Test Procedures For The Determination of Desiccant Adsorption Capacity
Test Procedures For The Determination of Desiccant Adsorption Capacity
The U.C. is only one of many parameters that are used to design adsorption systems, the
particulars of which are not important. What is important is that the drying system produces
“dry” air. How “dry” the air is can be readily determined by using a hygrometer to measure the
dew point of the air. The dew point will indicated how well the whole drying system is
operating, and not just the condition of one component.
Two test methods are for determining U.C. are shown below for informational purposes. Method
one is based on the fact that the desiccant gives off heat when it adsorbs water (Heat of
adsorption). This test is not 100% accurate because the desiccant does not adsorb liquid water in
the same way that it adsorbs gaseous water. A negative result (low temperature rise) does
indicate that the desiccant is contaminated, but a positive result (high temperature rise) does not
necessarily indicate that the desiccant is okay.
Method two is the test procedure used by some desiccant manufacturers when they test desiccant
samples. The normal value is 18% for molecular sieve using a 10% R.H. atmosphere.
Method one
Heat of Adsorption Desiccant Analysis
Apparatus
2 Beakers 250 ml
1 Glass thermometer 50F – 150F
150 ml Distilled water
1 Graduated cylinder
Place a sample of desiccant in an oven set at 550F. After 3 hours, remove the sample and seal
the container to prevent moisture adsorption while the desiccant cools to room temperature
(about 24 hours).
Pour 45 ml of water into a beaker and measure the temperature. In a second beaker, pour a
quantity of sieve which is about 10% more by volume than the water (50 ml). Dry the
thermometer and place it into the sieve. Pour the water into the beaker and observe the increase
in temperature of the mixture while stirring with the thermometer. Record the peak temperature
which will occur in about 20 seconds. Subtract the water temperature from the peak temperature
and record. Run at least 3 tests and determine an average value.
If the temperature difference is 40F or greater, the sieve may be in good condition. If the
temperature rise is less than 40 F, the desiccant should be replaced.
If the temperature rise is correct but the dryer still does not produce a low enough dew point, the
problem could be that the desiccant beads are physically breaking apart. Look for the presence of
powdered or chipped desiccant particles.
Method two
Analytical Procedure for the Determination
of Adsorption Capacity of Silica Gel and Molecular Sieve
Principle
Air saturated at 10-20-40-60 and 80 percent relative humidity is passed over the sample until
equilibrium is reached at pressure less that 1” Hg and at some temperature between 76 and 86
deg F. Once the test is started, the temperature must not vary more than + or – 1 deg F
throughout the test period.
This table of acid strengths by weight percent has been computed from the International Critical
Tables. The relative humidity represents values that would be obtained in a closed static system.
The actual effluent relative humidities will be below these values, decreasing as the back
pressure increases on the system.
The sample containing bottle should be a Nesbitt, Fleming or other standard form adsorption
bulb. The test room, the inlet air and the sulfuric acid solutions should be at the same
temperature and should not vary more than plus or minus 1.0 deg F during the test period.
Procedure
Refer to drawing SK-912A. The adsorption bulb is tightly packed with about a 1 inch layer of
Pyrex glass wool. A two to eight gram sample (2 grams of fine gels and 8 grams on large crystal
size) of the desiccant is weighed accurately in the tared adsorption bulb. Extreme care should be
exercised to insure exposure of the sample for a minimum amount of time. Connect the
adsorption bulb to the apparatus and adjust the air flow to four liters per minute with the by-pass
closed. If the manometer indicates the pressure to be greater than 1” of mercury, the bleed-off
line should be opened just enough to lower the pressure to less than 1” and that point maintained
for the test period. When using the bypass, an occasional check should be made of the amount of
air passing through the adsorption bulb by diverting the bypass air form the flowmeter to the
atmosphere through the “tee” connection. The flowmeter reading will then indicate the actual air
flow through the adsorption bulb. After 5 or 6 hours, the adsorption bulb is removed from the
train and weighed. The bulb is then placed back in the adsorption train for two more hours and
then it is again weighed. Repeat the two hour weightings until two successive weightings show a
weight gain not exceeding five milligrams.
Calculations
AxB=C
DxE=F
C x 1000 = ml. C.P. H2SO4 required to make 1 liter of R.H. required solution