Introduction To Measurement and Instrumentation System
Introduction To Measurement and Instrumentation System
Instrumentation
Sensor
Examples of Sensor
● Thermocouple
Input Output
Temperature e.m.f
Thermocouple
● Resistance thermometer
Output
Input Resistance
Temperature Resistance Change
Thermometer
Example of Signal Conditioner
● Amplifier
Output
Input Small Larger
e.m.f Voltage
Amplifier
● Wheatstone bridge
Input Larger
Resistance Voltage Voltage
Change Wheatstone Change Change
Amplifier
Bridge
Example of Display
● Display
Input Output
Current Movement
Change of Pointer
Scale
● Computer
Input
Analog Digital Output
Signal Video Display Readings
Data Acquisition Signal
Unit
System
(VDU)
Measurement Standards
● Three commonly used systems:
International System (SI: meter, kilogram, second, …)
CGS System (centimeter, gram, second, …)
English System (feet, pound, second, …)
● Basic units – length, time, mass, temperature, ...
● Derived units – area, volume, acceleration, force, ...
Base SI Units
SI is founded on seven SI base units for seven base
quantities assumed to be mutually independent.
Derived SI Units
Defined in terms of the seven base quantities via a system
of quantity equations.
Performance Characteristics
● 3 types of characteristics have been applied to define
the performance of a measurement system:
Operational characteristics
Static characteristics
Dynamic characteristics
● Operational characteristics include:
Range
Span
Sensitivity
Resolution
Dead band / threshold
● Static characteristics include:
Error
Accuracy
Precision
Repeatability
Linearity
Hysteresis
Calibration
● Dynamic characteristics include:
Rise time
Settling time
Maximum overshoot
Steady state errors
Operational Characteristics
● Range the minimum and maximum values.
● Span the difference between the maximum and the
minimum range.
● Sensitivity ratio of a change in output to the change in
input, in steady states condition.
● Resolution the least incremental value of input or
output that can be detected.
● Dead band/threshold the largest range of values of a
measured variable to which the instrument does not
respond.
Examples of Operational
Characteristics
Example 1: A specification of a thermometer reading is
given as follows:
Range and subdivision of temperature in °C → −0.5 to 40.5
°C
–Minimum = −0.5 °C
–Maximum = 40.5 °C
–Span = 41 °C
Example 2: A galvanometer has a sensitivity of 17 mm/µA.
Based on the input shown in the following figure, determine
the corresponding output.
µA mm
1 17
0 0 T
T
Input Signal Output Signal
Example 3: A force sensor measures a range of 0 to 150 N
with a resolution of 0.1% FS (full scale). Find the smallest
change in force that can be measured.
Smallest force change that can be measured = 0.1% × (150)
= 0.15 N
● Example 4: A sensor has a transfer function of 5 mV/°C.
Find the required voltage resolution of the signal
conditioning if a temperature resolution of 0.2 °C is
required.
Voltage resolution = T.F. × temperature resolution
= 5 × 0.2
= 1 mV
Static Characteristics
● Error the difference between the true value Yn and
instrument reading Xn (error = Yn – Xn)
– Types of errors: systematic error (bias error) and random error
(precision error).
– Bias (systemic) error, eb: consistent and repeatable error (eb =
average readings – true value).
– Random error, er: the lack of repeatability in the output of the
measuring system (er = reading – average reading).
● Accuracy reflects the ability of the system to respond
to a true value. Error limit of a measuring device, under
certain operating conditions can appear in several forms:
Measured variable: the accuracy is ±0.2 of the measurement. If the
temperature reading = 30.1 °C, the actual temperature lies between
29.9 °C and 30.3 °C.
Percentage of full scale (FS): accuracy ±1% f.s. If the full scale is 10
A, accuracy = ±0.1 A.
Percentage of instrument span: accuracy ±3% span. If the span for
pressure measurement is 20 – 50 psi, accuracy = ±0.03(50 – 20) =
±0.9 psi.
Percentage of actual reading: for a ±2%. If the true value of the
voltmeter is 2 V accuracy = (2 × ±0.02) = ±0.04 V.
Considering that the powers of ∆x, ∆y, and ∆z, which are
greater than 2, produce very small values, and can be
neglected.
xa yb ∆x ∆y ∆z
V ± ∆V ≈ c 1 ± a 1 ± b 1 ± (− c )
z x y z
Expand the above equation and neglect any multiplications
between the uncertainties (e.g., ∆x⋅∆y ≈ 0):