Manual Servicio
Manual Servicio
Table of contents
0-1
MGSS21-9B002 First Edition
1. General
1. Purpose
The purpose of this manual is to provide inspection procedures and replacement procedures of periodical
Replacement parts to be performed by the customer,(The “Customer” i.e. the individual who received the user
maintenance manual to maintain quality, effectiveness and safety of the equipment)
Refer to this manual when performing inspection, replacement of periodical replacement parts.
2. Notes
(Handling of this manual)
This manual contains confidential matters. Please handle it with care and do not copy / reprint, distribute/lend/transfer
to others. Moreover, the content of this manual is subject to change without notice. If needs, contact our representative
to confirm this manual being updated.
(General)
(1) Be sure to read the instruction manual as well as this manual before handling and performing work on the
equipment.
(2) Make sure that work is done in accordance with the instruction manual for this equipment. If there should be a
problem arising from improper handling, correct the cause and instruct proper handling procedures.
(3) Refer to this manual and perform troubleshooting accordingly if there should be an abnormal operation despite
proper handling.
(4)In case of repair / inspection request from clinical department, go over the request thoroughly before starting work
on the equipment.
(5)Contact our representative if the function in problem may not be restored in accordance with this manual.
(6)When returning the equipments for repair which might have been contacted with body fluid of patient,
Disinfection is required for infection prevention. Otherwise, we may request you again to disinfect it before our
repair.
(7)Be sure to make the work record and keep them.
(8)No warranty is given as to practice of inspection and repair in accordance with this manual.
Note that the equipment disassembled for parts replacement is out of our quality assurance.
(9)Do not take any action other than those listed in this manual. In addition, do not remodel / use non-specified
parts.Otherwise performance and safety may be impaired.
1-1
MGSS21-9B002 First Edition
2. Specifications
2-1
MGSS21-9B002 First Edition
Hands Free Bolus dose rate setting range 0.01 to 10.00 (0.01 step)
10.00 to 100.00 (0.10 step)
100.00 to 999.00 (1.00 step)
(In units such as mL and mg/kg etc.)
Hands Free Bolus dose time setting range 1 s to 60 min 00 s (1 s step)
Volume delivered display range 0.00 to 10.00 mL (0.01 mL step)
10.00 to 100.00 mL (0.10 mL step)
100.00 to 9999.00 mL (1.00 mL step)
When the dose rate is set in mass units, the volume is expressed in mass units.
(Volume delivered display range: 0.01 ng to 9999.99 kg)
(It will be 0.00 mL when the power is turned on.)
Flow rate Machine accuracy Within ±1% (flow rate ≥ 1.00 mL/h)
accuracy
Accuracy including Within ±2%
syringe (The hourly precision after one hour has passed since the start of solution
delivery, in the case where water or a physiological saline solution is used with
a TERUMO 50/60 mL syringe made in the Philippines (TERUMO_P) at a
flow rate of 1.00 mL/h or more (ambient temperature: 23±2°C) according to
IEC 60601-2-24:2012 (EN 60601-2-24:2015) standard.)
Within ±3%
(The hourly precision after one hour has passed since the start of infusion at a
constant flow rate of 1.00 mL/h or more according to IEC 60601-2-24:2012
(EN 60601-2-24:2015) standard.)
Occlusion detection pressure 10 to 120kPa (set value)
The occlusion detection pressure can be set in 10 levels.
Purge flow rate Approx. 150 mL/h (when using syringe of 5 mL)
Approx. 300 mL/h (when using syringe of 10 mL)
Approx. 400 mL/h (when using syringe of 20 mL)
Approx. 500 mL/h (when using syringe of 30 mL)
Approx. 1200 mL/h (when using syringe of 50/60 mL)
Alarms Occlusion alarm, Pressure alarm*1, Nearly Empty alarm, Slider Displacement
alarm, Syringe Barrel Detection alarm, Syringe Displacement alarm, Plunger
Displacement alarm, Battery alarm, Re-alarm, Shutdown Notice alarm, Power
Failure alarm, Start Reminder, No Flow Rate alarm, No VTBI alarm*2, Flow
Rate/VTBI Volume Judgment alarm*2, Completion alarm*2, Link Interruption
alarm*1.
*1: TE-SS830 only
*2: Only if the VTBI setting function is enabled (Default: disabled).
2-2
MGSS21-9B002 First Edition
Safety functions Remaining volume detection Sets the Nearly Empty alarm position based on the time before the syringe is
setting function empty.
Occlusion detection pressure Selects the occlusion detection pressure setting values in 10 levels.
select function
Bolus reduction function When an occlusion occurs, the slider is drawn back to reduce internal pressure.
Flow rate range setting The flow rate setting range can be controlled within the “flow rate setting
function A (Soft limits) range” prescribed for each type of syringe.
• When you try to set a value outside the range controlled by this function, a
message asking for confirmation is displayed.
If you confirm the setting, you can set the value which is outside the
controlled range and the display colour of the value changes.
Flow rate range setting The flow rate setting range can be controlled within the “flow rate setting
function B (Hard limits) range” prescribed for each type of a syringe.
• Values outside the range controlled by this function cannot be set.
Keypad lock function Disables switch operations. However, the Start switch, the keypad lock
function disable operation (press and hold the Back/Mute switch), and the
Power switch still function. When an alarm is activated, the keypad lock is
released.
Available Syringe installation guidance Displays a screen that indicates the syringe installation status on the LCD.
Functions function
Syringe brand display Displays the syringe brand and size on the LCD.
function
Syringe type switch function Switching syringes is possible by performing the specified operation after the
power is turned on.
VTBI setting function Sets the VTBI of dosage of drug solution.
VTBI time setting function Sets the VTBI time of dosage of drug solution.
Purge buzzer sound function When the Purge switch is being pressed, the buzzer sounds.
Bolus buzzer sound function During a bolus, the buzzer sounds.
Volume delivered clear Clears the volume delivered display to zero.
function
Standby function By pressing and holding the Back/Mute switch, the pump can be put in
standby mode. In the standby function, the Start Reminder is delayed.
Standby duration time select Changes the time to activate a Start Reminder in the standby mode.
function
Buzzer volume select Changes the buzzer volume.
function
Switch operation buzzer Buzzer sounds when a switch operation is performed.
sound function
Stop transition buzzer sound Buzzer sounds when the Stop switch is pressed during solution delivery.
function
LCD brightness select Changes the LCD brightness.
function
Maintenance timer function Time to next maintenance check can be set. That information is displayed on
the LCD.
Date and time setting Sets the date and time.
function
External communication Allows communication with a network system to output the status of the pump
function (RS-232C) (flow rate, alarms, etc.) via RS-232C.
(TE-SS732 only)
Free message function (TE- Displays a message on the pump via communication.
SS732 and 830 only)
External communication Allows communication with the optional TERUFUSION Communication
function (infrared Rack System to output the status of the pump (flow rate, alarms, etc.) via
communication (IrDA)) infrared.
(TE-SS830 only)
2-3
MGSS21-9B002 First Edition
Available External communication Allows communication with a network system to output the status of the pump
Functions function (wireless LAN) (flow rate, alarms, etc.) via wireless LAN.
(TE-SS830 only)
Nurse call function When an alarm is activated, relay contact operates through communication
(TE-SS732 only) unit.
History function Displays the operation history, etc.
AC power detector function Detects the presence or absence of AC power supply to the pump and displays
that status.
AC power detector buzzer Buzzer sounds when AC power supply is provided to or disconnected from the
sound function pump.
Purge volume upper limit Sets the upper limit of the purge volume.
setting function
Bolus volume upper limit Sets the upper limit of the hands on bolus volume.
setting function
Keep Vein Open Function After activation of the Infusion Completion alarm, solution delivery is
continued at the keep vein open flow rate.
Keep Vein Open flow rate Changes the keep vein open flow rate.
select function
During-start volume Clears the volume delivered to zero during solution delivery.
delivered clear function
During-start flow rate/ Changes the flow rate/dosage rate during solution delivery.
dosage rate change function
Selecting dose mode during Change the dose mode from mL/h to another mode during solution delivery.
the solution delivery function
Set value memory function Saves the current set values (e.g. flow rate, VTBI, dosage and weight) before
the power is turned off.
Alarm melody select function Changes the buzzing pattern (melody) of the alarm buzzer.
Setting dial acceleration Rotating the Dial while pressing the Stop switch accelerates the setting rate to
function set items such as the flow rate.
Hands On Bolus function After the bolus flow rate is set, a bolus is delivered while the Purge switch is
pressed.
Hands Free Bolus function When bolus volume and bolus time are set, the flow rate is calculated and a
bolus is delivered.
Shortcut call function Press multiple keys in combination to directly call the bolus flow rate setting
display.
Night mode function Changes the LCD brightness for a specified time period.
Power OFF buzzer sound Buzzer sounds as the power is turned off.
function
Dose mode select function The dose mode can be selected among mL/h, μg/kg/min, mg/kg/h and drug
library (TE-SS830 only).
Start-up dose mode select The initial dose mode after the power is turned on can be selected from mL/h,
function μg/kg/min, mg/kg/h, or drug library (TE-SS830 only).
Dose mode information When the AC power cable is connected, the version and the profile of the drug
display function library are displayed before the power is turned on.
Menu1 (User mode1) Enable to go directly to the dose mode select display when the menu screen is
shortcut function to dose selected.
mode
2-4
MGSS21-9B002 First Edition
Drug/patient information Displays drug information such as dilution, patient information including
display function patient name, age, gender, weight, height, etc. The dilution and weight can be
changed.
Dilution setting function by Selects how to specify the dilution from either Dilution or Amount/ Solution.
amount and solution
Interval/Intermittent dose The specified dosage rate is delivered at the specified regular intervals.
function (TE-SS830 only)
Multi-step dose function The speed of solution delivery increases or decreases at the specified levels
(TE-SS830 only) (up to 10 levels).
Delayed Start dose function Solution delivery starts after the specified time has elapsed.
(TE-SS830 only)
Link function (Switch Switch function A: Operations of two pumps are linked. When a Completion
function A, B) alarm occurs on the first pump, the second pump starts
(TE-SS830 only) solution delivery.
Switch function B: Interlocks two pumps. Starts solution delivery of the first
pump, and then starts solution delivery of the second
pump at the predetermined time. After solution delivery of
the second pump is started, each pump changes the
solution delivery speed according to respective settings.
The interlock of the pumps finishes upon completion of
the set solution delivery or at alarm timing.
Come and See ME function Automatic call function after a specified time elapses.
(TE-SS830 only)
Volume delivered unit select When the dose rate is set in mass units, the volume delivered unit can be
function changed to mass units.
Micro notation select Switches “μg” to “mcg” to be displayed in the drug library mode, μg/kg/min
function mode or mg/kg/h mode.
Use conditions Temperature: 5 to 40°C
Humidity: 20 to 90%RH (non-condensing)
Atmospheric pressure: 70 to 106kPa (700 to 1060hPa)
Storage conditions* Temperature: -20 to 45°C
(* A state in which the equipment is unpacked and Humidity: 10 to 95%RH (non-condensing)
stored for subsequent use with the main power Atmospheric pressure: 50 to 106kPa (500 to 1060hPa)
supply unplugged.)
Transport conditions Temperature: -20 to 60°C
Humidity: 10 to 95%RH (non-condensing)
Atmospheric pressure: 50 to 106kPa (500 to 1060hPa)
Power supply AC100 to 240V, 50 to 60Hz
Battery: Lithium ion battery
• Continuous use period: Approx. 12 hours (applies when solution is delivered
continuously at 5 mL/h in ambient temperature of
25°C with new fully charged battery)
• Charging time: ≥8 hours (when charged with AC power supply
with the power turned off)
The battery can be charged to approx. 80% in
3hours.
Sub-battery: Nickel-metal-hydride battery (for a Power Failure alarm in case
of both power supplies are lost)
• Charging time: ≥8 hours (when charged with a new battery and AC power
supply in ambient temperature of 25°C when the power is
off)
Power consumption 24VA
Classification Electric shock protection Class I equipment and internally powered equipment, Defibrillation-proof type
CF applied part
Mode of operation Continuous operation
Protection against harmful IP24
ingress of water or particulate
matter
2-5
MGSS21-9B002 First Edition
2-6
MGSS21-9B002 First Edition
2-7
MGSS21-9B002 First Edition
2-8
MGSS21-9B002 First Edition
3. Operating principles
This product is used to send chemical solution inside a syringe continuously at the specified flow rate (mL/h), based on the
syringe barrel pump method. Solution flow control is performed by a microcomputer that processes information and
controls motor rotation signals. The motors push the plunger inside the syringe and send solution through sliders and other
parts.
Block diagram
*1
*3
*1
*2
It is possible to use exchanging the normal battery cover instead either communication unit1 or communication
unit2.
*1:These terminal are not accessible metal parts.
*2:TE-SS830 only
*3:The accessible metal parts is protective earth.
3-1
MGSS21-9B002 First Edition
4. Description of modes
1. Description of icons
The following section describes the names of icons which appear in the document.
4-1
MGSS21-9B002 First Edition
Delayed start mode icon (*1) Advanced dose mode TE-LM830 only
4-2
MGSS21-9B002 First Edition
2. Description of modes
2. 1 Menu 1
* These setting items do not require restart of pump, and become effective after setting.
【Default setting: 6】
4-3
MGSS21-9B002 First Edition
4-4
MGSS21-9B002 First Edition
2.2 Menu 2
* These setting items do not require restart of pump, and become effective after setting.
4-5
MGSS21-9B002 First Edition
4-6
MGSS21-9B002 First Edition
5. Replacement Parts
Note ●Time of replacement may vary, and in some cases other parts may also require replacement, depending on the
frequency and environment of use.
●The service life of this product is 6 years when maintenance and inspection are conducted, and consumables are
replaced as specified(self-certified). This is based on the assumption that the product is used in a standard
manner. Those exceeding the service life will require inspection and repair work including a comprehensive
parts replacement.
●We retain periodical replacement parts and service parts for 8 years after the termination of manufacture of the
product. Some are off-the-shelf items, however, and those which have long been on the market may become
unavailable prior to the expiration of this period due to discontinued production by the parts manufacturer.
The product (or in some cases, on the packaging of the product) bears a label shown on the left. It
indicates that the parts used for the product (e.g. battery, cord) need to be “collected separately” as
classified waste material. Be sure to dispose of them properly in accordance with the disposal rules
of respectable districts.
Note Perform replacement work on a conductive sheet to protect product from damaging.
(Contents)
Replacement of battery 2.1
Replacement of handle unit 2.2
Replacement of dial 2.3
Replacement of Pole clamp 2.4
5-1
MGSS21-9B002 First Edition
Caution
・When you replace the battery, make sure that AC cable is disconnected. Otherwise, it may cause the device trouble or
incorrect indication of remaining battery level.
●Disassembly procedure
1. Remove the screws securing the battery cover on the rear case, and then remove the battery cover.
Securing screw (TE-SS730/ 830***, TE*SS730/ 830N)
Type of screw: Cross-recessed head screw with captive
spring washer and small washer M3-L8
(SUS)
Number of screws: 2
Battery cover
Washer (SUS)
Washer (plastic)
5-2
MGSS21-9B002 First Edition
Communication connector
Battery connector
Fig.5.2.1(3) TE-SS730/ 830***, TE*SS730/ 830N
Caution
・Exercise care when removing TE-SS732***,
TE*SS732N as there is a cable and FG wire
connected to the RS-232C PCB.
・Be careful not to detach the FG wire. If FG wire is
detached, "Continuity check" is necessary.
Battery connector
3. Disconnect sub battery connector and remove the sub battery from casing.
Caution
Sub battery connector The sub-battery connector locks in place.
Protective earth
blade
External metal part
(RS-232C metal part)
Fig.5.2.1(6) Fig.5.2.1(7)
5-3
MGSS21-9B002 First Edition
● Assembly procedure
1. Connect the sub battery connector and replace battery into the casing.
Fig.5.2.1(8)
2. Connect the battery connector and replace battery into the casing.
Caution
• Install the battery with the warning label facing the bottom of the pump.
• Route the battery cable on the upper side of pump.
• There is a cable connecting TE-SS830N to IrDA PCB on battery cover. Make sure that the connector is securely
connected.
• There is a cable connecting TE-SS732N to RS-232C PCB on battery cover. Make sure that the connector is securely
connected.
↓Upper side
Fig.5.2.1(9) TE-SS730/ 830***, TE*SS730/ 830N
↓Upper side
Fig.5.2.1(10) TE-SS732***, TE*SS732N
5-4
MGSS21-9B002 First Edition
Caution
Two types of washers, one made of plastic and one
made of metal (SUS), are used for the part shown in
yellow circle ②. Take care not to lose them.
Cross-recessed head screw with
captive spring washer and small
washer M3-L14 (SUS)
Washer (SUS)
Washer (plastic)
Caution
・Charge the battery after replacing.
・Maintenance must be performed after replacement.
・When you turn the power on after the battery was removed, be sure to connect the AC power cable. Otherwise,
the clock may be reset. If the clock has been reset, perform the clock setting.
5-5
MGSS21-9B002 First Edition
2. Remove the screw and spring washer (①) securing the stopper, and then pull out the handle (②).
Caution
A spring washer is attached. Be careful not to lose it.
Securing screw
Type of screw: Cross-recessed head screw M3-L4
Number of screws: 1
①
②
Fig. 5.2.2(2)
● Assembly procedure
1. Insert the handle (①), and secure it with the spring washer using the screw (②).
Securing screw
Type of screw: Cross-recessed head screw M3-L4
with spring washer
Number of screws: 1
Tightening torque: 0.64~0.68N・m
Screw locking compound: Necessary (head)
②
①
Fig. 5.2.2(3)
5-6
MGSS21-9B002 First Edition
Securing screw
Type of screw: Cross-recessed head screw with captive
spring washer and small washer M3-L8
Fig. 5.2.2(4) (SUS)
Number of screws: 1
Tightening torque: 0.27~0.34N・m
Screw locking compound: Unnecessary
Caution
・Align the groove to the back before doing this.
・Polyslider is attached to shaft. Be careful not to lose the
polyslider.
Groove
Fig. 5.2.3(1)
● Assembly procedure
1. Install the polyslider and dial.
Caution
・Polyslider is attached to shaft.
Polyslider
5-7
MGSS21-9B002 First Edition
Fig.5.2.4(1)
2. Remove the pole clamp without base from the clamp base.
Clamp base
Fig.5.2.4(2)
● Assembly procedure
1. Install the pole clamp without base to the clamp base.
Clamp base
Fig. 5.2.4(3)
2. Secure the pole clamp without base and clamp base using the screws.
Securing screw
Type of screw: Cross-recessed ultra-low head screws
M4-L10(SUS)
Number of screws: 4
Tightening torque: 0.64~0.68N・m
Screw locking compound: Necessary(screw coated)
Fig. 5.2.4(4)
5-8
MGSS21-9B002 First Edition
6. Maintenance
1. Precautions for maintenance and inspection
Caution
● Maintenance or inspection must be performed after disassembly operation.
● The AC power cable must be connected during maintenance and inspection.
Caution
● When performing maintenance and inspection, use designated syringes.
● Use syringes, etc. that are sterilized. Due to an influence on the measurement accuracy, do not re-use them. Use them
only once.
Caution Even if no abnormality is noted in the external appearance and operation of the main unit, stop using the device.
2. Jigs/tools
6-1
MGSS21-9B002 First Edition
3. Inspection procedure
6-2
MGSS21-9B002 First Edition
1) There must be no damage, etc. or defective operation due to adhesion of chemical solution to the main unit and
pole clamps.
1. Main unit
[1] There must be no damage, etc. to the exterior of the main unit.
[2] There must be no damage, looseness, and adhesion of chemical solution to the AC inlet.
[3] There must be no damage and adhesion of chemical solution to the slider part.
[4] The slider must operate smoothly from side to side in clutch manipulation.
[5] The setting dial part must have no adhesion of chemical solution, rotate smoothly, and enable setting of flow
rate while the power is on.
2. Pole clamps
[1] There must be no damage to the clamp part.
2) The power must be on by the internal battery while a syringe is not placed. Also, the following self-checking
operations must be done.
[1] The operation indicator must flash in green and red alternation.
[2] The message “During self-check” must be displayed in the LCD. The back light must flash and the buzzer must
sound.
4) Connect the AC power cable and make sure that the AC connecting icon [ ] and the battery charged
5) When the 5mL, 10mL, 20mL, 30mL, and 50/60mL made by the setting syringe manufacturer are set, each one
must be detected. Also, when the flange of the syringes is set without putting a flange holder (flange detection
6) When the slider is pushed through without setting the syringe, the [Nearly empty alarm] and the nearly empty icon
7) Remove the slider from the plunger while the device operates and make sure that [Plunger displacement alarm]
and the Clutch icon [ ] appear in the LCD, and the buzzer must sound.
8) Set the syringe and make sure that an arbitrary flow rate can be set by setting dial manipulation, and press the
[Start] and [Stop] switches and make sure that the following operations are done.
[1] When the [Start] switch is pressed, the operation indicator must flash in green.
[2] When the [Stop] switch is pressed, flashing must disappear.
6-3
MGSS21-9B002 First Edition
If occlusion detection is checked as periodic inspection, perform the inspection according to the following procedure.
[Inspection method]
[1] Connect the 50 mL syringe and the pressure gauge by using a tube, etc. (about 50 cm).
[2] Hold the clutch and move it until the slider touches the plunger of the syringe.
[3] Turn the power on and set up the flow rate and the set value of occlusion pressure as in the table below.
(Regarding how to change the set value of occlusion pressure, refer to the [section4 Description of modes])
[4] Press the "Start switch" and make the pump function until the occlusion alarm occurs and the operation is
stopped.
[5] Record the occlusion pressure when the occlusion alarm occurs, and make sure that the value is within the range
of occlusion detection pressure values
1) [Inspection method]
[1] Turn the power on, feed about 50mL of water (determine the amount by taking priming into account) to the
50mL syringe, connect the winged intravenous needle at the syringe tip and install it in the main unit.
[2] Press the "Purge switch" to fill the water up to the needle tip and clear the accumulated volume to "0".
[3] Set the flow rate as shown in the table below and insert the tip of the winged intravenous needle into an empty
measuring cup.
[4] Press the "Start switch" and start the stopwatch simultaneously, and press the "Stop switch" after two minutes to
stop the operation.
[5] Calculate the volume from the measured weight of water and the specific gravity (approx. 0.998g/mL at 23°C),
and make sure that the water fed volume is within the specified value.
*After the end of the inspection, return to the setting that recorded before inspection.
6-4
MGSS21-9B002 First Edition
[1] Turn the power on, feed about 50mL of water (amount by taking priming volume into account) to the 50mL
syringe, connect the winged intravenous needle at the syringe on pump.
[2] Press the "Purge switch" to fill the water up to the needle tip and clear the accumulated volume to "0".
[3] Set the flow rate as shown in the table below and insert the tip of the winged intravenous needle into an Graduated
cylinder.
[4] Press the "Start switch" and start the stopwatch simultaneously, and press the "Stop switch" after thirty minutes
to stop the operation.
[5] Make sure that the measured water volume is within the specified value.
*After completion of those inspection, confirm that the pump setting checked prior to inspection are surely configurated..
6-5
MGSS21-9B002 First Edition
4. Cleaning
Caution Before cleaning, be sure to turn the "Power switch" off and disconnect the AC power cable.
Disinfect the device periodically for the prevention of infection. Soak gauze, etc. in any of the disinfectant solutions
shown in the table below, and twist it softly and wipe the main unit softly. Regarding the dilution ratio, follow the
instructions for the chemical agent. After that, wipe again the main unit with gauze twisted softly after being soaked in
lukewarm water, etc.
Caution
● Avoid using a dryer, etc. and using alcoholic disinfectants because the functions and material quality of the
device may be deteriorated.
● Autoclave sterilization cannot be done for this machine.
● Clean the AC inlet, slider part, and clamp part periodically. During cleaning, do not remove the screw in each
part.
● Before using the AC inlet after being cleaned, make sure that it is dry well.
If the pump contacted with chemical solution and got dirty very much, wipe it immediately with gauze, etc. soaked in
water or lukewarm water.
If the inside of the AC inlet got dirty, wipe it with a cotton swab.
Caution
● Do not wipe it with an organic solvent such as alcohol and paint thinner.
● Do not rub with hard or sharp objects.
Ex) Metal tweezers
When the sliding portion of the One Touch Pole Clamp has dirt and an abnormal sound occurs, it cleans as follows.
●Cleaning procedure
1. Pull and hold the lever, pull the pole clamp knob to full.
4. Pull and hold the lever, a few times to side the knob
Clamp portion of the pole clamp from side to side, and adapt lubricant (mineral oil).
6-6
MGSS21-9B002 First Edition
Execution date
Customer name Catalogue number Serial number Operator
(mm/dd/yyyy)
/ /
●Inspection items
Check item Checked
Check before Inspection
- Check the Occlusion pressure setting value and VTBI setting.
Occlusion pressure setting value 1 ・ 2 ・ 3 ・ 4 ・ 5 ・ 6 ・ 7 ・ 8 ・ 9 ・ 10 □ Check
VTBI setting ON ・ OFF
Electrical safety inspection
1 Earth leakage current test
Measure the current that flows from N or L of the AC inlet to the ground (FG terminal) in the
normal status and single fault status. Pass Fail
Normal status (0.5mA or less) [mA]
Single fault status (1.0mA or less) [mA]
2 Touch current
Measure the current that flows from the external metal part of main unit (measurement point:
Handle part) to the ground in the normal status and single fault status. Pass Fail
Normal status (0.1mA or less) [mA]
Single fault status (0.5mA or less) [mA]
Basic performance inspection
1 There must be no abnormalities such as breakages in the main unit and pole clamps Pass Fail
2 The unit must operate by the battery and self-checking must be executed normally. Pass Fail
3 [Battery alarm] must not appear. Pass Fail
4 When the power is on by AC power, the [AC] icon and the [Battery] icon must appear. Pass Fail
[1] 5 mL, 10 mL, 20 mL, 30 mL, and 50/60 mL syringes must be detected accurately.
5 [2] When the flange of the syringe is set without being put in the flange holder, the syringe Pass Fail
detection icon must appear.
When the slider is pressed through without a syringe, the “Nearly empty alarm” and the Nearly
6 Pass Fail
empty icon must appear in the LCD and the buzzer must sound.
When the slider is put off from the plunger while the device operates, the [Plunger displacement
7 Pass Fail
alarm] and the clutch icon must appear in the LCD, and the buzzer must sound.
8 "Start" and "Stop" switches must operate properly. Pass Fail
9 Set the date and time. □ Check
Occlusion detection inspection
●Occlusion pressure measurement
Occlusion pressure setting Specified value Measured value
1 Setting 9 110±25 [kPa] [kPa] Pass Fail
Setting 6 70±15 [kPa] [kPa]
Setting 4 40±15 [kPa] [kPa]
Flow rate accuracy inspection
[Inspection method]Flow rate accuracy must be with the specified value range.
Inspection environment: flow rate 150mL/h, 2-minute measurement
1) Pass Fail
Specified value Measured value
4.85 to 5.15mL (5mL±3%) [mL]
[Graduated cylinder use] Flow rate accuracy must be with the specified value range.
Inspection environment: flow rate 70mL/h, 30minute measurement
2) Pass Fail
Specified value Measured value
34 to 36mL (5mL±3%) [mL]
Check after Inspection
- Confirm that set back to the setting of the before the inspection. □ Check
Overall evaluation (Repair is required if Fail is present for any one of the items) Pass Fail
Memo
6-7
UMSS21-9B002 First Edition
7-1
8. Service parts List
This section shows a table of parts that compose TE-SS730N/ TE-SS732N/ TE-SS830N.
8-1
6 Battery label - 06PP5015 180201 to 1 100 sheets
No. 1 2 3
Appearance
No. 4 5
Appearance
No. 5 6
Appearance
8-2
MGSS21-9B002 First Edition
No. 7 8 -
Appearance
Article name Pole clamp Pole clamp without base Pressure gauge
Parts code 05PP207 05PP210 05LD371
Remarks
8-3
FOR INFORMATION ABOUT TERUMO PRODUCTS
If this product should fail to perform as intended, immediately stop using the product and contact our representative.
Refer to Instruction Manual for contact address of TERUMO.