144 Watt DC Input Power Supply
USPA-DC 15 Isolated / Sequenced Outputs
Features:
R
C US H
TUV Rheinland
• Designed for Ultrasound Application
• Convection Cooled Package
• 15 Isolated Outputs
• Automatic Power Up Sequencing of all Outputs
• Measures 258 x 363 x 52mm / 10.15” x 14.29 “x 2.03”
• Works in conjunction with USPA-AC Power Supply
• Remote on/off Control
• Power Good / Fail Signal
Input Specifications: General Specifications:
Voltage 48VDC ±1% Input Range Operating Temp 5°C to 40°C
Current 10A at Input Range (max continuous) Storage Temp -40°C to +70°C
Start-Up Voltage 47VDC (minimum) Operating Humidity 10% to 90% Non Condensing
Shutdown Voltage 49VDC (maximum) Storage Humidity 10% to 95% Non Condensing
Cooling Convection
Size 10.15” x 14.29 “x 2.03” (L x W x H)
Output Specifications: Electromagnetic Compatibility:
Voltagte / current see output table Emissions EN55011 Level B (Conducted)
Minimum Load None Required Emmisions EN60601-1-2
Regulation ±2% maximum Emmissions EN50082-1
Ripple ±1% (20MHz) ESD Immunity EN61000-4-2
Over Current 115-135% (Auto-Recovery)
Over Voltage 105-115% (recycle input to reset)
Status & Control: Standards & Approvals:
/CPPSPOWEROFFOK Open Collector Output / Low = Power Off UL 2601-1
cppssw1+ Low = Power Key On CSA22.2 No.601.1
cppssleepok Open Collector Output / High = Power Sleep IEC60601-1
/pscppoweroff Open Collector Output / High = Power On EN60601-1
North America Asia
DongAh Elecomm USA DongAh Elecomm Corp. (Headquarters)
1300 E Lookout Dr, #210, 167 Namgok-ri, Yangji-myun,
Richardson TX 75082 Yongin-si Kyunggi-do 449-820 Korea
Phone: (972) 497-9990 Tel: 82-31-330-8805
Fax: (972) 497-9991 Fax: 82-31-330-8889
E-mail:
[email protected] E-mail:
[email protected] (Director)