0% found this document useful (0 votes)
145 views4 pages

EZ7800 Series Online Colorimetric Analyser For Total Phosphorus

The document summarizes the EZ7800 Series online colorimetric analyzer for measuring total phosphorus in water. It can measure total phosphorus in wastewater and surface water applications. The analyzer uses a proprietary sample digestion technique combined with automatic operation to provide fast, reliable total phosphorus measurements. It has options for multiple measurement ranges and streams and can output data via analog outputs and digital communication protocols.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
145 views4 pages

EZ7800 Series Online Colorimetric Analyser For Total Phosphorus

The document summarizes the EZ7800 Series online colorimetric analyzer for measuring total phosphorus in water. It can measure total phosphorus in wastewater and surface water applications. The analyzer uses a proprietary sample digestion technique combined with automatic operation to provide fast, reliable total phosphorus measurements. It has options for multiple measurement ranges and streams and can output data via analog outputs and digital communication protocols.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

EZ7800 Series Applications

• Wastewater

Online Colorimetric • Surface water

Analyser for
Total Phosphorus

Online, automatic monitoring of Total Phosphorus (TP)


in water
The EZ7800 Series of Online TP Analysers meet the needs • Full oxidation of all complexed phosphate
for fast, convenient and reliable monitoring of the regulatory
• Smart automatic features
sum parameter Total Phosphorus in municipal and industrial
wastewater as well as surface water applications. The EZ7800 • Control and communication via industrial panel PC
Series share similar technology with the EZ7700 Series which
monitor Total Nitrogen. • Standard 4 - 20 mA signal output with alarm processing

Phosphorus is often synonymous with phosphate but can be • Communication supporting Ethernet connectivity to
present in different forms: orthophosphate (O-phosphate), Modbus TCP/IP
condensed phosphate (polyphosphate) or organically bonded • Multiple stream analysis
phosphate. O-phosphate is often the only form that is
measured with traditional test kits or even in laboratory, yet the
organic form of phosphate can be an important constituent
of industrial discharges, often deriving from corrosion control
programs or use of fertilisers. In summary, knowing levels of TP
can be an important added value to waste water operations.
The EZ7800 Online Total Phosphorus Analyser allows operators
of industrial and municipal water works to monitor and control
in a convenient way the total portion of phosphorous, if needed
combined with a free phosphate measurement. The analyser’s
industrial mainframe harnesses a proprietary sample digestion
technique combined with seamless operation:
2 EZ7800 Series Online Colorimetric Analyser for Total Phosphorus

Technical Data*
Parameter Total Phosphorus, free PO4
Colorimetric measurement at 630 nm using ascorbic acid reduction and molybdate colour solution
Measurement method
after persulphate digestion in acidic medium, conform with APHA 4500-P
0.01 - 2 mg/L TP
0.025 - 5 mg/L TP
0.05 - 10 mg/L TP
Measuring range
0.1 - 20 mg/L TP
0.01 - 2 mg/L TP & 0.005 - 1 mg/L PO4-P
0.025 - 5 mg/L TP & 0.005 - 1 mg/L PO4-P
Precision Better than 2% full scale range for standard test solutions
Detection limit ≤ 10 µg/L
Arsenic (V), Chromium (VI), Copper (II) > 10 mg/L, Iron (III) > 10 mg/L, Sulphide > 2 mg/L, and Vanadium (V),
Interferences
Silica > 60 mg/L. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Cycle time 30 min including digestion of 10 min (standard)
Automatic cleaning Yes
Calibration Automatic, 2-point; frequency freely programmable
Validation Automatic; frequency freely programmable
Ambient temperature 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Reagent requirements Keep between 10 - 30 °C
Sample pressure By external overflow vessel
Flow rate 100 - 300 mL/min
Sample temperature 10 - 30 °C
Sample quality Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
230 VAC, 50/60 Hz
Power
Max. power consumption: 440 VA
Instrument air Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Demineralised water For rinsing
Drain Atmospheric pressure, vented, min. Ø 64 mm
Cooling water Flow rate approx. 5 L/h; temperature max. 30 °C; pressure max. 0.5 bar
Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Analogue outputs Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Digital outputs Optional: Modbus RS232, RS485
Alarm 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Protection class Analyser cabinet: IP55 / Panel PC: IP65
Hinged part: Thermoform ABS, door: plexiglass
Material
Wall section: Galvanised steel, powder coated
Dimensions (H x W x D) 690 mm x 465 mm x 330 mm
Weight 25 kg
Certifications CE compliant / UL certified
*Subject to change without notice.

www.hach.com
EZ7800 Series Online Colorimetric Analyser for Total Phosphorus 3

Dimensions

Hach Service
With Hach Service, you have a global partner who understands your needs and cares about delivering timely, high-quality service you
can trust. Our Service Team brings unique expertise to help you maximise instrument uptime, ensure data integrity, maintain operational
stability, and reduce compliance risk.

www.hach.com
4 EZ7800 Series Online Colorimetric Analyser for Total Phosphorus

Order Information - Part Number Configurator


Total Phosphorus, 0.01-2 mg/L TP EZ7800.99
Total Phosphorus, 0.025-5 mg/L TP EZ7801.99
Total Phosphorus, 0.05-10 mg/L TP EZ7802.99
X X X X X 2
Total Phosphorus, 0.1-20 mg/L TP EZ7803.99
Total Phosphorus, 0.01-2 mg/L TP / 0.005-1 mg/L PO4-P EZ7850.99
Total Phosphorus, 0.025-5 mg/L TP / 0.005-1 mg/L PO4-P EU7851.99

Measurement range settings / Dilution options


Standard range 0

Power supply
230 VAC, 50/60 Hz A
120 VAC, 50/60 Hz B

Number of sample streams


1 stream 1
2 streams 2
3 streams 3
4 streams 4
5 streams 5
6 streams 6
7 streams 7
8 streams 8

Outputs
1x mA 1
2x mA 2
3x mA 3
4x mA 4
5x mA 5
6x mA 6
7x mA 7
8x mA 8
RS232 A
Modbus TCP/IP B
Modbus RS485 C
1x mA + Modbus RS485 E
2x mA + Modbus RS485 F
3x mA + Modbus RS485 G
4x mA + Modbus RS485 H
DOC053.52.35160.Mar20

1x mA + Modbus TCP/IP I
2x mA + Modbus TCP/IP J
3x mA + Modbus TCP/IP K
4x mA + Modbus TCP/IP L

No adaption, standard version 0

www.hach.com