71% found this document useful (7 votes)
4K views

Problem ch1

The heritage of Lincoln's mother Nancy remains unclear, but it is widely assumed that she was the daughter of Lucy Hanks.[7] Thomas and Nancy married on June 12, 1806, in Washington County, and moved to Elizabethtown, Kentucky.[8] They had three children: Sarah, Abraham, and Thomas, who died an infant

Uploaded by

Mohammad
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
71% found this document useful (7 votes)
4K views

Problem ch1

The heritage of Lincoln's mother Nancy remains unclear, but it is widely assumed that she was the daughter of Lucy Hanks.[7] Thomas and Nancy married on June 12, 1806, in Washington County, and moved to Elizabethtown, Kentucky.[8] They had three children: Sarah, Abraham, and Thomas, who died an infant

Uploaded by

Mohammad
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 3

Problems

1.11 A temperature sensor has a range of 0 to 120°C and an absolute accuracy of ±3°C.
What is its FSD percent accuracy?
absolute accuracy ±3
% FSD= × 100 %= ×100 %=± 2.5 %
full scale 120

1.12 A flow sensor has a range of 0 to 25 m/s and a FSD accuracy of ±4.5 percent. What is
the absolute accuracy?
|.| Acc .
%FSD=± 4.5 %= 100 % →|.| Acc .=± 1.125 m/s
25

1.13 A pressure sensor has a range of 30 to 125 kPa and the absolute accuracy2 kPa. What is
its percent full-scale and span accuracy?
±2 |.| Acc . ±2
%FSD= 100 %=± 1.6 % Span Acc .= 100 %= =± 2.1 %
125 Span 125−30

1.14 A temperature instrument has a range −20°F to 500°F. What is the error at 220°F?
Assume the accuracy is (a) ±7 percent of FSD and (b) ±7 percent of span.
a ¿|.| Acc .=± 0.07∗500 F=± 35 F b ¿|.| Acc=± 0.07∗520 F=± 36.4
1.15 A spring balance has a span of 10 to 120 kg and the absolute accuracy is ±3 kg. What is
its %FSD accuracy and span accuracy?
±3 ±3
%FSD= ∗100 %=2.5 % Span Acc .= ∗100 %=±2.7
120 110
1.16 A digital thermometer with a temperate range of 129.9°C has an accuracy specification
of ±1/2 of the least significant bit. What is its absolute accuracy, %FSD accuracy, and its
resolution?
0.5
Resolution=±0.5 C |.| Acc=± 0.5 C %FSD= ∗100 %=0.38 %
129.9

1.17 A flow instrument has an accuracy of (a) ±0.5 percent of reading and (b) 0.5%FSD. If
the range of the instrument is 10 to 100 fps, what is the absolute accuracy at 45 fps?
a ¿|.| Acc at 45 fps=0.005∗45 fps=± 0.225 fps
b ¿|.| Acc=0.005∗100 fps=±0.5 fps

1.18 A pressure gauge has a span of 50 to 150 psi and its absolute accuracy is ±5 psi. What
is its %FSD and span accuracy?
±5 ±5
%FSD= ∗100 %=3.33 %FSD Span Acc .= ∗100 %=5 %
150 100

1.19 Plot a graph of the following readings for a pressure sensor to determine if there is
hysteresis, and if so, what is the hysteresis as a percentage of FSD?
Graph of Reading in Problrm 1.19
100

90

80

70
Gauge Reading (kPa)

60

50

40

30

20

10

0
0 10 20 30 40 50 60 70 80 90 100
True Reading (kPa)

±9 ± 10
%FSD= 100 %=9%FSD hysteresis %FSD= 100 %=10 % nonlinearity
100 100
1.20 Plot a graph of the following readings for a temperature sensor to determine the
linearity of
the sensor.
What is
the

nonlinearity as a percentage of FSD?

Graph of reading for problem 1.20


120

100

80
Gauge Pressure (kPa)

60

40

20

0
0 10 20 30 40 50 60 70 80 90 100