M96116 MF Operator
M96116 MF Operator
Spectroil M/F-W
Fuel Analysis Spectrometer
160 Ayer Road Littleton, MA 01460 U.S.A. • Tel. 978.486.0123 • Fax 978.486.0030
email: [email protected] • web page: www.spectroinc.com
Table of Contents
1.0 GENERAL OPERATING REQUIREMENTS ................................................................. 1
1.1 Power Application and Systematic Power Removal............................................................. 1
1.2 Rod Electrode Sharpening ................................................................................................. 3
1.3 Installing the Disc Electrode............................................................................................... 4
1.4 Installing the Rod Electrode and Setting the Gap............................................................... 4
1.5 Installing and Positioning the Sample Holder and Cover.................................................... 5
1.6 Cleaning the Sample Stand . .............................................................................................. 7
1.6.1 Cleaning After Each Burn Cycle ........................................................................................ 7
1.6.2 Cleaning After Each Operating Shift .................................................................................... 7
1.6.3 Cleaning the Quartz Window . .......................................................................................... 8
1.6.4 Cleaning Solutions ............................................................................................................ 8
1.7 Paper Tissue for Operating and Paper Towels for Cleaning ............................................... 8
1.8 Waste Oil Disposal Container ........................................................................................... 8
Modification Description
Mod 0 Original CID version of Spectroil M
Mod 1 Upgrade with SFTM port and frequency adjustment potentiometer
Mod 2 Addition of solid state excitation ignition module and SFTM port
Mod 3 Upgrade to combined solid state source
OilMWindows Upgrade to Windows hardware and software
Mod 4 Upgrade to panel PC hardware and Windows XP
Mod 5 CE version of the Spectroil M
Mod 6 CCD Optic and New Software v. 5, starting w/serial number 6001
Mod 7 C.E. version of Mod. 6
WARNING!!!
High Voltages are Present During the Operation of the
Spectroil M!
Definitions
WARNING!!!
An operating procedure or practice that may cause injury if not carefully observed or
followed.
CAUTION!!!
An operating procedure or practice that may cause damage to the LNF if not carefully
observed or followed
NOTE!!!
An operating procedure or practice that is essential to emphasize
Software CAUTION!!!
The Spectroil M computer is capable of running multiple software
applications and/or operating systems. However, as designed, the
computer processor is dedicated to the operation and control func-
tions of the Spectroil M. Do not attempt to add any software or al-
ter the original factory installed software without checking first with
the Spectro Inc. Service Department.
Although this manual frequently refers to the military “D” series of standards, the op-
erator procedures are identical for all types of Spectroil M spectrometers. Commercial
customers should substitute their equivalent “V” or “S” series of standards throughout
this manual.
WARRANTY
The warranty period of the Spectroil M family of spectrometers is twelve (12)
months from date of installation or fifteen (15) months from date of shipment,
whichever occurs first. Spectro warrants the Spectroil under conditions of opera-
tion against defects of materials and workmanship. All defective material will be
replaced providing damage was not caused by improper use. Warranty applies to
parts and labor only.
Fuel Analysis Spectrometer This manual provides the Spectroil M/F-W opera-
tor with routine instructions on how to set-up the
spectrometer, analyze fuel samples, and perform
routine maintenance. The instructions apply to
the Spectroil M/F-W Fuel Analysis Spectrom-
Electrode Sharpener
Power Connection
initialize and automatically load the application. maximum instrument stability. When not in
Simultaneously, the readout system will load use, the MODE switch should be placed in the
Windows XP® and start up the instrument’s appli- STANDBY position.
cation program called OILMWindows®. The in-
strument will boot directly to the Analysis Screen, To completely turn off the Spectroil M/F-W, first
Figure 1-2. shut down the OILMWindows software and then
remove power from the instrument.
If the system fails to establish communication, a
screen similar to Figure 1-3 appears to select the CAUTION: Utilizing the Windows XP® oper-
Correct Configuration File Path. This will hap- ating system dictates a specific series of steps to
pen if the system cannot find a file in case it hasbe performed in the process of shutting down
been moved, updated or is corrupted. Refer to the Spectroil M/F-W. If power is accidentally re-
the Maintenance Manual Section 2.4.4.6 for as- moved from the Spectroil M/F-W spectrometer
sistance to diagnose and correct this condition. or the circuit breaker CB1 is shut off while the
readout system is running Windows XP® or the
Next, move the MODE switch to the OPERATE OILMWindows® application, an orderly shut-
position and power will be applied to the excita- down would not be performed and as a result,
tion source and electrode sharpener. A noticeable the Windows XP® operating system must perform
increase in fan activity will be observed; this is hardware and software diagnostics when power is
normal. reapplied.
At this point, the instrument is ready to begin To prevent this from happening, always follow the
operation. Some time will be needed before the next steps to shut down the Spectroil M/F-W.
instrument stabilizes after power is applied. It is
recommended that the main power remain on 1. Shut down the OILMWindows® applica-
when the instrument is not in use to maintain tion by left clicking the close box (box with
an “X”)in the upper right corner of the
OILMWindows® header, or choose File/Exit
from the pull down menu options. This
will return the software to the Windows XP®
desktop.
er CB1 in the down position to remove all base of the electrode sharpener. The sound of the
power from the instrument. electric motor should be heard and a slight vibra-
tion should be felt through the motor. The mo-
If, by accident, power was removed from the in- tor will continue to run on a self-timed cycle for
strument, a series of diagnostics will automati- approximately 3 to 5 minutes. Insert the rod into
cally be performed when power is reapplied. It is the rotating electrode guide hole until it comes
extremely important to allow these diagnostics to in contact with the cutter blade. Apply inward
complete in their entirety before loading any ap- pressure until approximately 1/8 to 3/16 inch (3
plication software. If problems are experienced to 5 mm) is cut from the end of the rod. Slightly
during the process of running these diagnos- decrease the inward pressure on the rod electrode,
tics, contact the technical service department of but still maintain its contact with the cutter blade.
Spectro Incorporated for instructions on how to This will polish the rod electrode tip.
recover and proceed with normal operation.
Remove the rod electrode and visually inspect
To restore power to the instrument, follow the in- the tip. It should have a clean cut with no ap-
structions given above. parent chipping around the circumference of the
rod. The surface should be very smooth and have
a polished mirrored looking surface. If the qual-
1.2 Rod Electrode Sharpening ity of the surface is not as described, insert the
rod into the sharpener and repeat the cutting and
The rod electrode, along with the disc electrode, polishing procedure. Remove the rod, inspect
form the analytical gap through which the oil or the surface quality and if acceptable, place the
fuel sample are passed for analysis. An alternat- rod electrode into the original box for storage un-
ing current discharge will occur between the disc til ready for use. To prevent contamination of a
and rod electrode and vaporize the sample and the sharpened rod electrode, do not touch the tip or
metallic components in it. This is the basis of the edge of the tip of the sharpened electrode with the
arc emission technique. fingers or metallic surfaces or anything but a fresh,
clean laboratory grade paper towel. Do not use a
The preparation of the tip of the rod electrode rod electrode for analysis if the surface appears to
plays a significant role in obtaining repeatable an- have been damaged. Refer to Section 7.6.5.1 for
alytical data. The rod electrode must be cleaned the procedure to change/rotate the cutter blade.
prior to inserting it into the electrode sharpener.
T his is accomplished by taking a clean paper NOTE: The electrode sharpener can be a potential
towel and removing the components of the burn source of contamination that can result in degraded
residue from the previous analysis. Remove all accuracy. The electrode sharpener must be emptied
residue from the tip and sides of the electrode by and cleaned frequently.
rotating the rod in the paper towel while applying
pressure with the fingers of the opposite hand. The rod electrode sharpener has a graphite collec-
tor barrel assembly which is secured to the motor
NOTE: The paper towel should be laboratory grade mount and face plate by a rubber o-ring. After
and free of silicon. Do not use household grade approximately 250 sharpenings, this graphite col-
paper towels. lector will need to be emptied. To remove this
barrel assembly, locate the sharpener over a waste
With the spectrometer on, turn the MODE basket with the collector barrel pointing down-
switch to OPERATE. This applies power to J2, ward. Grasp the barrel with the opposite hand
the electrode sharpener power connector. Mo- and rotate it while pulling it away from the motor
mentarily press the power switch located on the mount and face plate. Once the o-ring disengages
from the collector barrel, it will be easy to sepa- See Section 1.7 for a description of the laboratory
rate and empty. grade paper towel. Pour out a few disc electrodes
onto a clean laboratory grade paper towel. Take
While the collector barrel assembly is removed, a laboratory grade paper towel and double it to
check the rod electrode cutter head and blade. be sure that no contamination from the fingers
There should not be an accumulation of oil or will be absorbed into the disc. If large size paper
carbon in this area. If an oily residue is present, towels are used, they should be cut with scissors
this is an indication that the rod electrodes are not into two inch squares to facilitate easy handling.
being adequately cleaned before insertion into Place the towel over the disc electrodes, and with
the electrode sharpener. To clean this assembly, the forefinger and thumb, grab one disc from the
remove the cutter blade and use general purpose pile and place the disc on the shaft, Figure 1-4.
spray and wipe detergent. Reassemble the cutter With firm pressure, push the disc electrode onto
blade and the graphite collector barrel assembly. the shaft until it comes to rest against the index
shoulder of the shaft. If the disc electrode does
The cutter blade, part number M90102, for the not offer some resistance to the shaft as it is be-
rod electrode sharpener is a three-sided tungsten ing inserted, remove and discard this electrode
cutter tool commonly used in machining opera- because the inner diameter has not been made to
tions. One side is capable of approximately 1000 the tolerances specified.
sharpenings. The edge of a cutter is worn when
the initial cut of the electrode requires an abnor- CAUTION: The disc electrode shaft is designed
mal amount of inward pressure, and/or a smooth to be replaced by the operator using a small
reflective surface cannot be achieved on the tip jeweler’s screwdriver. The shaft has right hand
of the rod electrode. Refer to Section 3 for the threads for tightening it into the commutator.
procedure to replace the cutter blade. When pushing the disc electrode on the shaft, do
not apply counter-clockwise rotation on the disc
The electrode sharpener power is on a timing electrode as this may cause the disc electrode shaft
circuit and will turn off after approximately 3 to to loosen.
5 minutes. The electrode sharpener can also be
turned off by placing the MODE switch in the NOTE: Loose disc electrodes will produce errone-
STANDBY position. ous results. If the disc electrode is too loose, arcing
will occur between the inner diameter of the disc
and the outer diameter of the shaft.
1.3 Installing the Disc Electrode
CAUTION: The disc electrode may be very hot
The disc electrode is the most significant contrib- to the touch.
utor to the accuracy and repeatability of the in-
strument. They are manufactured and then puri- Use a towel to remove a disc electrode from the
fied to strict specifications to ensure that they do shaft after an analysis, and to wipe away any oil
not contain unacceptable levels of trace element which may have spilled over from the burn.
contamination for the elements of interest. The
care taken to properly install the disc on the shaft
will help to ensure that excitation parameters will 1.4 Installing the Rod Electrode and Setting
be kept as constant as possible, thus resulting in the Gap
repeatable analytical data.
The rod electrode is installed after the disc elec-
To install the disc electrode on the shaft, a labo- trode is already in place. To install the rod elec-
ratory grade disposable towel is recommended. trode, take the rod in the fingers of the right hand
Rod Electrode
Sample Table
Positioning
Figure 1-5a, Sample Stand with Disposable Sample Figure 1-6a, Sample Stand with Disposable Sample
Holder in Place Holder and Cover in Place
WARNING
PROLONGED CONTACT WITH SOME SOL-
VENTS AND OILS MAY CAUSE CANCER!
Figure 1-6b, Sample Stand with Reusable Sample
Holder and Cover in Place
the sample holder with the fuel sample. Do
WARNING
not overfill the sample holder.
DO NOT USE ANY CHLORINATED SOL-
VENTS INTERNALLY OR EXTERNALLY ON
3. Place the cover over the reusable sample
THE INSTRUMENT!
holder and disc electrode. Note that the
cover only fits in one direction and has a
CAUTION: All chemicals should be used in ac-
cutout for the disc electrode shaft.
cordance with good laboratory practice. Proper
ventilation is required when using any solvent.
4. Install the rod electrode and set the gap, Sec-
Skin contact and prolonged exposure to fumes
tion 1.4.
produced by any solvent may be hazardous.
The sample is now ready for analysis.
1.6.1 Cleaning After Each Burn Cycle
Take the paper towel used to remove the disc elec-
When the analysis is complete, open the sample
trode from the shaft and clean the shaft, the sam-
stand door, remove the rod electrode, remove the
ple table, and the sample plate area between the
cover, lower the sample table, remove the sample,
disc electrode shaft and the rod electrode clamp.
and remove the disc electrode. The sample stand
is now ready for the next analysis.
1.6.2 Cleaning After Each Operating Shift
After 8 hours of operation, the complete sample
stand area must be wiped clean of the oil film
1.6 Cleaning the Sample Stand
buildup created by the burn cycles. If performed
routinely, the sample stand can be cleaned simply
The Spectroil M/F-W incorporates the rotat-
with paper towels and moderate rubbing. How-
ing disc arc emission technique for excitation of
ever, if this procedure is performed sporadically
the fluid sample. This technique produces a fine
or inadequately, an oil dispersant may be required
carbon residue which, when combined with oil
to remove the buildup. A general purpose foam
droplets, produces an oil coating over the sample
type spray detergent is recommended to dissolve
stand and door area. If allowed to accumulate, this
the oil film buildup. A spray detergent is capable
coating will collect the carbon particles and even-
of contacting those areas which are hard to reach.
tually produce a lower resistance path than the
Remove all detergent by wiping dry with paper
towel.
23 December 2009 / v.3.4.1
8 | Operator’s Manual
1.6.3 Cleaning the Quartz Window 1.7 Paper Tissue for Operating and Paper
The quartz window that protects the lens and fi- Towels for Cleaning
ber optic must be cleaned frequently depending
upon the type of fluid being analyzed. In general, Disposable paper tissues and towels are recom-
this should be done at least every 5 burn cycles. mended for use in the daily operation of the
To clean the protective quartz window, take a Spectroil M/F-W. The type of paper tissue used to
clean, soft, disposable laboratory tissue and wet handle the disc electrodes is very important. Most
one corner of the towel with isopropyl rubbing household tissue paper is treated with certain ele-
alcohol or ammonia based window cleaner. With ments to make it softer or more absorbent. If used
the forefinger, rub the wetted portion of the pa- to handle the disc electrodes, these elements will
per towel along the surface of the window while contaminate the electrodes and produce erratic
applying moderate clockwise pressure on the results, especially for silicon. Therefore, it is rec-
window. This will disperse the oil film. Now take ommended that a laboratory grade paper tissue be
the dry portion of the paper towel and repeat this used for this operation.
procedure until no oil can be seen on the tissue
paper. A cotton swab can also be used for this Paper towels are useful to clean the sample stand
purpose. A diluted solution of ammonia and wa- components and wipe spills which occur during
ter may be used. routine operation. The type of paper towel used
for this function is not critical. Typical household
CAUTION: The lens protected by the window towels or C-fold janitorial towels work best for
does not require cleaning and should only be this function because of their absorbent character-
disassembled by a qualified engineer. istics.
tissue.
2.0 DAILY OPERATION
3. Turn printer ON and check to see that suf-
This section details those procedures that will be ficient paper is available. If the printer has an
routinely used in the day-to-day operation of the ON LINE light, it should be illuminated.
Spectroil MF-LD. The operator must be familiar
with the general operating requirements described 4. Ensure that an ample supply of sample hold-
in Section 1.0. A flow chart of the normal daily ers, sharpened electrodes and discs are on
routines is shown on the next page in Figure 2-1. hand.
For convenience, the parentheses after each step
in the chart refer to the corresponding sections in 5. Select standards for daily use and shake vig-
this chapter. The various procedures are explained orously for at least 30 seconds.
in brief, easy to follow step-by-step instructions.
6. Have an oil waste container on hand (Sec-
The spectroil M/F-W may be equipped with the tion 1.8).
optional sulfur capability. The procedure to ana-
lyze samples for sulfur is described in Section 2.9 7. Have cotton swabs, contaminant free tissue
of this manual. paper and paper towels on hand (Section
1.7).
Daily
Daily Cleaning
Standardization Routine Sample
Check, (2.4)
OK Analysis, (2.3)
Prior to Securing,
(2.6)
Not OK
Standardization, (2.5)
Daily
Standardization
Check, (2.4)
OK
Not OK
Standardization, (2.5)
Daily
Standardization
Check, (2.4) OK
Not OK
Acceptable
Accuracy Test, OK
(4.2)
Contact Spectro
Not OK Incorporated
2.3 Routine Fuel Sample Analysis 2.3.1 Routine Fuel Sample Analysis of
Undiluted Samples
This paragraph gives the steps to follow to analyze The Spectroil M/F-W is set-up at the factory
or “burn” a fuel sample or a calibration standard with two routine analysis programs, one for light
using the disposable sample holders. Refer to the fuel and the other for heavy fuel. The appropri-
referenced sections for details. The various parts ate analysis program must be selected to coincide
referred to are shown with labels in Figure 2-1. with the type of fuel sample to be analyzed.
Refer to Section 2.9 if sulfur in fuel is to be ana-
lyzed. NOTE: When a new lot of disc electrodes is started,
either from a new manufacturer or a different lot
Most fuel samples can be analyzed without any from the same manufacturer, the disc electrode offset
preparation by pouring them into a ample holder. procedure of Section 4.4 must be performed.
However, #6 residual fuels must diluted in a 1:1
ratio with base oil prior to analysis, see Section NOTE: Rubber gloves or finger tip covers such
2.3.2. as Anti-static Nitrie COTS should be worn when
handling fuel samples, standards and consumables.
Note On Sample Contamination This will protect the user and minimizes contami-
nation.
The data generated by an analytical instru-
ment is only as good as the samples that are
The Spectroil M/F-W is set-up at the factory
provided. It is imperative that the fuel sam-
ple be taken from the same sampling point with two routine analysis programs, one for light
and in an approved and repeatable manner fuel and the other for heavy fuel. The appropri-
such as ASTM D 4057 Manual Sampling of ate analysis program must be selected to coincide
Petroleum and Petroleum Products. The with the type of fuel sample to be analyzed.
sampling container must be clean and free of
any contamination. In the laboratory, sample 1. The video monitor should display the Analy-
handling and analytical techniques are also sis Program screen, Figure 2-3. If a screen
important. Incorrect or careless procedures saver is in use, the Analysis Program screen
can influence analytical data that may lead to will not be displayed. Press any key on the
inaccurate data.
keyboard to terminate the screen saver and
re-display the Analysis Program screen.
Contamination through careless operation
can also be a source of error. The fuel sam-
ple and the graphite electrodes of the RDE The currently selected program is listed in the up-
spectrometer should never be touched by per left hand corner of the screen below the icons
hand. The electrodes must always be in-
stalled with a laboratory disposable towel to
avoid contamination from the fingers. Per-
spiration from the body will easily add sev-
eral parts per million of sodium to the analy-
sis. The the rod electrodes must be wiped
off before sharpening and the electrode
sharpener must be emptied and cleaned fre-
quently. Resulting errors, when not detected
can lead to erroneous conclusions and cost-
ly consequences about the status of the fuel
or the fuel treatment system. Although the
fuel analyzer is simple to operate, proper
housekeeping procedures must always be Figure 2-3, Sample Analysis Screen
followed.
23 December 2009 / v.3.4.1
12 | Operator’s Manual
3. Place a disposable sample holder on the 5. Raise the sample table using the sample table
sample holder adopter and fill the holder positioning lever (sample fluid will contact
with sample to be analyzed, Figure 2-5. Be bottom of disc) (Section 1.5). See Section
sure to always fill sample holders to the rim 2-8 if Sample ID’s are to be entered.
(Section 1.5).
6. Place the cover on the disposable sample
NOTE: The adopter is not required when reusable holder and disc electrode, Figure 2-6. Note
sample holders are used. that the cover fits only in one direction and
has a cutout for the disc electrode shaft.
4. Place the adopter with the filled sample
holder on the table and slide it back to the 7. To install the rod electrode, press inward on
end of the groove on the table (Section 1.5). the black plastic knob of the spring loaded
rod electrode clamp to open the jaws of the
clamp. Insert a graphite rod electrode until
the tip of the carbon rod is in contact with
the disc electrode, then release the knob to
secure the electrode in the clamp.
14. Using a tissue or paper towel, wipe ex- Heavy fuel analysis, such as #6 residual, can be
cess, spilled or splattered sample fluid from accomplished using the RDE/AES analytical
sample table and disc electrode shaft. technique of the Spectroil M/F-W. The RDE/
AES technique has sufficient sensitivity as well as
NOTE: The quartz protective window should be dynamic range to detect low trace elements such
cleaned every burn. as sodium, potassium and lithium as well as con-
centrates of vanadium, lead, calcium, nickel etc.
15. The Spectroil M/F-W is now ready to analyze
another sample by again following the above 14 Due to the complex matrix of #6 residual fuels, all
steps. samples must be properly prepared through sam-
ple dilution prior to analysis in order to achieve
To improve precision and to comply withe ASTM full excitation of the elemental constituents. This
D 6728, the sample should be analyzed at least procedure has been developed to provide step-
three to five times and averaged. The spectroil by-step instructions on sample preparation of
M/F-W will calculate an average for the analyses #6 residual fuels for elemental analysis using the
on the screen when function key 6 (F6) or the Spectroil M/F-W spectrometer.
average icon are pressed.
2.3.2.1 Pre-Dilution Sample Heating
NOTE: A copy of ASTM D 6728 Standard Test Depending on the source of the crude oil that re-
Method for Determination of Contaminants in Gas sults in refined products, the heavy volatiles that
constitute #6 residual fuels may or may not be the Spectroil M/F-W is also used as the diluent.
sufficiently viscous to pour at room temperature.
The dilution procedure can be performed by the
For good statistical analysis of the elemental con- gravimetric or volumetric method. Both tech-
tent of any #6 residual fuel, it is highly recom- niques will provide valid results, however, gravi-
mended that the sample of #6 fuel is preheated metric is preferred when absolute accuracy is a
and hand or mechanically shaken before decant- requirement. A laboratory balance is required if
ing for sample preparation and analysis. This pro- the dilution is to be performed by the gravimet-
cedure requires a variable temperature hot plate, ric method. For volumetric dilutions a graduated
and a metal container of sufficient size to hold cylinder or similar device can be used to measure
water plus the volume of the sample bottle con- equal volumes of diluent and sample.
taining the #6 residual fuel. A thermometer may
also be necessary if the hot plate does not have a The dilution procedure to prepare sufficient sam-
calibrated temperature control dial. ple for five analyses is as follows:
To heat the sample prior to dilution: 1. Remove the residual fuel sample bottle from
the temperature bath. Tighten the sample
1. Fill the metal container with water to a bottle lid to be sure the contents do not leak
height that will allow approximately 1/4 to out.
1/3 of the sample bottle to be immersed in
the water. Slightly crack the seal on the lid WARNING: The sample is hot! Use gloves to
of the fuel sample to permit gas pressure to protect hands and goggles to protect eyes. Wear a
escape as the sample heats. laboratory coat to protect against accidental spill-
age on body parts.
2. Place the metal container with water and
the fuel sample on the hot plate and adjust 2. Wipe the sample bottle dry then hand or
the temperature control on the hot plate mechanically agitate the fuel sample for
to a temperature of approximately 150° F 30-60 seconds.
(66°C). If the hot plate does not have a cali-
brated temperature dial, use a thermometer 3. Open the sample bottle to decant some of
to determine and control the temperature. the fuel sample.
3. Allow the sample to reach the set point tem- WARNING: Carefully crack open the sample
perature of 150° F (66°C) and stabilize for 5 container lid, contents may be under pressure.
minutes.
4. Using a new and clean disposable pipette,
4. Once the temperature has stabilized, it is decant a minimum of 50g of fuel sample
ready for sample agitation, decanting and into a new (clean) plastic sample container.
preparation. Follow the sample preparation
(dilution) procedure Section 2.3.2.2. 5. Using a new disposable pipette, add 50g of
75Cst base oil used for standardization of
2.3.2.2 Sample Dilution of # 6 Residual Fuel the Spectroil M/F-W spectrometer.
Once the fuel sample has reached the temperature
set point and has stabilized for at least 5 minutes, NOTE: For volumetric dilutions, measure and
it can be removed from the metal container, hand pour equal amounts of 75Cst base oil and fuel sam-
or mechanically shaken, then decanted for dilu- ple into a clean container. Approximately 15 ml of
tion. The 75Cst base oil used for calibration of base oil and sample is sufficient for five analyses.
2.5 Complete Standardization Ca, Zn and Mg are standardized using the 10/30
ppm standard. The 10/30 ppm standard contains
Complete standardization is a procedure per- 10 ppm of all these elements but 30 ppm of Mg.
formed to place the calibration of the instrument A third calibration point with a 100/300 ppm
as close to the standard values as the instrument standard is used to standardize all other elements.
originally produced during factory calibration. These concentration levels are programmed into
This procedure involves burning oil standards at the computer and are displayed at the appropriate
predetermined points along the calibration curve. time in the following procedure.
After these standards are analyzed, the computer
software will determine mathematical factors to NOTE: Calibration standards are not flammable
correct for any change in the calibration. Com- for normal analysis times and do not require a
plete standardization is performed under the fol- sample holder cover for analysis.
lowing conditions:
From the Analysis Program screen, press function
• When the instrument has been relocated to key 7 (F7), choose the Standardization icon, or se-
another site for operation. This is generally lect the Operations/Standardize pull down menu.
performed after the optical profile procedure The software will automatically clear all previous
has been completed. measurements from the video display. A dialog
• When results from the daily standardization with the name of the first calibration standard the
check fall outside of acceptable limits for instrument will expect to measure will appear. Re-
operation. fer to Figure, 2-7.
• After optical profiling procedure has been
performed. Three options exist when this dialog appears. The
first is to select the OK button which confirms
A complete standardization is performed by that the operator will begin to measure the 0 PPM
burning calibration standards that have been standard. The second option is to select the SKIP
pre-selected during factory calibration of the in- button indicating the operator does not intend or
strument. The concentration levels for complete need to measure the 0 PPM standard and wants to
standardization have been selected based on the increment to the next standard in the standardiza-
application and typical operating range for the el- tion process. The last option is to press CANCEL
ements of interest. and this action will terminate the standardization
routine completely and return the software to the
For the light fuel analysis application, all elements Analysis Program screen.
are standardized with a 0 ppm base oil standard
standard to determine the background level. For
a second calibration point, all other elements are
standardized using the 10/30 ppm standard. The
10/30 ppm standard contains 10 ppm of all ele-
ments but 30 ppm of Mg. These concentration
levels are programmed into the computer and are
displayed at the appropriate time in the following
procedure.
In most cases, the OK button will be selected and one time. This will highlight the burn in a black
the dialog will disappear. Centered just below the background. Pressing the DELETE key on the
tool bar will be the name of the standard, 0 PPM keyboard one time will remove this measurement
in a red banner. This banner will remain there off of the screen. In the event that the wrong
until an average is made and the next standard measurement has been highlighted, position the
will appear. Those elements to be standardized pointer over the measurement again and left click
at this concentration level will have their values the mouse a second time. This will remove the
appear and those elements not standardized will highlight, then select the proper measurement to
have no values appear. reject and press the DELETE key.
NOTE: Reference channels are not standardized If a group of sequential measurements are to be
and therefore will not appear highlighted. rejected, for example measurements 3 through 7,
place the pointer over the first measurement (#3)
1. Select the 0 ppm standard and fill five and left click the mouse one time to highlight the
sample holders. Take care to always fill the measurement. Then move the pointer over the
sample holder to the rim. last measurement (#7) and hold the SHIFT key
down and left click the mouse one time. All mea-
2. With an optical lens cleaning solution (not surements from number 3 through number 7 will
containing silicon), isopropyl alcohol, or a be highlighted. Press the DELETE key to remove
window cleaning solution with ammonia, all five measurements.
clean the quartz window attached to the
Fiber Optic Lens Holder in accordance with 5. After making as many measurements as
Section 1.6.3. necessary to obtain a good average, press
function key 6 (F6), click the average icon,
3. Following procedures set forth in Section or select Operations/Average from the pull
2.3, burn all five samples of the 0 ppm down menu. The average of each element
standard. will be calculated and displayed below each
element’s column of measurements. To
4. On completion of Step 4, look at the read- make a printout of the measurements and
ings on the video screen. If one of the five their average, press the print icon. Auto-
burns does not appear to represent the other matically, the next dialog will appear provid-
four, it may be rejected. ing instructions to make measurement of the
10/30 ppm standard, Figure 2-8.
NOTE: The decision to accept or reject burns dur-
ing this procedure is at the discretion of the opera- 6. Select the 10/30 ppm standard and fill five
tor. Quite often an operator will know the cause of
a rejectable burn and therefore reject it almost auto-
matically. Rejectable burns can be caused by incon-
sistencies in consumables such as excess variation in
the specific density (hardness) of the disc electrode,
a loose fitting disc electrode, a poorly sharpened rod
electrode, an under- or over-filled sample holder or
an analytical gapping error.
If the Light Fuel Program is selected, the standard- 15. After making as many measurements as
ization program will continue with step 16. If the necessary to obtain a good average, press
The Heavy Fuel Program is selected, to make a function key 6 (F6), click the average icon,
printout of the measurements and their average, or select Operations/Average from the pull
press the print icon. The procedure continues down menu. The average of each element
with the third calibration standard. will be calculated and displayed below each
element’s column of measurements. If there
Thee next dialog will appear providing instruc- are no additional standards to be measured
tions to make measurement with the next stan- as part of the standardization routine, next
dard. The name of the standard 100/300 ppm dialog to appear will indicate standardization
Figure 2-9. is complete and inquire if the average and
burns for the last standard measured should
11. Select the 100/300 ppm standard and fill be printed, Figure 2-10.
five sample holders. Take care to always fill
the sample holder to the rim. 16. Click on the Yes button if a printed copy of
the analyses is desired. After clicking on the
12. With an optical lens cleaning solution (not Yes or the No button, the Standardization
containing silicon), isopropyl alcohol, or a Values Screen, Figure 2-11 is displayed. This
table can also be displayed at any time by
6. Clean the quartz protective lens using a NOTE: it is important to only use the standard
clean soft disposable laboratory tissue. specified by the software profiling dialog for the
profiling procedure.
7. Wipe all oil standard bottles clean.
1. To begin the optical profile procedure press
8. Check supply of standards (don’t run out). function key 4 (F4), left click the profile
icon, or select Operations/Profile from the
9. Clean and wipe used oil container. pull down menu options.
10. Clean the working area. 2. The screen shown in Figure 2-13 appears
and calls for the first analysis of the profiling
11. Sharpen all rod electrodes and store them standard..
so they are protected from inadvertent con-
tamination. 3. Fill a sample holder with the profiling stan-
dard and following procedures set forth in
Section 2.3, to burn an oil sample.
2.7 Optical Profiling
4. At the completion of the burn, a screen
The Spectroil M/F-W optical system is shock similar to Figure 2-14 will appear with the
mounted in a light-sealed and environmentally profile log. The profile log shows the current
protected temperature stabilized enclosure. Con- profile for each chip, the previous profile and
sequently, the optics do not need to be profiled the difference between the two. The status
frequently. However, detection limit and repeat- for each chip should be OK, if not repeat
ability suffer when the optics are off profile. Un- the profile procedure one or two more times
fortunately, there is no one rule which ensures that until the status is OK for all chips. If this
the optics are on profile. The following guidelines condition cannot be achieved after three at-
are presented to indicate when profiling should tempts, contact Spectro for assistance.
be done:
• At least once every month. 5. Click on OK to return to the main analy-
• After the instrument has been transported to sis screen. The profiling procedure is now
a new location. complete.
• Whenever the instrument has been subjected
to temperature variations greater than 15° F
(10° C).
Figure 2-15, System/Sample ID Setup Menu Figure 2-16, Single Sample ID Entry Menu
Figure 2-17, Multiple Sample ID Entry Menu Figure 2-19, Sample ID File Selection Screen
to 50 sample identifications to be pre loaded to 2.9 Sulfur Analysis
facilitate rapid sample throughput. All sample ID
fields are configured at the system level through An optional sulfur analysis capability is available
the System/Sample ID menu option. Along the for the spectroil M/F-W. The optional capability
bottom of the Multiple ID dialog are buttons to consists of a separate sulfur optic attached to the
expedite the entry of sample numbers. The Copy right hand side of the sample stand and some mi-
button will copy the contents of one field and per- nor modifications to the sample stand. The sulfur
mit it to be copied into another field of equal or spectral line is in a region of the spectrum where
greater field size using the Paste button. Copy All any signal resulting from the excitation process
will copy one sample and insert it into all remain- is absorbed by air. For this reason, when sulfur
ing empty fields in that column. If that column analysis is required, the sample stand area and the
is set for auto increment, the sample number will optic must be purged with an inert gas so that the
increase one value per row. Insert, Delete, and CCD detector is able to see the sulfur signal gen-
Clear are self-explanatory. erated at the analytical gap. To solve this problem,
the Spectroil M/F-W uses nitrogen as a gas and a
Click load to proceed. The first time that this op- separate optic just for the sulfur capability.
tion is enabled, an input file layout screen, Figure
2-18 appears and must be filled in with the user 2.9.1 Set-up for Sulfur Analysis
layout preferences. Click OK when complete and Follow this procedure to prepare the sulfur optic
the a screen enabling the user to select the sample for routine analysis.
ID files appears, Figure 2-19.
1. Attach a low pressure regulator onto a
Nitrogen bottle or other nitrogen source.
The output of the regulator should have a
1/4 inch barbed quick disconnect, or ferrule
hose fitting
3. To activate the view to the sulfur optic, Figure 2-21, “L” Bracket Installed Covering Lens
the “L” bracket must be removed from the Assembly
sulfur optic light pipe. Unscrew the round knurled nut. Verify that the other “L” bracket over
knurled nut that holds the “L” bracket in the lens assembly is in the “down” position.
place and remove it from in front of the
sulfur light pipe, Figure 2-20 5. Open the nitrogen bottle to allow the nitro-
gen to pass through the regulator, adjust the
4. Relocate the “L” bracket over the lens as- regulator to approximately 25 PSI.
sembly of the routine optic located to the
left of the analytical gap, Figure 2-21. Use 6. Turn the main Nitrogen valve on the sulfur
the round knurled nut to fasten it in the optic control panel to “ON”, Figure 2-22.
new location. This will help to keep the lens
assembly clean during the sulfur analysis. 7. If the Spectroil M/F-WLD is OFF, turn it
ON by raising the circuit breaker (CB1)
NOTE: Some instruments may be delivered with on the right side of the instrument. The
two “L” brackets installed. When that is the case, spectrometer will power up and the OilM
loosen the round knurled nut, flip up the “L” brack- Windows software program will load.
et for the sulfur light pipe and tighten the round
8. Depress the “OPTIC PURGE” button, Fig- Note: During the profiling procedure, both optics
ure 2-22, on the sulfur optic control panel must see the analytical gap. Make sure the “L”
for 5 seconds, and release for 5 seconds. brackets are removed from the sulfur light pipe and
Repeat this process 5 times to flush the opti- the lens assembly. If two “L” brackets are installed,
cal system. they should be in the up position.
9. To get load the Sulfur program select File 2.9.3 Standardization of the Sulfur Optic
from the pull down menu options, then The software must be in the sulfur program to
Open. A list of available programs will be perform this function. If not, see step 9 in Sec-
displayed. Select the radio button for the tion 2.9.1.
Main Sulfur program and press OK. The
sulfur optic is now ready for operation. 1. From the main analysis screen, press func-
tion key 7 (F7), choose the Standardization
WARNING: Irritating and toxic hydrogen sul- icon or select the Operations/Standardize
fide gas may be found in confined vapor spaces pull down menu. The software will auto-
when samples with large percentages of sulfur are matically clear all previous measurements
analyzed. Greater than 15 - 20 ppm continu- from the video display. A dialog box with
ous exposure (in air, not to be confused with the the name of the first calibration standard the
liquid calibration standard) can cause mucous instrument will expect to measure will ap-
membrane and respiratory tract irritation. pear. A blank oil is used or this step and the
50 - 500 ppm can cause headache, nausea, and dialog box may call for “Base Oil”, “Blank
dizziness, loss of reasoning and balance, difficulty Oil”, or 0 PPM Oil.
in breathing, fluid in the lungs, and possible loss
of consciousness. Greater than 500 ppm can cause 2. Following the procedure set forth in Sec-
rapid or immediate unconsciousness due to respi- tion 2.3, burn three samples of Base/Blank
ratory paralysis and death by suffocation unless oil and press function key 6 (F6), click the
the victim is removed from exposure and success- Average icon or select Operations/Average
fully resuscitated. from the pull down menu to average them.
To make a printout of the measurements
The “rotten egg” odor of hydrogen sulfide is not a and their average, press the print icon. Au-
reliable indicator for warning of exposure, since tomatically, the next dialog box will appear
olfactory fatigue (loss of smell) readily occurs, es- providing instructions to make measure-
pecially at concentrations above 50 ppm. At high ments of the next standard. The concentra-
concentrations, the victim may not even recog- tion of this standard is 1,000 PPM The
nize the odor before becoming unconscious. name in the dialog box will vary depending
on the calibration of the Spectroil M/F-W
2.9.2 Profiling the sulfur Optic which can be or light fuel, heavy fuel, crude
The sulfur optic is profiled at the same time as the oil, residual oil or # 2 diesel fuel.
standard optic. Profiling should be performed if
the instrument has been relocated and or shut off 3. Following the procedure set forth in Section
for an extended period of time. 2.3, burn the three samples of the 1,000
PPM sulfur standard and press function key
The Nitrogen purge should be turned OFF at the 6 (F6), click the Average icon or select Op-
sulfur optic control panel during profiling. To erations/Average from the pull down menu
profile the optic, follow the instructions in Sec- to average them.
tion 2. 7 in this manual.
4. The next dialog box that appears will indi- pipe bracket and flip up the lens assembly
cate standardization is complete and inquire bracket.
if the average and burns for the last standard
measured should be printed, select YES. 3. Place the Mode/Operate switch on the
Spectroil M/F-W control panel to STAND-
5. After confirmation that the burn data and BY.
average are to be printed, the last dialog
box to appear will inquire if the Factors and 4. Thoroughly clean the sample stand area, es-
Offsets should be printed, again select YES. pecially the quartz window and gap sensors.
The Standardization procedure is now com- They will be extremely dirty since the nitro-
plete. Review the factor for sulfur; it should gen was blowing the sample flame towards
be close to one, +/- 20 %. the left side of the sample stand during the
sulfur analysis.
2.9.3 Verification
It is good practice to verify calibration after the
Spectroil M/F-W has been standardized. 2.10 Water Analysis
1. In PPM mode, run five samples of each cali- This section describes how to set up the Spectroil
bration standard in the set. Print all burns, M/F-W for the optional elemental analysis of tur-
averages and statistics. Remember the read- bine cleaning water and NOx suppression water.
ings are displayed in percentages.
The optional water method is used for the analysis
2. Depending on the type of standard analyzed, of corrosive elements in water in support of gas
the averages should be better than +/- 15%. turbine operations. The method provides rapid
If not, repeat the standardization procedure, analysis of sodium, lithium, potassium and other
Section 2.9.3 and if it does, proceed with elements in water that is introduced into the en-
step 3. gine combustor for the purpose of suppressing the
oxides of nitrogen (NOx) in the exhaust gases and
3. Routine samples are analyzed for sulfur water used for on-line compressor cleaning and
content per the procedure in Section 2.3, crank-soak compressor cleaning. All the elements
Routine Sample Analysis. are analyzed in less than one minute and without
auxiliary gases or the need for sample prepara-
2.9.4 Clean Up and Final Checks tion.
This section provides the steps that must be per-
formed when the sulfur analysis procedure is com- The RDE-AES technique of the Spectroil M/F-W
plete nd the Spectroil M/F-W will be used for the has been accepted by turbine and fuel treatment
analysis of routine furl samples. system manufactures as the ideal on-site analy-
sis instrument for metallic contaminants in fuel.
1. Turn the Main Nitrogen valve “OFF”. However, turbine manufacturers actually limit
the total concentration of alkali metals ingested
2. Remove the interchangeable “L” bracket by the turbine from all sources. This includes not
from the lens assembly and place over the just fuel, but also contamination in compressor
end of the sulfur light pipe in the sample inlet air, steam/water injection for NOx control
stand. This will prevent any oil from travel- and cleaning water. A need thus exists for the on-
ing down the pipe and contaminating the site analysis of alkali metals in water. .
sulfur optic lens assembly. If two “L” brack-
ets are installed, flip down the sulfur light
2.10.2 Verification
It is good practice to verify calibration after the
Spectroil M/F-W has been standardized.
To maintain the Spectroil M/F-W performance, This section of the manual details the mainte-
periodic maintenance must be performed by the nance action that is required to be performed
operator. This maintenance falls into two catego- on the Spectroil M/F-W at regularly scheduled
ries: intervals. Periodic maintenance will keep the
• Daily Maintenance instrument in good working condition and help
• Scheduled Maintenance to identify sources of future trouble before they
cause serious downtime. The following tables
separate periodic maintenance inspections by
3.1 Daily Operator Maintenance subassembly and item. Each item has instruc-
This section details the maintenance actions re- tions on what maintenance action is required,
quired of the operator on a daily basis. These the interval in which the maintenance should be
maintenance actions pertain mainly to the opera- performed and by which maintenance level.
tor accessible assemblies such as the sample stand,
the readout and control panel, and the automatic The scheduled maintenance tables are as fol-
printer. Any maintenance that is required to be lows:
performed in the excitation source, optics, or • Table 3-2, External Housing Inspections
computer electronics is strictly limited to techni- • Table 3-3, Internal Housing Maintenance
cally skilled personnel. Inspections
• Table 3-4, Excitation Source and Power
Table 3-1 lists each maintenance action that is au- Distribution Maintenance Inspections
thorized and required of the operator. • Table 3-5, Microprocessor Maintenance
Inspections
TABLE 3-1
DAILY OPERATOR MAINTENANCE
CAUTION: Do not use alcohol or Chlorinated Solvents to clean plastic or painted surfaces.
TABLE 3-2
EXTERNAL HOUSING MAINTENANCE INSPECTIONS
TABLE 3-3
INTERNAL HOUSING MAINTENANCE INSPECTIONS
TABLE 3-4
EXCITATION SOURCE AND POWER DISTRIBUTION MAINTENANCE INSPECTION
TABLE 3-5
MICROPROCESSOR MAINTENANCE INSPECTIONS
The cutter blade has three sharpened edges and The purpose of performing a calibration curve
can thus be used three times before it is replaced. verification is to determine if the instrument re-
peats the curve generated at the factory or by an
To replace or rotate the cutter blade to a new cut- authorized service representative.
ting edge, unplug the sharpener power connector
J2 at the power connection plate. Next, remove To perform a calibration curve verification, the in-
the graphite collector barrel assembly to empty strument must first be standardized. Refer to Sec-
out any accumulated graphite and to expose the tion 2.5 in this manual for a detailed procedure
cutter blade, Figure 3-1. on daily standardization. When the instrument
has been standardized using the calibration stan-
To remove the barrel assembly, locate the sharp- dards for the new program, the calibration curve
ener over a waste basket with the collector barrel verification can be performed.
pointing downward. Grasp the barrel with the
opposite hand and rotate it while pulling it away The calibration curve verification consists of per-
from the motor mount and face plate. Once the forming an analysis of each synthesized standard
O-ring disengages the face plate, it will be easy as if it were an unknown sample. It is recom-
to separate and empty. Use a flat blade screw mended that the operator conduct te n analyses
driver to remove the #4-40 screw which mounts of each standard and perform statistics on the
the cutter blade, see Figure 2-14. When replac- measurements to obtain the average and standard
ing or rotating the cutter blade, be sure to place deviation for each element. Instrument perfor-
the rear edge of the cutter blade tight against the mance for alkali metal analysis should be within
cutter head. This is the reference point to achieve the limits listed in Table 4-1.
the correct angle on the graphite rod electrode.
Replace the barrel assembly and reconnect the
sharpener to its power connector. 4.1 Repeatability Testing
where:
N = the number of analyses (normally 10)
Figure 3-1, Electrode Sharpener Blade
K 0.0 to 1.0 ppm, +/- 0.1 ppm 1.0 to 20 ppm, +/- 10% >20 ppm, +/- 15%
Na 0.0 to 1.0 ppm, +/- 0.1 ppm 1.0 to 20 ppm, +/- 10% >20 ppm, +/- 15%
Ca 0.0 to 2.0 ppm, +/- 0.2 ppm >2.0 ppm, +/- 10%
Mg 0.0 to 2.0 ppm, +/- 0.2 ppm >2.0 ppm, +/- 10%
V 0.0 to 8.0 ppm, +/- 0.8 ppm >8.0 ppm, +/- 10%
Pb 0.0 to 8.0 ppm, +/- 0.8 ppm >8.0 ppm, +/- 10%
this contamination.
3. After all five samples have been burned, NOTE: It is extremely important to pay attention
press function key 6 (F6), the average icon, to which elements this ratio is applied. Not all ele-
or Operations/Average to calculate the aver- ments have a background factor as indicated by the
age of the five measurements. Once the value 0.00000. For the elements specified below*,
average is calculated, a dialog with the title do not set the F/R RATIO value in the BKG FAC-
Background Correction Factors will appear TOR column. Failure to do so will adversely affect
on the screen. This table will vary from the analytical results of the instrument!
instrument depending on the analytical con- *Spectroil M/F - MgHi 518.36, C 387.10 and H 486.133.
figuration of the spectrometer; however, the
format is the same as shown in Figure 4-1. 6. The cursor will automatically be located
in the first row of the background factor
4. This dialog will display (from the left) column. To set the new F/R RATIO for all
ELEMENT in the second column, WAVE- elements, left click the Set All button. Au-
LENGTH in the third, FORWARD inten- tomatically, the new factors calculated and
sity in the fourth, REVERSE intensity in displayed in the F/R RATIO column will
the fifth, F/R RATIO in the sixth and BKG appear in the background factor column.
FACTOR in the last column. The cursor
will appear in the upper right corner of the 7. Once all of the elements that have a back-
screen under the column BKG FACTOR. ground factor have been updated to the new
The absolute value for this mode is 1.00000 F/R RATIO, a copy of this screen may be
which indicates that the intensity produced made for your records by left clicking the