IO72.00 Overlaod Device Manufacturer Calibration
IO72.00 Overlaod Device Manufacturer Calibration
IO72 Pag. 1 di 6
Introduction
Attention:
The calibration of the two channels using push-button S3-S4 is not more required (S3 and S4
are no more present on this board). The board performs automatically the calibration of the
second channel.
Attention:
Push-button 0 (zero) belonging to the previous procedures is no more present.
This board is provided with the following push-buttons: SW1, SW2, SW3.
Attention:
The new board is compatible with the previous ones so it is possible to install it on units with the
previous version of the overload board. During the replacement it is required to replace three
green connectors with the ones supplied inside the new board box.
IO72 OVERLOAD BOARD CALIBRATION PROCEDURE Pag. 2 di 6
1 User procedure: This procedure can be performed just if the payload value read by the
overload board is between +10% and -10% of the maximum payload around the original 0
(zero).
2 Manufacturer calibration: Access mode is different from the previous one, it is possible just
for the 10 seconds after the initial test is completed and following a precise order of keys.
Both procedures can be performed using push-buttons SW1, SW2,SW3 as for picture below:
IO72 OVERLOAD BOARD CALIBRATION PROCEDURE Pag. 3 di 6
Manufacturer calibration
1) To access the quick calibration, press the following combination of keys within 10 seconds
after start-up (during this phase the two lower display dashes are ON):
SW2 + SW3
SW1 + SW2
SW1 + SW3
During the calibration phases each time the push-button is pressed the horizontal display
dashes scroll to the top of the display.
2) If the sequence is performed correctly and during the first 10 seconds after start-up, the
calibration mode is accessed and the display shows a blinking “70”.
This message means “waiting for zero calibration”.
“Tare” calibration
3) Make sure that there’s no weight on the basket than press SW1 to confirm this
calibration.
When the minimum value is confirmed the display shows a fixed “70”.
“Pre-alarm” calibration
5) Make sure that the pre-alarm payload is put inside the basket (attention: the pre-alarm
signal is not used; the pre-alarm payload is the same as the alarm payload, please see
chart page 6 in order to find the test payload to use to perform the procedure), press
SW1 to confirm this calibration.
When the pre-alarm value is confirmed the display shows a fixed “7P”.
“Alarm” calibration
7) Make sure that the alarm payload is put inside the basket (please see chart page 6 in
order to find the test payload to use to perform the procedure), press SW1 to confirm this
calibration.
When the alarm value is confirmed the display shows a fixed “7A”.
After the procedure perform the payload tests on the basket base (left, right and central
part of the basket base) with maximum payload allowed, check that:
the overload indicator switches ON
the buzzer sounds
and that the unit stops.
IO72 OVERLOAD BOARD CALIBRATION PROCEDURE Pag. 5 di 6
Note
Push-button/keys:
Errors
In order to understand the error code check the following Leads flashings:
Flashings Legend
- No error
1 Sensor line error or channel 1
2 Sensor line error or channel 2
3 -
4 -
5 Signal value error (lower than the correct value)
Flashings Legend
- No error
1 Faulty board: replace the board
2 - wrong transducers wiring: check the electric connections
-faulty transducer: check the resistances inside the transducer
-faulty board: replace the board
3 Check the polarities of the connections to the Relay (connector J4)
4 -
5 CRC flashing error: replace the board
6 -
7 Difference between the sensors error. Check the sensor and eventually replace
the load sensor module. If the error stays, replace the board.