Control Box CB6S - OpenBusTM

Data sheet

LINAK.COM/MEDLINE-CARELINE
CB6S - OpenBusTM

CB6S is part of the LINAK OpenBus™ product range - that


provides more flexible solutions no matter which actuator
concept is preferred.
LA23, LA27C, LA31, LA34, LA40, LA43, LA44, BL1 with
mini-fit plug and std. end-stop switch / signal switch are all
compatible.

Features: Options:
• Battery lock: A new improved lock introduced to fix • External battery, connection (BA18):
the battery plug Capacity: 1.2Ah - must be ordered separately
• Channels: 4 • Ambient temp. : +5º to +40ºC
• Colour: Light grey (RAL7035) • Humidity: 20% - 90% @ 30ºC- not condensing.
• Control concept: FET
• Charging indicator: Supported Usage:
• Duty cycle : 10% - 2/18 min. on/off continuous use • Approvals: IEC 60601-1:2005, EN 60601-1:2006,
• EOP: Yes ANSI/AAMI ES 60601-1:2005 + A2:2010,
• Housing type: V - 0 and CAN/CSA-C22.2 No. 60601-1:08
• IP Rating: IPX6
• Mains voltage: 230V, 120V The OBF version supports:
• Max. current, total: 8A total -w/EOP • Charging indicator (for indication of battery status)
• Max. current, per ch.: 8A / CH -w/EOP • Service monitoring
• Operation: True parallel • Feedback for positioning (parallel operation)
• OpenBus accessories, current limit: Max. 200 mA • More OpenBus™ accessories connected to one system
(Vbus), must be ensured by the sales engineer
• Safety system: Controlled and monitored by Software The OBF platform is available with 1 standard software
• Service monitor: Supported package Z.
• Software: Predefined - 1 standard package available;
customised software available upon request
• Speed control: Support of speed control
• Start/Stop: Soft-start / Soft-stop
• Transformer: +35%

LINAK.COM/MEDLINE-CARELINE
CB6S Ordering example:

CB6S 6 1 4 + A 1 0 0 9

Cable connection: 9 = Cable socket without GND (Class 2)

Not used

Voltage input: 0 = 230V


1 = 120V

Effect: 2 = +35% transformer


- important, specify +35% vers. if possible

Software: Z = Standard OBF software (0078212)

Colour: + = Plastics RAL7035 (light grey)

Number of channels: 4 = CH 1-4

PCB type: 3 = OBF with external battery

Protection class: 6 = IPX6

Type: Control box CB6S


CB6S platform
Controls (= 1 pcs only) Controls (>1)*
HB70, HB80, ACO, FPP HB70, HB80, ACO, FPP, UBL2

Accessories
MJB5, MJB8

CB6 OBF
Transformer

Mains cable BA18

Ch. 1-4
Power switch
Ch. 1-4
LA, switch signal
LA, power switch
LA, power switch and Hall
LA, encoded * Max. 12 items or <200 mA consumption
LA, encoded and Hall (OBL = 25 mA)
User Functionality, OBF
CB6 OBF is available as standard with software item ‘Z’. This OBF software has software item no. = 0078212.

OBF - support of ACO (attendant control for nursing staff)


CB6 OBF also supports the ACO. However this set-up has more features which are:
• Locking of channels is improved to be locking of functions. It provides more possibilities for the user.
• The set-up also supports the CPR function. Therefore front covers incl. CPR can be used. E.g.:

• The set-up also supports a charging indicator.

OBF - support of Under Bed Light (UBL)


UBL is also supported by CB6S OBF.
Dimensions:

176.4

102.8

21.5

164.9
136.2

137.4
141
R4
9

226.5

49
25
100.5

43

Mounting of CB6S – article no. 9271233


Apart from LA27 and LA40, it is not possible to mount CB6S onto LINAK actuators. In such case you need a CB6 mounting bracket which can be
welded or likewise to the application.
This bracket is shown below:

6.8
17

28.25
22

13.15 17.8
80

11.5
6.8
40
35.8

12
14

85

128.3
Copyright © LINAK 2017.07 . MA M9-02-267-C . Chapter 6.13

Terms of use
The user is responsible for determining the suitability of LINAK products for specific application. LINAK takes great care in providing
accurate and up-to-date information on its products.
However, due to continuous development in order to improve its products, LINAK products are subject to frequent modifications and
changes without prior notice. Therefore, LINAK cannot guarantee the correct and actual status of said information on its products.
While LINAK uses its best efforts to fulfil orders, LINAK cannot, for the same reasons as mentioned above, guarantee the availability
of any particular product. Therefore, LINAK reserves the right to discontinue the sale of any product displayed on its website or listed
in its catalogues or other written material drawn up by LINAK.
All sales are subject to the Standard Terms of Sale and Delivery for LINAK. For a copy hereof, please contact LINAK.

FOR MOUNTING INSTRUCTIONS AND GUIDANCE IN USAGE, PLEASE SEE THE RELEVANT USER’S MANUALS