This document is the maintenance guide for the TERUFUSION Syringe Pump Type SS3. It provides instructions for maintenance and replacement of parts for the pump. The guide includes sections on general information, specifications, operating principles, modes, replacement parts, maintenance, error codes, and service parts. Specification sections list details on compatible syringes, dose modes, flow rate and volume delivered display ranges, and machine accuracy. The maintenance guide is intended to be used by customers performing quality maintenance and repairs on the pump.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (1 vote)
2K views

Manual Servicio

This document is the maintenance guide for the TERUFUSION Syringe Pump Type SS3. It provides instructions for maintenance and replacement of parts for the pump. The guide includes sections on general information, specifications, operating principles, modes, replacement parts, maintenance, error codes, and service parts. Specification sections list details on compatible syringes, dose modes, flow rate and volume delivered display ranges, and machine accuracy. The maintenance guide is intended to be used by customers performing quality maintenance and repairs on the pump.
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 38

MGSS21-9B002 First Edition

TERUFUSION Syringe Pump


Type SS3
Maintenance Guide

Product name TERUFUSION Syringe Pump Type SS3


TE-SS730
Model
TE-SS830
MGSS21-9B002 First Edition

Table of contents

Section No. contents


1. General
2. Specifications
3. Operating principles
4. Description of modes
5. Replacement Parts
6. Maintenance
7. Error Code Table
8. Service Parts List

0-1
MGSS21-9B002 First Edition

1. General

1. Purpose
The purpose of this manual is to provide inspection procedures and replacement procedures of periodical
Replacement parts to be performed by the customer,(The “Customer” i.e. the individual who received the user
maintenance manual to maintain quality, effectiveness and safety of the equipment)
Refer to this manual when performing inspection, replacement of periodical replacement parts.

2. Notes
(Handling of this manual)
This manual contains confidential matters. Please handle it with care and do not copy / reprint, distribute/lend/transfer
to others. Moreover, the content of this manual is subject to change without notice. If needs, contact our representative
to confirm this manual being updated.

(General)
(1) Be sure to read the instruction manual as well as this manual before handling and performing work on the
equipment.
(2) Make sure that work is done in accordance with the instruction manual for this equipment. If there should be a
problem arising from improper handling, correct the cause and instruct proper handling procedures.
(3) Refer to this manual and perform troubleshooting accordingly if there should be an abnormal operation despite
proper handling.
(4)In case of repair / inspection request from clinical department, go over the request thoroughly before starting work
on the equipment.
(5)Contact our representative if the function in problem may not be restored in accordance with this manual.
(6)When returning the equipments for repair which might have been contacted with body fluid of patient,
Disinfection is required for infection prevention. Otherwise, we may request you again to disinfect it before our
repair.
(7)Be sure to make the work record and keep them.
(8)No warranty is given as to practice of inspection and repair in accordance with this manual.
Note that the equipment disassembled for parts replacement is out of our quality assurance.
(9)Do not take any action other than those listed in this manual. In addition, do not remodel / use non-specified
parts.Otherwise performance and safety may be impaired.

(Cautions on performing work)


(1) Do not disassemble equipment or remove electronic parts with the power turned on or AC power cable connected
to the power supply.
(2) Check visually for any burnout, short-circuiting, disconnected connector, and damaged parts, as well as for
abnormal noise and odor (due to burning on electronic parts).
(3) Take precautions for electrostatic discharge failure when handling and working on electronic parts.
(4) Make sure that all settings on the equipment are restored, and clean the equipment and tidy up work area to check
for any loose parts, screws and so forth upon completion of work.

1-1
MGSS21-9B002 First Edition

2. Specifications

Product name TERUFUSION Syringe Pump Type SS3


Model TE-SS730/ TE-SS830
TE-SS730xxx, TE-SS732xxx, TE-SS83xxxx
TE Alphanumeric characters are entered for x in the catalogue number.
Catalogue number
TE*SS730xxx, TE*SS732xxx, TE*SS83xxxx
Other
Alphanumeric characters are entered for x in the catalogue number.
Compatible syringes Sterilized specified syringes for general use and specified pre-filled syringes
Syringe size 5mL, 10mL, 20mL, 30mL, 50/60mL
Syringe brand TERUMO or other specified brands
Dose mode mL/h mode
μg/kg/min mode
mg/kg/h mode
Library mode (TE-SS830 only)
Flow rate setting range • Setting range
0.01 to 150.00 mL/h
Note that the upper limit of the flow rate can be changed in the following
ranges:
0.01 to 150.00 mL/h (when using syringe of 5 mL)
0.01 to 300.00 mL/h (when using syringe of 10, 20, 30 mL)
0.01 to 1200.00 mL/h (when using syringe of 50/60 mL)
• Step
0.01 mL/h step ( 0.01 to 10.00 mL/h)*
0.10 mL/h step ( 10.00 to 100.00 mL/h)*
1.00 mL/h step ( 100.00 to 1200.00 mL/h)*
*: Step can be used when flow rate is set by the Dial.
(It will be 0.00 mL/h when the power is turned on.)
VTBI setting range 0.10 to 100.00 mL (0.10 mL step)
100.00 to 9999.00 mL (1.00 mL step)
When the dose rate is set in mass units, setting in mass units is possible.
(VTBI setting range: 0.01 ng to 9999.99 kg)
“----.--” (Unlimited)
(It will be 0.00 mL when the power is turned on.)
VTBI time setting range 1 min to 99 hours 59 min (1 min step)
“-- --” (Unlimited)
Dose rate setting range 0.01 to 10.00 (0.01 step)
10.00 to 100.00 (0.10 step)
100.00 to 999.00 (1.00 step)
(In units such as μg/kg/min and mg/kg/h etc.)
(It will be 0.00 when the power is turned on.)
Weight setting range 0.1 to 300.0kg (0.1kg step)
Dilution setting range 0.01 to 10.00 (0.01 step)
10.00 to 100.00 (0.10 step)
100.00 to 999.00 (1.00 step)
(In units such as mg/mL etc.)
Amount setting range 0.01 to 10.00 (0.01 step)
10.00 to 100.00 (0.10 step)
100.00 to 999.00 (1.00 step)
(In units such as μg and mg etc.)
Solution setting range 0.01 to 10.00 (0.01 step)
10.00 to 100.00 (0.10 step)
100.00 to 999.00 (1.00 step)
(Unit: mL)
Hands On Bolus flow rate setting range 100.00 mL/h (when using syringe of 5 mL)
100.00 to 300.00 mL/h (when using syringe of 10 mL, 20 mL, 30 mL)
100.00 to 1200.00 mL/h (when using syringe of 50/60 mL)
(100.00 mL/h step)

2-1
MGSS21-9B002 First Edition

Hands Free Bolus dose rate setting range 0.01 to 10.00 (0.01 step)
10.00 to 100.00 (0.10 step)
100.00 to 999.00 (1.00 step)
(In units such as mL and mg/kg etc.)
Hands Free Bolus dose time setting range 1 s to 60 min 00 s (1 s step)
Volume delivered display range 0.00 to 10.00 mL (0.01 mL step)
10.00 to 100.00 mL (0.10 mL step)
100.00 to 9999.00 mL (1.00 mL step)
When the dose rate is set in mass units, the volume is expressed in mass units.
(Volume delivered display range: 0.01 ng to 9999.99 kg)
(It will be 0.00 mL when the power is turned on.)
Flow rate Machine accuracy Within ±1% (flow rate ≥ 1.00 mL/h)
accuracy
Accuracy including Within ±2%
syringe (The hourly precision after one hour has passed since the start of solution
delivery, in the case where water or a physiological saline solution is used with
a TERUMO 50/60 mL syringe made in the Philippines (TERUMO_P) at a
flow rate of 1.00 mL/h or more (ambient temperature: 23±2°C) according to
IEC 60601-2-24:2012 (EN 60601-2-24:2015) standard.)
Within ±3%
(The hourly precision after one hour has passed since the start of infusion at a
constant flow rate of 1.00 mL/h or more according to IEC 60601-2-24:2012
(EN 60601-2-24:2015) standard.)
Occlusion detection pressure 10 to 120kPa (set value)
The occlusion detection pressure can be set in 10 levels.
Purge flow rate Approx. 150 mL/h (when using syringe of 5 mL)
Approx. 300 mL/h (when using syringe of 10 mL)
Approx. 400 mL/h (when using syringe of 20 mL)
Approx. 500 mL/h (when using syringe of 30 mL)
Approx. 1200 mL/h (when using syringe of 50/60 mL)
Alarms Occlusion alarm, Pressure alarm*1, Nearly Empty alarm, Slider Displacement
alarm, Syringe Barrel Detection alarm, Syringe Displacement alarm, Plunger
Displacement alarm, Battery alarm, Re-alarm, Shutdown Notice alarm, Power
Failure alarm, Start Reminder, No Flow Rate alarm, No VTBI alarm*2, Flow
Rate/VTBI Volume Judgment alarm*2, Completion alarm*2, Link Interruption
alarm*1.
*1: TE-SS830 only
*2: Only if the VTBI setting function is enabled (Default: disabled).

2-2
MGSS21-9B002 First Edition

Safety functions Remaining volume detection Sets the Nearly Empty alarm position based on the time before the syringe is
setting function empty.
Occlusion detection pressure Selects the occlusion detection pressure setting values in 10 levels.
select function
Bolus reduction function When an occlusion occurs, the slider is drawn back to reduce internal pressure.
Flow rate range setting The flow rate setting range can be controlled within the “flow rate setting
function A (Soft limits) range” prescribed for each type of syringe.
• When you try to set a value outside the range controlled by this function, a
message asking for confirmation is displayed.
If you confirm the setting, you can set the value which is outside the
controlled range and the display colour of the value changes.
Flow rate range setting The flow rate setting range can be controlled within the “flow rate setting
function B (Hard limits) range” prescribed for each type of a syringe.
• Values outside the range controlled by this function cannot be set.
Keypad lock function Disables switch operations. However, the Start switch, the keypad lock
function disable operation (press and hold the Back/Mute switch), and the
Power switch still function. When an alarm is activated, the keypad lock is
released.
Available Syringe installation guidance Displays a screen that indicates the syringe installation status on the LCD.
Functions function
Syringe brand display Displays the syringe brand and size on the LCD.
function
Syringe type switch function Switching syringes is possible by performing the specified operation after the
power is turned on.
VTBI setting function Sets the VTBI of dosage of drug solution.
VTBI time setting function Sets the VTBI time of dosage of drug solution.
Purge buzzer sound function When the Purge switch is being pressed, the buzzer sounds.
Bolus buzzer sound function During a bolus, the buzzer sounds.
Volume delivered clear Clears the volume delivered display to zero.
function
Standby function By pressing and holding the Back/Mute switch, the pump can be put in
standby mode. In the standby function, the Start Reminder is delayed.
Standby duration time select Changes the time to activate a Start Reminder in the standby mode.
function
Buzzer volume select Changes the buzzer volume.
function
Switch operation buzzer Buzzer sounds when a switch operation is performed.
sound function
Stop transition buzzer sound Buzzer sounds when the Stop switch is pressed during solution delivery.
function
LCD brightness select Changes the LCD brightness.
function
Maintenance timer function Time to next maintenance check can be set. That information is displayed on
the LCD.
Date and time setting Sets the date and time.
function
External communication Allows communication with a network system to output the status of the pump
function (RS-232C) (flow rate, alarms, etc.) via RS-232C.
(TE-SS732 only)
Free message function (TE- Displays a message on the pump via communication.
SS732 and 830 only)
External communication Allows communication with the optional TERUFUSION Communication
function (infrared Rack System to output the status of the pump (flow rate, alarms, etc.) via
communication (IrDA)) infrared.
(TE-SS830 only)

2-3
MGSS21-9B002 First Edition

Available External communication Allows communication with a network system to output the status of the pump
Functions function (wireless LAN) (flow rate, alarms, etc.) via wireless LAN.
(TE-SS830 only)
Nurse call function When an alarm is activated, relay contact operates through communication
(TE-SS732 only) unit.
History function Displays the operation history, etc.
AC power detector function Detects the presence or absence of AC power supply to the pump and displays
that status.
AC power detector buzzer Buzzer sounds when AC power supply is provided to or disconnected from the
sound function pump.
Purge volume upper limit Sets the upper limit of the purge volume.
setting function
Bolus volume upper limit Sets the upper limit of the hands on bolus volume.
setting function
Keep Vein Open Function After activation of the Infusion Completion alarm, solution delivery is
continued at the keep vein open flow rate.
Keep Vein Open flow rate Changes the keep vein open flow rate.
select function
During-start volume Clears the volume delivered to zero during solution delivery.
delivered clear function
During-start flow rate/ Changes the flow rate/dosage rate during solution delivery.
dosage rate change function
Selecting dose mode during Change the dose mode from mL/h to another mode during solution delivery.
the solution delivery function
Set value memory function Saves the current set values (e.g. flow rate, VTBI, dosage and weight) before
the power is turned off.
Alarm melody select function Changes the buzzing pattern (melody) of the alarm buzzer.

Setting dial acceleration Rotating the Dial while pressing the Stop switch accelerates the setting rate to
function set items such as the flow rate.
Hands On Bolus function After the bolus flow rate is set, a bolus is delivered while the Purge switch is
pressed.
Hands Free Bolus function When bolus volume and bolus time are set, the flow rate is calculated and a
bolus is delivered.
Shortcut call function Press multiple keys in combination to directly call the bolus flow rate setting
display.
Night mode function Changes the LCD brightness for a specified time period.
Power OFF buzzer sound Buzzer sounds as the power is turned off.
function
Dose mode select function The dose mode can be selected among mL/h, μg/kg/min, mg/kg/h and drug
library (TE-SS830 only).
Start-up dose mode select The initial dose mode after the power is turned on can be selected from mL/h,
function μg/kg/min, mg/kg/h, or drug library (TE-SS830 only).
Dose mode information When the AC power cable is connected, the version and the profile of the drug
display function library are displayed before the power is turned on.
Menu1 (User mode1) Enable to go directly to the dose mode select display when the menu screen is
shortcut function to dose selected.
mode

2-4
MGSS21-9B002 First Edition

Drug/patient information Displays drug information such as dilution, patient information including
display function patient name, age, gender, weight, height, etc. The dilution and weight can be
changed.
Dilution setting function by Selects how to specify the dilution from either Dilution or Amount/ Solution.
amount and solution
Interval/Intermittent dose The specified dosage rate is delivered at the specified regular intervals.
function (TE-SS830 only)
Multi-step dose function The speed of solution delivery increases or decreases at the specified levels
(TE-SS830 only) (up to 10 levels).
Delayed Start dose function Solution delivery starts after the specified time has elapsed.
(TE-SS830 only)
Link function (Switch Switch function A: Operations of two pumps are linked. When a Completion
function A, B) alarm occurs on the first pump, the second pump starts
(TE-SS830 only) solution delivery.
Switch function B: Interlocks two pumps. Starts solution delivery of the first
pump, and then starts solution delivery of the second
pump at the predetermined time. After solution delivery of
the second pump is started, each pump changes the
solution delivery speed according to respective settings.
The interlock of the pumps finishes upon completion of
the set solution delivery or at alarm timing.
Come and See ME function Automatic call function after a specified time elapses.
(TE-SS830 only)
Volume delivered unit select When the dose rate is set in mass units, the volume delivered unit can be
function changed to mass units.
Micro notation select Switches “μg” to “mcg” to be displayed in the drug library mode, μg/kg/min
function mode or mg/kg/h mode.
Use conditions Temperature: 5 to 40°C
Humidity: 20 to 90%RH (non-condensing)
Atmospheric pressure: 70 to 106kPa (700 to 1060hPa)
Storage conditions* Temperature: -20 to 45°C
(* A state in which the equipment is unpacked and Humidity: 10 to 95%RH (non-condensing)
stored for subsequent use with the main power Atmospheric pressure: 50 to 106kPa (500 to 1060hPa)
supply unplugged.)
Transport conditions Temperature: -20 to 60°C
Humidity: 10 to 95%RH (non-condensing)
Atmospheric pressure: 50 to 106kPa (500 to 1060hPa)
Power supply AC100 to 240V, 50 to 60Hz
Battery: Lithium ion battery
• Continuous use period: Approx. 12 hours (applies when solution is delivered
continuously at 5 mL/h in ambient temperature of
25°C with new fully charged battery)
• Charging time: ≥8 hours (when charged with AC power supply
with the power turned off)
The battery can be charged to approx. 80% in
3hours.
Sub-battery: Nickel-metal-hydride battery (for a Power Failure alarm in case
of both power supplies are lost)
• Charging time: ≥8 hours (when charged with a new battery and AC power
supply in ambient temperature of 25°C when the power is
off)
Power consumption 24VA
Classification Electric shock protection Class I equipment and internally powered equipment, Defibrillation-proof type
CF applied part
Mode of operation Continuous operation
Protection against harmful IP24
ingress of water or particulate
matter

2-5
MGSS21-9B002 First Edition

Wireless LAN Hereby, TERUMO CORPORATION declares that this TERUFUSION


Syringe Pump Type SS3 is in compliance with the essential requirements and
other relevant provisions of Directive 2014/53/EU.
Catalogue number: TE*SS830xxx*
• Transmit frequency or frequency band: 2412 to 2472 MHz
• Type of modulation and frequency characteristics:
• DBPSK (Differential Binary Phase Shift Keying)
• DQPSK (Differential Quadrature Phase Shift Keying)
• DSSS (Direct Sequence Spread Spectrum)
• CCK (Complementary code keying)
• Effective radiated power:
• IEEE802.11b 10.7 dBm
Catalogue number: TE*SS835xxx*
• Transmit frequency or frequency band: 2412 to 2472 MHz
• Type of modulation and frequency characteristics:
• DSSS (Direct Sequence Spread Spectrum)
• OFDM (Orthogonal Frequency-Division Multiplexing)
• Effective radiated power:
• IEEE802.11b 15dBm
• IEEE802.11g 13dBm
• IEEE802.11n 12dBm
* Alphanumeric characters are entered for x in the catalogue number.
Dimensions Without external communication (RS-232C):
381 mm (W)*1 x 120 mm (H)*1 x 112 mm (D)*2
With external communication (RS-232C):
381 mm (W)*1 x 120 mm (H)*1 x 131 mm (D)*2
*1: Excluding protrusions
*2: Excluding protrusions, pole clamp and moving range
Unit Weight Approx. 2.0 kg

2-6
MGSS21-9B002 First Edition

Units available (For TE market) (As of Mar., 2018)


Drug
Catalogue Weight
Model Labeling Plug type Computer interface library Remarks
number mode,
mode
TE-SS730EN1 European plug
English
TE-SS730EN2 UK plug
TE-SS730DE1 German
TE-SS730FR1 French
TE-SS730NL1 Dutch
-
TE-SS730RU1 Russian European plug
TE-SS730ES1 Spanish
TE-SS730PL1 Polish
TE-SS730IT1
Italian
TE-SS730IT3 Italian plug
TE-SS730 -
TE-SS732EN1 European plug
English
TE-SS732EN2 UK plug
TE-SS732DE1 German
TE-SS732FR1 French
TE-SS732NL1 Dutch
RS-232C
TE-SS732RU1 Russian European plug
TE-SS732ES1 Spanish
TE-SS732PL1 Polish
TE-SS732IT1
Italian
TE-SS732IT3 Italian plug
Yes
TE-SS830EN1 English European plug
TE-SS830EN2 English UK plug
TE-SS830DE1 German
TE-SS830FR1 French
Infrared
TE-SS830NL1 Dutch
communication(IrDA),
TE-SS830RU1 Russian European plug
Wireless LAN
TE-SS830ES1 Spanish
TE-SS830PL1 Polish
TE-SS830IT1
Italian
TE-SS830IT3 Italian plug
TE-SS830 Yes
TE-SS835EN1 English European plug
TE-SS835EN2 English UK plug
TE-SS835DE1 German
TE-SS835FR1 French
Infrared
TE-SS835NL1 Dutch
communication(IrDA),
TE-SS835RU1 Russian European plug
Wireless LAN
TE-SS835ES1 Spanish
TE-SS835PL1 Polish
TE-SS835IT1
Italian
TE-SS835IT3 Italian plug

2-7
MGSS21-9B002 First Edition

Units available (For other countries technicians) (As of Apr., 2019)


Drug
Catalogue Computer Weight
Model Plug type library Remarks
number interface mode,
mode
AC Power cable is not
TE*SS730N01 -
included.
TE*SS730N02 UK plug
-
TE*SS730N03 European plug
TE*SS730N05 -
TE*SS730N07 NBR14136 plug
TE-SS730 -
AC Power cable is not
TE*SS732N01 -
included.
TE*SS732N02 UK plug
RS-232C
TE*SS732N03 European plug
TE*SS732N05 -
TE*SS732N07 NBR14136 plug
Yes
AC Power cable is not
TE*SS830N01 -
included.
TE*SS830N02 UK plug
TE*SS830N03 European plug
TE*SS830N05 - Infrared
TE*SS830N07 NBR14136 plug communication
TE-SS830 Yes
(IrDA), Wireless AC Power cable is not
TE*SS835N01 -
LAN included.
TE*SS835N02 UK plug
TE*SS835N03 European plug
TE*SS835N05 -
TE*SS835N07 NBR14136 plug

* Describe the model as follows in this manual:


・TE-SS730xxx/ TE*SS730xx : TE-SS730
・TE-SS732xxx/ TE*SS732xx : TE-SS732N
・TE-SS83xxxx/ TE*SS83xxxx: TE-SS830

2-8
MGSS21-9B002 First Edition

3. Operating principles
This product is used to send chemical solution inside a syringe continuously at the specified flow rate (mL/h), based on the
syringe barrel pump method. Solution flow control is performed by a microcomputer that processes information and
controls motor rotation signals. The motors push the plunger inside the syringe and send solution through sliders and other
parts.

Block diagram

*1
*3

*1
*2

It is possible to use exchanging the normal battery cover instead either communication unit1 or communication
unit2.
*1:These terminal are not accessible metal parts.
*2:TE-SS830 only
*3:The accessible metal parts is protective earth.

3-1
MGSS21-9B002 First Edition

4. Description of modes

1. Description of icons

The following section describes the names of icons which appear in the document.

Display Icon name Item Remarks


Battery icon Status icon

AC icon Status icon

Occlusion icon Status icon

Occlusion icon Status icon

Syringe brand/Size icon Status icon

Keypad lock icon Status icon

Mute icon Status icon

Maintenance icon Status icon

Alarm icon Status

Wireless LAN icon Status icon TE-SS830N only

Communication icon Status icon

Syringe Displacement icon Sensor icon

Nearly Empty icon Sensor icon

Occlusion icon Sensor icon

Slider Displacement icon Sensor icon

Flow rate tab Tab display

Dose rate tab Tab display


Volume delivery clear tab Tab display
Info. tab Tab display
Hands on bolus tab Tab display
Hands free bolus tab Tab display
Startup icon Ready icon
Ready for bolus icon Ready icon
Dose mode Menu tab Menu 1
Occlusion detection pressure Menu tab Menu 1

Keypad lock Menu tab Menu 1

Bolus Menu tab Menu 1


Menu 1
Advanced dose mode Menu tab
TE-SS830N only
Menu 1
Combination dose mode Menu tab
TE-SS830N only

4-1
MGSS21-9B002 First Edition

Display Icon name Item Remarks


History Menu tab Menu 2

Sound volume tab Menu tab Menu 2

Brightness tab Menu tab Menu 2

Date and time tab Menu tab Menu 2


Menu 2
Profile tab Menu tab
TE-SS830N only
Menu 2
Communication environment tab Menu tab
TE-SS830N only
Password tab Menu tab Menu 2
Advanced dose mode
Interval/Intermittent mode icon (*1) TE-LM830 only
icon
Multi-step mode icon (*1) Advanced dose mode TE-LM830 only

Delayed start mode icon (*1) Advanced dose mode TE-LM830 only

VTBI icon VTBI display

VTBI time icon VTBI time display

Volume delivery icon Volume delivered display

Volume delivery clear icon Volume delivered display

Remaining time icon Remaining time display

4-2
MGSS21-9B002 First Edition

2. Description of modes

Operation of each mode is performed by combination of switches.

* Cursor operation method


The cursor operations in this section are performed as follows.
1. Dial operation (up and down movement)
2. Screen switch operation (down movement only)

2. 1 Menu 1

* These setting items do not require restart of pump, and become effective after setting.

Item number Setting item


U1-1 Dose mode select function
U1-2 Occlusion detection pressure select function
U1-3 Keypad lock function
U1-4 Hands on bolus function
U1-5 Hands free bolus function
U1-6 Advanced dose mode (TE-SS830N only)
U1-7 Combination dose mode (TE-SS830N only)

U1-1 Dose mode select function


● How to set
(1) While power is turned on, long press "Display select switch" to display the menu screen on the LCD.
(2) Select "Dose mode tab" with the cursor and press "Confirmation switch".
(3) Use the cursor to select dose mode, and press "Confirmation switch" to set.

Dose mode • mL/h


• µg/kg/min
• mg/kg/h
• mg/kg/h+DIPRIVAN
• Library (TE-SS830N only)

U1-2 Occlusion detection pressure select function


● How to set
(1) While power is turned on, long press "Display select switch" to display the menu screen on the LCD.
(2) Select "Occlusion detection pressure tab" with the cursor and press "Confirmation switch".
(3) Use the cursor to select a setting from [1⇔2⇔3] and press "Confirmation switch" to set.
* Select from [1 to 10] if 10 setting levels have been selected.

【Default setting: 6】

U1-3 Keypad lock function


● How to set
(1) While power is turned on, long press "Display select switch" to display the menu screen on the LCD.
(2) Select "Keypad lock tab" with the cursor and press "Confirmation switch".
(3) Use the cursor to select from [OFF⇔ON] and press "Confirmation switch" to set.
* While key is being locked, key lock can be released by pressing and holding "Back/ mute switch".
* At startup, "keypad lock release operation" and "power" operation can be performed while key lock is active.

4-3
MGSS21-9B002 First Edition

U1-4 Hands on bolus function


● How to set
(1) While power is turned on, long press "Display select switch" to display the menu screen on the LCD.
(2) Select "bolus tab" with the cursor and press "Confirmation switch".
(3) Use the cursor to select "Hands On Bolus" and then press "Confirmation switch".
(4) Use the dial to set the bolus flow rate.
(5) The bolus will only operate while the purge switch is held down.

U1-5 Hands free bolus function


● How to set
(1) While power is turned on, long press "Display select switch" to display the menu screen on the LCD.
(2) Use the cursor to select the bolus tab and press "Confirmation switch".
(3) Use the cursor to select "Hands Free Bolus" and then press "Confirmation switch".
(4) Use the dial to set the bolus VTBI and bolus VTBI time.
(5) Press the "purge switch" to start the bolus.
*: The bolus ends when the VTBI is reached.

【Default setting: Enabled】

U1-6 Advanced dose mode


* You will need to set a library to select the dose mode in advance. Refer to the instruction manual for details.

U1-7 Combination dose mode


* You will need to set a library to select the dose mode in advance. Refer to the instruction manual for details.

4-4
MGSS21-9B002 First Edition

2.2 Menu 2

* These setting items do not require restart of pump, and become effective after setting.

Item number Setting item


U2-1 History display function
U2-2 Sound volume select function
U2-3 Display brightness select function
U2-4 Date and time setting function
U2-5 Profile (TE-SS830N only)
U2-6 Communication environment (TE-SS830N only)

U2-1 History display function


● How to set
(1) While power is turned on, long press "Back/ mute switch + Display select switch" to display the menu screen
on the LCD.
(2) Select the history tab with the cursor and press "Confirmation switch" to enter the history screen.
(3) Use the cursor to select history No. and press "Confirmation switch" to display event content. By pressing
"Back/ mute switch", screen goes back to history list.
* History feature can memorize maximum about 10,000 items of administration setting, operation status, and
alert etc.
* When exceeding about 10,000 items of history, history items are deleted from the oldest ones, then the newest
history is recorded.
* The "AC connection history" operation history item shows AC connections during history registration
regardless of the current AC connection.
* The "remaining battery history" operation history item shows the current remaining battery amount—not the
remaining amount during history registration.
* Refer to the instruction manual for details.

The following history can be confirmed:


1. Operation history (administration setting, operation status and alarm)
2. Drug Lib. history (contents of selected drug Lib.) *1
3. Patient information (contents of patient information specified by communication) *1

*1 Can be read out only on TE-SS830N

4-5
MGSS21-9B002 First Edition

U2-2 Sound volume select function


● How to set
Sound volume can be set for alert, operation, and cable connection/disconnection sounds.
(1) While power is turned on, long press "Back/ mute switch + Display select switch" to display the menu screen
on the LCD.
(2) Select the sound volume tab with the cursor and press "Confirmation switch" to enter the setting screen.
(3) Use the cursor to select sound volume setting items, and press "Confirmation switch".
(4) Select volume, and press "Confirmation switch" to set.
Type Volume setting Default setting
Alarm volume 1(Low) ⇔ 2(Medium) ⇔ 3(High) ⇔ Step Up ⇔ 1(Low) 2(Medium)
0(Very low) ⇔ 1(Low) ⇔ 2(Medium) ⇔ 3(High) ⇔ 0(Very 1(Low)
Key operating volume
low)
0(Very low) ⇔ 1(Low) ⇔ 2(Medium) ⇔ 3(High) ⇔ 0(Very 0(Very low)
Cable insertion alarm
low)

U2-3 Display brightness select function


● How to set
(1) While power is turned on, long press "Back/ mute switch + Display select switch" to display the menu screen
on the LCD.
(2) Select the brightness tab with the cursor and press "Confirmation switch" to display the selection screen.
(3) Use the cursor to select from [1(dark) ~ 5(bright)], and press "Confirmation switch" to set.
【Default setting: 5】

U2-4 Date and time setting function


● How to set
(1) With the Back/Mute switch pressed and held, press and hold the Display select switch for 2 seconds or more.
Turn the Dial or press the Display select switch to select the date and time tab, and press the confirmation
switch.
(2) Turn the Dial to change the date, and press the confirmation switch.
(3) Turn the Dial to change the time, and press the confirmation switch.

U2-5 Profile (TE-SS830N only)


● How to set
(1) While power is turned on, long press "Back/ mute switch + Display select switch" to display the menu screen
on the LCD.
(2) Select the profile tab with the cursor and press "Confirmation switch" to display the selection screen.
* If no library or profile is registered, the profile tab is not displayed.
(3) Select an item from list of profiles that have already been set, and press "Confirmation switch" to set.
* Set profiles will be applied to the dose mode select function library.

U2-6 Communication environment (TE-SS830N only)


● How to set
(1) While power is turned on, long press "Back/ mute switch + Display select switch" to display the menu screen
on the LCD.
(2) Select the communication environment tab with the cursor and press "Confirmation switch" to display the
selection screen.
(3) Use the cursor to select setting items, and press "Confirmation switch".
(4) Use the cursor to select settings, and press "Confirmation switch" to set.

Setting item Contents


Select from the registered communication environment
Communication mode select
list.
Wireless LAN ON ⇔ OFF

4-6
MGSS21-9B002 First Edition

5. Replacement Parts

1. Parts requiring periodical replacement

Periodical replacement parts


Periodical replacement parts are those which deteriorate and wear gradually over time after their initial use, and require
replacement on a regular basis to maintain the accuracy and performance of the equipment. The following parts will
need to be replaced in accordance with the years elapsed since the initial use.

Parts Code Parts Name Elapsed Time Reason for Replacement


Even when charged, after a short period of time
06PT5001 Battery 2.5~3 years “Battery alarm” is displayed and the buzzer
sounds.
The Power Failure alarm does not function
06PT5014 Sub Battery 1~2 years
properly.

Note ●Time of replacement may vary, and in some cases other parts may also require replacement, depending on the
frequency and environment of use.
●The service life of this product is 6 years when maintenance and inspection are conducted, and consumables are
replaced as specified(self-certified). This is based on the assumption that the product is used in a standard
manner. Those exceeding the service life will require inspection and repair work including a comprehensive
parts replacement.
●We retain periodical replacement parts and service parts for 8 years after the termination of manufacture of the
product. Some are off-the-shelf items, however, and those which have long been on the market may become
unavailable prior to the expiration of this period due to discontinued production by the parts manufacturer.

The product (or in some cases, on the packaging of the product) bears a label shown on the left. It
indicates that the parts used for the product (e.g. battery, cord) need to be “collected separately” as
classified waste material. Be sure to dispose of them properly in accordance with the disposal rules
of respectable districts.

2. Replacement procedures for replacement parts

Note Perform replacement work on a conductive sheet to protect product from damaging.

(Contents)
Replacement of battery 2.1
Replacement of handle unit 2.2
Replacement of dial 2.3
Replacement of Pole clamp 2.4

5-1
MGSS21-9B002 First Edition

2.1 Replacement of battery

Caution
・When you replace the battery, make sure that AC cable is disconnected. Otherwise, it may cause the device trouble or
incorrect indication of remaining battery level.

●Disassembly procedure
1. Remove the screws securing the battery cover on the rear case, and then remove the battery cover.
Securing screw (TE-SS730/ 830***, TE*SS730/ 830N)
Type of screw: Cross-recessed head screw with captive
spring washer and small washer M3-L8
(SUS)
Number of screws: 2

Battery cover

Fig.5.2.1(1) TE-SS730/ 830***, TE*SS730/ 830N


Securing screw (TE-SS732***, TE*SS732N)
Type of screw ①: Cross-recessed head screw with
captive spring washer and small
washer M3-L8 (SUS)
Number of screws: 1
② ① Type of screw ②: Cross-recessed head screw with
captive spring washer and small
washer M3-L14 (SUS)
Number of screws: 1
Caution
Two types of washers, one made of plastic and one
made of metal (SUS), are used for the part shown in
yellow circle ②. Take care not to lose them.

Cross-recessed head screw


Fig.5.2.1(2) TE-SS732***, TE*SS732N
with captive spring washer
and small washer M3-L14
(SUS)

Washer (SUS)

Washer (plastic)

5-2
MGSS21-9B002 First Edition

2. Disconnect battery connector and remove the battery from casing.


Caution
Exercise care when removing TE-830***,
TE*SS830N as there is a cable (shown in yellow
circle) connected to the IrDA PCB.

Communication connector

Battery connector
Fig.5.2.1(3) TE-SS730/ 830***, TE*SS730/ 830N

Caution
・Exercise care when removing TE-SS732***,
TE*SS732N as there is a cable and FG wire
connected to the RS-232C PCB.
・Be careful not to detach the FG wire. If FG wire is
detached, "Continuity check" is necessary.

Battery connector

Fig.5.2.1(4) TE-SS732***, TE*SS732N

3. Disconnect sub battery connector and remove the sub battery from casing.
Caution
Sub battery connector The sub-battery connector locks in place.

Release the lock

Fig.5.2.1(5) TE-SS732***, TE*SS732N

<FG wire continuity check procedure>


* After attaching the battery cover, conduct "continuity check".
Check for continuity between the protective earth blade (GND) and any of three external metal parts (RS-232C
metal part).
If it is not conducting, make sure that the FG wire is properly connected.

Protective earth
blade
External metal part
(RS-232C metal part)

Fig.5.2.1(6) Fig.5.2.1(7)

5-3
MGSS21-9B002 First Edition

● Assembly procedure
1. Connect the sub battery connector and replace battery into the casing.

Place the cable in the groove

Fig.5.2.1(8)

2. Connect the battery connector and replace battery into the casing.
Caution
• Install the battery with the warning label facing the bottom of the pump.
• Route the battery cable on the upper side of pump.
• There is a cable connecting TE-SS830N to IrDA PCB on battery cover. Make sure that the connector is securely
connected.
• There is a cable connecting TE-SS732N to RS-232C PCB on battery cover. Make sure that the connector is securely
connected.

↓Upper side
Fig.5.2.1(9) TE-SS730/ 830***, TE*SS730/ 830N

↓Upper side
Fig.5.2.1(10) TE-SS732***, TE*SS732N

5-4
MGSS21-9B002 First Edition

3. Fix battery cover in place using 2 securing screws.


Caution
Be sure that battery cable is placed within the rim to prevent it from getting caught under the cover.
Securing screw (TE-SS730/ 830***, TE*SS730/ 830N)
Type of screw: Cross-recessed head screw with captive
spring washer and small washer M3-L8
(SUS)
Number of screws: 2
Tightening torque: 0.27~0.31N・m
Screw locking compound: Unnecessary

Fig.5.2.1(11) TE-SS730/ 830***, TE*SS730/ 830N


Securing screw (TE-SS732***, TE*SS732N)
Type of screw ①: Cross-recessed head screw with
captive spring washer and small
① washer M3-L8 (SUS)
② Number of screws: 1
Type of screw ②: Cross-recessed head screw with
captive spring washer and small
washer M3-L14 (SUS)
Number of screws: 1
Tightening torque: 0.27~0.34N・m
Fig.5.2.1(12) TE-SS732***, TE*SS732N Screw locking compound: Unnecessary

Caution
Two types of washers, one made of plastic and one
made of metal (SUS), are used for the part shown in
yellow circle ②. Take care not to lose them.
Cross-recessed head screw with
captive spring washer and small
washer M3-L14 (SUS)

Washer (SUS)

Washer (plastic)

Caution
・Charge the battery after replacing.
・Maintenance must be performed after replacement.
・When you turn the power on after the battery was removed, be sure to connect the AC power cable. Otherwise,
the clock may be reset. If the clock has been reset, perform the clock setting.

5-5
MGSS21-9B002 First Edition

2.2 Replacement of handle unit


●Disassembly procedure
1. Remove the screw securing the handle cover, and then remove the handle cover.
Caution
After removing the screw, slide the handle cover sideways
to remove it.

Fig. 5.2.2(1) Securing screw


Type of screw: Cross-recessed head screw with captive
spring washer and small washer M3-L8
(SUS)
Number of screws: 1

2. Remove the screw and spring washer (①) securing the stopper, and then pull out the handle (②).
Caution
A spring washer is attached. Be careful not to lose it.

Securing screw
Type of screw: Cross-recessed head screw M3-L4
Number of screws: 1


Fig. 5.2.2(2)

● Assembly procedure
1. Insert the handle (①), and secure it with the spring washer using the screw (②).

Securing screw
Type of screw: Cross-recessed head screw M3-L4
with spring washer
Number of screws: 1
Tightening torque: 0.64~0.68N・m
Screw locking compound: Necessary (head)



Fig. 5.2.2(3)

2. Put the handle cover and secure it using the screw.

5-6
MGSS21-9B002 First Edition

Caution Slide the handle cover sideways to attach it.

Securing screw
Type of screw: Cross-recessed head screw with captive
spring washer and small washer M3-L8
Fig. 5.2.2(4) (SUS)
Number of screws: 1
Tightening torque: 0.27~0.34N・m
Screw locking compound: Unnecessary

2.3 Replacement of dial


●Disassembly procedure
1. The dial is attached to the main unit by magnetic force. Put your finger into the groove in the dial and move it up and
down to remove the dial.

Caution
・Align the groove to the back before doing this.
・Polyslider is attached to shaft. Be careful not to lose the
polyslider.

Groove

Fig. 5.2.3(1)

● Assembly procedure
1. Install the polyslider and dial.
Caution
・Polyslider is attached to shaft.

Polyslider

・If the shaft needs to be greased, use the Silicone grease


Fig. 5.2.3(2)
(Shin-Etsu Chemical, Model G501).

Apply the grease


thinly.

5-7
MGSS21-9B002 First Edition

2.4 Replacement of Pole clamp


●Disassembly procedure
1. Remove the screws.
Securing screw
Type of screw: Cross-recessed ultra-low head screws
M4-L10(SUS)
Number of screws: 4

Fig.5.2.4(1)

2. Remove the pole clamp without base from the clamp base.

Clamp base

Pole Clamp without base

Fig.5.2.4(2)

● Assembly procedure
1. Install the pole clamp without base to the clamp base.

Clamp base

Pole clamp without base

Fig. 5.2.4(3)

2. Secure the pole clamp without base and clamp base using the screws.
Securing screw
Type of screw: Cross-recessed ultra-low head screws
M4-L10(SUS)
Number of screws: 4
Tightening torque: 0.64~0.68N・m
Screw locking compound: Necessary(screw coated)

Fig. 5.2.4(4)

5-8
MGSS21-9B002 First Edition

6. Maintenance
1. Precautions for maintenance and inspection

Caution
● Maintenance or inspection must be performed after disassembly operation.
● The AC power cable must be connected during maintenance and inspection.

1.1 Syringes to be used for maintenance and inspection

Caution
● When performing maintenance and inspection, use designated syringes.
● Use syringes, etc. that are sterilized. Due to an influence on the measurement accuracy, do not re-use them. Use them
only once.

1.2 If a fall or shock occurred

Caution Even if no abnormality is noted in the external appearance and operation of the main unit, stop using the device.

2. Jigs/tools

Prepare the jigs/tools listed in the table below before inspection.


Name Specification
1) Cup for measuring flow rate Approx. 200mL
2) Stopwatch
3) Terumo syringes Sizes of 5, 10, 20, 30, and 50 mL
4) Water
5) Winged intravenous needle
6) Pressure gauge Pressure range: 0 to 970 kPa (0 to 9.9 kgf/cm2)
7) Electronic balance Measuring power: Approx. 0.01 g
8) Graduated cylinder(50mL) 0.5mL graduated
9) Leakage current tester

6-1
MGSS21-9B002 First Edition

3. Inspection procedure

*Record the following setting before beginning inspection.


・Occlusion pressure setting value
・VTBI setting

3.1 Electrical safety inspection

(1) Leakage current tests


* Set the leakage current tester to the setting corresponding to "IEC 60601-1 (2005) 3rd".
For measurement, follow the instruction manual of the leakage current tester.
*In this test, attach any compatible disposable syringe that has been filled with water, and conduct measurement by
driving at the maximum flow rate.

[1] Earth leakage current test


Use the leakage current tester and connect AC100 to 240V to the main unit. Then measure the leakage current
between the power input terminal (N, L) and protective earth blade (GND) in the normal status and single fault
condition.
● Normal status : 0.5mA or less
● Single fault condition : 1.0mA or less

[2] Touch current test


Use the leakage current tester and connect AC100 to 240V and the internal power supply to the main unit. Then
measure the leakage current between the main unit’s external metal part (handle part) and ground (protective
earth blade [GND]) in the normal status and single fault status.
In the same way, when driven by the battery, the higher value is assumed to be the leakage current value.
● Normal status : 0.1mA or less
● Single fault condition : 0.5mA or less

6-2
MGSS21-9B002 First Edition

3.2 Basic performance inspection

Make sure the followings:

1) There must be no damage, etc. or defective operation due to adhesion of chemical solution to the main unit and
pole clamps.
1. Main unit
[1] There must be no damage, etc. to the exterior of the main unit.
[2] There must be no damage, looseness, and adhesion of chemical solution to the AC inlet.
[3] There must be no damage and adhesion of chemical solution to the slider part.
[4] The slider must operate smoothly from side to side in clutch manipulation.
[5] The setting dial part must have no adhesion of chemical solution, rotate smoothly, and enable setting of flow
rate while the power is on.
2. Pole clamps
[1] There must be no damage to the clamp part.

2) The power must be on by the internal battery while a syringe is not placed. Also, the following self-checking
operations must be done.
[1] The operation indicator must flash in green and red alternation.
[2] The message “During self-check” must be displayed in the LCD. The back light must flash and the buzzer must
sound.

3) [Battery alarm] must not appear.

4) Connect the AC power cable and make sure that the AC connecting icon [ ] and the battery charged

level/remaining level icon appear alternately with [ ] in the LCD.


[1] There are no feelings of looseness at the time of connecting AC power cable.

5) When the 5mL, 10mL, 20mL, 30mL, and 50/60mL made by the setting syringe manufacturer are set, each one
must be detected. Also, when the flange of the syringes is set without putting a flange holder (flange detection

part), the syringe detection icon [ ] must appear.

6) When the slider is pushed through without setting the syringe, the [Nearly empty alarm] and the nearly empty icon

[ ] must appear in the LCD, and the buzzer sounds.

7) Remove the slider from the plunger while the device operates and make sure that [Plunger displacement alarm]

and the Clutch icon [ ] appear in the LCD, and the buzzer must sound.

8) Set the syringe and make sure that an arbitrary flow rate can be set by setting dial manipulation, and press the
[Start] and [Stop] switches and make sure that the following operations are done.
[1] When the [Start] switch is pressed, the operation indicator must flash in green.
[2] When the [Stop] switch is pressed, flashing must disappear.

9) Set the date and time.


[1] Hold down the "Back/Mute" and "Display select" switches together until the menu screen appears. Select the
[Date/time] tab and press the "Confirmation switch".
・The currently set year, month, day, hour and minute must be displayed.

6-3
MGSS21-9B002 First Edition

3.3 Accuracy inspection

(1) Occlusion detection inspection

If occlusion detection is checked as periodic inspection, perform the inspection according to the following procedure.

[Inspection method]
[1] Connect the 50 mL syringe and the pressure gauge by using a tube, etc. (about 50 cm).
[2] Hold the clutch and move it until the slider touches the plunger of the syringe.
[3] Turn the power on and set up the flow rate and the set value of occlusion pressure as in the table below.
(Regarding how to change the set value of occlusion pressure, refer to the [section4 Description of modes])
[4] Press the "Start switch" and make the pump function until the occlusion alarm occurs and the operation is
stopped.
[5] Record the occlusion pressure when the occlusion alarm occurs, and make sure that the value is within the range
of occlusion detection pressure values

* In the same way, perform inspection for other occlusion settings.

Flow rate Set value of Distance from pump to Occlusion detection


setting occlusion pressure pressure gauge pressure value
9 110±25 [kPa]
100mL/h 6 Approx. 50cm 70±15 [kPa]
4 40±15 [kPa]

(2) Inspection for flow rate accuracy

1) [Inspection method]
[1] Turn the power on, feed about 50mL of water (determine the amount by taking priming into account) to the
50mL syringe, connect the winged intravenous needle at the syringe tip and install it in the main unit.
[2] Press the "Purge switch" to fill the water up to the needle tip and clear the accumulated volume to "0".
[3] Set the flow rate as shown in the table below and insert the tip of the winged intravenous needle into an empty
measuring cup.
[4] Press the "Start switch" and start the stopwatch simultaneously, and press the "Stop switch" after two minutes to
stop the operation.
[5] Calculate the volume from the measured weight of water and the specific gravity (approx. 0.998g/mL at 23°C),
and make sure that the water fed volume is within the specified value.

[Volume] mL = [Measured weight] /[Specific gravity]

Flow rate setting Syringe size Measurement time Specified value


150 mL/h 50 mL 2 minutes 5mL±3% (4.85 to 5.15mL)

*After the end of the inspection, return to the setting that recorded before inspection.

6-4
MGSS21-9B002 First Edition

2) [In case of Graduated cylinder use]

[1] Turn the power on, feed about 50mL of water (amount by taking priming volume into account) to the 50mL
syringe, connect the winged intravenous needle at the syringe on pump.
[2] Press the "Purge switch" to fill the water up to the needle tip and clear the accumulated volume to "0".
[3] Set the flow rate as shown in the table below and insert the tip of the winged intravenous needle into an Graduated
cylinder.
[4] Press the "Start switch" and start the stopwatch simultaneously, and press the "Stop switch" after thirty minutes
to stop the operation.
[5] Make sure that the measured water volume is within the specified value.

Flow rate setting Syringe size Measurement time Specified value


70 mL/h 50 mL 30 minutes 35mL±3% (34 to 36mL)

*After completion of those inspection, confirm that the pump setting checked prior to inspection are surely configurated..

6-5
MGSS21-9B002 First Edition

4. Cleaning
Caution Before cleaning, be sure to turn the "Power switch" off and disconnect the AC power cable.

4.1 Available disinfectant solutions

Disinfect the device periodically for the prevention of infection. Soak gauze, etc. in any of the disinfectant solutions
shown in the table below, and twist it softly and wipe the main unit softly. Regarding the dilution ratio, follow the
instructions for the chemical agent. After that, wipe again the main unit with gauze twisted softly after being soaked in
lukewarm water, etc.

Ingredient Name Dilution (e.g.)


Chlorhexidine gluconate 5%
Benzalkonium chloride 10%
Ethanol* 76.9 - 81.4vol%
*: Do not use ethanol to disinfect the following:
•One Touch Pole Clamp (Rubber cover and Sliding sites)

Caution
● Avoid using a dryer, etc. and using alcoholic disinfectants because the functions and material quality of the
device may be deteriorated.
● Autoclave sterilization cannot be done for this machine.
● Clean the AC inlet, slider part, and clamp part periodically. During cleaning, do not remove the screw in each
part.
● Before using the AC inlet after being cleaned, make sure that it is dry well.

4.2 Cleaning of the device

If the pump contacted with chemical solution and got dirty very much, wipe it immediately with gauze, etc. soaked in
water or lukewarm water.

4.3 Cleaning of the AC inlet

If the inside of the AC inlet got dirty, wipe it with a cotton swab.

Caution
● Do not wipe it with an organic solvent such as alcohol and paint thinner.
● Do not rub with hard or sharp objects.
Ex) Metal tweezers

4.4 Cleaning of the One Touch Pole Clamp.

When the sliding portion of the One Touch Pole Clamp has dirt and an abnormal sound occurs, it cleans as follows.

●Cleaning procedure
1. Pull and hold the lever, pull the pole clamp knob to full.

2. Dampen gauze with a lubricant (mineral oil) in small


quantities.

3. Wipe the place surrounded by a dotted line in Fig 6.1


using a gauze prepared in step2.

4. Pull and hold the lever, a few times to side the knob
Clamp portion of the pole clamp from side to side, and adapt lubricant (mineral oil).

Caution Do not spray directly the lubricant (mineral oil).

Fig.6.1 One Touch Pole Clamp

6-6
MGSS21-9B002 First Edition

TE-SS730/830 Maintenance Records

Execution date
Customer name Catalogue number Serial number Operator
(mm/dd/yyyy)
/ /

●Inspection items
Check item Checked
Check before Inspection
- Check the Occlusion pressure setting value and VTBI setting.
Occlusion pressure setting value 1 ・ 2 ・ 3 ・ 4 ・ 5 ・ 6 ・ 7 ・ 8 ・ 9 ・ 10 □ Check
VTBI setting ON ・ OFF
Electrical safety inspection
1 Earth leakage current test
Measure the current that flows from N or L of the AC inlet to the ground (FG terminal) in the
normal status and single fault status. Pass Fail
Normal status (0.5mA or less) [mA]
Single fault status (1.0mA or less) [mA]
2 Touch current
Measure the current that flows from the external metal part of main unit (measurement point:
Handle part) to the ground in the normal status and single fault status. Pass Fail
Normal status (0.1mA or less) [mA]
Single fault status (0.5mA or less) [mA]
Basic performance inspection
1 There must be no abnormalities such as breakages in the main unit and pole clamps Pass Fail
2 The unit must operate by the battery and self-checking must be executed normally. Pass Fail
3 [Battery alarm] must not appear. Pass Fail
4 When the power is on by AC power, the [AC] icon and the [Battery] icon must appear. Pass Fail
[1] 5 mL, 10 mL, 20 mL, 30 mL, and 50/60 mL syringes must be detected accurately.
5 [2] When the flange of the syringe is set without being put in the flange holder, the syringe Pass Fail
detection icon must appear.
When the slider is pressed through without a syringe, the “Nearly empty alarm” and the Nearly
6 Pass Fail
empty icon must appear in the LCD and the buzzer must sound.
When the slider is put off from the plunger while the device operates, the [Plunger displacement
7 Pass Fail
alarm] and the clutch icon must appear in the LCD, and the buzzer must sound.
8 "Start" and "Stop" switches must operate properly. Pass Fail
9 Set the date and time. □ Check
Occlusion detection inspection
●Occlusion pressure measurement
Occlusion pressure setting Specified value Measured value
1 Setting 9 110±25 [kPa] [kPa] Pass Fail
Setting 6 70±15 [kPa] [kPa]
Setting 4 40±15 [kPa] [kPa]
Flow rate accuracy inspection
[Inspection method]Flow rate accuracy must be with the specified value range.
Inspection environment: flow rate 150mL/h, 2-minute measurement
1) Pass Fail
Specified value Measured value
4.85 to 5.15mL (5mL±3%) [mL]
[Graduated cylinder use] Flow rate accuracy must be with the specified value range.
Inspection environment: flow rate 70mL/h, 30minute measurement
2) Pass Fail
Specified value Measured value
34 to 36mL (5mL±3%) [mL]
Check after Inspection
- Confirm that set back to the setting of the before the inspection. □ Check
Overall evaluation (Repair is required if Fail is present for any one of the items) Pass Fail
Memo

Warning If any abnormality is noted, stop using the device immediately.

6-7
UMSS21-9B002 First Edition

7. Error Code Table


Display Error name Detailed item
E1 RAM/register check error
E2 Program ROMSUM error - E2a: program data
- E2b: display data (font, icon, etc.)
E3 Motor rotation error
E4 Peripheral circuit error - E4a: GateArray(FPGA) error
- E4b: GateArray(FPGA) attachment memory
error
E5 Serial flash memory data error - E5a: disposable data
- E5b: base data such as adjustment value
- E5: data other than the above
E8 Switch short error - E8s: stop switch
- E8g: start switch
- E8p: purge switch
- E8e: confirmation switch
- E8m: Display select switch
- E8c: back/ mute switch
E10 Power source error - E10a: battery monitoring IC error
- E10b: reset error before power OFF
- E10c: battery temperature error (≧80℃)
- E10d: internal power supply failure

7-1
8. Service parts List

This section shows a table of parts that compose TE-SS730N/ TE-SS732N/ TE-SS830N.

Service parts revision history


No. Article name Other parts included Q’ty Remarks
No. of revision Parts code Applicable lot
1 Battery - 06PT5001 180201 to 1
2 Sub battery - 06PT5014 180201 to 1
3 Handle unit - 05PP003 180201 to 1
4 Dial - 05PP005 180201 to 1
5 AC power cable - 05XX02P 180201 to 1 American type

- 06LA532B 180201 to 1 UK type

- 06LA532C 180201 to 1 Europe type

8-1
6 Battery label - 06PP5015 180201 to 1 100 sheets

- 06PP5015C 190701 to 1 100 sheets

7 Pole clamp - 05PP207 180201 to 1


8 Pole clamp without base - 05PP210 180201 to 1

●Jig for maintenance


Service parts revision history
No. Article name Other parts included Q’ty Remarks
No. of revision Parts code Applicable lot
- Pressure gauge - 05LD371 180201 to 1
MGSS21-9B002
First Edition
MGSS21-9B002 First Edition

No. 1 2 3
Appearance

Article name Battery Sub battery Handle unit


Parts code 06PT5001 06PT5014 05PP003
Remarks

No. 4 5
Appearance

Article name Dial AC power cable AC power cable


Parts code 05PP005 05XX02P 06LA532B
Remarks American type UK type

No. 5 6
Appearance

Article name AC power cable Battery label


Parts code 06LA532C 06PP5015 06PP5015C
Remarks Europe type 100 sheets
100 sheets
Language: Portuguese

8-2
MGSS21-9B002 First Edition

No. 7 8 -
Appearance

Article name Pole clamp Pole clamp without base Pressure gauge
Parts code 05PP207 05PP210 05LD371
Remarks

8-3
FOR INFORMATION ABOUT TERUMO PRODUCTS

If this product should fail to perform as intended, immediately stop using the product and contact our representative.
Refer to Instruction Manual for contact address of TERUMO.

TERUFUSION Syrige Pump Type SS3 Maintenance Guide


Document No :MGSS21-9B002
Date:Jan.9, 2020 First Edition
TERUMO CORPORATION
After Service Department General Hospital Company

TERUMO CORPORATION
44-1, 2-CHOME, HATAGAYA, SHIBUYA-KU, TOKYO 151-0072, JAPAN