Wandong Products Specification
China Resources Wandong Medical Equipment Co., Ltd.
High Frequency 50kW, 150kV Radiography System
- HF50-R
Building 3, No.9, Jiuxianqiaodong Road, Chaoyang District, Beijing 100015, P.R. China
E-mail: [email protected] Tel/Fax: +86-10-84575792/3/5, 84575794
This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 1 of 7
Wandong Products Specification
WDM: A successful journey of 60 years
China Resources Wandong Medical Equipment Co., Ltd. (WANDONG, originally Beijing Wandong
Medical Equipment Co., Ltd.) was established in 1955, and has been listed on the Shanghai Stock
Exchange since 1997. Headquartering in High-Tech industrial
park in Beijing, Wandong occupies an area of more than
100,000 square meters as its modern production base and
R&D center, and has 30+ branches in China and a wide sales
and service network in about 70 countries all over the world.
With the 60 years dedication to the medical imaging,
Wandong has made brand name of “WDM”, a well-known
brand in the world.
As one of the largest radiology imaging equipment
manufacturers in the world, Wandong has wide range of
product lines including General Radiography, Mobile X-ray &
C-arm, Digital Radiography (DR), Digital Fluoroscopy (DRF),
Digital Mammography, Cath-Lab, Computed Tomography (CT)
and Magnetic Resonance Imaging (MRI) machine, and the
annual production capability is more than 6,000 units.
This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 2 of 7
Wandong Products Specification
Performance feature:
1) Excellent High-frequency Generator
High-frequency generator output is stable and precise. System effectively reduced radiation to
patient by much lower scattering dose than traditional machine
Transient exposure eliminates motion artifacts.
APR (Anatomical Program Radiography) of more than 600 set exposure programs can meet all
clinical requirements. (5 kinds of body size, 30 anatomical positions, 4 projection directions)
2) Humanization Design Console
Fast and reliable operation guaranteed by soft touch buttons and LCD display of parameters.
Categorized buttons in control panel match with doctor’s operation procedure.
Variety of console installation including wall-hanging, pedestal table, desktop placed fits for
different site requirements.
Operation fatigue will be reduced by ergonomic design.
3) Integrated Radiography table
The wide range of movement of Stand and radiography table to ensure patient to be covered
the whole body by X-ray tube, To fully meet the lateral position、oblique position and the head
of table and requirement of special position.
Stand rotation makes stretcher patients to do diagnostic without moving.
This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 3 of 7
Wandong Products Specification
1. Configurations:
Standard Configurations:
No. Component Part Number Brand Quantity
1 High Frequency Generator GFS501-1 WDM 1
2 X-ray Tube Assembly E7843X Toshiba 1
3 HV cable DL7512-1B WDM 1
4 Collimator XS1-2 WDM 1
5 Radiography Table SC4-4 WDM 1
6 Vertical Bucky Stand LS-3 WDM 1
Optional Configurations:
No. Component Part Number Brand Quantity
1 Ionization Chamber SSMC610 3 fields Claymount 1
2 Radiography Table SC4-1 WDM 1
This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 4 of 7
Wandong Products Specification
2. Technical Specification:
Standard configuration
Component Specifications
Part Number: GFS501-1
Power: 50kW
Constant Frequency of Inverter: 25kHz
KV Range: 40~150kV
mA Range: 25~630mA
High Frequency Time Range: 2ms~5s
Generator mAs:0.5mAs~500mAs
AEC interface, APR
Power Supply: 380VAC, three-phase
Fault Auto Detection and Diagnosis
Console: LCD Displays
APR:600 positions
Part Number: Toshiba E7843X
Focal Spot: 0.6mm/ 1.2mm
X-ray Tube
Power: 22 / 50kW
Assembly
Anode Heat Capacity: 150kHU
Target Angle: 12°
Part Number: XS1-2
External Dimension: 250mm×224mm×250mm
Collimator
Projection Field(SID=65cm) Max: 35cm×35cm; Min: 0cm×0cm
Operation Mode: Manual
This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 5 of 7
Wandong Products Specification
Part Number: SC4-4
External Dimension(Length* Width*Height): 2000mm*740mm*658mm
Tabletop Movement(Longitudinal):1200mm
Tabletop Movement(Transverse): 200mm
Longitudinal Travel of Tube Stand: 1200mm
Vertical Travel: 810mm~1710mm (Distance from X-ray Tube Focus to The Floor)
SID: 1100mm Max. (to Table Bucky)
Rotation Range of X-ray Tube Around Horizontal Arm: +120°~-120°
Tube Stand Self-Rotation Range: +180° ~ -180°, and the interval of 90°could be located.
Radiography
Tabletop Locking Mode: Electromagnetic lock, braking force≥100N
Table
Tabletop Locking Control: foot switch
Bucky Mode: Fixed
Grid Size: 356mm×432mm(14 ″×17″)
Bucky specification: f0=100cm, r=10, N=40L/cm
Cassette Size: Max 14″×17″
Distance between Tabletop and Bucky Film ≤65mm
Longitudinal Travel of Bucky(SID): 640mm
Weight Capacity: Max. 135 kg
Part Number: LS-3
Vertical Travel: ≥1000mm
Vertical Bucky
Maximum Dimension: 8″ ×10″-14″ ×17″
Stand
Grid (According to User’s Requirement):
N=40L/cm; r=8, 10 or 12; f0=150cm, 180cm or 200cm
This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 6 of 7
Wandong Products Specification
Optional Components
Component Specifications
Part Number: Claymount SSMC610
Ionization AEC function, 3 fields
Chamber Exposure dose range:1~100uGy
Exposure time range: 1ms~5s
Part Number: SC4-1
External Dimension (Length* Width*Height): 2000mm*730mm*690mm
Tabletop Movement(Longitudinal): 1200mm
Tabletop movement(Transverse): 200mm
X-ray tube Stand Travel Range (Longitudinal): 580mm
Rotation Range of X-ray Tube Around Horizontal Arm: +90°~-90°
X-ray Tube Stand Self-Rotation Range: +180°~-180°, interval of 90°could be located.
Radiography
Distance from X-ray tube focus to the floor ≥ 1030mm~1740mm
table
Tabletop Locking Mode: Electromagnetic lock, footswitch control, braking force≥100N
Gird: 388mmх464mm, f0: 100cm, r10, N40 L/cm
Cassette Size: Max. 14″×17″
Distance between Tabletop and Bucky Film ≤ 80mm
Bucky Longitudinal Travel: 580mm
Bucky locking: Electromagnetic, braking force≥100N
This document must not be used as contract file. The information provided in this paper is subject to change without further notice
Page 7 of 7