Impulse 6000D/7000DP: Sending Test Data To A PC: Application Note
Impulse 6000D/7000DP: Sending Test Data To A PC: Application Note
F ro m t h e F l u k e B i o m e d i c a l D i g i t a l L i b r a r y @ w w w. f l u k e b i o m e d i c a l . c o m / l i b r a r y
Advanced users The COM port should be set to: Hyperterminal
t 115,200 baud
Advanced users can find more t No parity To use HyperTerminal: Make a
information about the FT232R t 8 data bits connection for the COM port and
from the FTDI web site: www. t 1 stop bit set it to 115,200 baud, N81, with
ftdichip.com. New software t Hardware handshaking should hardware flow control. Open the
drivers, application notes, and be turned on. connection.
USB utilities are available. You When you press the “Send”
can learn how to view your USB button on your Impulse, you
connections and load and/or
Test data should see the test data on the
delete all FTDI drivers from your For most tests performed by terminal screen.
PC and obtain drivers for other the Impulse, the test data is After receiving the test data
operating systems. You can even displayed on one or more Sum- you can mark it and copy it to
use their D2XX direct interface mary screens. At each Summary the Clipboard, then paste it into
API to include in your own screen there is a “Send” button. a file somewhere else.
custom interface programs if you Press the “Send” button to send Or you can go to Transfer/Cap-
don’t want to use a COM port. that test data out via the USB ture Text and open a file. Then
Port. all data received will go into
COM port settings that file. Stop the capture when
Settings for the COM port should
Receiving data on a PC you are finished; you can then
use the data in the file as you
be made by the program that To get the test data on your PC,
please.
opens and uses the COM port you must open that COM port
such as: and receive the data when it is
t Terminal emulation program sent by the Impulse.
(HyperTerminal, Tera Term or You can use a terminal
other) emulation program such as
t Your custom Impulse connec- HyperTerminal to receive the
tion program data. Terminal emulation pro-
t Fluke Ansur Test Automation grams typically have ways to
System program save incoming data to a file. It
is also possible to write your
The settings in Device Manager
own PC program to receive the
are usually irrelevant because
data from the COM port for per-
they are overridden by the con-
sonalized applications.
trolling program.
Fluke Biomedical.
About Fluke Biomedical Better products. More choices. One company.
Fluke Biomedical is the world’s leading manufacturer of quality biomedical
test and simulation products. In addition, Fluke Biomedical provides the
latest medical imaging and oncology quality-assurance solutions for Fluke Biomedical
regulatory compliance. Highly credentialed and equipped with a NVLAP Lab 6045 Cochran Road
Code 200566-0 accredited laboratory, Fluke Biomedical also offers the best Cleveland, OH 44139-3303 U.S.A.
in quality and customer service for all your equipment calibration needs. Fluke Biomedical Europe
Today, biomedical personnel must meet the increasing regulatory Science Park Eindhoven 5110
pressures, higher quality standards, and rapid technological growth, 5692EC Son, The Netherlands
while performing their work faster and more efficiently than ever. Fluke For more information, contact us:
Biomedical provides a diverse range of software and hardware tools to In the U.S.A. (800) 850-4608 or
meet today’s challenges. Fax (440) 349-2307
In Europe/M-East/Africa +31 40 267 5435 or
Fluke Biomedical Regulatory Commitment Fax +31 40 267 5436
From other countries +1 (440) 248-9300 or
As a medical test device manufacturer, we recognize and follow certain
Fax +1 (440) 349-2307
quality standards and certifications when developing our products. We are Email: sales@flukebiomedical.com
ISO 9001 and ISO 13485 medical device certified and our products are: Web access: www.flukebiomedical.com
t $&$FSUJýFE
XIFSFSFRVJSFE
©2012 Fluke Biomedical. Specifications subject
t /*455SBDFBCMFBOE$BMJCSBUFE
to change without notice. Printed in U.S.A.
t 6-
$4"
&5-$FSUJýFE
XIFSFSFRVJSFE 10/2012 4280040A_EN
t /3$$PNQMJBOU
XIFSFSFRVJSFE
Modification of this document is not permitted
without written permission from Fluke Corporation.