0% found this document useful (0 votes)
285 views44 pages

Operating Manual: Swiss Made by Uwatec Ag

Uploaded by

Khodor Hamzi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
285 views44 pages

Operating Manual: Swiss Made by Uwatec Ag

Uploaded by

Khodor Hamzi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 44

Operating Manual

English

®
Smart TEC
SWISS MADE BY UWATEC AG
Safety considerations
You must carefully read and understand this entire manual before using your Smart TEC.

Diving has many inherent risks. Even if you follow the instructions of this
manual in a careful manner, it is still possible that you may be seriously injured
or die from decompression sickness, oxygen toxicity or some other inherent
WARNING
risk of scuba with Nitrox or compressed air. Unless you are fully aware of these
risks and are willing to personally accept and assume responsibility for those
risks, do not use Smart TEC!

Guidelines for the use of Smart TEC:


The following guidelines for using Smart TEC are derived from the latest medical research and the
recommendations of the American Academy of Underwater Sciences for diving with diving computers.
Following these guidelines will greatly increase your safety while diving, but cannot guarantee that
decompression sickness or oxygen toxicity will not occur.

• Smart TEC is designed for dives with Nitrox (to a max.100% O2) and compressed air (21%O2) only. Do
not use Smart TEC for dives made with other mixed gases.
• It is absolutely necessary to check the set mixture before each dive and to compare it to the gas mix-
ture currently used. Always remember: setting an incorrect mixture carries an inherent risk of decom-
pression sickness and/or oxygen toxicity! Maximum deviation from the measured mixture must not
exceed 1% O2. An incorrect gas mixture can be lethal!
• Only use Smart TEC with open circuit breathing systems. The gas mixture must be determined for each
pressure tank you use.
• Only use Smart TEC for diving with an independent breathing apparatus. Smart TEC is not designed
for long term exposures with Nitrox.
• Always observe the visual and audible alarm signals of Smart TEC. Avoid situations of increased risk
which are marked with a warning sign in this operating manual.
• If the ascent arrow appears, start to ascend.
• If the flashing ascent arrow appears, start to ascend immediately.
• Smart TEC has a ppO2 warning, the default limits of which are set at 1.4 bar ppO2max. This limit can
be adjusted via SmartTRAK. An alteration of the ppO2max to higher than 1.6 bar is dangerous and we
do not recommend this.
• Frequently check the "oxygen clock" (CNS O2 Limit), especially in the range higher than 1.4 bar ppO2.
Ascend and finish the dive if the CNS O2 exceeds 75%.
• Never dive deeper than the Maximum Operating Depth (MOD) pertinent to the gas mixture in use.
• Always check the diving limits considering the oxygen content and standard sports diving procedures
(decompression sickness, oxygen toxicity).
• In accordance with the recommended maximum diving limit of all instructional agencies, do not dive
deeper than 40 metres/130 feet.
• The danger of nitrogen narcosis has to be taken into consideration. Smart TEC gives no warning about
this.
• On all dives with Smart TEC, make a safety stop for at least 3 minutes at 5 metres (15 feet).
• All divers using dive computers to plan dives and indicate or determine decompression status must use
their own computer, which they take with them on all dives.
• If Smart TEC fails at any time during the dive, the dive must be terminated, and appropriate surfacing
procedures (including a slow ascent and a 3 to 5 minute safety stop at 5 metres /15 ft) should be initia-
ted immediately.
• Comply with the ascent rate and carry out any decompression stop required. If the computer should
fail for any reason, you must ascend at a rate of 10m (30 feet) per minute or less.
• On any given dive, both divers in a buddy pair must follow the most conservative dive computer for
that particular dive.
• Never dive without a buddy. Smart TEC does not substitute for a dive buddy.
• Only make dives that are appropriate to your level of dive training. Smart TEC does not increase your
knowledge of diving.

2 I UWATEC® Smart dive computers


Safety considerations I
• Always dive with back-up instruments. Make sure that you always use back-up instrumentation inclu-
ding a depth gauge, submersible pressure gauge, digital bottom timer or dive watch, and have access
to decompression tables whenever diving with a dive computer.
• Avoid repeated ascents and descents (yo yo diving).
• Avoid repeated heavy workload while at depth.

English
• Plan the dives to be shorter if they are made in cold water.
• After finishing the decompression or at the end of a no-stop dive, the final stage of the ascent should
be as slow as possible.
• You MUST be familiar with all signs and symptoms of decompression sickness before using Smart TEC!
Seek IMMEDIATE treatment for decompression sickness should any of these signs or symptoms occur
after a dive! There is a direct correlation between the effectiveness of treatment and the delay bet-
ween the onset of symptoms and the treatment for decompression sickness.
• Only dive with Nitrox after you have been thoroughly instructed by a recognised institution.

Repetitive dives
• Do not start your next dive before your CNS O2 % status has dropped below 40%.
• Diving with Nitrox: Make sure your surface interval is long enough (just like diving with compressed
air). Plan for a minimum surface interval of two hours. Oxygen, too, needs sufficient time to leave
the body.
• Match gas mixture to the intended dive.
• Do not attempt a repetitive dive if the microbubble warning is visible on the display.
NO

• Plan a day without diving once a week.


• If you have to change computers, wait at least 48 hours before carrying out your next dive.

Altitude and diving


• Do not dive at altitudes higher than 4000m (13000 feet).
• After a dive do not rise to altitudes that Smart TEC prohibits via the flashing altitude
segments. (->25)

Flying after diving


• After diving, wait at least 24 hours prior to flying.

Smart TEC dive instrument is a personal protective equipment in compliance with the essential safety
requirements of the European Union directive 89/686/EEC. RINA SpA, Via Corsica 12, I-16128 Genoa,
notified body no. 0474, have certified the conformity with the European Standard EN 250:2000 and EN
13319:2000.

EN250:2000 Respiratory equipment - Open circuit self contained compressed air diving apparatus -
Requirements, testing, marking (pressure gauge test).

EN13319:2000 Diving accessories - Depth gauges and combined depth and time measuring devices -
Functional and safety requirements, test methods. Any information on decompression obligation display-
ed by equipment covered by this standard is explicitly excluded from its scope.

I UWATEC® Smart dive computers


3
Introduction
Congratulations on purchasing Smart TEC and welcome to UWATEC. From now on you will enjoy the assi-
stance of the most extraordinary dive computer - equipped with UWATEC's most innovative technology -
while diving.
Smart TEC enables you to use up to three different gas mixtures during the same dive. However,
for an easy reading this manual mainly refers to dives with a single gas mixture. Information for diving with
several gas mixtures is marked with ( ) or has been summarised in special chapters.
We thank you for choosing Smart TEC and we hope you will enjoy safe dives in the future!
Further information on UWATEC Smart dive computers and other products by UWATEC can be found on
our web page at www.uwatec.com.
To make this manual easier to read we will use the term 'TEC' as an abbreviation for 'UWATEC Smart TEC
diving computer' throughout this booklet.
Safety considerations
Dive computers provide divers with data; they however, do not provide the knowledge how this data
should be understood and applied. Dive computers cannot replace common sense! You must therefore
carefully read and understand this entire manual before using your TEC.

Important remarks concerning signal words and symbols


This operating manual makes use of the following icons to indicate especially important comments:
Remarks Information and tips which are important for optimal use of the functions of
Smart TEC.

Danger!
Indicates a potentially hazardous situation which, if not avoided, could result
WARNING in death or serious injury.

The following symbols are used in the operating manual:

Flashing display Valid only for dives -> Page reference


with more than one e.g. ->10
gas mixture

Audible signals
4 sec. Audible attention Audible alarm signal
signal

Instructions for manual input


–+E Bridge contacts Push buttons
Example: bridging contacts B and E Press or press and hold (->9)
B e.g. press the upper left button

September 2004, Copyright© by UWATEC Switzerland

4 I UWATEC® Smart dive computers


Quick reference I

O2 mix icon (input) Microbubble level icon (input / MB level reduced)


Gauge icon Do not fly Icon
Logbook icon Do not dive icon
Dive planner icon Microbubble warning

English
Current depth Dive time / No-fly time / SOS duration
NO
Service icon Icon for stopwatch and safety stop
NO
CNS O2% Logbook indicator
Ascent obligation Total ascent time / Dive number
Altitude sections Ascent time icon
Too fast ascent DECO STOP Desaturation time indicator
LEVELSTOP No-stop time / Decompression duration / MB no-stop time
Maximum depth NO STOP
Temperature Level stop duration / Duration of Microbubble Warning
Microbubble level Desaturation duration / Duration of surface interval
Maximum Operating %
Depth (MOD) Increased workload warning
Remaining bottom time RBT / Number of active tanks
Transmitter warning
O2 mix Tank icons
Decompression stop obligation Tank pressure warning
Ignored decompression stop At the surface: several gas mixture have been set

Tank pressure / Gas consumed Decompression depth / Level stop depth


Partial pressure of oxygen ppO2
Oxygen toxicity CNSO2
Level stop indicator Ascent rate
No-stop indicator Battery capacity
Decompression stop indicator

Operating scheme
Select (press) and confirm (press and hold) gas mixtures
Start / Confirmation / Enter Activate the timer for safety stops (dive mode only, in
depths < 6.5 m / 20 ft) (press)
+ / Navigate Operate the stopwatch (gauge mode only) (press)
Set a bookmark (press)
– / Navigate
Toggle between maximum depth > temperature >
Display
MB level > …(press)
switched off Intermediate time and reset (in gauge mode only and
while stopwatch is active) (press)
Switching on the backlight (press and hold)
Switching on 10
Dive planner 37 Logbook 39 O2 mix 18 Microbubble level 28 Gauge mode 26

Navigation

( + Surface
– interval ) %

+ Percentage
%
Battery capacity 10 – of oxygen
+ + Gauge mode
Microbubble
on and off
level
NO STOP

%
% – –
+ Dive number
– Maximum
+ +
Bottom time operating
– – depth (MOD)
Exit (2 sec.) Exit Confirmation Confirmation Confirmation

Display switches off automatically after 3 minutes without operation.

I UWATEC® Smart dive computers


5
List of chapters
I Safety considerations ___________________________________2
Introduction ___________________________________4
Important remarks concerning signal words
and symbols ___________________________________4
Quick reference / Operating scheme ___________________________________5
List of chapters ___________________________________6

II System and operation ___________________________________8


1 System description ___________________________________8
2 Operation ___________________________________8
2.1 Operating elements ___________________________________8
2.2 SmartTRAK ___________________________________9
2.3 Switching on the display __________________________________10
2.4 Checking the battery capacity __________________________________10
2.5 Selection and activation of user functions __________________________________10
2.6 Active backlight __________________________________11
2.7 Switching off the display __________________________________11
3 SOS mode __________________________________11
4 Setting up Smart TEC __________________________________12
4.1 Mounting of transmitter __________________________________12
4.2 Pairing of transmitter and dive computer __________________________________13

III Diving with Smart TEC __________________________________15


1 Terminology / Symbols __________________________________15
1.1 General terminology / Display during no-stop phase __________________________________15
1.2 Display during decompression phase /
Remaining Bottom Time __________________________________15
1.3 Nitrox information (O2 information) __________________________________16
2 Attention messages and alarms __________________________________17
2.1 Attention messages __________________________________17
2.2 Alarms __________________________________17
3 Preparation for the dive __________________________________18
3.1 Setting the gas mixture and MOD __________________________________18
3.2 Setting the MB level __________________________________18
3.3 Preparation for diving with more than one gas mixture ________________________________18
3.4 Preparation for the dive and function check __________________________________18
4 Functions during the dive __________________________________19
4.1 Immersion __________________________________19
4.2 Setting bookmarks __________________________________19
4.3 Dive time __________________________________19
4.4 Current depth __________________________________19
4.5 Maximum depth / Temperature __________________________________19
4.6 Ascent rate __________________________________19
4.7 Partial pressure of oxygen (ppO2) / Maximum Operating Depth (MOD) ___________________20
4.8 Oxygen toxicity (CNS O2%) __________________________________21
4.9 Tank pressure __________________________________21
4.10 Remaining Bottom Time (RBT) __________________________________22
4.11 Decompression information __________________________________22
4.12 Safety stop timer __________________________________23
5 Functions at the surface __________________________________24
5.1 End of a dive __________________________________24
5.2 Desaturation time __________________________________24
5.3 No-fly time __________________________________24
5.4 Microbubble warning __________________________________24

6 I UWATEC® Smart dive computers


List of chapters I
6 Diving in mountain lakes __________________________________25
6.1 Altitude ranges __________________________________25
6.2 Prohibited altitude __________________________________25
6.3 Decompression dives in mountain lakes __________________________________25

English
IV Gauge mode __________________________________26

V Diving with microbubble levels (MB) __________________________________28


1 Comparison of dives with MB level L0 and MB level L5 __________________________________28
2 Terminology __________________________________29
2.1 Display during microbubble (MB) no-stop phase __________________________________29
2.2 Display during level stop phase __________________________________29
3 Preparation for a dive with microbubble levels (MB levels) __________________________________30
3.1 Setting the MB level __________________________________30
4 Functions during the dive with microbubble levels __________________________________30
4.1 Level stop information __________________________________30
4.2 Total time of ascent __________________________________31
4.3 Decompression obligation __________________________________31
4.4 Level stop and deco stop __________________________________32
5 Complete a dive with MB levels __________________________________32

VI Diving with more than one gas mixture __________________________________32

VII Dive planner __________________________________37


1 Planning a no-stop dive __________________________________37
2 Planning a decompression dive __________________________________38
3 Leaving the dive planner __________________________________38

VIII Logbook __________________________________39


1 Survey __________________________________39
2 Operation __________________________________39

IX Appendix __________________________________40
1 Technical information __________________________________40
2 Maintenance __________________________________40
2.1 Replacing the battery of the transmitter __________________________________40
3 Conversion of tank pressure __________________________________42
4 Warranty __________________________________42
5 FCC __________________________________43
6 Index __________________________________43

I UWATEC® Smart dive computers


7
II System and operation
1 System description
TEC displays all important dive and decompression data and comes with a unique multichannel receiver,
which can receive tank pressure data from up to 3 transmitters. Each transmitter is mounted at the high
pressure (HP) outlet of the regulator, where it measures the tank pressure and radio transmits the infor-
mation to TEC. UWATEC's specially patented transmission process prevents interference and ensures con-
tinuous and reliable reception
TEC has a data memory which stores the dive data. The data can be transmitted with an infrared inter-
face (IrDA) and the SmartTRAK software to a Windows® personal computer.
The SmartTRAK CD software is included with the SmartTRAK Package. Infrared interfaces are available in
PC stores. A list of recommended interfaces is available on the UWATEC homepage (www.uwatec.com).

Smart TEC

SmartTRAK
Infrared port

Infrared Interface
(option)

Pressure transmitter

2 Operation
On page 5 you will find an operating schematic.

2.1 Operating elements


Contacts TEC has 4 operating contacts B, E, +, – on the outside of the housing. For
manual operation, touch base contact B and any one of the other three contac-
- + E ts above the display with moistened fingers («bridging» contacts).
Contact B: Base contact, which has to be touched for all operations.

Contact E: Enter contact. It serves to switch on TEC and to confirm or enter the
CNS O2%
NO

NO
displayed value. It is therefore comparable to the ENTER or RETURN key of a
DECO STOP
LEVELSTOP
keyboard.
NO STOP

+ / – Contacts: These allow to navigate between menus and, once inside a


menu, to increase or decrease the indicated value.

8 II UWATEC® Smart dive computers


II System and operation I
Push buttons To use the push buttons TEC must be switched on. Operation of the push
buttons is divided into «press» and «press and hold (1 second)».
The push buttons , and activate the following functions:
Select the gas mixture (press)
Confirm the gas mixture (press and hold)

English
Activate the timer for safety stops (dive mode only, in depths < 6.5 m /
20 ft) (press)
Operate the stopwatch (gauge mode only) (press)
Set a bookmark (press)
Toggle between maximum depth > temperature > MB level > …
(press)
Intermediate time and reset (in gauge mode only and while stop-
watch is active) (press)
Switching on the backlight (press and hold)

2.2 SmartTRAK
With SmartTRAK you can transfer dive data to a personal computer and graphically display the data.
The following settings may be changed with SmartTRAK:
• Unit system metric/imperial
• Audible attention signal suppression selective
• Gauge mode
• Depth alarm 5 - 100 m (20 - 330 feet)
• Backlight illumination duration 2-12 sec.
• Maximum partial pressure of oxygen (ppO2 max) 1-1.95 bar
• Time limit to reset the O2 % mix to air no reset / 1 - 48 hrs.
• Minimum reserve pressure at the end
of the dive (basis for RBT calculation) 20 – 120 bar (300 - 1750 psi)
• Tank pressure alarm 50 - 200 bar (750 - 2900 psi)
• Replace audible RBT alarm by audible attention signal
• Workload sensitivity 25 steps
• Safety stop duration 1-5 min.

The following data may be recalled with SmartTRAK:


• Number of past dives
• Total length of past dives
• Atmospheric pressure
• Pairing information from up to 3 transmitters
• Dive profile
• Logbook
• Temperature curve
• Workload curve
• Alarms and attention messages
• Bookmarks

I UWATEC® Smart dive computers


9
2 Operation II
2.3 Switching on the display
• automatically, on submerging in water or when adaptation to atmospheric
pressure is necessary;
• manually, by bridging contacts on housing (B-E).

• When TEC is in state of rest no information is displayed but


the atmospheric pressure is continuously monitored. If a
NO higher altitude range is detected, TEC switches on for 3 minu-
tes automatically. -> 25
NO
CNS O2%

DECO STOP
LEVELSTOP
NO STOP • TEC switches on by bridging the contacts B and E. All
% segments light up for 5 seconds.

Afterwards the display shows the selected O2 mix, the


temperature and in certain circumstances an altitude range
Temperature ->25.
%
O2 Mix If the appertaining transmitter is switched on and located
within transmitting distance, the tank pressure is displayed,
otherwise <---> will be displayed. If no transmitter has been
Do not fly icon
paired yet, the display will be blank.
NO

No-fly time
If there is a remaining saturation due to the last dive or
Oxygen toxicity change of altitude, TEC also displays the remaining desatu-
Desaturation time ration time, the oxygen toxicity and the "no-fly time" ->24.
%

2.4 Checking the battery capacity


After switching on TEC you can check the battery capacity by bridging the
contacts B and E. The remaining capacity is displayed for 3 seconds as a per-
centage. If the value reaches 0%, the battery warning gets activated (->17) and
the battery has to be replaced by an authorised SCUBAPRO UWATEC dealer.
For a 7-day diving vacation TEC uses between 2 – 5 % of its battery capacity.
Battery capacity

2.5 Selection and activation of user functions


At the surface you can select the dive planner, the logbook
and gauge mode as well as the functions to enter the O2%
mix and the microbubble levels by bridging the contacts +
and B or – and B.

Gauge mode ->26


Input microbubble level ->30
Input O2 mix ->18
Logbook ->39
Dive planner ->37
After the selection of the desired function you can activate and deactivate it by
bridging the contacts B and E.

Details to the user functions are to be found on the pages mentioned above.

10 I UWATEC® Smart dive computers


2 Operation / 3 SOS mode
2.6 Active backlight
I
The display of TEC can be illuminated both on the surface and
underwater.
The backlight can be activated by pushing and holding button
for 1 second. The light will turn off automatically after 8

English
seconds or after the time selected via SmartTRAK.
The backlight can only be activated if the computer display is on.
The active backlight is no substitute
WARNING for a dive torch. When diving at
night or at increased depth we
recommend the use of a dive torch.

2.7 Switching off the display


On the surface TEC switches off automatically after 3 minutes without
operation.

3 SOS mode
Activation: automatic
Time remaining until If the diver remains above a depth of 0.8m (3 feet) for more than three minu-
SOS mode switches tes without observing a prescribed decompression, the computer will automa-
off automatically tically switch into SOS mode after the dive.
NO
The display shows the "SOS" sign and the remaining length of the SOS mode.
CNS O2%
NO
The dive will be entered in the logbook with "SOS". Other than that the SOS
mode has no further impact on the displays and functions on the surface.

%
• Serious injury or death may result if a diver does
WARNING not seek immediate treatment should any signs
or symptoms of decompression sickness occur
after a dive.
• Do not dive to treat symptoms of decompression sickness!
• Diving in SOS mode is extremely dangerous and you must assume full
responsibility for such behaviour. UWATEC will assume no liability.

Once in the SOS mode, the computer will lock up and will be inoperable as a
diving instrument for 24 hours.

A diving accident can be analyzed at any time in the logbook and downloaded
to a PC by means of the infrared interface (IrDA) and the SmartTRAK software.

I UWATEC® Smart dive computers


11
4 Setting up Smart TEC (transmitter and dive computer)
4.1 Mounting of transmitter
Each individual pressure tank requires a separate transmitter.
The transmitter is mounted at the high pressure (HP) outlet of the regulator’s first stage before the first
dive.

• Use air and Nitrox components in accordance with the law of the country.
WARNING

Procedure:
(Adaptor)

HP
Do not hold the transmitter Mount the transmitter at the HP Tighten the transmitter by
by its plastic part. oulet. If the threads do not match, means of a size 3/4" wrench.
you can get a fitting adaptor at your
diving equipment retailer.
The transmitter is best mounted laterally on the regulator.
If pressure tanks 2 and D are carried at the side, make sure the position of their transmitters in relati-
on to TEC resembles that of the transmitter of the main tank on the back.
It is advisable to mount the transmitter on the same side that the computer is located. This is the optimal
position for transmission.

Position of the transmitter Position of the transmitter


for left handers. for right handers.

Position of the transmitter for left Position of the transmitter for right
handers, if there is no connection handers, if there is no connection
possible on the left hand side. possible on the right hand side.

12 I UWATEC® Smart dive computers


4 Setting up (transmitter and dive computer) I
4.2 Pairing of transmitter and dive computer
To receive the data of the transmitters, each transmitter must be assigned to one tank symbol and paired
with TEC.
Pairing is necessary:
• before the first use of TEC with the transmitter;

English
• if you use a new transmitter or a new computer;
• after changing the battery;
• if a different tank symbol (1, 2, or D) is to be paired to the transmitter;
e.g. The transmitter is to be paired to tank symbol D instead of tank symbol 2.
To distinguish your tanks and transmitters you should regard the following assignments:

Bottom Mix Travel Mix Deco Mix


Transmitter 1 Transmitter 2 Transmitter 3
Lowest Intermediate Highest
oxygen % oxygen % oxygen %

Note that you can also set 2 or even 3 tanks to the same gas mixture. This comes
in handy when diving with twin tanks or with double outlet on a single tank.

Pairing procedure:
1. Shut the valve, depressurise the regulator and wait for 15 seconds.
2. Switch on TEC (bridge contacts B and E). Select the tank symbol you wish to
pair with a certain transmitter by pressing . Selected tank symbol will
flash. Confirm the selection by pressing for at least 1 second. Tank
symbol stops flashing.
Tank symbol 2 can only be selected after tank symbol D has been assig-
ned to a certain gas mixture.
3. Move the dive computer and transmitter into the
shown position.

Transmitter and dive computer must be in physical


contact during the entire pairing procedure.

4. Open the tank valve. The transmitter briefly sends a pairing sequence to the
computer.
5. The dive computer shows a flashing <PAIr> shortly after the valve has been
%
opened.
6. In order to confirm the pairing, the B and E contacts have to be bridged with-
in 5 sec. A beep confirms the input and the <PAIr> display remains still.

10 sec. 7. "PAIr" disappears after approx. 10 sec.

In case of faulty pairing, the display <FAIL> appears instead of <PAIr>. In that
case, completely purge the regulator and repeat the pairing procedure. This is
% only possible after at least 15 seconds.

I UWATEC® Smart dive computers


13
4 Setting up Smart TEC
• Pairing of transmitter and dive computer may already be carried out at home
and need only be executed once, before the first use.
• A transmitter can only be paired with a single tank symbol at a time. If
you try to pair the same transmitter with a second tank symbol, the first
pairing will automatically be deleted.
• The pairing of transmitter and tank symbol will remain valid, even if the
gas mixture is switched off (-> 34, switching off of gas mixtures).
• You may delete the pairing of transmitter and dive computer by means of
the SmartTRAK software.

How to check if transmitter and computer are paired correctly:


1. Switch on the computer manually (B and E)
2. To check the pairing of a certain tank symbol, press to select the tank
symbol. Confirm your choice by pressing for 1 second.
3. Move the computer into transmitting range of the transmitter.
4. Open the tank valve. The transmitter switches on automatically.
5. Check the display: Pairing has been carried out correctly if the pressure is
displayed within 5-10 seconds.

% Check the assignment of all tanks and transmitters as well as the pairing of
transmitters and dive computer!
Pairing ok

If the transmitter of a selected pressure tank has been correctly paired, but TEC
% cannot receive the tank pressure, "- - -" will appear on the display. In that case,
check the position of transmitter and dive computer.
Pairing present, no pressure
data available

If the specific channel of transmission (i.e. tank) has not been paired, or if an
%
existing pairing has been deleted via SmartTRAK, the display will remain empty.
In that case, transmitter and computer must be paired again.
Pairing not present

14 I UWATEC® Smart dive computer


III Diving with Smart TEC III
1 Terminology / Symbols
The information on the display of TEC varies depending on the kind of dive and the dive phase.
For information about diving with microbubble levels (MB level) see chapter IV
->28. Specific features of "Diving with more than one gas mixture", are descri-
bed in chapter VI -> 32.

English
1.1 General terminology / Display during no-stop phase

No-stop phase Decompression phase


Oxygen toxicity Dive time Dive phase during which surfacing is allowed Dive phase during which
CNS O2% Duration of without stop. surfacing is allowed only
Ascent rate the dive (min) after decompression
(Only while stops are completed.
%
ascending) SPEED

Depth Time
Current depth
In metres (feet)

No-stop time
CNS O2% Remaining time
at a given depth Current depth 24min
NO STOP
during which ascent
%
is allowed without
decompression
stop (min).
Tank 26.2 m Maximum depth
pressure RBT, Remaining
bottom time 37 min. Dive time No-stop time
O2% Mix
Elapsed bottom time
Selected oxygen fraction
tank 1
Maximum depth Temperature MB Level…
Maximum depth reached during the dive

1.2 Display during decompression phase / Remaining Bottom Time (RBT)

Decompression stop Total ascent time to surface (7 min)


All required decompression Time
stops must be observed.
Depth

3m
Deepest
6m
decompression
Decompressionstop duration 2 min stop depth
CNS O2%

DECO STOP

Decompression RBT 21 min


stop depth
Deepest stage is 35.7 m
displayed.
No-stop range Decompression range
Decompression
stop duration Total ascent time Remaining bottom time, RBT
Prescribed duration of the Including decompression Maximum remaining time at current
decompression stop at the stops in minutes. depth including all decompression
displayed decompression obligations (minutes).
stage (minutes).

III UWATEC® Smart dive computers


15
1 Terminology / Symbols
1.3 Nitrox information (O2 information)
For dives with compressed air in normal recreational diving, nitrogen is the decisive gas for the decom-
pression calculations. When diving with Nitrox, the risk of oxygen toxicity rises with the increase of the frac-
tion of oxygen and the increase of depth and can limit dive time and the maximum depth. TEC includes
this in the calculations and displays the necessary information:

<O2% MIX> Fraction of oxygen: The fraction of oxygen in the Nitrox mixture can be set bet-
ween 21% (normal compressed air) and 100% in 1% increments. Your
selected mix will be the basis for all calculations.

ppO2 max. Maximum allowed partial pressure of oxygen: the higher the fraction of oxygen
in the mixture, the shallower the dive depth at which this value of the
partial pressure of oxygen is reached. The depth at which ppO2 max. is reached
is called Maximum Operating Depth (MOD). Default setting is 1.4 bar, but it can
be set by means of SmartTRAK between 1.0 and 1.95 bar. When you enter the
settings for the gas mixture, TEC will display the ppO2 max. limit setting and the
corresponding MOD. TEC warns the diver audibly and visually once the depth
is reached at which the ppO2 reaches the maximum allowed value.

• The partial pressure limit set by means of SmartTRAK can be reduced manually
at TEC (->18, setting the gas mixture).
• The CNS O2% value/alarm is not influenced by the selected ppO2max. setting.

<CNS O2> Oxygen toxicity: With the increased ppO2


percentage of oxygen, the oxygen in 0,21 bar
the tissues (especially in the central
CNS O2%
nervous system (CNS)) becomes im- decreases
portant. If the partial pressure of oxy-
gen rises above 0.5 bar, the CNS O2 0,5 bar
value increases, if the partial pressu- CNS O2%
increases
re of oxygen is below 0.5 bar, the
CNS O2 value decreases. The closer
the CNS O2 value is to 100%, the
closer the limit where symptoms can
occur. See page 21.

Nitrox diving may only be attempted by experienced divers after proper training
WARNING from an internationally recognized agency.

16 III UWATEC® Smart dive computers


2 Attention messages and alarms III
TEC draws the diver’s attention to certain situations and warns the diver of unsafe diving practices. Atten-
tion messages and alarms are always visual and audible under water, only visual at the surface except the
decompression alarm.

The audible attention messages (but not the alarms) can be selectively

English
switched off with SmartTRAK.
2.2 Alarms
WARNING Serious injury or death may result from failing to
immediately respond to alarms given by TEC.
2.1 Attention messages
Attention messages are communicated to the Alarms are given to the diver visually by flashing
diver visually by symbols, letters or flashing figu- symbols, letters or figures. In addition, an audible
res. In addition, two short audible sequences can sequence in one frequency can be heard during
be heard (in an interval of 4 seconds) in two diffe- the whole duration of the alarm.
rent frequencies under water.

4 sec (can be switched off) (can not be switched off)

Attention messages come up in the following An alarm occurs in the following situations:
situations: (More information can be found on the listed
(More information can be found on the listed pages.)
pages.)
Page Page

• Maximum Operating Depth / max ppO2 • Oxygen toxicity reaches 100% 21


is reached 20 • Ignored decompression 23
• Set maximum depth is reached 19 • Remaining Bottom Time zero
• Oxygen toxicity reaches 75% 21 ( can be switched to attention beep) 22
• No-stop time = 2 minutes 22 • Exceeding the prescribed ascent rate 20
• Prohibited altitude* (surface mode) 25 (Particular scale of beeps, ->20)
• Entering decompression when diving with L0 23
• Remaining Bottom Time < 3 minutes 22 • Low battery alarm* see below
• Tank pressure has reached set warning level 21
• Increased workload 21 Low battery alarm TEC*
• Depth for tank switch has been reached 35 The service symbol appears if the battery capacity
reaches 0%.
Take the unit to your
Diving with microbubble levels (L1-L5): authorised SCUBAPRO
• MB no-stop time = 0 30 UWATEC retailer.
• MB level stop ignored 31
• MB level reduced 31
• Entering decompression when diving
with MB level L1-L5 31 Transmitter battery low: *
<bAt> flashing and alternating with the display of
the tank pressure.
NO STOP NO STOP Replace the battery in
% the transmitter. ->40

*without audible alarm


*without audible alarm

III UWATEC® Smart dive computers


17
3 Preparation for the dive
3.1 Setting the gas mixture and MOD
Before every dive and after changing the tank, make sure that the settings for the
WARNING gas mixtures correspond with the current mixtures used. An incorrect setting cau-
ses TEC to miscalculate this particular dive. If the fraction of oxygen is set too
low this can lead to oxygen poisoning without warning. If the value is set too
high decompression sickness may occur. Inaccuracies in the calculations are carried
over to repetitive dives.
For dives with a single gas mixture, assign the gas mixture to tank symbol 1 and shut off gas mixtures
2 and D. ->34
If more than one gas mixture has been selected, TEC will display at the surface and the number of
gas mixtures.
For dives with several gas mixtures please read pages 32 and 13 for further reference.
To set the gas mixture, TEC must be in user mode.
1. Press to choose the tank for the O2 mixture you are going to enter. The
tank symbol will start flashing. Press and hold to confirm your selection.
… 2. Bridge contacts B and + or B and – until the symbol for the setting of the
… O2 mixture appears.
3. Confirm that you wish to change the displayed oxygen fraction by bridging
B and E.
4. Change the oxygen fraction in increments of 1% by bridging B and + or B
+ – and –. TEC will display the current fraction of oxygen, the maximum partial
pressure limit, ppO2 max (as pre-set by means of SmartTRAK) and the MOD.
5. Confirm the selected percentage with B and E.
%
6. By bridging B and + or B and – you can change the
MOD for the chosen fraction of oxygen. TEC will now
O2% mix + –
display the corresponding partial pressure limit (ppO2
PpO2max. max) for the new MOD.
Max. Operating 7. Confirm your MOD settings with B and E.
Depth MOD % ppO2max.

• Without confirmation the display will disappear after 3 minutes and your entries will not be
accepted.
• Please refer to page 32 for the setting of gas mixtures for tanks 2 and D.
• The time to reset the O2 % mix to air can be set with SmartTRAK between 1 and 48 hours
or to "no reset" (default).

3.2 Setting the MB level See chapter V, ->28


3.3 Preparation for diving with more than one gas mixture See chapter VI, ->32
3.4 Preparation for the dive and function check
The following descriptions of the preparation are based on the assumption that all transmitters have been
correctly mounted at the HP outlet of the regulators (->12) and successfully paired with TEC (->13).
1. Mount the regulator together with the transmitter on the tank
2. If present, check the reserve valve of your tank, the reserve valve must be
WARNING open.
3. Switch on TEC (B-E) and check the test display: Are all elements of the dis-
NO play activated? Do not use TEC if the display does not show all elements.
4. Open the valve (transmitter will switch on automatically) and check the tank
NO
CNS O2%

DECO STOP
pressure (after approx. 10 seconds). If the pressure is insufficient, change the
LEVELSTOP
NO STOP tank.
% 5. Check the connections and instruments for leaks. Never dive with leaky
equipment!
Repeat 4 and 5 with all transmitters you are going to use.

18 III UWATEC® Smart dive computers


4 Functions during the dive III
4.1 Immersion
After immersion, starting at a depth of about 0.8 m (3 ft), TEC automatically selects gas mixture 1, all
diving functions are monitored, i.e. depth and dive time displayed, maximum depth stored, saturation of
tissues calculated, no-stop time or decompression prognosis determined, ascent rate controlled and dis-
played and the correctness of the decompression procedure supervised. In addition, TEC also shows the
tank pressure and about 2 minutes into the dive the Remaining Bottom Time (RBT) is displayed.

English
4.2 Setting bookmarks
During the dive you can create bookmarks in your dive profile by pressing . An audible signal con-
firms the creation of the bookmark. These bookmarks will be graphically displayed in the dive profile of
SmartTRAK.
To visualize the bookmarks on the dive profile, you must select the box “Gene-
rate bookmarks” under “Program options” in SmartTRAK.

4.3 Dive time


Dive time The whole time spent below a depth of 0.8m (3 feet) is displayed as dive time
in minutes. The time above 0.8m (3 feet) is counted as dive time only if the
diver descends again below 0.8m (3 feet) within 5 minutes.
While the dive time is running, the colons on the right of the figures are
flashing in one second intervals. Maximum dive time displayed is 999 minutes.
NO STOP
If a dive lasts longer than 999 minutes the dive time display starts
%
again at 0 minutes.

4.4 Current depth


Current depth is given in 10 cm increments (1 foot). On switching
on and at a diving depth of less than 0.8 m (3 ft) the display shows
<– – –>.
Depth measurement is based on freshwater. Therefore,
NO STOP TEC shows a slightly greater depth when diving in salt
% water, depending on the salinity of the water. No calcu-
lation however is affected.
%
Current depth
4.5 Maximum depth / Temperature
Maximum depth is only displayed if it exceeds the current depth by more than
CNS O2%
1 m (3 feet) (maximum indicator function). If maximum depth is not displayed,
TEC shows the temperature.
NO STOP
4 sec
% WARNING
Maximum depth CNS O2%

Set maximum depth reached


If the maximum depth set with SmartTRAK has
NO STOP
been reached (default 40m/130 feet), the display
%
will flash and the ascent arrow will be displayed.
Ascent arrow
Ascend until the ascent arrow disappears.
4.6 Ascent rate
Optimal ascent rate varies depending on depth between 7 and 20 m/min (23
and 67 ft/min). It is displayed as a percent of the reference variable ascent rate.
If the ascent rate is greater than 100% of the set value, the black arrow
<SLOW> appears. If the ascent rate exceeds 140%, the arrow starts flashing.
TEC provides an audible alarm if the ascent rate is 110% or greater. The inten-
NO STOP
sity of the alarm increases in direct proportion to the degree that the prescri-
%
Ascent rate bed ascent rate is exceeded.

III UWATEC® Smart dive computers


19
4 Functions during the dive

The prescribed ascent rate must be observed at all times! Exceeding the pre-
WARNING scribed ascent rate can lead to microbubbles in the arterial circulation which
can lead to serious injury or death due to decompression sickness.

• In case of an improper ascent TEC may require a decompression stop even


within the no-stop phase because of the danger of microbubble formation.
• The decompression duration necessary for the prevention of microbubbles can
increase massively if the ascent rate is exceeded.
• From great depth a slow ascent may cause heightened saturation of tissues
and an extension of both decompression duration and total ascent time.
At shallow depth, a slow ascent may shorten the decompression duration.
• Display of the ascent rate has the priority over <CNS O2>.

Ascent rate Visual Audible alarm


WARNING alarm
S
% L
O
W
SPEED

S
L
% O
W
SPEED

S
% L
O
W
SPEED

S
% L
O
SPEED W

Excessive ascent rates for longer periods are entered in the logbook.

4.7 Partial pressure of oxygen (ppO2 max) / Maximum Operating Depth (MOD)
The maximum partial pressure of oxygen ppO2 max (default 1.4 bar) determines
4 sec
the Maximum Operating Depth (MOD). Diving deeper than the MOD will expo-
se the diver to oxygen partial pressures higher than the set maximum level.
CNS O2%
The MOD and consequently the ppO2 max can be reduced manually at TEC
(->18, setting the gas mixture, point 6).
NO STOP In addition the maximum allowed ppO2 can be set by means of SmartTRAK
% between 1.0 to 1.95 bar.

The MOD is a function of ppO2 max and the mixture used. If during the dive
WARNING the MOD is reached or passed TEC sends an audible attention message, the
ascent arrow appears and the current depth display starts flashing.

Ascend to a shallower depth in order to diminish the danger of oxygen poiso-


ning.
Change to the gas mixture that is required at this depth by
selecting the appropriate tank symbol.

• The MOD should not be exceeded. Disregarding the warning can lead to
WARNING oxygen poisoning.
• ppO2 max should not be set higher than 1.6 bar.

20 III UWATEC® Smart dive computers


4 Functions during the dive III
4.8 Oxygen toxicity (CNS O2%)
CNS O2%
TEC calculates oxygen toxicity from depth values, time and the gas mixture and
displays it in place of the ascent rate. The toxicity is expressed in 1% increments
of a maximum tolerated value (O2 clock). The symbol <CNS O2> is displayed
NO STOP
together with the percentage.
%

English
Oxygen toxicity An audible attention signal goes off if oxygen toxi-
4 sec WARNING city reaches 75%. The symbol <CNS O2> flashes
and the ascent arrow appears.
Ascend to shallower depth to decrease oxygen
CNS O2%
loading.

When oxygen toxicity reaches 100%, an audible


WARNING alarm goes off every 4 seconds. <CNS O2>, the
CNS O2%
precentage value and the ascent arrow flash. Dan-
ger of oxygen toxicity!
NO STOP
Start ascent at once.
% • During an ascent and if the CNS O2 % value does not increase anymore (due
to a lower partial pressure of oxygen), the audible warning is suppressed.
• During the ascent, the display of the oxygen toxicity is replaced by the ascent
rate. If the ascent is stopped, the display changes back to the indication of the
CNS value.
• TEC will display CNS O2 % values exceeding 199 % with 199 %.

4.9 Tank pressure The tank pressure is also used for the calculation of
% the remaining bottom time (RBT) and the workload.

When the tank pressure reaches the set warning


4 sec
WARNING pressure (SmartTRAK) an audible alarm goes off
and the tank symbol is shown. Default value of
warning pressure: 100 bar (1450 psi)
Do not dive any deeper. Start to ascend soon.

In case of increased workload, TEC displays a lung


4 sec
WARNING symbol and an audible alarm goes off. (The sensi-
tivity of workload can be changed with Smart-
TRAK).
In order to prevent additional saturation, reduce
exertion, relax and breathe more slowly.

If TEC receives no data for 30 seconds, an audible


4 sec WARNING alarm goes off and the transmitter warning
appears.
% If TEC does not receive pressure data for another 40 seconds, it will activa-
te another audible alarm. RBT and transmitter warning will no longer be dis-
played. Instead of the tank pressure value TEC will display "- - -".
Transmitter warning Check the position of transmitter and TEC. Start ascending at once.
TEC will switch back to its normal display as soon as new data is received.

If the tank pressure is lower than 14bar/200psi the


WARNING transmitter switches off and TEC will display "---".
Do not let the tank pressure drop below
14bar/200psi.
If tank pressure
<14 bar/200psi
III UWATEC® Smart dive computers
21
4 Functions during the dive
4.10 Remaining Bottom Time (RBT)
RBT is the time left at the current depth until the point of time when the ascent
must be started. The RBT is calculated on the basis of the current tank pressure,
breathing rate, the temperature, and the dive data so far recorded. The RBT is
based on the assumption that the tank pressure should amount to the set pres-
sure (default 40 bar/600 psi) at the end of the dive. Alterations can be made with
SmartTRAK. A graphic representation of RBT is on page 15.
During dives with several gas mixtures, the calculation of the RBT is based
on the assumption that the diver will use the current gas mixture to finish the
dive.->34, 35
Never allow the RBT to go below three minutes. If the RBT goes below three
WARNING minutes there is a danger of insufficient supply of gas mixture for the ascent
as well as an increased risk of decompression sickness, and serious injury or
death may result!
Correct calculation of RBT when using a reserve or “J“ type valve is possible only if the reserve func-
tion of the valve is in the open (down) position during the dive.

4 sec If the RBT drops below three minutes, an audible


WARNING attention signal is activated, the ascent arrow is dis-
played and the tank icon start flashing.
Start ascent immediately.

NO STOP
The RBT value should never reach <0:>. With
%
WARNING RBT=0 the remaining tank reserve may not be suf-
ficient for the ascent.
RBT < 3 minutes
When the last minute has passed (RBT=0) an audible alarm is activated every
4 seconds. The RBT, the ascent arrow and the tank icon start flashing. The
audible alarm* on exceeding the RBT is suppressed at depths less than 6.5
m (21 ft) if TEC is in the no-stop phase.
Start ascent at once.
NO STOP
* The audible RBT alarm can be changed to an audible attention signal by
%
means of SmartTRAK. 4 sec
RBT = 0 minutes

4.11 Decompression information


No-stop time is displayed if no decompression stops are necessary. The arrow NO STOP is visible if no
decompression stops are necessary. The figures indicate no-stop time in minutes.

• No-stop display <199:> means remaining time of 199 minutes or


more.
• No-stop time is calculated on line and influenced by the current
NO STOP
workload and current water temperature
%

If no-stop time drops below 3 minutes, an audible


No-stop time attention signal is activated and the no-stop value
WARNING
4 sec begins to flash.
If no-stop time is less than 1 minute, the no-stop
display shows the flashing value "0".
CNS O2%
In order to prevent a decompression dive, ascend
slowly until the no-stop time is 5 minutes or more.
NO STOP

% Dives that require decompression stops are not


WARNING recommended.

22 III UWATEC® Smart dive computers


4 Functions during the dive III
Decompression values On entering the decompression phase, the arrow NO STOP disappears, the
arrow DECOSTOP appears and an attention beep goes off. Right beside the
arrow, the deepest decompression stage in metres (feet) is displayed. Next to
CNS O2%
the decompression stop depth, the decompression stop duration of the dis-
played stage appears in minutes. The display <3m 7:> means that a decom-

English
DECO STOP

pression stop of 7 minutes at a depth of 3m has to be made.


%
When a decompression stop has been finished, the next higher decompression
stop is displayed. When all decompression stops have been made, the arrow
Decompression DECOSTOP extinguishes and the arrow NO STOP reappears. The indication of
stop depth time on the lower right shows the no-stop time again.
Decompression
time Deco stop depths deeper than 27m (90 ft) are displayed as <– – : – – >
The decompression alarm is activated if the
WARNING decompression stop is ignored. The arrow
DECOSTOP , the decompression stop duration and
CNS O2%
decompression stop depth begin to flash and an
audible alarm goes off.
DECO STOP
Due to the formation of microbubbles decompression can increase massive-
ly if a decompression stop is ignored. When the surface is reached during
the decompression alarm, the arrow DECOSTOP , the decompression stop
duration and decompression stop depth continue flashing, in order to
Ignored decompression
point to the risk of a decompression accident. The SOS mode is activated 3
alarm
minutes after the dive if corrective action is not taken (->11).
If the total (cumulative) duration of the decompression alarm is longer than
a minute, it is entered in the logbook.
Descend to the prescribed decompression stop depth immediately!

Total time of ascent As soon as decompression stops are necessary TEC shows the total time of ascent.
This includes the ascent time from the current depth to the surface and all decom-
pression stop obligations.

DECO STOP The total time of ascent is calculated on the basis of the prescribed
ascent rate and a normal workload. Total time of ascent can be
%
subject to change if the ascent rate is not ideal (100%) or if TEC
detects a higher workload.
Total time of ascent Ascent time greater than 199 minutes is displayed as <– –>.

On all dives with TEC, make a safety stop for at least


WARNING three minutes at a depth of 5 m (15 feet).

4.12 Safety stop timer


The safety stop timer displays the time span a diver should spend at the safety stop depth at the end of
the dive. The timer is activated by the diver and counts back to zero, starting with the time that was
pre-set with SmartTRAK. It can be restarted any number of times. Range: 1 - 5 minutes (default 3 min.)
CNS O2%
The safety stop timer can be activated under the following conditions: Depth
< 6.5 m (21 ft), no-stop display 199 min., gauge mode is switched off.
STOP

Activate the safety stop timer by pressing . The timer begins to count
% backwards and a bookmark will be created in the dive profile.
safety stop The safety stop timer will switch off automatically if the depth exceeds 6.5 m
icon (21 ft) or the no-stop phase is shorter than 199 minutes.
remaining time span
minutes
seconds

III UWATEC® Smart dive computers


23
5 Functions at the surface
5.1 End of a dive
After reaching the surface (<0.8 m/3 ft) TEC remains in dive mode for 5 minutes.
The delay allows for surfacing for a short period for orientation.

After 5 minutes the dive is closed and it is entered into the logbook.

For the calculations of the desaturation and no-fly time it is assumed that the
WARNING diver breathes air while on the surface.

5.2 Desaturation time


After the dive has been closed DESAT desaturation time in hours and
NO
minutes and, if available, oxygen toxicity is displayed. Desaturation time is
CNS O2%
determined either by oxygen toxicity, nitrogen saturation or the regression of
microbubbles, depending on which requires the longer time. Oxygen toxicity
(<CNS O2>) is displayed and adjusted until the value becomes 0%.
% Desaturation time is indicated until the next dive or until it reaches zero.
The display is switched off to save energy three minutes after the last manipu-
Oxygen toxicity lation is made. The calculations are nevertheless continued in the background.
Desaturation time

5.3 No-fly time


No-fly time The <no-fly time> is indicated beside the icon <do not fly>. <no-fly time> is the
time in hours that should pass before a flight and is displayed and adjusted until
Do not fly icon
the value becomes 0 hours.
NO

CNS O2% Flying while TEC displays <do not fly> may lead to
WARNING serious injury or death from decompression sickn-
ess.
%

5.4 Microbubble warning


Through repetitive dives microbubbles accumulate in the lungs if the surface
NO interval is not long enough. Ignoring decompression stops or ascending at an
CNS O2%
NO
excessive rate can also lead to microbubbles in tissues. In order to reduce the
risk of decompression sickness for repetitive dives, the surface interval should
be planned long enough. If TEC calculates that the formation of microbubbles
occurs during the surface interval, it will advise a diver to extend the surface
interval via the microbubble warning. The duration of the microbubble war-
Microbubble
ning is visible by entering the dive planner -> 37.
warning
Desaturation time If the <microbubble warning (NO DIVE)> is visible
WARNING during the surface interval, the diver should not
undertake another dive.

If the dive is made in spite of the microbubble warning, the diver must cope
with a clearly shorter no-stop time or an extension of decompression. Also, the
duration of the microbubble warning at the end of the dive can increase con-
siderably.

24 III UWATEC® Smart dive computers


6 Diving in mountain lakes III
6.1 Altitude ranges TEC measures the atmospheric pressure every 60 seconds even while the dis-
play is switched off. If the computer detects a sufficient increase in altitude, it
switches on automatically and indicates the new altitude range (1-4) and the
desaturation time. Desaturation time indicated at this moment refers to
adaptation time at this altitude. If the dive starts within this adaptation time,

English
TEC treats it as a repetitive dive, since the body is offgassing.
%
Altitude is divided into five ranges, which are influenced by barometric pressu-
re. That is why the defined altitude ranges overlap on their fringes. If a moun-
Altitude range 1 tain lake is reached, the altitude range is indicated at the surface, in the log-
book and in the dive planner by a stylised mountain filled with one or more of
Desaturation time 4 segments representing the 4 ranges. Sea level to an altitude of approximately
Adaptation time 1000m (3300 feet) is not Switching
indicated. In the following at approx.
No deco data
diagram, you can see the gauge mode 4000 m
610 mbar
8.85 psi
Altitude ranges approximate breakdown of 13120 ft
the altitude ranges:
3000 m
725 mbar
9840 ft
10.51 psi
1 2 3 4 2000 m 815 mbar
6560 ft 11.82 psi

1000 m 905 mbar


3280 ft 13.12 psi

0m
6.2 Prohibited altitude

TEC shows via flashing altitude segments while at


WARNING the surface to which altitude the diver may not rise.

Ascent to altitude range 3


and 4 prohibited. Max. altitude: 850 m 1650 m 2650 m 4000 m
Max. allowed altitude: 2790 ft 5413 ft 8694 ft 13120 ft
2650 m (8694 ft).
The ascent prohibition can also be displayed together with an
altitude range:
Example: You are at 1200 m (3937 ft) (altitude
range 1) and you may ascend to range 2 only
(2650 m / 8694 feet). You may not rise to the alti-
tude range 3 or 4.

6.3 Decompression dives in mountain lakes


In order to assure optimal decompression even at higher altitudes, the 3m (10
ft) decompression stage is divided into a 4 m (13 ft) stage and a 2 m (7 ft) stage
in altitude ranges 1, 2 and 3. The prescribed decompression stop depths are, in
sequence, 2m / 4m / 6m / 9m… (7 ft / 13 ft / 20 ft / 30 ft…).

%
If atmospheric pressure is below 620 mbar (8.99 psi) (altitude higher than
4100 m / 13450 ft above sea level), no decompression data is calculated and
Dive at altitude range 4: displayed (automatic gauge mode). In addition RBT and the dive planner are
• no deco data not available anymore. The oxygen toxicity and the tank pressure are still indi-
• no RBT cated.

III UWATEC® Smart dive computers


25
IV Gauge mode
In gauge mode ALL audible and visual alarms and attention messages are
WARNING turned off. This includes ascent speed, low tank pressure and interrupted
signal from transmitter.
In gauge mode TEC will display depth, dive time and tank pressure, the maximum depth is stored, ascent
rate and tank pressure are monitored. By pressing you can switch from maximum depth to tempe-
rature and vice versa. You can also activate the stopwatch and check the split time. Gauge mode does not
support the calculation of no-stop time or the supervision of decompression. Supervision of ppO2 max and
CNS O2 % will also be switched off. TEC will display no information about RBT or microbubble develop-
ment. Bookmarks can be set. The settings for gas mixtures, MOD and microbubble level cannot be set and
the dive planner cannot be selected.
• Dives in gauge mode are performed at your own risk!
WARNING • After diving in gauge mode you should wait for at least 48 hours before
using a decompression computer.

Switching the gauge mode on and off


Gauge mode can be switched on and off at the surface, if no desaturation time
is being prescribed.
After diving in gauge mode, TEC can not be used as dive computer for 48
hours.

Procedure:
1. Bridge contacts B and + until the gauge symbol and "on" or "off" are dis-
played.
2. Confirm with B and E that you wish to activate or deactivate the gauge
mode. The gauge symbol starts flashing.
3. By bridging contacts B and + or B and – the gauge mode is switched on and
off.
4. Confirm your settings with B and E.

Without confirmation the display will disappear after 3 minutes and your
entries will not be accepted.

Diving in gauge mode


After immersion, starting at a depth of about 0.8m (3 ft), TEC automatically sel-
ects gas mixture 1.
The following information is displayed in gauge mode:

Gauge icon
Current depth Dive time

Maximum depth /
Temperature

Tank pressure

Changing the gas mixtures in gauge mode


In gauge mode it is up to the diver to decide on a time and depth to change
the gas mixture. TEC will not indicate a necessary change of gas mixture. For
more information about changing the gas mixture see page 35.

26 III UWATEC® Smart dive computers


IV Gauge mode III
Stopwatch
In gauge mode, after immersion, TEC will automatically monitor the dive time
and at the same time activate the stopwatch. If the gauge symbol and "on"
are displayed (-> 26, point 1) the stopwatch can also be activated at the surfa-
ce by pressing . The stopwatch will run for a maximum of 1 hour at the

English
surface and no longer than 24 hours underwater.
By pressing you can start, stop and restart the stopwatch.
Once the stopwatch has been stopped, you can reset the timer to zero with
Stopwatch icon .
Timing Hours
Minutes While the stopwatch is running you can measure split times and return to the
Seconds timer display by pressing . The timekeeping is not interrupted in the
meantime.

Stopwatch Example
will start 00h00 00 Stopwatch activated Split time
automatically (stopwatch continues to
after run in the background)
immersion 00h00 15 Stopwatch is running

00h02 35 Stopwatch halted

Automatically
at the end of 00h00 00 Time reset Icon for split time
dive
When split time is
Exit stopwatch taken, the temperature is
displayed for 5 seconds.

• By pressing you can select a different pressure tank even while the
stopwatch is running or the split time is being displayed.
• At a diving depth of less than 0.8m (3 ft) the stop watch will automatically
measure and display the split time. If the diver descends again the stopwatch
timer will be displayed again.

After diving in gauge mode


TEC shows the remaining time span during which it cannot be used in compu-
ter mode. Once the waiting period is over, the gauge mode can be switched
off manually. ->26
NO

The no-fly time after diving in gauge mode is 48 hours.

Desaturation time will not be displayed.

III UWATEC® Smart dive computers


27
V Diving with microbubble levels (MB)
The following chapter deals with the characteristics of diving with microbubble levels (MB level).
For general information about displays and features of diving with TEC see chapter III.
Microbubbles are tiny bubbles that can build up inside a diver's body during any dive and normally dissipate
naturally during an ascent and on the surface after a dive. Dives within no-stop time and observance of decom-
pression stops do not prevent the formation of microbubbles in the venous blood circulation.
Dangerous microbubbles are those migrating into the arterial circulation. The reasons for the migration
from the venous blood circulation to the arterial circulation can be a great many microbubbles collecting in
the lungs. UWATEC has equipped Smart dive computers with a new technology to protect from microbubbles.
The diver chooses – according to his/her needs – an MB level and influences through it the level of pro-
tection from microbubbles. Diving with MB levels requires additional ascent stops (level stops), the ascent is slo-
wed down and the body gets more time to desaturate. This works contrary to the formation of the microbub-
bles and increases the safety.
Smart TEC features 6 microbubble levels (L0-L5). Level L0 corresponds to UWATEC's well-known
decompression model ZH-L8 ADT and does not require level stops due to microbubble formation. Levels L1
to L5 offer additional protection from microbubble formation with level L5 offering the highest protection.
Similar to the display of information during decompression dives or dives within no-stop time, TEC displays
depth and duration of the first level stop as well as the total time of ascent as soon as the MB no-stop time
has run out. As the MB no-stop time is shorter than the ordinary no-stop time a diver will be required to carry
out a stop (level stop) sooner than a diver using level L0.
If a diver ignores a required level stop, TEC will change over to a lower MB level and the dive can not be
completed with the initially chosen MB level. E.g. If a diver sets level L4 on TEC prior to the dive and during the
dive ignores the stops recommended TEC will automatically adjust the setting to level L3 or lower.

1 Comparison of dives with MB level L0 and MB level L5


When two Smart TECs are used simultaneously, one unit is set for example to MB level L5, the other to
L0, the no-stop time will be shortened and level stops will be required before the diver has the obligation
of a decompression stop. These additional level stops help dissipate the microbubbles.

Microbubble accumulation
at the end of dive
Depth

Time
L0 3m/10ft

6m/20ft
L5
9m/30ft

12m/40ft

15m/50ft
Stop depth

ft
Microbubble

L0 DECO STOP
Decompression values
NO STOP NO STOP NO STOP m
level

L5 DECO

NO STOP
LEVELSTOP
m
LEVELSTOP
m
LEVELSTOP
m Level stop values
ft ft ft

Decompression obligation, ->31

28 V UWATEC® Smart dive computers


2 Terminology V
This chapter will exclusively deal with terminology and display features used while diving with MB levels.
All other features are described in chapter III (page 15).

2.1 Display during microbubble (MB) no-stop phase

English
MB no-stop phase Level stop phase
MB level Dive phase during which surfacing is
activated possible without MB level stop.
Level L2 is active

Time
Depth

CNS O2%

NO STOP
Dive time 21min MB no-stop time
%

MB no-stop time Current depth


Remaining time at a
respective depth allowing
ascent without level stop.

35.7 m Maximum depth


117 ft

2.2 Display during level stop phase

Total ascent time


including level stops
Total ascent time (14 min)

Time
Depth

3m/10ft
CNS O2% 6m/20ft
9m/30ft Lowest level
LEVELSTOP stop depth
% 3 min
Level stop duration
Level stop depth
The deepest level
stop is displayed.
Level stop duration
The duration of a level 35.7m
stop at a given level 117ft MB no-stop phase Level stop phase
stop depth is displayed. To complete the dive without being
reduced to a lower MB level all
requested level stops must be
observed.

V UWATEC® Smart dive computers


29
3 Preparation for a dive with microbubble levels (MB levels)
3.1 Setting the MB level

To change the MB level TEC must be in user mode.


1. Bridge contacts B and + until the symbol for MB levels appears.

2. Confirm that you wish to change the displayed MB level by bridging B and E.

+ – 3. Change MB level by bridging contacts B and + or B and – .

4. Confirm with B and E the selected MB level.


Without confirmation the display will disappear after 3 minutes and your
entries will not be accepted.
TEC will display the symbol to confirm that an MB level beyond L0 (L1-L5)
has been chosen. During the dive the MB level is shown by pressing briefly
. If however a level stop is ignored, the new MB level is permanently
shown (->31).
MB levels have an influence on the dive planner.

4 Functions during the dive with microbubble levels


4.1 Level stop information
Microbubble (MB) no-stop time
While diving with MB levels L1 to L5 TEC will display the MB no-stop time instead of the ordinary no-stop
time. Within the MB no-stop time no level stops are required.
The arrow NO STOP and the MB level symbol are visible. The remaining MB no-stop time is shown
in minutes.

• Information and alarms for MB no-stop time and ordinary no-


stop time are the same (->22).
• Regardless of the MB level, we generally recommend to perform
a slow ascent during the last few metres / feet.

NO STOP

MB no-stop time
Level stop On entering the level stop phase, the arrow NO STOP disappears and the
arrow LEVELSTOP appears. The LEVELSTOP arrow flashes for 8 seconds and an
Level stop icon audible alarm goes off. To complete the dive without being reduced to a lower
MB level, all requested level stops must be observed.
To the right of the LEVELSTOP arrow, the deepest level stop is displayed in met-
res/feet. The display <3m 2:> (<10ft 2:>) means that a level stop of 2 minutes
CNS O2%
at a depth of 3 metres (10ft) has to be observed.
When a level stop obligation is finished, the next higher level stop – if present
– is displayed. When all level stops have been observed, the arrow LEVELSTOP
LEVELSTOP extinguishes and the arrow NO STOP reappears. The indication of time shows
the MB no-stop time again.
Deepest level
stop depth
Level stop duration
Total ascent time

30 V UWATEC® Smart dive computers


4 Functions during the dive with microbubble levels V
4 sec The attention message "Level stop ignored" is
WARNING activated if the requested level stop is not observed.
An attention beep* goes off, the arrow LEVELSTOP ,
CNS O2%

the depth and duration of the ignored level stop


begin flashing.

English
LEVELSTOP
To complete the dive without being reduced to a
lower MB level, you must descend to the prescribed
depth immediately!
Level stop ignored
The warning "microbubble level reduced" is
Microbubble level reduced WARNING activated if the diver ascends more than 1.5m (5ft)
above the required level stop. TEC reduces the
4 sec microbubble level, an attention beep* goes off and
for the rest of the dive the new MB level is indica-
CNS O2% ted in the position of the maximum depth. The new
MB level stop is displayed.
LEVELSTOP
To complete the dive without being further reduced
to an even lower MB level the new level stop must
%
be observed.

New microbubble level


* Attention beeps can be suppressed via SmartTRAK.

4.2 Total time of ascent


TEC displays the level stop information and the total time of ascent. This inclu-
des the time of ascent as well as all level stops.
CNS O2%

The total time of ascent is calculated on the basis of the prescribed


LEVELSTOP
ascent rate and a normal workload. Total time of ascent can be sub-
ject to change if the ascent rate is not ideal (100%) or if TEC detects
a higher workload.
Total ascent time

4.3 Decompression obligation

Decompression obligation Avoid decompression dives when diving with MB


CNS O2%
WARNING levels.

DECO TEC calculates and displays level stops to reduce microbubble formation, but it
LEVELSTOP
also calculates the diver's decompression data. If decompression stops become
obligatory, the DECO symbol will be displayed. The total ascent time will now
Level stop information also contain a decompression stop.

You are close to entering decompression: At the


4 sec
WARNING beginning of a decompression phase an attention
beep goes off and the DECO symbol flashes for 8
CNS O2%
seconds.
DECO In order to prevent a dive with long decompression
LEVELSTOP

stops it is recommended that you ascend a few met-


%
res/feet on seeing this message.

V UWATEC® Smart dive computers


31
4 Functions during the dive with microbubble levels
4.4 Level stop and deco stop
When the level stop depth equals the
Time
depth of the first obligatory decompres-
Depth

L0 L5 sion stop and if you are within


1.5m/5feet of the stop depth itself, TEC
3m/10ft
Deco stop shows DECOSTOP and LEVELSTOP . The
duration indicated duration refers to level stop
duration.
Level stop duration
Since level stops are more restrictive
than decompression stops, when all
Microbubble

L0 decompression obligations have been


observed the display changes from
DECOSTOP LEVELSTOP to LEVELSTOP only.
level

L5
LEVELSTOP
LEVELSTOP

ft ft

decompression stop
depth = level stop depth Level stop duration

5 Complete a dive with MB levels

A dive with MB levels is completed the same way as a dive without MB levels
(L0) (-> 24), save for the following exceptions:

If the MB level has been reduced during the dive, TEC will display a flashing MB
level symbol and the current MB level for five minutes after reaching the surfa-
ce. The dive is then completed and TEC changes to user mode with the MB
level switching back to the original MB setting.

Repetitive dives and microbubble levels: If during a dive a level stop is being
ignored and the diver starts another descent shortly afterwards, TEC might
immediately request level stops. To complete the dive with the initially set MB
level all level stops must be observed.

VI Diving with more than one gas mixture VI


The following chapter deals with the characteristics of diving with more than one gas mixture.

Smart TEC enables you to use up to three different gas mixtures during the same dive. To measure the tank
pressure, each tank requires an individual transmitter. Tank 1 contains the bottom mix, tank 2 contains the
travel mix and tank D contains the deco mix. For dives with a single gas mixture you must use tank 1, for
dives with 2 gas mixtures you must use tanks 1 and D.
Time Time
Depth
Depth

Bottom mix Travel mix Deco mix Bottom mix Deco mix

Diving with 3 gas mixtures Diving with 2 gas mixtures

32 V UWATEC® Smart dive computers


VI Diving with more than one gas mixture VI
Preparation for a dive with several gas mixtures
Before the dive, all transmitters in use must be paired with TEC, the fraction of oxygen must be entered for
each tank and the depth for changing the gas mixture must be set for tanks 2 and D.

Setting the gas mixture and the depth for changing the gas mixture

English
During dives with more than one gas mixture the bottom mix
Range of O2 fraction
(tank 1) contains the lowest fraction of oxygen, while the deco
O2 fraction mix (tank D) contains the highest fraction. The additional travel
100% mix (tank 2) for dives with 3 gas mixtures holds a fraction of

D D oxygen that is between that of tank 1 and tank D. TEC will only
accept settings corresponding with this order. Note however
that you can also set 2 or even 3 tanks to the same gas mixture.
21% TEC will only accept gas mixture switching depths (MOD) as
input where the maximum partial pressure of oxygen (ppO2), as
3 gas mixtures 2 gas mixtures pre-set manually (-> 18 point 6) or by means of SmartTRAK, is
not exceeded.

For gas mixtures having an oxygen percentage of 80% or greater the ppO2 is
WARNING fixed at 1.6bar and cannot be altered in any way.

Procedure:
1. Enter the setting for the fraction of oxygen and the MOD for tank 1 (bot-
tom mix) according to the instructions on page 18 (points 1-7).
2. Enter the setting for the fraction of oxygen for tank D (deco mix) accor-
ding to the instructions on page 18 (points 1-5).
3. By bridging B and + or B and – enter the desired depth at which you will
change to tank D during the ascent phase. Confirm your settings with B and E.
The depth you have selected is the Maximum Operating Depth
(MOD) for this gas mixture and the indicated ppO2 becomes the
maximum tolerable partial pressure of oxygen (ppO2 max). During
the ascent an audible attention message will indicate that you
have reached the depth which requires the change of your gas
mixture. This applies to tanks D and 2 (see points below).

4. For dives with three gas mixtures enter the setting for the fraction of
oxygen for tank 2 (travel mix) according to the instructions on page 18
(points 1-5).
Note that tank symbol 2 can only be selected after tank symbol
D has been assigned to a gas mixture.
%
5. Now enter the maximum depth for the use of tank 2 by bridging B and
+ or B and -. Confirm your settings with B and E.
Depth for ppO2max The information given under point 3. does apply here also.
change of Without confirmation with B and E the display will disappear
gas after 3 minutes and your entries will not be accepted.
mixture, + –
MOD The time to reset the O2% mix to air can be set with SmartTRAK between 1 hour
and 48 hours or to "no reset" (default).
After the reset, the oxygen fraction of tank 1 is set to 21%, tanks 2 and D are
switched off (– – %O2).

VI UWATEC® Smart dive computers


33
VI Diving with more than one gas mixture
Switching off gas mixtures
Tanks which are not going to be used during the next dive (even though their
fraction of oxygen has been defined) must be "switched off". This is achieved
by switching off the appropriate gas mixture at TEC.
Gas mixtures which have been switched off cannot be selected during the dive
and TEC will ignore their transmitters.
Tank 1 cannot be switched off. If tank D (deco mix) is switched off, tank 2 will
automatically be switched off as well.
%
Procedure:
Mixture 2 1. Select the tank symbol of the gas mixture you wish to switch off and the set-
"switched off" ting of the O2 mix, according to the instructions on page 18 (points 1-3).
2. Reduce the fraction of oxygen with B and –, until "– – %O2" appears.
3. Confirm your settings with B and E.

The pairing of transmitter and tank symbol will remain valid even if the gas mix-
ture has been shut off.
The selected depth for the change of gas mixture (MOD) is deleted once the
gas mixture is switched off.

Functions during a dive with several gas mixtures


Diving with more than one gas mixture represents a much higher risk (than
WARNING diving with a single mixture), and mistakes by the diver may lead to serious
injury or death.
During dives with several gas mixtures, always make sure you are breathing
from the tank you intended to breath from.
Mark all your tanks and regulators, so they can under no circumstances be
mixed up! Before every dive and after changing a tank, make sure that each
gas mixture is set to the correct value for the corresponding tank.

Decompression prognosis
The calculation of decompression data (e.g. the total time of ascent) is based on the assumption that
changes of gas mixture will be performed at the previously selected depths (MOD gas mixture 2, MOD gas
mixture D). If a diver ignores a required change or changes the mixture belatedly, TEC will readjust the
decompression calculation accordingly. The dive computer will then base its calculations on the assumption
that the diver will either use the indicated gas mixture until he/she reaches the depth for the next change
or – if no further changes are planned – that he/she will ascend to the surface with the current mixture.

Remaining Bottom Time


In addition to calculating the decompression prognosis for all the activated tanks, in the background TEC
also calculates the decompression prognosis for tank 1 by itself, and (when diving with three tanks) for
tanks 1+2 only and for tanks 1+D only. The Remaining Bottom Time (RBT) while breathing from tank 1
considers that you will finish the dive with tank 1, using the decompression prognosis for tank 1 alone.
The RBT while breathing from tank 2 considers that you will finish the dive with tank 2 using the decom-
pression prognosis for tanks 1 and 2 only (see -> 35, 36). This allows, when starting the ascent before
RBT=0, to always be able to finish a dive if by accident you were to lose the travel and/or deco bottles,
find that they are empty or in case of regulator malfunction. Note that for any tank (1, 2 or D) when you
are shallower than 6m/20ft the RBT indicates the duration of the gas supply (minus the reserve) at the cur-
rent depth.
Since TEC needs to evaluate your breathing pattern and the corresponding
pressure drop in the tank, it takes about 2 minutes after you start breathing
from a tank before TEC shows the RBT.

34 VI UWATEC® Smart dive computers


VI Diving with more than one gas mixture VI
Changing the gas mixture
After immersion, TEC automatically selects gas mixture 1.
4 sec DECO STOP If during the ascent a diver reaches the depth
WARNING requiring a change of gas mixture (MOD gas mix-
% ture D or 2), the imminent change is brought to

English
the diver’s attention by an audible attention signal
and the flashing of the tank symbol of the new
oxygen fraction of tank 1 gas mixture, both lasting for 30 seconds.

DECO STOP
LEVELSTOP
NO STOP Procedure: • switch to the regulator with the new gas mixture and start brea-
% thing
• confirm the change by pressing and holding (long beep)
within 30 seconds. The tank symbol will stop flashing.
change of gas mixture
confirmed
No change of gas If a diver fails to confirm the change of gas mixture, TEC will continue to cal-
mixture: culate with the previous gas mixture and adapt the decompression calculation
accordingly.
Belated change of A diver can catch up on a required change of gas mixture until he reaches the
gas mixture: surface. By quickly pressing the tank symbol of the new gas mixture will
start flashing.
Switch to the regulator with the new gas mixture and start breathing. Confirm
the change by pressing and holding . The tank symbol will stop flashing
and the decompression calculation will readjust accordingly.
Submerging again If after a change of gas mixture the Maximum Operating Depth (MOD) of the
after a change current mixture is exceeded, the ppO2 max warning will appear. ->20
of gas mixture: Change to the gas mixture suited for this depth and select the appropriate tank
symbol or rise to the Maximum Operating Depth of the current gas mixture.
Failure to do so can result in oxygen poisoning.

Examples for clarification of RBT and change of gas mixtures


The following figures show a dive profile and the display of various TECs in the following settings:
TEC-1: set to 21% (tank 1) TEC-3: set to 21% (tank 1) and 100% (tank D)
TEC-2: set to 21% (tank 1) and 50% (tank D) TEC-4: set to 21% (tank 1), 50% (tank 2) and
100% (tank D)
1. Start of ascent: the TECs have different decompression prognosis, indicating the benefit of the travel
and/or decompression gases. However, the RBT is the same for all four, as it relates to tank 1 only. It
indicates that starting the ascent within the next three minutes the diver will have enough gas for the
ascent and the decompression obligations shown on TEC-1. At RBT=3 or less, the ascent arrow is sho-
wing: TEC suggests that the diver starts the ascent.
2. Slightly below MOD for 50% mix (MOD=22m): The slow ascent has caused significant build-up of
decompression obligations. There is a marked difference in ascent times, indicating the benefit of
decompressing at high oxygen percentage.
3. Above MOD for 50%, missed switch: TEC-2 adjusts its decompression prognosis to be same as TEC-
1. TEC-4 adjusts its decompression prognosis to be the same as TEC-3.
4. Again below MOD for 50%: TEC-2 and TEC-4 once again show credit for the presence of the 50%
mix in their calculations.
5. Above MOD for 50%, switch carried out: TEC-2 and TEC-4 show the tank pressure in the 50% tank.
The RBT agrees because also in TEC-4 it considers the 50% tank will be used to finish the dive. TEC-1
and TEC-3 are not receiving any signals from the transmitter on the 21% tank and show “- - -“.
6. Slightly below the MOD for 100% mix (MOD=6m): During the slow ascent from 22m, TEC-2 using
50% oxygen has caught up on the decompression schedule of TEC-3, which is still using 21% oxygen
but gives credit for the upcoming switch to 100% oxygen.

VI UWATEC® Smart dive computers


35
VI Diving with more than one gas mixture
7. Above MOD for 100%, missed switch: now TEC-3 gives the same decompression prognosis as TEC-
1. TEC-4 gives the same decompression prognosis as TEC-2.
8. Again below the MOD for 100%: TEC-3 and TEC-4 again show credit for the presence of the 100%
mix in their calculations
9. Above MOD for 100%, switch carried out: TEC-3 and TEC-4 show the tank pressure in the 100%
tank. The RBT agrees. TEC-2 is not receiving anymore signals from the 50% tank and shows “- - -“.
However, the slow ascent and delayed switch to 100% has caused the decompression prognosis to be
shorter on TEC-2 than TEC-3.

TEC-1 CNS O2% CNS O2% CNS O2%

21% tank 1 DECO STOP DECO STOP DECO STOP DECO STOP

% % % %

TEC-2 CNS O2% CNS O2% CNS O2%

21% tank 1 DECO STOP DECO STOP DECO STOP DECO STOP

50% tank D % % % %

TEC-3 CNS O2% CNS O2% CNS O2%

21% tank 1 DECO STOP DECO STOP DECO STOP DECO STOP

100% tank D
% % % %

TEC-4 CNS O2% CNS O2% CNS O2%

21% tank 1
DECO STOP DECO STOP DECO STOP DECO STOP

50% tank 2
100% tank D % % % %

1 2 3 4
0
Depth (m)

10
20
30
40
50
60
0 Time (min) 10 5 20 6 30 7 8 40 9 50

TEC-1 CNS O2% CNS O2% CNS O2%

21% tank 1 DECO STOP DECO STOP DECO STOP DECO STOP DECO STOP

% % % % %

TEC-2 CNS O2% CNS O2% CNS O2%

21% tank 1 DECO STOP DECO STOP DECO STOP DECO STOP DECO STOP

50% tank D % % % % %

TEC-3 CNS O2% CNS O2% CNS O2%

21% tank 1 DECO STOP DECO STOP DECO STOP DECO STOP DECO STOP

100% tank D
% % % % %

TEC-4 CNS O2% CNS O2% CNS O2%

21% tank 1
50% tank 2
DECO STOP DECO STOP DECO STOP DECO STOP DECO STOP

100% tank D % % % % %

36 VI UWATEC® Smart dive computers


VII Dive planner VII
TEC has a dive planner which allows the planning of no-stop dives as well as decompression dives with
freely determinable surface intervals and with up to three different gas mixtures.
Basis of the planning: • selected fraction of oxygen and MOD of all active gas mixtures
• selected microbubble level
• water temperature of the most recent dive

English
• altitude range (if any)
• status of saturation at the time the dive planner is selected
• assuming a normal workload of the diver and observance of the prescribed
ascent rates
• assumption: changes of gas mixture are performed at the selected MOD of
mixtures 2 and D.
If two or more divers using computers are planning a dive, planning for all
WARNING divers has to be based on the dive computer showing the shortest no-stop
times. Failure to do this may lead to serious injury or death from decompres-
sion sickness.
Before using the dive planner, the fraction of oxygen must be entered for each required tank and the
depths for a change of gas mixture for tanks 2 and D must be entered. Tanks, i.e. gas mixtures which are
not going to be used during the next dive must be switched off (->33-34). All 'activated' gas mixtures will
be indicated by their tank symbols and allowed for calculation.

1 Planning a no-stop dive


With the contacts B and – you can select the dive planner at the surface. (The
dive planner cannot be selected in gauge mode)
NO
The microbubble warning and
its duration are displayed if
NO
Microbubble warning TEC detects an increased risk
(Do Not Dive) due to the accumulation of
if NO

Duration of the microbubbles.


warning
Dive planner icon Enter the dive planner with B and E.
Enter the
dive planner The input window for the time interval is displayed
if there was a remaining desaturation (DESAT) befo-
re the Dive Planner has been selected. This surface
if rep. dive interval between now and the beginning of the dive
can be changed with the contacts + and – in steps
of 15 minutes.
NO

If a microbubble warning (no dive) and its dura-


Input of the + tion has been displayed, TEC proposes this time –
surface interval – rounded up to the next 15 minutes – as surface
Icon for interval. If the proposed interval is shortened, the
microbubble microbubble warning appears.
level 1-5 With B and E you confirm the displayed interval (if applicable), then TEC starts
scrolling the no-stop times. The no-stop times are displayed in 3 metre incre-
CNS O2%
ments (10 ft) and are displayed for every increment for about 2 seconds. The
process starts at 3 metres (10 ft).
If a microbubble level has been selected (L1-5), the MB no-stop time is
NO STOP
shown.
%
No-stop times will be displayed as long as the selected maximum operating
No-stop time or depth (MOD) of tank 1 is not exceeded.
MB no-stop time NO
On page 24 you will find further information and safety considerations
regarding the microbubble warning.

VI UWATEC® Smart dive computers


37
VII Dive planner
2 Planning a decompression dive
1. Activate the dive planner for a no-stop dive ->37.
NO STOP 2. Wait until the desired depth appears, then switch into decompression plan-
% ning by bridging contacts B and E. TEC shows the bottom time (no-stop time
+ 1 minute) and the appropriate decompression information or level stop
data respectively.
3. <Add> asks that you set the bottom time. This is done with contacts B and
+, B and – respectively. As soon as the contacts are no longer bridged, TEC
CNS O2%
calculates the decompression information or level stop data respectively for
DECO STOP
this set bottom time.
If you wish to plan a decompression dive at another depth, switch from decom-
%
pression planning to no-stop planning by means of B and E. TEC again shows
Bottom Time + the scrolling no-stop times. Now you can switch between no-stop planning and

decompression planning at will with contacts B and E.

If the calculated decompression information and the total ascent time exceed
199 minutes, or the CNS O2 % value exceeds 199%, the said values will start
flashing on the display or <– –> values appear and the decompression calcula-
tion is suspended until the bottom time has been reduced accordingly. CNS
O2% values higher than 199% will be displayed as 199 %.

3 Leaving the dive planner


With the contacts B and E (1-2 sec) you can exit the Dive Planner. This also
occurs after three minutes without operation.

38 VI UWATEC® Smart dive computers


VIII Logbook VIII
1 Survey
TEC records the profiles of about 100 hours of diving. This information can be transferred to a PC with the
standard infrared interface (IrDA) and the Windows® software SmartTRAK. All dives in the memory can be
displayed directly on the dive computer. Up to 99 dives can be displayed directly on the dive computer.
A dive is only entered in the logbook if the dive time is longer than 2 minutes. The following information

English
of the dive is displayed:

Logbook icon Diving in gauge mode


Ignored decompression stop* or MB level reduced*
diving in SOS mode
Dive time
CNS O2%

Too fast ascent* Dive number


Altitude range (if any) DECO
CNS value at the end of the dive
Duration of surface interval (only with repetitive dives)
Maximum depth / Lowest temp. / MB level
%
Increased workload warning*
(at the beginning of the dive) dP = Differential pressure (gas consumed)
Gas consumed (bar or psi) from tank 1 (displays only those gas
Fraction of oxygen tank 1 " tank 2 mixtures used during the dive)
" tank 2 " tank D
" tank D Ignored decompression stop*
*Alarms during the dive

If a dive is started within adaptation time (after a change of altitude), the


adaptation time is displayed instead of the surface interval.

2 Operation
With the contacts B and + or B and – you can select the logbook. With B and
E you enter the logbook.

If there was a remaining desaturation time (DESAT) before selecting the log-
book, the time since the last dive (surface interval) is displayed.

With B and E you get the most recent dive displayed (LOG 1).

Each bridging of B and + or B and – causes a jump to the next older or more
CNS O2% recent dive. Upon continuous bridging of the contacts all dives are displayed
successively.

With the contacts B and E you can exit the logbook.


The logbook closes automatically after 3 minutes without operation.

VIII UWATEC® Smart dive computers


39
IX Appendix
1 Technical information
Operating altitude: with decompression information: sea level up to approx. 4000 m (13120ft);
w/o decompression, w/o RBT information: usable in gauge mode (at any altitude)
Max. displayed depth: 120m (394 ft), resolution between 0.8 m and 99.9 m: 0.1 m, >99.9 m: 1m
The resolution in feet is always 1 foot
• Do not dive deeper than the limits given by the chosen fraction of oxygen
WARNING (decompression sickness, oxygen toxicity).
• Never dive deeper than your training qualification (experience) allows you.
• Always observe local dive depth restrictions.
Decompression calculation depth range: 0.8 to 120m (3 to 394 ft)
Maximum environment pressure: 13 bar (189 psi)
Clock: Quartz timer, display up to 999 minutes.
O2% Mix: Adjustable between 21%O2 (compressed air) and 100% O2
Operating temperature: -10° to +50°C (14°F to 122°F).
Power supply: Special battery UWATEC LR07
Life of the battery: 500 to 800 dives, depending on the quantity of dives per year and the use of
the backlight.
Transmitter: High pressure connection: Maximum working pressure: 300 bar (4350 psi)
Life of the battery: up to 1000 dives, max. 3 years without use.
Power supply: User replaceable battery CR2450
2 Maintenance
The tank pressure gauge and the parts of this product used to measure the tank pressure should be serviced
by an authorized SCUBAPRO UWATEC dealer every second year or after 200 dives (whichever comes first).
Aside from that your TEC is virtually maintenance free. All you need to do is to rinse it carefully with fresh
water after each use and to have the batteries changed when needed. To avoid possible problems with your
TEC, the following recommendations will help assure that it will give you years of trouble free service:
• Avoid dropping or jarring your TEC.
WARNING • Do not allow your TEC to be exposed to direct, intense sunlight.
• Rinse your TEC thoroughly with fresh water after each dive.
• Do not store your TEC in a sealed container; make sure there is free ventilation.
• If there are problems with operating the contacts, use soapy water to clean
TEC and dry it thoroughly. The surface of your TEC housing can be treated with
silicone grease. Do not apply grease to the water contacts!
• Do not clean TEC with liquids containing solvent (apart from water).
• If the service icon appears, TEC must not be used for any further dives. Take
your dive computer to an authorized SCUBAPRO UWATEC dealer.
Take the dive computer to an authorised SCUBAPRO UWATEC dealer in order
WARNING to change the batteries. The battery replacement is carried out by a SCUBA-
PRO UWATEC subsidiary. TEC is checked for its technical integrity at the same
time. Do not attempt to have the batteries changed by anyone other than an
authorised SCUBAPRO UWATEC dealer.

2.1 Replacing the battery of the transmitter


It is recommended to have the battery of the transmitter replaced by an
WARNING authorized SCUBAPRO UWATEC dealer. The change must be made with par-
ticular care in order to prevent water from seeping in.
The warranty does not cover damages due to an improper replacement of the
battery.

Transmitter battery set (PN 06.201.920): Includes a Type CR 2450 battery and a 26.00 x 2.00 mm trans-
mitter cap o-ring .

40 XI UWATEC® Smart dive computers


IX Appendix IX

Never touch the metal surface of the battery with bare fingers.
WARNING The two battery poles must never be short circuited.

Transmitter Battery Transmitter Cap

English
CR2450

HP port o-ring
8.73 x 1.78 mm

Transmitter screws
PT®KA 30 x 16 mm

Transmitter cap o-ring


26.00 x 2.00 mm
Procedure:
To replace the battery you need a Phillips screwdriver and a clean cloth.
• A leaking transmitter cap may lead to the destruction of the transmitter by
WARNING water seeping in or cause the transmitter to switch off without prior notice.
• Always open the transmitter in a dry and clean environment.
• Only open the transmitter to replace the battery.

1. Remove the transmitter from the HP outlet of the first stage of the regulator.
2. Dry the transmitter with a soft towel.
3. Remove the 3 screws with the Phillips screwdriver.
4. Remove the transmitter cap carefully.
5. Remove the transmitter cap o-ring carefully. Do not damage the sealing surfaces.
6. Remove the battery by holding it on both sides. Do not touch the contacts or electronic
parts.

Protect the environment and dispose the battery properly.

If you notice traces of seeping water, damages, or other defects on the o-ring,
WARNING do not use the transmitter for further dives. Take it to an authorized SCUBA-
PRO UWATEC dealer for check and repair.

7. Always insert a new o-ring when you replace the battery and dispose the old o-ring.
Make sure that the new, lubricated o-ring is in perfect condition, and that o-ring, o-ring groove and
the sealing surfaces of HP outlet and transmitter cap are free of dust and dirt.
If necessary, clean the parts with a soft cloth. Fit the o-ring in the o-ring groove.

8. Check the proper polarity of the battery. The transmitter can be damaged,
WARNING if you do not insert the battery correctly.

Wait for at least 30 seconds. Now insert the new battery, with "+" pointing upwards, into the battery
compartment.
9. After battery replacement the transmitter will perform an automatic test and switch into ready mode
after 60 seconds.

IX UWATEC® Smart dive computers


41
IX Appendix

10. The transmitter cap will only fit in one position. Check the proper position
WARNING of the guide slots on the electronic support and in the transmitter cap.

Slide the transmitter cap carefully back into its proper position.
11. Do not overtighten the screws! Fasten the transmitter cap with the 3 screws.
12. Mount the transmitter on the HP outlet of the first stage of the regulator and check transmission and
pairing. If you do not receive valid tank pressure data, transmitter and dive computer must be paired
again.

3 Conversion of tank pressure


Tank pressure indicated may differ from the information given by a manometer/pressure gauge. TEC dis-
plays pressure always converted to a temperature of 20°C / 68°F, whereas the mechanical pressure gauge
displays the current pressure influenced by temperature.

The figure below allows you to compare the information given by a conventional pressure gauge and by
TEC at six different temperatures.

bar psi
260 3770
35o C 95o F
240 3480 30o C 86o F
20o C 68o F
Display pressure gauge

220 3190 10o C 50o F


5o C 41o F
0o C 32o F
200 2900

180 2610

160 2320

140 2030

120 1740
150 160 170 180 190 200 210 220 230 bar
2175 2320 2465 2610 2755 2900 3045 3190 3335 psi
Display Smart TEC

4 Warranty
The warranty only covers dive computers which have been bought from an authorised SCUBAPRO
UWATEC retailer.
The warranty is given for a period of 2 years.
Repairs or replacements during the warranty period do not increase the warranty period.
In order to put forward a warranty claim: send the dive computer together with a dated receipt of the
purchase to your authorised retailer or an authorised servicing point.

UWATEC reserves the right to determine the merits of a warranty claim and to determine whether the
computer will be repaired or replaced.

42 IX UWATEC® Smart dive computers


IX Appendix IX
Excluded are faults or defects due to:
• excessive wear and tear;
• exterior influences, e.g. transport damage, damage due to bumping and hitting, influences of weather
or other natural phenomena;
• servicing, repairs or the opening of the dive computer by anybody not authorised by the manufactu-

English
rer. This especially concerns the change of battery;
• pressure tests which do not take place in water;
• diving accidents;
• improper placement of the transmitter cap.

5 FCC
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.

6 Index
Active backlight___________________________11 No-stop time ______________________15, 22, 28
Ascent rate________________________15, 17, 19 O2% mix, Set up… ________________________18
Battery alarm _____________________________17 O2 fraction ________________________15, 16, 18
Battery capacity, Checking the… ____________10 O2 partial pressure__________________16, 17, 20
Battery lifetime ___________________________40 O2 partial pressure, maximum ___________16, 20
Beep, Switch off the… _____________________17 O2 toxicity __________________15, 16, 17, 21, 39
Changing the gas mixture ______________26, 34 Operating TEC ________________________4, 8, 9
CNS O2 ________________2, 3, 15, 16, 17, 21, 39 Oxygen…, see "O2…"
Deco data during decompression phase ______15 PC, transfer to PC (logbook) ________________39
Deco data during no-stop phase ____________15 ppO2, see O2 partial pressure
Decompression stop, Ignored… __________17, 23 Premix, setting the gas mixture______________18
Depth, current ____________________________19 RBT ______________________________15, 17, 22
Desaturation time _________________________24 Safety stop timer__________________________23
Dive ____________________________________15 Setting the MOD ______________________18, 33
Dive planner _____________________________37 SmartTRAK ___________________8, 9, 16, 17, 39
Dive time ________________________________19 SOS mode _______________________________11
Dive, end of a dive ________________________24 Stopwatch _______________________________27
Fly, "no-fly time" ______________________10, 24 Surface interval________________________37, 39
Gas mixture, Setting… _____________________18 Switching off gas mixtures__________________34
Gauge mode _____________________________26 System ___________________________________8
Light ____________________________________11 Tank pressure ____________________________21
Logbook _________________________________39 Technical information ______________________40
Maintenance _____________________________40 Transmitter _________________________8, 12, 40
Maximum depth_______________________19, 39 Warnings ________________________________17
Microbubbles_____________________________28 Workload _______________________________21
Microbubbles, Warning of… ____________24, 37 Workload, Increased… ____________________ 21
Mountain lakes, Diving in… ____________25, 39
Mounting the transmitter to the first stage ___12
Nitrox ______________________________16

iX UWATEC® Smart dive computers


43
SCUBAPRO UWATEC Americas SCUBAPRO UWATEC France
(USA/Canada/Latin America) Les Terriers Nord
1166 Fesler Street 175 Allée Belle Vue
El Cajon, CA 92020 USA F-06600 Antibes
t: +1 619 402 1023 t: +33 (0) 4 92 91 30 30
f: +1 619 402 1554 f: +33 (0) 4 92 91 30 31
www.scubapro.com www.scubapro-uwatec.com

SCUBAPRO UWATEC Asia Pacific SCUBAPRO UWATEC Benelux


1208 Block A, MP Industrial Center Avenue des Arts, 10/11 Bte 13
18 Ka Yip St. 1210 Bruxelles
Chai Wan Hong Kong t: +32 (0) 2 250 37 10
t: +852 2556 7338 f: +32 (0) 2 250 37 11
f: +852 2898 9872 www.scubapro-uwatec.com
www.scubaproasiapacific.com
SCUBAPRO UWATEC España
SCUBAPRO UWATEC Australia Pere IV, nº359, 2º
Unit 21, 380 Eastern Valley Way 08020- Barcelona
Chatswood, N.S.W. 2067 t: +34 93 303 55 50
t: +61 2 9417 1011 f: +34 93 266 45 05
f: +61 2 9417 1044 www.scubapro-uwatec.com
www.scubapro.com.au
SCUBAPRO UWATEC U.K.
SCUBAPRO UWATEC Deutschland Vickers Business Centre
(Germany / Austria / Scandinavia) Priestley Road
Taucherausrüstungen GmbH Basingstoke, Hampshire RG24 9NP
Rheinvogtstraße 17 England
79713 Bad Säckingen-Wallbach t: +44 0 1256 812 636
t: +49 (0) 7761 921050 f: +44 0 1256 812 646
f: +49 (0) 7761 921051 www.scubapro.co.uk
www.scubapro.de
SCUBAPRO UWATEC Switzerland
SCUBAPRO UWATEC Italy Oberwilerstrasse 16
Via G.Latiro 45 CH-8444 Henggart
I-16039 Sestri Levante (GE) t: +41 (0) 52 3 16 27 21
t: +39 0185 482 321 f: +41 (0) 52 3 16 28 67
f: +39 0185 459 122 www.scubapro.de
www.scubapro.it

SCUBAPRO UWATEC Japan


4-2 Marina Plaza 5F
Kanazawa-Ku
Yokohama
Japan
t: +81 45 775 2288
f: +81 45 775 4420
www.scubapro.co.jp

www.uwatec.com

9.04

UWATEC® Smart dive computers

You might also like