Honeywell Smartline Calibration Report & Certificate of Conformance

The document provides a calibration report and certificate of conformance for a Honeywell Smartline pressure transmitter. It details the customer information, serial number, calibration results showing the applied pressures, outputs, and errors, and certifies that the instrument meets Honeywell requirements and was calibrated using NIST traceable equipment.

Uploaded by

abhishek ingale
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
640 views1 page

Honeywell Smartline Calibration Report & Certificate of Conformance

The document provides a calibration report and certificate of conformance for a Honeywell Smartline pressure transmitter. It details the customer information, serial number, calibration results showing the applied pressures, outputs, and errors, and certifies that the instrument meets Honeywell requirements and was calibrated using NIST traceable equipment.

Uploaded by

abhishek ingale
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

24-May-2021 Honeywell Smartline Calibration Report & Certificate of Conformance Honeywell

Model Number: STD725-E1AC4AS-1-J-CHT-11C-B-11A0-F1-0000

Customer: Artee Flow Controls Pvt Ltd


Customer Tag: Customer PO No: FID2122-2122000004
Serial Number: C4000040816433 Prom ID: S970492SJ0H1 Firmware. 118
Meter Body ID: S970492SJ0H Station: Pune-RDC-5 Slot: 6

Result: PASSED

Custom Cal Range: 0.000 to 100.000 mbar Tolerance +/-: 0.0088 mA

Applied Pressure Output Error Expected Actual Error


Pass/Fail
% Span mbar mbar mbar Output (mA) Output (mA) (mA)
100.00 100.000 100.005 0.005 20.0000 20.0008 0.0008 Pass
75.00 75.000 75.018 0.018 16.0000 16.0028 0.0028 Pass
50.00 50.000 50.018 0.018 12.0000 12.0028 0.0028 Pass
25.00 25.000 25.008 0.008 8.0000 8.0012 0.0012 Pass
0 0.000 0.002 0.002 4.0000 4.0004 0.0004 Pass

Transmitter Configuration: LRV to URV: 0.000 to 100.000 mbar

Comm Mode: Analog/HART Damping: 0.50 Sec

Failsafe Mode: High Conformity: Linear

We certify that this instrument has been tested in accordance with and meets all applicable Honeywell
requirements, and has been calibrated on equipment with accuracy traceable to the National Institute of
Standards and Technology (NIST).

Standards Traceability and Calibration Due Dates:


Pressure Standard: PPC4-2b Sr.No.1468 07-Jan-2022
Voltmeter Standard: K2002-5Sr.No.4129981 10-Jan-2022
Transmitter Calibrated by: SANDEEP K.

Honeywell Automation India Limited, Village Fulgaon, Tal-Haveli, 412216 PUNE, India F-3399 04/2017

Note:- This is computer generated document and does not require signature.