Calibration Procedure: 1-Purpose

This document provides calibration procedures for various measuring instruments and equipment. Key points include: - Measuring tapes must be calibrated annually against a master rule calibrated every two years. Pressure gauges require biannual calibration against a master gauge calibrated annually. - Temperature instruments need annual calibration against a master instrument calibrated every two years. Densitometers require calibration every 90 days against a national standard. - Step wedge films are verified prior to first use and annually. Periodic verification of densitometers and step wedges is also required. - Other equipment like hardness testers and micrometers require annual calibration. Calibration tolerances are defined for each instrument. Positive and negative calibration outcomes are also described.

Uploaded by

Hassan Sleem
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
80 views

Calibration Procedure: 1-Purpose

This document provides calibration procedures for various measuring instruments and equipment. Key points include: - Measuring tapes must be calibrated annually against a master rule calibrated every two years. Pressure gauges require biannual calibration against a master gauge calibrated annually. - Temperature instruments need annual calibration against a master instrument calibrated every two years. Densitometers require calibration every 90 days against a national standard. - Step wedge films are verified prior to first use and annually. Periodic verification of densitometers and step wedges is also required. - Other equipment like hardness testers and micrometers require annual calibration. Calibration tolerances are defined for each instrument. Positive and negative calibration outcomes are also described.

Uploaded by

Hassan Sleem
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 5

CALIBRATION SHEET 2 of 4

QUALITY CONTROL
PROCEDURE AR – TB – CAL001

1- PURPOSE

This Procedure describes the requirements for the calibration of length measures, instruments,
densitometer and step wedge comparison film, and pressure gauges. For reference, see section 10 of
the Quality Control Manual.

2- CALIBRATION

2-1 Measuring Tapes

The Steel Rule 1000 MM master shall be checked by the manufacturer or by external laboratory
every two years and shall be used for calibrating the production tapes. The production tapes shall
be calibrated every year .
The maximum allowed tolerances are :
± 1 mm for master rule.
± 1 mm/m for 3 to 5 meters tape.
± 5 mm for tapes over 5 meters. ,

2-2 Pressure Gauges

The Production Pressure Gauges shall be calibrated using calibrated master gauge each test 6
month’s and is valid for three months. The master gauge shall be calibrated once a year by the
manufacturer or by external laboratory. At least four readings between zero and end of scale and
shall be checked in increasing and decreasing directions .
The maximum allowed tolerances are :
1 % end of scale for master pressure gauge.
2% end of scale for production pressure gauge.

2-3 Temperature Measurement Instruments

The Production Temperature instrument shall be calibrated against master instrument once a year
The Master Temperature Instrument shall be calibrated by the manufacturer or by external
laboratory every two years .
The maximum allowed tolerances are:
± 2 °C for Master Instrument .
±10 °C for digital thermocouples. .
± 2°f for thermometer measuring temperature for impact test.

2-4 Densitometers and Step Wedge Film .

2-4-1 Densitometers. Densitometers shall be calibrated at least every 90 days during use as follows:
(a) A national standard step tablet or a step wedge calibration film, traceable to a national standard
step tablet and having at least 5 steps with neutral densities from at least 1.0 through 4.0, shall
be
used. The step wedge calibration film shall have been verified within the last year by
comparison
with a national standard step tablet.

SHEET 3 of 4
QUALITY CONTROL CALIBRATION
PROCEDURE AR – TB – CAL001

(b) The densitometer manufacture’s step-by-step instruction for the operation of the densitometer
shall
be followed.
(c) The density steps closest to 1.0, 2.0, 3.0, and 4.0 on the national standard step tablet or step
wedge
Calibration film shall be read.
(d) The densitometer is acceptable if the density readings don’t vary by more than ± 0.05 density
units
from the actual density stated on the national standard step tablet or step wedge Calibration
film

2-4-2 step Wedge Calibration Film. step wedge comparison films shall be verified prior to first use,
unless
performed by the manufacturer, as follows:
(a) The density of the steps on a step wedge comparison film shall be verified by a calibrated
densit-
ometer.
(b) The step wedge comparison film is acceptable if the density readings don’t vary by more than ±
0.1 density units from the density stated on the step wedge comparison film

2-4-3 Periodic Verification

(a) Densitometer. Periodic calibration verification checks shall be performed as described in T-


262.1
at the beginning of each shift, after 8 hr. of continuous use, or after change at apertures,
whichever
comes first. The densitometer is acceptable if the density readings are within ± 0.05 of the
calibration readings determined in T-262.1
(b) Step wedge comparison films. Verification checks shall be performed annually per T-262.2.

2-4-4 Documentation

(a) Densitometer calibration readings required by T-262.1 (c) shall be recorded in an appropriate
calibration log
. (b) Periodic verification readings required by T-262.3 do not have to be recorded.

2-5 Portable Hardness Tester


Portable Hardness Tester, shall be calibrated to give equivalent results to those of a standard
machine on a comparison test block of approximately the same hardness as the material to be
tested prior to each use. Test block must be in good condition. Calibration each time before use
according to tolerances stated at standard blocks attached acc. to manufacturer commendations.

2-6 Volt- Amp. Meters


A master Volt-Amp Meter shall be used to calibrate the production Volt-Amp Meter by
comparison readings and shall be calibrated every year by the manufacturer or by external
laboratory .At least four readings shall be checked. The allowed tolerance ± 7%. Of the reading
Production Volt-Amp Meter shall be calibrated every year and the allowed tolerance ± 10%.

2-7 Lifting Power of Yokes


The Magnetizing Force of yokes shall be checked at least once a year, or whenever a yoke has
been damaged . If a yoke has not been in use for a year or more, a check shall be done prior to
first use .Each alternating current electro magnetic yoke shall have a lifting power of at least 10 Ib
( 4.5 Kg ) at the maximum pole spacing that will be used. Each direct current or permanent
magnetic yoke shall have a lifting power of at least 40 Ib (18.1 Kg ) at the maximum pole
spacing that will be used. Each weight shall be weighted with a scale from a reputable
manufacturer and stenciled with the applicable nominal weight prior to first use .A weight need
only be verified again if damaged in manner that could have caused potential loss of material.

QUALITY CONTROL CALIBRATION SHEET 4 of 4

PROCEDURE AR – TB – CAL001

2-8 Calipers

The Calipers shall be calibrated every year using external laboratory / the Standard block Gauges.
The calibration shall be performed with a minimum of three readings, one near zero ,one in the
middle of Caliper range and one near the end.
The maximum allowed tolerance is ± 0.1 mm.

2-9 Micrometers
The micrometers shall be calibrated every year using external laboratory / the Standard block
Gauges. The calibration shall be performed with a minimum of three readings, one near zero ,one
in
the middle of Caliper range and one near the end.
The maximum allowed tolerance is ± 0.1 mm.

2-10 Any equipment or instrument other than above ( such as alloy analyzer ,
magnetomet......etc,) Shall be calibrated as recommended by manufacturer instruction manual
and the values haven't undergone a significant change with original calibration values .

3- Calibration with Positive Outcome


The Calibration Eng. certifies that the calibration was satisfactorily carried out by updating the list
of equipment that can be used in production and applying a label on each calibrating equipment
.Calibrated equipment only shall be used in production .The proper functioning of all instrument
shall be go over ( brush upon) every year.

4- Calibration with Negative Outcome


Calibration reports are valid for the interval calibration unless the instrument is suspected of being
any damaged or at any time that there is reason to believe that it is in error. Whenever a unit or
instrument or equipment is found out of calibration , the Calibration Eng. shall be responsible for :

A- Segregating the unit in an area identified as area of "OUT OF CALIBRATION” equipment.


B- Issuing a Non-Conformity report for the unit in object which shall be submitted to the Q.C.
Manger for disposition

C -It is Q.C Manger responsibility to resolve the non-conformity as described in the Q.C.
Manual, Sec. 6.

5- CERTIFICATION
1- The certification shall include the following data:

A- Equipment description and Serial Number.


And date.

B- Description and serial number of the standard used for calibration or calibrated against
nationally Traceable Standards ( NIS, NATIONAL INSTITUTE FOR STANDARD )
Egypt.
C- Date of calibration and operator signature.
D- Date of calibration expire.

2- All the calibration of units listed in this procedure shall be carried out by the equipment manufacturers
or by external laboratory. in this case a certificate of calibration shall be issued by the calibrating
laboratory and shall be accepted by the Q.C Manger by signing and dating such certificates.
All Master Equipment shall be with calibration certificate accepted by Q.C Manger by signature.

SHEET 1 of 4
WELDED STEEL
QUALITY CONTROL
PIPE PILES FAB. AR – TB – EDKU - 001

PROCEDURE
WELDED STEEL PIPE
PILES FABRICATION
PROCEDURE

Project:- Egypt LNG Project - EDKU

Client:- CONEX CANTRACTING CO.

REV. DATE Q.C ENG. DATE Q.C MGR. DATE Client

21/4/2003 Adel Hamza 21/4/2003 M. Fouad