Ca 620
Ca 620
03-11-24 3
4 03-11-24
03-11-24 5
1177en.gif
Operator training
Additonal operator training is not required under the following conditions:
1. The operator must have basic scills in working under laboratory conditions.
2. The operator must have basic scills in hematology.
3. The operator must be aware of IVD(EU) and FDA(USA) requirements re-
garding laboratory equipment for In Vitro Diagnostic purposes.
4. The operator must read and understand this manual.
1176en01 03-11-24 7
Standards
EN591:2001
IVD 98/79/EG
SSEN 61010-2-101 (Low Voltage 73/23/EEC)
EN 61326 (1997) with amendment EN 61326/A1 (1998) (EMC 89/336/EEC)
Standards harmonized with FDA/UL
Manual number
1504062en
Date of issue
August 2003
8 03-11-24 1176en01
1070en01 03-11-24 9
Electrical hazard
• Do not spill blood, reagent, drop metal objects such as wire staples or paper
clips into the instrument. This might cause a short circuit.
If this should occur, turn off the power immediately and contact your serv-
ice department.
Note: The cover should only be opened/removed by authorised service
personnel.
• Do not touch the electrical circuits inside the cover.
There is a hazard of electrical shock.
Note: The cover should only be opened/removed by authorised service
personnel.
Contamination hazard
• Always wear protective gloves and/or goggles when handling infectious or
potentially infectious materials.
• Handle samples with great care.
There is a risk of infection if contaminated blood splashes.
- If blood splashes, enter your eye or a cut, wash it off with plenty of water.
• Do not touch the waste liquid when discarding waste or disassembling/as-
sembling the related parts outside the instrument.
Risk of infection from contaminated blood.
-If you should touch the waste liquid inadvertently, wash off with disinfect-
ant first, then wash it off with soap.
10 03-11-24 1070en01
Crush hazard
• Always switch off the instrument when opening the front “door”. crush
hazard can occur if the option “Cap Piercing Device” is installed in the
equipment. See Signs on Equipment on page 12.
Note: The cover should only be opened/removed by authorised personnel
during operation.
Warning
Caution
Important
1070en01 03-11-24 11
1013.jpg
1022.jpg
1040.jpg
1. This label indicates that the safety instructions in the manual must be read
before opening the front door of the instrument.
2. This label indicates a potential bio hazard due to blood exposure or possible
contaminated waste.
3. This label indicates a potential crush hazard. Disconnect the instrument
from the mains outlet before opening the front door. Only authorized per-
sonell are allowed to open front door.
4. Serial Number, Voltage/Fuse specifications and CE / UL markings.
12 03-11-24 1070en01
1001en01 03-11-24 13
The instrument employs the electronic impedance principle for cell counting and
sizing, and a colorimetric method for measuring hemoglobin. A microprocessor
is used to measure the parameters and to size the cells. During the count the pro-
cessor is checking the analysing process for any irregularities. Size distributions
are printed for all populations (RBC, PLT and WBC).
The instrument has as standard a parallel and a serial output, which is user pro-
grammable. From a built-in selectable program the user can choose between dif-
ferent print formats.
A bar-code reader (scanner), available as an option, can be connected to enter the
sample ID automatically.
The instrument is fully automatic, which means that, the system is always pow-
ered on and performs automatically preforms check- and cleaning cycles to min-
imize user maintenance. Important
Applicable for CA620-
Note:
CellGuard:
PCT, LPCR, RDWa and PDWa are not established parameters. Their use should
be restricted to research or investigational use only. The MPA is intended to be
used for pre-donor screen-
2.3 Short-List of Specifications ing. Using the MPA for
PLT concentrate may re-
Measuring principle Impedance duce the accuracy of the
RBC, WBC, PLT result.
Measuring principle HGB Cyanide free method 540nm
Discriminator Floating programmable
Sampling system Closed shear valve
Parameters reported RBC, MCV, HCT, PLT, MPV HGB, MCH,
MCHC, WBC,
RDW%, LYMF abs, MID abs, GRAN
abs., LYMF%, MID%, GRAN%
RDW abs, PDW abs, LPCR, PCT
Size distributions printed for RBC, PLT and WBC diff.
Aspirated blood volume approx. 125µl
(open tubes)
Aspirated blood volume approx. 200 µl
(closed tubes)1
Blood volume using the 20 µl
Micro Pipette Adapter
Pre-diluted mode 1:200 to 1:250 using min. 20µl blood
e.g. 20 µl to 5 µl diluent
30 µl to 6 µl diluent
40 µl to 8 µl diluent
CA530 Screen 2x40 character LCD
CA620 Screen Graphic LCD display
Keyboard Numerical
Total cycle time approx. 73 seconds
14 03-11-24 1001en01
1001en01 03-11-24 15
WBC 0 - 99.9
RBC (CA620-Cellguard) 0 - 13.99
RBC (CA620/530) 0 - 13.99
MCV 15 - 250
PLT 0 - 6999
HGB 0 - 99.9
PLT Concentrate mode 0-6999
(applicable on CA620-Cellguard only)
Correlation1
16 03-11-24 1001en01
15
10
5
0
0 50 100 150 200
Number of Blood_samples/Day
1014.eps
Diluent Consumption
20
18
Lyzer consumption /
16
sample (ml)
14
12
10
8
6
4
2
0
0 50 100 150 200
Number of Blood_samples/Day
1015.eps
Lyzer consumption
1001en01 03-11-24 17
2.6 Limitations
Verification of any abnormal test result (including flagged results or results out-
side of the normal range) should be performed using reference methods or other
standard laboratory procedures for the conclusive verification of results. The sec-
tion below list known limitations of automated blood cell counters in general as
well as specific issues for cell counters using the impedance technology.
NRBC
Immature Nucleated Red Blood Cells will be counted in the WBC parameter. If
the number of the NRBC is sufficient to activate the DE alarm, such interference
will be detected. However, the manual differential blood cell count, performed
on a stained blood film, will reveal the presence of NRBCs.
Multiple myeloma
The precipitation of proteins in multiple myeloma patients may give elevated
WBC counts.
Hemolyzis
Hemolyzed specimen contains red cell stroma (debris), which may elevate the
WBC and/or PLT count.
Leukemias
This disease state may result in a spurious low WBC count because of the pos-
sible increased fragility of the leukocytes leading to some destruction of these
cells during counting. The cell fragments will also interfere with the white cell par-
tial differential parameters (LYMF, GRAN and MID). A spurious low WBC
count may also be seen in patients with lymphocytic leukemias due to the pres-
ence of abnormally small lymphocytes, which may not be counted by the instru-
ment.
Chemotherapy
Cytotoxic and immunosuppressive drugs may increase the fragility of the leuko-
cytes, which may cause low WBC counts.
Cryoglobulins
Increased levels of cryoglobin that may be associated with myeloma, carcinoma,
leukemias, macroglobulinemia, lymphoproliferative disorders, metastatic tu-
mours, auto immune disorders, infections, idiopathic disease, aneurism, pregnan-
cy, thromboembolic phenomena, diabetes etc. and may cause elevated levels of
WBC, RBC or PLT counts as well as HGB. The specimen can be warmed up to
37°C and re-analysed immediately or a manual WBC, RBC or PLT count can be
performed.
18 03-11-24 1001en01
Cold agglutinins
IgM immunoglobulins which are elevated in cold agglutinin disease may lower
RBC and PLT counts and increase the MCV.
HGB (hemoglobin)
Turbidity of the blood sample due to any number of physiological and/or thera-
peutic factors may produce falsely elevated HGB results. The CA620/530 how-
ever is compensated for this effect with high WBC counts up to WBC approx. 60
10e3/µl. In case of extreme WBC counts, the following is recommended:
1. Elevated WBC
The diluted sample should be centrifuged and the supernatant fluid checked
on a spectrophotometer for turbidity.
2. Elevated lipids
Elevated lipids in the blood sample will give the plasma a “milky” appear-
ance. This condition can occur with hyperlipidemia, hyperproteinemia and
hypobilirubinemia. Accurate HGB determination can be achieved by using
reference methods and a plasma blank.
Increased turbidity may also be seen in cases where the red blood cells are resis-
tant to lyzing. This condition will cause a falsely elevated HGB result but can be
detected by monitoring the MCHC.
Fetal blood
The mixing of fetal and maternal bloods may produce a falsely elevated HGB val-
ue.
HCT (Hematocrit)
As HCT is the product of MCV x RBC, any erroneous result in MCV and/or
RBC will produce an equal error in the HCT parameter.
WBC
An excessive number of WBCs might cause interference within the RBC popula-
tion and therefore a false MCV value.
PLT
Excessive numbers of PLT, in most cases, do not interfere with the MCV param-
eter due to the use of the floating discriminator technology in the instrument.
1001en01 03-11-24 19
Agglutinated RBCs
Agglutinated RBCs might trap platelets and may give an erroneous low PLT
count. The presence of agglutinated RBCs is detected by monitoring the MCHC
parameter and by careful examination of the stained blood film.
Chemotherapy
Cytotoxic and immunosuppressive drugs may increase the fragility of these cells,
which may cause low PLT counts. Reference (manual) methods may be necessary
to obtain an accurate platelet count.
Hemolyzis
Hemolyzed specimens contain red cell stroma, which may elevate platelet counts.
A.C.D. blood
Blood anti coagulated with Acid Citrate Dextrose may contain platelet aggre-
gates, which could depress the platelet count.
RBC inclusions
Erythrocyte inclusions may also produce a spuriously increased platelet count.
(e.g. Howell-Jolly bodies, siderotic and basophilic granules)
Platelet agglutination
Clumped platelets due to poor collection techniques or platelet satellitosis caused
by EDTA activation of immunoglobulins may cause a decreased platelet count
and/or an elevated WBC count. The specimen should be recollected in sodium
citrate anticoagulant and re analyzed for only the platelet count. The final PLT
result must be corrected for the sodium citrate dilution effect.
20 03-11-24 1001en01
Small erythrocytes
Very small RBCs might fall into the PLT region and might be counted as PLTs
and therefore influence the MPV parameter. (This effect is minimized in the in-
strument due to its floating threshold system)
Agglutinated erythrocytes
This may trap platelets and therefore affect the MPV parameter. Note that agglu-
tinated erythrocytes may be detected by carefully examine the MCHC parameter
and/or the stained blood film.
Chemotherapy
May also effect the size of the PLTs
EDTA
Note that all samples collected in EDTA will not maintain a stable MPV. The
PLTs will swell as a function of time and temperature.
LYMF (Lymphocytes)
The lymphocyte count is derived from the WBC count. The presence of nucleat-
ed red cells (NRBC), certain parasites and erythrocytes that are resistant to lysis
may interfere with LYMF count.
GRAN (Granulocytes)
The granulocyte cell count is derived from the WBC count. The presence of ex-
cessive numbers of metamyelocytes, myelocytes, promyelocytes, blasts and plas-
ma cells may interfere with an accurate granulocyte count.
1001en01 03-11-24 21
22 03-11-24 1001en01
The menu numbering for the CA620, CA620-CellGuard and CA530 is identical.
This user manual is therefore applicable for both type of instruments.
The following table shows which parameters are available on the 9, 16 and 20 pa-
rameter models:
CA530- CA530- CA530- CA620- CA620-
9 16 20 16 20
1001en01 03-11-24 23
24 03-11-24 1001en01
Sampling valve
1017.wmf
Two separate measuring chambers and transducers are used, one for RBC/PLT
and one for WBC/HGB analysis. This excludes any possibility of cross contam-
ination between the lyzer and the RBC/PLT dilution.
A pressure is applied on top of the diluted sample and the diluted sample is
pressed through an orifice (aperture) of 80 µm diameter. Each side of the orifice
is equipped with a platinum electrode and an electrical current is applied between
the electrodes.
When a cell is drawn into a constant current, flowing from an electrode through
the orifice to a second electrode, the electrical conductivity changes. This gener-
ates an equivalent voltage pulse.
The amplitude of the pulse is directly proportional to the volume of the repre-
sented cell. The number of pulses corresponds to the number of cells detected.
Coincidence corrections are made within the software of the instrument, giving
the instruments a full linear range within the specifications.
1016en01 03-11-24 25
electrical current
electrodes
cells
Flow direction
A=Aperture 1018.tif
The PLT, RBC and WBC parameters are measured on a precise aliquot of the
sample. The amount of sample measured is determined by the volume of a pre-
cise glass column; called a metering tube. Two optical detectors are used to start
and stop detection.
The start detector is activated by the flow of the isotonic diluent through the me-
tering tube. As the meniscus passes the optical path it causes a voltage change that
activates the count and sizing circuitry of the system.
As the isotonic diluent continues to flow upward into the metering tube it passes
the stop detector. This action stops the count and analysing process and the pa-
rameters and distribution curves are displayed (and automatically printed if so
programmed by the user). Due to this principle, the CA instruments perform ab-
solute counts related to fixed volumes.
glass tube
stop detector
start detector
1019.tif
The instruments use a lower discriminator level at approx. 2.5 fl. All cells over this
level are analyzed and the counts are stored. The user defines, by means of setting
the discriminator level(s), which cell is seen as a PLT or as a RBC cell. A red blood
cell is defined as a cell with a volume larger than the upper PLT programmable
discriminator level.
26 03-11-24 1016en01
All cells smaller than the upper PLT discriminator level will not be recognised as
red cells. In case the user has selected a “floating” discriminator; RBC cells are
counted from the set-point of this variable discriminator, which varies from sam-
ple to sample between the volume limits set by the user.
The number of cells for determining RBC is counted from a suspension of
1:40 000 dilution ratio of whole blood.
Let us suppose that a sample contains 5 000 000 cells/µl. A dilution of 1:40 000
then give a final concentration of 5 000 000 divided by 40 000 = 125 cells/µl, so
each µl drawn through the aperture generates 125 pulses.
The measured volume drawn through the aperture is 270 µl (factory calibrated),
so the system will count 270*125 = 33 750 pulses. The analyzer uses a fixed divi-
sion factor of 67.5, so the display will show 33 750/67.5 = 500, which is the cor-
rect value.
With the size of the orifice and the concentration of cells, there will be a certain
coincidence in the orifice system. However this is a constant system factor and
compensated within the software by a correcting algorithm to make the RBC
count linear within the stated specifications.
Hence, the total number of cells passing through the aperture when determining
the RBC is the value on the display multiplied by the factor 67.5. Therefore, a
sample which gives 5.00 in the RBC display field has been analyzed by measuring
500 * 67.5 = 33 750 cells.
The reproducibility is directly related to the total number of cells entering the or-
ifice. The higher the concentration, the better the reproducibility. The instrument
has a dilution ratio for RBC of 1:40 000 and the CV will therefore be less than
1% for samples with an RBC number within the normal range.
1016en01 03-11-24 27
Note:
The 'Counting time' is not related to the actual result. Atmospheric pressure vari-
ations, protein built-up within the orifice (aperture) and other secondary effects
that might cause pressure changes will NOT influence the counted parameters
RBC, PLT and WBC.
28 03-11-24 1016en01
n
4 hours old
fixed discriminators
fresh” blood
LYMF GRAN
Volume (fl)
1020.tif
MID cells
LYMF GRAN
Volume (fl)
1021.tif
1016en01 03-11-24 29
30 03-11-24 1016en01
Let us suppose that a sample contains 200 000 platelets/µl in whole blood. After
a dilution of 1:40 000 the sample contains 200 000 divided by 40 000 = 5 cells/
µl. So, each µl drawn through the aperture gives 5 pulses. As the counting volume
(the volume of the metering glass tube) is 270 µl, the total number of cells that
are analyzed will be 5*270=1350 cells.
In other words, the total number passing through the orifice when determining
the PLT is the value shown on the display screen without decimals multiplied by
the factor 6.75.
The reproducibility is directly dependent on the total number of cells entering the
orifice. In the instrument, measuring PLT from the same dilution as RBC, the CV
will be less than 3.5% for most of the samples within normal range. The CV will
be lower for most samples, but on some samples it might be slightly higher. A
'mean' CV of about 3.2% is expected for well-treated fresh EDTA whole blood
samples within the range of 200-350 103/µl. As the system uses an orifice size of
80 µm diameter, coincidence losses will take place with extreme sample RBC/
PLT counts. The system has a software with a well-balanced mathematical cor-
rection algorithm to minimize these effects
Please note that if a floating discriminator is used and no well-defined minimum
is found between the RBC and PLTs, the reproducibility of mainly the PLT is af-
fected. To check the reproducibility of the low PLTs, it might be wise to put the
analyzer in a fixed discriminator mode to exclude any error introduced by an ill-
defined RBC-PLT population.
Note:
In case high numbers of RBCs are present using the PLT concentrate mode, PLT
will show a flag “RP” (Red cells Present) to warn the operator that the PLT pa-
rameter may not be correctly analyzed.
1016en01 03-11-24 31
32 03-11-24 1016en01
1002en01 03-11-24 33
Note:
The instrument performs an auto-clean of the orifice after a TU is displayed; a
re-analyze of the sample might cancel this warning in most cases.
See Maintenance, Shut-Down & Transport on page 95. See also Trouble
Shooting on page 99.
Note:
That the orifice is cleaned automatically and that in most cases a re-analyze of the
sample will cancel this clogging.
See Maintenance, Shut-Down & Transport on page 95. See also Trouble
Shooting on page 99.
34 03-11-24 1002en01
Usually, with correct settings of the discriminator limits, the PLT parameter is still
measured correctly.
1002en01 03-11-24 35
NM'No Mode'
No significant population was found. There is no valid 3 part WBC population.
OM'One Mode'
Only one population is found. This flag might be displayed in case of Granulo-
cytosis or Lymphocytosis or if the Granulocytes are collapsed into the Lympho-
cytes (old blood).
TM'Triple Mode'
There are more than 2 main populations. This flag might be displayed in case of
Monocytosis, heavy mains disturbance or wrong installation due to poor ground-
ing of the instrument and/or connected printer.
BD'Bad Distribution'
The LYMF and GRAN population are too overlapped. This flag is displayed if
the Granulocyte population is collapsed into the Lymphocyte area. The reason
might be:
36 03-11-24 1002en01
Note:
The control must have been defined previously with the bar-code reader in menu
7.4.1.
1002en01 03-11-24 37
38 03-11-24 1002en01
1003en01 03-11-24 39
Location
The instrument should be placed on a clean and level table or work station. Avoid
exposure to sunlight. Make sure that the instrument has access to proper ventila-
tion and that at least ca. 10 cm free space must be left at the rear of the analyser,
marked as “A” in the picture below..
Place the instrument with easy access to the mains outlet and the mains cord. In
case that emergency shut-down is needed, due to an obvious malfunction of the
instrument and the instrument needs to be powered off follow the instructions
below:
1. Switch off the instrument immediately by:
Pulling out the mains cord from the mains supply.
2. Contact your authorized distributor’s service department immediately. Important
Operating the instrument
The instrument can work correctly within the stated temperature and humidity in an environment over
range however, to minimize the risk of bacterial growth in the reagent packages, +32°C (90°F) increases
it is strongly recommended to keep the laboratory at a temperature of approx. service-needs as well as
+22°C (72°F). degradation of sample
) specimen.
1178.bmp
Reagent packages
Reagent packages are preferably placed at the same level as the instrument (e.g.
the same table). In case such conditions cannot be met, place the Diluent con-
tainer at the floor level (maximum1 meter below the instrument level) but keep Warning
Contamination hazard
the Lyzer container preferably at the same level as the instrument (e.g. the same
table). Because no test method
can offer complete assur-
ance that HIV, Hepatitis B
Waste connection or C viruses, or other in-
The instrument has an open waste outlet and can be connected to a central waste fectious agents are absent,
system within the laboratory. The waste outlet, or container, must always be at a the waste should be han-
lower level than the instrument. National and/or local regulations must be fol- dled at the Biosafety Level
2 as recommended for any
lowed in all cases. infectious human blood
specimens in Protection of
Laboratory Workers From
Infectious Disease Trans-
mitted by Blood, Body
Fluids and Tissues- 2nd
Edition, Tentative Guide-
lines (1991) Document
M29-T2 promulgated by
the National Committee
for Clinical Lab. Standards
in the U.S.A. (NCCLS)
40 03-11-24 1003en01
Instrument installation
1. Lift the instrument and place it on the chosen location as shown below.
Be careful not to stress the front door of the instrument.
1023.tif
1024.tif
1003en01 03-11-24 41
1025.tif
7. Remove all transport guides at the pinch valves. These guides are red co-
loured plastic clamps inserted in each valve. Save them for later use in case
of a re-installation.
Transport guides
Important
One valve is located be-
hind the pre-dilute start le-
ver MPA. Do not forget to
remove the transport
guides in this valve.
The instrument will not
1026.tif
operate correctly in case a
transport guide is not re-
moved. Possible conse-
8. Re-assemble the front cover in reverse order. Be sure to reconnect the
quences of not removing
ground-wire!
all transport guides are:
Reagent pack installation Indication numbers, incor-
1. Place the reagent pack in the proper place. See Reagent packages on page rect parameter results and/
40. or incorrect counting
2. Put the probes as shown below (the example shows the lyzer container). times.
1027.bmp
42 03-11-24 1003en01
3. Locate the reagent inlets at the rear of the instrument, see picture below.
1028.tif
4. Connect the corresponding tubes from the reagent probes and the level de-
tector cables as shown in the picture below.
Caution 1029.tif
Bio hazard
The end of the waste tube 5. Connect the waste outlet to a proper container or open drain.
must be at a lower level See Waste connection on page 40.
than the instrument itself.
Final set-up
Not following this might 1. Locate the serial number plate at the rear and check that the mains voltage
lead to improper instru- and frequency corresponds to your local mains supply outlet.
ment functions and/or
waste liquid flowing back- 2. Insert the power cable and connect to the mains supply. A short beep is
wards into the instrument. heard where after the instrument will perform a self-check that normally
takes 2 minutes. (Maximum15 minutes in case of condensation).
1003en01 03-11-24 43
1030.tif
3
4
5
1031.tif
44 03-11-24 1003en01
1032.tif
[Print] will send the sample data to either the printer and/or serial output, de-
pending on how the instrument is configured by the end-user.
[Menu-Operate] is used to switch between Menus and the operational mode. The
operational mode is defined as the instrument status where the next sample can
be entered at the aspiration pipette.
1033.tif
The arrow keys are used to scroll through the Menu system, switch between dif-
ferent display modes and to display alternative parameters. The right arrow key is
also used to enter either a + or a - within the calibration menu.
1034.tif
1035.tif
1003en01 03-11-24 45
Note:
Aspiration command keys are not located on the keyboard. Aspiration of a sam-
ple is started by pressing the start-lever behind the aspiration needle.
1036en.gif 1037en.gif
The system is now filled with reagents. This cycle lasts for ca. 6 minutes.
When the system is finished with this cycle, the cursor field returns to the field
8.2. Press [Menu] to return to the main menu.
46 03-11-24 1003en01
1038en.gif 1039en.gif
1004en01 03-11-24 47
1038en.gif 1041en.gif
6.4 CA620/530
The CA620/530 is configured as default to the most common use. This applies
to “Normal Ranges”, “Floating Discriminator Settings” and “Units”. Default set-
tings of the instrument can be printed out after the installation procedure by se-
lecting menu 5.9.
Menu 5.8 is set to ‘1’ as shown in Setting up the CA620-CellGuard for Pre-
Donor Screening on page 47.
48 03-11-24 1004en01
1042en.gif 1043en.gif
As seen on the LCD screen, 4 different date formats can be set. For EU, set date
format 0, for the U.S. set date format 2.
1. Use the up-down arrow keys to scroll to the date format field.
2. Enter the correct setting.
3. Scroll to the line “Date'” and enter the correct date using the previous set
date format.
4. Scroll to the line “Time” and set the correct time in 24 hours notation.
In case a different date separator is required:
1. Scroll to the line “Date separator”.
2. Press [Enter]
From the following menu, different “date separator” signs can be chosen.
3. Use the scroll arrow keys to select.
4. Press [Enter] to validate.
1044en.gif
1045en.gif
Point 2, 3 and 4 describes the settings of the print commands and format.
1004en01 03-11-24 49
1046en.gif
Note:
If no external computer is connected to the instrument, do not select settings 3-
8 on the Main Menu.
For detailed information regarding printers and printing options, please refer to
section Printer setup menu CA620 on page 73 and Printer and Serial Output
on page 93. For advanced printer options, refer to appendix 530-30-205 available
from your distributor (in English only).
50 03-11-24 1004en01
1047en.gif
1048en.gif
1004en01 03-11-24 51
Select 1 2 3 4 5 6
number:
HGB expr g/dl mmol/l g/l g/l g/dl g/dl
HCT expr. % l/l % l/l % %
WBC 103/mm3 109/l 109/l 109/l 103/µl 109/l
RBC 106/mm3 1012/l 1012/l 1012/l 106/µl 1012/l
HGB g/dl mmol/l g/l g/l g/dl g/dl
HCT % l/l % l/l % %
MCV µm3 fl fl fl fl fl
MCH pg fmol pg pg pg pg
MCHC g/dl mmol/l g/l g/l g/dl g/dl
PLT 103/mm3 109/l 109/l 109/l 103/µl 109/l
MPV µm3 fl fl fl fl fl
RDW(CV) % % % % % %
RDWa fl fl fl fl fl fl
PDW fl fl fl fl fl fl
PCT % % % % % %
LPCR % % % % % %
LYMF 103 3
/mm 10 /l 9
10 /l9
10 /l9 3
10 /µl 109/l
GRAN 103/mm3 109/l 109/l 109/l 103/µl 109/l
MID 103/mm3 109/l 109/l 109/l 103/µl 109/l
LYMF% % % % % % %
GRAN% % % % % % %
MID% % % % % % %
52 03-11-24 1004en01
1049en.gif 1050en.gif
1051en.gif
1005en01 03-11-24 53
The background should not be higher than the figures below, assuming that at
least 2 blanks are run after a sample.
( CGS units) (“Units =1”:
RBC 0.01
WBC 0.1
HGB 0.1
PLT 10
1052en.gif
Warning
Contamination hazard if
contaminated blood enters
into open cut.
Always wear protective
gloves when handling
blood samples.
1053.bmp
(Gloves in the picture are
The display shows the following sequences: not shown for clarity rea-
sons only.)
Last Sample (= blank) Aspirating sequence
1052en.gif 1055en.gif
54 03-11-24 1005en01
Important
Remove the sample from
the open inlet aspiration
needle when the display
1056en.gif
has changed from “Aspi-
rating” to “Counting cycle
Although not mandatory, enter the ID of the sample with the numerical keys.
in progress”.
Note that a positive identification of a sample is highly recommended to avoid
Not removing the sample
erroneous patient parameter reporting. A maximum of 15 digits are allowed in the
could result in incorrect
ID field. In case the optional Bar-Code reader is installed, simply scan in the ID
washing sequence of the
barcode from the sample tube.
aspiration needle.
After approx. 30 seconds; the CA620 will switch to the INTRA mode as shown
below:
1057en.gif
Note:
The INTRA mode is only available on the CA620-series
The display shows the counting rate on each counted parameter in real-time. The
horizontal lines in the “boxes” represent the parameter normal range as set for
PROG1 in menu 5.4.1, which is assigned to 'blood mode'. The drawn line during
the analysis is slightly irregular due to the counting statistics. However if the
drawn line is within the parameter limit lines, the sample is probably within the
normal range. In case the drawn line is outside the limits, the sample is probably
pathological. With this tool, the operator is alerted of a suspected pathological
case before the sample is analyzed completely. The vertical dotted lines represent
the expected normal counting time. See below.
Rate
1005en01 03-11-24 55
Important
The instrument is ready to
accept the next sample af-
ter the flashing display cur-
1059en.gif sor in the upper left display
corner disappears.
Entering the next sample
7.3 Analyzing the Sample (Pre-Dilute) before the flashing cursor
Any dilution rate between 1:200 and 1:250 can be used under the condition that
disappears will result in er-
the minimum total volume is 5ml and the maximum volume 8 ml. It is obvious
roneus parameter results
that always the same dilution rate must be chosen as any reproducibility error in
for the next sample.
an external dilution will directly affect the counted parameters.
Examples:
• 20 µl and 5 ml diluent, 30 µl and 6 ml diluent or 40 µl and 8 ml diluent
The sample should be analyzed as soon as possible. Prolonged waiting increases
the inaccuracy of the MCV and WBC differential parameters.
• Place the cup with the pre-diluted sample under the aspiration pipette for
pre-diluted samples and press the lever behind the pre-dilute aspiration pi-
pette.
Note that the instrument aspirates more than 5 ml. Volumes over 5ml are not
specified. This means that in most cases the whole sample (volume) is aspirated
into the analyzer.
• Remove the sample cup after aspiration.
The display sequence on the LCD screen is identical to the description in Rou-
tine Operation on page 53 above.
If no sample was entered and the pre-dilute start lever was pressed anyhow, the
system will recognise this and restore the liquid flow system automatically.
1060en.gif
56 03-11-24 1005en01
Warning
Contamination hazard if
contaminated blood enters
into open cut.
1061en.gif
Always wear protective
gloves when handling
blood samples. 3. Puncture, using the Boule Micro Lancet.
1062.jpg
Caution
Always insert the capillary
tube in the longest metal
tube of the MPA.
Inserting the capillary tube
in the shortest metal tube
1063.jpg
of the MPA might damage
the MPA or break the cap- 5. Insert the micro pipette (capillary) in the longest tube of the MPA device.
illary.
1064.jpg
1005en01 03-11-24 57
6. Insert the MPA into its holder and the instrument will automatically start
the analyzing sequence.
Important
Fill the micro pipette com-
pletely and wipe off the ex-
cessive blood on the
outside surface.
1065.jpg Ignoring this instruction
might cause incorrect and
Note: non-reproducable results.
7. Detailed movie shots are available on Boule’s support server how to use the
MPA system in detail. Please put your browser to the following address and
download the MPEG movie shots. The movie shot shows also how to clean
the Micro Pipette before it is loaded into the MPA.
www.medonic.se/MPA/
Warning
Contamination hazard if
contaminated blood enters
into open cut.
1066.bmp
Always wear protective
gloves when handling
blood samples
(Gloves in these pictures
are not shown for clarity
reasons only.)
58 03-11-24 1005en01
2. Push (rotate) the protection cover to the left and insert the tube up-side
down.
Caution
Insert the sample tube
with lid facing downwards.
Ignoring this instruction
may damage the aspiration
needle.
1067.bmp
1068.bmp
The aspiration cycle is now started following the sequence as described in Anal-
ysing the Sample (Open Tube) on page 54.
Remove the tube after the sample parameters are displayed in reversed order. The
adapter is ready to accept the next tube when the sample parameters are displayed
and the flashing cursor in the parameter-field of the display disappears.
1005en01 03-11-24 59
1069en.gif
2. Aspirate the sample by pressing the start lever or pull out the MPA.
3. After the aspiration of the sample or insertion of the MPA, enter the sample
ID if required.
The sample is now processed as “Blood” and all PROG 1 settings are applied to
this sample.
Note:
The next sample will “fall back” to the default PROG mode unless again changed
by the user.
60 03-11-24 1005en01
1071en.gif
1072en.gif
In the display example above it is seen that a total of 19 samples are in memory
but none (0) during the current date. The whole memory, in this example 19 sam-
ples, is selected by just pressing [Enter] at the “From Date” line in menu 1.4.
1073en.gif
Use the arrow down key to select other condition-selections. The selected num-
ber of samples is always displayed as well as the total number in memory. Press
[Menu] to return to the previous menu 1.4 to 1.8.
Please find below some examples how the sample memory can be used.
1006en01 03-11-24 61
Question 1 (example):
To-days date is 20 July 2003. We have a sample in memory tagged as ID = 1021.
This sample was analyzed 9 times during this day. We want to determine the CV
of this sample.
Answer:
1. Enter the “Sample Memory”.
2. Scroll to “ID =”
3. Type 1021
4. Press [Enter]
Now, the display shows SEL “9 of 75”, confirming that 9 samples with this ID
were found in memory of total 75 samples.
5. Scroll down to “Statistical Calculations”.
6. Press [Enter]
7. Press [Print] if a printout is required.
Note:
Statistical calculations are done on both Normal+Abnormal and Normal param-
eter values only.
“Abnormal parameter values” are defined as being all values except 0 or out of
measuring range.
Normal parameter values are values within the parameter normal range as set in
menu 5.4.
SD, CV, X and n are displayed for each parameter.
Use the “left/right”- keys to scroll between the parameters and use the “up/
down”-keys to select SD, CV, X or n for NORM+ABN or NORM values only.
Pressing [Print] will print all available data to the selected printer. This is the rec-
ommended way of listing the statistical calculations.
Below is an example of a Statistical calculations display with no specific ID selec-
tion:
1074en.gif 1075en.gif
The example above shows that on RBC, 16 samples are present in the selected
memory whereof 9 within the normal range.
Question 2 (example)
We want to have access to all samples in memory and scroll backwards through
all samples.
62 03-11-24 1006en01
Answer:
1. Enter the “Sample Memory” menu 1.
2. Scroll down to line 1.4 and press [Enter].
3. Scroll to line “From Date=” and press [Enter].
Now, the bottom line will change to “198 OF 198”, confirming that all samples
in memory are selected. (the number “198” is an example only).
4. Press [Menu] to return the menu 1
5. Scroll to line 1.5 and press [Enter].
The display shows now the last sample in memory.
1. Use the “up/down” keys to scroll through the sample memory.
2. Use the “right/left” keys to display different parameters.
The [Print] key can be used to print a specific sample.
1076en.gif
The letter 'M' is displayed in the upper right corner to indicate that the display
shows a sample in memory and not the last run.
Answer:
Enter the “Sample Memory” from the main menu 1.
1. Scroll down to line 1.4 and press [Enter].
2. Scroll to the line “From Date=” and press [Enter].
3. Scroll down with the “up/down” key to the line PROG and select “2”. As
“2” is assigned to PLT concentrate mode.
The display might change to “22 OF 200” confirming that 22 samples were tested
in PLT concentrate mode.
4. Press [Menu]to return to the sample memory menus.
5. Scroll down to Q/C calculations and press [Enter].
6. Press [Print] to get a print out of the statistics.
1006en01 03-11-24 63
1077en.gif 1078en.gif
1081en.gif 1082en.gif
Note:
Each PROG has its own set of “Normal Range settings”, “Floating PLT/RBC
discriminator settings” and “WBC differential settings”. The Calibration is the
same for all PROGs
64 03-11-24 1006en01
CA620-CellGuard
1. Enter PROG 1 to set normal ranges for “Blood mode” as shown above.
or
2. Enter PROG 2 to set normal ranges for “PLT concentrate mode”.
The following example sets the normal range limits for “PLT concentrate mode”.
1083en.gif 1084en.gif
Note
that several parameters do not have normal ranges set as they are blocked from
being displayed and printed. See also menu 5.10.12 'Block parameters'.
Enter the required ranges by using the numerical key pad.
3. Validate with [Enter].
4. Scroll through this menu using the arrow keys.
5. Press [Menu] to return.
Blood mode
In the example below, PROG 1 is shown by pressing [Enter] and the settings for
the floating discriminator are displayed.
1085en.gif 1086en.gif
Note:
Applicable for CA620-CellGuard
1006en01 03-11-24 65
c) PLT conc. mode is set to “0” as PROG 1 is assigned to “Blood mode” only.
Important
Do not alter PLT conc.
mode. This must be “1” to
enable the correct sample
aspiration mode.
1083en.gif 1088en.gif
Setting PLT conc. mode to
“0” will aspirate the sam-
The displayed fields are:
ple for 10 seconds and
a) PLT-L. The lowest possible setting for the floating threshold between PLT macro Platelets up to 60fl
and RBC in fl (=µm3). As default this is set to 15 fl will not be counted.
b) PLT-H. The highest possible setting for the floating threshold between PLT
and RBC in fl (=µm3). As default this value is set to 60 fl, which is the max-
imum level available in this menu.
c) PLT conc. Mode is set to “1” as PROG 2 is assigned to “PLT concentrate
mode” only. Setting this menu to “1” enables the PLT-H setting to a maxi-
mum level of 60 fl. This setting (60fl) is chosen as default in PLT concen-
trate mode to include possible macro PLTs in the total PLT count.
This setting “1” also enables the aspiration mode to switch to a time aspiration
instead of using the Blood detector.
Also, if PLT-H is set to 30 fl instead of 60 fl, all PLTs larger than 30 fl will be
reported by the system as RBCs, which might introduce a slightly false RBC pa-
rameter value. Therefore, it is recommended not to change this setting without a
compelling reason.
1089en.gif 1090en.gif
66 03-11-24 1006en01
1091en.gif
1092en.gif
1006en01 03-11-24 67
1093en.gif
Within setup menu 2, less common functions are displayed and altered, such as
language, units etc.
To set language and units; please refer to Initial System Configuration on page
47.
Machine ID
Refer to Setup Menu 2 on page 68 above and scroll down with the arrow keys
to the line “Machine ID”.
This menu is only of use in case the instrument is connected to a computer sys-
tem. On the serial output a Machine ID number is sent to the connected com-
puter to identify the instrument in case that several Boule analyzers are used in a
lab. In such an environment the Machine ID can be set to the serial number of
the instrument to give the host a positive identification from which system data
was sent.
1094en.gif 1095en.gif
68 03-11-24 1006en01
1096en.gif
Change the setting, if required, by using the “left/right” arrow keys and jump to
another line within this menu with the “up/down” arrow keys.
See further in section Printer and Serial Output on page 93.
Barcode reader
In case a Barcode reader is connected (available from Boule as optional device),
the device must be enabled in this menu.
1097en.gif
1006en01 03-11-24 69
1098en.gif
Scroll down to this line using the arrow keys and set '1' in case the instrument is
located above 1500 meters (4500 feet) MSL.
This setting will only extend some of the washing sequences in the instrument
due to the capacity reduction of the waste (drain) pump when used at high alti-
tudes.
No other instrument specifications will be affected.
1000en.gif
This menu prints out a list of defined quality control-blood types. Type [Enter] at
this menu line to print out the list. As quality control-blood is not comparable
with human blood regarding the structure of the WBC differential mainly, certain
discrepancies in the total WBC count and/or differential count might occur
whenever using such fixed cells as a normal sample. Entering the defined ID
numbers as obtained from the list, terminated with a “+” sign (right arrow key),
will instruct the instrument to shift it's discriminator range automatically so that
it fits the specific control-blood see also section Calibration and controls on
page 81, in this manual.
Example of a printout:
70 03-11-24 1006en01
Hence, enter such control blood by typing 888 and press the “right” arrow key.
Note:
Entering a specific control blood Using 777+ to 999+ Boule control bloods, are
stored as default in PROG 8.
Reagent type
This menu line is only used to initialise the basic set up at the factory. It defines
for which purpose the instrument is used and stores the correct default values.
This menu is password protected and cannot be altered by the user. It should
show “15” for the CA620-CellGuard and “11” for the CA620/530, in all cases.
CA620-CellGuard CA620/530
1100en.gif 1014en.gif
1101en.gif 1102en.gif
In the above example, we want to set the name “Blood” for PROG1.
Press [Enter] and the sub menu is displayed asking to define the text to be shown
on the display and printer.
1006en01 03-11-24 71
Press [Enter] to set the required text on the display or scroll to the next line to
alter the printed name on the printout instead.
The following is displayed.
1103en.gif
1. Scroll with the “up/down” arrow keys through the displayed field.
2. Press any digit to select the required character.
3. Press [Enter] to return.
Changing both PROG1 and PROG2; the list will now show:
CA620-CellGuard CA620/530
1104en.gif 1146en.gif
Note
That the PROG names are not changed in any menu to “Blood” or ”PLT-C” but
clarified only as shown above.
The display and printout however will show “Blood” or “PLT-C” instead of
PROG1 or PROG2.
Block parameters
The CA620 allows the operator to block certain parameters to be displayed and
printed. This is especially useful when the instrument is used for PLT concentrate
control only. Several parameters are of no interest in such case, like HGB, HCT,
MCHC etc.
To do so:
1. Scroll with the arrow keys to line 5.10.12 “Block parameters”.
2. Press [Enter]
72 03-11-24 1006en01
1105en.gif
1. Scroll down to the line PROG 2 in case some of the parameters in the PLT
concentrate mode should be blocked. See below.
2. Press [Enter] and the next sub menu is displayed.
1106en.gif 1107en.gif
In this menu, setting a “1” to a parameter means that it will be displayed and
printed. Setting “0”, will block the parameter from the display and printout.
In the above example it is seen that all parameters are blocked except RBC, PLT,
MPV, WBC and LPCR whenever the CA620-CellGuard is used in PLT concen-
trate mode.
1108en.gif
1006en01 03-11-24 73
In such a case, one must keep the pre-dilute start lever pressed for 5 seconds to
start the pre-dilute aspiration.
1109en.gif
Service menus
Service menus, starting at line 6 of the main menu, are used by your authorized
distributor to check and maintain the instrument. These menus are described in
the Service Manuals available to your authorized distributor on CD-ROM. Sev-
eral of these menus are password protected. For clarity reasons and back report-
ing to the manufacturer, service menus and the proper description of each is
available in the English language only.
74 03-11-24 1006en01
1110en.gif
This is a normal warning and indicates that the MPA was pulled out. The instru-
ment is waiting for the MPA to be inserted where after a count cycle starts auto-
matically.
1111en.gif
The MPA cycle was aborted by pressing the [CE] key. Push back the MPA. The
system will not start a count cycle.
1112en.gif 1113en.gif
The MPA was pulled out before the cycle was finished. Push back the MPA
adapter and the instrument will restore the flow system with a prime cycle. See
menu message above.
1114en.gif
1008en01 03-11-24 75
The MPA was pulled out in a wrong mode (a menu was displayed). The MPA may
only be pulled out in “Operational mode” where the last run sample is displayed.
The operational mode is selected by pressing the [Menu/Operate] key.
1115en.gif
The above warning indicates that the Pre dilute Sample inlet was activated but no
sample was aspirated. This might be caused by:
a) No sample is entered.
b) An attempt is made to run on distilled water in the pre-dilute inlet.
1116en.gif 1117en.gif
If the instrument is in Stand-by mode, the text displayed might not be clear as the
back lighter of the LCD display is switched off. The instrument will now perform
an auto check and cleaning cycle every 4th hour.
1118en.gif 1119en.gif
76 03-11-24 1008en01
In case one of the reagent containers are empty within the Stand-by mode, one
of the above warnings will be displayed. It is essential that new reagent containers
are installed, as the 4-hour checking cycle cannot be performed by the instru-
ment.
1120en.gif
This warning indicates that an error indication occurred and it was not cleared by
the user before the system entered the Stand-by mode.
Restart the instrument, cancel indications by pressing [CE] and perform a
“Prime” from menu 8.1
1121en.gif
When the instrument is in Stand-by mode, and the [Menu] key is pressed, the sys-
tem will perform a “Prime cycle” to restore the flow system. This lasts for approx.
2 minutes.
1122en.gif 1123en.gifs
The above warnings are displayed in case of an empty reagent container. Replace
and perform a “Start prime cycle” menu 8.1 from the “System flow menu”,
menu 8.
1008en01 03-11-24 77
1124en.gif 1125en.gif
The warning messages above are related to the connected printer. Either the
printer is not connected or busy.
1126en.gif 1127en.gif
1127en.gif
The above warning is displayed in case the connected computer serial line is busy
and not ready to accept data.
78 03-11-24 1008en01
1129en.gif 1130en.gif
The above messages are displayed from the sample memory only. Printing of 19
samples is selected. The second warning shows a question in case a deletion of 19
samples is requested from memory.
1131en.gif
1008en01 03-11-24 79
80 03-11-24 1008en01
1009en01 03-11-24 81
.
Do not push the sample tube against the Correct sample handling
washing device
Important
Dry the aspiration pipette
manually on the outside at
each control blood run
and do not push the con-
trol or calibrator tube
1135.bmp 1136.bmp against the upper washing
device. See pictures.
Not following this disci-
10.3 Calibration on a Known Sample pline might give dropping
The following procedure can be used in case no Hematology Blood Calibrator is values for the RBC, HGB,
available or if the recommended transport conditions of such cannot be met. PLT and WBC parameters.
The RBC, WBC and PLT values can be obtained by using a reference analyzer or
a microscope method.
The hematocrit should be checked with a micro centrifuge method and the he-
moglobin with the cyan meta-hemoglobin method on a reference photometer
system.
Calculate the MCV by: MCV =HCT/RBC.
Procedure:
1. Make at least 3 HCT determinations with a micro centrifuge and calculate
the mean value.
2. Make at least 3 HGB determinations with the cyan meta-hemoglobin meth-
od and calculate the mean value.
3. Perform blank counts on the instrument and check that the background is
according to the limits stated in chapter Analysing the Sample (Open
Tube) on page 54.
4. Aspirate and analyze the known sample at least six times. Print the parame-
ter results of each count.
82 03-11-24 1009en01
5. Check that the CV values1 of the measured parameters are within the fol-
lowing ranges:
6. Calculate the difference as per cent between the mean of the results ob-
tained by the instrument and the known values of the sample.
7. Select “Calibration Whole Blood” from the “Setup menu”.
8. Scroll to the parameter that has to be adjusted and enter the percentage that
was calculated. Use the -/+ key (=right arrow key) to select a - or a + cor-
rection.
In the example below, the RBC is adjusted with +2.0%
1137en.gif 1138en.gif
1009en01 03-11-24 83
The following calibration procedure is used for the CA620 and the CA620-Cell-
Guard, using a Boule certified whole blood calibrator:
1. Run the calibrator 5 times at the open tube inlet. Follow the guidelines
shown in Use of Calibrators and Controls on page 81 above. Enter the
ID 888+ for each run; entered as 888 and terminated with the right-arrow
key. See Print control blood id:s on page 70. In case the calibrator was read
by a bar-code reader as outlined in Initializing the Levey-Jennings Plots
and Functions on page 87, use the calibrator ID instead of the 888+ entry.
To display a quick overview which controls and calibrators are defined, go
to menu 5.10.9 and press [Enter] to print a list with control/calibrator def-
initions.
2. Go to menu 5.1. The following is displayed:
1139en.gif
WBC < 4%
RBC < 1.8%
PLT < 6%
MCV < 1%
HGB < 1.8%
MPV < 5% (if applicable)
In case one run should be excluded from the auto calibration to reduce the CV,
move the cursor to the specific sample and press [Enter]. The display of “Use”
will show “0” on such sample, indicating that it is not included in the auto cali-
bration.
It is also possible to print out a sample by moving the cursor to a sample line and
pressing the [Print] button.
1. The CV values are stated higher than the typical CV of the instrument due to
the limited number of sample runs.
84 03-11-24 1009en01
4. Enter the target value read from the assay sheet in the field “Target” and
press [Enter]. The analyzer will calculate the new calibration factor and dis-
play this in the field “New”.
1140en.gif 1141en.gif
Note:
Sample runs with (SE, DE etc.) are not included in the 5 samples displayed in
menu 5.1-5.3. This is shown in the calibration menus as =0 at the sample display.
1009en01 03-11-24 85
1142en.gif
In the example above, the last sample showed RDW=13.8. The new target value
is entered as 12.5%. The last sample is recalculated to the target value.
Please note that RDW% obtained from blood samples is depending on:
a) How the blood was collected.
b) Temperature conditions in the lab.
c) Mixing conditions of the sample.
d) Type and condition of EDTA in the sample tube.
It is therefore recommended to adjust the RDW calibration within the local en-
vironment.
86 03-11-24 1009en01
1157en.gif 1158en.gif
Important Note:
Do not forget to dry the Refer to section “Calibration” how to handle the blood control and how to dry
aspiration pipette manually the aspiration pipette manually in order not to contaminate the blood controls.
on the outside at each con- See Use of Calibrators and Controls on page 81.
trol blood run and do not CA620 and CA530 without bar-code reader.
push the control sample
tube against the upper Blood controls must be identified with the ID number 777+ (low), 888+(norm)
washing device. and 999+ (high). The “+” sign is entered by pressing the right arrow key.
Not following this disci- PROG8 (CNTRL8) is used for the definition of discriminator settings and “nor-
pline might give dropping mal range” settings. Note that the Levey-Jennings plots are not available in case
values for the RBC, HGB, the bar-code reader is not installed.
PLT and WBC parameters.
11.2 Initializing the Levey-Jennings Plots
and Functions
To initialize the L-J functions for the blood control(s), the bar-code reader must
be installed.
Enter menu 7.4 and 7.4.1.
Follow the instructions on the control assay sheet how to read the blood control
data.
1159en.gif 1160en.gif
1156en01 03-11-24 87
After the blood control data is read from the assay sheet. Go to menu 7.4.2 to
view the blood control assay values. See example below. Use the left or right ar-
row keys to switch between the 2 displays, see picture below.
12 different blood controls from Boule can be defined and stored simultaneously.
Use the up/down arrow keys to display another defined control blood.
.
1161en.gif 1162en.gif
Press [Enter] at menu 7.4.3 to print out all blood control definitions or press
[Enter] at menu 5.9 to obtain a short list.
Note:
In case all 12 control blood definitions/pages are in use, the next control that is
read via the bar-code scanner will ask the operator to delete the oldest control
samples first before continuing. Also, proper warnings are displayed in case a
wrong bar-code is read or if a bar-code is read out of sequence.
88 03-11-24 1156en01
To refine the search and selection criteria of the control, enter menu 7.2.1. See
example below.
1163en.gif 1164en.gif
After the select conditions are defined, enter menu 7.2.2 to display an overview
of the blood control. Use the left or right arrow keys to display the L-J plots. See
example below. The L-J plots are displayed for all parameters defined in the con-
trol assay sheet except the WBC differential parameter “MID”.
1165en.gif 1166en.gif
To print out the L-J plots, go to menu 7.2.3 and press [Enter]. The number of
points printed will be depending on the type and resolution of the installed print-
er.
The figures printed on the vertical scale points to the first plotted horizontal line.
E.g. in the HGB plot in the figure above, the mean value is 12.7. The first hori-
zontal line above the mean value is 13.1 (the max limit for this control) and the
second horizontal line above the mean value is 13.5.
The number of control runs displayed or printed is always counted from today's
date back in time. In the example above, the last run 167 control runs are dis-
played.
Note:
In case a control shows an error or warning flag SE, DE, FD, OF, LO, HI, NG,
TU, TL or TB; the parameter values of such control run will not be included in
the L-J plots.
1156en01 03-11-24 89
Enter menu 7.4.4 to alter the expected mean values for the MCV, MCH and
MCHC. It is recommended that at least the MCV and MCH parameters are set
to the expected patient mean values with a range of +/- 3%.
1167en.gif 1168en.gif
To display the X-B Plots, enter menu 7.3 see example below.
1169en.gif 1170en.gif
By default, all sample data is selected. Refine the display criteria if requested by
entering a date window in menu 7.3.
Go to menu 7.3.1 to display the X-B data summary and use the left or right arrow
key to display the plots.
1171en.gif 1172en.gif
The data points displayed are always from today's date back in time. In other
words, the last data points are displayed. In the example above, 200 points are dis-
played giving the mean of 4000 samples as each point is the mean value of 20
samples.
The figures printed on the vertical scale points to the first plotted horizontal line.
E.g. in the MCHC plot in the figure above, the mean value is 34.0. The first hor-
izontal line above the mean value is 35.0 (the max limit set in menu 7.4.4) and the
second horizontal line above the mean value is 36.0.
Go to menu 7.3.2 and press [Enter] to printout the X-B Plots. The number of
points printed will be depending on the type and resolution of the installed print-
er.
Note:
Samples with an error or warning flag SE, DE, OF, LO, HI, NG, TU, TL or TB
are not included in the X-B Plots.
90 03-11-24 1156en01
1173en.gif 1174en.gif
Enter menu 7.1.5 to view the controls and use the up/down arrow keys to scroll
through the control memory. Size distributions are not saved and displayed for
blood controls.
Go to menu 7.1.6 to display the mean, SD and CV of the selected data and use
menu 7.1.7 to print out the selected controls.
Menu 7.1.8 is used to delete the selected control samples from memory.
1156en01 03-11-24 91
92 03-11-24 1156en01
Note:
A printer or serial connection must conform to EN 60950
Important
If an IBM, Epson or HP
compatible printer is used, 1150en.gif
it is of great importance
for the correct printing
Press [Print] to printout a list of available formats.
that the actual printer is
supporting the selected Note:
data format. Incorrect Extended printer format settings and user definable print layouts are available.
printing might follow if an Please refer to appendix 530-30-205 for detailed information how to set up a user
improper format or printer definable format. This appendix is available from your authorized distributor in
is selected. the English language only and requires basic understanding of printer protocol,
formats and font definitions.
1010en01 03-11-24 93
Use a correct cable that is specified for RS232 data transmission and don't use
unnecessary high-speed set-ups on the output. This to avoid checksum errors and
to secure a correct data transfer. Enter the menu 5.10.5 and select the proper data
transmission speed and hand shaking protocol.
Note:
To enable the serial output of the instrument, the Print functions in point 2 and/
or 3 of the main-menu must be set properly see Printer Configuration on page
49.
Connections:
PC Computer using a 25 pin RS232
Cable end instrument Cable end PC
9 Pin Female DSUB 25 Pin Female DSUB
2------------------------------------------------------------------- 3
3------------------------------------------------------------------- 2
5------------------------------------------------------------------- 7
7------------------------------------------------------------------- 4
8------------------------------------------------------------------- 5
94 03-11-24 1010en01
1011en01 03-11-24 95
96 03-11-24 1011en01
Note:
Customers are advised to be knowledgeable of applicable local, state and federal
requirements, and the contents of effluent streams, before disposing of waste in
public sewer systems.
1011en01 03-11-24 97
98 03-11-24 1011en01
Note:
In case frequently a TU or TL warning is displayed on the WBC parameters, this
might indicate that the pre-dilute inlet is clogged.
1. Attach a syringe with hypochlorite, approved by Boule, to the pre-dilute as-
piration inlet and rinse from any possible clogs.
2. Afterwards, enter menu 8.1 (Prime) to restore the flow system.
Note:
In case the MPA is installed, such blockage might be caused by broken capillary
tubes within the MPA and/or connecting tubes. Call your service dept. in such
case, see also Analysing the Sample (Micro Pipette Adapter, MPA) on page
57.
1012en01 03-11-24 99
The lamp voltage is usually ca. 4.5 Volts and the photometer output ca. 3.5 volts.
These values are not critical as long as the photometer output voltage is between
3.2 and 3.8 volts.
If the photometer output voltage is lower than 2.5 volt, a “LO” indication will
follow on the HGB parameter value.
An automatic adjustment will be performed when digit [1] is pressed on the key-
board.
The system will measure the offset, blank and automatically adjust the lamp volt-
age to obtain a correct reference. This procedure takes ca. 5 minutes.
3. In case the above does not help to reduce the PLT background to accept-
able limits, contact your authorized distributor and request bulletin 000-80-
020 where instructions are given how to decontaminate the instrument.
Indication = 1
Date and Time not set. Set within the SETUP Menu.
Indication = 2
The Cap Piercing Device was started and immediately aborted by turning the
safety cover Counter-Clock-Wise.
Press [CE] and the instrument will be operational.
Indication = 9
Auto HGB adjustment failed. The cause might be a broken lamp. Contact service
department.
Indication = 100-199
One of the internal motors or motor drivers failed. Restart the instrument, cancel
any indication with [CE]. If error remains, contact service department. In case of
a wrong installation (no CVT in use) this indication might occur due to excessive
variation of the mains supply. See Mains Supply Environment on page 39.
Indication = 200-299
The diluent status in the internal mixing cup was incorrect. Restart the instru-
ment, cancel any indication with [CE] and do a “Prime” from menu 8.1
These errors might be displayed in case there was an Indication 300-399 previ-
ously.
Restart the instrument by disconnecting the power and switch on again. Cancel
the indication with [CE] and do a “Prime” from menu 8.1.
Indication = 900-999
Memory failure. Call the service department.
Indication =
-1099 and 2000-2999
Hardware or software failure. Contact the service department.
Warning
Contamination hazard if
contaminated blood enters
into open cut.
Always wear protective
gloves when handling 1143en.gif 1144en.gif
blood samples or parts of
the instrument that might
2. Press digit [1]The motor will now move the needle to its far down position
be contaminated with
(if it was not there already).
blood.
3. Switch off the instrument and open the front “door”.
4. Remove the upper needle locking ring only as indicated below.
Needle
Aspiration tube
1145.bmp
Note:
In the displayed menu 6.9.3, the positions 1, 2 and 3 stand for:
1 = Lowest position equals to open tube selection.
2 = Middle position equals to closed tube selection (Can only be reached from
position 3).
3= Upper position equals to penetrating position.
105 03-11-24
T
TB 36
Thrombocytes 20
TL 34, 99
TM 36
Transport 95, 96
TU 34, 99
tubing 102
U
Units 51
W
Warning flags 15, 33
Warranty 9
Waste connection 40
waste container 97
waste liquid 10, 43
waste outlet 43
Waste pump 101
Waste tube 39
WBC 18, 19, 25, 27, 28, 36, 66, 81
WBC discr. Program 71
whole blood 9
106 03-11-24