Commissioning Procedures: Vegaflex 61
Commissioning Procedures: Vegaflex 61
Commissioning Procedures
VEGAFLEX 61
4 … 20 mA/HART two-wire
Contents
Contents
1 About this document
1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Symbolism used. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 Installation Check
3.1 PruProductdescription
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 6
3.2 General instructions . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.3 Mounting instructions . . . . . . . . . . . . . . . . . . . . . . . . 8
Ex applications
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
This arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbitrary
conversions or modifications are explicitly forbidden.
2.6 CE conformity
This device fulfills the legal requirements of the applicable EC
guidelines. By attaching the CE mark, VEGA provides a confirmation
of successful testing. You can find the CE conformity declaration in the
download area of www.vega.com.
3 Installation Check
3.1 Product description
1
2
3
Fig. 1: VEGAFLEX 61
1 Housing cover with integrated indicating and adjustment module (optional)
2 Housing with electronics
3 Process fitting
Welding work Before beginning the welding work, remove the electronics module
from the sensor. By doing this, you avoid damage to the electronics
through inductive coupling.
Moisture You can give your instrument additional protection against moisture
penetration by leading the connection cable downward in front of the
cable entry. Rain and condensation water can thus drain off. This
applies mainly to outdoor mounting as well as installation in areas
where high humidity is expected (e.g. through cleaning processes) or
on cooled or heated vessels.
1 2
Voltage supply Power supply and current signal are carried on the same two-wire
cable. The voltage supply range can differ depending on the
instrument version.
The data for power supply are specified in chapter "Technical data".
Provide a reliable separation between the supply circuit and the mains
circuits according to DIN VDE 0106 part 101. The VEGA power supply
units VEGATRENN 149A Ex, VEGASTAB 690 as well as all
VEGAMETs and VEGASCANs meet this requirement.
Keep in mind the following additional influences on the operating
voltage:
l Output voltage of the power supply unit can be lower under
nominal load (with a sensor current of 20.5 mA or 22 mA in case of
fault message)
l Influence of additional instruments in the circuit (see load values in
chapter "Technical data")
Connection cable The instrument is connected with standard two-wire cable without
screen. If electromagnetic interference is expected which is above the
test values of EN 61326 for industrial areas, screened cable should be
used.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm
(0.2 … 0.35 in) ensures the seal effect of the cable gland. If you are
using cable with a different diameter or cross-section, exchange the
seal or use a suitable cable gland.
We generally recommend the use of screened cable for HART
multidrop mode.
Cable gland ½ NPT On the instrument with cable entry ½ NPT and plastic housing there is
a metallic ½" threaded insert moulded into the plastic housing.
8 Press down the opening levers of the terminals, you will hear the
terminal spring closing
Display
I²C 1
1 2 5 6 7 8
4 2
Fig. 11: Electronics and connection compartment with single chamber housing
1 Plug connector for VEGACONNECT (I²C interface)
2 Spring-loaded terminals for connection of the external indication VEGADIS
61
3 Ground terminal for connection of the cable screen
4 Spring-loaded terminals for voltage supply
Wiring plan
Display
I2C
1 2 5 6 7 8
Electronics compart-
ment
Display
1
I²C
1 2 5 6 7 8
3 2
Connection compart-
ment
Display
1 2 I²C
Wiring plan
I2C
1 2
Display
1
I²C
1 2 5 6 7 8
3 2
Connection compart-
ment
1 2
Wiring plan
1 2
2
1
1.1
Adjustment system The sensor is adjusted via the four keys of the indicating and
adjustment module. The LC display indicates the individual menu
items. The functions of the individual keys are shown in the above
illustration. Approx. 10 minutes after the last pressing of a key, an
automatic reset to measured value indication is triggered. Any values
not confirmed with [OK] will not be saved.
2 Select the menu item "Basic adjustment" with [->] and confirm with
[OK]. Now the menu item "Min. adjustment" is displayed.
Min. adjustment
0.00 %
=
10.000 m(d)
8.000 m(d)
3 Prepare the % value for editing with [OK] and set the cursor to the
requested position with [->]. Set the requested percentage value
with [+] and save with [OK]. The cursor jumps now to the distance
value.
4 Enter the suitable distance value in m for the empty vessel (e.g.
distance from the sensor to the vessel bottom) corresponding to
the percentage value.
5 Save the settings with [OK] and move to "Max. adjustment" with [-
>].
Carry out max. adjust- Proceed as follows:
ment
Max. adjustment
100.00 %
=
1.000 m(d)
2.000 m(d)
1 Prepare the % value for editing with [OK] and set the cursor to the
requested position with [->]. Set the requested percentage value
with [+] and save with [OK]. The cursor jumps now to the distance
value.
17 VEGAFLEX 61 • 4 … 20 mA/HART two-wire
5 Set up with the indicating and adjustment module PLICSCOM
Application Each product has different reflective properties. In addition, there are
various interfering factors which have to be taken into account:
agitated product surfaces and foam generation (with liquids); dust
generation, material cones and echoes from the vessel wall (with
solids). To adapt the sensor to these different conditions, you should
first select in this menu item under "Medium" either "Liquid" or "Solid".
Application
Liquid
Standard
(DK ≥ 2)
Damping
0s
Enter the requested parameters via the appropriate keys, save your
settings and jump to the next menu item with the [->] key.
Linearisation curve A linearization is necessary for all vessels in which the vessel volume
does not increase linearly with the level - e.g. in a horizontal cylindrical
or spherical tank - and the indication or output of the volume is
required. Corresponding linearization curves are preprogrammed for
Linearisation curve
linear
Enter the requested parameters via the appropriate keys, save your
settings and jump to the next menu item with the [->] key.
Sensor-TAG In this menu item you can enter an unambiguous designation for the
sensor, e.g. the measurement loop name or the tank or product
designation. In digital systems and in the documentation of larger
plants, a singular designation should be entered for exact identification
of individual measuring points.
Sensor-TAG
Sensor
With this menu item, the Basic adjustment is finished and you can now
jump to the main menu with the [ESC] key.
Gating out of false sig- High sockets or vessel installations, such as e. g. struts or agitators as
nals well as buildup and weld joints on the vessel walls, cause interfering
reflections that can impair the measurement. A false echo storage
detects and marks these false echoes, so that they are no longer taken
into account for the level measurement. A false echo memory should
be created with low level so that all potential interfering reflections can
be detected.
Change now?
Proceed as follows:
1 Move from the measured value display to the main menu by
pushing [OK].
2 Select the menu item "Service" with [->] and confirm with [OK].
Now the menu item "False signal suppression" is displayed.
3 Confirm "False signal suppression - Change now" with [OK] and
select in the below menu "Create new". Enter the actual distance
from the sensor to the product surface. All false signals in this area
are detected by the sensor and saved after confirming with [OK].
Copy sensor data This function enables reading out parameter adjustment data as well
as writing parameter adjustment data into the sensor via the indicating
and adjustment module. A description of the function is available in the
operating instructions manual "Indicating and adjustment module".
The following data are read out or written with this function:
l Measured value presentation
l Adjustment
l Medium
l Vessel form
l Damping
l Linearisation curve
l Sensor-TAG
l Displayed value
l Display unit
l Scaling
l Current output
l Unit of measurement
l Language
l Sensitivity
The following safety-relevant data are not read out or written:
l HART mode
l PIN
l SIL
l Sensor length/Sensor type
l Gating out of false signals
Copy sensor data
Min. adjustment 1.1 Max. adjustment 1.2 Damping 1.3 Linearisation curve 1.4
0.00 % 100.00 %
= = 0s linear
10.000 m(d) 1.000 m(d)
8.000 m(d) 2.000 m(d)
Sensor-TAG 1.5
Sensor
Display
Basic adjustment 2
▶ Display
Diagnostics
Service
Info
Diagnostics
Basic adjustment 3
Display
▶ Diagnostics
Service
Info
Pointer 3.1 Sensor status 3.2 Curve selection 3.3 Echo curve 3.4
Service
Basic adjustment 4
Display
Diagnostics
▶ Service
Info
Sensor 4.1 Application 4.2 Gating out of false signals 4.3 Current output 4.4
Liquid ▼ Output mode: 4-20 mA ▼
5.00 m(d) ▼ Change now? Failure mode: 20.5 mA ▼
Rod ▼ Standard min. current: 4 mA ▼
(DK ≥ 2) ▼ max. current: 20.5 mA ▼
SIL 4.9 HART mode 4.10 Copy sensor data 4.11 PIN 4.12
Standard
Activated! ▼ Address 0 Copy sensor data? Enable?
Info
Basic adjustment 5
Display
Diagnostics
Service
▶ Info
Instrument type 5.1 Date of manufacture 5.2 Last change using PC 5.3 Sensor characteristics 5.4
12. Dec. 2005
Serial number Software version 04. March 2004 Display now?
12345678 3.22