CTM MC calibration-PHPL

This calibration certificate summarizes the calibration of a compression testing machine. The machine was calibrated for its 1000 kN range using a 1000 kN force proving instrument. Readings were taken at several load points and positions and were all found to be within the specified tolerance of 1%. The machine was determined to be operating properly and has a measurement uncertainty of 0.83% at a 95% confidence level. It is due for recalibration on August 23, 2021.

Uploaded by

Rajeev Kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
907 views

CTM MC calibration-PHPL

This calibration certificate summarizes the calibration of a compression testing machine. The machine was calibrated for its 1000 kN range using a 1000 kN force proving instrument. Readings were taken at several load points and positions and were all found to be within the specified tolerance of 1%. The machine was determined to be operating properly and has a measurement uncertainty of 0.83% at a 95% confidence level. It is due for recalibration on August 23, 2021.

Uploaded by

Rajeev Kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

pvffi[r&

BALITIKURI, CHAKPARA, HOWRAH 711 113.


Servicing, Repairing & Calibration of Testing Machines, Lab Equipments,
Temperature lnstruments & all Type of Measuring lnstruments.
Customer Service Cell No. : 0332653 0204, 7059626385
CIN No. U73'1 00WB2O1 4Pr C2OO542

TCLPU5.'1oiTEC/20.
CALTBRATTON CERTIFTCATE uLR No.: cc2o752oooooo65o8F

Customer's Details: M/s. Tuaman Engineering Limited.


BS-VI Project at Paradeep Refinery.

Gertificate No: T1 02t 137 I 08t 20 -21 tF O -01 Details of Machine

Certificate Date: 25.O8.2020 Name: Gompression Testing Machine


Date of Calibration: 24.O8.2020 Make: Masta

Suggested due on: 23.O8.2021 Range: 1000 kN

Environmental Temperature: 30r2"c Sr. No: HOT-204


Condition:
Relative Humidity: (50r10)% Model No: -

Calibration Method: lS 1828 (Part 1)2015. ld. No. TEWCTM - 01

Mode of calibration: ln Compression Location: LAB

Details of Standard Equipments used for Calibration


Name of lnstrument I Certificate No.
I Valid up to I Calibrate by
1000 kN FRt/08/18/10797 I 24.10.2020 FRI
Contd. - page 2

.- souu( 'qq
RATION ENGINEER {"9 TECHNICAL

Mobile : 983112'1015
e.mail : [email protected] & [email protected], Website : www.technocalibrationlaboratory.co.in
WAN
BALITIKURI, CHAKPARA, HOWRAH 711 113,
T@RYPVT,EEB.
Servicing, Repairing & Calibration of Tes
Temperature lnstruments & all Type of Measuring lnstruments.
Customer Service Cell No. : 0332653 0204, 7059626385
CIN No. U731 00WB2O1 4Pr CZOOS4Z

TCLPUS.'10/TEC/20. CALIBRATION CERTIFICATE ULR No.: CC207520000006508F

Certificate No: T 1 02t 1 37 I 08 I 20 -21 I F O -Ol Certificate Date: 25.08.2020


Calibration Date: 24.08.202A
Calibration Due On: no.
23.08.2021 Page 2 Total Pages 2

Result

Standard Temp.
Actual Reading obtained P
qr= (, co-o
Load in Relative >O L
reading Correction Average Accuracy EGIO
(E= F E=L
KN (!= O
(Drv) (Drv) Position Position Position o olu
Ettot o/o
t4 & t3u:
0o 120 0
240" o
e.
By Using 1000 kN Force proving lnstrument

Scale:0-1000kN L.C-5 kN
_.------r----_-__....1
200 I 142.1 142.4 142.0 142.5 142.0 142.17 0.14 0.35 I 0.025

400 282.7 283.2 284.0 284.5 284.0 284.17 -0.33 0.18 0.013

600 425.5 426.3 428.5 427 .5 428.5 428.17 -0.43 0.008

800 568.3 569.4 572.0 573.0 572.0 572 33 -0.52 0.17 0.006

1,000 712.1 713.4 715.0 716.0 715.0 715.33 -0.26 0.14 0.00 5

Class of the machine: Class 1

uncertainty of Measurements : t0.83% (coverage factor k =2 ar gio/o confidence lever)


Remarks: The above compression Testing Machine has been calibrated for its range. The reading
observed is
tabulated above in respect of its accuracy. The reading with in the toleran ce (!1%). T+|e
liachine is working ok.

#
*-sotMKtdl
Calib iON ENGTNEEI
TECHNICI.,"l