Taurus Healthcare Pvt. Ltd.

COPYRIGHT NOTICE
Copyright© 2015 Taurus Healthcare Pvt. Ltd., New Delhi

This work is protected under section of copyright act of India and is the sole property of
Taurus Healthcare Pvt. Ltd. No part of this document may be copied or reproduced, or
stored in any electronic information retrieval system, except as specially permitted under
copyright act of India, without the prior written consent of Taurus Healthcare Pvt. Ltd.
For more information, details are mentioned below:

Head Office:

TAURUS HEALTHCARE PVT. LTD.

A5-A6,Pragati Chamber, Ranjit Nagar


Commercial Complex, New Delhi 110008,India
Ph. : 011-45087820, Telefax : 011-25702368
Mobile : +91-9899966945,9711203605,9711203625
Email : [email protected],[email protected]
Web : www.taurushealthcare.co.in, www.medicalcompressor.net

TRADEMARK NOTICE
Ventiair Air Compressor is a deemed registered trademark of Taurus Healthcare Pvt.
Ltd. All other brand names and product names mentioned in this manual are
trademarks, registered trademarks, or names of their respective holders.

WARRANTY
The Ventiair Air Compressor is warranted to be free from defects in material and
workmanship to meet the published specification for one (1) year or 5,000 hours
whichever occurs first or as per terms agreed with the customer with the following
exceptions:-

1. Components for monitoring of physical variable such as temperature, pressure or


flow warranted for ninety (90) days from date of receipt.
2. Elastomeric components and other parts or components subject to deterioration
over which Taurus Healthcare Pvt. Ltd., has no control, are warranted for sixty
(60) days from the date of receipt.

The liability of Taurus Healthcare Pvt. Ltd., under this warranty is limited for replacing,
repairing or issuing credit, at the discretion of Taurus Healthcare Pvt. Ltd., for parts
those become defective or fail to meet published specification during the warranty
period.
2
Taurus Healthcare Pvt. Ltd.
Any authorization of Taurus Healthcare Pvt. Ltd., for repair or alteration by the buyer
must be in writing to prevent the warranty. In no event shall Taurus Healthcare Pvt. Ltd.,
be liable to the buyer for loss of profit, loss of use, consequential damage or damages of
any kind based upon a claim for breach of warranty, other than the purchase price of
any defective product covered hereunder.

Taurus Healthcare Pvt. Ltd., warranties as herein and above set forth shall not be
enlarged, diminished or affected by, and no obligation or liability shall arise or grow out
of the rendering of technical advice or service by Taurus Healthcare Pvt. Ltd. or its
agents in connection with the buyer‟s order furnished hereunder.

LIMITATIONS OF LIABILITIES
This warranty does not cover normal maintenance such as cleaning, adjustment or
lubrication of equipment parts. This warranty shall not apply if the equipment is used
with accessories or parts not manufactured by Taurus Healthcare Pvt. Ltd. or authorized
used without taking authorization in writing by Taurus Healthcare Pvt. Ltd. or if the
equipment is not maintained in accordance with a prescribed schedule of maintenance.

EMC NOTICE
The equipment generates, uses, and can radiate frequency energy. If not installed and
used in accordance with the instructions in this manual, electromagnetic interference
may result. The equipment has been tested and found to comply with the limits set
forth in IEC 601-1-2 and EN 60601-1 for medical products. These limits provide
reasonable protection against electromagnetic interference when operated in the
intended use environments described in this manual.

MRI NOTICE
This equipment contains electromagnetic components whose operation can be affected
by intense electromagnetic fields. Do not operate the air compressor unattended in an
MRI environment or in the vicinity of high frequency surgical diathermy equipment,
defibrillators, or short wave therapy equipment. Electromagnetic interference could
disrupt the operation of air compressor.

3
Taurus Healthcare Pvt. Ltd.

CONTENTS
1. GENERAL INFORMATION 05

Purpose 05

Information search made easy 05

Operator‟s Safety information 05

Warning 06

Caution 06

Equipment Symbols 08

2. INTRODUCTION 09

3. PNEUMATIC DIAGRAM 10

4. ELECTRIC DIAGRAM 11

5. EQUIPMENT DESCRIPTION 12

VA-11/VA-12 12

VA-09 13

6. INSTALLATION OF COMPRESSOR 15

7. SETUP AND OPERATING PROCEDURE 18

8. MAINTENANCE 20

Maintenance Schedule 20

9. TROUBLESHOOTING 22

 APPENDIX A - GLOSSARY
 APPENDIX B - PRODUCT SPECIFICATION
 APPENDIX C - SPARE PARTS LIST
 APPENDIX D -LIST OF ACCESSORIES FOR ALL MODELS

4
Taurus Healthcare Pvt. Ltd.

GENERAL INFORMATION

ABOUT THIS MANUAL

Purpose

• This section can be referred for setup, operation and maintenance of the Ventiair
compressor. After reading this section, you will understand :
• Intended use of the air compressor
• Controls and Indicators
• Operational use of the air compressor
• Setup of the compressor and how to attach a ventilator

Information search made easy

The manual has been specially designed to make the information search easy.

The manual features the following:

√ Table of contents: It shows the topics and the order in which they are covered under
this manual
√ Index: The index contains all the topics. If you cannot find any topic in the table of
contents, just scroll through the index. All the entries have been cross-referenced
for easy reference.
Operator’s safety information

Terms

√ WARNING: Identify conditions or practices that could result in serious adverse


reactions or potential safety hazards.
√ CAUTION: Identify conditions or practices that could result in damage to the
compressor or other equipment.
√ NOTE: Identify supplemental information to help you better understand how the
compressor works.

Symbols

This symbol indicates a WARNING.


WARNING!

This symbol indicates a CAUTION.


CAUTION!

This symbol indicates a NOTE

5
Taurus Healthcare Pvt. Ltd.

WARNING

Listed below are the following warnings those apply all the time during the lifetime of the
compressor whether operating or servicing:

• Possible explosion hazard if used in presence of flammable anesthetics.


• Electrical shock hazard, do not remove panels. Refer qualified service personnel.
• All electromechanical systems are subject to malfunction from both internal and
external causes. Although the various protections are used and the alarms have
been incorporated to detect and notify the operator of these faults. In case of fault,
shut down the system and refer to the authorized technician.
• Operating an improperly functioning compressor could be hazardous to the operator
as well as to the patient. In case of malfunction remove the compressor from service
and contact Taurus Healthcare Pvt. Ltd., immediately.
• Do not operate the air compressor unless qualified/trained personnel are in
attendance to promptly respond to alarms, inoperative conditions or sudden
malfunction.
• Never ignore any audible alarm, respond immediately.
• The air compressor system is grounded. Do not plug it into ungrounded source.
• Disconnect the power before removing panels for servicing.
• All the design specifications should be met before the operation. Contact Taurus
Healthcare Pvt. Ltd., in case the specifications are not met.
• The pump must be disconnected from power supply before service can begin.
Failure to do so may result in severe personal injury.
• Never use parts, accessories, or options those have not been authorized for use
with the air compressor.

CAUTION

Listed below are the following cautions those apply anytime you operate or service the
air compressor:

• A protective ground connection by way of grounding conductor in the power cord is


essential for safe operation .The absence of it may result in electric shock. To avoid
the shock, plug the power cord into a properly wired receptacle and make sure the
power cord supplied with compressor is always used and is in fine condition.
• Use circuit breakers to avoid fire hazard. Make sure the circuit breaker should
comply with current, voltage and other ratings as specified in the manual.
• Perform preventive maintenance at the minimum recommended intervals.
• The air compressor is of the oil-less type and do not lubricate any of the parts with
oil, grease or petroleum products.
• The filters should be maintained properly as the lack of it may result in pressure
drop, over heating or other severe damages to the compressor.
• Preventive maintenance may be required more frequently if operated in an
6
Taurus Healthcare Pvt. Ltd.
environment with a relative humidity in excess of 70%.
• Do not use excessive force in attaching equipment to the compressed fittings.
• Allow a minimum of 12‟‟ clearance from the front & 3‟‟ clearance from the bottom of
compressor.
• Before removing the doors of the air compressor, remove all equipments mounted to
the top of compressor cabinet.

The following cautions apply when cleaning the air compressor

• Abrasive agents such as scouring pads should not be used when cleaning the
exterior surface.
• Caustic cleaning agents or solvents should not be used to clean any internal or
external part of the air compressor.
• Do not sterilize the Ventiair compressor in a sterilization chamber.

7
Taurus Healthcare Pvt. Ltd.

Equipment Symbol

The following symbols may appear on the air compressor:

S.N Symbol Compliance Meaning

1. IEC 878 Symbol # 03-02 This Symbol indicates CAUTION &


WARNING
.

2. IEC 878 Symbol # 01-01 This Symbol indicates the condition


and connects the equipment from the
MAINS SUPPLY

3. IEC 878 Symbol # 01-24 This symbol indicates protective


EARTH(ground)

This symbol indicates the


EQUIPOTENTIAL connection used to
4. IEC 878 Symbol # 01-24 connect various parts of the
equipment or of a system to the same
Potential, not necessarily being the
earth (ground) potential.

5 This symbol indicates the equipment


IEC 878 Symbol # 01-02 is suitable for alternating current.

8
Taurus Healthcare Pvt. Ltd.

INTRODUCTION
 Intended Use of Air Compressor

The Ventiair compressor is designed to supply clean, dry, filtered, compressed air
for ventilator or other respiratory care devices. The ambient air, ranging from a
low relative humidity to 85%, is compressed and delivered at a working pressure
of 50 PSI (3.5 bar). The air compressor is capable of delivering a continuous flow
of 50 LPM.
 What to do if the compressor fails

If a condition is detected that could possibly prevent the compressor from


continuing to operate safely, it will sound an alarm. Contact a Taurus Healthcare
Pvt. Ltd., certified technician.

WARNING
WARNING! WARNING!

Do not operate the medical air compressor unless


qualified personnel are in attendance to promptly
respond to alarms, inoperative conditions, or sudden
malfunctions. Patients on life support equipment
should be visually monitored all the times.

9
Taurus Healthcare Pvt. Ltd.

PNEUMATIC DIAGRAM

Air from Ambient through Front


cabinet filter

Cooling Coil Auto Drain


Compressor Compressor motor
Water Trap 2
Suction Filter Exhaust

Water Output
Solenoid valve
Water
Water Tray Air Storage
Accumulator Tank
Exhaust

Accumulator
Water Auto Drain
Pressure Water Trap 1
Relief
Output

Pressure Display

To Ventilator/ Bubble CPAP Pressure


Regulator
Compressor Outlet

Air Compressor Pneumatic Diagram

Note: - 1) Dotted Lines used for Input & Output.

2) Solid lines Used for Exhaust

10
Taurus Healthcare Pvt. Ltd.

ELECTRICAL DIAGRAM

Main Supply 220VAC/50 Hz

ON/OFF Solenoid Compressor


Upper Fan Motor
Switch Valve

Display

Lower Fan
Hour Meter

11
Taurus Healthcare Pvt. Ltd.
EQUIPMENT DESCRIPTION

VA-11/VA-12

(A)

(D)

(B) (E)

(C)

(H)

(F)

(I)

(G)

12
Taurus Healthcare Pvt. Ltd.

VA-09

(B)

(A)

(J)

(D)

(E) (K)

(C)

(L)

(I)

(F)

(G)

13
Taurus Healthcare Pvt. Ltd.
(A) Power Switch
Turn the Air Compressor to ON position. A red light in the switch will indicate that
the compressor is ON. Air Compressor equipped with standby /back up mode
(central distribution line) will supply compressed air only if the primary air source
fails. If the compressor is not connected to a primary air source, the power switch
will turn the compressor on.

(B) Pressure Gauge


It indicates outlet pressure.

(C) Air Intake Filter


This acts as the first source of filtration and prevents cabinet contamination.

(D) Out -Compressor Air Outlet


This provides the oil free compressed air for ventilator.

(E) External Water Trap (optional)


It provides the final stage filtration.

(F) Compressor Motor


It intakes the natural air from the environment and generates high pressure
compressed air. The output of the compressor motor is hot high pressure air.

(G) Cooling Coil


The cooling coil helps in reducing the temperature of the hot compressed air. It
receives the hot high pressure air from compressor motor. When this air passes
through it, hot high pressure air gets cooled.

(H) Cooling Fan


The cooling fan is mounted over the compressor motor pump to cool down the
temperature of the unit.

(I) Safety Valve


It opens when compressor motor provides air pressure more than 150 PSI.

(J) IEC Inlet for Power Cord


It is used for connecting power cord.

(K) Heated Humidifier Clip


It is used for inserting Heated Humidifier.

(L) Basket
For storage purpose only.

14
Taurus Healthcare Pvt. Ltd.

Installation of Compressor

 Choose the given installation procedure according to air compressor


model.

Method-I

General Instructions
 Remove all internal as well as outer packing before switching on the unit as
shown in further pictures.
 This is done as the compressor pump inside the compressor is mounted on the
teflon orange o-rings for lower vibrations as well as noise level.

Method-II

VA-11/VA-12

1) Slide down the unit horizontally on a smooth base.


2) Two tie-wraps and four LN screws are required to be removed as marked in picture
(b).
3) Cut the tie-wraps using cutter and remove the LN screw using 4 mm LN key as
shown in pictures.

(a) (b)

15
Taurus Healthcare Pvt. Ltd.

(c) (d)

4) Now, unit is ready for operation and it can be operated by connecting to power
supply and controlled by ON/FF switch.

16
Taurus Healthcare Pvt. Ltd.
VA-09

1) Slide down the unit horizontally on soft base so that cabinet is not scratched as
shown in picture.
2) Open 4 LN screws (packing screws) as screwed to base trolley by 4mm LN key
as shown in picture (b).

(a) (b)

3) Pick up the compressor carefully.


4) Now, it can be operated by using ON/OFF switch.
5) To connect ventilator to compressor, push the ventilator nozzle into the DISS
outlet i.e. “OUT” of compressor.

17
Taurus Healthcare Pvt. Ltd.

Setup and Operating Procedure


Operational Use

The Ventiair compressor is intended to be used as both a compressed air source and, if
configured, as an emergency backup air source. The air compressor will sense pressure
drop in a primary air source, and at a preset level, will override the primary air source to
maintain air pressure.

Set up and Operation Instruction

1. Position air compressor so that all four sides are unobstructed.

WARNING
WARNING! WARNING!

Particular attention should be exercised to keep


the front of the Compressor free of any obstruction
allow a minimum of 12” clearance at front of
compressor and 3‟‟ from the bottom surface.

2. Connect the power cord to a grounded AC electrical outlet.

3. If desired, connect air inlet to 55 +/- 5 PSIG (3.6+/-.34 bar) air supply.
4. Attach equipment to air outlet.

5. Turn ON/OFF switch to the “ON” position.

6. While the compressor is operating, observe the pressure gauge to verify operating
pressure (50-65 PSI).
7. For Ventilator Air Compressor, if the pressure of 50 PSI is not obtained at 50 LPM
flow .Contact Taurus Healthcare Pvt. Ltd.
8. For Bubble CPAP, if the pressure of 40 PSI is not obtained at 20 LPM flow. Contact
Taurus Healthcare Pvt. Ltd.

18
Taurus Healthcare Pvt. Ltd.

COMPRESSED AIR CONNECTION

The medical compressor is equipped with quick couplings „„AIR INLET‟‟


and “OUT‟‟ -compressor air outlet on the rear of the cabinet. Connect the
pressure hose from the given equipment/respiratory device to the quick
coupling “OUT‟‟ –outlet compressed air. Connect the compressed air line
from the central distribution line to the quick coupling air inlet compressed
air. Air from the distribution system is automatically connected through the
compressor to the “OUT‟‟ outlet port. In this configuration, the compressor
serves as a backup source of compressed air. If the air pressure from the
central distribution system is reduced, the compressor automatically
switches on and there is no interruption in the supply of continuously
pressurized air at the outlet of the compressor.

Please note that the air supply from the central distribution line entering
the compressor must be medical grade air (particle size, humidity). The
compressor does not modify the air form the central distribution line.

The hose connecting the compressor to the respiration equipment must


be as short as possible with no kinks (Long hose may cause water to
condense inside the hose).

The compressor comes with a plug containing an appropriate protective


contact (ground). The voltage and frequency of the mains must comply
with the specifications on the data label.

The power cord must not be stressed or have any tension exerted upon it,
and must always be free.

19
0434

Taurus Healthcare Pvt. Ltd. 0434

0434
Maintenance

MAINTENANCE SCHEDULE

Maintenance Time interval To be performed by


Clean air cabinet filter At least once a week Staff
*Clean suction filter Once in a year Qualified expert
Once in 2 years, if required
*Calibration Qualified expert
Once in a year, if required
Safety valve operation Qualified expert
After every 10,000 hours or
⃰ Replace water trap filter (25
24 months whichever Qualified expert
micron)
comes first
After every 10,000 hours or
Replace water Trap filter (5
24 months whichever Qualified expert
micron)
comes first
Every 12,000 hours or 30
Motor overhauling months whichever comes Qualified expert
first
Change felt filter Every 6 months Qualified expert
Replace suction filter Every 8000 hours Qualified expert
⃰ Fan checking Every 6 months Qualified expert

* Very important.

Shut Down

If the compressor is not going to be used for long period of time disconnect it from the main
power supply.

20
0434

Taurus Healthcare Pvt. Ltd. 0434

I. Clean Air Intake Filter

0434
Air intake filter should be cleaned at least once a week located on the front side of the unit.
Wash the filter in warm soapy water, rinse thoroughly and allow it to dry.

II. Clean Suction Filter, suction filter with PVC module assembly (Drawing and Pictorial
view)

PVC module

Felt

Suction Filter

Felt inside the PVC module should be replaced once a year. Suction filter to be cleaned after
7000 hrs or once a year. (Depending on environmental conditions)

21
0434

Taurus Healthcare Pvt. Ltd. 0434

0434
TROUBLESHOOTING

Prior to servicing the equipment, disconnect the equipment from the main power
supply. Only trained service personnel can perform the activities listed in the
troubleshooting guide.

PROBLEM POTENTIAL CAUSE SOLVING COMMON PROBLEMS


Compressor does not start No main power voltage Check supply. Power supply should
be 220VAC± 20VAC. Recommend
the use of voltage stabilizer in case of
large voltage fluctuation.
Loose wire terminal-tighten it.
Power cord defective-replace it.
Interrupted winding of Replace motor.
motor
Defective capacitor Replace capacitor.
Jammed piston or other Contact manufacturer/authorized
part dealer.
Compressor is making Damaged motor piston or Contact manufacturer/authorized
noise(knocking metal noise) bearing dealer.
Loose orange o -rings of Replace damaged orange o-rings.
the compressor motor
Fan is making noise Check its connection to mains if that
is OK then replace with new one
Low pressure at pressure Open the unit; visually inspect the
gauge and leakage in the component from where noise is
Pneumatic circuit can coming. Use the Teflon tap/sealant
cause loud noise. and reconnect the component. If till
there is leakage, replace the
corresponding component.
Voltage fluctuation Check the input supply. If there is
voltage fluctuation, use a voltage
stabilizer.
Solenoid valve not working Check power connection, if proper
voltage is there, replace the solenoid
valve.
Water coming out of outlets Dirty cartridge in water trap Clean or replace the water trap or
and filter regulator filter cartridges.
Unit out of calibration Long run usage Recalibrate the unit and check the
performance for 5-6 hrs.

22
0434

Taurus Healthcare Pvt. Ltd. 0434

0434

High humidity Pressure Relief or Pressure Calibrate the unit again and check the
output Regulator is out of calibration or performance for 5-6 hrs.
broken
Leakage of air from Pneumatic Inspect pneumatic distribution system
distribution system –seal loose connection and other
components.
Dirty air suction filter Replace the dirty air suction filter
Dirty air intake filter Clean or replace the dirty air intake
filter
Dirty cartridge of the water trap Dip the cartridge in hot water & clean
through normal brush. If cartridge still
seems to be dirty, replace them.
Low Pressure Pressure gauge not working Check pressure gauge & replace it.
Compressor pump not making Check the pressure at the outlet of the
enough pressure due to wear pump using some external pressure
and tear of piston rings or gauge in parallel with flow meter. It
bearing should be 50 LPM at 50 PSI or
visually feel the pressure at outlet. If
pressure is low, contact
manufacturer/authorized dealer.
Leakage in the unit Open the unit; visually inspect the
various components, if there is
leakage, use the Teflon tap/sealant
and reconnect the component. If still
there is leakage, replace the
corresponding component.
Pressure regulator/Pressure Recalibrate the unit. Check the
relief valve not working working for some time. If again
calibration is out. Replace the
components
Solenoid valve not working First check the wire connection. If
connection is OK then replace
solenoid valve.
Pressure switch not working Replace the pressure switch.
Dirty air suction filter Replace dirty air suction filter.
High temp. Non-functioning of fans Replace the fan which is not working
indicator is Alarm malfunctioning Check high temperature sensor/PCB
activated and replace corresponding part.

23
0434

Taurus Healthcare Pvt. Ltd. 0434

0434

How to call for Service


To get help to perform any of the preventive maintenance routines described in this chapter, or
request service, contact Taurus Healthcare Pvt. Ltd., Technical Service Department
TAURUS HEALTHCARE PVT. LTD...

A5-A6,Pragati Chamber, Ranjit Nagar


Commercial Complex, New Delhi 110008,India
Ph. : 011-45087820, Telefax : 011-45087810
Mobile : +91-9899966945,9711203605,9711203625
Email : [email protected]
Web : www.taurushealthcare.co.in, www.medicalcompressor.net

APPENDIX A: GLOSSARY

A Amps-A unit of electric current


AC Alternating Current
Bar Unit of Pressure
°C Degree Celsius-A unit of temperature
DB Decibels-A unit of Sound (noise)
DISS Diameter Index Safety System
Cm Centimeter- A unit of distance
°F Degree Fahrenheit-A unit of temperature
Hp Horse Power-A unit of power
Hz Hertz-A unit of frequency
I/O On/Off
L Liter-A unit of Volume
LPM Liter Per minute-A unit of Flow
MI Milliliters
O2 Oxygen
PCB Printed Circuit Board
PSI Pound per Square inch; A unit of Pressure
PSIG Pound per Square inch gauge
V Volts-A unit of electric voltage
THC Taurus Healthcare Pvt. Ltd.

24
0434

Taurus Healthcare Pvt. Ltd. 0434

0434

APPENDIX B: PRODUCT SPECIFICATION

Technical specification VA-09 VA-11 VA-12

Nominal voltage and 220VACV±10%,50 Hz 220VACV±10%,50 Hz 220VACV±10%,50Hz


frequency
Current and Power
2.38 A,500W 2.28 A,500W 2.28 A,500W
consumption
Compressor efficiency 50 lpm;3.5 bar 50 lpm;3.5 bar 20 lpm;3.5 bar
Air tank capacity 2 liter 2 liter 2 liter
Filtration of air 5 µm 5 µm 5 µm
Dew point depression at 3 5⁰C under the ambient 5⁰C under the ambient 5⁰C under the ambient
bar,20⁰C temperature temperature temperature
DISS 1160-A(3/4‟‟-16 DISS 1160-A(3/4‟‟-16 DISS 1160-A(3/4‟‟-16
Outlet connection
UNF) UNF) UNF)
Mode of operation Continuous Continuous Continuous
Automatic Back-up to On request On request On request
central pipeline
Noise level < 49 db(A) < 49 db(A) < 49 db(A)
Dimensions( l x b x h ) 57 x 48 x 87 cm 43.7 x 38.5 x 44 cm 43.7 x 38.5 x 44 cm
Dimensions( l x b x h ) with
57 x 48 x 98 cm 43.7 x 38.5 x 54.5 cm 43.7 x 38.5 x 54.5 cm
castors
Implementation according
Type B class I Type B class I Type B class I
to EN 60601-1,EN 12021
Classification according to
II A II A II A
MOD 93/42 EEC,2007/47 EC

Note: - Product specifications may change without notice.

25
0434

Taurus Healthcare Pvt. Ltd. 0434

0434

APPENDIX C: Parts List

SPARE PARTS VA 11, VA 12

S.No. Item Code Item Name


1 THC/CA-005 Cabinet
2 THC/MA-004 Castor wheel- 3" (Supo)
3 THC/MA-013 Castor wheel- 3" (Supo) with break
4 THC/PN-122 Cooling Coil (8 X 8 X 3)
5 THC/MT-001 Motor
6 THC/PN-060 SS-Tubing (18")
7 THC/PN-055 Solenoid Valve
8 THC/PN-034 Plastic-L
9 THC/PN-026 Hose Clamp-2"
10 THC/EL-004 Fan
11 THC/EL-026 Thimble Female
12 THC/EL-027 Thimble Male
13 THC/EL-015 Power Cord-(3MTR.)
14 THC/BC-008 BC-S(F1/4X1/4)
15 THC/BC-012 BC-S(MF1/4X1/8)
16 THC/BC-004 BC-L(MF1/4)
17 THC/BC-030 Dummy (M1/4)
18 THC/PFC-010 PFC-L(08,1/8)MALE
19 THC/PFC-008 PFC-L(08,1/4)MALE
20 THC/PFC-007 PFC-L(08,1/4)FEMALE
21 THC/PFC-027 PFC-T(08,Equal)
22 THC/CO-020 LN-SCREW (6 X 20)
23 THC/CO-030 SS-SCREW (6 X 25)
24 THC/CO-014 Wire-Black
25 THC/CO-016 Wire-Red
26 THC/CO-013 Washer Metal 1/4

26
0434

Taurus Healthcare Pvt. Ltd. 0434

SPARE PARTS VA 09

0434
S.No. Item Code Item Name

1 THC/CA-004 Cabinet
2 THC/ACC-012 Ventiair - Base / Basket / Basket Clip
3 THC/MA-005 Twin Wheel Castor - 4'' (Supo)
4 THC/MA-014 Twin Wheel Castor - 4'' (Supo) without Break
5 THC/MA-001 Back Handle
6 THC/MA-010 Chassis Plug
7 THC/MT-001 Motor
8 THC/EL-060 Fan
9 THC/EL-031 Socket 6 amp
10 THC/EL-016 Power Cord-(3mtr.)
11 THC/EL-048 Switch-Red
12 THC/PN-042 Pressure Gauge (0-100 PSI)
13 THC/PN-118 Cooling Coil (7x7x4)
14 THC/PN-060 Steel Braided Tubing -18"
15 THC/PN-080 SS-Cylinder
16 THC/PN-112 Water Trap Auto Drain AL1 (5micron)
17 THC/PN-069 Water Trap Auto Drain AL2 (25micron)
18 THC/PN-045 Relief Regulator
19 THC/PN-124 Pressure Regulator
20 THC/PN-055 Solenoid Valve
21 THC/PN-065 External Water Trap (Janatic)
22 THC/PN-008 Diss. Fitting (Janatic)
23 THC/PN-006 Air Outlet Nozzle (Janatic)
24 THC/PN-035 Plastic-T
25 THC/BC-037 BC-Safety Valve - 150 PSI
26 THC/CO-021 Module
27 THC/PM-019 Foam Filter Jali (Thin)
28 THC/PM-020 Foam Filter Jali (Thick)

27
0434

Taurus Healthcare Pvt. Ltd. 0434

APPENDIX D: List of accessories for all models


0434
S.No. Photo Part Name Ordering Part No.

Nozzle THC/PN-001

01

02
OUT-Compressor THC/PN-008
Air Outlet (J)

03 Cabinet Filter THC/PN-077

THC/PN-020
Compressor
04 Suction Filter

28
0434

Taurus Healthcare Pvt. Ltd. 0434

TEST REPORT 0434

Name : Medical Grade Air Compressor

Model : Ventiair VA-09/VA-11/VA-12

Sr. No. : THC/

Test conclusion : OK Tested

SPECIFICATION TEST RESULT

PHYSICAL INSPECTION

OUTPUT FLOW AND PRESSURE

CONNECTORS AND FITTINGS

MOTOR SR .NO.

SAFETY VALVE RELEASE AT (100 PSI)

TESTED HOURS FOR COMPRESSOR

NOISE LEVEL (<49 dBA)

REMARKS (IF ANY)

Authorized by :

29
0434

Taurus Healthcare Pvt. Ltd. 0434

Model No :
0434
Serial No :

Motor Serial No:

WARRANTY OF PERFORMANCE

Medical Air Compressor Ventiair is backed by all of the skill and


practical experience of users. Its reputation for quality has been built by
multitude of discerning buyers who measure value in terms of efficient
design, long life and satisfactory, trouble free performance.

Medical Air Compressor Ventiair is warranted to be free from defects


in material and workmanship and meet the published specification for one
year or 5000 hours whichever comes first.

However warranty is not included on accessories, rubber spare parts


and other consumable items.

For Taurus Healthcare Pvt. Ltd.

Authorized Signatory

30