Brightness
control dial
brightness control dial is turned downward.
NOTICE: Turn off the light when the filmviewer is not used.
- 14 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
Coolant switch
Cooling spray works when the coolant switch is turned to the
ON side and a handpiece is then actuated. Increase or decrease
by the pedal
Chip blow button
depressing extent
Chip blow button Pedal
Chip air is delivered out of the turbine when the chip blow
button is depressed.
Type SE (E Type) E A
Pedal
The pedal depressing extent can control the turbine rotation
speed and air scaler output.
- The motor rotation is activated when the pedal is depressed.
The rotation increases when the pedal is slided to the right, and
decreases when the pedal is slided to the left. Coolant Switch
The electric scaler output also increases when the pedal is
slided to the right, and decreases when the pedal is shifted to
the left. The P-MAX scaler does not permit output control
during operation. Once stop the operation, control the output,
and restart the operation.
Decrease
Coolant switch Chip Blow Button
Depressing the coolant switch turns spray on/off like
Pedal
membrane switch on the table.
Increase
4-mode (Water -> Air -> OFF -> Spray) operation can be
selected for the motor.
- 15 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
16 11 10 12 2 1 7
E A 15 4 3 1
16 2 9 5 6
- 16 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
CAUTION
When operating the chair by the auto return switch, be careful not to nip the stool or the like by the
chair backrest. Damage to the backrest, stool and/or doctor table could be caused. Furthermore, at
the time of motion of the chair, ensure that the patient is sitting in a normal position and keep your
eye on the patient while the chair is in motion.
- 17 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
6 Cupfiller switch EA
E A
Water is supplied through the cupfiller nozzle when this switch is
E A E A
pressed. Simultaneous with start-up of cupfiller motion, water comes
out of the bowl flush nozzle and flushes the cuspidor bowl.
* The cupfiller is timed for 3 to 4 seconds, and press the switch again
to stop in the middle. When the cupfiller switch is pressed, water
is supplied automatically for 3 to 4 seconds irrelevant to the water
volume in the cup. Watch the overflow. Do not press the switch while
the cup is not in place.
- 18 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
- 19 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
13 Function switch E A
Use this switch for setting various working conditions.
The setup function changes in the sequence indicated below each
time when this switch is pressed.
1. Dental timer time setup and motion
2. Doctor number selection
3. Flushout selection and operation
4. Key touch tone selection and change
5. Handpiece light switching-on timing selection
6. Dental timer time-out alarm tone selection
7. Micromotor limit rotation speed selection
8. Micromotor spray mode selection
9. Cupfiller and bowl flush interlocked action selection
* Do not keep this switch pressed.
14 Store switch E A
Use this switch for storing data and so forth:
1. Store micromotor preset mode. See page 45 for the setup procedure.
2. Store preset timer.
3. Store micromotor recurrence mode at the time of power on.
- 20 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
The limit mode is a rotation mode providing an upper limit speed in SET2
the rotation.
SET3
The preset mode (SET1 to SET3) keeps the rotation speed fixed
regardless how the the foot controller is operated.
Pressing the switch each time changes the speed mode:
Limit speed -> SET1 -> SET2 -> SET3 -> Limit speed.
- 21 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
Mode switching
Sliding the foot controller pedal right and left can change the output
of the VARIOS scaler or EMS scaler during operation.
When the pedal slides to the right-hand end, the output becomes the
maximum, and the output is displayed by a bar graph on the indicator
during the operation.
CAUTION
Carefully read the instruction manual and other documents attached to each scaler before use,
and correctly use the scaler
- 22 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
Indicator
Function Switch
The contents of display on the indicator on the doctor membrane switch panel change sequentially
each time when the function switch is pressed to select a desired setup. Contents and sequence of
function switch setup are described below.
= START
2 Selection of doctor number
② Dr NUMBER =0 ↓
BEEP=ON F=↓
④
5 Selection of handpiece light switching-on timing
(-)=OFF (+)=ON
(Example) Case of "with switch tone"
(-)=FOOT (+)=HOLD
(Example) Case of holder mode
7 Selection of number of steps of micromotor limit rotation speed ⑥ ALARM TYPE =0 F=↓
0 1 2 LP (3) PUSH
(-)=3 (+)=5STEP
9 Selection of interlocked action of cupfiller and bowl flush
⑧ SPRAY=4MODE F=↓
⑨ * =LINK F=E
(-)=IND (+)=LINK
- 23 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
. = START 01:10
[Timer cancellation]
This is the procedure for cancellation of the dental timer during Press once
countdown action.
1. Press the function switch once. Press once
2. Press rotation mode select switch .
TIMER SER F=↓
= STOP 01:10
- 24 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
3. Pick up the handpiece to be flushed, and set it over the PUSH FOOT PEDAL
cuspidor bowl or the like.
▲
1 2 3 4 F 00:40
4. Depress the foot controller pedal. Then release the pedal, and ▲
The ID number of the picked up HP is displayed.
▲
the water jets out of the handpiece for 40 seconds .
(HP numbers are 1, 2, 3 and 4 from front left.)
1 2 3 4 F 05:00
minutes for cuspidor bowl flushout.
▲
* To stop flushout action in mid-way, press any switch or The residual time is displayed. (Starts from 5
minutes, and after termination of handpiece
depress the foot controller pedal.
flushout, starts from 5 minutes again.)
pressed.
- 25 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
With or without beep tone that is issued when a switch is BEEP=ON F=↓
pressed can be selected, and the beep tone frequency (tone (-)=OFF (+)=ON
- 26 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
- 27 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
CAUTION
Do not change the rotation while the micromotor is running.
- 28 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
④ ③ ② ①
4 6 7 8 9
⑥
⑤
Increase Decrease
W A
⑦ ⑧ ⑨
NOTICE: Water drips from the cupfiller nozzle when the water in the water heater
OFF ON is heated. This phenomenon is not a sign of failure.
CAUTION
To prevent damage to the water heater caused by heating without water, ensure that the cup is
filled with water before turning on the water heater switch. The heating element will burn out if
the water heater is switched on without water.
- 29 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
OPEN
immediately after the vacuum handpiece was returned to the
holder, but suction will continue for about 3 seconds by the CLOSE
function of delay circuit. Valve
Opening or closing the valve can control the suction flow rate.
opening or closing the valve can control the suction flow rate. Valve
- 30 -
5. CARE AND MAINTANANCE
1. Chair Section
Chair Cleaning
- The surface of the chair’s seating area is made of
synthetic leather. Apply dry wiping or wipe the surface
with cloth moistened with either water or diluted neutral
detergent for the care.
- If the color of clothing or belt remained on the
synthetic leather, wipe it off with cloth moistened with
diluted neutral detergent as soon as possible, to avoid its
penetration caused by plasticizer.
- In case the synthetic leather is wiped with a wet cloth,
fully wipe off the moisture. If it remains, hydrolytic
degradation may be accelerated. Do not use solvent or
bleach.
- Apply dry wiping using a dry and soft cloth to metallic
areas. If any metallic area is wetted, wipe off the
moisture as soon as possible. It will rust otherwise.
- Wipe the resin cover with a wet and soft cloth.
Chair Sterilization
- Use FD333 made by Durr or ethanol as sprayed to a soft
cloth or paper towel for cleaning and sterilization of the
product exterior. Do not operate the product until the
liquid used for sterilization has fully dried up.
CAUTION
Do not place any hard and heavy article or any article having a sharp tip or edge on the chair’s seating
area. The synthetic leather may be damaged otherwise.
For cleaning the resin cover, do not use solvent or detergent containing abrasive agent. In addition, do
not use any chemical that is other than specified for this unit. The resin cover may crack otherwise.
Use a soft cloth moistened with either water or diluted neutral detergent.
Never use detergent of strong acidity or alkali pipe detergent, or materials including metal may be
corroded.
- 31 -
5. CARE AND MAINTANANCE
Membrane Panel
MA
IN
LO
CK
LP
0
1
2
A
W
Doctor Instruments holder
LP
0
1
2
①
A
W
occasions of cleaning.
ドクター側ホルダー
汚物入れ
Oil mist separator (optional) オイルミスト
The oil mist separator (optional) collects the oil contained in
セパレーター
- 32 -
5. CARE AND MAINTANANCE
Handpiece hose
Carefully wipe away the filth from the handpiece hose
using soft cloth moistened with alcohol or the like.
CAUTION
Assure that the nut is positively tightening before use.
- 33 -
5. CARE AND MAINTANANCE
treatment.
CAUTION
Never use emery paper, metallic brush or detergent containing grinding agent for cleaning.
Use of strong acid detergent or alkali pipe detergent may corrode metals. Never use detergent of
such a nature.
CAUTION
Make sure to turn off the main switch before dismounting or mounting of the cuspidor bowl or
cupfiller tray. If the main switch is not turned off at such an occasion, the cupfiller sensor may be
activated by hand or any other article during mounting/dismounting work, water for bowl flush
may jet out, and corrosion to metallic components may arise.
The cuspidor bowl is made of porcelain, and it may be cracked if impact is applied to it.
Furthermore, cleaning of the cuspidor bowl with hot water should be avoided because cracking
may result.
The cuspidor bowl can be easily removed. Be careful not to apply excessive force at occasions
other than cleaning. (Be careful not to apply impact to it, not to drive it and not to allow it to fall.)
- 34 -
5. CARE AND MAINTANANCE
Vaseline
* The saliva ejector handpiece body also permits
disassembly and cleaning by observing the procedure
described above.
The vacuum tip, saliva ejector nozzle, cap and filter can
Vacuum tip
be sterilized with autoclave. Sterilize the body with (optional) Saliva
ejector
autoclave in the assembled state. nozzle
(optional)
Sterilization with autoclave is permitted up to 250 times.
However, sterilization of the valve with autoclave is Cap Cap
place. Body
Valve
Valve
Saliva
ejector
Vacuum hose
hose
Hose connector
- 35 -
5. CARE AND MAINTANANCE
running water.
For cleaning the vacuum hose and saliva ejector hose,
turn off the main switch for the unit. For fitting the Joint
hose, insert the joint section and then twist it clockwise Saliva ejector hose
Vacuum hose
for locking it.
If the hose can be hardly removed or hardly locked, turn
the hose while forcing it.
- 36 -
5. CARE AND MAINTANANCE
Drain Valve
NOTICE: Wipe off water and residual disinfectant immediately. Corrosion and rusting may arise
otherwise.
- 37 -
6. STORAGE, LIFETIME, DISPODAL AND RESTRICTIONS OF USE
Lifetime
The lifetime of this product is ten years after production and shipping.
- 38 -
7. BEFORE ASKING FOR REPAIRS
If any of phenomena described below has occurred, make the following checks before asking for repairs.
The product Air compressor power is not on. Turn on the power.
does not
work at all. Equipment circuit breaker in the clinic
cabinet panel is not on. Turn on the equipment circuit breaker.
Locking device for motion stop is activated. Unlock the device. See page .
The chair does
not work.
Saliva ejector
Ifhandpiece
the unit does not normally
does Filter ofwork even
saliva if actions
ejector were taken
handpiece upon checkup
is filthy. stated
Clean the filter.above, then stop using the
unit,
not turn
make offsuction.
the main switch and contact your dealer or our office.
- 39 -
8. DIMENSION AND SPECIFICATION
Specification
Rated power supply ------------------------------------------------- AC.115V 7.7A/6.3A 50/60Hz
AC.220V 3.5A/3.5A 50/60Hz
AC.230V 3.4A 50Hz
AC.240V 3.6A 50Hz
Fuse ------------------------------------------------------------------- 115V F10A 125V
220 - 240V F5A 250V
Air main pressure --------------------------------------------------- 0.45 to 0.5MPa
Water main pressure ------------------------------------------------ 0.1 to 0.2MPa
Mass ------------------------------------------------------------------ Cuspidor unit section : 74kg
Doctor unit section : 40kg
Chair section : 150kg
Dental light ---------------------------------------------------------- AL-720S / AL-520
AL-820S / 900 Dental Light (Type 920)
Working environment ---------------------------------------------- Temperature : 10 to 40 ℃
Humidity : 30 to 75%
Atmospheric pressure : 700 to 1060hPa
Maximum Load ----------------------------------------------------- 135kg
Classification
Operation mode ------------------------------------------------------ Short-time operating equipment (chair)
Classification of foot controller: ----------------------------------- IPX1(applicable standard IEC60529)
Classification of protection against electric shock -------------- Class I equipment, B type mounting section
( Upholstery and Handpieces )
Equipment that is not suitable for use in air, flammable anesthetic gas, oxygen or nitrous oxide
- 40 -
8. DIMENSION AND SPECIFICATION
Dimension (Numerical values shown here represent standard values.) Unit: mm
Base mount type Dimensional tolerance: ±10%
1710
594
462
398
460
(1939 for 920 light)
584
Max.2456
188
1871
848
790~900
483~748
822
959
594
636
584
(1939 for 920 light)
Max.2456
188
1871
296
848
750~1000
483~748
570
- 41 -
9. MAINTANANCE AND INSPECTION
Guide for daily maintenance and inspection (Maintenance and inspection by user)
Management of maintenance and inspection of medical equipment should be implemented by the user (medical
保 守 点 検
institution). In case the user does not implement such management, it is permitted that such management is
outsourced to a qualified entity such as a medical equipment repair company.
For safe use of this product, it is necessary that inspection should be conducted in the specified frequency on
the items described below.
No. Item Frequency Inspection method Influence if inspection Maintenance required in case
and diagnosis not conducted of nonconformity
1 Check Before ①Chair operation switches Unexpected personal Contact your dealer or our
of safety start shall not work when foot injury and troubles may office if any abnormality arises.
functions controller pedal is arise due to motion of
depressed. the chair during medical
②When cuspidor section is treatment and due to
turned to the patient side, pinching between
cuspidor indicator shall doctor section and
light up in pink, and chair chair.
operation switches shall
not work.
2 Check for Before Leakage of water and air The product will not Contact your dealer or our office
leakage of start shall not be observed normally work, and if any abnormality arises.
water and around the product. troubles may arise.
air
3 Cupfiller Before When a paper cup is placed Cupfilling may not be Execute re-inspection in accor-
start on the cupfiller, the cup executed. dance with "Method for operation"
shall be detected and described in the instruction
cupfilling shall be executed. manual.
* Malfunction may arise if Contact your dealer or our office if
the cup is of another recovery is not achieved as a
material grade (such as result of re-inspection.
stainless steel and plastics)
or if the paper cup is of dark
color or pattern.
4 Check of Before ①Air turbine revolution, Troubles such as injury Control the water flow in accor-
motions of water flow, air flow and so in patient's oral cavity dance with "Control of compo-
start forth shall be free of
equipment and equipment failure nents" described in the instruction
abnormality. may arise. manual. If any other abnormality
②Micromotor revolution, arises, refer to the instruction
water flow and so forth shall
be free of abnormality. manual attached to individual
③Scaler vibration, water equipment.
flow and so forth shall be Contact your dealer or our office if
free of abnormality. recovery is not achieved.
5 Check of For each Appropriate bar shall be The bar will not normally If abnormality such as flaw and
air turbine patient positively mounted. work and troubles may deformation is found on the bar,
bar Make sure to refer to the arise. replace the bar in accordance with
instruction manual attached the instruction manual attached to
to individual equipment. individual equipment.
6 Check of For each Appropriate tip shall be The tip will not normally If the tip was worn or deformed,
scaler tip patient positively mounted and be work and troubles may replace the tip in accordance with
correctly used. arise. the instruction manual attached to
Make sure to refer to the the scaler.
instruction manual attached Contact your dealer or our office if
to the scaler. any other trouble arises.
7 Check of For each The nut for fixing the nut of Troubles may arise if Turn and positively retighten the
tightness of patient Type 77, 3-way syringe shall the nut comes off. nut that fixes the nozzle.
syringe nut be positively tightened.
8 Matters After Excessive handpiece oil or The motor section will Execute care in accordance with
attached to closing the like shall not be attached not work normally and the instruction manual attached to
micromotor to the motor section. troubles may arise. individual micromotor.
- 42 -
保 守 点 検
9. MAINTANANCE AND INSPECTION
No. Item Frequency Inspection method Influence if inspection Maintenance required in case
and diagnosis not conducted of nonconformity
9 Care After Flush the suction line, and Faulty suction may Clean the suction line and filter in
Vacuum closing then clean the filter of the arise. accordance with "Method for
and saliva vacuum or saliva ejector care" described in the instruction
ejector handpiece. manual.
handpiece
10 Care After Clean the cuspidor and Faulty water drainage Execute dust removal and
Cuspidor closing dust filter. may arise. cleaning in accordance with
section "Method for care" described in
the instruction manual.
11 Care After Clean the filter of the solid Vacuum suction will Clean the filter in accordance with
Solid closing collector. become weak. "Method for care" described in the
collector instruction manual.
12 Care After Chemical, filthy water and Discoloration and Execute wiping in accordance
Exterior closing so forth shall not be found deterioration to the with "Method for care" described
(attached or remaining) on exterior, and corrosion in the instruction manual.
the product exterior. and rusting to metallic
components may arise.
13 Check of After The product main switch Product failure and Contact your dealer or our office
main switch closing shall be off, and water/air troubles may arise. if the main switch cannot be
and main main valves shall be turned off or if the main valve
valves closed. cannot be closed.
14 Product's Once No abnormal noise or the The product will not Contact your dealer or our office if
moving every like shall be produced from normally work and any abnormality arises.
parts week product's moving parts troubles may arise.
when the product is
operated.
15 Care Water may enter the air Drain the water from Drain the water from the air filter
Drain line, and equipment failure the air filter drain valve. in accordance with "Method for
valve may arise. care" described in the instruction
manual.
16 Check of Check the water pressure The product will not Contact your dealer or our office
water and air pressure by reading normally work, and if the set pressure is abnormally
pressure pressure gauges in the troubles may arise. high or low.
and air cuspidor section.
pressure Basic set pressure :
Water: 0.1 to 0.2 MPa
Air: 0.45 to 0.5 MPa
17 Check of The table shall be free of Injury caused by falling Contact your dealer or our
conditions inclination, and water shall of goods located on the office if any abnormality arises.
of table not flow on the table. table and other troubles
section may arise.
18 Care The oil level in the oil mist Normal output will not be Discharge the oil in accordance
Oil mist separator shall be lower produced due to inferior with "Method for care" described
separator than the red line. handpiece exhaust. in the instruction manual.
- 43 -