Electronic Instruments: Working of D'Arsonval Meter
Electronic Instruments: Working of D'Arsonval Meter
com
Electronic Instruments
www.allsyllabus.com
ohm meters is a current-sensing device called a D'Arsonval meter movement.
The D'Arsonval movement is a DC moving coil type movement in which an
electromagnetic core is suspended between the poles of a permanent magnet. The
current measured is directed through the coils of the electromagnet so that the
magnetic field produced by the current opposes the field of the permanent magnet
and causes rotation of the core. The core is restrained by springs so that the needle
will deflect or move in proportion to the current intensity.
The D'Arsonval movement is a DC device and can only measure DC current
or AC current rectified to DC.
The next few sections illustrate the modifications used for measuring AC current with
a D’Arsonval meter.
vtu.allsyllabus.com
www.allsyllabus.com
Average-Responding AC voltmeters
In these meters, the meter scale of an average responding meter is calibrated
in terms of the rms value of a sine wave. As most of the wave-forms in electronics
are sinusoidal, this is an entirely satisfactory solution and certainly much less
expensive than a true rms-responding voltmeter.
Nonsinusoidal waveforms, however, will cause this type of meter to read high
or low, depending on the form factor of the waveform.
vtu.allsyllabus.com
www.allsyllabus.com
Fig. 4
These meters are used when required to measure the peak value of a waveform
instead of the average value. As shown in Fig. 4 the rectifier diode charges the
small Capacitor to the peak of the applied input voltage and the meter will indicate
the peak voltage.
In most cases, the meter scale of an average responding (DC) meter is calibrated
in terms of both the rms and peak values of the sinusoidal input waveform. The
www.allsyllabus.com
rms value of a voltage wave that has equal positive and negative excursions is
related to the average value by the form factor
Form Factor
• The form factor, is the ratio of the rms value to the average value of this
waveform,
• for a sinusoid it is expressed as
vtu.allsyllabus.com
www.allsyllabus.com
www.allsyllabus.com
Since Erms = Eav , the calibrated meter scale (Erms = 1.11 Eav) reads high
ksin e _ wave
By a factor = 1.11
k square _ wave
vtu.allsyllabus.com
www.allsyllabus.com
The derivation of the voltage e, its rms value, etc is shown below.
www.allsyllabus.com
Since Erms = 1.155Eav, the calibrated meter scale (Erms = 1.11 Eav) reads low By a
ksin e _ wave 1.11
factor = = 0.961 .
k saw _ tooth _ wave 1.155
vtu.allsyllabus.com
www.allsyllabus.com
These meters are used to provide accurate rms readings of complex waveforms
i.e., non-sinusoidal waveforms having a crest factor of 10:1. Some of the
applications are:
– Measurements of electrical or acoustical noise
– Low duty cycle pulse trains
– Voltages of undetermined waveforms
Problems with Waveforms having a DC Component
Many waveforms have a DC component, which usually shows as the wave not
being symmetrical above and below ground level. Determination of the rms level of
such a waveform sometimes requires separate dc and ac measurements. First the
www.allsyllabus.com
DC level is measured on a DC voltmeter with the AC quantity filtered out. Then the
AC rms level is measured on a capacitor-coupled AC voltmeter. The rms value of
the original waveform is then determined as the square root of the sum of the
squares of the two readings,
The procedure above is not necessary with the true-rms instrument described next
vtu.allsyllabus.com
www.allsyllabus.com
Working:
The 2 thermocouples form part of a bridge in the input circuit of the DC
www.allsyllabus.com
amplifier. The input voltage is amplified and fed to the heating element of the
thermocouple. The heat produced by the wire is sensed by the measuring
thermocouple which produces a proportional DC voltage. This DC voltage upsets
the bridge balance. The unbalance voltage is amplified by the DC amplifier and
fed back to the heating element of the balancing thermocouple. Bridge balance is
reestablished when the two thermocouples produce the same output voltages. At
this point the DC current in the heating element of the feedback thermocouple is
proportional to the AC current in the input thermocouple i.e., the DC is
proportional to the rms value of the input AC signal. This DC value is
indicated by the meter movement in the output circuit
vtu.allsyllabus.com
www.allsyllabus.com
The crest factor for a pure sine wave is 1.414, but non-sinusoidal waveforms can
have much larger crest factors. The rms level of waveforms with a crest factor of 2
or 3 can be determined by most rms measuring instruments. Waveforms with
higher crest factors are more difficult to measure. The maximum waveform crest
factor is usually specified for all rms measuring instruments
The accuracy of this technique has been difficult to control because of the non-
linear behavior of the thermocouple which complicates the meter calibration,
Thermal variations & Sluggish response of the thermocouple which are also
susceptible to burnout also aggravate the problem.
Thermal variations are reduced by installing the heater and the thermo couple
in an evacuated glass bulb and by using fine wires of low thermal conductivity.
Use of null balance techniques reduces the effect of non linear behavior. Generally
the Nonlinear behavior of the measuring and feedback (balancing) thermocouples
cancel each other.
www.allsyllabus.com
vtu.allsyllabus.com
www.allsyllabus.com
Electronic Multimeters
One of the most versatile general-purpose instruments capable of measuring
dc and ac voltages as well as current and resistance is the solid-state electronic
multimeter or VOM.
Although circuit details will vary from one instrument to the next, an electronic
multimeter generally contains the following elements:
• Balanced-bridge dc amplifier and indicating meter
• Input attenuator or RANGE switch, to limit the magnitude of the input
voltage to the desired value
• Rectifier section, to convert an ac input voltage to a proportional dc value
• internal battery and additional circuitry, to provide the capability of
resistance measurement
• FUNCTION switch, to select the various measurement functions of the
instrument
In addition, the instrument generally has a built-in power supply for ac line
operation and, in most cases, one or more batteries for operation as a portable test
www.allsyllabus.com
instrument.
Fig.6 Balanced Bridge dc amplifier with input attenuator and indicating meter
vtu.allsyllabus.com
www.allsyllabus.com
To return the meter to zero, the circuit is balanced by ZERO adjust control
R3 for a true null indication.
www.allsyllabus.com
Output Indication
When a positive voltage is applied to the gate of input transistor Q1, its drain
current increases which causes the voltage at the source terminal to rise. The
resulting unbalance between the Ql and Q2 source voltages is indicated by the
meter movement, whose scale is calibrated to agree with the magnitude of the
applied input voltage.
Typical input voltage attenuator for a VOM is shown in Fig.7. The RANGE
switch on the front of the panel of the VOM allows selection of the desired
voltage range. The maximum voltage that can be applied to the gate of Q1 is
determined by the operating range of FET (usually a few volts). The range of
input voltages can easily be extended by an input attenuator or RANGE switch,
as shown in Fig.7. The unknown dc input voltage is applied through a large
resistor in the probe body to a resistive voltage divider. Thus, with the RANGE
vtu.allsyllabus.com
www.allsyllabus.com
switch in the 3-V position as shown, the voltage at the gate of the input FET is
developed across 8 MΩ of the total resistance of 11.3 MΩ and the circuit is so
arranged that the meter deflects full scale when 3 V is applied to the tip of the
probe. With the RANGE switch in the 12-V position, the gate voltage is
developed across 2 MΩ of the total divider resistance of 11.3 MΩ and an input
voltage of 12 V is required to cause the same full-scale meter deflection.
www.allsyllabus.com
Fig. 7
vtu.allsyllabus.com
www.allsyllabus.com
Resistance Ranges
When the function switch of the multimeter is placed in the OHMS position, the
unknown resistor is connected in series with an internal battery, and the meter
simply measures the voltage drop across the unknown. A typical circuit is shown
in Fig.8, where a separate divider network, used only for resistance measurements,
provides for a number of different resistance ranges. When unknown resistor Rx is
connected to the OHMS terminals of the multimeter, the 1.5-V battery supplies
current through one of the range resistors and the unknown resistor to ground.
Fig.8
www.allsyllabus.com
Voltage drop Vx across Rx is applied to the input of the bridge amplifier and
causes a deflection on the meter. Since the voltage drop across Rx is directly
proportional to its resistance the meter scale can be calibrated in terms of
resistance which in an Electronic multimeter is from left to right (vice versa for
ordinary multimeter). i.e., High Rx => High Vx ( in ordinary multimeter High Rx
=> Low current I)
vtu.allsyllabus.com
www.allsyllabus.com
Digital Voltmeters
Digital voltmeters are essentially analog-to-digital converters with digital displays
to indicate the measured voltage. DVM displays measurements of ac or dc
voltages as discrete numerals instead of a pointer deflection on a continuous scale.
The development of ICs has lead to drastic reduction of the size, power
requirements and cost of DVM. Hence DVMs can actively compete with
conventional analog instruments, both in portability and price.
CONSIDERATIONS IN CHOOSING AN ANALOG VOLTMETER
The most appropriate instrument for a particular voltage measurement depends on
the performance required in a given situation. Some important considerations are
– Input Impedance
– Voltage Ranges
– Decibels
– Sensitivity Versus Bandwidth
– Battery Operation
vtu.allsyllabus.com
www.allsyllabus.com
Classification of DVMs
1. Ramp-type DVM
2. Integrating DVM
3. Continuous-balance DVM
4. Successive-approximation DVM
Ramp-type DVM
Its Operating principle is based on the measurement of the time it takes for a linear
ramp voltage to rise from 0 V to the level of the input voltage, or to decrease from
the level of the input voltage to zero. This time interval is measured with an
electronic time-interval counter, and the count is displayed as a number of digits
on electronic indicating tubes.
www.allsyllabus.com
The working principle i.e., the Conversion from a voltage to a time interval is
illustrated by the waveform in Fig. 10
vtu.allsyllabus.com
www.allsyllabus.com
Fig.10
At the start of the measurement cycle, a ramp voltage is initiated; this voltage can
be positive-going or negative-going. The negative-going ramp, (see Fig. 10) is
continuously compared with the unknown input voltage. At the instant that the
www.allsyllabus.com
ramp voltage equals the unknown voltage, a coincidence circuit, or comparator,
generates a pulse which opens a gate. The ramp voltage continues to decrease with
time until it finally reaches 0 V (or ground potential) and a second comparator
generates an output pulse which closes the gate. An oscillator generates clock
pulses which are allowed to pass through the gate to a number of decade counting
units (DCUs) which totalize the number of pulses passed through the gate. The
decimal number, displayed by the indicator tubes associated with the DCUs, is a
measure of the magnitude of the input voltage.
vtu.allsyllabus.com
www.allsyllabus.com
Staircase-Ramp DVM
www.allsyllabus.com Fig. 11
It contains a 10-MΩ input attenuator, providing five input ranges from 100
mV to 1,000 V full scale. The dc amplifier with a fixed gain of 100, delivers 10 V
to the comparator at any of the full-scale voltage settings of the input
divider. The comparator senses coincidence between the amplified input voltage
and the staircase-ramp voltage which is generated as the measurement proceeds
through its cycle . A Clock (4.5 kHz oscillator) provides pulses to three DCUs
in cascade. The units counter provides a carry pulse to the tens decade at every
tenth input pulse. The tens decade counts the carry pulses from the units decade
and provides its own carry pulse after it has counted ten carry pulses. This carry
pulse is fed to the hundreds decade which provides a carry pulse to an over-
range circuit. The over range circuit causes a front panel indicator to light up,
warning the operator that the input capacity of the instrument has been
exceeded. The operator should then switch to the next higher setting on the input
attenuator.
Each decade counter unit is connected to a digital-to-analog (D/A) converter.
vtu.allsyllabus.com
www.allsyllabus.com
The outputs of the D/A converters are connected in parallel and provide an output
current proportional to the current count of the DCUs. The staircase amplifier
converts the D/A current into a staircase voltage which is applied to the
comparator. When the comparator senses coincidence of the input voltage and the
staircase voltage, it provides a trigger pulse to stop the oscillator. The current
content of the counter is then proportional to the magnitude of the input voltage.
The sample rate is controlled by a simple relaxation oscillator. This oscillator
triggers and resets the transfer amplifier at a rate of two samples per second. The
transfer amplifier provides a pulse that transfers the information stored in the
decade counters to the front panel display unit. The trailing edge of this
pulse triggers the reset amplifier which sets the three decade counters to zero
and initiates a new measurement cycle by starting the master oscillator or clock.
The display circuits store each reading until a new reading is completed,
eliminating any blinking or counting during the computation.
The ramp type of A/D converter requires a precision ramp to achieve accuracy.
Maintaining the quality of the ramp requires a precise, stable capacitor and
resistor in the integrator. In addition, the offset voltages and currents of the
www.allsyllabus.com
operational amplifier used in the integrator are critical in the accurate ramp
generator. One method of reducing the dependence of the accuracy of the
conversion on the resistor, capacitor, and operational amplifier is to use a technique
called the dual-slope converter.
Dual-Slope Converter
In the dual-slope technique, an integrator is used to integrate an accurate
voltage reference for a fixed period of time. The same integrator is then used to
integrate with the reverse slope, the input voltage, and the time required to return
to the starting voltage is measured. The Order of integrations does not matter.
Consider the integration of the unknown first as shown in Fig. 12
vtu.allsyllabus.com
www.allsyllabus.com
Fig. 12
If the integration were allowed to continue for a fixed period of time T1, the
output voltage would be
www.allsyllabus.com
Notice that the integrator output has gone in the opposite polarity as the input.
That is, a positive input voltage produces a negative integrator output. If a reference
voltage Vref, were substituted for the input voltage Vx, as shown in Fig.13, the
integrator would begin to ramp toward zero at a rate of Vref/ RC assuming that the
Vref was of the opposite polarity as the unknown input voltage. The integrator for
this situation does not start at zero but at an output voltage of V1, and the output
voltage Vout is
vtu.allsyllabus.com
www.allsyllabus.com
Fig.13
Setting the output voltage of the integrator to zero and solving for Vx yields
where Tx is the time required to ramp down from the output level of V1 to zero volts.
Notice that the relationship between the reference voltage and the input voltage does
not include R or C of the integrator but only the relationship between the two times.
Because the relationship between the two times is a ratio, an accurate clock is not
www.allsyllabus.com
required but only that the clock used for the timing be stable enough that the
frequency does not change appreciably from the up ramp to the down ramp.
Fig. 14
vtu.allsyllabus.com
www.allsyllabus.com
The dual-slope type of A/D conversion is a very popular method for digital
voltmeter applications. When compared to other types of ADC techniques, the
dual-slope method is slow but is quite adequate for a digital voltmeter used for
laboratory measurements. For data acquisition applications, where a number of
measurements are required, faster techniques are recommended. Many
refinements have been made to the technique and many large-scale-integration
(LSI) chips are available to simplify the construction of DVMs.
When a dual-slope A/D converter is used for a DVM the counters may be
www.allsyllabus.com
decade rather than binary and the segment and digit drivers may be contained in
the chip. When the converter is to interface to a microprocessor, and many high-
performance DVMs use microprocessors for data manipulation, the counters
employed are binary.
One significant enhancement made to the dual-slope converter is automatic
zero correction. As with any analog system, amplifier offset voltages, offset
currents, and bias currents can cause errors. In addition, in the dual-slope A/D
converter, the leakage current of the capacitor can cause errors in the integration
and consequentially, an error. These effects, in the dual-slope AID converter, will
manifest themselves as a reading of the DVM when no input voltage is present.
Fig.15 shows a method of counteracting these effects.
vtu.allsyllabus.com
www.allsyllabus.com
Fig.15
The input to the converter is grounded and a capacitor, the auto zero capacitor, is
connected via an electronic switch to the output of the integrator. The feedback of
the circuitry is such that the voltage at the integrator output is zero. This
effectively places an equivalent offset voltage on the automatic zero capacitor so
that there is no integration. When the conversion is made, this offset voltage is
present to counteract the effects of the input circuitry offset voltages. This
automatic zero function is performed before each conversion, so that changes in
the offset voltages and currents will be compensated.
www.allsyllabus.com
vtu.allsyllabus.com
www.allsyllabus.com
Successive-approximation DVM
Fig. 17
www.allsyllabus.com
than" or "less than" decision is made by a comparator. The D/A converter
provides the estimate and is compared to the input signal. A special shift register
called a successive-approximation register (SAR) is used to control the D/A
converter and consequentially the estimates. At the beginning of the conversion all
the outputs from the SAR are at logic zero. If the estimate is greater than the input,
the comparator output is high and the first SAR output reverses state and the
second output changes to a logic "one." If the comparator output is low, indicating
that the estimate is lower than the input signal, the first output remains in the logic
one state and the second output assumes the logic state one. This continues to all
the states until the conversion is complete.
vtu.allsyllabus.com
www.allsyllabus.com
Q-Meter
The Q meter is an instrument designed to measure some of the electrical
properties of coils and capacitors. The operation of this useful laboratory
instrument is based on the characteristics of a series-resonant circuit, i.e., that the
voltage across the coil or the capacitor is equal to the applied voltage times the Q
of the circuit. If a fixed voltage is applied to the circuit, a voltmeter across the
capacitor can be calibrated to read Q directly.
Fig. 19
vtu.allsyllabus.com
www.allsyllabus.com
Measurement Methods
There are three methods for connecting unknown components to the test terminals of
a Q meter: direct, series, and parallel.
The type of component and its size determine the method of connection
Direct connection
Most coils can be connected directly across the test terminals, exactly as shown in the
basic Q-circuit of Fig.19. The circuit is resonated by adjusting either the oscillator
frequency or the resonating capacitor. The indicated Q is read directly from the
vtu.allsyllabus.com
www.allsyllabus.com
"Circuit Q" meter, modified by the setting of the "Multiply Q by" meter. When the
"Multiply Q by" meter is set at the unity mark, the “Circuit Q" meter reads the correct
value of Q directly.
Series connection
Low-impedance components, sub as low-value resistors, small coils, and large
capacitors, are measured in series with the measuring circuit. Fig.20 shows the
connections. The component to be measured here indicated by [Z], is placed in
series with a stable work coil across the test terminals. (The work coil is usually
supplied with the instrument)
www.allsyllabus.com
Fig.20 Q-meter measurement of a low-impedance component in series connection.
vtu.allsyllabus.com
www.allsyllabus.com
For the second measurement, the reactance of the unknown can be expressed in terms
of the new value of the tuning capacitor (C2) and the in-circuit value of the inductor
(L). This yields
www.allsyllabus.com
If the unknown is a small inductor, the value of the inductance is found from Eq. and
is
vtu.allsyllabus.com
www.allsyllabus.com
Parallel connection
High-impedance components, such as high-value resistors, certain inductors, and
small capacitors are connected them in parallel with the measuring circuit.
www.allsyllabus.com
vtu.allsyllabus.com
www.allsyllabus.com
When the unknown impedance is connected into the circuit and the capacitor is tuned
for resonance, the reactance of the working coil (XL) equals the parallel reactances of
the tuning capacitor (Xc2) & the unknown (Xp)
1
If unknown is inductive, X P = ωLP and LP = 2
ω (C1 − C 2 )
If the unknown is capacitive, Xp = 1/ωCp and CP = C1 – C2
The resistance (Rp) of the unknown impedance is found by using the conductances in
the circuit
• GT -- total conductance of the resonant circuit
•
•
www.allsyllabus.com
Gp -- conductance of the unknown impedance
GL -- conductance of the working coil
GT = GP + GL
GP = GT – GL
vtu.allsyllabus.com
www.allsyllabus.com
Sources of Error
www.allsyllabus.com
Fig.22
Working: make two measurements at different frequencies. The coil under test is
connected directly to the test terminals of the Q meter, as shown in Fig.22. The
tuning capacitor is set to a high value, preferably to its maximum position, and the
circuit is resonated by adjusting the oscillator frequency. Resonance is indicated
by maximum deflection on the "Circuit Q" meter. The values of the tuning
capacitor ( C1) and the oscillator frequency (f1) are noted. The frequency is then
increased to twice its original value = 2f1 and the circuit is returned by adjusting
the resonating capacitor (C2).
vtu.allsyllabus.com
www.allsyllabus.com
After the oscillator and the tuning capacitor are adjusted, the capacitance of the circuit
is C2 + Cd , and the resonant frequency equals
www.allsyllabus.com
vtu.allsyllabus.com