better analysis counts

Application Areas:
• Total sulfur analysis from ultra low sulfur

7039 G3
fuels to crudes.
• For use in refinery labs, pipeline terminals,
additive plants, testing vans and inspection
laboratories.
• Complies with ASTM D7039 and ISO 20884.

Features and Benefits:


• LOD: 0.15 ppm at 300 s.
• Dynamic Range:
Standard: 0.15 ppm to 3000 ppm
XR package: 0.15 ppm to 10%
• Available with 8 cell auto-sampler.
• Fits on any bench and compatible for use in
mobile labs/vans.
• Plug-it-in and measure: power is only utility.
• Touch screen user interface.
• Utilizes ACCU-CELL pre-assembled and
pre-vented sample cups for enhanced preci-
sion, extreme ease-of-use and enhanced
productivity.
• User programmable measurement time:
30-900 s.
• One calibration curve will run both diesel and
gasoline up to 3000 ppm.
• No conversion gasses, heating elements,
quartz tubes or columns.
• 75 W air-cooled excitation tube.
Sulfur Analysis with Unprecedented Precision • Robust polyamide window for easy cleaning.
No compromises in detection, performance and reliability - the SINDIE 7039 G3
analyzer is the ideal sulfur analytical solution for the energy industry. From ultra Options:
low sulfur diesel and gasoline to heavy fuel oil and crudes, the analyzer delivers • Extended Range package: from 0.15 ppm
unprecedented precision and accuracy. The Sindie 7039 G3 bench-top analyzer is up to 10%.
easy-to-use, robust and complies with both ASTM D7039 and ISO 20884 methods. • 8 cell Autosampler.
Plug it in and measure. Results with one touch. Unrivaled precision. • LIMS data output software capability.
MWD XRF
Monochromatic Wavelength Dispersive X-Ray Fluorescence (MWD XRF) utilizes state-of-the-art
focusing and monochromating optics to increase excitation intensity and dramatically improve signal-
to-background over high power traditional WD XRF instruments. This enables significantly improved
detection limits and precision and a reduced sensitivity to matrix effects. A monochromatic and focused
primary beam excites the sample and secondary characteristic fluorescence x-rays are emitted from the Sulfur
sample. A second monochromating optic selects the sulfur characteristic x-rays and directs these x-rays Polychromatic
Incident X-rays Optic 1 Optic 2
to the detector. MWD XRF is a direct measurement technique and does not require consumable gasses or from tube
sample conversion.

SINDIE Autosampler Precision


• 8 sample cell capacity Typical repeatability (r) and reproducibility (R) values in
diesel fuel, at 95% confidence. 300 s measurement time.
• Increases productivity
• Utilitizes XOS Accu-Cell cups Sulfur
Concentration r R
(ppm)
2 0.3 0.7
ACCU-CELL Sample Cups 5 0.5 0.8
• No assembly of separate film & cup components
8 0.6 1.0
• Pre-vented sample cups 15 0.8 1.4
• Eliminates sample & cup contamination 100 2 4
• One discharge of 1 ml pipette will fill the cup 500 5 10

SINDIE-XR Low Range


Calibration Calibration
Curve

Product Specifications
Test Method ASTM D7039 and ISO 20884
Dimensions 37 cm (w) x 50 cm (d) x 34 cm (h)
Power 100-120 VAC, 47-63 HZ at 6.0 Amps/200-240 VAC, 47-63 HZ at 6.0 Amps
Sample Cup Volume 1 ml
I/O Ports Ethernet 10/100 base T, RS232
Optional Computer Interface Pentium, 100 MHz, 32 MB RAM/Windows 98 or newer operating system
Ambient Temperature Requirements 5-40o C (40-104o C)
Dynamic Range Standard: 0.15 – 3000 ppm, XR Package: 0.15 ppm – 10%
Measurement User selectable: 30-900 s
Calibration 8 calibration curves. Automatic and Manual Calibration functionality

better analysis counts


15 Tech Valley Drive • East Greenbush, New York 12061, USA • 518.880.1500 • Fax: 518.880.1510
e-mail: [email protected] • website: www.xos.com
© XOS (2012) MAR2012