Drilling Sequences for TSIV Fixture
Taper KIT User ∅4.0mm Fixture ( Length : 10mm )
10mm
Bone quality ∅2.0 drill F3.5 taper drill 4.0 Fixture
D4 ▶
Implant placement
Soft ▶ ▶
∅4.5mm Fixture ( Length :10mm )
10mm
Bone quality ∅2.0 drill ∅3.0 drill F3.5 taper drill F4.0 taper drill 4.5 Fixture
D4 ▶
Implant placement
Soft ▶ ▶ ▶
∅5.0mm Fixture ( Length :10mm )
10mm
Bone quality ∅2.0 drill ∅3.0 drill F3.5 taper drill F4.5 taper drill 5.0 Fixture
D4 ▶
Implant placement
Soft ▶ ▶ ▶
Straight Drill User ∅4.0mm Fixture ( Length : 10mm )
10mm
Bone quality ∅2.0 drill Pilot drill (2.0/3.0) ∅3.0 drill ∅3.6 drill (Half) 4.0 Fixture
D4 ▶
Implant
Soft ▶ ▶ ▶ ▶ placement
∅4.5mm Fixture ( Length :10mm )
10mm
Bone quality ∅2.0 drill Pilot drill ∅3.0 drill Pilot drill ∅3.8 drill ∅4.1 drill (Half) 4.5 Fixture
(2.0/3.0) (3.0/3.8)
D4 ▶
Implant
Soft ▶ ▶ ▶ ▶ ▶ ▶ placement
∅5.0mm Fixture ( Length :10mm )
10mm
TSIV
Implant System
Bone quality ∅2.0 drill Pilot drill ∅3.0 drill Pilot drill ∅3.8 drill ∅4.6 drill (Half) 5.0 Fixture
(2.0/3.0) (3.0/3.8)
D4 ▶
Implant
Soft ▶ ▶ ▶ ▶ ▶ ▶ placement
※ Because TS IV fixture is especially developed for maxilla and soft bones, there is no surgical sequence for normal bones.
CAS KIT User Catalog & Manual
Fixture Bone Quality ∅2.0 Drill CAS ∅2.8 CAS ∅3.1 CAS ∅3.3 CAS ∅3.6 CAS ∅3.8 CAS ∅4.1
Soft ▶ ▶
Normal ▶ ▶
∅4.0
Soft ▶ ▶ ▶
Normal ▶ ▶ ▶
∅4.5
Soft ▶ ▶ ▶
Normal ▶ ▶ ▶
∅5.0
HEAD OFFICE
8th FL, WorldMeridian Venture CenterⅡ, 426-5, Gasan-dong, Geumcheon-gu, Seoul, Korea
Tel : 02-2016-7000 Fax : 02-2016-7001 www.osstem.com Edition 03 / 2012
TSIV Implant TSIV Implant
Characteristics Specifications
Hex 1.2
TSIV Fixture Order Code
Fixture Only
L
- Fixture : Product Code [ex : TS4S4010S]
Pre-Mounted Fixture [Simple Mount]
- Fixture + Mount + Cover Screw : B + Product Code [ex : BTS4S4010S]
Implant that specially performs in the weaker bone
D
�Specially developed for maxilla and soft bones �Recommended insertion torque : below 40Ncm
* High success rate even with poor bone quality. �Submerged-type implant with morse tapered internal hex
Regular Diameter ∅4.0 Connection Regular
�Improved design for initial stability and simplified surgical sequences connection
L D ∅4.0(Pitch 0.8)
* Improved the initial stability with improved application of helical �Excellent self-tapping ability
7 TS4S4007S
cutting, corkscrew thread, and sharp and rounded apex design �Implant that does not need osteotome Hex 2.5
where implant can be placed with minimal drilling. (∅2 or ∅3mm ∅4.45 8.5 TS4S4008S
can be used on D4 bone) 10 TS4S4010S
11.5 TS4S4011S
L 13 TS4S4013S
SA surface
∅1.8
D
OSSTEM’s SA surface morphology and roughness increased by 45% compared to RBM treatment. 7 8.5 10 11.5 13
Regular Diameter ∅4.5 Connection Regular
L D ∅4.5(Pitch 1.0)
7 TS4S4507S
Hex 2.5
∅4.85 8.5 TS4S4508S
10 TS4S4510S
11.5 TS4S4511S
L 13 TS4S4513S
∅2.0
D
7 8.5 10 11.5 13
Connection Regular
Regular Diameter ∅5.0
L D ∅5.0(Pitch 1.2)
RBM SA ‘S’Co.SA 7 TS4S5007S
∅5.3
Hex 2.5 8.5 TS4S5008S
SEM (Scanning Electron Microscope) image
10 TS4S5010S
● SA : Sand blasted with alumina and Acid etched surface 11.5 TS4S5011S
- Optimal morphology : combine crater and micro-pit L 13 TS4S5013S
- Optimal surface roughness : Ra 2.5~3.0㎛
- Early cell response : 20% higher than RBM ∅2.2
D
- Early bone healing : 20% higher than RBM
7 8.5 10 11.5 13
- Early loading possible after 6 weeks after implantation
※ For posterior single tooth case, ∅4.5mm or larger is recommended / ※ Actual dimensions may be different
02 TSIV Implant System Catalog & Manual TSIV Implant 03