MDF-DC700VXC-PA - 86°C Ultra-Low Temperature Chest Freezer: 25.3 Cu - Ft. 715 L

The document describes an ultra-low temperature chest freezer with a dual cooling system that maintains temperatures of -86°C. It has two independent refrigeration circuits that work together to provide continuous temperature protection. The freezer allows storage of up to 520 sample boxes and has security and monitoring features like temperature logging and alarms.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
64 views2 pages

MDF-DC700VXC-PA - 86°C Ultra-Low Temperature Chest Freezer: 25.3 Cu - Ft. 715 L

The document describes an ultra-low temperature chest freezer with a dual cooling system that maintains temperatures of -86°C. It has two independent refrigeration circuits that work together to provide continuous temperature protection. The freezer allows storage of up to 520 sample boxes and has security and monitoring features like temperature logging and alarms.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

MDF-DC700VXC-PA

-86°C Ultra-Low Temperature Chest Freezer 25.3 cu.ft. | 715 L

The patented TwinGuard -86°C chest freezer is powered by a unique concept


of two independent auto-cascade cooling circuits that work together, both
independently and concurrently, to provide continuity in stored product
protection over a range of real world conditions.

MDF-DC700VXC-PA

Ultimate Efficient Security Control


Sample Protection Sample Storage and Monitoring

The dual cooling system offers the The use of VIP ® Plus cabinet and lid A color LCD touchscreen allows full
highest level of sample protection insulation, combined with a newly user control while the USB port
through the use of two independent designed refrigeration system layout makes transferring logged data to
refrigeration systems that maintain ensures a capacity of 520, 2" storage a PC convenient. All performance
-86°C temperature. The two circuits boxes, utilizing less floor space than attributes are displayed and an
work together for fast temperature competitive units. internal time/temperature log is
recovery, while the secondary system expressed in graphic form.
keeps the temperature at -70°C if a
service event occurs.

Valuable Sample Storage Predictive Maintenance Ease of Use


With two independent systems, each Performance monitoring ensures The display panel includes a
sustaining ultra-low temperatures, the freezer is operating within comprehensive alarm system with
your samples remain safe. Optional specification. This protects the audible and visual warnings, unit
CO2 backup provides an extra layer compressors and condenser fans. diagnostics and running temperature
of protection. External stress on the compressors log.
or fans, due to ambient temperature
abnormality or excess load, is
displayed for immediate notification.
-86°C ULT Chest Freezer MDF-DC700VXC-PA
Model Number MDF-DC700VXC-PA
External Dimensions (W × D × H) 1) inches | mm 90.6 × 37.6 × 42.1 I 2300 × 955 × 1070
Internal Dimensions (W × D × H) inches | mm 58.3 × 25.2 × 29.8 I 1480 × 640 × 756
Volume cu.ft. | liters 25.3 I 715
Net Weight lbs | kg 807 I 366
Capacity (2" boxes) qty 520
Capacity (3" boxes) qty 360
Inventory Racks (2" boxes) qty 40
Inventory Racks (3" boxes) qty 40

Performance
Warranty 2) 5 years parts and labor
Cooling Performance 3) °C -86
Temperature Setting Range 3) °C -50 to -90 in 1° increments
Temperature Control Range 3) °C -50 to -86 in 1° increments
Cooling Performance 3)
°C -70
Service Event (System A or B)

Control
Controller Microprocessor, touchscreen data entry, password protected
Display LCD color touchscreen
Temperature Sensor Pt-1000

Refrigeration

Refrigeration System qty 2 (auto-cascade)


Model MDF-DC700VXC-PA
Compressors W (2) 1100
Refrigerant Mixed HC/HFC blended
2.8 I 70, foamed-in-place + VIP Plus
Improved Temperature Performance Insulation Thickness, Material inches I mm
vacuum insulated panels
Construction
The TwinGuard -86°C chest freezer design provides
Exterior Material Painted steel
optimal temperature stability and uniformity during
Interior Material Stainless steel
routine lid openings, minimizing the loss of cold air
Lid qty 1 (key lock)
even when the freezer lid is open. Sub Lids qty 3 (insulated, foamed styrene)
Access Ports qty 1: back wall
Access Port Diameter inches | mm 0.6 | 17

Temperature Uniformity Data: Casters qty 6 (3 leveling feet)

Alarms (V = Visual Alarm, B = Buzzer Alarm, R = Remote Alarm)


Power Failure V-B-R
-80°C setpoint at 20°C ambient temperature, empty freezer
60Hz AT 20º SetP -80º Temperature (High and Low) V-B-R
-68
-68
-70 Sensor Failure V-B-R
-72 -72
Door Open V-B-R
Temperature (°C)

-74
Internal Tempareture


-76 -76
⑩ Battery Check V
-78

Internal

-80 -80
⑯ Fan Motor Check V
-82 ⑰
-84
-84 Cooling Circuit Abnormality V-B-R
-86
-88
-88 -90
Remote Alarm Contacts DC 30V 2A (normally open, normally closed, common)
0:00
0:00 0:300:30 1:00
1:00 1:30
1:30 2:00
2:00 2:30
2:30 3:00
3:00
Electrical and Noise Level
Time scale[hour]
Time (hours:minutes)
208V-230V, 1Ø, 60Hz, NEMA 6-15P,
Power Supply
requires NEMA 6-15R receptacle
Typical uniformity profile. For detailed information on actual data points
Voltage Booster Standard, automatic, built-in
contact PHC Corporation of North America.
Noise Level 4) dB(A) 52

Options

Liquid CO2 Back-Up System MDF-UB5-PW

Dimensions Unit : inch | mm


Liquid Nitrogen Back-Up System MDF-UB5-PW-LN2

Temperature Recorder
90.6 | 2300 33.3 | 845 4.1
Circular Type 5) 6", 7 day circular MTR-85H-PW
58.3 | 1480 21.7 | 550 2.1 105
25.2 | 640
54 Chart Paper 52 charts per box RP-G85-PW
Ink Pen pack of 6 PGRPW
Stand Alone Recorder 6", 7 day circular MTRC954
29.8 | 756

42.1 | 1070

Optional Communication System


Wireless, Cloud-Based,
LabAlert® Monitoring System
Automatic Data Management
1)
Exterior dimensions of cabinet excluding handle, rear stand-off brackets and other external projections.
Consult sales rep for doorway entry instructions, less than 37.6".
2)
Current warranty offered at time of printing and may be subject to change; US and Canada only
3)
Air temperature measured at freezer center, ambient temperature +30°C, no load
4)
Nominal value, background noise 20 dB(A)
5)
Requires sensor cover MTR-DU700SF-PW

Specifications are subject to change without notice.


For latest specification information contact PHC Corporation of North America at [email protected].

PHC Corporation of North America


1300 Michael Drive, Suite A, Wood Dale, IL 60191
Toll Free USA (800) 858-8442, Fax (630) 238-0074
www.phchd.com/us/biomedical
Printed in USA | 09 | 19 | 2020 | OW11520.4 | vf

You might also like