Pressure Monitoring System in Gastro-Intestinal Tract
Pressure Monitoring System in Gastro-Intestinal Tract
Abstract - Diseases in the gastro-intestinal tract are on an modules in order to realize the functional endoscopes of
increasing trend. In order to diagnose a patient, various temperature, pressure, pH, and so on.
signals of the digestive organ, such as temperature, pH, and
pressure, can offer the helpful information. Among the above For in-vivo telemetric studies of the gastro-intestinal
mentioned signals, we choose the pressure variation as a tract, Johannessen [3] suggests an analytical microsystem,
monitoring signal. The variation of a pressure signal of the which incorporates a four sensing channel. The measuring
gastro-intestinal tract can offer the information of a digestive signals are temperature, pH, conductivity and oxygen.
trouble or some clues of the diseases. In this paper, a pressure However, they did not deal with a pressure signal in the
monitoring system for the digestive organs of a living pig is body.
presented. This is why a pig’s gastro-intestinal tract is very
similar as human’s. This system concept is to transmit the Especially, because the pressures in the human body or
measured biomedical signals from a transmitter in a living a living body is necessary as a part of clinical examinations
pig to a wireless receiver that is positioned out of body. The and for physiological studies, a new inspection and new
integrated solution includes the following parts: (1) the devices have been developed for a monitoring of a gastro-
swallow type pressure capsule, (2) the receiving set consisting intestinal tract. Firstly, the intra gastric and intra intestinal
of a receiver, decoder box, and PC. The merit of the pressures are measured in the gastro-intestinal tract [4].
proposed system is that the monitoring system can supply the These pressure signals are considered as the sum of the
precise and repeatable pressure in the gastro-intestinal tract. pressures developed by the peristaltic motions of the
In addition, the design of low power consumption enables it stomach or the intestine and the pressure in the peritoneal
to keep sending reliable signals while the pressure capsule is cavity. Recently, there are many endeavor and attempts for
working in the digestive organ. The subject of the study for an increasing accuracy of the diagnosis in a bio-medical
the pressure monitoring system is in-vivo experiments for a field for this environment. For the long-term monitoring of
living pig. We achieved the pressure tracings in digestive the pressure in the digestive system, it is preferable to use a
organs and verified the validity of system after several in- pressure transducer that is separated from external
vivo tests using the pressure monitoring system. As a result, instruments. Therefore, the pressure sensor module can be
we found each organ has its own characterized pressure
either surgically implanted (as in the case of the
fluctuation.
intracranial pressure measurements) or swallowed (for
Index Terms - Pressure monitoring system; pressure pressure measurements in the gastrointestinal tract) [4].
capsule; gastro-intestinal tract; in-vivo test The swallowable capsule including gastro-intestinal
pressure telemetry has been studied by many researchers
I. INTRODUCTION [5-9]. In addition, Mackey has suggested the pressure
Recently, the endoscopy has been developed since sensor, which was enclosed in a rubber balloon and
diseases in the gastro-intestinal tract have gradually measured the pressure signal from a human stomach, a
increased. Among the endoscopic technologies, a capsule small intestine, and a colon [10]. However, the above
type endoscope is highlighted for the convenience of a mentioned results are insufficient to differentiate each
patient. While the capsule endoscope moves along the digestive organ.
gastro-intestinal tract, it transmits the various visual images In this paper, we report the concept of the telemetry
of in the digestive organs [1] [2]. The device, known as the system about the proposed pressure monitoring system as
M2A Imaging System (Given Imaging), can supply reliable shown in Figure 1. This concept is to transmit the measured
images of the gastro-intestinal tract [1]. biomedical signals from the transmitter in the body to the
Except the visual images of the digestive organs, we wireless receiver out of the body. The RF carrier
can know that the various signals in the organs, such as frequency is 1.2 GHz (Frequency Shift Keying) for the
temperature, pH, and pressure, can offer the useful transmission of the signals from inside the living body. The
information for the diagnosis of a patient. In addition, we pressure monitoring system comprises: (1) the swallowable
have an interest in the miniaturized wireless telemetry type pressure capsule and (2) the receiving set including
1334
The interface program displays the pressure measurement
results.
III. DEVELOPMENT OF PRESSURE CAPSULE This capsule consists of a pressure sensor, a digital
controller by Verilog HDL, a 1.2 GHz transmitter, and a
A. Pressure Capsule battery as shown in Figure 7. The capsule also includes a
Figure 6 shows an electronic circuit block diagram in 10mm diameter loop antenna for transmitting the measured
the pressure capsule. For a pressure sensor, a pressure sensing data. The digital controller of the pressure
commercialized pressure sensor (MS5535A, Intersema of capsule is implemented as a single module that consists of
Switzerland) was used. The pressure sensor module [11], the controller for the pressure sensor, the encoder, and a
MS5535A, is a SMD-hybrid device including a piezo- double-sided single PCB (Printed Circuit Board). The
resistive pressure sensor and an ADC-interface IC. The specifications of this controller are summarized in Table I.
resolution of this sensor is typically 1.2 mbar.
A four-wire interface is used for the communication TABLE I. SPECIFICATION OF DIGITAL CONTROLLER.
between the sensor and the digital controller unit. The
SCLK (Serial Clock =125 kHz) signal initiates and
Feature Value
synchronizes the data transference. In addition, the data
signal in Figure 6 is the ADC conversion result for the Device XCR3064-10CP56C
measured data by the pressure sensor. The selection of the
output data is dependent on the control signal. The signal Voltage 3.3 V
MCLK (Master Clock) indicates the AD-converter clock
Current 8 ~ 10mA
(32.768 kHz frequency), that is also supplied from the
digital controller unit. The digital controller is designed Used I/O 10
using Verilog HDL in ISE-WebPACK (Xilinx). The
selected chip is XCR3064-10CP56C (Xilinx) for the digital Frequency 4MHz
control unit. This unit generates MCLK signal, SCLK
signal, and the control signal in order to operate the
pressure sensor and transmit the measured pressure data to The dimension of this controller with an oscillator is
the encoder. The output of the digital controller is encoded 4.5 mm in thickness and 10 mm in diameter. The pressure
to suit the wireless transmission and the reconstruction of sensor and the electronic circuits are connected to a 3 V Li-
the pressure measurement data. battery. The Li-battery uses a lithium metal alloy in its
negative electrode (anode) and manganese a dioxide in its
1335
positive electrode (cathode) [12]. The consumable power has an omni directional antenna characteristic in order to
from the battery is an average level of 48 mW. In order to transmit the measured signals continuously. The wireless
investigate the battery lifetime, the supplied voltage of the transmitter module totally consumes the current of about
battery is measured and the result is shown in Figure 8. 13 ~ 14mA.
The energy output rating for the Li-battery used is typically
indicating 160 mA*hr. The pressure capsule should last IV. EXPERIMENTAL VERIFICATION
more than 11 hours to guarantee a sufficient measuring
through a whole gastro-intestinal tract because the A. Experimental Setup
operating voltage of the capsule is 2.2 V at the minimum.
The in-vivo test of the monitoring system was carried
out at Yonsei medical center in Korea under anesthesia
condition. The objective of the in-vivo test was to check
the operation of the proposed pressure monitoring system.
To verify the pressure capsule, the various experiments for
the reliable data acquisition, transmission of an acquired
data through a living medium were performed. By these
experiments, the operation of the receiving set was tested
for receiving, sorting, storing and displaying the signals. In
addition, our research is intended to verify the miniaturized
wireless telemetry modules for other function such as the
recognition of the location based on the pressure difference
of each digestive organ.
1336
pressure data at the esophagus. Next, the pressure capsule TABLE II. MEASURING PRESSURE DATA.
reaches at the stomach and transmits the pressure data in
the stomach. Finally, the capsule is deeply put in the colon
through the anus (Figure 11-(d)) and the pressure data of Mean Max Min Max-Min
the colon can be received. (kPa) (kPa) (kPa) (kPa)
1337
in the organs such as digestive troubles and breathing ACKNOWLEDGMENT
malfunctions.
Authors wish to thank Prof. Hyunchul Choi in
Kyungpook National University for the development of
V. CONCLUSIONS telemetry module and many valuable discussions. This
work was supported by the 21st Century's Frontier R&D
Projects, under the contract number MS-02-324-01,
The pressure monitoring system with the pressure sponsored by the Ministry of Science and Technology,
capsule suitable for a digestive tract is presented. Through Korea.
the pressure measurements of the digestive tract using the
pressure capsule, the reliability and validity of the pressure
REFERENCES
monitoring system are verified. Consequently, the
proposed pressure monitoring system can supply precise
and repeatable pressure signals in the gastro-intestinal tract. [1] G.A. Douglas and J. G. Christopher, “State of the Art - Wireless
Capsule Endoscopy,” Hospital Physian, May, 2003, pp. 14-22.
Based on the design of low power consumption of the
pressure capsule, this capsule can keep sending reliable [2] http://www.rfnorika.com/e_system/e_system_001.html
pressure data until it comes out through anus. In addition, [3] E.A. Johannessen, L. Wang, L. Cui, T.B. Tang, M. Ahmadian, A.
Astaras, S.W. Reld, P.S. Yam, A.F. Murray, B.W. Flynn, S.P.
we could find each organ has its own characterized Beaumount, D.R. Cumming, and J.M. Cooper, “Implementation of
pressure fluctuation. Therefore, the location of the pressure Multichannel Sensors for Remote Biomedical Measurements in a
capsule in the digestive tract can be obtained based on the Microsystems Format,” IEEE Tran. Biomedical Engineering, Vol.
continuous monitoring of the pressure signal while the 51, No. 3, 2004, pp. 525-535.
capsule is moving from mouth to anus. [4] T. Togawa, T. Tamura, and P.A. Oberg , “Biomedical Transducers
and Instruments,” CRC press, 1997, pp. 13-71
The in-vivo experiment was carried out under an [5] J.T. Farrar, V.K. Zworykin, and J. Baum, “Pressure-sensitive
anesthesia condition and the breathing machine was used telemetering capsule for study of gastrointestinal motility,” Science,
for the forced breathing. Therefore, the measured breathing Vol. 126, pp. 975~976, 1957.
frequency is determined by that of the breathing equipment. [6] R.S. Mackay, “Radio telemetering from within the human body,”
For the insertion of the pressure capsule, the endoscope is IRE Trans. Med. Electron., Vol. 6, pp. 100~105, 1959.
used and thus its air insufflation pressurizes each organ. [7] J.T. Farrar, C. Berkley, and V.K. Zworykin, “Telemetering of
inraenteric pressure in man by an externally energized wireless
Consequently, it is estimated that the measuring the capsule,” Science, Vol. 131, pp. 1814, 1960.
pressure signals are slightly changed by the breathing
[8] C.N. Smyth, and H.S. Wolff, “Application of endoradiosound or
machine and the air insufflation of the endoscope. "wireless pill" to recording of uterine contractions and foetal heart
sounds,” lancet, Vol. 2, pp. 412, 1960.
In order to reduce the above-mentioned problems, in
[9] M.B. Rappaport, E.H. Bloch, and J.W. Irwin, “A manometer for
the future, we aim at in-vivo test in a real active measuring dynamic pressures in the microvascular system,” J. Appl.
environment of a living pig. In addition, through many Physiol., Vol. 14, pp. 651, 1959.
clinic experiments by using the pressure capsule, we are [10] R.S. Mackay, “Bio-medical telemetry,” IEEE press, 2nd Edition,
going to get the useful information of the diseases in the 1993.
organs. [11] http://www.intersema.ch/site/technical/ms5535.php
[12] http://www.sanyo.com/batteries/lithium_ind.cfm
1338