Troubleshooting Flow Chart
Troubleshooting Flow Chart
Have a cold
beverage.
Revised and updated on January 16, 2003, by FACScan(FC)\FACSort(FST)
Tony Leger and Automation Lab Technology. FACScalibur (FCB)
Fluidics/Pneumatics DCM is the acronym for YES
troubleshooting guide droplet containment module.
SIP is the acronym for sample
injection probe
NO Is the
Is there a No Is Sample Is leak YES No YES YES instrument
puddle of injection coming from Are there Is there a Is the
passing the
fluid under probe sheath or any fluid sample flow rate
daily QC
the dripping? waste tank leaks? flow rate? stable?
protocol
Cytometer ? area?
No No
YES
YES SIP may be
YES clogged See
YES See trouble Is Instrument
shooting status NOT
Section 6 NO
See trouble
Open fluidic No shooting
section 4 for READY? step1 for help
Is DCM pump help
door (FC/FST) running when section 1
and remove right the sample
side panel. See
arm is to the YES troubleshoot
(FC/FST/FCB) side?
NO ing
Section
Problem
YES 7 for help
solved?
See
See trouble Yes troubleshooting
Is instrument
shooting aids staying in
section 1 for Section 5 for STANDBY No
Assistance Remove right side help
Check in all cover and See
locations shown troubleshooting Replace the
in section 1 for three
section 2 for help hydrophobic
leaks. No filters. See
figs.3.2 and
Proceed to 5.1 for
troubleshooting
locations
section 6 for help
It sounds like
you have a clog
in you waste line.
Remove right YES
side cover and
Problem
See trouble
solved?
shooting section
3 for help.
No
Call for further
The pictures in the proceeding troubleshooting section are of assistance
the FACScan. The sub-assemblies of the FACSort and
FACScalibur are functionally similar, however, some are
located in different locations.
Troubleshooting Section 1
Figures 1.1-1.4 show where leaks may be possibly coming from. Before disconnecting fittings make sure to de-pressurize the
sheath tank. Follow all general Lab safety precautions when working on instrument.
Fig. 1.1 Sheath/Waste Fluid Area Fig 1.2 Flow Cell Area**
Check around sheath Check male and female, Remove waste tank Check to see if waste
tank neck, and level white and orange quick and check for leaks. (black line) barb fitting Check sides of
sensor cap for air leaks disconnect fittings. Check Check around the is broken. viewing orifice for
(hissing sound). fittings to the right of waste seam area of the drips. If drips are
tank (dashed line). tank. present suspect
black line barb
fitting leaking or o-
rings on top and
bottom of viewing
orifice to be
defective.
** Follow
manufactures safety
precautions when
Check for viewing flow cell with
pooling optics cover removed
leaks here.
Check clear tubing For viewing ease, the sample injection probe (SIP) has been removed.
in pinch valves. When checking this area for leaks using a flashlight and inspection mirror
Both sides. will be of assistance.
Home
Troubleshooting Section 2
Fig 2.1 SIP (bottom view) Fig. 2.2 DCM pump
Differential pressure regulator DCM is the acronym for droplet
Step 3. Loosen one containment module.
Step 1. Remove C-clamp side of clamp with SIP is the acronym for sample
DCM tubing from thumb screw or injection probe
barb fitting and pull using hex key.
it gently enough to
Air
get syringe on the
pump
end.
Droplet moves
up the outer Droplet
sleeve of the SIP from
to the barb fitting pump to
waste Tubing
on the rear of clamp
flow cell.
In order to troubleshoot in this section you will need a syringe, DI water, 10% Fig. 2.3 Sheath/Waste Fluid Area
bleach, and a 3/32” hex key.
Home
Troubleshooting Section 3
Fig 3.1 Flow Cell Area**
Checking for clog in waste line Black line to waste during run
1. While performing drain and fills (prime for FCB), and fill. Pressurizes during
drain Remove here in
observe flow cell area below viewing orifice (fig 3.1), accordance with step 3.
and the waste line going to the waste tank (fig. 2.3). You should see
fluid drain from the flow cell to the waste tank during the Viewing
drain cycle (first half of the prime cycle). You should see fluid orifice
fill the flow cell and the waste line during the fill. Check that the rate
of fluid flow into the waste tank increases when the mode changes
from FILL(PRIME) to RUN. If you observe the above the problem
may be gone. Try running calibration particles. If the above is not
Check here to observe
observed proceed with step 2. flow cell draining and
2. Remove clear silicone tubing from pinch valve (fig 3.2) to see filling
if this corrects problem. If problem is corrected you may have Blue line to
waste during
a problem with the pinch valve or the electronics that control it. drain
3. If draining is not happening try removing the black line
(fig 3.1) from the top of the flow cell and check to see if it
is blowing air during the drain cycle. If the line is blowing
air check for clogs by following the blue line (fig 3.1) to the
orange line to the waste tank (fig 2.3). Chances are the clog
is happening at one of the quick disconnect fittings between
the flow cell and waste tank.
Fig 3.2 Fluidics Box (Facscan)
Remove Hydrophobic
silicone tubing filter for air line
from pinch
valve here
IAW step 2. Pressure transducer
Home
Troubleshooting Section 4 Fig. 4.1 Laser Shutter/Flow cell
The following steps were photographed on an instrument that had been fully decontaminated prior to refurbishment. Manufacturer recommends the
wearing of protective gloves while performing the following steps.
Fig. 5.2 Changing the Bal-seal (steps 1-5)
Step 1. Remove drip Step 2. Remove Bal-seal Step 3. Attach Bal-seal to SIP, Step 4. Bal-seal shown installed Step 5. Reinstall drip tube
tube and retaining nut (shown removed) Attach at an angle, and then in place and retaining nut
push lower side of bal-seal
into place
Home
Troubleshooting section 6
Fig 6.1 Cleaning the Sample injection Probe (SIP) with a cleaning wire
Home
Troubleshooting Section 7
Fig. 7.1 Exploded view of flow cell viewing orifice
Instrument is failing daily QC procedure
1. Before proceeding verify particles are not expired, sheath fluid is
and tubes are clean and you have tried a new batch of beads.
2. Instrument is failing FSC sensitivity test. This is caused by a
high signal to noise ratio detected by the FSC detector.
Solution: first perform a few drain and Fills (primes), recheck. If Exploded view of viewing
still failing try monthly cleaning procedure, recheck. If still failing orifice
perform flow cell flush with syringe and 50% bleach(see fig. 7.2).
Laser
3. Instrument failing SSC sensitivity test. This is caused by a
high signal to noise ratio detected by the SSC detector. 2nd Laser Flow
Solution: Bleed off excess air in saline filter (fig.4.2), recheck. If Channel
still failing perform monthly cleaning, recheck. If still failing
Check for
perform flow cell flush with syringe and 50% bleach(see fig. 7.2). Bubbles here
4. Instrument failing Fluorescence sensitivity test. This is
caused by a high signal to noise ratio detected by the FL(X )
detector. cells
Solution: If beads are good, check for bubble in flow cell (fig. 7.1
SIP clogged or clog in the waste lines (Sec. 3). If the flow cell is
bubble free, and the SIP and waste lines are not clogged suspect
poor optical alignment.
Before performing the steps in figure 7.2, drain flow cell (FACScalibur press prime, wait for flow cell to drain ,
Fig. 7.2 Flushing flow cell with syringe then put instrument in STBY), de-pressurize sheath tank, then turn instrument off.
Step 1. Remove optics cover Step 2. Remove the drain line (Black line) from
the barb fitting on top of the flow cell
Home
The following steps in Fig 7.2 were photographed on an instrument that had been fully decontaminated prior
to refurbishment. Manufacturer recommends the wearing of protective gloves and clothing while performing
Fig. 7.2 Flushing flow cell with syringe the following steps.
(Cont.)
Before performing the steps in figure 7.2, drain flow cell (FACScalibur press prime, wait for flow cell to drain ,
then put instrument in STBY), de-pressurize sheath tank, then turn instrument off.