0% found this document useful (0 votes)
665 views29 pages

Deckma OMD2008 15ppm Bilge Alarm

The document is an instruction manual for a 15ppm Bilge Alarm made by DECKMA HAMBURG GmbH. It contains information about the principle of operation, specifications, construction, installation, wiring, power supply, commissioning, operation, maintenance, fault finding, calibration, spare parts, and remarks. The manual provides guidance to users on set up and proper use of the bilge alarm unit.

Uploaded by

Luke Wilson
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
665 views29 pages

Deckma OMD2008 15ppm Bilge Alarm

The document is an instruction manual for a 15ppm Bilge Alarm made by DECKMA HAMBURG GmbH. It contains information about the principle of operation, specifications, construction, installation, wiring, power supply, commissioning, operation, maintenance, fault finding, calibration, spare parts, and remarks. The manual provides guidance to users on set up and proper use of the bilge alarm unit.

Uploaded by

Luke Wilson
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 29

INSTRUCTION MANUAL

15ppm Bilge Alarm

OMD-2008
Versions with
Electric Valve (EV)
Flow Control (FC)
Manual Cleaning Unit (MCU)

DECKMA HAMBURG GmbH


Kieler Straße 316, D-22525 Hamburg - Germany
Tel.: +49 (0) 40 54 88 76-0, Fax: +49 (0) 40 54 88 76-10
Internet: www.deckma.com eMail: [email protected]

MOTEC
Marine & Offshore Technology

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

IMPORTANT NOTICE
Replacement components for 15ppm Bilge Alarms.
General
All monitors in our range are inspected and tested to the related I.M.O. requirements at
our factories prior to delivery.
In normal use the units should operate correctly and without fault over a long period of
time requiring only small amounts of maintenance to be carried out as outlined in the
instruction manuals.

Service Exchange Units


In the event of a monitor malfunction due to electrical or electronic component failure it
is our recommendation that a service exchange unit be ordered.
The defective instrument should be returned to our works within 30 days of supplying
the service exchange unit, then only the repair charge is payable. Otherwise the whole
cost of a service exchange unit becomes payable.
This procedure is by far the easiest and most cost effective way of ensuring the monitor
on board conforms to I.M.O. resolution MEPC.107(49).

Remark:
According the MEPC.107(49) § 4.2.11 the unit has to be checked at IOPP Certificate
renewal survey by the manufacturer or persons authorized by the manufacturer.
Alternatively the unit may be replaced by a calibrated 15 ppm Bilge Alarm.
The OMD-2008 EVFC is designed in that way, that only the measuring cell needs to be
changed, as this unit carries the Calibration Certificate. The Calibration Certificate with
the date of the last calibration check should be retained onboard for inspection
purposes.

If for some reasons the computer unit needs to be changed, it has to make sure, that
the memory card will remain on board for at least 18 month. The new computer unit will
carry its own memory card. The old card can be insert into the new unit only for reading.
Writing is only possible with the card delivered with the new computer unit. For details
see section 13.1.

Warranty
Our warranty terms are12 months after installation but maximum 18 months after
delivery ex works. The maker undertakes to remedy any defect resulting from faulty
materials of workmanship except wearing parts.
The maker's obligation is limited to the repairs or replacement of such defective parts by
his own plant or one of his authorized service stations.
The purchaser shall bear the cost and risk of transport of defective parts and repaired
parts supplied in replacement of such defective parts.

ANY DISMANTLING OR BREAKING OF A SEAL WILL VOID THE WARRANTY


Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 1 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

SECTION TITLE PAGE


1 Introduction 3
2 Important Notes 3
3 Principle of Operation 4
3.1 Measuring Principle 4
3.2 Features 4
3.3 Adjustment 4
3.4 Displays and Alarms 4
3.5 FC (Flow Control) 5
3.6 EV (Electric clean water Valve) 5
3.7 MCU (Manual Cleaning Unit) 5
4 Specification 6
5 Construction 7
6 Installation 8
7 Piping 10
8 Wiring 11
8.1 Typical Control System 13
9 Power Supply 13
10 Commissioning 14
10.1 Electrical and Piping 14
10.2 Flow Rate Adjustment 14
10.3 Functional Tests 15
10.4 Programming Mode 16
11 Operating Instructions 20
11.1 Operator Notes 20
12 Operator Maintenance 22
12.1 Manual Cell Clean Unit 21
Important Information: Cleaning of Glass Tube 23
13 Fault Finding 24
13.1 Memory Card 25
14 Calibration 25
14.1 Calibration and Repeatability Check 26
14.2 Function Test at Classification Survey 26
and Port State Control
15 Spare Parts 27
15.1 Recommended On Board Spares 27
16 Remarks 28

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 2 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

1 Introduction
The OMD-2008 EV-FC Bilge Alarm Unit has been designed specifically for use in
conjunction with 15 ppm oil-water separator units and has a specification and
performance which exceeds the requirements of the International Maritime
Organization specifications for 15ppm Bilge Alarms contained in Resolution
MEPC. 107 (49).
The unit is supplied with 2 works-adjusted alarms at 15 ppm. Other set points
(10 ppm or 5 ppm) are possible and can be adjusted on site at any time by using
the buttons at the front panel.
Instruments with maximum Alarm set-points of 10 ppm or 5 ppm respectively are
available.
If an alarm set point is exceed, the alarms are visible at the front panel and the
appropriate relays are switched. In case of malfunction the System LED at the
front panel will change from blinking green to permanent red.
For the data logging function the unit requires a status input from the separator.
A 0(4) - 20 mA (equal to 0 - 30 ppm) signal output is available for driving a
recorder or external meter.
The OMD-2008 EV-FC is be equipped with an electric switchover valve for clean
water usage and a Flow Sensor. A Manual Cleaning Unit is available as an
option (MCU). Different instrument versions are covered in this document. Please
verify instrument version on site and refer to the relevant sections of this manual.

2 Important Notes
a) This equipment must be installed and operated in strict accordance with the
instructions contained in this manual. Failure to do so will impair the protection
provided.
b) Installation and servicing must be undertaken by a competent and suitable
skilled person.
c) The equipment must be connected to the ground according relevant
requirements.
d) The unit must be isolated from the electrical supply before any maintenance of
the equipment is attempted.
e) All National or local codes of practice or regulations must be observed and,
where applicable, are deemed to take precedence over any directive or
information contained in this manual.
f) In case of freezing conditions the measuring cell should be emptied
completely.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 3 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

3 Principle of operation

3.1 Measuring Principle


An optical sensor array measures a combination of light scattered and absorbed
by oil droplets in the sample stream. The sensor signals are processed by a
microprocessor to produce linearized output.
If an alarm (works set point 15 ppm, or 10 ppm or 5 ppm respectively) occurs, the
two oil alarm relays are activated after the adjusted time delay.
The microprocessor continuously monitors the condition of the sensor
components and associated electronics to ensure that calibration accuracy is
maintained over time and extremes of environmental conditions.

3.2 Features
 Robust construction
 Solid suppression capability
 Low maintenance
 Easy installation
 Constant readiness
 Low spare part stock holding
 Works adjustment
 Easy settings via menu

3.3 Adjustment
The unit is delivered with a works calibration according the IMO-requirements.
The alarm points are set to 15 ppm (or 10 ppm, or 5 ppm for certain versions).
The "Zero" point is also works calibrated and can be re-adjusted on site by using
the programming mode and clean water. See Section 10.4 “Settings-Offset”. A
calibration is not permitted. This has to be done according IMO Regulations by
the manufacturer or persons authorized by the manufacturer.

3.4 Displays and Alarms


In the unit there are two independent oil alarm circuits available. Both can be set
separately from 1 to 15 ppm. From the manufacturing both alarms are set to
15 ppm (according IMO). The set points can be changed according to the
requirements on site, for example to 10 ppm or 5 ppm. An alarm point setting
above 15 ppm is not possible. The adjustment can be done in the programming
mode as described in Section 10.4.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 4 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

In this mode also the individual adjustment of the time delays for the alarms can
be done.
Both alarm circuits are also related to an alarm LED on the front panel.
In case of malfunction the “System” LED will indicate any type of internal fault of
the unit. This LED is flashing green in normal conditions and is red in alarm
conditions.
Additional to the alarm LEDs each alarm circuit is equipped with a relay with
potential free alarm contacts. These contacts can be used for external processing
of the signal or for control of further functions.
If a malfunction or failure of the power supply occurs, all three relays (both alarm
relays and the SYSTEMFAULT relay) will switch to alarm condition.

3.5 FC (Flow Control)


OMD-2008 EV-FC instruments are equipped with a Flow Sensor. The Flow
Sensor is connected in the Drain line of the Measuring Cell. If the flow rate
through the Measuring Cell is too low, or if the sample is not flowing at all, the
Instrument will go to Alarm condition and issue a “Status: Flow?” message.

3.6 EV (Electric clean water Valve)


OMD-2008 EV instruments are equipped with an electric switchover valve for the
clean water stream. The valve can be controlled via the menues, or remotely
using the control input.

3.7 MCU (Manual Cleaning Unit)


Optionally the instruments can be fitted with a Manual Cleaning Unit. The Manual
Cleaning Unit allows to clean the sample glass tube without opening the
Measuring Cell Head Screw, and without interrupting the normal sample flow.
Maintenance is made easier with the MCU. Please note that operating the MCU
may set the instrument to alarm condition for a few seconds.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 5 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

4 Specification OMD-2008 EV FC

Range: 0 – 30 ppm, Trend indication 50ppm


Accuracy According IMO MEPC. 107(49)
Linearity Up to 30 ppm better than ± 2 %
Display Yellow Graphic Display
Power Supply: 24 V - 240V AC or DC
Automatic Voltage selection
Consumption: < 10 VA
Alarm Points 1 + 2: Adjustable between 1 - 15 ppm
(Works adjustment 15 ppm)
Alarm 1 Operating Delay: Adjustable between 1 – 540 sec.
(for annunciation purpose) (Works adjustment 2 sec)
Alarm 2 Operating Delay: Adjustable between 1 – 10 sec.
(for control purposes) (Works adjustment 10 sec)
System Fault Alarm: Red LED
Alarm Contact Rating: Potential free 1 pole change over contacts,
3 A / 240 V
Alarm Indication: Red LEDs
Output Signal: 0 – 20 mA or 4 – 20 selectable
ext. Load < 150 
Sample Water Pressure: 0,1 – 10 bar
Sample Flow: Approx. 0,1 - 3 l/min depending on pressure
Ambient Temperature: + 1 to + 55° C
Sample Water Temperature: + 1 to + 65° C
Roll: Up to 45°
Size (Computer Unit) 200 mm W x 200 mm H x 100 mm D
Size (EV-FC-Arrangement) 150 mm W x 290 mm H x 140 mm D
Degree of Protection: IP 65
Weight: 5,9 kg
Pipe Connections: for 6mm OD Cu or SS pipe
Technical specifications are subject to change without notification

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 6 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

5 Construction
OMD-2008 EV-FC consists of two main components:
The computer unit contains the display PCB with the data logger and the
terminals for external connections.
The measuring cell is built out of an all-aluminium, nickel plated body with inlet
and outlet block from stainless steel. It contains optics and electronics and is
connected with the computer unit via a plugged data cable. It is mounted onto a
stainless steel support that also holds the valve assembly and the Flow Sensor.
An electric valve controls sample water and clean water usage.
The OMD-2008 EV-FC can easily be mounted in wall or bulkhead installation.

1 Computer Unit 4 Flow Sensor


2 Measuring Cell 5 Clean Water Valve EV
3 Manual Cleaning Unit
(MCU) (Option)

Fig. 1

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 7 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

6 INSTALLATION (REFER TO FIG. 2 AND FIG. 3)


See Section 2 for important notes concerning installation.
The OMD-2008 EV-FC Monitor should be located as close as possible to the oily
water separator to minimize response delays. According MEPC.107(49) the
layout of the installation should be arranged so that the overall response time
(including the response time of the 15 ppm Bilge Alarm, which is less than 5 s.)
between an effluent discharge from the 15 ppm Bilge Separator exceeding 15
ppm, and the operation of the Automatic Stopping Device preventing overboard
discharge, should be as short as possible and in any case not more than 20 s.
Mount the OMD-2008 EV-FC Computer Unit by means of M6 or M8 screws on to
a rigid vertical surface and preferably with the display panel of the monitor at eye
level. For service and maintenance sufficient space to all sides should be
available.

Fig. 2a: Dimensions of Computer Unit


Computer Unit and Measuring Cell assembly should be mounted close to each
other.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 8 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

Fig. 2b Dimensions of Measuring Cell assembly


Pipe connections are prepared for 6mm OD stainless steel or coper pipes.
Adapter fittings for other pipe diameters are available upon request. Working
space for maintenance and operation is recommended in front of and above the
instrument.
Care must be taken at mounting of the pipes connections to avoid any torsion of
the housing and damage of the instrument.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 9 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

7 Piping (REFER TO FIG. 3)


Connect the OMD-2008 EV-FC Monitor to the sample point of the oily-water
separator outlet and to a source of oil free water. Fittings on the Instrument are
for 6 mm OD copper or stainless steel pipe. Adapter fittings for other pipe
diameters are available upon request.
The sample point should be located on a vertical section of the separator outflow
piping to minimize the effects of any entrained air. The tapping point should be at
a level above the outlet of the monitor to ensure the sample cell is flooded at all
times.
If connection to a vertical section of the separator outlet piping is impractical, the
tapping may be made into the side of the horizontal pipe. Avoid top or bottom
entry.
For separator discharge pipes up to 75 mm OD a standard "T"-type junction of
the welded or screwed type is satisfactory for the tapping point. For the separator
discharge pipes of 80 mm OD and above a sample probe should be employed
which protrudes into the discharge piping by approx. 25 % of the ID of the pipe.

Fig. 3

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 10 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

8 Wiring (REFER TO FIG. 4 + 5)


See Section 2 for important notes concerning wiring.
This unit must be connected to the mains supply via a suitable rated and
approved fused isolator unless such fusing / isolation is provided by associated
equipment. When fitted, the isolator should be close, readily accessible and
marked as to function.
Electrical connections are made through the metric cable gland openings
prepared underneath the instrument. Not all openings may be free.

Fig. 4

Precise wiring details will vary dependent upon the control system to be
employed but the most frequently used systems employ alarm relay 1 for alarm
only and alarm relay 2 for control purposes.
Electrical connections are made to the terminal blocks inside the computer
housing. Wires are connected to the terminals by pushing a suitable screwdriver
into the clamp holes to release the internal spring loaded clamps. After the wire is
inserted to the terminal and the screwdriver is removed, the wire is fixed.
If the instrument is operated at high voltages, additional care has to be taken to
provide reliable ground connections. Ground (PE) can be connected direct to the
terminal or, if this is not sufficient according local rules, to the computer housing
(PE bolt inside, right side).
The instrument provides a pilot voltage output at Terminals 3&4. This is internally
connected to the power supply input (Terminals 1&2 via Fuse T2A). The pilot
voltage can be used to supply additional external circuitry, e.g. alarm lamps or
electrical valves.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 11 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

Please note: any device connected to the pilot voltage output must be rated for
the voltage the instrument is supplied with. Do not use the pilot voltage for driving
motors, heaters or other high load devices. The pilot voltage is intended for alarm
purposes only.

Fig. 5: Connection Example


Close front cover completely after electrical installation. Water inside the
instrument may result in corrosion and malfunction. Alarm contacts description is
in alarm (non-energized) condition.
Note: In case the OMD-2008 EV-FC is installed as replacement to a
MEPC.60(33) approved separator and no “Status Separator” signal is available, a
link across terminal 11 + 12 can be installed. Alternatively a “Status Separator”
signal can be generated from the Separator control, and an auxiliary relay be
operated with this signal. The Contacts of this relay then can operate the
“STATUS” input. Please refer to the Separator manual, or contact the separator
maker for information regarding the separator.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 12 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

8.1 Typical Control System


The installation on site has to make sure that in case of any loss of power supply
and/or loss of air supply for the automatic stopping device the overboard
discharge valve close the overboard line and open the re-circulating line.
The system showed in the example, employs alarm relay 2 to control a
pneumatic solenoid valve which energises or de-energises a pneumatically
operated 3 - way valve as depicted in Fig. 5.
The separation process will continue until such time as the pollution level falls
below the alarm set point at which time the discharge will be directed overboard.
A pump stop system is according MEPC.107 (49) not allowed.

8.2 Remote EV Electric Valve control


It is possible to remotely control the EV valve. Connect a switch to Terminals
19&20. If the Terminals are connected (linked trough the switch) the EV valve will
switch to clean water. If the Terminals are open/disconnected, the EV valve will
switch to sample stream.

9 Power Supply
See Section 2 for important notes.
The unit is designed for a power supply of 24 V to 240 V AC or DC. It has an
automatic voltage selection. The power supply must have a fuse rated no more
then 2A.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 13 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

10 Commissioning
See Section 2 for important notes.
On completion of the installation, wiring and piping carry out the following checks:

10.1 Electrical and Piping


a) Check that the power supply is connected to the terminals 1 + 2 of the
terminal block.
b) Check that the wiring of the automatic stopping device (back-to-bilge valve)
and to the alarm system is according the IMO Requirements.
c) Check that the grounding (PE) has been made according to the relevant
regulations.
d) Check all piping connections for leaks and rectify as appropriate

10.2 Flow rate adjustment


a) Run oil free water through the instrument to purge the system. Measure the
amount of water flowing through the instrument per Minute. Flow rate for both
smaple stream and clean water stream should be approx. 2 Liters per Minute.
b) On OMD-2008 EV-FC instruments adjust the flow rate through the unit by
using the small O-Rings in the cell cap.

LOW FLOW MEDIUM FLOW HIGH FLOW

O-Ring 4.5x2 inside O-Ring 8.5x2 outside Main O-Ring 11.5x3 always present

NB: The flow rate should be checked on both, the clean water supply and the separator sample
supply. If the clean water supply is obtained from a high pressure source, the flow rate will be
higher than from the sample point.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 14 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

The flow rate adjustment is important for the time delay between the sample
point and the monitor. If the sample flow rate is too small, the instrument will
be in Alarm and display “Status: Flow”.
On instruments fitted with Manual Cleaning Unit
(OMD-2008 EV-FC MCU), sample a higher or a
lower flow rate can be adjusted. To get the
higher flow rate, remove the smaller O-Ring
(9.5x2) from the Manual Cleaning Unit.

MCU
(Manual Cleaning Unit)

10.3 Functional Tests


a) Run oil free water through the instrument to purge the system
b) Verify that the water is flowing with the proper flow rate trough the instrument.
c) Switch on the instrument and make sure, that the Power LED is illuminated
and the display is
showing the
initializing display for
about 15 sec. After
that time it will
change to the
standard display, showing the actual measurement.
d) Switch the Electric Valve to clean water. While oil free water is running
through the monitor check the Zero adjustment. The display should be "0" to
“2” and the status will show “FW”. If the display varies by greater amounts, it
may be that air entrainment is present. If this is the case, the cause must be
located and rectified.
f) If the Zero need to be adjusted, this can be done in the programming mode as
described in section 10.4. (Settings – Offset)
g) If the flow rate is not sufficient, the instrument will display “Status: Flow”. Both
sample flow rate and clean water flow rate are recommended to be approx.
2 Liters per Minute.
h) Switch the electric Valve (EV) back to sample stream.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 15 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

10.4 Programming Mode


There are 3 groups of push buttons to
control the functions of the display.
Navigation buttons are in group 1.
Functional buttons are group 2. Group 3 is
ESC
+ 1 for data logger operation.
- OK
In the programming mode the alarm set
TEST
ON AL1 AL2 SYS
SET
2 points, the time delays, and the offsets can
be modified. It is also possible to reset to
the factory default values at any time.
LOG
3
The clock is factory set for GMT,
Greenwich Mean Time, and cannot be
OMD-2008
changed.

Initial Display. Normal Operation Display. Exit from SYSTEM-info menu


+ by pressing the ESC button.
Will disappear a few seconds Pressing the button will display
after power up. additional information.
Refer to Fault finding table in
Pressing the O K button will manual for explanations of
display more detailed information status information.
about the current status.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 16 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

AL1 AL2

Pressing the AL1 button leads To change the value, press


into SETTINGS menu, Alarm1 the “+” or “-“ button. Confirm
settings preselected. At the SETTINGS menu the
with “OK”.
alarms, time delays, the Offset
Pressing the Al2 button leads and optionally the output signal
into SETTINGS menu, Alarm2 can be modified within the
settings preselected. limitations. Select the required
point by using the „+“ or „-„
button. To modify settings press
the button.
SET

To change the value, press


the “+” or “-“ button. Confirm
with “OK”.
SET

Pressing the SET button from


Normal Operations Display leads
into SETTINGS menu, set
default option preselected.
To change the value, press
the “+” or “-“ button. Confirm
with “OK”.

At the SETTINGS menu the all To change to “yes”, press the


settings can be reset to the “+” button. Confirm with “OK”
factory default values. To reset to to reset all settings to the
factory values once again press factory default settings.
the button.
SET

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 17 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

SYS

Pressing the SYS button directly Select if you want information Exit from SYSTEM-info menu
leads into SYSTEM menu. about the instrument or by pressing the ESC button.
information about the measuring
cell.

Exit from
MEASURING CELL menu by
pressing the ESC button.

ON

Pressing the ON button directly To change to “ON”, press the


leads into the Select if you want to activate the “+” button. Confirm with “OK”
SYSTEM-OPTIONS menu. clean water valve or if additional to activate the optional clean
information should be displayed. water valve for a limited time.

Flow rate and status of the


meter is displayed. Exit from
Flow display by pressing the
ESC button.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 18 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

TEST

Select if you want to activate the Wait until Alarms Test is


Pressing the TEST button Alarms Test or if Desiccator completed, as indicated by
directly leads into the status information should be countdown value and
SYSTEM-TESTS menu. displayed. progress bar.

Exit from information


Desiccator status display by
pressing the ESC button.

LO G

The LOG button leads into the


data logger function.
Initially the data logger displays
the live data. With the LO G

button it can be switched to the


graphical display mode.

LO G LO G

By pressing the LOG button


twice the recorded data display
mode is invoked. The data logger displays The data logger displays
recorded data. With the LO G recorded data. With the LO G
button it can be switched to the button it can be switched to
non-graphical display mode. the graphical display mode.

In both data display modes the arrow buttons can be used to


navigate to another date/time of recorded data.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 19 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

NB: Changed values have to be confirmed by pressing the " OK " button.
Otherwise the existing values remain valid.

11 Operating instructions
The OMD-2008 EV has an electric switchover valve to switch over from sample
water to clean water. Whenever clean water is used, the instrument will be in
Alarm condition.
The OMD-2008 EV-FC has an additional Flow Sensor to detect the flow rate of
the water through the instrument. If the sample does not flow, the instrument
will be in alarm condition.
The OMD-2008 EV-FC MCU has the Flow Sensor, and additionally it has the
Manual Cleaning Unit.

Instrument start-up sequence:


a) Switch on the power supply.
b) Allow a period of time for water entering the sample tube.
c) Switch the Electric Valve to clean water stream and let oil free water flow
through the system for a few minutes. Check that the display shows 0 to 2
ppm. If not, clean Measuring Cell sample Glass Tube properly.
d) Switch the instrument back to sample stream from to the separator sampling
point connection.
e) The instrument is now ready for use.

11.1 Operator Notes


a) When oily water flows through the instrument the display will show the actual
value of oil content.
b) If the oil concentration exceeds the adjusted threshold (works adjustment
15 ppm), the alarm indicator 1 will be illuminated in intervals during the
selected time delay before it changes to steady light and the associated alarm
relay will operate. Accordingly also the alarm indicator 2 will be illuminated
and its associated alarm relay will take the appropriate shut down action.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 20 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

12 Operator Maintenance
See Section 2 for important notes.
AT WEEKLY INTERVALS:
a) Flush the cell with oil free water.
b) Stop sample and oil free water flow.
c) Unscrew and remove the cell cap. For instruments equipped with Manual
Cleaning Unit (MCU), unscrew the MCU instead.
d) Insert a suitable Cell Cleaning brush (Art. No. 77555) into the cell and clean it
with upwards and downwards motion through the entire length of the cell
several times.
e) Remove the Cell Cleaning brush and replace the cell cap.
f) Open clean water valve and allow oil free water to flow through the instrument
for a few minutes.
g) Observe that the display is showing "0" to “2”. If not, clean again.
h) Examine the status of the desiccator (Chapter 10.4, TEST button). The
Desiccator status display indicates if the desiccator is worn out and working
insufficient. If the desiccator status is any other then OK, the desiccator should
be replaced. Additionally, the Measuring Cell Dew Point Temperature can be
checked. The Dew Point Temperature should be below 0°C, otherwise the
desiccator needs to be exchanged.
Insufficient desiccator performance could result in condensation inside the
measuring cell and wrong measurement and/or damage to optical
components. Desiccator can easily be exchanged. Just unscrew the
desiccator, replace the desiccator with a new one (Art. No. 79550). Make sure
to close the desiccator cap properly. Allow the new desiccator some time to
absorb the humidity inside the measuring cell. Please note that a desiccator
package should never be opened unless it is installed immediately.
j) Switch electric valve to SAMPLE setting.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 21 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

12.1 Manual Cell Clean Unit (MCU) DH 77780


Optional item if fitted
This unit facilitates cleaning of the cell without the need of removing the cell cap.
Regular use of this device should prevent malfunction of the monitor due simply
to fouling of the sample tube and all the inconvenience which this can cause.

Operating Instructions
a) Ensure that there is a clean water supply through the cell.
b) Activate the manual cell clean unit by pressing the handle several times.
c) Check the reading is between 0 to 2 ppm.
d) Repeat a) to c) at least once a week or as necessary.
e) Additional conventional cleaning with the cleaning brush may be necessary.
NB: The Manual Cell Clean Unit may also be used during normal operation with
sample water, but in this case an alarm occurs because the wiper is passing the
light source.
Spares: Wiper Seal DH 77606

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 22 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

Important Information!
Cleaning of Glass Tube at 15 ppm Bilge Alarms OMD-2008 series

IMPORTANT:
NEVER DISASSEMBLE THE UNITS AS THIS MAY VOID THE
CALIBRATION AND THE CERTIFICATION!

CLEANING HAS ONLY TO BE DONE TROUGH THE REMOVED CELL CAP BY

USING THE CLEANING BRUSH!

In most cases of high reading with clean water the measuring cell has a problem with
internal coating of the glass tube. Just cleaning with brush and clean water will not help
in this case.
Please carry out the following instructions to make sure, that the glass tube is really
clean. Then the unit will show 0 to 2 ppm with clean water.

Check Measuring cell humidity readings and desiccator status. Desiccator status must
be OK and dewpoint should be considerably lower then both sample temperature and
clean water temperature. If not, change desiccant container and allow new desiccator to
absorb the humidity inside the measuring cell.
Clean the glass tube by using the cleaning brush under assistance from some cleaner.
In certain cases iron oxide can be deposited inside the glass tube (brownish surface
deposit on the glass tube), depending on environmental conditions on site. In this case
some citric acid, juice from a fresh lemon may help, if you fill it into the glass tube and
leave it at least over night before using the cleaning brush for removing the last dirt from
the glass tube. Also, in cases of calceous deposits in the glass tube, treatment with
some mild acidic cleaner, citric acid, may allow removal of the deposits. Make sure, that
the cleaning fluid will stay in the tube and is not draining. Sometimes the cleaning with
citric acid has to be done 2 or 3 times for at least 12 hours, depending on the thickness
of the coating.
Additional use of tooth paste may also assist in cleaning as a last resort. Do not use
abrasive cleaners containing hard particles. Hard abrasive particles may scratch the
glass surface, permanently damaging the instrument.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 23 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

13 Fault finding
See Section 2 for important notes.
The OMD-2008 EV-FC will indicate several malfunctions in the status line of the
display. Pressing the “OK” button will lead into an information window, similar to
the items listed in the table below.
Status Reading System- Alarm- Reason Servicing
Alarm- circuit 1,2
LED
OK 0..49 Green / Normal Normal operation -
Blinking operation
OK EE Green / Alarm Sample reading is out of range: Wait until oil content is
Blinking Oil content too high, dirty within the range, clean
sample tube sample tube

FW ! 0..49 / EE Green / Alarm Freshwater is enabled -


Blinking
FW ! 0..49 / EE Green / Alarm remote freshwater input is remove/check external
Blinking linked (Terminals 19&20) wiring
EVALVE? any red Alarm FW valve position not detected check connectors and
wiring
Sample? EE Red / Alarm Meter is not able to measure the Check sample, clean
Steady sample: no water in, oil content sample tube according
much too high, no light Page 20
transmission possible
Flow! 0..49 / EE Green / Alarm Low Sample flow rate Check sample flow rate
Blinking Flow Input (Terminals 15&16)
open
Com? EE Red / Alarm No communication between Check connection
Steady computer unit and measuring between computer unit
cell and measuring cell
Datalog? 0..49 / EE Red / Alarm Datalogging is not possible: no Insert the active memory
Steady DECKMA memory card card
inserted
Datalogging is not possible: a Insert the active memory
read only card has been card
inserted
Datalogging is not possible: a Activate card or insert the
new DECKMA memory card active memory card
has been inserted, but has not
been activated
Desicc 0..49 / EE Green / Normal Measuring Cell humidity Check/Replace
Blinking operation critically high (>40%rH) Desiccator
Humid 0..49 / EE Green / Normal Sample temperatur below Check/Replace
Blinking operation dewpoint. Instantaneous Desiccator
condensation possible
Int.Err Red / Alarm Internal error Restart the system
Steady

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 24 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

13.1 Memory Card (refer to Fig. 6)


The Memory Card is on the backside of the Display PCB inside the computer
housing. It is suitable for the life of the instrument, as it is calculated according
MEPC 107(49) to the required storage time of at least 18 month. When the card
is completely used, the oldest entry will be overwritten, so that a replacement is
not necessary. Under normal use the card should not be taken out, as this is
linked with the specific system. The card can be read in other OMD-2008 units,
but writing is only possible in the related system.

1: Display PCB 2: Memory Card


Fig. 6

If no Memory Card is mounted or a card from another system is mounted, the


unit will be in alarm condition.

14 Calibration
15 ppm Bilge Alarms built according MEPC.107(49) have to be protected against
access beyond the checks of instrument drift, repeatability of the instrument
reading and zero adjustment. For this reason the instrument is electronically
sealed, so that only the manufacturer or his authorized persons, equipped with
the related tools, are able to get access for changing the calibration.
To provide a simple procedure to check the instrument aboard ship, the OMD-
2008 EV-FC is constructed in that way, that the zero check also confirms the
instrument drift within the specifications.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 25 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

14.1 Calibration and repeatability check


a) Switch off the power supply and stop any water flow.
b) Clean the sample tube accurately by using a suitable cell cleaning brush as
described under Section 12.0. Make sure, that the offset is correct at ± 0.
c) Run clean water through the instrument.
d) If it is sure, that non aerated, clean water is in the instrument, the reading
should be 0 ppm ± 2 ppm.
e) Continue as described under Section 11.0.
Note § 4.2.11 of MEPC. 107(49):
The accuracy of the 15 ppm Bilge Alarms should be checked at IOPP
Certificate renewal surveys according to the manufacturers
instructions. Alternatively the unit may be replaced by a
calibrated 15 ppm Bilge Alarm. The calibration certificate for
the 15 ppm Bilge Alarm, certifying date of last calibration
check, should be retained onboard for inspection purposes. The
accuracy checks can only be done by the manufacturer or persons
authorized by the manufacturer.

14.2 Function Test at Classification Survey and Port State Control

All 15 ppm Bilge Alarms leaving our works are calibrated according the
requirements with an accuracy of better than +/- 5 ppm within the measuring
range. The alarm points are pre-set to 15 ppm (or 5 ppm for instruments limited
to 5 ppm), and can only be changed to a lower value on site. A setting to a higher
value is not possible.

To provide a simple procedure for check the instrument aboard ship, the OMD-
2008 EV-FC is constructed in that way, that the zero check also confirms the
instrument drift within the specifications. The Test button starts a self test routine
and allows to put both alarms contacts into alarm condition. The instrument will
count down from an assumed high reading (30ppm) downwards until the
assumed value is equal to the actual measured ppm value. Note that this test will
only switch the alarm contacts to non-alarm condition, if the sample contains less
than 15ppm oil content and all other conditions for proper measurement are OK.

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 26 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

15 Spare Parts
When ordering spares, it is important to supply details of the type of monitor, part
number of each spare required, its description and any relevant serial number.

DESCRIPTION ART-NUMBER
Service Kit 18570
Desiccator 79550
Cell Cleaning Brush 77555
O-Ring Set 77775
Memory Card 79186

Service Exchange Measuring Cell 2277500


(including new Calibration Certificate)

Optional Items
Manual Cell Clean Unit 77780
Wiper Seal for Cell Clean Unit 77606
OMD-CR Memory Card Reader 75185
Function Test Kit 18510
Flow Sensor 16700

15.1 Recommended On Board Spares

DESCRIPTION ART-NUMBER
1 off Service Kit 18570

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 27 of 28

Downloaded from www.Manualslib.com manuals search engine


DECKMA HAMBURG GmbH

16 REMARKS

All the modifications and deviations from the standard form, which have to be
carried out in the supply, should be attached at this paragraph.

Commissioned on: ............................. by: ..........................................


Date Firm's Name

Remarks:

Issue: 08.07.2015 Instruction Manual OMD-2008 (EV, FC, MCU Versions) Page 28 of 28

Downloaded from www.Manualslib.com manuals search engine

You might also like