93300
Vital Signs Monitor
5/ 5i / 5c/ 7/ 7i/ 7c/ 5 elite / 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite
Summary
The élance 5 / 5i / 5c / 7 / 7i / 7c /5 elite/ 5i elite / 5c elite / 7 elite / 7i elite /7c elite are integrated vital signs monitors with
widescreen resistive touchscreen display. The monitors raise a new standard of style, convenience, economy, ease of
use and reliability in patient monitors. Weighing not more than 2.4kg for the élance 5 series and not more than 2.7kg for
the élance 7 series, including an up to 5-hour lithium-ion battery, they are one of the lightest full-featured vital signs
monitors. With its 7cm depth, the élance vital signs monitors fit easily into even the most crowded bedside
environments.
www.spacelabshealthcare.com 061-2181-00 Rev J
93300
Vital Signs Monitor
5/ 5i / 5c/ 7/ 7i/ 7c/ 5 elite / 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite
Features
Parameters Supported 2 LCD sizes are available:
Standard parameters include 12.1 inch Display: 10.2 inch Display:
Electrocardiogram (ECG), Respiration,
Resolution 1280 x 800 pixels Resolution 800 x 480 pixels
Non-Invasive Blood Pressure (NIBP),
Pulse Oximetry (SPO2), Dual Size (Horizontal) 262 mm (10.3 inches) Size (Horizontal) 222 mm (8.7 inches)
Temperature. Additional parameters (Vertical) 164 mm (6.5 inches) (Vertical) 134 mm (5.3 inches)
such as Dual Invasive Blood Pressure
and Capnography are also available
depending on the model configurations.
User Interface All controls are on-
screen touch keys, with the exception
of ON/OFF key. A tethered stylus is
also available for function navigation.
Display Widescreen, color thin film
transitive (TFT) liquid crystal display Sweep Speed: A variety of speeds
(LCD) with resistive touchscreen. are available under parameter control.
Dimensions Parameters and Waveforms Power An internal power supply
provides power to the monitor
12.1 inch Display: The élance 5 will display up to four
Width 291 mm (11.5 inches) waveforms and up to six numerics at
Height 221 mm (8.7 inches) the display zones. Advanced Power Management
A built-in lithium-ion battery maximizes
Depth 75 mm (3.0 inches) All the other élance 5 model battery performance for up to 5 hours
Weight 2.7kg (6.0 lbs) configurations (élance 5i and 5c) will during transport.
display up to five waveforms and up to
seven numerics at the display zones. Connectors RJ45 (10/100 Base T
10.2 inch Display:
Width 252 mm (9.9 inches) All élance 7 model configurations will Ethernet), USB (for software updates,
Height 193 mm (7.6 inches) display up to six waveforms and up to keyboard/mouse interface) and RJ11
seven numerics at the display zones (for nurse alert)
Depth 75 mm (3.0 inches)
Weight 2.4kg (5.3 lbs)
Trends Displays up to 120 hours Languages Supported
of tabular trend data. Save at 1 English, Spanish, Portuguese, Italian,
minute trend record time period. French, German, Russian, Chinese
simple, Chinese traditional.
Time interval of 1, 5, 10, 15, 30, 60,
120-minutes
Options
Ethernet (Option P) Printing Method Paper Speed 25 mm per second
Provides Central surveillance capability Thermal array print head
from any monitored patient on the Alarm Record Records any
network. Print Resolution 8 dots per mm parameters in an alarm state when the
10/100 Base T connector (RJ45) (vertical) and 32 dots per mm alarm recording is active
provided (horizontal) at 25 mm per second
sweep speed Indicators Paper out
Recorder (Option U)
Dual channel recording capability via Prints Manual and automatic alarm Record Enables selection of up to
external Recorder recordings for waveforms, vital signs two active monitor channels plus trends
data and trends
www.spacelabshealthcare.com 061-2181-00 Rev J
93300
Vital Signs Monitor
5/ 5i / 5c/ 7/ 7i/ 7c/ 5 elite / 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite
Product Specifications
Refer to the specific parameter section for the appropriate specifications.
ECG
Detects cardiac QRS complexes and ventricular fibrillation; evaluated using ANSI/AAMI EC57:1998/(R)2003
with AHA, CU, MIT, and NST databases.
Input 5-lead or 3-lead ECG cable Signal Width 70ms to 120ms (Q to S) Display Bandwidth
Extended 0.05 to 100Hz
Input Dynamic Range Input Impedance ≥5 M at 10 Hz
5 mV AC, 300mV DC Filter 0.5 to 30Hz
Display Sweep Speeds 6.25, 12.5, Monitor 0.5 to 40 Hz
Voltage Range +/-0.5 to +/-5 mV 25.0 mm/sec
Defibrillator Discharge Recovery Heart Rate Alarm Limits Channel Crosstalk
5 seconds <2.5 mm p-p per AAMI EC13
High 20 to 300 bpm
Sample Rate Low 15 to 295 bpm Common Mode Rejection Ratio
250 samples per second (sps) <1 mm, reference to input (rti) per
AAMI EC13
Pacer Detection
Heart Rate Range 15 to 300 bpm Detects pacer pulses of ±2 mV to ±700
mV with pulse widths of 0.25 to 2 msec Baseline Stability
Accuracy and rise times 10% of width not to 0.5 mV
±1% or ±2 bpm (whichever is greater) exceed 100 microsec.
Calibration Signal
Heart Rate Resolution 1 bpm Pacer Rejection 1 mV p-p 10%
Single and double pulse pacers with
less than 4 msec tails.
ST Segment Analysis (Option T)
Resolution 0.08 mm Displays 5-lead numerics
Trends
Threshold Range 0.25-8 mm Display Bandwidth Up to 120 hours of trend data.
(1mV=10mm) ST 0.05 to 40 Hz
Alarms Single lead
Arrhythmia Analysis (Option A)
Standard Leads Sample Rate 250 sps Arrhythmia Analysis
I, II, III, V, AVR, AVL,AVF Asystole, VTach, VFib, PVC,
Sensitivity 5mm/mV Couplets, Pacer non-capture, high/
Input Channels low rate violation
Simultaneous acquisition of all
standard leads
www.spacelabshealthcare.com 061-2181-00 Rev J
93300
Vital Signs Monitor
5/ 5i / 5c/ 7/ 7i/ 7c/ 5 elite / 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite
Respiration
Input 5-lead or 3-lead ECG cable Respiration Resolution Accuracy ±3 breath per minute
1 breath per minute
Measurement Technique Numeric Update Rate
Trans-thoracic Impedance Respiration Rate Alarm Limits Every 1 second or immediately at
the onset of an alarm
High 6 to 120 breaths per minute
Respiration Rate Range
0 to 120 breaths per minute Low 5 to 119 breaths per minute Waveform Sweep Speeds
6.25, 12.5 and 25.0 mm/sec
Detection Sensitivity
0.5 at 500 input source impedance.
NIBP
Measurement Method Oscillometric Pulse Rate 30 to 240 bpm Measurement Time
30 seconds (typical); < 135
Resolution 1 mmHg seconds (maximum)
Measurement Units mmHg
Measurement Ranges Automatic Cuff Deflation
Measurement Mode Automatic or Manual
Systolic Pressure Range Measurement time exceeding
Automatic Measurement Intervals Adjustable at 30 to 250 mmHg 135 sec in adult/pediatric or
intervals of 3, 4, 5, 10, 15, 30, 60, 90, 120 or 240 maximum pressure value
minutes; exceeding 300 mmHg
Diastolic Pressure Range
Blood Pressure Measurement Accuracy 10 to 210 mmHg
Overpressure Protector
Satisfies ANSI/AAMI SP10: 2002 Mean Pressure Range 290 mmHg at normal condition
mean error standard deviation 20 to 230 mmHg 300 +/- 10 mmHg at single fault
systolic +1.7 mmHg 6.3 mmHg
Measurement Time condition
diastolic - 3.2 mmHg 6.8 mmHg
30 seconds (typical); < 135 seconds
(maximum)
Temperature
Probe Type Thermistor probe YSI 400 Accuracy Resolution 0.1º C (±0.2º F)
The maximum permissible error of
Units ºC or ºF a complete thermometer is ±0.2º C Numeric Update Rate Every 3 seconds
(±0.4º F) in the temperature range
Measurement Range from 5.0º C to 50.0º C (41.0º F to Alarm Limits
5.0º C to 50.0º C (41.0º F to 122.0º F) 122.0º F)
High 32.1º to 42.0º C (89.7º to 107.6º F)
Display Accuracy ±0.1º C plus probe
Display Parameters Low 32.0º to 41.9º C (89.6º to 107.5º F)
accuracy
T1, T2, and delta temperature (DT)
(two probes attached) Probe Accuracy ±0.1º C (±0.2º F)
www.spacelabshealthcare.com 061-2181-00 Rev J
93300
Vital Signs Monitor
5/ 5i / 5c/ 7/ 7i/ 7c/ 5 elite / 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite
SPO2
Spacelabs SpO2 (Option S)
Measurement Method
Functional saturation (oxygen Saturation Resolution 1% Pulse Rate Resolution 1 bpm
saturation of functional hemoglobins) Measurement Accuracy
Saturation Alarm Limits
Measurement Range O2 Saturation High 51% to 100%
Low 50% to 99%
O2 Saturation 30% to 100% Without Motion 70 to 100% ±2%
1 to 69% unspecified
Pulse Rate 30 to 250 bpm Numeric Update Every 3 seconds
Low Perfusion 70% to 100% ±2%
Display Sweep Speed 1% to 69 unspecified ®
TruLink Sensors
6.25, 12.5, 25.0 mm/sec Pulse Rate Operate at or near 660 nm and 940 nm;
total radiated optical power from 500 to
Sample Rate 50sps Without Motion: 30 to 250BPM ±3BPM 1,000 nm does not exceed 60 mW
Low Perfusion: 30 to 250BPM ±3 BPM
Masimo SET SpO2 (Option M)
Measurement Method Pulse Rate Resolution 1 bpm Saturation Alarm limits
Functional saturation (oxygen High 51% to 100%
saturation of functional hemoglobins) Pulse Rate Accuracy Low 50% to 99%
No Motion ±3 bpm
Measurement Range
Motion ±5 bpm
O2 Saturation 1% to 100% Masimo Sensors
Low Perfusion ±3 bpm Operate at or near 660 nm and 905 nm;
Pulse Rate 25 to 240 bpm
total radiated power from 500 nm to 1000
Numeric Update Every 3 seconds nm does not exceed 0.79 mW.
Saturation Resolution 1%
Measurement Accuracy (Arms) * The accuracy specification under motion These sensors have been clinically
conditions is ±3%. Motion is defined as validated by Masimo against the
Saturation Accuracy
Sensor Weight 70 to 100%
continuous rubbing and tapping motions at Masimo MS-11 oximetry board.
Models Range 2 to 4Hz, at amplitude of 1 to 2 cm, and
No Motion Low
Perfusi continuous random frequency motion
on† between 1 to 5Hz, at amplitude of 2 to 3
LNCS Reusable Sensors cm. No Implied License
LNCS DC-I >30 kg ±2% ±2%
Possession or purchase of this
** These sensors were not validated under device does not convey any express
motion conditions.
LNCS DC-IP 10 to 50 kg ±2% ±2%
or implied license to use the device
LNCS TC-I >30 kg ±3.5% ±3.5% † Pulse amplitude >0.20%; % transmission with unauthorized sensors or cables
>5% (LNOP Y-I sensor was not validated that would alone, or in combination
LNCS TF-I >30 kg ±2% ±2%
for low perfusion).
LNCS Adhesive Sensors with this device, fall within the scope
LNCS Adtx >30 kg ±2% ±2% †† Saturation accuracy under no motion for of the Masimo patent rights.
LNCS Pdtx 10 to 50 kg ±2% ±2% neonatal, infant, or pediatric patients with
congenital cyanotic cardiac lesions ±3% for
80 to 100%, ±4% for 60 to 80%, and ±3.3%
for 70 to 100%.
www.spacelabshealthcare.com 061-2181-00 Rev J
93300
Vital Signs Monitor
5/ 5i / 5c/ 7/ 7i/ 7c/ 5 elite / 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite
Nellcor OxiMax SpO2 (Option N)
Measurement Method Measurement Accuracy (Arms)
Functional saturation (oxygen * The accuracy specification under
saturation of functional motion conditions is ±3%.
hemoglobins) Saturation Accuracy
Sensor Models ** The accuracy specification has
70% to 100% been determined between
Measurement Range saturations of 80% and 100%.
O2 Saturation 1% to 100% OxiMax Sensor Models, Single Patient Use
† The MAX-N and the OxiCliq N were
tested on patients >40 kg
Pulse Rate 25 to 300 bpm MAX-A* MAX-AL* ±2%
MAX-N*† (Adult) ±2% These sensors have been clinically
Saturation Resolution 1%
validated by Nellc or against the
MAX-P* ±2%
Pulse Rate Resolution 1 bpm Nellcor NELL-1 oximetry board.
MAX-FAST ±2%
Pulse Rate Accuracy MAX-R** ±3.5% Nellcor Sensors
25 – 250 bpm ±3 bpm Operate at or near 660 nm and 880
OxiCliq Sensor Models, Single Patient Use
nm; total radiated optical power from
OxiCliq A ±2.5% 500 to 1,000 nm does not exceed 15
Saturation Alarm Limits mW
High 51% to 100% OxiCliq P ±2.5%
Low 50% to 99% OxiCliq N† (Adult) ±2.5%
Reusable Sensor Models
Numeric Update
Every 3 seconds DS-100A ±3%
OXI- AN(Adult) ±3%
Invasive Blood Pressure
(élance 5i / 5c / 7i / 7c / 5i elite / 5c elite / 7i elite / 7c elite)
Transducer Type Measurement Range Display Parameters
Strain-gauge, standardized to 5 -50 to +300 mmHg Systolic, diastolic, and mean pressures
μV/mmHg/V (±1%) displayed for arterial, pulmonary artery,
Accuracy umbilical artery, umbilical venous, and
±2 mmHg or ±2 % generic pressure; mean pressures
Transducer Excitation Voltage (whichever is greater)
5.00 VDC ±1% displayed for all others
Filter Frequency 12Hz Labels Arterial (ART), Central
Dynamic Waveform Range
-10 to +300 mmHg Venous (CVP), Left Atrial (LAP),
Alarm Limits Pulmonary Artery (PA), Right Atrial
High -5 to +250 mmHg (RAP), Umbilical Artery (UA), Umbilical
Signal Bandwidth 0 to 40 Hz
Low -10 to +245 mmHg Venous (UV), and Generic Pressure
(PRS)
Sample Rate 50sps Waveform Sweep Speeds
6.25, 12.5, 25.0 mm/sec Numeric Update Rate
Measurement Units mmHg Every 3 seconds
www.spacelabshealthcare.com 061-2181-00 Rev J
93300
Vital Signs Monitor
5/ 5i / 5c/ 7/ 7i/ 7c/ 5 elite / 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite
EtCO2
(élance 5c / 7c / 5c elite / 7c elite)
Parameter Displayed Alarm Limit Ranges EtCO2 Accuracy
CO2 waveform, EtCO2, FiCO2
EtCO2 0 to 38 mmHg, ± 2 mmHg
CO2 Measurement Range High 7 mmHg to 80 mmHg
0.8 kPa to 10.6 kPa 39 to 99 mmHg,
EtCO2 0 mmHg to 99 mmHg ± (5% of reading + 0.08% for every 1
0 kPa to 13.2 kPa Low 5 mmHg to 60 mmHg mmHg above 38 mmHg)
0.6 kPa to 7.9 kPa
Units mmHg, kPa
FiCO2 0 mmHg to 24 mmHg Total System Response Time
Display Sweep Speeds 0 kPa to 3.2 kPa < 5 seconds
6.25, 12.5 and 25.0 mm/sec
Classification
MDD Class IIb
EN 60601-1
Class 1 Requires outlet with safety ground (Protective Earth) conductor Rated for continuous operation
CISPR11, Group 1, Class A Suitable for use in all establishments other than domestic that are connected to a low supply
network
Electrical Specifications
Mains Power Fuses Isolation EN 60601-1
AC Mains, 100 - 240 V, 50 - 60 Hz, 2- Qty 2, T1.0 A, 250 volts,
1A IEC (5×20 mm) Compliance
91/157/EEC Battery Declaration
Internal Power Supply Battery Directive
30W Medical Grade A single Li-ion battery provides up to 5
hours
Environmental Requirements
Storage Operating
Temperature −20 º to 60 º C Temperature 0º to 40º C
(−4º to 140º F) (32º to 104º F)
Humidity 15% to 95% Humidity 15% to 95%
(non-condensing) (non-condensing)
Altitude Up to 12,200 meters Altitude Up to 4,572 meters
(Up to 40,000 feet) (Up to 15,000 feet)
www.spacelabshealthcare.com 061-2181-00 Rev J
93300
Vital Signs Monitor
5/ 5i / 5c/ 7/ 7i/ 7c/ 5 elite / 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite
Accessories
External Recorder, two-channel Mounting Kits
A variety of mounting (wall mounts, P/N 016-0794-01
tabletop mounts) and mobility P/N 016-0795-01
solutions (roll stands, bedrail hook) 10.2” Tabletop mount, Black
are available from Spacelabs
Healthcare. P/N 016-0798-00
P/N 016-0799-00
P/N 016-0796-00 12.1” Tabletop mount, White
P/N 685-0254-00 P/N 016-0796-01
élance Recorder P/N 016-0797-00 P/N 016-0798-01
P/N 016-0797-01 P/N 016-0799-01
P/N 685-0254-01 Wall mount 12.1” Tabletop mount, Black
élance elite Recorder P/N 016-0791-00
P/N 016-0794-00
P/N 016-0795-00 Roll stand
10.2” Tabletop mount, White
Documentation
élance Vital Signs Monitor Documentation CD-ROM (P/N 084-1963-00)
Regulatory Approvals
Meets EN 60601-1. CE marked in accordance with the Medical Device Directive 93/42/EEC.
élance and TruLink are trademarks of Spacelabs Healthcare .
Brands and product names are trademarks of their respective owners
www.spacelabshealthcare.com
061-2181-00 Rev J