Manual Usuario
Manual Usuario
HB-7021
Automated Hematology
Analyzer
Operation Manual
55
CATALOGUE
56
3.8Startup 21
3.11Calibration 24
3.16 Shutdown 29
57
4.7.7 Three-part Setting 40
4.7.8 Cleaning Time 41
4.7.9 Report Format Setting 41
4.7.10 Display Brightness 42
4.7.11 Landmark Setting 42
4.7.12 Time Setting 43
4.8 Prime 43
4.9 External Keypad 43
4.10 Tubing Clean 44
4.11 Shutdown 45
Chapter 6 Calibration 52
6.1 Pre-calibration 52
6.1.1Calibrator 52
6.1.2Background Test 53
6.1.3Evaluation of Repetition Precision 53
6.2Calibration Coefficient Modification 54
6.3 Calibration Review 55
Chapter 8 Maintenance 58
8.1 Daily Maintenance 59
8.1.1 Rinse 59
8.1.2 Shutdown Prime 59
8.2Weekly Maintenance 59
Chapter 9 Troubleshooting 61
9.1 Troubleshooting Guidance 61
58
9.2Seeking for Technical Assistance 62
9.3 Troubleshooting Disposal 63
9.3.1 WBC clog or RBC clog 63
9.3.2 WBC Data Change Abnormally 64
9.3.3 WBC BUBBLES or RBC BUBBLES 65
9.3.4 VACUUM is Low 66
9.3.5 WASTE is full 66
9.3.6 HGB Problem 66
9.3.7 DILUENT Empty 66
9.3.8 LYSE Empty 66
9.3.9 Detergent Empty 67
9.3.10 Time Error 68
9.3.11 No Paper in Recorder 68
9.3.12 Recorder Temp is high 68
9.3.13 Recorder Head Error 69
9.3.14 Recorder Error 69
9.3.15 Recorder off-line 69
9.3.16 Printer Paper Out 69
59
Copyright and declaration
Copyright©, SINNOWA Medical Science & Technology Co., LTD2005, all
rights reserved.
Congratulations to you for becoming the valued customers of SINNOWA Medical
Science & Technology Co., LTD. Welcome to the HB-7021 automated hematology
analyzer, it will bring you a new experience and convenience.
This statement is based on China's relevant laws, regulations and the specific
circumstances of the HB-7021 automated hematology analyzer designed by
SINNOWA Medical Science & Technology Co., LTD. It applies to the HB-7021
automated hematology analyzer whose installation and use in the People's Republic of
China (excluding Taiwan, Hong Kong and Macao) only.
The statement includes the latest information right before the brochures printed.
SINNOWA Medical Science & Technology Co., LTD takes full responsibility for
simplified Chinese version and the amendment of the statement, and reserves the right
of changing the contents without notice after printing the brochures.
The picture concerned in this manual is sketch, for reference purposes only, if the
picture does not match with real, in-kind basis.
All the information in the brochures is protected by copyright law. Without prior
written consent of SINNOWA Medical Science & Technology Co., LTD, not to the
content of any statement in any form of reproduction, copying, translating into other
languages.
The operator must operate strictly in accordance with the guidance of operating
manuals. Or SINNOWA Medical Science & Technology Co., LTD is not responsible
for any errors and equipment failure caused by illegal operation.
Limited responsibility for quality assurance:
1
SINNOWA Medical Science & Technology Co., LTD is not liable for the
following direct, indirect or eventually damage. Since the following cases happen,
SINNOWA Medical Science & Technology Co., LTD will not be responsible for
safety, reliability of the equipment. Quality assurance responsibility, the rights and
obligations of the after-sales service including the agreement produced and terminated
by it and all the power of free services agreed will be deemed to unconditionally give
up permanently at the same time.
① Equipment with non-normal use, without maintenance or has been damaged.
② The use of reagents and the annex which is non-recognition or not provided by
SINNOWA.
③ As users or others not in accordance with the <HB-7021 automated
hematology analyzer Manual> operation which led to misuse, negligence
caused damage to the equipment.
④ Replacement parts which is not allowed by SINNOWA Medical Science &
Technology Co., LTD or unauthorized personnel to equipment for
maintenance, repairs or alterations to equipment.
⑤ Components were dismantling, stretching, re-adjustment.
All hospitals or institutions use this equipment would cause equipment failure if
not achieve a standard maintenance plan.
SINNOWA Medical Science & Technology Co., LTD will be paid to provide
circuit at the request of the user, also list the calibration methods and other
information, to help users by qualified technical personnel to repair an equipment part
which is allowed to be repaired and maintained by users themselves.
Caution
SINNOWA Medical Science & Technology Co., LTD does not promise to any
implied warranties to make some kind of special-purpose equipment including its
marketability, the applicability.
We encourage you to call us when you encounter problems.
User Services Division opens user technology, services hotline to provide users
with technical support and troubleshooting.
If the equipment is failure, please keep the samples and to immediately notify the
authorized dealers of SINNOWA Medical Science & Technology Co., LTD or
directly contact with User Services Division of SINNOWA Medical Science &
Technology Co., LTD.
If the equipment failure can not be resolved by telephone guidance, SINNOWA
Medical Science & Technology Co., LTD will send a professional engineer or
regional service representatives to provide you with timely, comprehensive services.
SINNOWA Medical Science & Technology Co., LTD is referred to SINNOWA
for short hereinafter.
Production license:
Registration No.:
2
Standard ID: YZB/0511-2006HB Series of automated blood cell analyzer
Production enterprises: SINNOWA Medical Science & Technology Co., LTD
Production add: Qilin Industrial Park, Nanjing, Jiangsu, China (211135)
After-sales service units: SINNOWA Medical Science & Technology Co., LTD
TEL :( 025)84127188
FAX :( 025)84127188-2222
E-MAIL: [email protected]
ZIP: 211135
3
How to use this manual
This manual contains general information on the use of equipment; this is
information guidance to the new user to use instruments. Many users read brochures
all over before their first use. You will learn about the characteristics of the
instruments and methods of operation after you read the manual. You can use the
directory to quickly find the necessary information in the course of day-to-day use.
The brochures include equipment operation, maintenance instructions and
attention to issues, in order to maintain the performance of this equipment must be in
accordance with the specification of the equipment to operate, maintain.
Practice in this manual:
Warning: information that you should be aware of how to avoid the operator or others
suffered injuries
Cautions: information that you should be aware of how to avoid equipment damage
and lead to unreliable results of the analysis
4
Appendix 1: Symbol note
Equipotential
Refer to manual
5
Chapter 1 Instrument Introduction
1.2Measurement Parameters
6
RDW-CV the coefficient of variation of RBC distribution %
width
RDW-SD standard deviation of RBC distribution width fl
PLT total number of platelets
109cells/L
MPV The average size of platelet fL
PDW Platelet distribution width fL
PCT Platelet pressure plot %
排堵:Enforcement 排堵 procedures
7
Display
and
keyboar
d
Power
indicator
Add
needle
Printer
Figure 1-1
Printer: start recording Built-in devices to print the test results
Menu button: function menu button, menu function calls or out of the current
selection and returned to the main window of the blood cells.
Dilution key: liquor drainage key, discharge dilute solution from the sample needle,
mainly for the pre-diluted of the peripheral blood;
Delete key: delete the number and text messages
Cleaning key: the implementation of cleaning procedures
Enter key: Enter key, to confirm the content of choice
ESC: In any interface, you can click here to exit the current menu
Mode keys: working methods conversion keys, the whole blood and pre-dilution
mode conversion
↑↓←→ key: direction Key, you can move your cursor; select the menu or project
you need.
7. Display screen
5.5-inch LCD screen for liquid crystal display screen is divided into four regions, as
shown in Figure 1-2:
8
Current number
Current number Repeat Mode Time/Date Functional
mode
Results display
Figure 1-2
The district of the working status: show the "Please wait count" when counting,
counting completed, show the counting time of WBC, RBC; when equipment failure,
show fault alarm.
System time zone: showed that the current system date and time.
Current code area: showed that the ID number of the current testing results.
Results: show that the current test results, also give three histograms.
Model: shows that the current test mode.
Repeat: the implementation of retry, ID is not changed.
D earthing contact
9
c. SENSOR waste water sensor interface to connect waste water sensors.
d. equipotential ground terminals, used to connect the grounding of the hospital and
the system.
Instrument includes host, stylus printer (optional) and reagents three major
components.
1.5.1 Host
Completion of the blood sample collection and the whole process of analysis,
automatically displays test results after counting on the LCD screen. Host mainly
includes the following components:
Central board is the control center of equipment, it controls the following parts:
The switch of liquid solenoid valve, the suction and perfusion of reagent, and the
discharging of the waste.
The operation force pump and vacuum air pump provide the motion for suction and perfusion.
The control of the stepper motor is for inhalation and dilution of samples and adding
reagents;
Control A /D transform of WBC, RBC/PLT, HGB, to provide front-end service for
data processing; detect all of the actions of the photoelectric switch.
WBC survey unit is composed of hard wares such as the signal collection plate,
electroplate, mipor sensors, and liquid road.
Signal collection plate --- to provide a constant source for electrodes, will provide the
collected pulse signal with amplification and processing to CPU Central Processing
Unit.
Electrode --- WBC has two electrodes: one is internal, the other is external, the
internal one is fixed in the internal of WBC probe, the external is fixed in the external,
when the inner and outer electrodes immersed in the miscible liquid which is
composed by diluents, constant flow source on electrode to form galvanic circle
through Millipore sensors.
Millipore sensor --- WBC's Millipore sensors fixed in the front of the probe, its
Millipore diameter is of 100 microns. To take a sample survey, the particle samples
pass trough from the porous.
Road --- Use the negative pressure as the driving force, inhale the diluted liquid, fluid,
washing liquor and test samples from their own containers to measuring pipe, and
then discharge the measured waste liquid. In front of the WBC ;liquid road fixed the
10
hemolytic agent adding and blending part, through control of the stepper motor
control board, join the appropriate amount of hemolytic agent in the WBC samples
Cup, and then mix air by the compressed gas produced by the force pump.
RBC / PLT survey unit is composed of hard wares such as the signal collection plate,
electroplate mipor sensors, and liquid road:
Signal collection plate --- to provide a constant source for electrodes, will provide the
collected pulse signal with amplification and processing to CPU Central Processing
Unit.
Electrode --- RBC/PLT has two electrodes: one is internal, the other is external, the
internal one is fixed in the internal of RBC/PLT probe, the external is fixed in the
external, when the inner and outer electrodes immersed in the miscible liquid which is
composed by diluents, constant flow source on electrode to form galvanic circle
through Millipore sensors.
Millipore sensor --- RBC/PLT's Millipore sensors fixed in the front of the probe, its
Millipore diameter is of 100 microns. To take a sample survey, the particle samples
pass trough from the porous.
Road --- Use the negative pressure as the driving force; inhale the diluted liquid, fluid,
washing liquor and test samples from their own containers to measuring pipe, and
then discharge the measured waste liquid.
1.5.1.4 Tubing
Tubing is composed of solenoid pilot actuated valve, breaking valves, force pumps,
vacuum pumps, and plastic piping: solenoid pilot actuated valve: the one with feeler
microscope two-way or tee joint T to control the circulate of the liquid road.
Breaking valves --- control the flow of road by breaking the plastic tube; vacuum
pump - inhale dilute solution in the fluid reservoir;
Liquor drainage pump ---- inhale the waste liquid by counting into waste liquor pail.
Force pump --- to provide positive pressure impetus for reverse washing and
hemolytic blending.
Plastic tube - carrier of reagents, and the flow of waste water;
Vacuum chamber --- cause subpressure, and as a temporary waste water containers
after sample test complete.
1.5.1.5 Display
HB-7021 adopts the display of 5.5-inch LCD liquid crystal display, can show 19
parameters and three histograms. Details, see § 1.3 "instrument front panel."
1.5.2 Reagents
11
In order to maintain the good performance of equipment, SINNOWA configuration
on a special reagent for HB-7021. These reagents are took delivery inspection based
on product standards and tested all the pre-qualified, the nominal metrics are obtained
under the conditions that in the use of the reagents for HB-7021. Non-SINNOWA's
reagent will affect the performance of equipment, lead a serious measurement error,
and could lead to accidents.
All reagents must be kept at room temperature in order to maintain the ideal chemical
properties, in the preservation process should avoid the cold, overheating and direct
sunlight, when the temperature is below 0 ℃, the reagent will be frozen, the chemical
properties, electrical conductivity of reagents will be changed. In the process of the
use, in order to minimize the evaporation of reagents, reduce pollution outside to the
reagent, the reagent containers should be covered, the pipeline is put into reagents
through the cover of container. However, the mass of reagents with the time passing
will change, all reagents should be used in period of validity.when the reagent is
replaced, can not pour unspent reagents into the new reagent barrels (bottles), prevent
the new reagent from contamination.
1.5.2.1 Diluent
1.5.2.2Lyse
Lyse solution is a new type of reagent without double nitrogen and cyanide. To meet
the test requirements as follows:
a) quickly dissolve red blood cells, produce the matrix composites at least;
b) change the WBC epicyte, cytoplasm spread slowly, shrink the epicyte around the
nucleus, WBC was granular.
c) Transformation of hemoglobin to form hemoglobin complex, suitable for testing
under the optical wavelength of 540 nm;
d) non-cyanide, to avoid humans and the environment serious pollution
1.5.2.3Detergent
12
1.5.2.4 Probe Detergent
Probe cleaning fluid containing efficient oxides for cleaning equipment probe, mainly
used for solution of WBC, RBC probe intractable plugging holes.
Instrument equips with internal memory, can store not less than 200,000 samples test
results.
1.11 Accuracy
Measurements of WBC, RBC, PLT and HGB should be in the range of whole blood
control standard value. The range of error should be consistent with the requirements
of Table 1-3.
Table 1-3 Requirement of test results accuracy
13
Measurement Project Measurement accuracy tolerances
WBC ≤±4%
RBC ≤±3%
HGB ≤±3%
MCV ≤±3%
PLT ≤±8%
WBC
0×10 /L ~6.0×10 /L ±0.3×10 /L
9 9 9
RBC
0×1012/L ~0.99×1012/L ±0.05×1012/ L
HGB
0 g/L ~99 g/L ±2g/L
PLT
0×109/L ~99×109/L ±10×109/L
14
1.15 Duration of Service
Instruments used the classic method to measure cells and count. As shown in Figure
2-1, in the conductive liquid (mainly is dilute solution), provide constant current
source to the electrode so that the circuit can form a steady impedance loop. When the
cells through the porous, conductive fluid is replaced by cells, the circuit resistance
changes, cause electrical pulse, when the cells of different size through the porous,
their Vpp of electrical pulse are different, so that we can confirm the number and the
volume of porous cells through the amount and Vpp of electrical pulse.
15
Figure 2-1
As the number of pulses is corresponding with the cells which pass through the
porous, amplitude of pulse has the same size with the cells through the porous,
instrument measures each of the cells, and make size classification in accordance with
the size of them, equipment classify RBC, WBC and platelet in accordance with the
software pre-classification procedure, usually classified according to the specific size
of the following:
WBC 35-450 fL
RBC 30-100 fL
PLT 2-30 fL
The WBC which is dealt with hemolytic agent can be divided into lymphocytes (LY),
MO, neutrophilic granulocyte (GR) in accordance with its size, its interval of volume:
LY 35-98 fL
MO 98-135 fL
GR 135-450 fL
When added to blood samples of hemolysis, hemolytic agents can quickly destroy the
red blood cell membrane and combined with the hemoglobin as a complex which has
an absorption characteristic of a wavelength of 540 nm, through the compare between
the absorbance of dilution and the samples, the concentration of sample hemoglobin
can be calculated.
16
2.2 Approach of Equipment Parameters Measuring
Parameters from blood sample test are composed of three forms: one is the direct
parameters from the equipment measurement, such as the WBC, RBC, PLT, HGB,
MCV; another is the parameters according to the histogram, such as LY%, MO%,
GR%, HCT, RDW-CV, RDW-SD, MPV, PDW; the third is parameters derived from
certain formula, such as LY #, MO #, GR #, MCH, MCHC, PCT.
Formula derivation is as follows:
HCT(%)= RBC×MCV/10
MCH(pg)= 10×HGB/RBC
MCHC(g/L)= 100×HGB/HCT
PCT(%)=PLT×MPV/10
GR(%)= 100×AG/(AL+AM+AG)
Figure 2-2
AL: cells quantity within the lymphatic region;
AM: cells quantity within the monocytes cells region between lymphocyte cells and
granulocyte;
AG: cells quantity within the granulocyte region.
Calculation formula for absolute value of Lymphocyte LY #, monocyte MO #,
granulocyte GR # as follows:
17
Lymphocyte(109L) LY# = LY%×WBC/100
RBC distribution width of the coefficient of variation (RDW-CV) is derived from red
blood cells histogram, it said the volume of red blood cell volume distribution
coefficient of variation, and the unit is%.
RBC distribution width of standard deviation (RDW-SD) is derived from red blood
cells histogram, it said red blood cell changes in the size of standard deviation, and
the unit is fL.
Platelet distribution width (PDW) is derived from the platelet histogram, it said the
volume distribution of platelet groups.
a) Carefully takes the instrument and enclosure out from packing chest,
b) Check the annex according with the packing list inventory
c) check into whether there is an incoming-stream or water logging.
d) check whether there were any mechanical defects.
e) Check all the exposed wires, the inserted parts and annexes. If you have any
questions please immediately contact the company customer service agent or the
Department.
Please refer to Chapter 10 of the brochure “precaution, restrict and harm” 10.2.
18
WARNING: This equipment is not treatment equipment.
NOTE: Equipment working environment should avoid the cold and over-heated.
NOTE: Do not use mobile phones, cordless phones near the equipment, and so too
have the radiation field equipment. A strong radiation field would interfere with the
normal work of equipment.
Before installing equipment, have to check whether the power ground of the
instruments is be up to the mustard, power specific requirements are as follows.
Voltage: AC220 (1 ± 10%) V Frequency: (50 ± 1) Hz
Warning: The grounding-line after the instrument board must be grounded by
connecting directly with the ground. The user places the obligation to ensure that to
protect the reliability of power supply.
NOTE: inspection before equipment connected to power to ensure that all electrical
equipment connected is plugs right and reliable.
NOTE: the regular fluctuations in voltage would result in equipment performance and
reliability of the reduction, the user should resolve this issue before use the
instruments, such as the installation of the exchange regulator (to be equipped with
their own users).
NOTE: frequent disruptions will lead to electricity equipment performance and
reliability of the serious decrease, or even damage to equipment, users should use
instruments to resolve this issue before, such as the installation of an uninterruptible
power supply (UPS) (to be equipped with their own users).
The latter Host panel equipped of four road interface. Equipment manufacturers in
order to prevent pollution in the pipe line interface, installed the pipeline plug, it
should be disconnected from the interface in the first installation and carefully
conserved.
Removed from the reagent box with red joints of hemolysis (LYSE) catheters,
19
connectors connect with the road connecting latter host board labeled LYSE, the other
end of catheter insert into the hemolytic bottle, and screw the cap of hemolytic bottle.
Hemolysis bottle must be placed in the same plane with equipment.
Remove the DILUENT tube with blue faucet from reagent kit and attach it to the
connector marked “DILUENT” on the rear panel. Place the other end into the
DILUENT container. Twist the cap until secure. Place the container on the same level
as the analyzer.
Remove the WASTE tube with black faucet from reagent kit and attach it to the
connector marked “WASTE” on the rear panel, connect BNC plug with the socket
marked “SENSOR” on the rear panel. Twist the tube’s cap clockwise onto the waste
container until secure. Place the container on the level at least 50cm lower than the
analyzer.
Remove the DETERGENT tube with yellow faucet from reagent kit and attach it to
the connector marked “DETERGENT” on the rear panel. Place the other end into the
DETERGENT container. Twist the cap until secure. Place the detergent container on
the same level as the analyzer.
CAUTION:
Keep all the tubes loose and relaxed after installation, no contortion or folding.
CAUTION:
All the tubes should be installed manually, no not utilize any tool.
CAUTION:
If any damage or leakage occurs in the reagent container, or the reagents have
exceeded expiry date, it can not be used online, please contact SINNOWA agency or
suppliers for replacement, or contact with SINNOWA Customer Service.
WARNING:
The waste must be handled with biochemical or chemical methods before outlet to the
drainage, or it will cause contamination to the environment. Hospital and laboratory
have obligation to follow the environmental regulation of municipal government.
20
3.5 Recorder Paper Installation
Remove the printer from the shipping carton along with the manual, according to
Section3.1to inspect the printer carefully and perform the following procedures:
Make sure the power switch on the backboard is off (O), and connect the analyzer with the power
supply. The earth device on the rear panel should be well grounded.
NOTE:
The power cord should be connected to the special electric outlet.
3.8Startup
Turn on the power switch on the rear panel, then the indicator light on the front panel
will become orange, the analyzer start initialization, and automatically absorb diluent,
lyse and detergent, then rinse the tubing. The Blood Cell Analyze window appears
after initialization (See Figure 3-1). Under this window, next sample number could be
altered. Detailed as follows:
a) At the window as figure 3-1, shift ↑↓to select “Serial-number”, print ENTER,
window shown as figure 3-2 appears:
21
Number Mode
Section
Case Bed No.
office
Proofer
送检者 审 核 者
assessor
Figure 3-1
b) Shift ← → to adjust the position of cursor, input the serial number, press ENTER,
then edit the patient’s data. Press ENTER, the serial number is altered successfully,
and this serial number is that of next sample; if not want to alter serial number, select
ESC, return to the Blood Cell Analyze window, current serial number plus 1
automatically is that of sample.
Background test should be performed after startup and before blood sample test, the
procedures are as follows:
a) Present clean tube beneath the probe. At Blood Cell Analyze window, press
DILUENT to dispense the diluent into the tube.
b) Press FUNC in instrument surface, shift ↑↓, select “Serial number” menu, press
ENTER to modify the serial number to 0.
c) Put the tube beneath the probe; make sure the probe touch tube bottom lightly.
d) Press Run key on the front panel, the tube can be moved away only after hearing
“Di”, instrument start to counting and measure automatically.
e) After counting, background test result display on screen, the acceptable range of the
22
sensor clog, please consult Chapter 9 «Troubleshooting».
parameter background u
≤ 1
nit
WBC
0.3 ≤ 09/L1
RBC
0.03 ≤ 012/L1
PLT
10 ≤ 2 09/L
HGB g
≤
HCT /L %
0.5
If the test results are out of this range, repeat the above a) ~d) step until the test results
are within the expected range. Refer to Chapter 9 for “Troubleshooting” if the results
still do not meet performance expectation after 5 times’ repetition.
NOTE:
Serial number for background test is set to be 0 by the software. After set to 0, test
result will not be stored.
NOTE:
The serial number for blood sample test can NOT be set to 0.
23
Figure 3-11
Quality control should be performed before daily test or on the initial installation.
Please refer to Chapter 5 “Quality control”.
3.11Calibration
SINNOWA measure the analyzer severely at the factory before shipment. On the
initial installation, if the test results of background and quality control are normal,
recalibration is not necessary. If not and there are shifts or trends in some parameters,
recalibrate the analyzer referring to Chapter 6 “Calibration”.
24
CAUTION:
Consider all the clinical specimens, controls and calibrators etc, that probably contain
human blood or serum as being potentially infectious, wear standard laboratory
clothing, gloves and safety glasses and follow required laboratory or clinical
procedures when handling these materials.
CAUTION:
Blood collection and disposal should be performed according to the municipal
government or laboratory’s requirements.
NOTE:
Be sure the blood collection clean and contamination-free. Eligible gelation inhibitor is
required.
NOTE:
Do not shake the sampler violently.
NOTE:
Venous blood can only be stored for 4 hours at room temperature. SINNOWA
recommends the blood sample be kept at temperature between 2-8℃ for longer
storage, if the blood sample can not be disposed within short time.
Collect venous (whole) blood sample through vein-puncture and store in a clean
sample tube with EDTA-K2· 2H2O, which can keep the configuration of WBC, RBC
and control platelets aggregation, gently shake the tube 5~10 times to make it well
mixed.
25
the tissue fluid from mixing with blood, thus causes error in results.
CAUTION:
Must insure blood collection is 20ul when collect blood sample by fingertip puncture, to
avoid incorrect results due to blood volume insufficient.
OK ESC
Figure 3-2
Shift ← → to select “OK”, press ENTER, instrument returns to the Blood Cell
Analyze window, current operational mode switches to “pre-diluent” mode.
Method of switching from “pre-diluent” to “whole blood” mode is the same as stated
above.
Blood sample counting and analysis should be quickly done after collection, suggest
within 3-5 minutes.
■ Pre-diluted Mode
a) Present the empty clean tube beneath the probe. At Blood Cell Analyze window,
presses DILUENT; the diluent will be dispensed into the tube. Remove the tube.
b) Add 20ul blood sample in the constant volume blood collection tube to the
prepared tub filled with diluent quickly, and gently shake the tube to make them well
mixed.
c) Present the tube filled with blood sample under the probe (keep probe touch the
tube bottom lightly). Press Run key on the front panel and the working status indicator
becomes orange. Only remove the tube after buzzer gives “Di”.
d) Instrument starts analyzing the sample automatically, please wait for the result.
26
■ whole blood mode
a) Gently shake the tube to well mix the blood sample, then present the sample tube
under the probe (keep the probe at the bottom of tube). Press Run key and the
working status indicator in the front panel become orange. Only remove the tube after
buzzer gives “Di”.
b) Instrument starts analyzing the sample automatically, please wait for the result.
The test results will be displayed behind relevant parameter at the Blood Cell Analyze
Window after counting and analysis, and relative size histograms of WBC, RBC and
PLT will be drawn. (As Figure 3-1 shown).
If problems like clogs or bubbles occur during the counting and analysis procedures,
the analysis will be stopped automatically, the alarm rings and indication is given on
the screen, the test results are invalid. Troubleshooting please refer to Chapter 9
“Troubleshooting” for solution.
If the test results for parameters exceed the expected limits, an “L” or “H” will appear
in front of the parameter which appears in the screen and is printed out. “L” means
test result is lower than the lower limit while “H” means test result is higher than
upper limit.
If test result is ***, indicates that this data is invalid.
CAUTION:
Insure correctly set work mode “whole blood” and “Pre-diluent” before counting,
specific consult 3.13.
CAUTION:
Analyze in “Pre-diluent” mode, count the remaining sample again when necessary.
CAUTION:
Analyze in “Pre-diluent” mode, must use one-off tube and fix-capacity sample tube,
reusage will cause cross contamination and lead to incorrect result.
CAUTION:
According to the requirement of clinical medicine, blood cell does not distort, work
If some of the cells in the blood sample are immature, abnormal or untypical, the
warning indication is given in the form of histogram after blood sample test.
According to different abnormal situations, warning information such as R1, R2, R3,
R4, RM and PM will be displayed in the right of the histograms.
27
3.14.2.1 WBC Histogram Alarm
胞、异常淋巴细胞、冷凝球蛋白及脂类颗粒等。
RBC not lysed thoroughly, platelets aggregated, huge platelets, plasmodiums,
nucleolate RBC, abnormal lymphocytes, condensed globins and lipid particles, ect.
R2 :It indicates that the area between lymphocyte and monocyte is abnormal.
Probable cause:、原幼细胞等存在。
Allotypic.lymphocytes, abnormal lymphocytes, plasma cells, eosinophilic granulocyte
increased basophilic granulocytes, blasts and so on.
③R3:It indicates the area between monocyte and granular cell is abnormal.
⑤ RM: It indicates that in the sub area of WBC, many areas except for above areas
are abnormal.
When PLT histogram appears unusual, Pm alarm will appear in its right. Pm specific
meaning is:
28
Pm:It indicates that the broader area between platelet and RBC is abnormal.
Probable cause:The clotting of platelets, large platelets, small RBC, cell debris and
fibrin exist.
Inner thermal recorder and the outer printer are optional according to customers’
needs. The followings happen after blood sample analysis.
①If Auto Print is on, parameter language is Chinese; Recorder will print parameter
test report in Chinese.
②If Auto Print is off, press REC, blood cell analyze report of current sample can be
printed by inner recorder.
③If Auto Print is off, press PRINT, blood cell analyze report of current sample can
be printed by outer printer.
The modes of recorder, printer, and test reports are set up at Settings. Refer to section
4.5 in Chapter 4 for specific procedures.
CAUTION:
Do NOT pull the recorder paper violently but push it at an even pace(匀速的送出)
when the recorder is running, or it may cause damage to the recorder.
3.16 Shutdown
Shutdown procedure is performed after daily operation and before turning the
analyzer off. Daily maintenance and tubing-rinse is needed to ensure no protein
aggregation exist during non-working and keep sampling system clean. Specific
shutdown steps as follow:
b)Press FUNC, shift ↑↓, select “Shutdown “menu, press ENTER, “Shutdown
“window pop-up in the screen, as Figure 3-3 shown.
29
Figure 3-3
c) Press OK to turn off the instrument, present sample tube with detergent to the
probe (insure probe touch tube bottom lightly). Press Enter and the working status
indicator in the front panel will become orange. Remove the tube until hear “Di”.
d) Instrument begins cleaning and maintaining the tubing.
e) After turning off the instrument, “Thank you and now turn off power” will appear
to instruct the operator to turn off the switch on the rear panel.
f) Tidy the work platform and dispose waste liquor.
g) Press Esc if the operator does not want to turn off the analyzer, it will return to
User can inquiry the parameter and histogram of tested blood sample, and can print
out test data and histogram through outer printer. Details as follows:
30
a) At Blood Cell Analyze window, press FUNC, shift ↓to select “Data Inquiry”,
press ENTER, instrument enter into data inquiry window, as Figure 3-4 shown.
b) Press ↑↓to review the data backwards or forwards. The data of 5 samples at most
c) Press ← → on keyboard to select single sample, move cursor to the sample No.
press ENTER into Details Inquiry window to browse specific historical sample
data and print it out by recorder or printer.
d) Press FUNC, return to Test Result Inquiry window.
When the number of test samples reaches to certain quantity, data volume stored in
instrument is too large; it will take a long time to review the data since there are so
many pages. If necessary, all of the stored data can be deleted periodically by the
customers. Deletion is divided into auto deletion and manual deletion.
31
Figure 3-4
3.17.2.1Auto Deletion
When storage test data nearly reach 2,000, instrument will indicate “memory full,
please backup”, hereafter, user still can count for 50 times. After 50 times counting,
instrument will delete all the data automatically, so should backup the stored test data
after appear “memory full, please backup”.
Manual deletion is divided into single data deletion and all data deletion. Details as
follows:
All data deletion
a)At blood cell analyze window, press FUNC, shift ↑↓, select “Data review”, press
b)Confirm that cursor is at the last sample ID of first page; Press DILUENT, then
press DEL, OK, ESC and all deletion will appear at the lower left, as Figure 3-5
shown.
c) Shift ← →, select “OK,” all the stored historical data will be deleted, instrument
will return to Blood Cell Analyze window, select “ESC” back to Test Result Inquiry
window.
32
Figure 3-5
a) At blood cell analyze window, press FUNC, shift ↑↓, select “Data Inquiry”,
press ENTER into Test Result Inquiry window, as figure 3-4 shown.
c)Select “OK” to delete the last data; Select “ESC” to return to Test Result Inquiry
window.
NOTE:
After all data deletion, one historical data still displays at Blood Cell Analyze
Window, which can’t be inquired at Test Result Inquiry window. This data will be
deleted automatically when a new data is stored.
NOTE:
Single data deletion, only can delete from the last one.
NOTE:
When “memory full, please backup” is indicated in the instrument, stored test data
should be backed up immediately.
33
Chapter 4 FUNC introduction
At Blood Cell Analyze window, press FUNC into function menu, as 4-1 shown.
1 Maintenance 6 Calibrations
menu
5 Normal
range setting
10 Shutdown
Note :Operation introduced in this chapter will change some run actions of the
instrument, please confirm the necessity of these changes.
User can press “OK” to enter into this menu, so as to do "daily maintenance", "weekly
maintenance", "yearly maintenance" for the instrument and pipe cleaning and
emptying. As follow 4-2 shown:
34
Daily maintenance Circle of the white
blood cells pipeline
Weekly maintenance
Emptying of the red
blood cells
Yearly maintenance
Circle of the red blood
cells pipeline
Emptying of the
white blood cells Sole timing
pipeline
Figure 4-2
4.2 Flush
User can kick back the jewel hole under this window, so as to discharge blockage.
Attention: it is forbidden that changing the blood sample number into 0, because if the number is 0,
the result will not be kept.
Users can inquiry the tested data, parameter and histogram of blood sample, and by the
internally installed grapher recording meter or external connection printer, the data and
parameter can be outputted. If necessary, users can delete selected test parameters or
all test parameters. Specific operations consult 3.17 “Query Result Inquiry”
The analytic system in the instrument gives 19 upper and lower limits of parameters.
If the tested data range exceeds the setting value range, system will mark “H” or “L,
H” besides the tested data, so as to indicate that it exceeds the upper limit of the
parameters. If “L”, it means it exceeds the lower limit of that. Specific details consult
chapter 7 “Parameter Boundary”. Users can set the normal ranges of people, children
and animals. Attention: the software default it as “ordinary person”. If you want to
choose other ranges, enter into it, and then ESC, it will be confirmed. As 4-3 shown:
35
Ordinary person
Horse
Children
Others
Girl
Others
Dog
Others
Figure 4-3
4.6 Calibrations
At window in Figure 4-1, select “System setting”, press ENTER, system setting window appears,
see figure 7-1.
7 Three–part
1 Printer
setting
2 Measuring
unit 8 Cleaning time
36
3 Hospital name 9 Report form
setting
5 Display 11 Landmark
language setting
Figure 7-1
6 Measurement
12 Time setting
gain
Users can enter this menu to choose self-contained printer or off-line one, auto print.
At system setting window as figure 7-1, shift ↑↓ to select Auto print, press ENTER,
and shift ← →to select ON or OFF, press ENTER.
Select ON, after counting, test report will be printed by recorder (printer)
automatically.
Users can alter instruments time, specific methods as follows:
a) At system setting window as figure 7-1, shift ↑↓ to select “Time setting”.
b) Press ENTER, cursor will move to date automatically. You can choose date format
as “YY-MM-DD”, “MM-DD-YY” and “DD-MM-YY”.
c) Press ← → to control the position of cursor. Input 0-9, you can alter Y, M, D and
time.
d) After alteration, press ENTER, System setting window will come back.
At system setting window as figure 4-7, shift ↑↓ to select “Alarm”, press ENTER, and
shift ← →to select ON or OFF, press ENTER.
At system setting window as figure 4-3, shift ↑↓ to select “Key tone”, press ENTER,
and shift ← →to select ON or OFF, press ENTER.
Select ON, you will hear bumming when press keyboard; select OFF, no bumming.
Users can enter this menu to amend or increase hospital names in the printed report.
First, at system setting window as figure 4-1, shift ↑↓ to select “System setting”, press
ENTER, enter figure 7-1, choose “hospital name” with the direction key, then press
“OK”, see figure 7-2
37
Hospital name:
ENTER
Figure 7-3
Note: At this window, you can only input Chinese or English. After input, press
ENTER to conserve it.
用户在此菜单下可以选择缺稀释液,缺溶血素,缺清洗剂报警???
Users can choose alarm of lacking diluent, lyse and detergent.
At system setting window as figure 4-7, shift ↑↓ to select “Alarm”, press ENTER, and
shift ← →to select ON or OFF, press ENTER.
Select “ON”, after counting, when reagent is not enough, there is letter twinkling hint
on top right corner of screen, at the same time; buzzer gives “Di” for alarm.
Select OFF, after counting, when reagent is not enough, there is no hint.
First, choose “Liquid level sensor” with the direction key, press “OK”, enter figure
7-3. You can choose all or certain ones. First, choose that with the direction key, then
pitch on it with pressing “OK”
Diluent alarm
Hemolysin alarm
Figure 7-4
Parameter language
At system setting window as figure 4-7, shift ↑↓ to select “Display language”, press
38
ENTER, shift ← →to choose “Chinese” or “English”, and press ENTER.
Choose “Chinese”, parameter at the blood cell analyze window will display Chinese,
as figure 4-5 shown. The printed report is Chinese format.
Figure 4-5
Choose “English”, parameter at the blood cell analyze window will display English,
as figure 3-1 shown. The printed report is English format.
First, at system setting window as figure 4-1, shift ↑↓ to select “System setting”, enter
figure 7-1, then select “Measurement gain”, enter figure 7-4. Users can magnify white
cell or red cell one time or two times according to need.
39
WC gain
RC gain
1.25
1.5
2.0
图7-6
40
a, press → on the keyboard, , you can shift the position of cursor.
b, press the number key on the keyboard, you can input number.
c, “." key on the keyboard is used for letters input: single-click to input A,
Figure 7-8
41
Report format 1 with Report format 1 without
histogram histogram
Users can set minimal and maximal cells of WBC RBC PLT which participated in
counting originally here.
First, at system setting window as figure 4-1 shown, shift ↑ ↓to select “system
setting”, press ENTER, enter into figure 7-1. In 7-1 interface, shift ↑ ↓ to select
“landmark setting”, press ENTER to enter it, as figure7-11 shown. WBC_L: minimal
WBC volume participated in counting; WBC_L: maximal WBC volume participated
in counting; LY-MO:Delimitation landmark between lymphocyte and neutrophile cell;
MO-GR: Delimitation landmark between neutrophile cell and granular cell; RBC_L:
42
WBC-L 25FL WBC-H 400FL
98FL 110FL
LY-MO MO-GR
30FL 150FL
RBC-L RBC-H
Users can set or alter time and date or date format here.
First, at system setting window as figure 4-1 shown, shift ↑ ↓to select “system setting”,
press ENTER, and enter into figure 7-1. In 7-1 interface, shift ↑ ↓ to select “time
setting”, press ENTER to enter to select “time format ” or “time setting”, as
figure7-12 shown.
Time format is divided into three types: YMD, DMY and MDY.
Time format
Time setting
Figure 7-12
NOTE:
After the instrument goes into auto dormancy, press any key on the keyboard to esc
dormancy.
4.8 Prime
Prime and rinse the whole tubing to keep the whole tubing full of diluents.
“Diluent prime”, “Lyse prime” and “Detergent prime” can be selected.
Users can select “External keyboard”, then press ENTER to change it into PS2.
43
4.10 Tubing Clean
Clean the obstinate stem in probe with the probe detergent. Specific as follow:
a) Put about 4 ml probe detergent to the clean test tube.
b) At the Blood Cell Analyze window, press FUNC into function menu window.
c) At function menu window, shift ↑ ↓to select “Tubing clean”, press ENTER,
Figure 4-6
d) Put the tube with probe detergent beneath sample probe, shift → to move cursor to
“OK”, press ENTER, analyzer inhale probe detergent automatically into WBC and
RBC probe cup separately, then soak for 6 minutes. Then progress bar will appear on
the screen to show the progress, number on progress bar descending with course, as
Figure 4-7 shown. Instrument will clean tubing automatically after 6 minutes,
schedule bar disappear, back to Blood Cell Analyze window.
Figure 4-7
e) If users want to exit “Tubing clean” program in midway, Can press FUNC after
the number on progress bar descending to 5, “OK” and “ESC” will appear. Select
“OK”, number on progress bar turns to 1, instrument clean the tubing automatically,
44
after progress bar disappears, back to Blood Cell Analyze window; select “ESC”,
instrument continues soaking the tube.
4.11 Shutdown
Shutdown procedure must be performed after daily operation and before turning the
analyzer off. Instrument does daily maintenance and tubing-rinse to avoid protein
aggregation in tubing during non-working and keep sampling system clean.
。
Specific step consults §3.16《shutdown》
NOTE:
Wrong operation on turnoff procedure is prohibited, nor will it decrease reliability and
performance of the analyzer. Any problems derived from that will NOT be guaranteed
free by SINNOWA.
NOTE:
Wrong turnoff leads to system data loss easily and causes system operating failure.
NOTE:
Wrong turnoff will cause instrument does not Prime and clean the tubing, easily lead
to albumen aggradations of blood sample in tubing, cause clog.
5.1 QC Material
45
Because of the QC material produced by different manufacturers, there should be
variations for their mixture and treatment. To ensure the result accuracy, users should
pay attention to the following matters when dealing with QC material:
a) Stored under the manufacturer's recommended storage conditions. It's better to
place the QC material in the central part of the refrigerator. Do not leave it on
refrigerator door or other places near the door, as these areas represent the extremes of
temperature within the refrigerator.
b) Check the status of QC material and ensure there is no crack and breakage on it's
container.
c) Be sure to follow the user's guide of QC material. Mix and warm it up slowly.
d) Check the expiry date of QC material. The use of any expired product is
prohibited.
e) Any opened QC material over the manufacturer's recommended time is
prohibited.
f) Intensive heat or vigorous shake of QC material is prohibited;
g) A new batch of high, medium and low levels of QC materials is used and each of
them should be tested for three times. Compare the data obtained with the last batch
and check variance of the test results.
CAUTION:
As all human derived products including clinical specimen, control and calibrator
should be treated as a potential biohazard, all staff must wear appropriate personal
protective equipment (eg, protective clothing, gloves, and goggles) and follow safe
laboratory practices for dealing with biological materials.
5.2 Edit
46
1 Run QC 6 Relative QC Chart
3 QC Levels
4 QC Data
5 Absolute QC Chart
Fig. 5-1
b) Select "QC Groups" and press "confirm" to access the submenu. Press "→" to
select group 1,2 or 3. After selection, press "confirm" to return to QC screen.
c) Move the cursor to "QC Levels" by pressing "↑"/"↓" and press "confirm" to access
the submenu. Press "→" to select level high, normal or low. After selection, press
"confirm" to return to QC screen.
d) Move the cursor to "QC Values Setup” by pressing "↑"/"↓" and press "confirm" to
access the submenu (fig. 5-2) for QC parameters setup. Move the cursor to "Lot" item
and input QC material batch number by pressing numerical key on the keyboard.
e) Press "→" to move the cursor to "QC material reference values". Input QC material
reference values by pressing numerical key on the keyboard.
f) Press "→" to move the cursor to "QC material reference ranges". Input QC material
reference ranges by pressing numerical key on the keyboard.
47
Fig. 5-2
g) While edit is completed, press "menu" button and two icons of "confirm" and
"cancel" appear at the bottom of the screen. Press "→" to select.
h) Return to QC screen for other QC operations.
Fig. 5-3
48
NOTE:
If there is a illegal data input for one parameter, the reference value and range of
which is a invalid data and will not be displayed. User should correct it. Otherwise,
only effective parameters can be monitored.
5.3 Run QC
6 Relative QC Chart
2 QC Groups
3 QC Levels
7 QC Values Setup
4 QC Data
49
Fig 5-5
f) While the counting process is completed, return to the window shown in fig 5-4.
Press "Menu" button to return to the QC window.
g) If blood cell counting is not required, Press "Menu" button after the step d (see fig
5-4) and return to QC window for other operations (see fig 5-1).
NOTE:
If total number of QC data reaches 31, a tip of “Memory full, Please backup” appears
in the window. User should enter the QC data window and backup stored QC data.
After deleting all QC data, run the QC program again.
5.4 QC Chart
Assigned values (target value and accepted range) as well as averages, standard
deviations (SD) and coefficients of variation (CV) are calculated based on the QC
measurements and are displayed together with LEVEY-JENNINGS chart which is a
Gaussian distribution. On the x-axis the number of the control run is plotted. A mark
is made indicating how far off the actual result was from the mean. The central line
represents the mean which is the expected value for the control. The dotted line on
either side of the mean represents the +2 and –2 SD which is the reference range for
the control. The high, medium and low levels of LEVEY-JENNINGS charts are
displayed together. Each LEVEY-JENNINGS chart contains 31 QC data at most.
Select "QC data" and access the QC data window. The current LEVEY-JENNINGS
chart can be sent to an external printer by pressing "Print" button.
5.5 QC Data
QC input data and output results are auto-stored in the memory in chronological order
and they can be retrieved at any time. Users can check the system precision and
accuracy by comparing the actual QC results with the reference values and ranges.
The detailed operation process shows below:
a)From the blood cell analysis window, press the "menu" button and then "↑"/"↓" to
50
access quality control item. Press "confirm" to execute to display the QC window as
shown in fig.5-1.
b)Move the cursor to "QC Data" by pressing "↑"/"↓" and press "confirm" to access the
QC Data Inquiry window (see fig 5-6).
Fig. 5-6
The current items can be sent to an external printer by pressing "print" button.
When the QC data reached a large amount, it is time-consuming work for user to
search one certain result by browsing many pages. Therefore, if required, users are
advised to delete useless data routinely.
5.5.2.1 Auto-Delete
If total number of QC data reaches 31, a tip of “Memory full, Please backup” appears
in the window. while the next QC measurement is completed and regard as the first
one, all previous QC data will be deleted automatically.
51
5.5.2.2 Manual-Delete
Chapter 6 Calibration
During daily operation, excursion may occur gradually in results for many reasons, so
it is necessary to recalibrate certain parameters. To ensure the analyzer’s precision and
obtain reliable test results, the parameters (WBC, RBC, PLT, HGB, and MCV) should
be calibrated in the following situations:
a)Working environment changes greatly.
b)One or multiple parameters’ test results are of excursion.
c)Any major component that could affect the measurement is replaced.
d)Requirement of the clinic or the laboratory.
e)The reagent has been replaced.
MCV, HCT parameters are relative, thus one can be obtained from given value of the
other. Only MCV will be calibrated by the analyzer.
WARNING:
Before make all the parameters calibration accurate, the tested data should not be used
in clinic or laboratory.
6.1 Pre-calibration
6.1.1Calibrator
52
the calibration.
WARNING:
Consider all the clinical specimens, controls and calibrators etc, that contain human
blood or serum as being potentially infectious, wear standard laboratory clothing,
gloves and safety glasses and follow required laboratory or clinical procedures when
handling these materials.
CAUTION:
Only calibrators recommended by SINNOWA can be used to
CAUTION:
Follow the recommendations provide by calibrator-maker to store the calibrators.
CAUTION:
Check if the container is broken or cracked before using the calibrator.
CAUTION:
Make sure the calibrators are brought to room temperature and well mixed slowly
before use.
CAUTION:
Make sure the calibrators are within the expiry date.
6.1.2Background Test
53
ITEM
ERROR(%)
WBC ≤±4
RBC ≤±3
HGB ≤±3
MCV ≤±3
PLT ≤±8
NOTE:
When whole blood and capillary blood should be used, each of them should be
calibrated; the calibration should not be performed until blood-collecting method is
confirmed. Details consult item 1.13 of Chapter 1.
CAUTION:
Measurement should be performed under the same blood-collecting method when the
method is confirmed.
CAUTION:
If any malfunction occurs during measurement, the test results are invalid. Repeat the
measurement after troubleshooting.
a) At Blood Cell Analyze window, press FUNC, shift ↑↓to select Calibration; press
ENTER into Calibration window, as shown in Figure 6-1. Choose “Auto
calibration” or “Manual calibration”. If choose “Auto calibration”, target value should
be firstly input, then test it for 4 times with the normal liquor, new factors will come
out automatically.
1 Manual calibration
2 Auto calibration
Figure 6-1
If choose Manual calibration, operate as follow:
54
WBC 100% RBC 100% HGB 100%
parameters in turn.
e) Press OK after input, until cursor select ENTER, by now select ESC and OK.
Select ESC to cancel current calibration result, select OK, instrument back to Blood
Cell Analyze window after storing current calibration result.
NOTE:
The calibration coefficient is allowed in the range of 70%~130%, if the test new
calibration coefficient exceeds the limit, the maximum and minimum of coefficient
range should be selected as new coefficient for calibration.
NOTE:
The analyzer calibrates not only single parameter (WBC, RBC, HGB, MCV .HCT
and PLT), but also all the parameters.
NOTE:
Pre-calibration and calibration should be performed with the same blood-collecting
method.
NOTE:
If any malfunction hint occurs during measurement, the test results are invalid. Repeat
the measurement after troubleshooting.
At Blood Cell Analyze window, press FUNC, shift ↑ ↓ to select calibration into
calibration window, as figure 6-2 shown. Instrument goes into Review state.
55
Information or indication is given if the test values exceed the range. The limits of 19
parameters are given in this instrument. Any results exceeding the range will be
marked “H” or “L” besides the test results. “H” means the results are higher than the
upper limits, while “L” means the results are lower than the lower limits.
NOTE:
Parameter limit is inner setting of instrument, is the important reference gist for clinic
diagnosis.
The shift in parameter limit may cause changes in abnormal indication of hematology
index. Please confirm the necessity for changing.
Detail step as follow:
a) At Blood Cell Analyze window, press FUNC into FUNC menu window.
b) Shift ↑↓to select Para setting; press ENTER into Para limit window, as shown
in figure 7-1.
Children Chicken
Female Others
Dog Others
Horse Others
56
c) Choose your needed ranges, press “OK” to enter, as shown in figure 7-2.
Figure 7-2
d) Shift ← →to select the parameter that need to set, press 0~9 to input its upper and lower limit
value.
57
9.0 14.0
Figure 7-3
Chapter 8 Maintenance
Routine care and regular maintenance are essential to keep the best status and reach
the designed precision, to minimize system problems, as well as to prolong the life
span. Procedures and instructions for preventive maintenance are discussed in this
chapter. More information is available at SINNOWA Customer Support Centre.
Preventive maintenance should be performed daily, weekly, monthly and yearly,
according to maintenance requests in use process. Pertinent maintenance is also
included according to actual requirement.
WARNING:
Analyzer failure will occur unless a normative maintenance criterion is performed
strictly in all hospitals or institutions which use the instrument.
The operator must wear powder-free gloves when performing the maintenance
procedures. If powder-free gloves are not available, rinse the gloves before
58
performing the maintenance procedures. Powder from the gloves may cause analyzer
problems.
Instrument will prime automatically to rinse on every startup, the whole process is
about 2 minutes. In order to reduce the percentage of plugged hole, automatic prime
is performed when the instrument in running. Automatic rinse can be performed
every two hours according to the user’s need; “automatic maintenance” will appear
in the screen. The instrument will rinse automatically if the count exceeds 25 times
within two hours. Also can perform artificial control prime during using the
instrument, details as follow:
Press FUNC into FUNC menu window, shift ↑↓, and select Prime at FUNC menu window, press
ENTER, instrument will prime to rinse whole tubing system. Prime directly if eliminate tubing
bubble, that is using diluents to rinse.
若进行消除微孔堵塞和仪器除菌的灌注冲洗,可用探头清洗液执行“液路清洁”程序。Use
Must perform this procedure before turn off the instrument, specific method for
making probe cup full refer to section 3.16 of Chapter 3.
NOTE:
Wrong turnoff or not turnoff, instrument will not prime the tubing, easily lead to
albumen aggradations of blood sample in tubing that cause clog.
8.2Weekly Maintenance
Clean the smudge on the surface, especially the spilt-blood on the sampler, to prevent
the protein from deposition, molding or contaminating. Wipe the outside of the
sampler with gauze soaked by neutral detergent before cleaning other parts.
CAUTION:
Never use corrosive acids, alkali or volatile organic solvent (such as acetone, aether
and chloroforms) to wipe the outside of the analyzer, but only neutral cleaner.
59
8.2.2 Tubing Maintenance
Do use detergent to rinse measure tubing once a week to ensure there is no albumen
aggradation in tubing. Details as follow:
a) Take out blue connecter of diluent pipe from mainframe.
b) At Blood Cell Analyze window, press FUNC into function menu window, shift
↑↓, select Prime, then press ENTER.
c) Repeat step b) until top full corner of screen alarm DILUENT EMPTY.
d) Take out diluent pipe from diluent barrel, repeat step b), until DILUENT EMPTY
disappears.
e) 20 minutes later, repeat step a)-c).
f) Connect blue connecter of diluent pipe to corresponding position on mainframe, the
other end still in diluent barrel, perform Prime three times
CAUTION:
As executing step (f), users should wash the pipe with distilled water before inserting the
pipe into the diluent bucket, invoiding the contamination of diluent by the residual
detergent.
Rinse WBC and RBC probe once every week at least, details consult section 4.9
Tubing Clean of Chapter 4
Good yearly maintenance will keep the analyzer in best status and prolong the
lifespan. Because of the strict requirements, the maintenance should be performed by
the engineer authorized by SINNOWA. Please contact customer service office of
SINNOWA before the yearly maintenance.
Inactivity or shipping for three months because of various causes, please do as follows:
a) Remove the diluent tubing with a blue connector on the rear panel from the diluent
container, and empty the liquid downward.
b) Remove the hemolytic agent tubing with a red connector on the rear panel from the
diluent container, and empty the liquid downward.
60
c) Cap the diluent and hemolytic agent containers and keep them according to
instruction. Efficient action should be taken to prevent the materials from
deterioration, misapplication and misusing. Reagents should be free from temperature
extremes.
d) Hang in the air the diluent and hemolytic agent tubing.
e)Perform Prime for several times until “No Diluent”, “No Hemolytic agent” and
“No washing liquid “appears on top right corner.
f) Insert the diluent and hemolytic agent tubing into the distilled water.
g)Perform Prime for several times until “No Diluent”, “No Hemolytic agent” and
“No washing liquid “disappears on top right corner.
h) Press RUN on the front panel at Blood Cell Analyze, press RUN again after
measurement.
i) Remove the diluent and lyse tubing and rinse them with distilled water. Dry them in
j) Perform Prime for several times until “No Diluent”, “No Hemolytic agent” and
off the power” appears, then turn off the power switch on the rear panel.
l) Pluck the waste tubing, and rinse it with distilled water, then dry it in shady place
m)Install the caps which are removed in the initial installation into the tubing faucet
n) Pluck the power cord, and pack it in a plastic bag after cleaning.
o) Pack the analyzer and parts in plastic bag and put them into the carton.
Chapter 9 Troubleshooting
This chapter describes the method for instrument troubleshooting. If the failure can
not be eliminated in accordance to this chapter or more technical files are required,
please contact SINNOWA custom service division.
61
The troubleshooting guide is designed for supporting users to find and clear faults
during the operation of the instrument, as well as providing the information for users
about how to receive the timely technical support and help from SINNOWA Customer
Service.
Excellent problem cleaning and eliminating skills root from deep understanding
about the instrument and experience accumulated day by day.
In order to find and eliminate problems accurately and quickly, users should read this
manual and know instrument normal operation and contingency maintenance.
Generally, Logical troubleshooting may be divided into three steps:
1. Problem Identification
2. Problem classification
3. Trouble clearing
Step 1: Problem Identification means not only identifying what is wrong but also
what is right. The investigation should identify the problem area and eliminate areas
that are right. Once this done, the trouble shooting process moves quickly to next step.
Step 2: Problem classification Analyzer problems are generally divided into three
categories:
1. Hardware component related
2. Software computer programs related
3. Measurement related to sample analysis
Hardware and software problems can only be corrected by a SINNOWA authorized
engineer. The operator can correct sample measurement problems with assistance
from SINNOWA engineers.
Step 3: Trouble clearing Maintenance engineers can take appropriate action to
correct the problem. If the operator can correct the problem, with or without technical
assistance from SINNOWA, waste time will be decreased sharply.
If assistance from SINNOWA is needed when the instrument goes wrong, please
calling the SINNOWA Customer Service or contact with the agency. The phone and
fax number is shown in preface. Detailed and clear problem description and related
data must be provided when seeking for technical assistance. Details as follow:
a) The analyzer model
b) Serial number and version number
c) Detailed and clear description of problem phenomenon and operation surroundings.
(As what is done with what window and status)
d) The lot numbers of the reagents (lyse, diluent and detergent)
e) Related data and report.
Familiar problems list and disposals are given in this Chapter. The operator can
identify the cause according to the warning information and operate according to the
troubleshooting provided in the list.
62
9.3 Troubleshooting Disposal
Familiar problems and disposals are listed as follows. If the problems can not be
corrected, or technical assistance is needed, please call SINNOWA Customer Service.
If the counting time exceeds upper limit during measurement, the alarm rings and
warning information is given: “WBC CLOG” or “RBC CLOG”.
Disposal ways and steps:
a) Press any key in the front panel to stop the alarm.
b) At Blood Cell Analyze window, select “Maintenance”, then select “Daily
Maintenance”, then makes the probe in the bottle which is filled with concentrate
abluent, press “OK”.
c) If the above-mentioned are not applicable to serious clog, perform as follow step.
d) Add approximately 4ml probe cleaner into clean tube.
e) At Blood Cell Analyze window, press FUNC into Function menu, shift ↑↓, select
“Daily Maintenance”.
f) Put the tube with probe detergent beneath sample probe (insure sample probe touch
bottom lightly), press ENTER. In case above method still can not settle, solve it by
adding probe detergent manually.
g)Perform “Tubing clean” with probe detergent.
Figure 9-1
b) As Figure 9-1 shows, open the left-side door.
c) After sample adding organ stop running, suck the probe detergent with the injector,
63
as figure 9-2 shows, inject 3ml probe detergent into corresponding probe cup which
appear clog, probe cup position as figure 9-2 shows.
Figure 9-2
If above still cannot settle, please call SINNOWA Customer Service or contact with
the agency.
NOTE:
Personnel unauthorized or untrained by SINNOWA can not perform above operation, any
error and instrument malfunction caused by incorrect operation, SINNOWA Company will
not take on any responsibility.
NOTE:
Please do self-protection when perform above operation.
Abnormal sample;
abnormal lyse;
Lyse can not be added
in ;
a) Replace blood sample (or using medium controls instead) and perform the
measurement again.
b) If problem can not be corrected, remove the lye tubing with red connector on
rear panel from the lyse, empty the tubing and hang it up.
c) At Blood Cell Analyze window, press FUNC, shift ↑↓, select Prime several
times until appear “No Lyse” in the screen right corner.
64
d) Replace lyse.
e) Perform step c) several times until No Lyse disappear.
f)As for disposing the failure of lyse addition, please refer to 9.3.8 problem disposal
methods.
9.3.2.2 Data abnormally change
Probable causes: Filter under the probe cup is jammed.
Disposal methods:
a) As figure 9-1 shows, open the left-side door.
b) There are two filters under the probe cup at right down corner, as Figure 9-3 shows.
Figure 9-3
d) Open the connection pipeline on two ends of filter under the WBC probe cup, take
out filter, examine whether there is jam in filter, and replace it if there is.
If above still cannot be settled, please call SINNOWA Customer Service or contact
with the agency.
NOTE:
Personnel unauthorized or untrained by SINNOWA can not perform above operation, any
error and instrument malfunction caused by incorrect operation, SINNOWA Company will
not take on any responsibility.
NOTE:
Please do self-protection when perform above operation.
The warning information appears when the air enters WBC or RBC tubing.
Required actions:
a) Press MUTE in the front panel to stop the alarm.
b) At Blood Cell Analyze window, press FUNC, shift ↑↓to select Prime, press
ENTER.
65
9.3.4 VACUUM is Low
The warning information of low vacuum appears if the analyzer can’t supply rating
negative voltage within the set time.
Required actions:
a) Press MUTE in the front panel to stop the alarm.
b) At Blood Cell Analyze window, press FUNC, shift ↑↓to select Prime, press
ENTER.
If above still cannot be settled, please call SINNOWA Customer Service or contact
with the agency.
Required action:
a) Press MUTE in the front panel to stop the alarm.
b) After perform “Prime”, perform the measurement again.
Probable cause: No diluent or diluent prime pump can not establish negative pressure
Required action:
a) Replacing diluent
b) At Blood Cell Analyze window, press FUNC, shift ↑↓to select Prime, press
ENTER.
If above still cannot be settled, please call SINNOWA Customer Service or contact with the
agency.
Probable cause:
No lyse
Lyse sensor is dirty
Correlative pipeline is conglutination together due to long-playing extrusion.
Required actions:
a) Press MUTE in the front panel to stop the alarm.
b) Replace the lyse
66
c) At Blood Cell Analyze window, press FUNC, shift ↑↓to select Prime, press
ENTER.
a) As figure 9-1 shows, open the left-side door.
b) Find solenoid valve V14 behind the vacuum chamber as Figure 9-6 shows.
If above method can not solve, maybe Correlative pipeline is conglutinated together,
please settle it according to follow step.
Figure 9-6
c) If there are lyse in pipeline, instrument still alarm, it indicates sensor is dirty; you
should get rid of dust inside and outside of the sensor.
d) If lyse can not be added in, maybe pipeline is conglutinated together, you can take
out pipeline in solenoid valves which near valve board, disjoin the conglutination
together part, change another position, then repress the pipeline into the solenoid
valve.
If above still cannot settle, please call SINNOWA Customer Service.
NOTE:
Personnel unauthorized or untrained by SINNOWA can not perform above operation, any
error and instrument malfunction caused by incorrect operation, SINNOWA Company will
not take on any responsibility.
NOTE:
Please do self-protection when perform above operation.
Required action:
a) Replace detergent.
67
b) At blood cell analyze window, press Select Prime, press
FUNC, shift ENTER
↑ ↓
b) At Blood Cell Analyze window, press FUNC, shift ↑↓to select Prime, press
ENTER.
Required action:
a) Press MUTE in the front panel to stop the alarm.
b) At Blood Cell Analyze window, shift ↑↓to select System Setting, shift ↑↓to select
Date Setting
c) Press ENTER, the cursor moves to Date automatically, date format is
MM-DD-YY.
d) Press → ←to control cursor; change time into current time by input 0-9.
If above still cannot settle, please call SINNOWA Customer Service.
WARNING:
Qualified recorder paper is required; otherwise unqualified recorder paper will lead to
recorder failure, inferior record quality or thermal head damage.
CAUTION:
Gently replace the paper to avoid impacting the thermal head.
CAUTION:
Do no open the recorder cover but for paper replacement or problem correction.
68
Probable cause: Thermal head is overheated.
Required action:
Please call SINNOWA Customer Service or contact with the agency.
Probable cause: No paper in printer or install printer without following the correct
step.
Required action:Install print paper following the method offer in instruction; correct
printer installation step consult §3.6.
Improper operation will never attain optimal performance; even cause damage to the
69
operator or others. So to avoid the damage and get a successful measurement, a
criterion should be designed to perfect the service conditions.
by SINNOWA, if any problems derived from that, SINNOWA will not offer free
warranty.
e) Follow the recommended maintenance schedules and procedures as outlined in
Chapter 8. Any incompliance will shorten the life span and affect the test results, or
cause incidents, thus lose the free warranty.
70
tubing is not twisted and the waste tubing is connected to the appropriate outlet and
routed to a suitable waste container or drain.
h) Do not disconnect any electrical connection while the power is ON. Verify the
analyzer is well connected with the ground to prevent electrical interfere and ensure
the safety.
CAUTION:
Anyone without authorization of SINNOWA should NOT remove the screws on the
cover, or the customer must take all the responsibility.
reagent pollution.
e) Follow the manual to store reagent, calibrators and controls. The customer have
obligation to take actions and management to prevent the reagent, calibrators and
controls from deterioration, misapplication or eating by mistake. The reagent should
be away from temperature extremes.
CAUTION:
Reagent will freeze when it is below 0℃, for which the reagent can not be used.
CAUTION:
Keep the reagents away from direct sunlight to avoid evaporation and contamination.
Seal the cap of the container. Minimize the diameter of the hole to avoid evaporation
and contamination.
71