Bipap Focus Ventilator: Service Manual
Bipap Focus Ventilator: Service Manual
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 1-I
Copyright © 2006-2014. Respironics California, Inc.
ALL RIGHTS RESERVED.
This work is protected under Title 17 of the United States copyright code and is the sole property of Respironics.
No part of this document may be copied or otherwise reproduced, or stored in any electronic information
retrieval system, except as specifically permitted under United States copyright law, without the prior written
consent of Respironics.
Manufacturer
Respironics California, Inc.
2271 Cosmos Court
Carlsbad, CA 92011
USA
+1-800-345-6443
Australian sponsor
Philips Electronics Australia Ltd
65 Epping Road
North Ryde, NSW 2113
Australia
II BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Table of Contents
Table of Contents
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. iii
Table of Contents
iv BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Table of Contents
10 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
10.1 Ventilator Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
10.2 Patient Circuit Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
10.3 Maximum Pressure Drop for Disconnect . . . . . . . . . . . . . . . . . . . . . . . . . 10-7
10.4 Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7
10.5 Date of Manufacture. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
10.6 Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-9
10.7 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-13
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. v
Table of Contents
vi BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 1. Introduction and Intended Use
The BiPAP Focus Ventilator System provides noninvasive ventilation for adult
(over 30 kg or 66 lbs.) patients for the treatment of respiratory insufficiency
and distress, and obstructive sleep apnea. The BiPAP Focus is appropriate for
acute, sub-acute, and intra-hospital transport settings.
The BiPAP Focus is intended for use with nasal and full face masks, and is not
intended to be a life-support device.
Review the BiPAP Focus Operator’s Manual thoroughly before running tests,
checking operational readiness, or initiating patient use. The operator’s
manual includes important information about ventilator safety, operation, and
applicable warnings and cautions.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 1-1
Chapter 1
Introduction and Intended Use
1.1 Tools and Test Table 1-1 lists the tools, test equipment, and materials required to service and
maintain the BiPAP Focus ventilator.
Equipment
Table 1-1: Test Equipment, Tools, and Materials
Test Equipment
Digital multimeter (DMM) accurate to three decimal Local supplier
places
Electrical safety analyzer Dale LT 5440 or equivalent
Flow source capable of providing up to 120 LPM of air
PC or laptop with: Local supplier
• Windows XP or compatible operating system
• 9-pin serial communication port COM1 (or USB to
serial adapter assignable to COM1)
• USB port
• CD-ROM drive
• Internet connection
Pneumatic calibration analyzer capable of measuring Respironics P/N 1040310 or equivalent
low pressure (cmH2O), flow rate (LPM), volume (liters
BTPS), and respiratory rate (/min).
Respironics utility software (required for calibrating Log on to
and testing BiPAP Focus ventilator) http:// my.respironics.com
to download utility software to a PC.
Ventilator Accessories
Connector, 22-mm (quantity 2) Respironics P/N 1002505 or equivalent
Coupling, silicone (qty. 2) P/N 500-1000-43
Flow control valve Respironics P/N 1006120
Inspiratory bacteria filter, 22-mm M/F (qty. 1) Respironics P/N 1014047
Oxygen enrichment adapter Respironics P/N 312710
Oxygen safety valve Respironics P/N 302418
Patient tubing, reusable, 18-in. (qty. 1) Respironics P/N 1003649
Patient tubing, reusable, 42-in. (qty. 1) Respironics P/N 1003643
Silicone cork Respironics P/N 1001735
Test orifice, BiPAP, 0.25-in. (qty. 1) Respironics P/N 332353
1-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 1
Introduction and Intended Use
Unit of
Range Accuracy
Measure
Pressure -25 to 150 cmH20 +/- 1% of reading or +/- 0.20 cmH20 (whichever is
0-100 psi greater).
Flow (standard) 0-300 SLPM +/- 2% of reading +/- 0.20 SLPM (whichever is greater)
Torque 5 to 25 in-lbs. +/- 1 in-lbs. of reading
Voltage DC: +/- 5 to 50V +/- 2% of reading
AC: 2 to 300 V
1.2 Where to Go for For customer service and product support contact:
Help
Respironics California, Inc.
USA telephone: +1-800-345-6443
Fax: +1-800-886-0245
International telephone: +1-724-387-4000
International fax: +1-724-387-5012
[email protected]
[email protected]
www.philips.com/healthcare
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 1-3
Chapter 1
Introduction and Intended Use
1-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 2. Warnings and Cautions
Warnings, Cautions, and Notes are used throughout this manual to identify
possible safety hazards, conditions that may result in equipment or property
damage, and important information that must be considered when performing
service procedures on the BiPAP Focus.
CAUTION: A condition that could cause damage to, or shorten the service life of,
the device.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 2-1
Chapter 2
Warnings and Cautions
2-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 2
Warnings and Cautions
2.2 Cautions • Federal law (US) restricts this device to sale by, or on the order of,
a physician.
• Failure to replace a dirty filter may cause the device to operate at
higher than normal temperatures and damage the device.
• Use only Respironics-approved filters.
• Use only Respironics-approved batteries.
• The gray reusable air inlet filter must be completely dry before
use. Never place a wet filter into the device.
• Never install more than one gray reusable air inlet filter into the
ventilator. The ventilator is designed to accommodate one gray
reusable air inlet filter and one white disposable ultra-fine filter.
• During battery charging, disconnecting AC mains power before
charging completes may result in a Battery Depleted or Missing
alarm. If a Battery Depleted or Missing alarm occurs, connect AC
power immediately to avoid total loss of power.
• To prevent the premature depletion and reduced life expectancy of
the battery, store the ventilator with AC power connected and the
ON/OFF switch in the ON (I) position. When the ventilator is
stored disconnected from AC power with the power ON/OFF switch
ON ( I ) and in standby, the battery may become depleted within
two weeks. When the ventilator is stored with the ON/OFF switch
in the OFF (O) position, the battery typically becomes depleted in
three months.
If a 343 error code is displayed, remove the battery and charge it
using the BiPAP Focus battery charging cable (P/N 1040421).
This cable provides the initial charge required so that the
ventilator can recharge a depleted battery. Refer to the battery
charge instructions supplied with the cable.
• Replace any filter that is damaged.
• Do not attempt to clean or reuse disposable air inlet filters.
• Electronic components used in this device are subject to damage
from static electricity. Perform repairs only in an antistatic,
electrostatic discharge (ESD)-protected environment.
• Avoid exposing the ventilator to extreme environmental conditions.
Allow the ventilator to return to room temperature before
operation.
• Never place liquids on or near the BiPAP Focus.
• The information in this manual is provided for service personnel
reference.
• Do not pull or crimp cables, tubes, or wires.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 2-3
Chapter 2
Warnings and Cautions
2.3 Notes • Refer to the BiPAP Focus Operator’s Manual for product use,
additional warnings, cautions, and notes.
• Service calibration procedures require a printer to be assigned to
the service PC: this can be an internal print location (for example,
using Adobe PDF) or a printer connected to the PC. If a print error
occurs at the end of the calibration, the calibration table is
cleared and the calibration must be repeated.
2-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 3. Theory of Operation
The BiPAP Focus ventilator provides noninvasive breathing support for adult
patients weighing 30 kg (66 lbs.) or more.
The BiPAP Focus offers two ventilation modes:
• Continuous positive airway pressure (CPAP), which provides a single
level of positive pressure to the patient.
• Spontaneous/timed (S/T), which provides two levels of positive
pressure (one during inspiration and one during exhalation), and
delivers timed breaths if the patient does not initiate a breath.
Ventilator alarms occur when high or low pressure, apnea, patient disconnect,
low power, or loss of AC power conditions occur. The system displays a real-
time estimated delivered pressure bar graph.
The BiPAP Focus includes alarm silence and alarm pre-silence features. A fully
charged backup battery provides a minimum of 45 minutes of backup power at
default settings in case AC power is not available (for example, during
transport within the hospital).
The BiPAP Focus features Digital Auto-Trak, which the ventilator to detect and
compensate for unintentional leaks, and adjust its trigger and cycle algorithms
to maintain optimum performance.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 3-1
Chapter 3
Theory of Operation
3.1 Pneumatic Figure 3-1 shows a pneumatic diagram of the BiPAP Focus ventilator.
System
BiPAP Focus ventilator Diffuser
Bacteria filter
Air inlet (mandatory)
filter(s) Flow sensor
Reusable
3.1.1 Air Inlet Filters
particulate filter
(gray) Air inlet filters:
• The gray reusable filter removes coarse particulate from ambient air
entrained by the blower.
• The white disposable filter is optional, and is recommended for
enhanced particulate filtration.
Disposable particulate
filter (white)
3.1.2 Muffler
The muffler reduces the noise of air flow into the blower. This is done by
channeling the air through a baffled system lined with sound absorbing
material. The channel also provides a path for air recirculation.
Muffler
3-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 3
Theory of Operation
3.1.3 Blower
The blower consists of a speed-controlled DC motor coupled to a single stage
Blower impeller. The motor speed is monitored and adjusted by the controller PCB
based on patient settings. A pressure tap at the blower outlet enables
monitoring and comparison with the proximal (patient) pressure.
Air flow
The air flow sensor measures flow from the flow valve. The ventilator uses this
sensor measurement to provide closed loop control of the flow valve to interpret
patient demand and calculated leak rate.
3.1.6 Diffuser
The diffuser is located on the output of the flow valve and is designed to
provide a reduction of air turbulence.
Diffuser
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 3-3
Chapter 3
Theory of Operation
3.2 Electronic System The BiPAP Focus electronic system controls, monitors, and displays patient
and ventilator data. Two microcontrollers (on the user interface and controller
PCBs) interpret user and sensor inputs while controlling inspiratory and
expiratory pressures according to the selected ventilator settings. Figure 3-2
shows an interconnect block diagram of the BiPAP Focus ventilator.
Exter
nal
AC/
DC
Motorola 68332 microcontroller,
A/D, primary alarm, temperature
monitor, primary alarm driver,
voltage monitoring,
3-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 3
Theory of Operation
J9
VGA/UI
ribbon
cable
J13
switch control
cable
J12
primary alarm
connector
J8
battery cable
J1-interprocessor
communications
cable J10
interconnect
Connects to J1 power cable
and J10 of the
controller PCB 18 VDC to
controller PCB
S1
ON/OFF
switch
J3 J4 J6
serial USB power supply
port port connection
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 3-5
Chapter 3
Theory of Operation
Switch control
cable
Backup alarm
cable
Blower
connection J5
Interprocessor Interprocessor
communications communications
cable J1 cable J10
Connects to user Connects to user
interface PCB (J1) interface PCB (J1)
3-6 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 3
Theory of Operation
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 3-7
Chapter 3
Theory of Operation
3.3 Back Panel The back panel has three connection points and the ON/OFF switch.
• Nurse call station/serial communications port
• USB port
• AC/DC power supply input and power cord strain relief clip
• ON/OFF switch
ON/OFF switch
USB AC/DC power shown in the OFF
port supply input (O) position
NOTE: The ON/OFF switch controls all power management function. Internal
battery charging and ventilator operation are disabled in the OFF (O)
position.
3-8 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 4. Periodic Maintenance
4.1 Preliminary WARNING: To prevent disease transmission, use personal protective equipment when
handling contaminated bacteria filters or other patient accessories. Follow
Cleaning and manufacturers' labeling and institutional guidelines for disposal of
Inspection contaminated accessories.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 4-1
Chapter 4
Periodic Maintenance
4.2 Cleaning the Follow these steps to clean the reusable air inlet filter (Figure 4-1):
Reusable Air Inlet 1. Open the filter door and remove the filter. Inspect for damage or
Filter debris.
2. Wash the filter using one of these solutions, then rinse thoroughly:
• Palmolive® or Dawn® dishwashing detergent
• White distilled vinegar (5% acidicy)
• Isopropyl alcohol (99.9%)
• Hydrogen peroxide (3%)
• Clorox Ultra® Bleach
• Sodium hypochloride (6%)
Filter door
Reusable air
inlet filter
(gray)
WARNING: Failure to replace a dirty filter may cause the device to operate at higher than
normal temperatures and damage the device.
CAUTION: The gray reusable air inlet filter must be completely dry before use. Never
place a wet filter into the device.
4.3 Replacing Air The ventilator is designed to accommodate 1 reusable (gray) filter and 1
disposable (white) filter at one time. The disposable filter is optional but
Inlet Filter(s)
recommended for ultra-fine particulate filtration.
CAUTION: Replace the reusable air inlet filter every year or if damage is visible.
Install only 1 reusable filter.
CAUTION: Replace the disposable air inlet filter if debris or damage is visible. Do not
attempt to clean or reuse.
4-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 4
Periodic Maintenance
4.4 Battery WARNING: To avoid the risk of fire, use only Respironics-approved batteries. All battery
connectors are keyed to ensure proper connection. Do not attempt to connect
Maintenance a battery incorrectly.
NOTE: Recharge the battery if a Low Battery or Battery Depleted alarm occurs
during a pre-operational check.
NOTE: If a Batt. Charge Failure alarm occurs or the battery does not begin to
charge after 1 hours, the battery may be fully depleted: see section 4.4.4.
NOTE: Battery alarms sound near the end of the battery test. Gas delivery
continues until the battery is depleted.
Test the battery once per month. A fully charged battery is designed to support
ventilation for 45 minutes at default settings, and should be replaced if it
cannot.
Approximate Battery
CPAP or IPAP Setting
Operating Time
10 cmH20 60 minutes
15 cmH20 50 minutes
20 cmH20 40 minutes
25 cmH20 30 minutes
30 cmH20 20 minutes
WARNING: When running on battery, the high flows that occur during a patient
disconnect can deplete the battery in 2 minutes. Connect AC power
immediately if the Battery Depleted alarm occurs to avoid total loss of power.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 4-3
Chapter 4
Periodic Maintenance
The nickel metal hydride (NiMH) internal battery automatically charges when
the ventilator is connected to AC power only if the power switch is in the ON (I)
position. The ON AC Power LED lights to indicate that AC power is connected
and the Battery Charging LED lights to indicate that the charging cycle is in
progress. The battery is fully charged when the Battery Charging LED turns off.
The charging cycle operates in Normal and Standby mode. Battery charge time
is typically less than 5 hours, but can be longer depending on ambient
temperature.
4-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 4
Periodic Maintenance
1. Turn the system OFF (O) and disconnect the DC power supply.
2. Remove the two screws that secure the battery compartment door.
3. Remove the old battery: remove from the compartment and unplug the
cable.
4. Install the new battery: plug in the cable and install the battery into
the compartment.
7. Allow sufficient time to fully charge the new battery (Battery is fully
charged when the Battery Charging LED turns off).
Battery compartment
Battery cable
connector
Battery
compartment
door
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 4-5
Chapter 4
Periodic Maintenance
1. Turn the ventilator OFF (O) and disconnect the power cord from the
rear of the unit (Figure 4-3).
ON/OFF switch
AC/DC power Shown in the
supply input OFF position
4-6 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 4
Periodic Maintenance
Battery
compartment
screws
3. Remove the internal battery and disconnect its cable (Figure 4-5).
Disconnect
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 4-7
Chapter 4
Periodic Maintenance
Internal
battery
Battery charging
cable
P/N 1040421
9V
battery
6. Disconnect the battery charging cable and reinstall the BiPAP Focus
battery within 30 seconds.
7. Connect the ventilator AC power cord to wall power and turn the ON/
OFF switch ON (I).
8. Confirm that the green Battery Charging LED lights (Figure 4-7). If
not, repeat the process from step 5 one additional time. If the Battery
Charging LED still does not light after a second attempt, replace the
battery.
Battery
Charging LED
NOTE: A fully-depleted battery may require several charge cycles to reach full
capacity.
4-8 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 4
Periodic Maintenance
4.5 Storing the Store the ventilator in a location that meets environmental specifications. To
avoid premature battery depletion, ensure that the ventilator power switch is in
Ventilator the OFF (O) position.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 4-9
Chapter 4
Periodic Maintenance
4-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5. Diagnostic Mode and
Troubleshooting
This section describes how to use Diagnostic mode for troubleshooting. Use
Diagnostic mode to:
• View product information, including serial numbers and versions of
software and hardware.
• View and adjust system settings, including language, time since last
service, date and time.
• View and adjust preferences, including units of pressure, alarm
volume, contrast, brightness, reverse video and screen lock.
• View the significant event log.
• View real-time sensor readings and status of specific controls.
• Activate and confirm audible, visual, and remote alarms.
• Activate specific controls to test circuits.
WARNING: Diagnostic mode suspends normal ventilation: disconnect the patient from
the ventilator before entering Diagnostic mode.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-1
Chapter 5
Diagnostic Mode and Troubleshooting
3. Press the arrow keys to highlight the menu selection, then press Enter.
4. To exit Diagnostic mode at any time, turn the ventilator off using the
STANDBY key.
5-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
5.2 Product The product information screen (Figure 5-3) displays serial number and
version number information for hardware and software components of the
Information ventilator.
To view the product information screen, use the arrow keys to highlight
PRODUCT INFORMATION on the diagnostic mode main menu, then press
Enter. Press Menu to return to the main menu.
Example 1:
This serial number indicates a manufacture date of 13 March 2009:
S/N 8772 09031301
P/N 1028772
Example 2:
This serial number indicates a manufacture date of 29 July 2006:
S/N 8772 060729-01
P/N 1028772
Date: YY/MM/DD Unit number
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-3
Chapter 5
Diagnostic Mode and Troubleshooting
5.3 System Settings Use the System Settings screen (Figure 5-4) to select language, operating time
(total and service), and date and time. Follow these steps:
1. Use the arrow keys to highlight the parameter to change, then press
Enter.
2. Use the arrow keys to adjust a value or select a setting from a drop-
down menu, then press Enter (or press Cancel to exit without making a
change).
5-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
5.4 Preferences Use the Preferences screen (Figure 5-5) to select display units, alarm volume,
contrast, brightness, reverse or normal video, and the screen lock timer
setting.
1. Use the arrow keys to highlight the button to change, then press Enter:
2. Adjust the value and press Enter or press Cancel to exit without
making a change.
• Use Display Units to select cmH2O, hPa, or mbar as the unit of
pressure.
• Use Alarm Volume to adjust the volume from 1 to 5. The alarm
volume cannot be turned completely off.
• Use Contrast to adjust the contrast from 20% to 100%.
• Use Brightness, to adjust the contrast from 10% to 100%
• Use Reverse Video to toggle between reverse and normal video.
• Use Screen Lock Timer to turn the screen lock on or off.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-5
Chapter 5
Diagnostic Mode and Troubleshooting
5.5 Significant Event The significant event log (Figure 5-6) displays a log of the 200 most recent
system events, including settings changes, alarms, and codes that can be used
Log for troubleshooting (section 5.10 describes significant event codes in detail).
Use the arrow keys to scroll up and down. Press Menu to return to the main
menu.
5-6 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
5.6 Sensor Readings The Sensor Readings screen (Figure 5-7) displays real-time system voltages,
A/D counts, temperature, and status (1=On, 0=Off)
Parameter Description
+3.3V +3.3 V supply measurement.
Range: 883 to 976 counts
Voltage (V) = counts * 0.003662
+5V +5V measurement.
Range: 884 to 977 counts
Voltage (V) = counts *0.005371
+12V +12V supply measurement.
Range: 889 to 983 counts
Voltage(V) = counts *0.012817
+18Vin +18V (mains input) measurement.
Range: 882 to 1008 counts
Voltage(V) = counts * 0.019043
+18V Boost +18V_Boost supply measurement.
Range: 923 to 1020 counts.
Voltage(V) = counts * 0.019043
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-7
Chapter 5
Diagnostic Mode and Troubleshooting
Parameter Description
VBATT/VTEMP Battery temperature measurement when unit operating on AC.
Battery voltage measurement when operating on battery.
Range (voltage): 624 to 936 counts
Voltage (V) = counts * 0.012817
Range (temperature): 79 to 421 counts (battery temperature
is OK if counts are within limits)
VOptoCap Measurement of backup capacitor voltage on user interface
PCB. Capacitor powers backup alarm for minimum 3 minutes
upon system power failure.
Range: 616 to 782 counts
Voltage(V) = counts * 0.005371
Board Temp Measurement of temperature on the user interface PCB.
Range: 233 to 540 counts
Temperature (oC) = ((counts * 0.00244) - 0.6) * 100
Ambient Temp Measurement of ambient temperature within the unit.
Range: 0 to 60 oC.
+18VINCMP Status of high incoming mains voltage (15.64V typical
threshold).
1: Above threshold
0: Below threshold
+18VBSTCMP Status of high 18V_Boost voltage (15.64V typical threshold).
1: Fault
0: No Fault
BattChrgFault Status of battery charging fault.
1: Fault
0: No Fault
BattOnChrg Status of battery charging.
1: Battery charging
0: Battery not charging
BattChrgToc Status of battery top-off charge.
1: Battery NOT in TOC state
0: Battery in TOC state
BattChrgRdy Status of battery ready to charge.
1: Ready for charging
0: Not ready for charging
5-8 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
Parameter Description
PORLATCH Status of processor reset due to a drop in the +3.3V rail. This
signal must be primed by software, with the
EXT_WDOG_LATCH_CLR signal. This can be done using the
PORLATCLK diagnostics label in the System Test screen.
1: Loss of +3.3V supply caused last reset.
0: Last reset not caused by loss of +3.3V supply. Probably
caused by watchdog (software reset).
PWRSWMON Status of user interface STANDBY key monitor.
1: STANDBY key is inactive (released)
0: STANDBY key is active (pressed)
PWRSWLATCH Status of latched STANDBY key monitor. This signal must be
primed by software with the PWR_SW_CLK signal. This can be
done using the PWRSWCLK diagnostics label in the System
Test screen.
1: STANDBY key has NOT been pressed.
0: STANDBY key has been pressed.
5.7 Setting Up the Follow these steps to connect the ventilator and a service PC to create a
diagnostic report (DRPT):
Serial Interface for
DRPT 1. Use the serial cable from communication kit P/N 1039800 to connect
between the PC and BiPAP Focus ventilator. Connect the male end
labeled Diagnostics to the back of the ventilator and the female end to
the PC as described in Chapter 6.
4. Enter a name for the connection (Figure 5-8 shows BiPAP Focus),
choose an icon, then click OK.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-9
Chapter 5
Diagnostic Mode and Troubleshooting
Figure 5-8: Entering a Name for the Connection to the BiPAP Focus Ventilator
5. Select the appropriate serial port from the Connect using: drop-down
list (Figure 5-9 shows COM1), then click OK.
6. Enter these settings for the serial port (Figure 5-10), then click OK.
8 Data bits
Parity = None
1 Stop bit
5-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
11. Select underline and blink for cursor settings then click OK.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-11
Chapter 5
Diagnostic Mode and Troubleshooting
13. Match the ASCII Setup screen (Figure 5-13), then click OK.
14. Save this configuration. Select File > Save As and save to the desktop
if you want to create an icon for BiPAP Focus Communications on the
Windows desktop.
5-12 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
5.8 Generating a 1. With the BiPAP Focus ventilator in Diagnostic mode and the
Diagnostics end of the serial communications cable connected to the
Diagnostic Report ventilator, start HyperTerminal, open the BiPAP Focus
(DRPT) Communications file, or click on the BiPAP Focus Communications
icon on the PC (if created).
7. Type DRPT (all caps) in the dialog box, and press the Enter key. This
retrieves information from the significant event log.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-13
Chapter 5
Diagnostic Mode and Troubleshooting
NOTE: The DRPT capture file is a TXT file by default: use any word processing
program to view the file.
Figure 5-14 shows how to interpret each line of the DRPT report.
V or T - System errors
V= voltage, T= temperature
(if applicable)
20:42 - Time stamp of occurrence
6/19/09 - MM/DD/YY
NOTE: The last two fields (if applicable) can also show mode/setting changes
and their from/to values.
10. To save the DRPT, highlight the appropriate fields (or Edit>Select All),
copy (Edit>Copy) and then paste into any text editor/word processing/
email program.
5-14 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
5.9 System Test The System Test screen (Figure 5-15) includes:
• Alarms: toggle the primary alarm speaker, backup alarm speaker, and
remote alarm relay ON or OFF.
• Indicators: toggle the Alarm LED, Silence LED, and On Battery LED
ON or OFF.
• Miscellaneous: toggle miscellaneous settings ON or OFF (Table 5-2).
• When V BattMeasEn is On, the Sensor Readings screen displays
battery voltage; when Off, the Sensor Readings screen displays the
internal battery thermistor temperature.
Setting Description
Blower On = Blower on.
Off = Blower off.
V BattMeasEn On = Sensor Readings screen displays battery voltage.
Off = Sensor Readings screen displays battery temperature.
V Out Enable Reserved for future use.
PWRSWCLK On = Clocks “1” to the STANDBY switch latch.
Off = No change.
PWRTUIOFF On = Enters STANDBY mode (removes 18V_Boost, shutting down power).
Off = No change.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-15
Chapter 5
Diagnostic Mode and Troubleshooting
Setting Description
PORLATCLK On = No change.
Off = Clocks “1” to power on reset latch.
Can be used to determine if an external power on reset occurred due to +3.3 V supply drop.
WDSTROBE On = External watchdog strobe on, resets the processor to confirm that system reset occurs correctly.
(The external watchdog strobe should reset the system and log two diagnostic codes in memory:
903 - Ext Wdog Unexpected and 310 POST Abnormal Reset.)
Off = No change.
5.10 Significant The BiPAP Focus Ventilator generates a log of system events that can be used
for troubleshooting, including settings changes, alarms, and codes. If a
Event Codes ventilator inoperative condition or unexpected reset occurs, the significant
event log provides a record of any ventilator setting changes, alarm conditions,
or error conditions that were in effect at the time.
The significant event log records up to 200 of the most recent events, with the
most recent events listed first. The ventilator maintains the significant event
log regardless of whether there is power to the ventilator.
5-16 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
Use the arrow keys to scroll through the log. Table 5-3 lists event IDs and
corrective actions.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-17
Chapter 5
Diagnostic Mode and Troubleshooting
5-18 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-19
Chapter 5
Diagnostic Mode and Troubleshooting
5-20 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
1. Enter Diagnostics mode and use the System Test screen to check if the
backup alarm can be turned on and off.
2. Replace the controller PCB.
3. Retrieve the DRPT and contact technical support.
44 12V REFERENCE OUT OF Controller processor detected that the limits test on the 12V reference voltage
RANGE signal failed.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-21
Chapter 5
Diagnostic Mode and Troubleshooting
5-22 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-23
Chapter 5
Diagnostic Mode and Troubleshooting
1. Enter Diagnostics mode and use the System Test screen to check if the
backup alarm can be turned on and off.
2. If the problem persists replace the controller PCB.
3. Retrieve the DRPT and contact technical support.
77 HIGH PRESSURE CONDITION Controller processor detected a high pressure condition and annuciated the Hi P
Reg alarm. No action necessary. If this condition persists for greater than 10
seconds the system declares error code 41
78 LOW PRESSURE CONDITION Controller processor detected a low pressure condition and annuciated the Lo P
Reg alarm. No action necessary.
79 UNABLE TO MAINTAIN Unable to maintain Pressure Support. System error declared after condition is
PRESSURE SUPPORT present for one minute.
5-24 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-25
Chapter 5
Diagnostic Mode and Troubleshooting
1. No action required if due to the system being powered off using the ON/OFF
(I/O) switch.
2. Otherwise, replace user interface PCB.
2XX: boot block error codes. Not logged to the significant event log because they are detected very early in the power on
sequence. The system annunciates a system error alarm and displays the error code and the text on the screen.
5-26 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
1. Ensure that the download of the released software completes without any
errors reported.
303 POST TCB NOT RESPONDING During POST the user interface processor detected that the controller processor
timed out.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-27
Chapter 5
Diagnostic Mode and Troubleshooting
1. Enter Diagnostics mode and use the System Test screen to test and identify
The screen should also display
the faulty LED.
INDICATOR FAILURE
2. Replace the user interface PCB.
310 POST ABNORMAL RESTART The system has detected an unexpected soft reset.
1. Examine the significant event log for recent error codes that might indicate a
controller PCB or user interface PCB problem.
2. Replace the user interface PCB.
3. Replace the controller PCB.
311 - 331 Unused
332 POST VBATTERY MON FAIL User interface processor detected that the VBATT test failed during POST.
1. If running on battery, ensure that the battery is fully charged.
The screen should also display
BATTERY VOLTAGE FAILURE 2. Check if the problem occurs during AC and battery operation.
3. Enter Diagnostics mode and use the Sensor Readings screen to check the
voltage and ADC readings.
4. Replace the user interface PCB.
333 POST 3_3VMON FAIL User interface processor detected that the 3.3V test failed during POST.
5-28 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
The screen should also display 1. If running on battery, ensure that the battery is fully charged.
18VBOOST VOLTAGE RAIL 2. Check if the problem occurs during AC and battery operation.
FAILED 3. Enter Diagnostics mode and use the Sensor Readings screen to check the
voltage and ADC readings.
4. Replace the user interface PCB.
338 POST SUPERCAPMON FAIL User interface processor detected that the SUPERCAP test failed during POST.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-29
Chapter 5
Diagnostic Mode and Troubleshooting
5-30 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-31
Chapter 5
Diagnostic Mode and Troubleshooting
1. Enter Diagnostics mode and use the System Test screen to check if the blower
can be turned on and off. Also check that the backup alarm can be turned on
and off.
2. Check the interprocessor communications cable.
3. Replace the user interface PCB.
442 BOARD UNDER User interface processor detected that the PCB low temperature test failed
TEMPERATURE during CBIT.
1. Ensure that the system is at room temperature or within operating
specifications.
2. Enter Diagnostics mode and use the Sensor Readings screen to check
temperature and ADC readings.
3. Replace the user interface PCB.
5-32 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-33
Chapter 5
Diagnostic Mode and Troubleshooting
5-34 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 5
Diagnostic Mode and Troubleshooting
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 5-35
Chapter 5
Diagnostic Mode and Troubleshooting
5-36 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6. Ventilator Communications
This chapter describes how to install operating software onto the BiPAP Focus
ventilator.
CAUTION: To avoid possible interruption due to power loss, ensure that the ventilator
is running on AC power (not battery) during this process.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-1
Chapter 6
Ventilator Communications
If your service PC does not have a serial port you can use a USB to serial port
adapter such as Keyspan model USA-19HS. You must install the USB to serial
port drivers included with the adapter before installing the adapter.
NOTE: Do not connect or use the BiPAP Focus USB cable while connected to a
docking station. Do not connect the BiPAP Focus USB cable through a
USB hub. Using docking stations or USB hubs causes communication
errors.
1. Connect the serial cable from BiPAP Focus communications cable kit
(P/N 1039800):
• Connect the double ended serial cable (Figure 6-1) labeled
SOFTWARE/RASP to the ventilator serial port.
• Connect the other end of the cable to the service PC serial port.
6-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
2. Connect the USB cable (from BiPAP Focus communications cable kit
P/N 1039800): insert the Type B end into the ventilator USB port,
and the Type A end to the service PC (Figure 6-2).
3. Ensure that the DC power supply and AC power are connected, then
turn the ON/OFF switch to the ON (l) position.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-3
Chapter 6
Ventilator Communications
4. Press the STANDBY key and then immediately press and hold the UP
and DOWN arrow keys (Figure 6-4) until the message WAITING FOR
DOWNLOAD SERVER appears.
1. Press the
STANDBY key
6-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
6.2 Loading USB Before installing ventilator operating software, USB drivers must be installed
onto the service PC. This is only done once per service PC. The USB drivers are
Drivers to the Service on software CD P/N 1063919.
PC
1. Insert the software CD (P/N 1063919) into the service PC CD-ROM
drive.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-5
Chapter 6
Ventilator Communications
5. At the prompt indicating that the software you will be installing has
not passed Windows logo testing, click Continue Anyway (Figure 6-7).
6-6 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
6. At the Files Needed window (Figure 6-8), click Browse and select the
CD-ROM drive.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-7
Chapter 6
Ventilator Communications
10. Use the file path suggested then click OK (Figure 6-12).
6-8 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
12. At the Files Needed window, click Browse and select the CD-ROM
drive (Figure 6-14).
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-9
Chapter 6
Ventilator Communications
15. Click on the BiPAPFocus.sys file then click Open (Figure 6-17).
16. Use the file path suggested then click OK (Figure 6-18).
6-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
17. The Found New Hardware Wizard begins installing the software
(Figure 6-19).
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-11
Chapter 6
Ventilator Communications
6.3 Installing Follow these steps to install operating software onto the ventilator.
Operating Software 1. With the software CD (P/N 1063919) into the CD-ROM drive, use
on the Ventilator Windows Explorer to view the CD files (Figure 6-21).
6-12 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
3. Select Install new BiPAP Focus Software (Figure 6-23), then click
Next.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-13
Chapter 6
Ventilator Communications
5. When Phase 1 BiPAP Focus Wizard appears, click Next (Figure 6-25).
The software erases the controller PCB software (Figure 6-26), and
then installs new software (Figure 6-27).
6-14 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-15
Chapter 6
Ventilator Communications
6-16 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
9. The Wizard verifies that it can communicate with the ventilator. When
the Next button is enabled click Next (Figure 6-31).
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-17
Chapter 6
Ventilator Communications
6-18 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
12. At the prompt that asks what operation to perform, select I want to
install new software into the ventilator and click Next (Figure 6-34).
13. At available software window, select (in this example) Version 2.00
(File: D:\Loads\P2.00.004. TUI.scl) then click Next (Figure 6-35).
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-19
Chapter 6
Ventilator Communications
14. At the Verify Software Download Settings window (Figure 6-36), click
Next.
6-20 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
17. When BiPAP Focus Wizard Phase 2 Completion window appears, click
Finish (Figure 6-39).
18. The ventilator restarts in Normal mode using the default settings.
Depending on the ventilator configuration, you may need to perform a
calibration as described in Chapter 7.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-21
Chapter 6
Ventilator Communications
6.4 Reprogramming The serial number of each BiPAP Focus Ventilator is stored in non-volatile
memory on the user interface PCB, and must be reprogrammed if you replace
the BiPAP Focus the PCB.
Serial Number
1. Write down the serial number from the bottom of the unit, include
hyphens if the serial number includes them. Do this before
reprogramming the serial number.
6-22 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
5. When the BiPAP Focus wizard asks what operation to perform (Figure
6-41), select I want to perform advanced configuration operations,
then click Next.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-23
Chapter 6
Ventilator Communications
6-24 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 6
Ventilator Communications
10. The ventilator restarts in Normal mode using the default settings.
Depending on the ventilator configuration, you may need to perform a
calibration as described in Chapter 7.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 6-25
Chapter 6
Ventilator Communications
6-26 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7. Testing, Calibration, and
Performance Verification
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-1
Chapter 7
Testing, Calibration, and Performance Verification
7.1 Test Equipment Table 7-1 lists required test equipment and service accessories. Check the
calibration status of all test equipment before use.
and Accessories
Table 7-1: Required Test Equipment and Service Accessories
7-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
7.2 Recommended Table 7-2 summarizes recommended testing following part replacement.
Testing
Table 7-2: Recommended Testing After Parts Replacement
NOTE: Replacing the controller PCB and/or user interface PCB clears the calibration tables. Perform the system
run-in once, between the two calibration procedures. (Do not perform system run-in before the first
calibration.)
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-3
Chapter 7
Testing, Calibration, and Performance Verification
7.3 Ventilator Run-in Perform the ventilator run-in procedure for at least 120 minutes to avoid
possible calibration failures.
Required equipment:
• Test orifice, BiPAP, 0.25-in. (P/N 332353)
Follow these steps:
7-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
7.4 Battery Test WARNING: To avoid the risk of fire, use only Respironics-approved batteries. All battery
connectors are keyed to ensure proper connection. Do not attempt to connect
a battery incorrectly.
NOTE: A Low Battery or Battery Depleted alarm during the pre-operational check
indicates that the battery must be charged. A Batt. Charge Failure alarm
indicates that the battery may be fully depleted. Recharge the battery
before proceeding. If the battery does not begin to charge after one hour
and the Batt. Charge Failure persists, replace the battery.
NOTE: Battery alarms sound near the end of the battery test. Gas delivery
continues until the battery is depleted.
Test the battery once a month. A fully charged battery is designed to support
ventilator operation for 45 minutes at default settings:
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-5
Chapter 7
Testing, Calibration, and Performance Verification
7.5 Real-time Clock Calibration software compares the ventilator real-time clock to service PC time
to determine exactly when an event occurs, Because the calibration fails if the
Calibration times do not match, perform the real-time clock calibration before you proceed
to the calibration procedure.
7-6 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-7
Chapter 7
Testing, Calibration, and Performance Verification
NOTE: The utility software requires you to use COM1 on the service PC. Do not
connect or use the BiPAP Focus USB cable while connected to a docking
station. Do not connect the BiPAP Focus USB cable through a USB hub.
Use of docking stations or USB hubs cause communication errors.
1. Record the serial number and model number exactly as they appear on
the bottom of the ventilator for use during calibration. This
information is required during system calibration (section 7.6).
4. On the service PC, open the Service Center Tools and select Real-time
Clock Calibration, then click the Execute Tool (Figure 7-4).
7-8 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-9
Chapter 7
Testing, Calibration, and Performance Verification
7-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
7.6 System Calibration software compares the ventilator real-time clock to service PC time
to determine exactly when an event occurs, Because the system calibration
Calibration fails if the times do not match, perform the real-time clock calibration (section
7.5) before you start this calibration procedure.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-11
Chapter 7
Testing, Calibration, and Performance Verification
7-12 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
NOTE: The real-time clock calibration (section 7.5) must be complete before
attempting the system calibration procedure.
Required equipment:
• Utility Software, Respironics Utility Tools
(http://my.respironics.com and click on BiPAP Focus Calibration
Software)
• Pneumatic calibration analyzer (P/N 1040310 or equivalent) capable
of measuring low pressure (cmH2O) and flow rate (LPM).
• Flow source capable of providing up to 120 LPM of air to provide
negative/reverse flow
• Windows-compatible personal computer running Windows XP version
software.
• A working default printer connected to the computer before beginning
the calibration procedure.
• BiPAP Focus communications cable kit (P/N 1039800)
• Flow control valve (P/N 1006120)
• Oxygen enrichment adapter (P/N 312710))
• Pressure tubing
• Smooth-bore patient tubing
NOTE: If not already done, perform the ventilator run-in (section 7.3).
1. Record the serial number and model number exactly as they appear on
the bottom of the ventilator.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-13
Chapter 7
Testing, Calibration, and Performance Verification
Pressure calibration:
7-14 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
8. Connect the pressure tubing from the oxygen port of the oxygen
enrichment adapter to the pneumatic calibration analyzer
(Figure 7-12).
9. On the service PC, open the Service Calibration Software and start the
calibration by clicking on the white Run arrow in the upper left corner
of the window (Figure 7-13).
Run
arrow
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-15
Chapter 7
Testing, Calibration, and Performance Verification
10. At the prompt, enter the ventilator serial number exactly as it appears,
then click OK.
11. At the prompt, enter the ventilator model number and click OK.
13. Verify that the label is BiPAP Focus and click PASS.
14. The program runs an alarm test: follow the onscreen instructions.
15. At the prompt (Figure 7-14), attach the manometer and occlude the
test orifice, then click OK to start sensor calibration.
7-16 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
16. Allow pressures to stabilize for at least one minute, then note the
pneumatic calibration analyzer pressure reading; click DECREASE
BLOWER SPEED or INCREASE BLOWER SPEED onscreen
(Figure 7-15) to achieve a pressure of 4.00 cmH2O, then click DONE.
17. Enter in the pressure reading from the analyzer and press ENTER on
the keyboard.
18. The program prompts you to calibrate at various pressure levels using
the onscreen buttons to increase or decrease blower speed to achieve
the correct pressures.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-17
Chapter 7
Testing, Calibration, and Performance Verification
19. The window (Figure 7-17), indicates the start of the flow calibration,
beginning with positive flow.
7-18 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
23. Note the flowmeter reading and adjust the flow control valve to
achieve a flow of 146 LPM, then click OK.
24. The screen then prompts you to adjust the flow valve to achieve
various flow settings. Ensure that each flow setting is achieved before
clicking OK.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-19
Chapter 7
Testing, Calibration, and Performance Verification
25. The window (Figure 7-20) indicates the start of the negative flow
calibration.
26. Connect a length of patient tubing between the flow control valve
outlet and the negative flow source outlet (Figure 7-21). Turn on the
negative flow source and click OK.
NOTE: For the negative flow source, a REMstar device may be used with the
pressure set at 20 cmH2O
7-20 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
27. Adjust the flow control valve until 10 LPM is displayed on the
flowmeter, then click OK (Figure 7-22).
As the program calibrates the negative portion of the flow table, the
flow meter may not display a negative flow. However, pressure against
the normal flow direction is considered a negative flow.
28. Enter the flow reading (Figure 7-23). Be sure to enter the negative or
minus symbol first. The software rejects a positive value at this point.
29. At the prompts, adjust the flow control valve to achieve various
negative flows, enter the flowmeter readings, then click OK.
30. At the Remove Negative Flow Resource prompt, remove the negative
flow source and click OK.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-21
Chapter 7
Testing, Calibration, and Performance Verification
31. Positive flow accuracy: completely open the flow valve and follow the
onscreen prompts to achieve the various flow settings. Click OK only
after the flow has been set.
32. Negative flow accuracy: connect the patient tubing between the flow
valve outlet and the negative flow source. Turn on the negative flow
source and follow the onscreen prompts to achieve the various flow.
Click OK only after the flow has been set.
33. Turn off the negative flow source. Disconnect the patient tubing and
occlude the O2 port at the ventilator outlet port.
35. At the green PASS prompt on service PC, click OK to view the test
data sheet (Figure 7-24). Click OK on the print results button to print
the test results. Figure 7-25 shows an example of a print error.
7-22 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-23
Chapter 7
Testing, Calibration, and Performance Verification
7.7 Electrical Safety The electrical safety test verifies that the forward and reverse leakage current
are within specified limits.
Test
Required equipment:
• Electrical safety analyzer
NOTE: During this test the neutral switch should be closed, not open.
IAW IEC 60601-1 test procedures.
Follow these steps:
1. Attach the test orifice to the gas outlet port, and turn the ventilator on
and to begin gas delivery.
NOTE: Ground impedance testing (25 A for US and Europe, 40 A for Canada) is
not applicable to the BiPAP Focus ventilator system. The earth
connection from the rear panel of the ventilator is a functional earth for
EMC purposes only. The Protective Earth system is within the power
supply unit as it has no conductive exposed parts. The accessible
functional earth is allowed by IEC 60601-1 because the BiPAP Focus
meets enclosure leakage and separation requirements.
It should be noted that the ground impedance test per IEC 601 is a type
test and not a production line or routine safety test. Respironics
recommends against testing Ground Impedance since it would require
opening the power supply. The functional earth from the ventilator should
be checked by continuity testing (ohmmeter or battery/buzzer) to assure
the ground is intact. The shield of the power supply output cable provides
the EMC ground connection to the protective earth conductor inside of
the power supply unit.
7-24 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
7.8 Alarm/LED Test The Alarm/LED Test confirms that alarms and LEDs are working properly.
3. Remove the AC power from the back of the ventilator for at least one
minute then reconnect the AC power to the back of the ventilator.
8. Remove the DMM leads and return to the System Test screen.
9. Select Alarm LED On, then Silence LED On, and then On Battery LED
On to test LEDs.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-25
Chapter 7
Testing, Calibration, and Performance Verification
7.9 Pressure The pressure accuracy test verifies that pressures are within specified limits.
Accuracy Test 1. Attach an oxygen enrichment adapter to the ventilator outlet port and
occlude.
4. Select CPAP mode and set the pressure to 4 cmH2O. Verify the
pressure reading and record the results.
5. Set the pressure to 20 cmH2O. Verify pressure reading and record the
results.
7. Verify that the BiPAP Focus switches between IPAP and EPAP. Record
the results.
8. Activate the ramp time setting and set ramp time to 5 minutes.
9. Observe the display and the analyzer reading. Verify that the pressure
increases to the target pressure in 5 minutes and record the results on
the data form.
7-26 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 7
Testing, Calibration, and Performance Verification
7.10 Performance A performance verification consists of all the electrical safety, alarm/LED, and
pressure tests.
Verification Data
Complete this form at every performance verification. Make copies of this form
Form for data collection.
Value after
Electrical safety test Passed value Failed value
repair
Forward leakage current: uA uA uA
< 100 microamperes (100-120 VAC)
< 300 microamperes (230-240 VAC)
Reverse leakage current: uA uA uA
< 300 microamperes (100-120 VAC)
< 500 microamperes (230-240 VAC)
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 7-27
Chapter 7
Testing, Calibration, and Performance Verification
Value after
Pressure accuracy test Passed value Failed value
repair
When CPAP = 4 cmH2O, measured pressure = 3.6-4.4 cmH2O
cmH2O cmH2O cmH2O
7-28 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 8. Component Removal/Installation
CAUTION: Electronic components used in this device are subject to damage from
static electricity. Perform repairs only in an antistatic, electrostatic
discharge (ESD)-protected environment.
NOTE: Always retrieve and save the DRPT if possible before replacing parts.
Removing the user interface PCB deletes the serial number, significant
event log entries, and ventilator preference settings. Review the
significant event log before removing the user interface PCB.
NOTE: When reconnecting the tubing to the controller PCB sensors, ensure that
the tubing fits tightly. If the tubing has become loose, clip a short length
(2 mm) from the end of the tubing to ensure a snug fit.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 8-1
Chapter 8
Component Removal/Installation
8.1 Disconnecting Before removing any components from the ventilator, disconnect AC and
battery power sources (Figure 8-1).
Power
1. Turn the ventilator OFF (O) and disconnect the AC power cord.
Battery compartment
Screws (x2)
(5 lbf-in., 57 N cm)
Battery cable
connector
8-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 8
Component Removal/Installation
8.2 Top Cover Follow these steps to remove the top cover (Figure 8-2, Figure 8-3, and Figure
8-4). Reverse to install.
1. From the underside of the ventilator, remove the screws that hold the
side covers (2 screws on each side cover) to the ventilator and remove
the side covers.
2. Remove the 4 screws that hold the display assembly to the ventilator
(2 screws at either side of the display and 2 screws under the display
assembly).
Display
assembly
Screws (x2)
Side panel (8 lbf-in., 90 N cm)
Screws (x2)
(5 lbf-in., 57 N cm)
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 8-3
Chapter 8
Component Removal/Installation
3. Use a 5-mm nut driver remove the 2 jack screws from the 9-pin serial
port on the back panel.
5. Tilt the bottom of the display assembly outward to separate it from the
bottom and top covers.
8-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 8
Component Removal/Installation
6. Remove the four screws holding the top and bottom covers together (8
lbf-in., 90 N cm). Lift front edge of top cover and remove, ensuring
that the back panel cut-outs clear the serial port and switch.
Top cover
Screws (x4)
(8 lbf-in., 90 N cm)
Bottom cover
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 8-5
Chapter 8
Component Removal/Installation
8.3 Display Assembly Follow these steps to remove or replace the display assembly (Figure 8-6,
Figure 8-7, Figure 8-8, Figure 8-9, Figure 8-9). Reverse to install.
2. Disconnect the VGA/UI ribbon cable from the user interface PCB.
8-6 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 8
Component Removal/Installation
Side cover
Screws (x5)
(5 lbf-in., 57 N cm)
4. Remove the screw from the ground cable retainer on the display
assembly.
7. Carefully open the display assembly, taking care not to damage the
ribbon cable and ground cable between the VGA LCD and VGA
interface PCB.
8. Pull the ground cable through the rear display cover and completely
open the display assembly.
9. Remove the 4 screws that secure the VGA LCD and VGA interface
PCB.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 8-7
Chapter 8
Component Removal/Installation
Screws (x4)
(5 lbf-in., 57 N cm)
VGA LCD
VGA interface
PCB
NOTE: Tuck ribbon cables under VGA interface PCB before securing screws. Do
not twist or pinch cables between covers.
8-8 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 8
Component Removal/Installation
8.4 User Interface Follow these steps to remove the user interface PCB (Figure 8-10). Reverse to
install.
PCB
1. Remove top cover (section 8.2).
Primary alarm
connector
Primary alarm
Battery cable
Interprocessor
communications
cable
Interconnect
power cable
Retaining screw
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 8-9
Chapter 8
Component Removal/Installation
8.5 Controller PCB Follow these steps to remove the controller PCB (Figure 8-11, Figure 8-12).
Reverse to install.
Controller
PCB Low-profile
service clip
Interconnect
power cable
8-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 8
Component Removal/Installation
MT2-P2 connects
to air flow sensor
nipple (away from
patient outlet) MT2-P1 connects
to air flow sensor
Switch control
nipple (closest to
cable
patient outlet)
MT3 blower
sensor port
Interconnect
power cable J7
Backup alarm
cable
Interprocessor
Blower cable
communications
J5
cable J1 and J10
6. Disconnect as needed:
• Flow valve cable
• Switch control cable
• Backup alarm cable
• Blower cable
• Interprocessor communications cable
• Interconnect power cable
• Tubing from the pressure transducers (MT2-P1, MT2-P2, MT3,
MT4): label tubing to ensure correct reassembly.
NOTE: When reconnecting the tubing to the controller PCB sensors, ensure that
the tubing fits tightly. If the tubing has become loose, clip a short length
(2 mm) from the end of the tubing to ensure a snug fit.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 8-11
Chapter 8
Component Removal/Installation
8.6 Flow Valve Follow these steps to remove the flow valve (Figure 8-13). Reverse to install.
Valve
strap
Rubber
elbow
Primary alarm
Tubing connects to
controller PCB MT3
Blower
Screws (x3)
(6 lbf-in., 68 N cm)
Grommets (x3)
Tubing connects to
controller PCB MT4
Outlet port
Flow valve
Diffuser
4. Remove the 3 screws, 3 grommets, and valve strap from the valve
assembly.
CAUTION: Do not pull the rubber elbow away from the flow valve. If the rubber elbow
is separated from the flow valve, you must replace the entire assembly.
8-12 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 8
Component Removal/Installation
8.7 Blower Follow these steps to remove the Blower (Figure 8-14, Figure 8-15). Reverse
to install.
Screws (x2)
(5 lbf-in., 57 N cm)
Diffuser
Blower
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 8-13
Chapter 8
Component Removal/Installation
Blower tab
Silicone gasket
NOTE: When replacing the blower and silicone blower gasket, align the gasket
tabs with the cut-outs in the blower muffler housing, then install the
blower tabs into the gasket cut-outs. Align the blower screw holes with
the screw holes in the muffler.
8-14 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 8
Component Removal/Installation
8.8 Backup Alarm Follow these steps to remove the backup alarm (Figure 8-16). Reverse to
install.
4. Move the outlet port and air flow sensor, without disconnecting them,
to access the backup alarm (Figure 8-13).
Screws (x2)
(3 lbf-in., 34 N cm)
Backup
alarm Diffuser
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 8-15
Chapter 8
Component Removal/Installation
8.9 Primary Alarm Follow these steps to remove the primary alarm (Figure 8-17). Reverse to
install.
Screws (x2)
(2 lbf-in., 23 N cm)
Primary
alarm
connector Primary User
alarm interface
PCB
2. Disconnect the primary alarm connector from the user interface PCB.
8-16 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 8
Component Removal/Installation
8.10 Bottom Cover Follow these steps to remove the bottom cover and muffler (Figure 8-18).
Reverse to install.
and Muffler
1. Remove top cover (section 8.2).
4. Remove the flow valve (section 8.6) and the outlet port, air flow
sensor, silicone connector, and diffuser.
6. Remove the four screws from the muffler that anchor it to the bottom
cover.
7. Remove the foam if necessary, and clean any adhesive residue with
isopropyl alcohol. Install new foam if needed, and apply product
labeling if replacing the bottom cover.
Screws (x4)
8 lbf-in., 90 N cm
Low-profile service
clip
Muffler Foam
Bottom
cover
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 8-17
Chapter 8
Component Removal/Installation
8.11 Interconnect This section shows power (Figure 8-19) and communication (Figure 8-20)
interconnections.
Diagrams
8-18 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 8
Component Removal/Installation
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 8-19
Chapter 8
Component Removal/Installation
8-20 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 9. BiPAP Focus Replacement Parts
List
This chapter lists and illustrates the major components of the BiPAP Focus
ventilator, including:
• Exterior
• Front panel/user interface
• Ventilator interior
• Electronic parts
NOTE: When replacing the bottom cover, order the main cover assembly
(P/N 1052538) and a main label (P/N 1031229). Contact product
support (+1-800-345-6443) to request a replacement serial number
label.
9.1 Complete Parts Table 9-1 is a complete list of all BiPAP Focus repair parts.
List Table 9-1: Complete BiPAP Focus Service Parts List
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 9-1
Chapter 9
BiPAP Focus Replacement Parts List
9-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 9
BiPAP Focus Replacement Parts List
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 9-3
Chapter 9
BiPAP Focus Replacement Parts List
9.2 Exterior
Washer (x4) P/N 1031240
Spacer (x4) P/N 1031235
Place standoffs after installing controller PCB
BiPAP Focus top cover assembly NOTE: When replacing the bottom cover, order the
(P/N 1052538) includes the main cover assembly (P/N 1052538) and a
following individual pieces. main label (P/N 1031229). Contact product
• Bottom cover support (+1-800-345-6443) to request a
• Top cover replacement serial number label.
• Side covers (2)
• Rear access door
• Battery cover
Side cover
M3 x 8 screws (x4)
P/N 1030148
Battery cover
Rear access door
P/N 1025761
P/N 1025766
9-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 9
BiPAP Focus Replacement Parts List
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 9-5
Chapter 9
BiPAP Focus Replacement Parts List
9-6 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 9
BiPAP Focus Replacement Parts List
Blower assembly
P/N 1023818
Flow valve seal, foam
P/N 1011998
Grommet (x3)
Flow valve
P/N 1017912
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 9-7
Chapter 9
BiPAP Focus Replacement Parts List
Muffler assembly
P/N 1017913
9-8 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 9
BiPAP Focus Replacement Parts List
9.5 Communications
Cables
BiPAP Focus communications cable kit
P/N 1039800
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 9-9
Chapter 9
BiPAP Focus Replacement Parts List
Power cord
Denmark P/N 1029991
Europe P/N 1029985
India/RSA P/N 1034954
Italy P/N 1034953
Switzerland P/N 1034955
UK/Ireland P/N 1029982
USA P/N 1029989
9-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 9
BiPAP Focus Replacement Parts List
Interprocessor
communications
cable
P/N 1040722
Battery cable
P/N 1025776
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 9-11
Chapter 9
BiPAP Focus Replacement Parts List
9-12 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 10. Specifications
10.1 Ventilator
Specifications
Default Settings
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 10-1
Chapter 10
Specifications
Measured Data
10-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 10
Specifications
NOTE: Est. Vt and Est. MV estimates are at ambient temperature and pressure, dry
(ATPD). Pressure signals are filtered using a 50-Hz low-pass Butterworth filter.
Alarm Settings
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 10-3
Chapter 10
Specifications
Options
Pin Signal
Serial Communications
NOTE:
• When pins 1 and 6 are used, the relay is open during normal operation,
and closed during an alarm condition including loss of power.
• When pins 1 and 9 are used, the relay is closed during normal
operation, and open during an alarm condition including
loss of power.
• Female DB9 connector pin configuration:
WARNING: Use only Respironics-supplied cables with the nurse call station/serial
communications connector.
10-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 10
Specifications
Power
Battery backup 3.7 amp-hour (Ah) nickel metal hydride (NiMH) battery
provides 45 minutes of operation at default settings.
System automatically activates the battery charge cycle
when connected to AC power and the ON/OFF switch is
ON ( | ) (during normal operation or Standby mode).
Recharge time is typically under 5 hours, but may
extend further depending on machine settings if
operating on battery prior to recharge, or if operating in
elevated ambient temperature (above 28°C). Charge is
complete when the green CHARGING LED extinguishes.
Nurse call station The nurse call station port is intended to connect only to
relay a SELV (Safety Extra-Low Voltage) and ungrounded
system with basic insulation to ground, in accordance
with IEC 60601-1. To prevent damage to the nurse call
circuit, the signal input should not exceed the maximum
rating of 24 VAC or 36 VDC at 500 mA, with a minimum
current of 1 mA.
Power cord Use the BiPAP Focus power cords listed in Chapter 9.
Physical
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 10-5
Chapter 10
Specifications
10.2 Patient Circuit The BiPAP Focus System automatically compensates for pressure drops
associated with a Vision single patient use circuit and the mandatory bacteria
Details filter. The pressure drop increases as restrictive elements (for example,
humidifiers) are added to the circuit. Always use a manometer to verify patient
mask pressure.
10-6 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 10
Specifications
10.3 Maximum The Disconnect alarm relies on a fixed relationship between the patient
Pressure Drop for pressure settings and the open circuit flow of the Vision single patient use
circuit. The alarm works correctly unless the circuit is excessively restrictive.
Disconnect Use the pre-operational check to confirm that the Disconnect alarm operates
properly with the circuit.
Volume-based trigger
Flow reversal cycle
Shape signal
Timed trigger
IPAP maximum of 3.0 seconds
Peak flow cycle
Supplemental oxygen
10.4 Environmental
Environmental
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 10-7
Chapter 10
Specifications
10.5 Date of The 12-digit serial number on the base of the unit indicates the date of
manufacture. From left to right, the fifth and sixth digits indicate the year; the
Manufacture seventh and eighth, the month; the ninth and tenth, the day.
Example 1:
This serial number indicates a manufacture date of 13 March 2009:
S/N 8772 09031301
P/N 1028772
Date: YY/MM/DD Unit number
This serial number has no hyphen
Example 2:
This serial number indicates a manufacture date of 29 July 2006:
S/N 8772 060729-01
P/N 1028772
Date: YY/MM/DD Unit number
10-8 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 10
Specifications
If you are subject to the WEEE directive, refer to www.respironics.com for the
passport for recycling this product.
NOTE: The backlight display includes a cold cathode fluorescent light (CCFL),
which contains mercury. Follow all applicable regulations for disposal.
Electromagnetic Emissions
WARNING: Using cables other than those supplied by Respironics may result
in increased emissions and/or decreased immunity of the BiPAP Focus
System.
Electromagnetic Environment
Emissions Test/Compliance
Guidance
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 10-9
Chapter 10
Specifications
Electromagnetic Environment
Emissions Test/Compliance
Guidance
10-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 10
Specifications
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 10-11
Chapter 10
Specifications
The BiPAP Focus System is suitable for use in the specified electromagnetic
environment. The customer and/or the user of the BiPAP Focus System should
ensure that it is used in an electromagnetic environment as described below:
IEC
Immunity 60601-1- Compliance
Electromagnetic Environment Guidance
Test 2 Test Level
Level
d = 1.2 P
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast, and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field
strength in the location in which the BiPAP Focus System is used exceeds the
applicable RF compliance level above, the BiPAP Focus System should be
observed to verify normal operation. If abnormal performance is observed, addi-
tional measures may be necessary, such as reorienting or relocating the BiPAP
Focus System.
b. Over the frequency range 150kHz to 80MHz, field strengths should be less
than 3 V/m.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
10-12 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Chapter 10
Specifications
10.7 Warranty Respironics warrants the BiPAP Focus Ventilator to be free from defects in
material and workmanship for a period of 12 months from the date of
purchase, provided that the unit is operated under conditions of normal use as
described in this manual. At its discretion, Respironics will make
replacements, repairs, or issue credits for equipment or parts that are found to
be defective.
THE WARRANTY SET FORTH ABOVE IS THE SOLE AND EXCLUSIVE
WARRANTY WITH RESPECT TO THE PRODUCT AND RESPIRONICS DOES
NOT MAKE, AND HEREBY SPECIFICALLY DISCLAIMS, ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION,
ANY IMPLIED WARRANTY OF MERCHANTABILITY OR IMPLIED WARRANTY
OF FITNESS FOR A PARTICULAR PURPOSE. THE REPAIR, REPLACEMENT
OR CREDIT REMEDY SET FORTH ABOVE WILL BE THE SOLE REMEDY FOR
BREACH OF WARRANTY. IN NO EVENT SHALL RESPIRONICS BE LIABLE
FOR LOST PROFITS, LOSS OF GOOD WILL, OR INCIDENTAL OR
CONSEQUENTIAL DAMAGES EVEN IF RESPIRONICS HAS BEEN ADVISED
OF THE POSSIBILITY OF THE SAME.
Exclusions
This warranty does not apply to any unit or individual parts which have been
repaired or altered in any way that in Respironics’ judgement, affect its ability
or reliability, or which has been subjected to misuse, negligence, abuse, or
accident.
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. 10-13
Chapter 10
Specifications
10-14 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Index
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. Index-1
Index
Filter, reusable
N
periodic maintenance, 4-1
Negative flow calibration, 7-18
Flow and pressure accuracy, 7-22
Notes, cautions, and warnings, 2-1
Flow valve
description, 3-3 Nurse call/serial port, 3-8
part number, 9-7 pinout, 10-4
removal/installation, 8-12
O
ON/OFF switch bezel
part number, 9-5
Index-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D
Index
REF 1029568 Rev D BiPAP Focus Ventilator Service Manual © Respironics, Inc. Index-3
Index
U
UI/VGA ground cable
part number, 9-11
UI/VGA ribbon cable
part number, 9-11
Unit of pressure, selecting, 5-5
USB cable
installing, 6-3
part number, 9-9
USB drivers, installing to service PC, 6-5
USB port, 3-8
User interface PCB
description, 3-5
part number, 9-10
removal/installation, 8-9
V
Ventilation modes, 3-1
Ventilator storage, 4-9
VGA display
description, 3-7
VGA interface PCB
description, 3-7
part number, 9-6
removal/installation, 8-6
VGA LCD
removal/installation, 8-6
Index-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev D