Manual Testo-435 PDF
Manual Testo-435 PDF
Instruction manual en
Contents
General notes ......................................................2
1. Safety instructions ..............................................4
2. Intended purpose ................................................5
3. Product description ............................................6
3.1 Display and control elements ..........................................6
3.2 Interfaces ......................................................................8
3.3 Voltage supply ................................................................8
4. Commissioning ..................................................9
5. Operation ..........................................................10
5.1 Connecting a probe ....................................................10
5.2 Switching on/off ..........................................................10
5.3 Display light ................................................................11
6. Setting the instrument ......................................12
6.1 Configuration menu ......................................................12
6.1.1 Profile ............................................................................12
6.1.2 Units ..............................................................................13
6.1.3 Device ............................................................................13
6.1.4 Probe ............................................................................15
6.1.5 Language ......................................................................18
6.2 Main menu ..................................................................18
6.2.1 Memory (435-2 / -4 only) ................................................20
6.2.2 Measuring program (435-2 / -4 only) ..............................21
6.2.3 Mean ............................................................................22
6.2.4 P = 0 (435-3 / -4 only) ....................................................22
6.2.5 Calculation ....................................................................23
6.2.6 Parameter / Area ............................................................24
6.2.7 Cyclical Print (435-1/-3 only) ..........................................25
7. Measuring ........................................................26
8. Care and maintenance ......................................29
9. Questions and answers ....................................30
10. Technical data ..................................................31
11. Accessories / spare parts ..................................33
2 General notes
General notes
This chapter provides important advice on using this documentation.
The documentation contains information that must be applied if the product is
to be used safely and efficiently.
Please read this documentation through carefully and familiarise yourself with
the operation of the product before putting it to use. Keep this document to
hand so that you can refer to it when necessary.
Identification
Representation Meaning Comments
Warning advice: Warning! Read warning advice carefully and take the precautionary measures
indicated! Serious physical injury could occur if you do not take the
precautionary measures indicated.
Warning advice: Caution! Read warning advice carefully and take the precautionary measures
indicated! Slight physical injury or damage to equipment could occur
if you do not take the precautionary measures indicated.
Short form
de
This document uses a short form for describing operating steps (e.g. calling up
a function).
en
Example: Calling up the “Instrument data” function
fr
Short form: Device OK Inst.data OK .
(1) (2) (3) (4)
es
Steps required:
it
1 Press / to select the Device function.
2 Confirm selection with OK .
pt
3 Press / to select the Inst.data function.
sv
4 Confirm selection with OK .
nl
??
??
4 1. Safety instructions
1. Safety instructions
This chapter gives general rules which must be followed and observed if the
product is to be handled safely.
de
Warning: Changes or modifications not expressly approved by the party responsible for compliance could void
the user's authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part
en
15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in
fr
accordance with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning the equipment off
es
and on, the user is encouraged to try to correct the interference by one or more of the following measures:
· Reorient or relocate the receiving antenna.
· Increase the separation between the equipment and receiver.
it
· Connect the equipment into an outlet on a circuit different from that to which the receiver is needed.
· Consult the dealer or an experienced radio/TV technician for help.
pt
Operation is subject to the following two conditions:
· this device may not cause harmful interference, and
· this device must accept any interference received, including interference that may cause undesired operation.
sv
nl
2. Intended purpose
??
This chapter gives the areas of application for which the product is intended.
??
Use the product only for those applications for which it was designed. Ask
Testo if you are in any doubt.
testo 435 is a compact, multifunction measuring instrument for measuring
temperature, humidity and flow rate.
The product was designed for the following tasks/applications:
· Measuring room climate
· Regulating and inspecting ventilation and air-conditioning installations
· Measuring the pressure dew point in compressed air systems
· Assessing the quality of room air with the help of the IAQ probe
The product should not be used in the following areas:
· Areas at risk of explosion.
· Diagnostic measurements for medical purposes.
6 3. Product description
3. Product description
This chapter provides an overview of the components of the product and their
functions.
Strong magnets
Damage to other instruments!
Keep a safe distance from products which
could be damaged by magnetism (e.g.
monitors, computers, pacemakers, credit
cards).
Probe socket(s)
Button functions
Button Functions
Function button (3x): The function depends on the button assignment
at the time
Change display of the 1st reading line
In configuration mode: Increase value, select option
Change display of the 2nd reading line
In configuration mode: Decrease value, select option
Print data
435-1/-3 only: If the Cyclical Printing function is activated, the
programmed measuring program is started.
Switch instrument on, switch display light on/off;
switch instrument off (press and hold)
3. Product description 7
de
Button Functions
Open (main) menu
en
OK Enter confirmation
ESC Cancel
/ Hold value/display current measurement value
fr
Hold ACT
Reset Reset max./min. values to current measurement value
Open menu item “Multi-point mean calculation“
es
Mean
Measp Open menu item “Measuring program“ (435-2/-4 only) 435>
Start Start test series (435-2/-4 only)
it
End End test series (435-2/-4 only), End Cyclical Print (435-1/-3 only)
Save Save values (435-2/-4 only)
pt
Turb Activate “Turb“ test series (435-2/-4 only with attached turbulence probe)
Area Open menu item “Area”
sv
P=0 Zero internal pressure sensor (435-3/-4 only)
nl
Important displays
Display Meaning
??
Battery capacity (only for operation by battery/rechargeable battery):
· 4 segments in the battery symbol are lit: Instrument battery is fully charged
· No segments in the battery symbol are lit: Instrument battery is almost spent
??
(flashing) Print function: Data are sent to the printer
435-3/-4 only: Measurement channel differential pressure (internal sensor)
, Measurement channel no.: Channel 1, channel 2.
If a measurement channel is a radio channel, the radio symbol lights up as well as the
measurement channel no.
8 3. Product description
3.2 Interfaces
Infrared interface
Measurement data can be sent to a Testo printer via the infrared interface on
the head of the instrument.
USB interface
The mains unit (accessory part) can be connected to the head of the
instrument via the USB interface to power the instrument.
Instruments with a memory: Measurement/instrument data can be exchanged
with a PC via the USB interface.
Probe socket(s)
Plug-in measuring probes can be connected via the probe socket(s) on the
base of the instrument. The instrument is a HighPower device, possibly an
additional USB-Hub is required!
4. Commissioning
de
en
This chapter describes the steps required to commission the product.
fr
Inserting batteries/rechargeable batteries and a radio module (accessory
part):
es
1 Undo the two screws on the rear of the instrument and lift off the battery
compartment cover.
it
2 Insert batteries/rechargeable batteries (3x mignon) into the battery
compartment. Observe the polarity!
pt
3 Push the radio module (accessory part) into the radio module
compartment until it engages in place. Note the guide groove.
sv
4 Replace the battery compartment cover, press down and secure by
tightening the two screws.
nl
??
??
10 5. Operation
5. Operation
This chapter describes the steps that are executed frequently when using the
product.
de
Press and hold (for approx. 2s) until the display goes out.
en
5.3 Display light
fr
Switching the display light on/off:
es
The instrument is switched on.
Press .
it
pt
sv
nl
??
??
12 6. Setting the instrument
press ESC several times until the instrument changes to measurement view.
6.1.1 Profile
The instrument has predefined measurement profiles that are tailored to
specific areas of application.
The profile setting influences the following points in measurement mode:
· Assignment of the function buttons
· Number of predefined functions
· Structure of the main menu
All functions are available in the standard profile. In the application-specific
measurement profiles, the available functions are reduced to only those that are
needed to ensure speedier access.
Setting a profile:
The configuration menu is open, config. is displayed.
1 Profile .OK
6.1.2 Units
de
Predefined systems and individual setting options:
en
Parameter ISO system US system Individual setting options
Temperature °C °F °C, °F
fr
Pressure hPa inchH2O mbar, Pa, hPa, kPa, inchH2O
Velocity m/s fpm m/s, fpm
es
Volumetric flow rate m³/h ft³/min m³/h, l/s, ft³/min
Length mm inch mm, inch
Power kW BTU/h kW, BTU/h, TONS
it
Setting units:
pt
The configuration menu is open, config. is displayed.
sv
1 Units OK .
2 Press / ISO/US (to set the system) or a parameter (to set
nl
individually) and confirm with OK .
3 Set the system of units or the desired unit with / and confirm
??
with OK .
??
6.1.3 Device
Instrument data
Displaying instrument data:
The configuration menu is open, config. is displayed.
1 Device Inst.data OK .
OK
- The firmware version and the serial number of the instrument are
displayed.
Date / Time
Setting the date/time:
The configuration menu is open, config. is displayed.
1 Device date/time OK .
OK
Battery type
To ensure that the battery capacity is displayed correctly, the battery type used
must be set.
Setting the battery type:
The configuration menu is open, config. is displayed.
1 Device OK Bat-type OK .
2 Press / Battery or ReBa and confirm with OK .
Auto OFF
If Auto OFF is switched on, the instrument switches itself off automatically after
10min if no button is pressed. Exception: Cyclical printing (instruments without
a memory) or a measuring program (instruments with a memory) is active.
Switching Auto OFF on/off:
The configuration menu is open, config. is displayed.
1 Device OK Auto OFF OK .
2 Press / to select On or Off and confirm with OK .
Reset
When a reset is carried out, the instrument is reset to the default settings, all
settings / data are deleted. Exception: Language, Date / Time.
Resetting:
The configuration menu is open, config. is displayed.
1 Device OK reset OK .
2 Reset with OK or cancel the reset with ESC .
Setting min. / max. printing function
If pr MinMaxAuto is activated, minimum and maximum values are also printed
with the measurement readings.
Switching off pr MinMax:
The configuration menu is open, Config. is displayed.
1 Device pr MinMax
OK OK .
2 Choose On or Off with / and confirm with OK .
6. Setting the instrument 15
K-factor
de
The querying of the K-factor (correction factor for areas) when entering the
parameter „Area“.can be switched on/off.
en
See chapter PARAMETER / AREA, p. 24.
fr
Switching K-ffactor on/off:
The configuration menu is open, config. is displayed.
es
1 Device OK K-factor OK .
it
2 With / , select On or Off and confirm with OK .
Num holes
pt
The querying of the number of holes (number of measurement points) in the
sv
calculation of a multi-point mean value can be switched on/off. The number of
holes is required to allocate the readings to the number of measurement points
nl
for later evaluation via the PC software.
See Chapter MEAN, p. 22 and MEASURING, p. 26.
??
Switching number of holes on/off:
??
The configuration menu is open, config. is displayed.
1 Device OK Num holes OK .
2 With / , select On or Off and confirm with OK .
6.1.4 Probe
RadioC
Radio probes may only be used in countries in which they have been Type
Approved (see application information of the radio probe).
A radio module (accessory part) is required for the use of radio probes. The
instrument can establish a connection with a maximum of three radio probes.
Each radio probe has a probe ID (RF ID). This consists of the last 3 digits of the
serial no. and the position of the slide switch (H or L) in the radio probe.
Setting up a radio probe:
A radio module (accessory part) is inserted in the instrument.
See chapter COMMISSIONING, p. 9.
The configuration menu is open, config. is displayed.
The radio probe is switched on and the transfer rate is set to 2 readings
per second (see the advice on using the radio probe).
16 6. Setting the instrument
1 Probe OK RadioC OK .
2 Press / to select the desired channel no. for the radio probe (P.1,
P.2 or P.3) and confirm with OK .
- The instrument searches for switched-on radio probes in the receiving
range.
- The probe IDs of the radio probes found are displayed.
If no radio probes were found, this may be because of the following:
· The radio probe is not switched on or the battery of the radio probe is
spent.
· The radio probe is outside the range of the measuring instrument.
· Sources of interference are influencing the radio transmission (e.g.
reinforced concrete, metal objects, walls or other barriers between
transmitter and receiver, other transmitters of the same frequency,
strong electromagnetic fields).
If necessary, rectify the possible causes of the fault in radio
transmission.
Alternatively, the probe ID can also be entered manually.
MAN Press / to enter the probe ID.
3 Press / to select the probe that is to be assigned to the chosen
channel no.
4 Assign the radio probe to the chosen channel no. with OK or exit the
function with ESC , without changing the probe configuration.
Humidity probe calibration (435-2/-4 only)
This function is only available if a humidity probe is plugged in.
The calibration values can be reset to the default settings (Reset). A 2-point
calibration can be performed.
Resetting the calibration values:
The configuration menu is open, config. is displayed.
1 Probe OK Calibr. OK .
2 Press / to select Reset and confirm by pressing OK twice.
- The calibration values are reset to the default settings.
6. Setting the instrument 17
Calibrating:
de
The configuration menu is open, config. is displayed.
1 Probe OK Calibr. OK .
en
2 Press / to select calibration point P1 or P2 and confirm by
pressing OK twice.
fr
3 Put the humidity probe into the reference medium and wait for the
es
equalisation period to elapse.
- The current humidity reading and the calibration point (nominal value) are
it
displayed.
4 Start the calibration menu with OK .
pt
5 Save calibration with OK or cancel calibration with ESC .
sv
P internal (435-3/-4 only)
The internal pressure sensor can be switched on/off.
nl
Switching the internal pressure sensor on/off:
??
The configuration menu is open, config. is displayed.
??
1 Probe OK P intern OK .
2 Press / On or Off and confirm with OK .
Te-Type
The probe characteristic curves stored in the instrument can be set for the
probe type used.
Setting probe type:
The configuration menu is open, Config. is displayed.
1 Probe Te-Type OK .
OK
Pressure
The function is only available when an absolute pressure probe is attached.
You can set whether the absolute air pressure (measured with an absolute
pressure probe) or the barometric air pressure (calculated from the measured
absolute pressure and the input of the altitude above sea level).
For the input of abs alt. for the calculation of the barometric air pressure, see
chapter PARAMETER / AREA, p. 24.
Setting measurement parameter:
The configuration menu is open, config. is displayed.
1 Probe OK Pres. OK .
2 With / , select the desired measurement parameter and confirm
with OK .
6.1.5 Language
Setting the language:
The configuration menu is open, config. is displayed.
1 Language OK .
2 Select the desired language with / and confirm with OK .
de
Profile menu items Function
Standard P = 0 (nur 435-3) Zero internal pressure sensor
en
Area Set form, cross-section area, K-factor
Calc. De/activate volume flow, differential temperature, dewpoint
temperature, psychrometric temperature calculation;
fr
435-3 additionally; De/activate flow calculation
Parameter Set reference pressure, abs. altitude;
es
435-3 additionally; Set reference temperature/humidity.
cyc. Print De/activate cycle printing
Ductm. P = 0 (nur 435-3) Zero internal pressure sensor
it
Velocity (nur 435-3) De/activate flow calculation
Vol. De/activate volume flow calculation
pt
Parameter (nur 435-3) Set reference pressure, abs. altitude, P-factor, Set reference
temperature/humidity.
sv
Pres. (nur 435-1) Set reference pressure
cyc. Print De/activate cycle printing
nl
Menu overview testo 435-2/-4
??
Profile menu items Function
Standard Memory Info, Activate/set measurement locality, print report, delete store
??
Meas. Prog Set/ de/activate measurement program
Mean Time/point mean calculation
Calc. De/activate volume flow, differential temperature, dewpoint tempera-
ture, psychrometric temperature, Enthalpy calculation; Set heat
transfer coefficient alpha
435-3 additionally; De/activate flow calculation
P = 0 (nur 435-4) Zero internal pressure sensor
Parameter Set reference pressure, abs. altitude, Set area form/cross-section;
435-4 additionally: Set reference temperature/humidity.
Ductm. P = 0 (nur 435-4) Zero internal pressure sensor
Memory Info, Activate/set measurement locality, print report, delete store
Velocity (nur 435-4) De/activate flow calculation
Vol. De/activate volume flow calculation
Parameter Set reference pressure, abs. altitude
435-4 additionally: Set reference temperature/humidity.
IAQ Mean Time mean calculation
Pres. Set reference pressure
Memory Info, Activate/set measurement locality, print report, delete store
abs alt. Set abs. altitude
P = 0 (nur 435-4) Zero internal pressure sensor
20 6. Setting the instrument
Location
The active location can be changed. Up to 99 locations can be created. The
numerical location designations (01-99) can be changed into any text (max.
10 characters) using the PC software.
Changing an active location:
The main menu is open, Menu is displayed.
1 Memory OK Location OK .
2 Press / to select the location to be activated and confirm with
OK .
Protocol
Saved measurement protocols can be printed out on a Testo printer (accessory
part) via the infrared interface.
Printing a measurement protocol:
The main menu is open, Menu is displayed.
1 Memory OK Protocol OK .
2 Press / to select the measurement protocol that is to be printed.
3 Press to start printing out the measurement protocol.
Delete
The entire memory with all measurement protocols can be cleared.
6. Setting the instrument 21
de
The main menu is open, Menu is displayed.
1 Memory OK Delete OK .
en
2 Press OK to clear the entire memory.
fr
6.2.2 Measuring program (435-2/-4 only)
es
A measuring program can be programmed and activated/deactivated:
it
Designation Description
Off Measuring program switched off: Readings can be stored manually
pt
AUTO Automatic measuring program: The measuring cycle (min. 1s) and the number of
readings can be set freely.
Turb Automatic measuring program for measuring turbulence (only if a turbulence probe
sv
is available and plugged in): The measuring cycle (1/5s) and duration (180s) are
preset.
nl
Deactivating a measuring program:
??
The main menu is open, Menu is displayed.
1 Meas.Prog . OK
??
2 Press / to select Off and confirm with OK .
- The instrument returns to measurement view.
Programming and activating the AUTO measuring program:
The main menu is open, Menu is displayed.
1 Meas.Prog OK .
2 Press / to select AUTO and confirm with OK .
The measuring cycle is set in the order: hours / minutes / seconds.
3 Press / to set the measuring cycle in hours and confirm with
.
OK
6.2.3 Mean
The menu item Mean value calculation is only available in the instrument
testo 435-2/-4. In the instrument testo 435-1/-3, the function Mean value
calculation is called up with the function button Mean .
For carring out Mean value calculation see chapter Measuring, p. 26.
6.2.5 Calculation
de
If calculation is switched on, additional parameters with calculated values can
en
be displayed from the readings of one probe. These are then displayed as
additional measurement channels in measurement view.
fr
Particular measurement channels must be available in order to perform the
calculation.
es
Additional calculation parameters need to be set for some calculated variables.
See the chapter PARAMETER / AREA, p. 24.
it
The following variables can be calculated:
pt
· Flow velocity (435-3/-4 only)
· Volumetric flow rate
sv
· Dew point (below 0°Ctd / 32°Ftd frost point temperatures are displayed)
· Psychrometric temperature
nl
· Enthalpy (Heating/cooling performance of assemblies)
??
The heat transfer coeffiient (alpha) required for the calculation of the U-value can
be set.
??
It is also possible to calculate the difference between two measurement
channels (Delta). This is only possible if the selected measurement channels
have the same unit.
Activating/deactivating reading calculation:
The main menu is open, Menu is displayed.
1 Calc. .
OK
6.2.6 Parameter/Area
Some calculated variables relate to particular reference values (ambient
conditions or factors for certain probes). These can be entered by means of
calculation parameters.
Parameters used for calculating variables:
Parameters Calculated measuring variables
Temp. (reference temperature) Flow velocity, volumetric flow rate (for measurement with a
(435-3/-4 only) pitot tube)
Humid. (reference humidity) Flow velocity, volumetric flow rate (for measurement with a
(435-3/-4 only) pitot tube)
Pres. (reference pressure) Flow velocity, volumetric flow rate (for measurement with a
pitot tube or hot wire probe), heating/cooling performance (enthalpy)
Area (cross-section area) Volumetric flow rate
P-factor (pitot tube factor) Flow velocity and volumetric flow rate (for measurement with a
(435-3/-4 only) pitot tube)
abs alt. Barometric air pressure
de
Five areas can be stored for the “Area” parameter. Three shapes are defined
in the default setting (one rectangles: edge length a and b, one circle:
en
diameter d, any shaped area: cross-section area A). The dimensions of the
areas can be adapted in the instrument. It is possible to reassign the shapes
fr
using the PC software (435-2/-4 only).
When K-factor activated (see chapter DEVICE, p. 13): An offset factor is stored
es
for each area. If parts of an area are covered (e.g. grilles on ventilation
openings), this can be calculated out using the offset factor. What must be
it
indicated is the free part of the area (20% covered --> 80% free area -->
offset factor 0.8).
pt
For measurements made at outlets and volume flow regulators with defined
differential pressure measurement points, a component-specific correction
sv
factor (k-Vol) prescribed by the manufacturer can be entered instead of the
area input.
nl
For measurements at ventilation outlets with a funnel the parameter Funnel
must be activated. The funnel set (order no. 0563 4170) consists of a funnel
??
for measurements at plate outlets (200 x 200mm) and a funnel for
measurements at ventilation outlets (330 x 330mm) in combination with the
??
testo 435 and with the 100mm vane probe 0635 9435.
The main menu is open, Menu is displayed.
1 testo 435-2/-4 only: Parameter OK .
2 Press / to select Area and confirm with OK .
3 Press / to select the shape by which the area is to be described
and confirm the input with OK .
4 Press / to set the parameter(s) and confirm each one with OK .
- The settings are applied and the last shape to be set is activated.
7. Measuring
This chapter describes the steps that are required to perform measurements
with the product.
Particular probes must be plugged in or switched on and registered (radio
probes) according to the variable that is to be measured.
Some probes require a warming-up phase until they are ready to measure.
For some variables additional calculation parameters have to be set if correct
measurements are to be obtained.
See PARAMETER / AREA, p. 24.
For the calculation of the U-value, please refer to the documentation included
with the U-value temperature probe (0614 1635).
Required for the calculation of the heating/cooling performance of assemblies:
· A vane probe (for determining the volume flow) must be connected.
· 2 wireless humidity probes (for determining enthalpy at the input and output
of the assembly and for calculating tightness).
· The parameter pressure (for the calculation of tightness) must be entered.
· The calculation factor enthalpy must be activated. The enthalpy is integrated
in the calculation of performance, an enthalpy value cannot be displayed..
· The wireless humidity probe allocated to radio channel 1 must be positioned
next to the vane probe, as its readings are automatically used to calculate
the mass flow.
7. Measuring 27
de
· The ambient CO probe must be situated in fresh air (CO-free) during the
zeroing phase.
en
· Cigarette smoke influences the measurement by more than 50ppm. The
breathed air of a smoker influences the reading by approx. 5ppm.
fr
· The flow impact of the gas influences the measurement accuracy. Frontal
impact on the sensor leads to higher measurement values. The best
es
measurement results are acheived with a slight movement of the probe back
and forth.
it
Taking a measurement:
pt
The instrument is in measurement view.
The measuring program AUTO or TURB is not activated (435-2/-4 only).
sv
Put the probe in position and take the readings.
nl
Changing the upper measurement channel line display:
Press .
??
Changing the lower measurement channel line display, showing the
??
max./min. value of the variable in the upper measurement channel line:
Press .
- The following are displayed in consecutive order:
· Available measurement channels
· Maximum value of the variable in the upper display line
· Minimum value of the variable in the upper display line
· Lower measurement line not shown
Resetting max./min. values:
The minimum or maximum values of all measurement channels are reset.
1 Press several times until the maximum or minimum value is displayed.
2 Reset the max./min. values with Reset .
Holding readings:
Press Hold .
Press Act to change back to displaying the actual reading.
Saving readings (435-2
2/-4
4 only):
Press .
Save
de
en
This chapter describes the steps that help to maintain the functionality of the
product and extend its service life.
fr
Cleaning the housing:
es
Clean the housing with a moist cloth (soap suds) if it is dirty. Do not use
aggressive cleaning agents or solvents!
it
Changing the battery/rechargeable battery:
pt
Instrument is switched off.
1 Undo the two screws on the rear of the instrument and lift off the battery
sv
compartment cover.
2 Remove spent batteries/rechargeable batteries and insert new
nl
batteries/rechargeable batteries (3x mignon) into the battery
compartment. Observe the polarity!
??
3 Replace the battery compartment cover and tighten the two screws.
??
30 9. Quenstions and answers
If we are unable to answer your question, please contact your dealer or Testo
Customer Service. For contact data, see back of this document or web page
www.testo.com/service-contact
10. Technical data 31
de
en
Measuring ranges and accuracies
fr
Parameter/Probe type Measuring range Accuracy (± 1 Digit) Resolution
Temperature/NTC -50...+150°C ±0.2°C (-25.0...+74.9°C) 0.1°C
±0.4°C (-50.0...-25.1°C)
es
±0.4°C (+75.0...+99.9°C)
±0.5% of reading (rest of range)
-58...+302°F ±0.4°F (-13.0...+166.9°F) 0.1°F
it
±0.8°F (-58.0...-13.1°F)
±0.8°F (+167.0...+211.9°F)
pt
±0.5% of reading (rest of range)
Temperature/ -200...+1370°C (Type K) ±0.3°C (-60.0...+60.0 °C) 0.1°C
Type K/T -200...+400°C (Type T) ±0.2°C+0.5% of reading (rest of range)
sv
-328...+2498°F (Type K) ±0.6°F (-76.0...+140.0°F) 0.1°F
-328...+752°F (Type T) ±0.4°F+0.5% of reading (rest of range)
nl
Relative humidity/ 0...+100%RH Depends on probe 0.1%RH
humidity probe
Flow velocity/vane Vane 16mm: Depends on probe 0.1m/s
??
0.6...+40m/s
Vane 60mm: Depends on probe 0.01m/s
??
0.25...+20m/s
Vane 100mm: Depends on probe 0.01m/s
0.3...+20m/s
Flow velocity/ 0...+20m/s Depends on probe 0.01m/s
hot wire probe
Pressure/absolute 0...+2000hPa Depends on probe 0.1hPa
pressure probe
CO2/IAQ probe 0...+10000ppm Depends on probe 1ppm
Lux/Lux probe 0...100000Lux Depends on probe 1Lux
(testo 435-2/-4 only)
Pressure/internal 0...+25hPa ±0.02hPa (0...+2hPa) 0.01hPa
differential pressure probe (Overload: 200hPa) ±1% of reading (rest of range)
(testo 435-3/-4 only)
CO/Ambient CO probe 0...500ppm ±5ppm (0...100ppm) 1) 1ppm
±5% of reading (101...500ppm) 1)
1) at 10...30°C, outside this range additionally ±0.2% of reading / °C
32 10. Technical data
de
en
This chapter gives important accessory and spare parts for the product.
Name Part no.
fr
Probes
Water-proof immersion/penetration probe, TC type K 0602 1293
es
Water-proof surface probe with widened measurement tip for smooth surfaces, TC type K 0602 1993
Robust affordable air probe, TC type K 0602 1793
it
Vane probe, 1000mm diameter 0635 9435
Vane probe, 60mm diameter, telescopic to max. 910mm 0635 9335
pt
Vane probe, 16mm diameter, telescopic to max. 890mm 0635 9535
Humidity/temperature probe, 12mm diameter (testo 435-2/-4 only) 0636 9735
sv
Handle for humidity/temperature probe, for connection to the measuring instrument,
including probe cable, for measures/calibrates of humidity probehead (testo 435-2/-4 only) 0430 9735
Absolute pressure probe 2000hPa 0638 1835
nl
Pressure dew point probe for measurements in compressed air systems (testo 435-2/-4 only) 0636 9835
Hot wire probe for m/s and °C, probehead 7.5mm diameter, telescopic to max. 820mm 0635 1025
??
IAQ probe to assess Indoor air quality, CO“, humidity, temperature and
absolute pressure measurement 0632 1535
Thermal flow probe with integrated temperature and humidity measurment,
??
12mm diameter, with telescope max. 745mm 0635 1535
Comfort probe for degree of turbulence measurement, with telescopic handle and stand,
fulfills DIN EN 13779 requirements (testo 435-2/-4 only) 0628 0109
Lux probe, probe for measuring luminous intensity (testo 435-2/-4 only) 0635 0545
Ambient CO probe 0632 1235
Miscellaneous
Plug-in mains unit, 5VDC, 500mA with European plug 0554 0447
Funnel set sonsisting of funnel for plate outlets and funnel for vents. 0563 4170
External recharger incl. 4 Ni-MH rechargeable batteries with built-in, international plug,
100-240V, 300mA, 50/60Hz, 12VA/instrument 0554 0610
For a complete list of all accessories and spare parts, please refer to the
product catalogues and brochures or look up our website: www.testo.com
34 Notes
Notes 35
?? ?? nl sv pt it es fr en de
testo AG
www.testo.com
Postfach 11 40, 79849 Lenzkirch
Testo-Straße 1, 79853 Lenzkirch
Telefon: (07653) 681-0
Fax: (07653) 681-100
E-Mail: [email protected]
Internet: http://www.testo.com