Manual of M6000C Central Monitoring System
Manual of M6000C Central Monitoring System
PREFACE
Our liaison-way:
Address: No.2 Innovation First Road, Technical Innovation
Coast, Hi-tech Zone, Zhuhai , P.R.China
Post code: 519085
Fax: +86-756-3399919
Toll-free consultation hot line: +86-400-8818-233
I
Central Monitoring System Software User’s Manual
Statement
This manual contains the proprietary information protected by
the manufacturer Law and the copyright is reserved. Without prior
written agreement from manufacturer no parts of this manual can
be photocopied, copied or translated into other languages.
The contents of the manual are subject to changes without
notice.
II
Central Monitoring System Software User’s Manual
Attention
III
Central Monitoring System Software User’s Manual
CONTENTS
IV
Central Monitoring System Software User’s Manual
V
Central Monitoring System Software User’s Manual
Chapter1 Overview
1-1
Central Monitoring System Software User’s Manual
1-2
Central Monitoring System Software User’s Manual
- SPO2
- PULSE
- NIBP
- TEMP
- IBP
- CO2(EtCO2、FiCO2)
- Anesthetic gas (N2O, O2, AA)
z Maximum of 96-hour holographic ECG waveform
storage and replay;
z Each bedside unit is capable of 10-day trend data
storage and replay;
z Each bedside unit is capable of storage and replay of
1,000 records of alarm messages;
z Alarm limits can be set up for all parameters, and
sound and light alarms are provided;
Sound and light alarms for exceeding the limit of HR
Sound and light alarms for exceeding the limit of RESP
Sound and light alarms for exceeding the limit of SPO2
Sound and light alarms for exceeding the limit of
PULSE
Sound and light alarms for exceeding the limit of BP
Sound and light alarms for exceeding the limit of
TEMP
Sound and light alarms for exceeding the limit of CO2
z It provides indication of probe detachment;
1-3
Central Monitoring System Software User’s Manual
1-4
Central Monitoring System Software User’s Manual
1-5
Central Monitoring System Software User’s Manual
1-6
Central Monitoring System Software User’s Manual
1-7
Central Monitoring System Software User’s Manual
Bed info
Alarm info
1-8
Central Monitoring System Software User’s Manual
1-9
Central Monitoring System Software User’s Manual
1. Function control
a. Alarm sounding state: Indicate the state of alarm sound.
Alarm sound open
Alarm sound closed
b. Exit system
Click the “Exit” button to exit the Central monitoring system
software.
2. System info and hospital name
System info, hospital name and alarm info of non-on-site beds
3. Indicating lamp of bed status (Fig. 1-4)
1-10
Central Monitoring System Software User’s Manual
1-11
Central Monitoring System Software User’s Manual
Chapter2 Installation
2-1
Central Monitoring System Software User’s Manual
Installation of Hardware
According to actual situations, select the placement location of the
Central Unit and lay out the electric wires and network wires.
Place the printer properly.
2-2
Central Monitoring System Software User’s Manual
Fig. 2-1
Fig. 2-2
2-3
Central Monitoring System Software User’s Manual
Fig. 2-3
Fig. 2-4
2-4
Central Monitoring System Software User’s Manual
Fig. 2-5
Fig. 2-6
2-5
Central Monitoring System Software User’s Manual
Fig. 2-7
Fig. 2-8
2-6
Central Monitoring System Software User’s Manual
2-7
Central Monitoring System Software User’s Manual
2-8
Central Monitoring System Software User’s Manual
2-9
Central Monitoring System Software User’s Manual
¾ Notes on Single-beds
¾ Notes on Printing
3.1 Overview
3-1
Central Monitoring System Software User’s Manual
Bed Number
Bed number is the only way for the Central monitoring system
software to differentiate bedside units. In the System, it has 2
meanings, i.e. physical equipment number (which is set by
manufacturer at the time of leaving factory and cannot be
modified by user, ranging from 1 to 66) and logical bed number
(which may be defined by hospital according to its actual
situations of beds in actual use.
Physical bed number is unique to the Central monitoring system
software. Each bedside unit in the same monitoring network has its
unique machine number. The Central monitoring system software
does not have any requirement on logical bed number. It is allowed
that multiple bedside units have the same logical bed number at the
same time. The display of bed number info at the Central
monitoring system software is in the following format: for example,
<27> Room 101 Bed 02, standing for the bedside unit whose
physical bed number is 27, room number is 101 and logical bed
number is 02. In this case, doctors shall determine the unique bed
number on the basis of the physical bed related to the bed during
the observation of the monitoring info displayed on the screen of
the Central monitoring system software.
Physical bed number, room number and logical bed number are
modified on bedside units. Modifications cannot be conducted on
the Central monitoring system software.
3-2
Central Monitoring System Software User’s Manual
Startup
Turn on the power of the monitor (if the power plug of the
monitor is connected to the main unit, please skip this step).
Turn on the power of the main unit.
Switch-off
3-3
Central Monitoring System Software User’s Manual
Startup:
After entering the enabling interface of system, the user may
left-click to set system printer (for details, see the following
system setup); click to set system time; click to enter
the system (or automatically enter the system after waiting for
about 10 seconds without pressing any key); and click to exit
the system. (See Fig. 3-1)
3-4
Central Monitoring System Software User’s Manual
Switch-off
To switch off the central monitor instrument, press the Exit
function button on the main dispaly screen. When a system
indication box (Fig. 3-4) of “Do you really want to exit the
system?” pops up, click “OK” to exit the system and switch it off.
3-5
Central Monitoring System Software User’s Manual
z Function Buttons
Click the downward arrow on each bed, the function menu of
single-bed pops up. (Fig. 3-5)
Patient info Open the dialogue box of patient’s info to input and
modify the patient’s info.
Spread Switch-type key, capable of extending and resuming
beds.
Freeze Switch-type key, capable of freezing and unfreezing
waveforms.
Measure NIBP Switch-type key, capable of commencing and
ending measurement of BP.
Single-bed Open the window of single-bed observation to
show more waveforms and parameters of the bed.
If data is presently saved, browsing of ECG replay, trend data, and
trend chart and alarm incident is possible.
Parameter Open the dialogue box of parameter setting to set
parameters of alarm, ECG, BP, etc.
Bed switching Open the dialogue box of bed switching to conduct
3-6
Central Monitoring System Software User’s Manual
3-7
Central Monitoring System Software User’s Manual
3-8
Central Monitoring System Software User’s Manual
3-9
Central Monitoring System Software User’s Manual
Waveform Display
When the unit is not connected to the bedside units, “Not
networked” will be displayed in the bed info zone; when the
bedside units are connected but patient’s info has not been entered,
the waveform display zone will indicate “Please enter patient’s
info”. It is suggested that patient’s info be entered and monitoring
data be saved so as to facilitate the replay and search in the future.
Under the window for displaying the 4th bed, a maximum of 6
channels of waveforms can be displayed in the waveform display
zone.
Under the window for displaying the 8th bed, a maximum of 3
channels of waveforms can be displayed in the waveform display
zone.
Under the window for displaying the 12th bed, a maximum of 2
channels of waveforms can be displayed in the waveform display
zone.
Under the window for displaying the 16th bed, a maximum of 2
channels of waveforms can be displayed in the waveform display
zone. Users can set up the beds and waveforms to be displayed
according to needs.
1 ) Under general conditions, the monitoring window for
single-bed patient occupies 1/16 of the screen and it can display 2
channels of waveforms (Fig. 3-8).
3-10
Central Monitoring System Software User’s Manual
3-11
Central Monitoring System Software User’s Manual
3-12
Central Monitoring System Software User’s Manual
3-13
Central Monitoring System Software User’s Manual
Bed Extension
3-14
Central Monitoring System Software User’s Manual
beds, and more parameter and waveform info will be displayed (as
shown in Fig. 3-12).
Measure NIBP
3-15
Central Monitoring System Software User’s Manual
Bed view
Press the button of Bed view to enlarge the single bed window
to the display of 1/2 or full window and it is possible to display a
maximum of 12 channels of ECG waveforms and 1 channel of
SpO2, RESP, CO2 or IBP waveform.
Single Bed view is applicable to the patients who need key
monitoring.
Single Bed view may browse all the ECG waveforms or optional
SpO2, RESP, CO2 or IBP waveform, and display all the parameter
info of the bed. After data is saved at the bed, it is possible to
replay the trend data, trend chart, and waveform and alarm incident.
The operations of replay of waveform, trend data and alarm
incident are the same as those of replay of historic data (for details
of replay, see “Data Replay”).
It is possible for the bed on the current screen to begin single
3-16
Central Monitoring System Software User’s Manual
Bed view by clicking the menu of the bed. The on-line beds which
are not on the current screen may conduct single bed observation
in the following 2 ways, i.e. click the status button of the bed and
select to begin single bed observation; or switch bed in the single
bed observation interface (Fig. 3-13).
Bed switching of single Bed view may also be conducted by
clicking the selection box of bed switching on the interface.
Selection of bed
switching
3-17
Central Monitoring System Software User’s Manual
Parameter Setup
3-18
Central Monitoring System Software User’s Manual
3-19
Central Monitoring System Software User’s Manual
2)Alter Parameters
3)Factory setup
Press “Factory setup” button and the system will automatically
restore the setup when the instrument was released from the
factory.
Press “Cancel” button to close the dialogue box and no
operation will be performed.
For example: The user has set up new alarm limits, ECG
parameters and BP parameters according to actual needs. When
the button of “Factory setup” in “Alarm parameter setup” is
pressed, not only will the alarm limits return to the setup before
the monitoring system left the factory, the ECG parameters and
BP parameters will also restore to the setup before it left the
factory.
3-20
Central Monitoring System Software User’s Manual
Fig. 3-15
The alarm levels of parameters include 4 levels, i.e.
Low/middle/high level and switch-off. After setting the alarm
levels, when the parameter sends out alarm, the system will prompt
yellow for low-level alarm and yellow for middle-level alarm and
red for high-level alarm. At the time of saving data, the alarm level
will also be saved. (System will not alarm if the parameters exceed
the limit while the alarm is switched off.)
3-21
Central Monitoring System Software User’s Manual
3-22
Central Monitoring System Software User’s Manual
3-23
Central Monitoring System Software User’s Manual
Bed Switching
Press the Bed switching button, and the dialogue box of “Bed
switching” will pop up (Fig. 3-18).
3-24
Central Monitoring System Software User’s Manual
Click the printing button in the bed menu to begin printing. The
waveforms currently drawn and all the parameters of the bed and
the patient’s info will be printed. The printed reports and detailed
information will be described in the following notes on printing.
After the Data Review key in the main window is pressed, the
data review window will pop up for reviewing the historic data. Or,
click Single-bed Observation for the patient under monitoring to
review the current monitoring data. Data replay includes trend
replay, ECG replay, alarm replay and abnormal ECG replay. The
replay interfaces of dual-screen Central monitoring system
software and single-screen Central monitoring system software are
different, as indicated below.
3-25
Central Monitoring System Software User’s Manual
Setup of review of
Pull-down menu of trend
beginning time
interval
Parameter value of
trend curve
Trend scale
3-26
Central Monitoring System Software User’s Manual
3-27
Central Monitoring System Software User’s Manual
The window of trend chart can, at the same time, display the
trend charts of HR, PR, SPO2, TEMP, RESP, DIA, SYS, MEAN
and CO2 at a given time.
3-28
Central Monitoring System Software User’s Manual
17:43:42 on July 14, 2004 and the Review interval is 240 hours,
then all the trend information from 17:43:42 on July 4, 2004 to
17:43:42 on July 14 will be reviewed.
3)Update Display
Click update, and the trend chart will be displayed according to
the current replay time and interval.
5)Trend Scale
Double-click the trend chart to switch the scale display range of
the trend chart.
Double-click the trend charts of HR and PR to realize the switch
between the ranges of “0~300BPM” and “60~100BPM”.
Double-click the trend charts of SPO2 and CO2 to realize the
switch between the ranges of “60~100%” and “90~100%” of the
SPO2 scale on the left as well as the switch between the ranges of
“0~100” and “20~60” of the CO2 scale on the right.
Double-click the trend charts of TEMP and RESP to realize the
switch between the ranges of “0~50℃” and “30~45℃” of the
3-29
Central Monitoring System Software User’s Manual
TEMP scale on the left as well as the switch between the range of
“0~150BPM” and “20~60BPM” of the RESP scale on the right.
Double-click the BP scale to realize the switch between the
ranges of “-50~300 mmHg” and “60~140mmHg”.
3-30
Central Monitoring System Software User’s Manual
Browse page by
page
Page info
3-31
Central Monitoring System Software User’s Manual
3-32
Central Monitoring System Software User’s Manual
Page info
3-33
Central Monitoring System Software User’s Manual
3-34
Central Monitoring System Software User’s Manual
3-35
Central Monitoring System Software User’s Manual
Display zone of
magnified waveform
3-36
Central Monitoring System Software User’s Manual
The central station can record, search and classify the abnormal
ECG Events from bedside units as shows in the following figure:
3-37
Central Monitoring System Software User’s Manual
3-38
Central Monitoring System Software User’s Manual
3-39
Central Monitoring System Software User’s Manual
3-40
Central Monitoring System Software User’s Manual
1)Query Conditions
Select the search conditions according to needs. Press query
button, and the records conforming to the search conditions will be
displayed in the display zone of search results.
2)Data Type
After “Data type” is selected, ECG data, trend data and alarm
incident will become selectable items. After the single selection
3-41
Central Monitoring System Software User’s Manual
3-42
Central Monitoring System Software User’s Manual
4)Exit Search
“Search ” window.
3-43
Central Monitoring System Software User’s Manual
3-44
Central Monitoring System Software User’s Manual
Color Setup
3-45
Central Monitoring System Software User’s Manual
3-46
Central Monitoring System Software User’s Manual
1)Unit Setup
Click the selection items to set up TEMP unit, BP unit and
scanning speed.
2)Color Printing
Select color printing to print colored waveforms in the color set
on the screen.
3) Setup of Hospital Name
Enter the hospital name in the editing box of “Hospital”. The
hospital name will be displayed in the hospital info zone in the
upper part of the screen. When no hospital name is entered, the
system will prompt “Please enter the hospital info”.
4)Print Setup
Press the print setup button and the dialogue box of “Print
setup” will pop up (Fig. 3-24). Then set up the printer.
3-47
Central Monitoring System Software User’s Manual
5)Alarm Setup
3-48
Central Monitoring System Software User’s Manual
Startup Setup
3-49
Central Monitoring System Software User’s Manual
About
Click “About” and it is possible to browse the related info of
the system (Fig. 3-37).
3-50
Central Monitoring System Software User’s Manual
3-51
Central Monitoring System Software User’s Manual
3-52
Central Monitoring System Software User’s Manual
3-53
Central Monitoring System Software User’s Manual
Caution:
The correctness of the parameters entered for calculation
must be ensured. Before using the calculation results of
medicine, the displayed results must be verified by the final
calculating time. The Company will not be held liable for all
the consequences resulting from error of entry and operation.
Before calculation of medicine, the user must ensure that the
formulas provided in 6) are the calculation formulas he needs.
The Company will not be held liable for all the consequences
resulting from the inappropriate use of calculation formulas.
The results of titrator are related to the calculation results of
medicine concentration and control parameters and
calculation formulas of titrator. In use, the above operations
must be conducted correctly. The Company will not be held
liable for all the consequences resulting from wrongdoing.
3-54
Central Monitoring System Software User’s Manual
3-55
Central Monitoring System Software User’s Manual
3-56
Central Monitoring System Software User’s Manual
4-1
Central Monitoring System Software User’s Manual
4-2
Central Monitoring System Software User’s Manual
4-3
Central Monitoring System Software User’s Manual
4-4
Central Monitoring System Software User’s Manual
Chapter5 Alarm
Alarms in the system are divided into three priorities, that is:
high level, medium level and low level.
High level alarm indicates the patient’s life is in danger. It is the
most serious alarm.
Medium level alarm means serious warning.
Low level alarm is a general warning.
5-1
Central Monitoring System Software User’s Manual
Auditory prompt
Alarm sound is given by the speaker,
The high/medium/low-level alarms are indicated by the system
software in following different audio ways:
Visual prompt
For the parameters that trigger the alarm, their values will
constantly flash to send out alarm in the preset color in the
parameter display zone. For the beds displayed on the current
screen, the alarm info of parameters and lead detachment will flash
to send out alarm above the waveform display zone.
Physiological alarm: The parameter, which triggers the alarm,
splashes on the screen. The physiological alarm area on the screen
displays alarm message, and red indicates high priority alarm,
yellow indicates medium or low priority alarm.
When Technical alarm or General alarm occurs, the Technical
alarm area displays alarm message, red indicates high priority
alarm, yellow indicates medium or low priority alarm, blue
5-2
Central Monitoring System Software User’s Manual
5-3
Central Monitoring System Software User’s Manual
5-4
Central Monitoring System Software User’s Manual
In the menu of parameter for each bed, you can check the alarm
limit, alarm level and alarm status.
When a parameter alarm of one bed is switched off, a symbol
“ ” displays on the alarm info area of the bed.
5-5
Central Monitoring System Software User’s Manual
6-1
Central Monitoring System Software User’s Manual
8 1
6.2.1 There are three types of cables as A、B、C for central System,
A and B cables are supplied with this unit. The C type cable can
be made and set on the basis of the hospital situation.
6-2
Central Monitoring System Software User’s Manual
6-3
Central Monitoring System Software User’s Manual
2 Orange Orange 2
4 Blue Blue 4
55 Blue white
Blue white 5
6 Green Green 6
7 Brown white
Brown White 7
8 Brow
Network Cable is about3 m Brown 8
RJ- 45 8 pin
RJ- 45 8pin
plug
plug
Remarks:
1. Orange white 2. Orange 3. Green white 4. Blue 5. Blue white 6.
Green 7. Brown white 8 Brown.
6-4
Central Monitoring System Software User’s Manual
2 Orange Green 2
4 Blue Blue 4
6 Green Orange 6
7 Brown white
Brown white 7
8 White Network cable is about 1m Brown 8
Remarks:
The left: 1.Orange white 2. Orange 3. Green white 4. Blue 5.
Blue white 6. Green 7. Brown white 8. Brown
The right: 1.Green white 2. Green 3.Orange white 4.Blue 5.
Blue white 6. Orange 7. Brown white 8.Brown
unit B
6-5
Central Monitoring System Software User’s Manual
Central station
A Cable
HUB
B Cable … B Cable
HUB HUB
…
…
C Cable C Cable … C Cable C Cable
A Cable… … A Cable …
A Cable A Cable
6-6
Central Monitoring System Software User’s Manual
Attentions:The user can set IP address for this central station and
bedside units within above range. However, any central station and
any bedside unit should not be set with the same IP address. Or the
whole system will not work properly. The same Unit No is not
allowed to allocate to any bedside unit.
6-7
Product name: Central Monitoring System Software
Product type: M6000C
Address: No.2 Innovation First Road, Technical Innovation Coast,
Hi-tech Zone, Zhuhai , P.R.China
T Post code: 519085
PN: 22-010-0005