This document provides instructions for transferring data from a U120 Ultra Urine Analyzer to an external computer or LIS system using different connection methods. Data can be sent via Bluetooth, LAN, WLAN or RS-232 cable after testing. Communication software on the receiving system must be compatible and recognize/capture transmitted data at the appropriate port or cable. The data format is also described.

Uploaded by

Eddie Zambrana
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
427 views3 pages

U120 Ultra Data Transfer Package Insert With Bluetooth Instructions 112...

This document provides instructions for transferring data from a U120 Ultra Urine Analyzer to an external computer or LIS system using different connection methods. Data can be sent via Bluetooth, LAN, WLAN or RS-232 cable after testing. Communication software on the receiving system must be compatible and recognize/capture transmitted data at the appropriate port or cable. The data format is also described.

Uploaded by

Eddie Zambrana
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

U120 Ultra Urine Analyzer (Urine) been set up and ready, you can find” and Next → choose

u can find” and Next → choose the relevant Bluetooth and Next → enter the password Line 2: …
“1234” → after connected press “Finish”. Line n: Mark Positive n & Test Abbr.n & Result n
Data Transfer Package Insert
5. Begin urine strip testing according to the U120 Ultra User’s Manual.
2. Invalid Test Strip Data Structure:
REF U124-131 LAN Line 1 to Line n is blank (space)
1. Connect the USB LAN adapter (DM 9601) in the USB port on the analyzer.
For data transfer to Personal Computer (PC) or Laboratory Information System (LIS). 2. Turn on the analyzer. Set the Type of Connectivity to LAN. SETUP USING BLUETOOTH, LAN, OR WLAN CONNECTIVITY
For professional in vitro diagnostic use only. 3. Setup the data as:
INTENDED USE IP set according to network environment A PC running on Windows XP, Windows 7, or Windows 8 is required. The directions are as follows.
This communication protocol defines the format and timing for data transmission from the U120 Ultra Urine Mask set according to network environment
DIRECTIONS FOR USE
Analyzer. Data available from the analyzer includes the specimen ID, date, time, serial number of the analyzer, Gate Way set according to network environment
patient information, operator ID, type of strip, color, clarity, and urine analysis results. This insert provides Host Name set according to network environment
Host Port set according to network environment Bluetooth Connectivity
information required to develop a suitable computer interface to transfer analyzer data to an external computer Note: Bluetooth Version 2.0 is the most current
with suitable PC software or a laboratory LIS system. 4. Begin urine strip testing according to the U120 Ultra User Manual.
Windows XP:
In the U120 Ultra, we have different data transfer method, ACON standard interface, HL7 interface and others. 1. Open the Control Panel. Double click on the Bluetooth Devices option highlighted in blue. This will open
Before using this insert, please check the method first. WLAN
the Bluetooth Devices box.
SUMMARY 1. Connect the USB WLAN adapter (AR9271) into the USB port on the analyzer.
2. Turn on the analyzer. Set the Type of Connectivity to LAN.
Data will be sent to the analyzer communication ports or USB port after testing. If the analyzer has a Bluetooth, 3. Setup the data as:
it may also be used for data download using software which is compatible with this port. The analyzer can also
ESSID set according to network environment
transmit data using LAN or WLAN.
Key set according to network environment
Communication software residing on the LIS or locally connected Personal Computer must look for and
Host Name set according to network environment
recognize data present at the communications port and capture it at the time of transmission.
Host Port set according to network environment
MATERIALS 4. Begin urine strip testing according to the U120 Ultra User Manual.
Materials Provided
• Data Transfer Cable (for SETUP
• Package insert DATA AND FORMAT
USING AN RS232 CABLE only)
The following data and formats are used for printing and transmission to the communication interface.
Materials Required But Not Provided
• A record starts with <STX> (0x02).
• Applicable PC or LIS software 2. Choose the “Devices” option and click on “Add” to enter the “Add Bluetooth Device Wizard”
SETUP USING AN RS232 CABLE 2014-11-03 11:17:59
NAME: SMITH
A null modem cable is required for connection to a personal computer. The pin configuration for the cable is as ID: 000004
follows:
Col Yellow
RS232 Cable Pin Assignment Cla Clear
Analyzer Computer LEU - Neg
2 3 NIT - Neg
3 2 URO +- 1 mg/dL
* PRO +- 15 mg/dL
5 5 pH 7.0
DIRECTIONS FOR USE BLO - Neg
SG 1.030
Download to PC KET - Neg
RS-232 BIL - Neg
1. Connect the RS232C port on the analyzer to the COM port of the computer. GLU - Neg
2. Turn on the analyzer. * ASC 1+ 10 mg/dL 3. Check the box next to “My device is set up and ready to be found.” Click Next.
3. Open the communication software in the PC, such as Hyperterminal.
4. Enter the following PC data interface:
# COM Port the analyzer can be used • All data fields are separated by<CR>(0x0D) <LF>(0x0A).
1 1200, 2400, 4800, 9600, 19200, 57600, or 115200 as settings • Segments of a field are separated by blanks.
2 5, 6, 7, or 8 Data bits as settings • A Test Number is sent with each test result and is in the range of 00001-99999. The leading number
indicates the mode of test, i.e. 0 for routine, 1 for STAT, and 2 for QC.
3 Even, Odd, or None parity as settings • The date and time fields are sent in the format defined by the current instrument set-up.
4 1 or 2 Stop Bit as settings • When a field contains no data or is not enabled via the set up, a blank record is sent.
5 None, Xon, or Xoff Data stream as settings • The Mark Positive 1 field is empty for a normal result or contains an ‘ * ’ for an abnormal test result,
otherwise it is left blank. There is no blank between Mark Positive and Test Abbr.
5. Begin Urine Strip testing according to the U120 Ultra User Manual.
• A <ETX> (0x03) is sent at the very end of each completed record data stream.
• Note that the result output strings are modified to avoid the use of non-ASCII.
Bluetooth 4. After searching, the new equipment “JBM-141” will appear. Highlight JBM-141and click “Next >”.Choose
1. Turn on the analyzer. Set the Type of Connectivity to Bluetooth. the “Input the key” or “Let me choose my own passkey” option. Input the key as “1234”. Click “Next”.
Header Data Structure
2. Check the data as:
The following data structures indicate the testing information provided after data transfer.
Baud rate 9600 Data Bit 8 First line: Date Time
Parity None Stop Bit 1 Second line: Patient Name
Data Stream None Third line: Test Number
Fourth line: Color
Fifth line: Clarity
3. Connect the Bluetooth adaptor to PC.
4. Setup the Bluetooth data in the computer.
Results Data Structure
Windows XP: My Computer→ Control Panel → Bluetooth Settings → Equipment → Add → Choose
The following data structures indicate the testing results depending on the number of parameters tested.
“The equipment has been set up and ready, you can find” and Next → choose the relevant Bluetooth and Next →
1. Valid test strip data structure:
enter the password “1234” → after connected press “Finish”.
Line 1: Mark Positive 1 & Test Abbr.1 & Result 1
Windows 7,8: Start→ Control Panel → Hardware and Sound →Add Equipment→ Choose “The equipment has
5. The Bluetooth Device Wizard will then establish a connection. Once this is complete, you will receive a 2. When the device is set up and ready to be found, the new equipment “JBM-141” will appear, click it then 5. Download the software “HyperTerminal” from “http://www.aconlabs.com/downloads/hyperterminal.zip”
message saying the device was successfully connected to your computer. The COM (serial) ports will be click “Next”. Choose the “Enter the device’s pairing code” option. Then extract the “hyperterminal.zip”.
displayed on the screen. Record these numbers. Click “Finish”.

6. Open the folder “hyperterminal”, double click the hyperterminal.exe. The next steps are same as the steps 6-8
6. From the Start Menu select -> All Programs -> Accessories -> Communications –> HyperTerminal, set the 3. Input the key as “1234”. Click “Next. Click “Close” to complete setup. from the Windows XP instructions.
information to the settings on the following screens. Click OK.

4. Right click the Computerselect the Properties optionclick Device Manageropen the project Ports(COM
&LPT), you will find the Standard Serial Over Bluetooth Link(COMx) LAN Connectivity
1. For this method, please connect the Ethernet USB cable to the RJ45 adapter (DM9601) in the USB port on
the analyzer. Follow with the USR-TCP232-Test(Download from http://www.usr.cn/Down/Software/USR-
TCP232-Test.rar)
2. To Setup the Analyzer,from the Setup Menu select Analyzer Configuration Connectivity. Choose the
LAN option then navigate back to the initial screen.
• IP: Set according to network environment
7. To setup the Mission U120 Ultra Analyzer,from the Setup Menu select Analyzer • Mask:Set according to network environment
ConfigurationConnectivity. Choose the “Bluetooth” option. Navigate back to the initial screen. • Gate Way:Set according to network environment
8. After you reset the information for the Bluetooth, press Records Review Review All Data,press Send • Host Name:Set according to network environment
Out. After “Send Out” is pressed, the data will be sent to the computer. During testing, once the results are • Host Port:Set according to network environment
displayed on the screen, the analyzer will also transmit the data to the computer. • If connected, the analyzer will navigate back to the Analyzer Configuration screen. If not, please
reset the parameters.
Windows 7 windows 8:
1. Open the Control Panel. Click on the “Add a devices” option ,
4. Input “ipconfig” to check the IP address: 8. After the information for WLAN has been set, select Records ReviewReview All Data,Press Send Out.
The data will be sent to the computer. During testing, when the results are displayed on the screen, the
3. Open the USR-TCP232-Test. There is no need to input information in boxes (1) and (2). For box (3), Local analyzer will also transmit the data to the computer.
Host Port, input the same Host Port information as seen in the picture below Box(3). Click the “Listening”
button.

5. Open the USR-TCP232-Test.


6. There is no need to input information in Boxes (1) and (2). For box (3), Local Host Port, input the
4. After the information for the LAN is reset, press Records Review Review All Data. Press Send Out. The information as seen in the picture below in Box (3). Click the “Listening” button.
data will be sent to the computer. During testing, when the data results are displayed on the screen, the Index of Symbols
analyzer will also transmit the data to the computer. Attention, see
Lot Number REF Catalog #
instructions for use
For in vitro Authorized
Manufacturer
diagnostic use only Representative

ACON Laboratories, Inc. MDSS GmbH


10125 Mesa Rim Road, Schiffgraben 41
30175 Hannover, Germany
San Diego, CA 92121, USA

WLAN Connectivity
1. For this method, please connect the USB WLAN adapter (AR9271) into the USB port on the analyzer. Follow 1150685401
with the USR-TCP232-Test. 7. To Setup the Analyzer,from the Setup Menu select Analyzer ConfigurationConnectivity. Choose the Effective date: 2012-xx-xx
2. Windows XP Pro (32bits), Windows 7 Pro (32&64bits), and Windows 8 Pro (64 bits) have been successfully WLAN option. Navigate back to the initial screen.
connected. If necessary, the firewall can be disabled or an “Exception” can be added to the firewall rules.
3. From the Start menu, type the word “Run” into the search field to open the Run Application. Input  ESSID: set according to network environment
“cmd” and then press “ Enter” key:  Key: set according to network environment
 Host Name: set according to network environment
 Host port: set according to network environment

You might also like