BRO Mass Comparators e PDF
BRO Mass Comparators e PDF
Page
Automatic Mass Comparators and Robots
Determination of Mass to the Very Highest Standards 4
From the 1 kg Prototype to Mass Standards 5
Robot for Fully Automated Determination of Mass from 1 mg to 1 kg 6
Automatic Mass Comparators with Load Alternator 8
Accessories
Susceptometer – A Complete Solution for Testing the Magnetic Properties
of Weights 17
Volume and Density Determination up to 1 kg 18
Density Determination up to 50 kg using the YP50K Pycnometer 19
Software for Mass Metrology – One Software
Program for All Equipment ScalesNet32 20
Software for the Comparison of Mass – YPR02C 21
Accessories for Mass Metrology 22
Technical Specifications
1 kg Prototype Mass Comparator and Robot 24
Mass Comparators with Load Alternator 25
Mass Comparators up to 1 kg 26
Mass Comparators from 1 kg to 10 kg 27
Mass Comparators from 10 kg to 60 kg 28
Mass Comparators from 60 kg to 300 kg 29
Mass Comparators from 600 kg to 3000 kg 30
Determination of Volume, Density, Susceptibility and Magnetism 31
3
Automatic Mass Comparators and Robots
Determination of Mass to the Very Highest Standards
Sartorius masters the core disciplines of The unique three-point holder is capable The driving force of these motors is trans-
weighing like no other and sets new of handling both cylindrical weights and ferred into the interior of the vacuum
standards in mass metrology. In coopera- silicon spheres efficiently. chamber by rotary axes and vacuum cou-
tion with the Bureau International des plings. The load alternator is driven by a
Poids et Mesures* and the Institute for The equipment thus meets all of the pulley hoist system.
Process Measurement and Sensor Technol- requirements for the new definition of
ogy of the Technical University of Ilmenau, the kilogram using a silicon sphere. All components inside the vacuum chamber
Sartorius has developed a mass comparator The scientific tests necessary to determine are completely resistant to high vacuum
– the CCL1007 – that is capable of deter- the Avogadro constant can therefore be and, therefore, will not outgas under
mining differences in mass to an accuracy performed conveniently and precisely. vacuum. This prevents any contamination
of 0.1 μg for weights of 1 kg – even under of the weights. All materials used in the
high-vacuum conditions. The weigh cell is located in the upper vacuum chamber are solid materials with
section of the vacuum chamber. The a low vapor pressure. They are free of
The mass comparator is protected from weighing pan for the substitution weights oil and grease and thus ideally suited to
environmental effects by a pressure-stable is accessed via the upper loading hatch. high-vacuum applications.
aluminum chamber. In the vacuum cham-
ber, measurements can be performed at The motors that drive the load alternator The more than 20 vacuum flanges enable
high vacuum up to 10-6 mbar. Unlike con- and the weight changing device are located additional sensors to be introduced into
ventional stainless steel vacuum bell jars, outside the vacuum chamber. This reduces the vacuum chamber, which has sufficient
the aluminum chamber does not cause any interference caused by temperature space to accommodate additional
magnetic reactions that would affect the fluctuations. measuring instruments.
weighing system.
4
From the 1 kg Prototype to Mass Standards
Areas of Application
– Use as a 1 kg prototype weighing system
for the subdivision of the mass of the
primary reference standards of National
Metrology Institutes (NMIs) with national
mass reference standards (kilogram
prototype)
s* The standard deviation "s" is the repeatability calculated from 6 ABBA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
5
Robot for Fully Automated Determination of Mass
from 1 mg to 50 kg
Robot systems with latest weighing The comb-type weight grabbers are
technology guarantee highest accuracy designed so that any desired types of
for the mass determination. weight – wire or leaf weights or knob,
cylindrical or disk weights – can be handled
The fully automated mass comparison efficiently.
simplifies work in the mass laboratory.
This enables complete sets of weights The number of magazine positions is
to be measured efficiently within the specified by the customer. Up to
shortest time possible. 164 magazine positions can be provided
Comb-type weight grabber with 4 weights for the range of 1 mg to 1 kg.
Reference and test weights are positioned
in the magazine that accommodates The robot systems are supplied with
a high number of weights. The weighing a PC and user-friendly control software.
mode is entered on the PC, the number of
comparison cycles is defined and a direct Naturally, all Sartorius robot systems can
1:1 mass comparison or subdivision (dis- be controlled directly by ScalesNet 32 (see
semination of mass for class E1 weights) YSN01C on pages 20-21). This is a software
is initiated. The robot retrieves the weights program designed to cover all requirements
from the magazine and starts mass com- and to network all equipment in the mass
parison. Prior to mass comparison, the standards laboratory.
robot automatically combines the groups
of weights required for dissemination of
mass according to the weighing design.
6
Technical Specifications CCR10|CCR1000|CCR10-1000
CCR10 CCR1000 CCR10-1000
Application range (E1) 1 mg – 10 g 10 g – 1 kg 1 mg–1 kg
Maximum capacity 10.5 g 1002 g 10.5 g|1002 g
Readability 0.1 μg 1 μg 0.1 μg|1 μg
Repeatability, s* < 0.3 μg 0-100 g: < 3 μg < 0.3 μg|0–100 g : < 3 μg
< 8 μg – |< 8 μg
Repeatability (typical), s* < 0.2 μg < 2 μg < 0.2 μg|< 2 μg
Linearity 1 μg 20 μg 1 μg|20 μg
Electronic weighing range 3.5 g 2.1 g 3.5 g|2.1 g
Magazine positions 39 21 60
Optional magazine positions 26-65 2-39 2–104
s* The standard deviation "s" is the repeatability calculated from 6 ABBA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
7
Automatic Mass Comparators with Load Alternator
8
Mass comparators with a load alternator The integrated centering equipment is
are ideally suited for determining the mass positioned below the weighing pan to give
of class E1 weights. the user the best possible access.
The generously sized weighing pan pro- These mass comparators can also be used
vides sufficient space for the calibration as stand-alone instruments without a PC.
of groups of weights and large-diameter Networking and PC control are possible
weights. by using external programs.
CCE10000S-L
Maximum capacity: 10.05 kg
Electronic weighing range: 60 g
Readability: 0.1 mg
Repeatability (typical), s*: 0.05 mg
CCE20000S-L
Maximum capacity: 20.05 kg
Electronic weighing range: 60 g
Readability: 0.1 mg
Repeatability (typical), s*: 0.1 mg
s* The standard deviation "s" is the repeatability calculated from 6 ABBA cycles after elimination of drift. Prerequisites for accurate calculation
of the standard deviation are good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
9
Manual Mass Comparators up to 1kg
The new CCE generation of manually- All manual CCE comparators have a fully Below-balance weighing equipment in
operated weighing equipment is the electronic weighing range or several Sartorius mass comparators makes density
ultimate in manual mass comparators. overlapping weighing ranges. As a result, determination easy. This equipment is
any weight value between 0 g and the standard on nearly all mass comparators in
With up to 61 million weighing steps, these maximum capacity can be attained. this product range.
instruments are world class. The monolithic
weigh cells that are manufactured The CCE products have an electronic In addition to the below-balance weighing
as a single structure from a high-quality draft shield and below-balance weighing equipment, Sartorius mass comparators are
aluminum alloy deliver results with equipment. also available with the YDK01 density
outstanding repeatability. determination kit.
s* The standard deviation "s" is the repeatability calculated from 6 ABA cycles after elimination of drift. Prerequisites for accurate calculation
of the standard deviation are good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
10
Manual Mass Comparators up to 20 kg
Developed for mass metrology, the inte- The majority of Sartorius manual mass Additional draft shields are available for
grated application software facilitates daily comparators are designed so that they do nearly all models. They reduce air
work routines. This software guides the not require space-consuming centering movement caused by air conditioning
user, while mass comparator takes over the equipment within the weighing chamber. systems and, in this way, enable excellent
timing and, at the end of the ABA or ABBA repeatability to be achieved even under
cycles, delivers the difference in mass, The complete electronics and the power unfavorable ambient conditions (for a list,
standard deviation and individual values. supply for display and evaluation are see page 23).
separate from the weighing system in
All CCE mass comparators up to a load order to prevent heat from affecting
range of 5 kg have a built-in motorized the results.
calibration and adjustment weight.
CCE2004 CCE20000
Maximum capacity and Maximum capacity: 20.05 kg
electronic weighing range: 2.5 kg Electronic weighing range: 60 g
Readability: 0.1 mg Readability: 1 mg
Repeatability (typical), s*: 0.1 mg Repeatability (typical), s*: 1 mg
CCE5003
Maximum capacity and
electronic weighing range: 5.1 kg
Readability: 1 mg
Repeatability (typical), s*: 0.5 mg
CCE5004
Maximum capacity and
electronic weighing range: 5.1 kg
Readability: 0.2 mg
Repeatability (typical), s*: 0.3 mg
s* The standard deviation "s" is the repeatability calculated from 6 ABA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
11
Manual Mass Comparators up to 300 kg
All mass comparators from a load range The single-point monolithic weigh cell Depending on the ambient conditions, the
of 40 kg have a continuous, full-resolution based on electromagnetic force compen- excellent repeatabilities can be increased
weighing range and large weighing pan. sation supports over 60 kg. As a result of still further by using a Centermatic and a
This makes them ideal for universal use. the direct introduction of force into the draft shield.
weigh cell, unprecedented repeatabilities
If necessary for below-balance weighing, a of just a few milligrams are achieved with
hanger can be attached directly to the port this weighing system. Due to the direct
on the weigh cell. In this way, for example, introduction of force, the off-center load-
a simple, yet highly accurate, density meas- ing error (eccentricity) is extremely low,
urement setup can be achieved. The inte- and the stated specifications are also
grated software supplied for density deter- achieved without the use of a Centermatic.
mination on all CCE models simplifies work
routines.
CCE40K3 CCI60K2
Maximum capacity: 41 kg Maximum capacity: 64 kg
Readability: 2 mg Readability: 50 mg
Repeatability (typical), s*: 3 mg Repeatability (typical), s*: 100 mg
CCE60K2 CCI100K2
Maximum capacity: 64 kg Maximum capacity: 151 kg
Readability: 10 mg Readability: 50 mg
Repeatability (typical), s*: 7 mg Repeatability (typical), s*: 200 mg
CCE60K3 CCI300K
Maximum capacity: 64 kg Maximum capacity: 303 kg
Readability: 2 mg Readability: 1000 mg
Repeatability (typical), s*: 4 mg Repeatability (typical), s*: 500 mg
s* The standard deviation "s" is the repeatability calculated from 6 ABA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
12
Manual Mass Comparators up to 3000 kg
All Sartorius heavy-duty mass comparators Models of the CCS range are equipped The innovative sturdy three-point frame of
are made of high quality stainless steel. with four high-resolution strain-gauge the CCT models with three high-resolution
This design eliminates the possibility of type load cells which are aligned with strain-gauge type load cells guarantees a
magnetic reactions from magnetic weights one other. Draft shields are available stable and distortion-free setup. Potential
having an effect on comparison weighing. for all high-capacity mass comparators off-center loading errors are minimized
(see page 23). due to the large distances between the
The unique sandwich construction weigh cells.
with integrated, strain-free weigh
cells guarantees excellent repeatabilities
even when weights are loaded in a
"heavy-handed" way.
CCT1000K CCS600K
Maximum capacity: 1200 kg Maximum capacity: 605 kg
Readability: 1g Readability: 1g
Repeatability (typical), s*: 2g Repeatability (typical), s*: 2g
Dimensions in cm, Dimensions in cm,
W+D+H: 151+131+30 W+D+H: 80+100+30
CCT2000K CCS1000K
Maximum capacity: 2010 kg Maximum capacity: 1510 kg
Readability: 1g Readability: 5g
Repeatability (typical), s*: 5g Repeatability (typical), s*: 5g
Dimension in cm, Dimensions in cm,
W+D+H: 192+166+30 W+D+H: 80+100+30
CCS3000K
Maximum capacity: 3010 kg
Readability: 10 g
Repeatability (typical), s*: 12 g
Dimensions in cm,
W+D+H: 150+125+30
s* The standard deviation "s" is the repeatability calculated from 6 ABA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
13
Summary and Product Recommendations
Application Range of Mass Comparators in Accordance with OIML R111
Robots and automatic mass comparators
CCL1007 E1
CCR10-1000 E2
CCR10 F1
CCR1000
CCE1000S-L
CCE10000U-L
CCE10000S-L
CCE20000S-L
CCE50001S-L
CCE6 E1
CCE36 E2
CCE66 F1
CCE111 F2
CCE605 M1
CCE1005
CCE1004
CCE2004
CCE5004
CCE5003
Manual mass comparators
CCE10000S
CCE10K3
CCE20000
CCE40K3
CCE60K3
CCE60K2 F1
CCI60K2 F2
CCI100K2 M1
CCI300K M2
CCS600K
CCT1000K
CCS1000K
CCT2000K
CCS3000K
100
200
500
100
200
500
100
200
500
1
2
5
1
2
5
1
2
5
1000
2000
5000
10
20
50
10
20
50
10
20
50
Mass
mg g kg
The chart shows the suitability of mass comparators for the OIML R111 accuracy classes.
The results shown in the bar chart were The diagram shows the suitability of Automated mass comparators even notice-
obtained by performing at least 3 ABBA comparators at a confidence level of 95% ably exceed the stringent requirements for
cycles with automatic comparators and 3 (K=2) for the application ranges. class E1 measurement uncertainty, which
ABA cycles with manual comparators. makes them suitable for use in national
The stated application ranges can be primary standard laboratories.
extended to include smaller weights by
increasing the number of comparison
measurements.
14
Application Range of Mass Comparators in Accordance with ASTM E617
Robots and automatic mass comparators
CCL1007 Class 0
CCR10-1000 Class 1
CCR10 Class 2
CCR1000
CCE1000S-L
CCE10000U-L
CCE10000S-L
CCE20000S-L
CCE50001S-L
CCE6 Class 0
CCE36 Class 1
CCE66 Class 2
CCE111 Class 3
CCE605 Class 4
CCE1005
CCE1004
CCE2004
CCE5004
CCE5003
Manual mass comparators
CCE10000S
CCE10K3
CCE20000
CCE40K3
CCE60K3
CCE60K2
CCI60K2 Class 3
CCI100K2 Class 4
CCI300K Class 5
CCS600K
CCT1000K
CCS1000K
CCT2000K
CCS3000K
100
200
300
500
100
200
300
500
100
200
300
500
1
2
3
5
1
2
3
5
1
2
3
5
1000
2000
3000
10
20
30
50
10
20
30
50
10
20
25
30
50
Mass
mg g kg
The chart shows the suitability of mass comparators for the ASTM E617 accuracy classes.
15
Selection of Comparators for Equipping a Mass Laboratory
1 mg 1 mg 1 mg 1 mg
10 mg 10 mg 10 mg 10 mg
Manual mass comparators (blue)
CCE6
1 g 1 g 1 g 1 g
CCE66
CCE66
2 g 2 g 2 g 2 g
CCE111
CCE111
CCE111
5 g 5 g 5 g 5 g
CCE605
CCE66
CCR10-1000
10 g 10 g 10 g 10 g
20 g 20 g 20 g 20 g
CCE2004
Robot systems (red)
50 g 50 g 50 g 50 g
CCE1005
CCE1004
100 g 100 g 100 g 100 g
CCE1000S-L
1 kg 1 kg 1 kg 1 kg
CCE5003
CCE10000U-L
CCE10000S
CCE10K3
2 kg 2 kg 2 kg 2 kg
CCE40K3
5 kg 5 kg 5 kg 5 kg
CCE40K3
CCE20000S-L
CCE60K2
1 0 kg 1 0 kg 1 0 kg 1 0 kg
CCE60K3
CCE60K2
CCE50001S-L
2 0 kg 2 0 kg 2 0 kg 2 0 kg
5 0 kg 5 0 kg 5 0 kg 5 0 kg
16
Accessories
Susceptometer – A Complete Solution for Testing the Magnetic Properties of Weights
The OIML R111 stipulates that magnetic The susceptometer method measures the Application Software
properties of weights be determined prior interaction between a permanent magnet The software that comes standard with
to mass calibration. This confirms that the and the weight to be tested as a function the susceptometer is exceptionally easy to
requirements placed on weights no longer of the force exerted on a mass standard. use and enables data to be automatically
only apply to geometric quantities, materi- A high-resolution balance is used for this transferred from the unit. It also supports
al density, weight, or surface qualities. In test. The desired magnetic properties of the the user through all required steps and
fact, many scientific evaluations support weight can be calculated on the basis of provides assistance with settings before
the demand that both the susceptibility the known properties of the test magnet, and during a measurement. Menu prompts
and magnetization of weights be the distance of the weight from the test request all relevant data and guide the user
considered. magnet and the weight's geometry. through the steps necessary for performing
a measurement. The results of the test
Susceptometer Method The Sartorius susceptometer stands out are recorded and can be exported to
There are several methods for defining against other well-known instruments by higher-level databases.
the magnetic properties of weights. virtue of its highly compact design and
The susceptometer method is stipulated in integrated components. An innovative A User-friendly, All-in-one Solution
OIML R111 for testing weights of classes turning mechanism for adjusting the posi- The YSZ01C|YSZ02C susceptometer
E1, E2, F1 and F2. The new YSZ01C |YSZ02C tion of the test magnet and simple adjust- from Sartorius rounds off the range of
susceptometer from Sartorius lets you ment of the height of the loading platform accessories for high-resolution balances
easily and conveniently determine the also make working with the susceptometer and mass comparators. Along with its
susceptibility and magnetization of simple and effective. This design signifi- specialized software, the susceptometer
weights in accordance with OIML R111. cantly improves the repeatability of mea- provides a unique, complete solution that
surements. The susceptometer is available enables you to ensure that the limits of
in two resolution levels. The YSZ01C fea- the magnetic properties of weights are
tures a 10-μg resolution; the YSZ02C has maintained.
a 1-μg resolution.
Technical Specifications
Susceptometer YSZ01C YSZ02C
Readability of the susceptometer 10 μg 1 μg
Application range in accordance E2|F1|F2 E1|E2|F1|F2
with OIMLR 111
Dimensions 338 x 286
Height 249 mm
Maximum capacity 50 kg
Dipole moment of the magnet m ~ 0.1 Am2
Geometry ratio of the magnet Height to diameter = 0.87
Susceptometer in a choice of versions YSZ01C, YSZ02C
Z0, distance from center of magnet Adjustable in fixed steps:
to the bottom of weight in mm Z1 = 18 | Z2 = 20 | Z3 = 27 | Z4 = 35 | Z5 = 43
Field strength due to different
heights Z0, in A/m 2700 | 2000 | 800 | 360 | 200
Turning mechanism for magnet Easy to operate by external rotary knob;
marking for N-S orientation of the magnet
Software Convenient application software; standard settings
and easy, user-defined configuration possible;
quick check function; printing of reports, exporting
of results
Data transfer protocol HTML mode and data transfer to user-specific
metrology software via CSV file format
Set of 3 reference magnets YSZ01RMC YSZ01RMC 3 reference magnets in a wooden case for Sartorius
susceptometer YSZ01C and YSZ02C
YSZ01RSC 1 kg reference susceptibility standard in a wooden
case for Sartorius susceptometer YSZ01C and
YSZ02C with PTB magnetic susceptibility certificate
17
Volume and Density Determination up to 1 kg
Density determination is often required in The load alternators and weighing pans are
metrology and industry. The most accurate designed as comb-type weight grabbers so
method for determining the density of that weights and groups of weights from
solid objects in accordance with OIML R111 1 g to 1 kg can be transferred directly from
is hydrostatic mass comparison in liquid. the load alternator to the suspended weigh-
In this procedure, a volume reference of ing pan. Adapter plates are not required.
known mass and volume is compared in liq- This has a positive effect on the accuracy
uid of known density with a test object of of the overall measurement system.
known mass. The volume or density of the
test weights is calculated from the differ- After the data has been entered, a built-in
ent buoyancies of the weights in the liquid. PC with user-friendly software takes over
fully automatic control of the volume
Sartorius has integrated this OIML R111 comparator and evaluation of density
method into a fully automatic volume measurement.
comparator.
Technical Specifications:
VD1005|VL1005
Density uncertainty* 1 kg/m3
Volume uncertainty* 0.00015 cm3
VD1005|VL1005, Volume comparator
Weight diameter 6…100 mm
Maximum sphere diameter 95 mm
Application
Density deterimnation of weights
in accordance with OIML R111,
Class E1 1 g…1 kg
Comparator Technical Specifications
Maximum capacity 1125 g
Readability 0.01 mg
Repeatability, s* < 40 μg
Repeatability (typical), s* < 20 μg
Weighing range, electronic 350 g
Tare range (subtractive) 350 g
* Partial uncertainty of the volume comparator (without references and test weights)
s* The standard deviation "s" is the repeatability calculated from 6 ABBA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
18
Accessories for Density Determination
YVT01C Thermostat
YWT20C Table for VD1005 and VL1005
YDR1000SIC 1 kg silicon sphere with PTB density certificate
YCS31-612-09 Set of 1g to 1kg weights with PTB density certificate
YCS51-612-02 Set of E2 substitution weights with E1 DKD certificate
19
Software for Mass Metrology –
One Software Program for All Equipment ScalesNet32
Careful acquisition of weighing data and The following functions are available – Number of weighing cycles and type of
associated parameters is one of the most to the user: weighing procedure (ABA or ABBA) can
important criteria in mass metrology. be configured by class
– Calibration of customer weights
To ensure that ScalesNet32 meets this
– Selection of classes in accordance with
requirement, it was continuously tested – On-location calibration of customer
OIML R111, ASTM 617 or other national
throughout the entire development phase weights
standards
in the DKD laboratory for correct function-
– Calibration of reference weights
ality and suitability in practical daily use. – Simultaneous testing of weights from a
These accompanying checks were carried – Subdivision of weights particular weight set on multiple mass
out in an accredited DKD laboratory and comparators in the laboratory
– Quick comparison of weights without
are not only substantially responsible for
printed report – Plausibility testing after the reference
the high degree of functionality achieved,
weight set and weighing instrument have
but also provide the user with the neces- – Calibration of weights with raw data
been selected (i.e., “Are both the instru-
sary peace of mind. ScalesNet32 meets the output
ment and the reference suitable for this
requirements of all quality management
– Manual input of weighing data for com- class?”)
systems that use calibrated weights to
parators without RS-232 connectivity
supply evidence of traceability to national – Test records generated for each weight
standards. – Calibration of mass comparators, tested; records include all data acquired
balances and scales during testing (reference weight,
The calibration of a weight is always weighing instrument used, temperature,
– Adjustment of mass comparators,
allocated to an order or customer, the humidity, air pressure, etc.)
balances and scales
serial number, manufacturer, shape and
– The history of each weight tested can
features of the weight. These details are – Recording and graphic plotting of
be viewed
assigned to the test object and thus guar- external parameters
antee clear identification. The weight data – User-configurable certificates of verifica-
is stored in a database; therefore, the The principal features of ScalesNet32: tion, inspection, calibration or DKD cali-
history of the weight can be viewed at any bration as Word templates; the data and
– Central SQL database for storing all
time. test results are positioned in the Word
measurement values and acquired infor-
template with bookmarks, which are
mation
The mass comparators used to calibrate replaced by measured values or other
weights are monitored by ScalesNet32 and – Automatic loading of weighing data over data when the report is printed; bilingual
calibrated and adjusted at specified inter- the RS-232 interface of the weighing DKD certificates can be generated.
vals. The calibration results obtained are instrument. The parameters of the inter-
– Automatic generation of inventory
stored in the database. face can be generated in accordance with
lists of the weighing instruments and
the weighing instrument manufacturer’s
reference weights used
ScalesNet32 also monitors the mass data.
standard and climate stations used, along – Monitoring of calibration intervals for
– Automatic detection of parameters
with their associated sensors. The software the mass comparators, reference weight
defining ambient conditions during
informs the user of the calibrations that sets and climate stations used
weighing cycles
are due. The reference weight standards,
climate stations and mass comparators are
blocked from use once the tolerance inter-
val has expired. The test intervals for the
mass comparators, reference weights and
climate equipment are set by the user in
accordance with the particular
requirements.
ScalesNet32, YSN01C
20
Components of ScalesNet32
ScalesNet32-Desk
Module for the administration of all master
data and orders, including all printouts.
Connectivity to the ScalesNet32 database
server is provided via Ethernet.
ScalesNet32-Mass
This module is designed for handling the
calibration of test objects and mass stan-
dards as well as calibration and adjustment
of comparators. Connectivity to the
ScalesNet32 database server is provided
RS-232C RS-232C
via Ethernet or WLAN.
This application software is designed for The time at which the mass comparison
highly accurate online determination of takes place and the maximum uncertainty
mass using Sartorius manual mass compara- for automatic mass comparison are
tors and for controlling automatic compar- selectable.
isons of mass in conjunction with Sartorius
automatic mass comparators. It runs on The standard software contains a database
Microsoft Windows 2000 or XP. for reference and test weights, and
facilitates online calculation of the
The software facilitates the determination standard deviation and mass difference
of the difference between two masses and and the uncertainties at either the sigma 1,
the determination of absolute or apparent sigma 2 or sigma 3 level.
mass (conventional mass value), taking
buoyancy errors into account. The program The number of ABBA or ABA cycles can be
automatically implements buoyancy cor- selected in the range from 01–99. Software, YPR02C
rections after manual entry of the climate
data or automatic acquisition of climate
data using the climate station YCM02C or
YSM03C. The clima datas will be collected
and stored on the PC.
21
Accessories for Mass Metrology
Mass Standards and Weights Class E1 and E2 sets are available in various
Sartorius offers weights and mass stan- weight combinations.
dards in conformance with OIML R 111
in all accuracy classes ranging from E1 All weights are available individually, with
to M2 and from 1 mg to 1000 kg. or without DKD or PTB certification.
M class mass standards are available up
to 1000 kg in a variety of shapes, sizes Please consult our separate brochure
and materials. for details or visit our website at
www.sartorius.com.
Nominal Value Order No. Order No. Order No. Set of weights, YSC011-611-00
DKD-calibrated DKD-calibrated
E1 E1 E2
1 mg – 5 g YCS011-351-00 YCS011-351-02 YCS011-352-00
1 mg – 100 g YCS011-511-00 YCS011-511-02 YCS011-512-00
1 mg – 200 g YCS011-521-00 YCS011-521-02 YCS011-522-00
1 mg – 1 kg YCS011-611-00 YCS011-611-02 YCS011-612-00
1 mg – 5 kg YCS011-651-00 YCS011-651-02 YCS011-652-00
1 g – 1 kg YCS31-611-00 YCS31-611-02 YCS31-612-00
1 g – 5 kg YCS31-651-00 YCS31-651-02 YCS31-652-00
1 g – 10 kg YCS31-711-00 YCS31-711-02 YCS31-712-00
23
Technical Specifications:
1 kg Prototype Mass Comparator and Robot
CCR10-1000
CCL1007 CCR10 CCR1000
Maximum capacity 1031 g 10.5 g 1002 g
Application range 1 mg…1 kg 1 mg…10 g 10 g…1 kg
Readability 0.1 μg 0.1 μg 1 μg
Repeatability, s* < 0.2 μg < 0.3 μg 0-100 g: < 3 μg
< 8 μg
Repeatability (typical), s* < 0.1 μg < 0.2 μg < 2 μg
Electronic weighing|
taring range 2g 3.5 g 2g
Magazine positions 8 39 21
Optional magazine
positions – 26-65 2-39
Substitution weights External 2+3.5 g 30|40|50|100|100|300|400 g
Linearity 1 μg 1 μg 20 μg
Range sensitivity 0.2 μg /500 mg 0.1 μg/10 mg 2 μg/500 mg
Stabilization time in s 60 15 25
Cycle time ABBA in s 480 240 240
Calibration weight 1 g Built-in 2 g|E2 2 g|E2
optional YCW322-00 YCW322-00
Standard Accessories
Interfaces LAN|USB| LAN|USB| LAN|USB|
RS-232 RS-232 RS-232
Centermatic Yes No Yes
Draft shield Vacuum chamber Yes, 2+ Yes, 2+
PC|Software Yes|Yes Yes|Yes Yes|Yes
Optional Accessories
Climate station Climate station for an E1 lab.: YCM02C; climate station for an E2 lab.: YCM03C
Software for PC ScalesNet32, metrological control and evaluation software with database, YSN01C
Miscellaneous Vacuum pump
Dimensions
Weighing pan 3-point 52+24 mm 104+68 mm
Sample size (D+H) 34-95+100 mm 18+20 mm 100+120 mm
(W+D+H) in mm 940+840+1100 1800+1150+1800 1800+1150+1800
s* The standard deviation "s" is the repeatability calculated from 6 ABBA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
24
Technical Specifications:
Mass Comparators with Load Alternator
s* The standard deviation "s" is the repeatability calculated from 6 ABBA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
25
Technical Specifications:
Mass Comparators up to 1kg
s* The standard deviation "s" is the repeatability calculated from 6 ABA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
** isoCAL fully automatic internal calibration and adjustment
26
Technical Specifications:
Mass Comparators from 1 kg to 10 kg
s* The standard deviation "s" is the repeatability calculated from 6 ABA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
** isoCAL fully automatic internal calibration and adjustment
27
Technical Specifications:
Mass Comparators from 10 kg to 60 kg
s* The standard deviation "s" is the repeatability calculated from 6 ABA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
28
Technical Specifications:
Mass Comparators from 60 kg to 300 kg
s* The standard deviation "s" is the repeatability calculated from 6 ABA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
29
Technical Specifications:
Mass Comparators from 600 kg to 3000 kg
s* The standard deviation "s" is the repeatability calculated from 6 ABA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
30
Technical Specifications:
Determination of Volume, Density, Susceptibility and Magnetism
s* The standard deviation "s" is the repeatability calculated from 6 ABBA cycles after elimination of drift. Prerequisites for accurate calculation of the standard deviation are
good ambient conditions in accordance with OIML R111 for an E1 mass standards laboratory.
31